Thermo NanoDrop One User Manual

NanoDrop Micro-UV/Vis Spectrophotometers
NanoDrop One
User Guide
269-309101 Revision B July 2016
©2015- 2016 Thermo Fisher Scientific Inc. All rights reserved.
DYMO and LabelWriter are either trademarks or registered trademarks of Newell Rubbermaid in the United States and/or other countries. Wi-Fi is either a trademark or a registered trademark of Wi-Fi Alliance in the United States and/or other countries. Bluetooth is either a trademark or a registered trademark of Bluetooth Special Interest Group. Windows is either a trademark or a registered trademark of Microsoft Corporation in the United States and/or other countries. All other trademarks are the property of Thermo Fisher Scientific inc. and its subsidiaries.
For U.S. Technical Support, please contact:
Unity Lab Services Part of Thermo Fisher Scientific 5225 Verona Road
For International Support, please contact:
Thermo Fisher Scientific Telephone: +1 608 273 5017 E-mail: support.madison@thermofisher.com
Madison WI 53711-4495 U.S.A. Telephone: 1 800 532 4752 E-mail: us.techsupport.analyze@thermofisher.com
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents.
For Research Use Only. This instrument or accessory is not a medical device and is not intended to be used for the prevention, diagnosis, treatment or cure of disease.
WARNING Avoid an explosion or fire hazard. This instrument or accessory is not designed for use in an explosive atmosphere.
C

Contents

Chapter 1 About the NanoDrop One Spectrophotometer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Instrument Models and Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Register Your Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Update Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Chapter 2 Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Detection Limits for All Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Measure dsDNA, ssDNA or RNA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Measure dsDNA, ssDNA or RNA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Nucleic Acid Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Setting for Nucleic Acid Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Calculations for Nucleic Acid Measurements. . . . . . . . . . . . . . . . . . . . . . . . . 18
Measure Microarray. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Measure Microarray Samples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Microarray Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Settings for Microarray Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Calculations for Microarray Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Measure using a Custom Factor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measure Nucleic Acid using a Custom Factor . . . . . . . . . . . . . . . . . . . . . . . . 35
Custom Factor Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Settings for Nucleic Acid Measurements using a Custom Factor . . . . . . . . . . 39
Detection Limits for Nucleic Acid Measurements using a Custom
Factor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Measure Oligo DNA or Oligo RNA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Measure Oligo DNA or Oligo RNA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Oligo Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Settings for Oligo DNA and Oligo RNA Measurements . . . . . . . . . . . . . . . . 47
Detection Limits for Oligo DNA and Oligo RNA Measurements . . . . . . . . . 48
Calculations for Oligo DNA and Oligo RNA Measurements . . . . . . . . . . . . 49
Thermo Scientific NanoDrop One User Guide iii
Contents
Measure Protein A280. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Measure Protein Concentration at A280 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Protein A280 Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Settings for Protein A280 Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Detection Limits for Protein A280 Measurements. . . . . . . . . . . . . . . . . . . . . 64
Calculations for Protein A280 Measurements . . . . . . . . . . . . . . . . . . . . . . . . 65
Measure Proteins and Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Measure Labeled Protein Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Proteins & Labels Reported Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Settings for Proteins and Labels Measurements . . . . . . . . . . . . . . . . . . . . . . . 74
Detection Limits for Proteins and Labels Measurements . . . . . . . . . . . . . . . . 76
Calculations for Proteins and Labels Measurements. . . . . . . . . . . . . . . . . . . . 77
Measure Protein A205. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Measure Protein Concentration at A205 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Protein A205 Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Settings for Protein A205 Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Calculations for Protein A205 Measurements . . . . . . . . . . . . . . . . . . . . . . . . 85
Measure Protein BCA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Measure Total Protein Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Protein BCA Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Settings for Protein BCA Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Measure Protein Bradford . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Measure Total Protein Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Protein Bradford Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Settings for Protein Bradford Measurements . . . . . . . . . . . . . . . . . . . . . . . . 109
Measure Protein Lowry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Measure Total Protein Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Protein Lowry Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Settings for Protein Lowry Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 118
Measure Protein Pierce 660. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Measure Total Protein Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Protein Pierce 660 Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Settings for Protein Pierce 660 Measurements . . . . . . . . . . . . . . . . . . . . . . . 127
Measure OD600 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
Measure OD600 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
OD600 Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Settings for OD600 Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Calculations for OD600 Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Measure Custom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Measure using a Custom Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Delete Custom Method. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Custom Method Reported Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
iv NanoDrop One User Guide Thermo Scientific
Contents
Measure UV-Vis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Measure UV-Vis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
UV-Vis Reported Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Settings for UV-Vis Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Measure Kinetics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Measure Kinetics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155
Create Kinetics Method. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Edit Kinetics Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
Kinetics Reported Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Settings for Kinetic Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Chapter 3 Learning Center . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .175
Micro-Volume Sampling—How it Works. . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
Set Up the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
Measure a Micro-Volume Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Measure a Sample Using a Cuvette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
Prepare Samples and Blanks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204
Basic Instrument Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210
NanoDrop One Home Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 211
NanoDrop One Measurement Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215
NanoDrop One Data Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
NanoDrop One General Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
Instrument Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236
Acclaro Sample Intelligence. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240
NanoDrop One Viewer Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248
Viewer Home Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 249
Manage Experiments and Associated Data. . . . . . . . . . . . . . . . . . . . . . . . . . 251
Manage Identifiers on a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260
Manage Custom Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265
Multimedia . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 277
Chapter 4 Maintaining Your Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .279
Maintenance Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 280
Cleaning the Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281
Maintaining the Pedestals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 282
Cleaning the Pedestals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 282
Reconditioning the Pedestals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285
Decontaminating the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287
Maintaining the Cuvette Sampling System . . . . . . . . . . . . . . . . . . . . . . . . . . . 290
Instrument Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 291
Intensity Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 291
Performance Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 293
Pedestal Image Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 298
Thermo Scientific NanoDrop One User Guide v
Contents
Chapter 5 Safety and Operating Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .301
Operating Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302
Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 303
Chapter 6 About this Help System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .311
Chapter 7 Contact Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .313
vi NanoDrop One User Guide Thermo Scientific
1

About the NanoDrop One Spectrophotometer

The Thermo Scientific™ NanoDrop™ One is a compact, stand-alone UV-Visible spectrophotometer developed for micro-volume analysis of purified nucleic acids and a wide variety of proteins. The patented sample retention system enables the measurement of highly concentrated samples without the need for dilutions.
The NanoDrop One system comes with preloaded software and a touchscreen display. The instrument can be connected to an optional USB label printer.
NOTICE Before operating a NanoDrop One instrument, please read the safety and
operating precautions and then follow their recommendations when using the instrument.
Instrument Models and Features
There are two models available for the NanoDrop One spectrophotometer...
Optional Accessories
A number of accessories are available for the NanoDrop One instruments...
Thermo Scientific NanoDrop One User Guide 1
Register Your Instrument
Register your instrument to receive e-mail updates on software and...
Update Software
Quickly and easily download the latest NanoDrop One software...
1
NanoDrop One Spectrophotometer NanoDrop OneC Spectrophotometer
Arm
Pedestal
Cuvette holder
About the NanoDrop One Spectrophotometer

Instrument Models and Features

Instrument Models and Features
There are two models available for the NanoDrop One spectrophotometer—the NanoDrop One and the NanoDrop One models include the patented
micro-volume sample retention system and general
features. The
NanoDrop One
features a cuvette holder for analyzing dilute samples using standard UV-visible cuvettes.
Both instruments come with a built-in, 7-inch Android high-resolution touchscreen preloaded with easy-to-use instrument control software. The NanoDrop One software is loaded with features to integrate with and simplify your daily workflows.
C
. Both
C
model also
1
Locate the instrument away from air vents and exhaust fans to minimize evaporation
2 NanoDrop One User Guide Thermo Scientific
Touchscreen
USB-A port
1
Pedestals
Touchscreen
1
About the NanoDrop One Spectrophotometer
Instrument Models and Features
Touchscreen can slide left or right to accommodate personal preference, and tilt forward or back for optimal viewing
1
Two more USB-A ports are located on instrument back panel
Thermo Scientific NanoDrop One User Guide 3
1
Instrument light path
Cuvette holder
About the NanoDrop One Spectrophotometer
Instrument Models and Features
NanoDrop One Software with Acclaro Sample Intelligence Technology
The Thermo Scientific™ Acclaro™ Sample
Intelligence technology built into the
NanoDrop One instruments provides these exclusive features to help you assess sample integrity:
• contaminant analysis to help qualify a sample before use in downstream applications
• on-demand technical support for measurements that are atypical or very low concentration
• invalid result alerts (a column sensor monitors for the presence of bubbles or reflective particles that can compromise measurement results)
NanoDrop OneC Model Additional Features
C
The NanoDrop One includes a cuvette holder for measuring dilute samples, colorimetric assays, cell cultures and kinetic studies. The cuvette system has these additional features:
• extended lower detection
limits
• 37 °C heater option for temperature-sensitive samples and analyses
• micro-stirring option to ensure sample homogeneity and support kinetic studies
For details, see Measure a
Sample using a Cuvette.
model
4 NanoDrop One User Guide Thermo Scientific

Optional Accessories

A number of accessories are available for the NanoDrop One instruments. To order an accessory, contact your local distributor or visit our website.
DYMO™ LabelWriter™ 450 USB Label Printer
Prints two 5/16-in x 4-in self-adhesive labels for transferring sample data directly into laboratory notebooks or posting on bulletin boards. The software allows printing of data from each sample measurement or from a group of samples logged and measured together.
The printer connects to the instrument (front or back) via a USB cable (included).
PR-1 Pedestal Reconditioning Kit
1
About the NanoDrop One Spectrophotometer
Optional Accessories
Specially formulated conditioning compound that can be applied to the pedestals to restore them to a hydrophobic state (required to achieve adequate surface tension for accurate sample measurements). The kit includes conditioning compound and applicators. For more information, see
Reconditioning the Pedestals.
PV-1 Performance Verification Solution
Liquid photometric standard used to check instrument performance. For more information, see Performance Verification.
Thermo Scientific NanoDrop One User Guide 5
1
About the NanoDrop One Spectrophotometer

Register Your Instrument

Register Your Instrument
Register your instrument to receive e-mail updates on software and accessories for the NanoDrop One instruments. An Internet connection is required for registration.
To register your instrument
1. Do one of the following:
–From the NanoDrop One Viewer software running on a personal
computer (PC) that is connected to the Internet, open the Help menu and choose NanoDrop One Website.
From any PC that is connected to the Internet, use any web browser to
navigate to our website.
2. On the website, locate NanoDrop One Registration and follow the instructions to register the instrument.
6 NanoDrop One User Guide Thermo Scientific

Update Software

1
About the NanoDrop One Spectrophotometer
Quickly and easily download and install the latest NanoDrop One software and release notes from our website. Follow the steps to update or upgrade the software on your local instrument and/or install or update the NanoDrop One Viewer software on a personal computer (PC). An Internet connection is required to download software.
Update Software
To install or update NanoDrop One Viewer software
1. Do one of the following:
• To install the Viewer software on a computer for the first time, open any
web browser and find the NanoDrop website.
• To update or upgrade the Viewer software, from the Viewer Home screen,
open the Help menu and choose NanoDrop One Website to open our website.
2. On the NanoDrop website, locate the software downloads page.
3. Select to download NanoDrop One (PC) Viewer software (English version) and follow the instructions to download and run the installer. (A computer restart is required after the installer completes.)
4. To add a language, including software and Help systems, download and run the language pack installer (English must be installed first). (No computer restart is required after a language installer completes.)
Thermo Scientific NanoDrop One User Guide 7
1
About the NanoDrop One Spectrophotometer
Update Software
To update or upgrade NanoDrop One instrument software
1. Do one of the following:
2. Insert a USB device such as a memory stick into a USB port on the computer.
3. On the NanoDrop website, locate the software downloads page, select to
4. To add a language, including software and Help systems, download the
5. Insert the USB device into any USB port on the NanoDrop One instrument.
–From the NanoDrop One Viewer software, open the Help menu and
choose NanoDrop One Website to open our website.
From any personal computer that is connected to the Internet, navigate to
the NanoDrop website.
update or upgrade NanoDrop One operating software (English version) and follow the instructions to download the installer to the USB device.
language pack installer(s) to the USB device.
6. From the instrument Home screen, tap (Settings) > System > Update Software.
If the USB device contains more than one version of the installer, a message is displayed. Select the version to install (English installer must be run first) and tap Update. (An instrument restart is required after the English installer completes.)
When the installation is complete, a message similar to the following appears next to the Update Software button:
Version: 1.2.0 (currently installed version of instrument operating software) Database version: 1 (version of NanoDrop One database on this instrument)
7. To add a language, including software and Help systems, tap Update Software again, select the language and version to install and tap Update. (No instrument restart is required after a language installer completes.)
Note: To change the language, tap Language, select an installed language and tap OK. (An instrument restart is required after you change the language.)
8 NanoDrop One User Guide Thermo Scientific
2

Applications

Detection Limits for All Applications

Note Detection limits provided in the tables below are approximate and apply to
micro-volume measurements only; they are based on the instrument’s photometric absorbance range (10 mm equivalent) of 0–550 A. For measurements with 10 mm pathlength cuvettes, the photometric absorbance range is 0–1.5 A.
Detection limits for standard applications
Sample Type Lower Detection Limit Upper Detection Limit Typical Reproducibility
dsDNA 2.0 ng/μL (pedestal)
0.20 ng/μL (cuvette)
ssDNA 1.3 ng/μL (pedestal)
0.13 ng/μL (cuvette)
RNA 1.6 ng/μL (pedestal)
0.16 ng/μL (cuvette)
Thermo Scientific NanoDrop One User Guide 9
27,500 ng/μL (pedestal)
75 ng/μL (cuvette)
18,150 ng/μL (pedestal)
49.5 ng/μL (cuvette)
22,000 ng/μL (pedestal)
60 ng/μL (cuvette)
±2.0 ng/μL for sample concentrations between 2.0 and 100 ng/μL samples; ±2% for samples >100 ng/μL
±2.0 ng/μL for sample concentrations between 2.0 and 100 ng/μL samples; ±2% for samples >100 ng/μL
±2.0 ng/μL for sample concentrations between 2.0 and 100 ng/μL samples; ±2% for samples >100 ng/μL
a
2
Applications
Detection Limits for All Applications
Sample Type Lower Detection Limit Upper Detection Limit Typical Reproducibility
DNA Microarray (ssDNA)
Purified BSA by Protein A280
IgG by Protein A280
Purified BSA by
1.3 ng/μL (pedestal)
0.13 ng/μL (cuvette)
0.06 mg/mL (pedestal)
0.006 mg/mL (cuvette)
0.03 mg/mL (pedestal)
0.003 mg/mL (cuvette)
0.06 mg/mL (pedestal)
495 ng/μL (pedestal)
±2.0 ng/μL for sample concentrations between 2.0 and 100 ng/μL samples;
49.5 ng/μL (cuvette)
825 mg/mL (pedestal)
±2% for samples >100 ng/μL
±0.10 mg/mL (for 0.10–10 mg/mL samples); ±2% for samples >10 mg/mL
402 mg/mL (pedestal)
19 mg/mL (pedestal) ±0.10 mg/mL for 0.10–10 mg/mL samples
Proteins & Labels
0.006 mg/mL (cuvette)
Protein BCA 0.2 mg/mL (20:1
8.0 mg/mL (pedestal)
2% over entire range
reagent/sample volume)
0.20 mg/mL (cuvette)
0.01 mg/mL over entire range
0.01 mg/mL (1:1 reagent/sample volume)
Protein Lowry 0.2 mg/mL (pedestal) 4.0 mg/mL (pedestal) 2% over entire range
Protein Bradford 100 μg/mL (50:1
reagent/sample volume)
8000 μg/mL
±25 μg/mL for 100–500 μg/mL samples ±5% for 500–8000 μg/mL samples
a
15 μg/mL (1:1 reagent/sample volume)
Protein Pierce 660 50 μg/mL (15:1
reagent/sample volume)
25 μg/mL (7.5:1 reagent/sample volume)
a
Based on five replicates (SD=ng/μL; CV=%)
Note To minimize instrument error with highly concentrated samples, make dilutions to ensure that measurements are made within these absorbance limits:
• For micro-volume measurements, maximum absorbance at 260 nm (for nucleic acids) or 280 nm (for proteins) should be less than 62.5 A.
• For measurements with 10 mm pathlength cuvettes, maximum absorbance at 260 nm (or 280 nm for proteins) should be less than 1.5 A, which is approximately 75 ng/μL dsDNA.
100 μg/μL
2000 μg/mL
1000 μg/mL
±4 μg/mL for 15–50 μg/mL samples ±5% for 50–125 μg/mL samples
±3 μg/mL for 50–125 μg/mL samples ±2% for samples > 125 μg/mL
±3 μg/mL for 25–125 μg/mL samples ±2% for samples >125 μg/mL
10 NanoDrop One User Guide Thermo Scientific
Detection limits for pre-defined dyes
2
Applications
Detection Limits for All Applications
Sample Type Lower Detection Limit Upper Detection Limit
Cy3, Cy3.5, Alexa Fluor
0.2 pmol/μL (pedestal) 100 pmol/μL (pedestal) ±0.20 pmol/μL for sample
555, Alexa Fluor 660
a
Typical Reproducibility
concentrations between 0.20 and 4.0
b
pmol/μL; ±2% for samples >4.0 pmol/μL
Cy5, Cy5.5, Alexa Fluor 647
0.12 pmol/μL (pedestal) 60 pmol/μL (pedestal) ±0.12 pmol/μL for sample concentrations
between 0.12 and 2.4 pmol/μL; ±2% for samples >2.4 pmol/μL
Alexa Fluor 488, Alexa Fluor 594
0.4 pmol/μL (pedestal) 215 pmol/μL (pedestal) ±0.40 pmol/μL for sample concentrations
between 0.40 and 8.0 pmol/μL; ±2% for samples >8.0 pmol/μL
Alexa Fluor 546 0.3 pmol/μL (pedestal) 145 pmol/μL (pedestal) ±0.30 pmol/μL for sample concentrations
between 0.30 and 6.0 pmol/μL; ±2% for samples >6.0 pmol/μL
a
Values are approximate
b
Based on five replicates (SD=ng/μL; CV=%)
Thermo Scientific NanoDrop One User Guide 11

Measure dsDNA, ssDNA or RNA

Measures the concentration of purified dsDNA, ssDNA or RNA samples that absorb at 260 nm.

Measure dsDNA, ssDNA or RNA

Reported Results
Settings
Detection Limits
Calculations
Measure dsDNA, ssDNA or RNA
Use the dsDNA, ssDNA and RNA applications to quantify purified double-stranded (ds) or single-stranded (ss) DNA or RNA samples. These applications report nucleic acid concentration and two absorbance ratios (A260/A280 and A260/A230). A single-point baseline correction can also be used.
To measure dsDNA, ssDNA or RNA samples
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently damage the quartz fiber optic cables.
Thermo Scientific NanoDrop One User Guide 13
Measure dsDNA, ssDNA or RNA
Typical nucleic acid spectrum
Comparison of nucleic acid spectra with and without two common contaminants
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new laboratory wipe. For more information, see Cleaning the Pedestals.
To measure nucleic acid
1. From the Home screen, select the Nucleic Acids tab and tap dsDNA, ssDNA or RNA, depending on the samples to be measured.
2. Specify a baseline correction if desired.
3. Pipette 1–2 μL blanking solution onto the lower pedestal and lower the arm, or insert the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette
light path with the instrument light path.
4. Tap Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you lower the arm. (This option is not available for cuvette measurements.)
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the blanking cuvette.
6. Pipette 1-2 μL sample solution onto the pedestal and lower the arm, or insert the sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if
Auto-Measure is off, lower arm and tap Measure.
Cuvette: Tap Measure.
When the sample measurement is completed, the spectrum and reported values are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
14 NanoDrop One User Guide Thermo Scientific
Measure dsDNA, ssDNA or RNA
Best practices for nucleic acid measurements
• Isolate and purify nucleic acid samples before measurement to remove impurities. Depending on the sample, impurities could include DNA, RNA, free nucleotides, proteins, some buffer components and dyes. See Preparing Samples for more information.
Note Extraction reagents such as guanidine, phenol, and EDTA contribute absorbance between 230 nm and 280 nm and will affect measurement results if present in samples (even residual amounts).
• Ensure the sample absorbance is within the instrument’s absorbance detection limits.
• Blank with the same buffer solution used to resuspend the analyte of interest. The blanking solution should be a similar pH and ionic strength as the analyte solution.
•Run a blanking cycle to assess the absorbance contribution of your buffer solution. If the buffer exhibits strong absorbance at or near the analysis wavelength (typically 260 nm), you may need to choose a different buffer or application. See Choosing and Measuring a
Blank for more information.
• For micro-volume measurements:
Ensure pedestal surfaces are properly cleaned and conditioned.
If possible, heat highly concentrated or large molecule samples, such as genomic or
lambda DNA, to 63 °C (145 °F) and gently (but thoroughly) vortex before taking a measurement. Avoid introducing bubbles when mixing and pipetting.
Follow best practices for micro-volume measurements.
Use a 1-2 μL sample volume. See Recommended Sample Volumes for more
information.
C
• For cuvette measurements (NanoDrop One
instruments only), use compatible cuvettes
and follow best practices for cuvette measurements.
Related Topics
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
• Best Practices for Micro-Volume Measurements
• Best Practices for Cuvette Measurements
• Prepare Samples and Blanks
• Basic Instrument Operations
Thermo Scientific NanoDrop One User Guide 15
Measure dsDNA, ssDNA or RNA
Tap row to select sample and update spectrum; tap more rows to overlay up to five spectra. Press and hold sample row to view measurement details.
Drag tab down/up to see
more/less sample data
Nucleic acid concentration
UV spectrum
Tap to select unit
Purity ratios
Menu of options; tap to open
Sample name;
tap to edit
Swipe screen left to
view table with more measurement results
Pinch and zoom to adjust axes; double-tap to reset
Tap to end experiment and export data
Menu of options;
tap to open

Nucleic Acid Reported Results

dsDNA measurement screen
For each measured sample, the dsDNA, ssDNA and RNA applications show the UV absorbance spectrum and a summary of the results. Here is an example:
Note Micro-volume absorbance measurements and measurements taken with nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
16 NanoDrop One User Guide Thermo Scientific
Measure dsDNA, ssDNA or RNA
dsDNA, ssDNA and RNA reported values
The initial screen that appears after each measurement (see previous image) shows a summary of the reported values. To view all reported values, press and hold the sample row. Here is an example:
• sample details (application and sampling method used, i.e., pedestal or cuvette)
•sample name
• created on (date sample measurement was taken)
• nucleic acid concentration
• A260/A280
• A260/A230
•A260
•A280
•factor
• baseline correction
Related Topics
• Basic Instrument Operations
• Nucleic Acid Calculations

Setting for Nucleic Acid Measurements

To show the dsDNA, ssDNA or RNA settings, from the dsDNA, ssDNA or RNA measurement screen, tap > Nucleic Acid Setup.
Thermo Scientific NanoDrop One User Guide 17
Measure dsDNA, ssDNA or RNA
Setting Available Options Description
Baseline Correction On or off
Enter baseline correction wavelength in nm or use default value (340 nm)
Optional user-defined baseline correction. Can be used to correct for any offset caused by light scattering particulates by subtracting measured absorbance at specified baseline correction wavelength from absorbance values at all wavelengths in sample spectrum. As a result, absorbance of sample spectrum is zero at specified baseline correction wavelength.
Related Topics
Instrument Settings

Calculations for Nucleic Acid Measurements

The nucleic acid applications use the Beer-Lambert
equation to correlate absorbance with concentration.
Solving Beer’s law for concentration yields the equation at the right.
Beer-Lambert Equation (solved for concentration)
c = A / ( * b)
where:
A = UV absorbance in absorbance units (AU)
= wavelength-dependent molar absorptivity coefficient (or extinction
coefficient) in liter/mol-cm
b = pathlength in cm
c = analyte concentration in moles/liter or molarity (M)
Note: Dividing the measured absorbance of a sample solution by its molar extinction coefficient yields the molar concentration of the sample. See
Published Extinction Coefficients for more information regarding molar
vs. mass concentration values.
18 NanoDrop One User Guide Thermo Scientific
Measure dsDNA, ssDNA or RNA
The Nucleic Acid applications use a modification of the Beer-Lambert equation (shown at right) to calculate sample concentration where the extinction coefficient and pathlength are combined and referred to as a “factor.”
For the dsDNA, ssDNA and RNA applications, the generally accepted factors for nucleic acids are used in conjunction with Beer’s Law to calculate sample concentration. For the Custom Factor application, the user-specified factor is used.
Extinction Coefficients vs Factors
Using the terms in the Beer-Lambert equation, factor (f) is defined as:
factor (f) = 1/( * b)
where:
= wavelength-dependent molar extinction coefficient in ng-cm/μL
b = sample pathlength in cm
As a result, analyte concentration (c) is calculated as:
c = A * [1/( * b)]
or
c = A * f
where:
c = analyte concentration in ng/μL A = absorbance in absorbance units (A)
f = factor in ng-cm/μL (see below)
Factors Used
dsDNA (factor = 50 ng-cm/μL)
ssDNA (factor = 33 ng-cm/μL)
RNA (factor = 40 ng-cm/μL)
Custom Factor (user entered factor between 15 ng-cm/μL and 150 ng-cm/μL
Thermo Scientific NanoDrop One User Guide 19
Measure dsDNA, ssDNA or RNA
Calculated nucleic acid concentrations are based on the absorbance value at 260 nm, the factor used and the sample pathlength. A single-point baseline correction (or analysis correction) may also be applied.
Concentration is reported in mass units. Calculators are available on the Internet to convert concentration from mass to molar units based on sample sequence.
Absorbance values at 260 nm, 280 nm and sometimes 230 nm are used to calculate purity ratios for the measured nucleic acid samples. Purity ratios are sensitive to the presence of contaminants in the sample, such as residual solvents and reagents typically used during sample purification.
Measured Values
Note: For micro-volume absorbance measurements and measurements
taken with nonstandard (other than 10 mm) cuvettes, the spectra are normalized to a 10 mm pathlength equivalent.
A260 absorbance
• Nucleic acid absorbance values are measured at 260 nm using the normalized spectrum. This is the reported A260 value if Baseline Correction is not selected.
•If Baseline Correction is selected, the absorbance value at the correction wavelength is subtracted from the absorbance at 260 nm. The corrected absorbance at 260 nm is reported and used to calculate nucleic acid concentration.
A230 and A280 absorbance
• Normalized and baseline-corrected (if selected) absorbance values at 230 nm and 280 nm are used to calculate A260/A230 and A260/A280 ratios.
Sample Pathlength
• For micro-volume measurements, the software selects the optimal pathlength (between 1.0 mm and 0.03 mm) based on sample absorbance at the analysis wavelength.
• For cuvette measurements, pathlength is determined by the cuvette Pathlength setting in the software (see General Settings).
• Displayed spectra and absorbance values are normalized to a 10 mm pathlength equivalent.
20 NanoDrop One User Guide Thermo Scientific
Measure dsDNA, ssDNA or RNA
Reported Values
Nucleic acid concentration. Reported in selected unit (i.e., ng/μL, μg/uL or μg/mL). Calculations are based on modified Beer’s Law equation using corrected nucleic acid absorbance value.
A260/A280 purity ratio. Ratio of corrected absorbance at 260 nm to corrected absorbance at 280 nm. An A260/A280 purity ratio of ~1.8 is generally accepted as “pure” for DNA (~2.0 for RNA). Acidic solutions may under represent the reported value by 0.2-0.3; the opposite is true for basic solutions.
A260/A230 purity ratio. Ratio of corrected absorbance at 260 nm to corrected absorbance at 230 nm. An A260/A230 purity ratio between
1.8 and 2.2 is generally accepted as “pure” for DNA and RNA.
Note: Although purity ratios are important indicators of sample quality, the best quality indicator quality is functionality in the downstream application of interest (e.g., real-time PCR).
Thermo Scientific NanoDrop One User Guide 21

Measure Microarray

Measures the concentration of purified nucleic acids that have been labeled with up to two fluorescent dyes for use in downstream microarray applications.

Measure Microarray Samples

Reported Results
Settings
Detection Limits
Calculations
Measure Microarray Samples
Use the Microarray application to quantify nucleic acids that have been labeled with up to two fluorescent dyes. The application reports nucleic acid concentration, an A260/A280 ratio and the concentrations and measured absorbance values of the dye(s), allowing detection of dye concentrations as low as 0.2 picomole per microliter.
To measure microarray samples
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently damage the quartz fiber optic cables.
Thermo Scientific NanoDrop One User Guide 23
Measure Microarray
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new laboratory wipe. For more information, see Cleaning the Pedestals.
24 NanoDrop One User Guide Thermo Scientific
To measure a microarray sample
Dye absorbance peak used to calculate dye concentration
A260 absorbance peak used to calculate nucleic acid concentration
Typical microarray spectrum
1. From the Home screen, select the Nucleic Acids tab and tap Microarray.
2. Specify the sample type and factor and the type of
dye(s) used.
Tip: Select a dye from the pre-defined list or add a custom dye using the Dye/Chromophore Editor.
3. Pipette 1–2 μL blanking solution onto the lower pedestal and lower the arm, or insert the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette
light path with the instrument light path.
4. Tap Blank and wait for the measurement to complete.
Measure Microarray
Tip: If Auto-Blank is On, the blank measurement starts automatically after you lower the arm. (This option is not available for cuvette measurements.)
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the blanking cuvette.
6. Pipette 1-2 μL sample solution onto the pedestal and lower the arm, or insert the sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if
Auto-Measure is off, lower arm and tap Measure.
Cuvette: Tap Measure.
When the sample measurement is completed, the spectrum and reported values are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
Thermo Scientific NanoDrop One User Guide 25
Measure Microarray
Related Topics
• Best Practices for Nucleic Acid Measurements
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
• Best Practices for Micro-Volume Measurements
• Best Practices for Cuvette Measurements
• Prepare Samples and Blanks
• Basic Instrument Operations
26 NanoDrop One User Guide Thermo Scientific

Microarray Reported Results

Drag tab down/up to
see more/less sample data
Dye concentration(s)
Tap row to select sample and update spectrum; tap more rows to overlay up to five spectra. Press and hold sample row to view measurement details.
Nucleic acid concentration
Tap to select unit
UV-visible spectrum
Menu of options; tap to open
Sample name;
tap to edit
Swipe screen left to
view table with more measurement results
Pinch and zoom to adjust axes; double-tap to reset
Tap to end experiment and export data
Microarray measurement screen
For each measured sample, this application shows the absorbance spectrum and a summary of the results. Here is an example:
Measure Microarray
Note
• A baseline correction is performed at 750 nm (absorbance value at 750 nm is subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements and measurements taken with nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
Thermo Scientific NanoDrop One User Guide 27
Measure Microarray
Microarray reported values
The initial screen that appears after each measurement (see previous image) shows a summary of the reported values. To view all reported values, press and hold the sample row. Here is an example:
• sample details (application used and pedestal or cuvette)
•sample name
• created on (date sample measurement was taken)
• nucleic acid concentration
•A260
• A260/A280
• dye 1/dye 2 concentration
• sample type
• analysis correction
•factor
Related Topics
• Basic Instrument Operations
• Microarray Calculations

Settings for Microarray Measurements

Microarray settings
The Microarray Setup screen appears after you select the Microarray application from the Nucleic Acids tab on the Home screen. To show the Microarry settings from the Microarray measurement screen, tap > Microarray Setup.
28 NanoDrop One User Guide Thermo Scientific
Setting Available Options Description
Measure Microarray
Sample type and Factor
Dye 1/Dye 2 Type
dsDNA (with non-editable factor of 50 ng-cm/μL)
ssDNA (with non-editable factor of 33 ng-cm/μL)
RNA (with non-editable factor of 40 ng-cm/μL)
Oligo DNA with non-editable calculated factor in ng-cm/μL
Oligo RNA with non-editable calculated factor in ng-cm/μL
Custom (with user-specified factor in ng-cm/μL)
a
Cy3, 5, 3.5, or 5.5, Alexa Fluor 488, 546, 555, 594, 647, or 660
Widely accepted value for double-stranded DNA
Widely accepted value for single-stranded DNA
Widely accepted value for RNA
Factor calculated from user-defined DNA base sequence. When selected, available DNA base units (i.e., G, A, T, C) appear as keys. Define sequence by tapping appropriate keys. Factor is calculated automatically based on widely accepted value for each base unit.
Factor calculated from user-defined RNA base sequence. When selected, available RNA base units (i.e., G, A, U, C) appear as keys. Define sequence by tapping appropriate keys. Factor is calculated automatically based on widely accepted value for each base unit.
Enter factor between 15 ng-cm/μL and 150 ng-cm/μL
Select pre-defined dye(s) used to label sample material, or one that has been added using Dye/Chrom. Editor.
Dye 1/Dye 2 Unit picomoles/microliter (pmol/uL),
micromoles (uM), or millimoles (mM)
Analysis Correction
b
On or off
Enter analysis correction wavelength in nm or use default value (340 nm)
a
To add a custom dye or edit the list of available dyes, use the Dye/Chromophore Editor.
b
The Analysis Correction affects the calculation for nucleic acid concentration only.
Select unit for reporting dye concentrations
Corrects sample absorbance measurement for any offset caused by light scattering particulates by subtracting absorbance value at specified analysis correction wavelength from absorbance value at analysis wavelength. Corrected value is used to calculate sample concentration.
Tip: If the sample has a modification that absorbs light at 340 nm, select a different correction wavelength or turn off Analysis Correction.
Thermo Scientific NanoDrop One User Guide 29
Measure Microarray
Tap to add custom dye
Tap to edit selected custom dye
Tap to delete selected custom dye
Locked dye (pre-defined; cannot be edited or deleted)
Selected dyes (will appear in Dye1 and Dye2 lists in Microarray Setup or Proteins & Labels Setup)
Tap to close Dye/Chrom. Editor
Custom dye (user-defined; can be edited or deleted)
Dye/Chromophore Editor
Dye/chromophore editor
Use the Dye/Chromophore Editor to add a custom dye to the list of available dyes in
Microarray Setup or Proteins & Labels Setup. You can also specify which dyes are available in
that list.
To access the Dye/Chromophore Editor:
• from the Home screen, tap > Dye/Chrom. Editor
• from the Microarray or Proteins & Labels measurement screen, tap > Settings
> Dye/Chrom. Editor
30 NanoDrop One User Guide Thermo Scientific
These operations are available from the Dye/Chromophore Editor:
Measure Microarray
Add or remove a dye
To add or remove a dye from the Dye1 or Dye2 drop-down list in Microarray Setup or
Proteins & Labels Setup:
select or deselect corresponding checkbox
Add custom dye
tap to show New Dye box
enter unique Name for new dye (tap field to display keyboard, tap Done key to close
keyboard)
–select default Unit that will be used to display dye concentration
–enter dyes Extinction Coefficient (or molar absorptivity constant) in L/mole-cm
(typically provided by dye manufacturer)
–specify Wa ve le ng th in nm (between 450 nm and 700 nm) that will be used to
measure dye’s absorbance
specify dye’s correction values at 260 nm and 280 nm
–tap Add Dye
Note To determine dye correction values (if not available from dye manufacturer):
use instrument to measure pure dye and note absorbance at 260 nm, 280 nm
and at analysis wavelength for dye (see above)
calculate ratio of A
260/Adye wavelength
and enter that value for 260 nm
Correction
calculate ratio of A
280/Adye wavelength
and enter that value for 280 nm
Correction
When a custom dye is selected before a measurement, the dye’s absorbance and concentration values are reported and the corrections are applied to the measured sample absorbance values, and to the resulting sample concentrations and purity ratios.
Edit custom dye
Tip Dyes pre-defined in the software cannot be edited.
tap to select custom dye
–tap
Thermo Scientific NanoDrop One User Guide 31
Measure Microarray
edit any entries or settings
–tap Save Dye
Delete custom dye
Tip Dyes pre-defined in the software cannot be deleted.
tap to select custom dye
–tap
NOTICE Deleting a custom dye permanently removes the dye and all associated information from the software.
Related Topics
Instrument Settings

Calculations for Microarray Measurements

As with the other nucleic acid applications, the Microarray application uses a modification of the
Beer-Lambert equation to calculate sample
concentration where the extinction coefficient and pathlength are combined and referred to as a “factor.” The Microarray application offers six options (shown at right) for selecting an appropriate factor for each measured sample, to be used in conjunction with Beer’s Law to calculate sample concentration.
If the factor is known, choose the Custom Factor option and enter the factor in ng-cm/μL. Otherwise, choose the option that best matches the sample solution.
Tip: Ideally, the factor or extinction coefficient should be determined empirically using a solution of the study nucleic acid at a known concentration using the same buffer.
Available Options for Factors
dsDNA (factor = 50 ng-cm/μL)
ssDNA (factor = 33 ng-cm/μL)
RNA (factor = 40 ng-cm/μL)
Oligo DNA (calculated from user entered DNA nucleotide sequence)
Oligo RNA (calculated from user entered RNA nucleotide sequence)
Custom Factor (user entered factor between 15 ng-cm/μL and 150 ng-cm/μL
Note: See Sample Type for more information.
32 NanoDrop One User Guide Thermo Scientific
Measure Microarray
Calculated nucleic acid concentrations are based on the absorbance value at 260 nm, the factor used and the sample pathlength. A single-point baseline correction (or analysis correction) may also be applied.
Concentration is reported in mass units. Calculators are available on the Internet to convert concentration from mass to molar units based on sample sequence.
Absorbance values at 260 nm, 280 nm and sometimes 230 nm are used to calculate purity ratios for the measured nucleic acid samples. Purity ratios are sensitive to the presence of contaminants in the sample, such as residual solvents and reagents typically used during sample purification.
Measured Values
A260 absorbance
Note: The absorbance value at 750 nm is subtracted from all wavelengths
in the spectrum. As a result, the absorbance at 750 nm is zero in the displayed spectra. Also, for micro-volume absorbance measurements and measurements taken with nonstandard (other than 10 mm) cuvettes, the spectra are normalized to a 10 mm pathlength equivalent.
• Nucleic acid absorbance values for all Microarray sample types are measured at 260 nm using the 750-corrected and normalized spectrum.
•If Analysis Correction is selected, the absorbance value at the correction wavelength is subtracted from the absorbance at 260 nm.
• If one or more dyes are selected, the dye correction values at 260 nm are also subtracted from the absorbance at 260 nm.
• The final corrected absorbance at 260 nm is reported and used to calculate sample concentration.
A280 absorbance
Dye concentrations are calculated from the absorbance value at the dye’s analysis wavelength, the dye’s extinction coefficient, and the sample pathlength. A sloped-line dye correction may also be used.
• 750-corrected and normalized absorbance value at 280 nm (minus the A280 dye correction) is used to calculate an A260/A280 ratio.
Dye absorbance
• Dye absorbance values are measured at specific wavelengths. See
Dye/Chromophore Editor for analysis wavelengths used.
• If Sloping Dye Correction is selected, a linear baseline is drawn between 400 nm and 750 nm and, for each dye, the absorbance value of the sloping baseline is subtracted from the absorbance value at each dye’s analysis wavelength. Baseline-corrected dye absorbance values are reported and used to calculate dye concentrations.
Dye correction
• Pre-defined dyes have known correction values for A260 and A280. See Dye/Chromophore Editor for correction values used.
• A260 dye corrections are subtracted from the A260 absorbance value used to calculate nucleic acid concentration, and from the A260 absorbance value used to calculate the A260/A280 purity ratio.
Thermo Scientific NanoDrop One User Guide 33
Measure Microarray
Sample Pathlength
• For micro-volume measurements, the software selects the optimal pathlength (between 1.0 mm and 0.03 mm) based on sample absorbance at the analysis wavelength.
• For cuvette measurements, pathlength is determined by the cuvette Pathlength setting in the software (see General Settings).
• Displayed spectra and absorbance values are normalized to a 10 mm pathlength equivalent.
Reported Values
Nucleic acid concentration. Reported in selected unit (i.e., ng/μL, μg/uL or μg/mL). Calculations are based on modified Beer’s Law equation using corrected nucleic acid absorbance value.
A260/A280 purity ratio. Ratio of corrected absorbance at 260 nm to corrected absorbance at 280 nm. An A260/A280 purity ratio of ~1.8 is generally accepted as “pure” for DNA (~2.0 for RNA). Acidic solutions may under represent the reported value by 0.2-0.3; the opposite is true for basic solutions.
Dye1/Dye2 concentration. Reported in pmol/μL. Calculations are based on Beer’s Law equation using (sloping) baseline-corrected dye absorbance value(s).
Note: Although purity ratios are important indicators of sample quality, the best indicator of DNA or RNA quality is functionality in the downstream application of interest (e.g., microarray).
Related Topics
• Calculations for Nucleic Acid Measurements
34 NanoDrop One User Guide Thermo Scientific

Measure using a Custom Factor

Measures the concentration of purified nucleic acids using a custom factor for the calculations.
Measure using Custom Factor
Reported Results
Settings
Detection Limits
Calculations

Measure Nucleic Acid using a Custom Factor

Use the Custom Factor application to quantify purified DNA or RNA samples that absorb at 260 nm with a user-defined extinction coefficient or factor. The application reports nucleic acid concentration and two absorbance ratios (A260/A280 and A260/A230). A single-point baseline correction can also be used.
To measure nucleic acid samples using a custom factor
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently damage the quartz fiber optic cables.
Thermo Scientific NanoDrop One User Guide 35
Measure using a Custom Factor
Typical nucleic acid spectrum
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new laboratory wipe. For more information, see Cleaning the Pedestals.
To measure using a custom factor
1. From the Home screen, select the Nucleic Acids tab and tap Custom Factor.
2. Enter the factor to be used for the calculations and specify a baseline correction if desired.
3. Pipette 1–2 μL blanking solution onto the lower pedestal and lower the arm, or insert the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette
light path with the instrument light path.
4. Tap Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you lower the arm. (This option is not available for cuvette measurements.)
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the blanking cuvette.
6. Pipette 1-2 μL sample solution onto the pedestal and lower the arm, or insert the sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if
Auto-Measure is off, lower arm and tap Measure.
Cuvette: Tap Measure.
When the sample measurement is completed, the spectrum and reported values are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
36 NanoDrop One User Guide Thermo Scientific
Related Topics
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
• Best Practices for Micro-Volume Measurements
• Best Practices for Cuvette Measurements
• Prepare Samples and Blanks
• Basic Instrument Operations

Custom Factor Reported Results

Measure using a Custom Factor
For each measured sample, this application shows the absorbance spectrum and a summary of the results. Here is an example:
Note The Custom Factor measurement screen is identical to the measurement screen for
the other nucleic acid applications except the Custom Factor is reported in the lower left
corner (see image below).
Thermo Scientific NanoDrop One User Guide 37
Measure using a Custom Factor
Custom factor used to calculate nucleic acid concentration
Related Topics
• Basic Instrument Operations
• Nucleic Acid Reported Results
• Nucleic Acid Calculations
38 NanoDrop One User Guide Thermo Scientific

Settings for Nucleic Acid Measurements using a Custom Factor

To show the Custom Factor settings, tap > Custom Factor Setup.
Setting Available Options Description
Measure using a Custom Factor
Custom Factor Enter an integer value
between 15 ng-cm/μL and 150 ng-cm/μL
Baseline Correction On or off
Enter baseline correction wavelength in nm or use default value (340 nm)
Related Topics
Constant used to calculate nucleic acid concentration in
modified Beer’s Law equation. Based on extinction coefficient
and pathlength:
f = 1/(
* b))
260
where: f= factor
= molar extinction coefficient at 260 nm in ng-cm/μL
b = sample pathlength in cm (1 cm for nucleic acids measured with the NanoDrop One instruments)
Optional user-defined baseline correction. Can be used to correct for any offset caused by light scattering particulates by subtracting measured absorbance at specified baseline correction wavelength from absorbance values at all wavelengths in sample spectrum. As a result, absorbance of sample spectrum is zero at specified baseline correction wavelength.
Instrument Settings

Detection Limits for Nucleic Acid Measurements using a Custom Factor

The lower detection limits and reproducibility specifications for nucleic acids are provided
here. The upper detection limits are dependent on the upper absorbance limit of the
instrument and the user-defined extinction coefficients.
To calculate upper detection limits for nucleic acid samples
To calculate upper detection limits in ng/μL, use the following equation:
(upper absorbance limit
Thermo Scientific NanoDrop One User Guide 39
instrumen
* extinction coefficient
t
sample
)
Measure using a Custom Factor
For example, for a sample measurement using an extinction coefficient of 55, the equation looks like this:
(550 AU * 55 ng-cm/μL) = 30,250 ng/μL
Note For measurements with 10 mm pathlength cuvettes, the upper absorbance limit is
1.5 AU, which is approximately 75 ng/μL for dsDNA.
Related Topics
• Detection Limits for All Applications
40 NanoDrop One User Guide Thermo Scientific

Measure Oligo DNA or Oligo RNA

Measures the concentration of purified ssDNA or RNA oligonucleotides that absorb at 260 nm.
Measure Oligo DNA or RNA
Reported Results
Settings
Detection Limits
Calculations

Measure Oligo DNA or Oligo RNA

Use the Oligo DNA and Oligo RNA applications to quantify oligonucleotides that absorb at 260 nm. Molar extinction coefficients are calculated automatically based on the user-defined base sequence of the sample. These applications report nucleic acid concentration and two absorbance ratios (A260/A280 and A260/A230). A single-point baseline correction can also be used.
Note If the oligonucleotide has been modified, for example with a fluorophore dye, check with the oligo manufacturer to determine if the modification contributes absorbance at 260 nm. If it does, we recommend using the Microarray application to quantify nucleic acid concentration. The Microarray application includes a correction to remove any absorbance contribution due to the dye from the oligo quantification result.
Thermo Scientific NanoDrop One User Guide 41
Measure Oligo DNA or Oligo RNA
To measure Oligo DNA or Oligo RNA samples
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently damage the quartz fiber optic cables.
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new laboratory wipe. For more information, see Cleaning the Pedestals.
42 NanoDrop One User Guide Thermo Scientific
To measure an oligonucleotide sample
Example Oligo DNA spectrum
1. From the Home screen, select the Nucleic Acids tab and tap either Oligo DNA or Oligo RNA, as needed.
2. Specify the Oligo base sequence and a baseline
correction if desired.
3. Pipette 1–2 μL blanking solution onto the lower pedestal and lower the arm, or insert the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette
light path with the instrument light path.
4. Tap Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you lower the arm. (This option is not available for cuvette measurements.)
Measure Oligo DNA or Oligo RNA
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the blanking cuvette.
6. Pipette 1-2 μL sample solution onto the pedestal and lower the arm, or insert the sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if
Auto-Measure is off, lower arm and tap Measure.
Cuvette: Tap Measure.
When the sample measurement is completed, the spectrum and reported values are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
Related Topics
• Best Practices for Nucleic Acid Measurements
Thermo Scientific NanoDrop One User Guide 43
Measure Oligo DNA or Oligo RNA
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
• Best Practices for Micro-Volume Measurements
• Best Practices for Cuvette Measurements
• Prepare Samples and Blanks
• Basic Instrument Operations
44 NanoDrop One User Guide Thermo Scientific

Oligo Reported Results

Nucleic acid concentration
Tap row to select sample and update spectrum; tap more rows to overlay up to five spectra. Press and hold sample row to view measurement details.
Drag tab down/up to see
more/less sample data
UV spectrum
1
Purity ratios
Tap to select unitSample name;
tap to edit
Menu of options;
tap to open
Swipe screen left to
view table with more measurement results
Pinch and zoom to adjust axes; double-tap to reset
Tap to end experiment and export data
Oligo DNA measurement screen
For each measured sample, the Oligo DNA and Oligo RNA applications show the UV absorbance spectrum and a summary of the results. Here is an example:
Measure Oligo DNA or Oligo RNA
1
Measured oligo: TTT TTT TTT TTT TTT TTT TTT TTT
Note Micro-volume absorbance measurements and measurements taken with nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
Thermo Scientific NanoDrop One User Guide 45
Measure Oligo DNA or Oligo RNA
Oligo DNA and Oligo RNA reported values
The initial screen that appears after each measurement (see previous image) shows a summary of the reported values. To view all reported values, press and hold the sample row. Here is an example:
• sample details (application and sampling method used, i.e., pedestal or cuvette)
•sample name
• created on (date sample measurement was taken)
• nucleic acid concentration
• A260/A280
• A260/A230
•A260
•A280
•factor
• oligo sequence
• baseline correction
•stirrer status
Note The five nucleotides that comprise DNA and RNA exhibit widely varying A260/A280 ratios. See Oligo Purity Ratios for more information.
Related Topics
• Basic Instrument Operations
•Oligo Calculations
46 NanoDrop One User Guide Thermo Scientific

Settings for Oligo DNA and Oligo RNA Measurements

Add guanine
Add adenine
Add thymine (DNA) or uracil (RNA)
Add cytosine
Remove most recent base
The Oligo setup screen appears after you select the Oligo DNA or Oligo RNA application from the Nucleic Acids tab on the Home screen. To show the Oligo settings from the Oligo DNA or Oligo RNA measurement screen, tap > Oligo DNA Setup (or
Oligo RNA Setup).
Setting Available Options Description
Measure Oligo DNA or Oligo RNA
Oligo Base Sequence
for DNA: Use the G, A, T and C keys to specify the DNA base sequence
for RNA: Use the G, A, U and C keys to specify the RNA base sequence
Specify your DNA or RNA base sequence by tapping the corresponding keys:
Each time a base is added to the sequence, the software calculates the following:
Factor. Constant used to calculate oligonucleotide concentration in modified
Beer’s Law equation. Based on extinction coefficient and pathlength:
f = 1/(
* b)
260
where: f= factor
= molar extinction coefficient at 260 nm in ng-cm/μL
b = sample pathlength in cm (0.1 cm for nucleic acids measured with the NanoDrop One instrument)
Thermo Scientific NanoDrop One User Guide 47
Measure Oligo DNA or Oligo RNA
Setting Available Options Description
Molecular Weight of oligo calculated from user-defined base sequence.
Number of Bases entered.
Molar Ext. Coefficient (260 nm). Molar extinction coefficient of oligo (in ng-cm/μL) at 260 nm calculated from entered base sequence.
%GC. Percentage of guanine and cytosine residues in total number of bases entered.
Baseline Correction On or off
Enter baseline correction wavelength in nm or use default value (340 nm)
Corrects for any offset caused by light scattering particulates by subtracting measured absorbance at specified baseline correction wavelength from absorbance values at all wavelengths in sample spectrum. As a result, absorbance of sample spectrum is zero at specified baseline correction wavelength.
Tip: If the sample has a modification that absorbs light at 340 nm, select a different correction wavelength or turn off Baseline Correction.
Related Topics
Instrument Settings

Detection Limits for Oligo DNA and Oligo RNA Measurements

The lower detection limits and reproducibility specifications for the oligonucleotide sample types (ssDNA and RNA) are provided here. The upper detection limits are dependent on the
upper absorbance limit of the instrument and the extinction coefficients for the user-defined base sequences.
To calculate upper detection limits for nucleic acid samples
To calculate upper detection limits in ng/μL, use the following equation:
(upper absorbance limit
For example, for a sample measurement using an extinction coefficient of 55, the equation looks like this:
(550 AU * 55 ng-cm/μL) = 30,250 ng/μL
Note For measurements with 10 mm pathlength cuvettes, the upper absorbance limit is
1.5 AU, which is approximately 75 ng/μL for dsDNA.
48 NanoDrop One User Guide Thermo Scientific
instrumen
* extinction coefficient
t
sample
)
Related Topics
260
1
2
3
1
N
+
2
N1
1
N1
=
• Detection Limits for All Applications

Calculations for Oligo DNA and Oligo RNA Measurements

Measure Oligo DNA or Oligo RNA
As with the other nucleic acid applications, the Oligo applications use the Beer-Lambert equation to correlate absorbance with concentration based on the sample’s extinction coefficient and pathlength. Because oligonucleotides are short, single-stranded molecules (or longer molecules of repeating sequences), their spectrum and extinction coefficient ( dependent on base composition and sequence.
(The generally accepted extinction coefficients and factors for single-stranded DNA and RNA provide a reasonable estimate for natural, essentially randomized, sequences but not for short, synthetic oligo sequences.) To ensure the most accurate results, we use the exact value of
The NanoDrop software allows you to specify the base sequence of an oligonucleotide before it is measured. For any entered base sequence, the software uses the equation at the right to calculate the extinction coefficient.
Tip: The extinction coefficient is wavelength specific for each oligonucleotide and can be affected by buffer type, ionic strength and pH.
to calculate oligonucleotide concentration.
260
) are closely
Extinction Coefficients for Oligonucleotides
The software uses the nearest neighbor method and the following formula to calculate molar extinction coefficients for specific oligonucleotide base sequences:
where:
= molar extinction coefficient in L/mole-cm
=
1
nearest neighbor
=
2
individual bases
=
3
modifications, such as fluorescent dyes
Thermo Scientific NanoDrop One User Guide 49
Measure Oligo DNA or Oligo RNA
Calculated nucleic acid concentrations are based on the absorbance value at 260 nm, the factor used and the sample pathlength. A single-point baseline correction (or analysis correction) may also be applied.
Concentration is reported in mass units. Calculators are available on the Internet to convert concentration from mass to molar units based on sample sequence.
Absorbance values at 260 nm, 280 nm and sometimes 230 nm are used to calculate purity ratios for the measured nucleic acid samples. Purity ratios are sensitive to the presence of contaminants in the sample, such as residual solvents and reagents typically used during sample purification.
Measured Values
A260 absorbance
Note: For micro-volume absorbance measurements and measurements
taken with nonstandard (other than 10 mm) cuvettes, the spectra are normalized to a 10 mm pathlength equivalent.
• Nucleic acid absorbance values are measured at 260 nm using the normalized spectrum. This is the reported A260 value if Baseline Correction is not selected.
•If Baseline Correction is selected, the absorbance value at the correction wavelength is subtracted from the sample absorbance at 260 nm. The corrected absorbance at 260 nm is reported and used to calculate nucleic acid concentration.
A230, A280 absorbance
• Normalized absorbance values at 230 nm, 260 nm and 280 nm are used to calculate A260/A230 and A260/A280 ratios.
Sample Pathlength
• For micro-volume measurements, the software selects the optimal pathlength (between 1.0 mm and 0.03 mm) based on sample absorbance at the analysis wavelength.
• For cuvette measurements, pathlength is determined by the cuvette Pathlength setting in the software (see General Settings).
• Displayed spectra and absorbance values are normalized to a 10 mm pathlength equivalent.
50 NanoDrop One User Guide Thermo Scientific
Measure Oligo DNA or Oligo RNA
The five nucleotides that comprise DNA and RNA exhibit widely varying A260/A280 ratios. Estimated A260/A280 ratios for each independently measured nucleotide are provided below:
Guanine: 1.15 Adenine: 4.50 Cytosine: 1.51 Uracil: 4.00 Thymine: 1.47
The A260/A280 ratio for a specific nucleic acid sequence is approximately equal to the weighted average of the A260/A280 ratios for the four nucleotides present.
Note: RNA will typically have a higher 260/280 ratio due to the higher ratio of Uracil compared to that of Thymine.
Related Topics
• Calculations for Nucleic Acid Measurements
Reported Values
Nucleic acid concentration. Reported in selected unit (i.e., ng/μL, μg/uL or μg/mL). Calculations are based on modified Beer’s Law equation using corrected nucleic acid absorbance value.
A260/A280 purity ratio. Ratio of corrected absorbance at 260 nm to corrected absorbance at 280 nm.
A260/A230 purity ratio. Ratio of corrected absorbance at 260 nm to corrected absorbance at 230 nm.
Note: The traditional purity ratios (A260/A280 and A260/A230), which are used as indicators of the presence of various contaminants in nucleic acid samples, do not apply for oligonucleotides because the shapes of their spectra are highly dependent on their base compositions. See side bar for more information.
Thermo Scientific NanoDrop One User Guide 51

Measure Protein A280

Measures the concentration of purified protein populations that absorb at 280 nm.
Measure A280 Proteins
Reported Results
Settings
Detection Limits
Calculations

Measure Protein Concentration at A280

Use the Protein A280 application to quantify purified protein populations that contain amino acids such as tryptophan or tyrosine, or cys-cys disulfide bonds, which exhibit absorbance at 280 nm. This application reports protein concentration measured at 280 nm and one absorbance ratio (A260/A280). A single-point baseline correction can also be used. This application does not require a standard curve.
Note If your samples contain mainly peptide bonds and little or no amino acids, use the
Protein A205 application instead of Protein A280.
To measure Protein A280 samples
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently damage the quartz fiber optic cables.
Thermo Scientific NanoDrop One User Guide 53
Measure Protein A280
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new laboratory wipe. For more information, see Cleaning the Pedestals.
54 NanoDrop One User Guide Thermo Scientific
To measure a Protein A280 sample
High concentration BSA sample
Low concentration BSA sample
1. From the Home screen, select the Proteins tab and tap Protein A280.
2. Specify a sample type and baseline correction if desired.
3. Pipette 1–2 μL blanking solution onto the lower pedestal and lower the arm, or insert the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette
light path with the instrument light path.
4. Tap Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you lower the arm. (This option is not available for cuvette measurements.)
Measure Protein A280
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the blanking cuvette.
6. Pipette 2 μL sample solution onto the pedestal and lower the arm, or insert the sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if
Auto-Measure is off, lower arm and tap Measure.
Cuvette: Tap Measure
When the sample measurement is completed, the spectrum and reported values are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
Thermo Scientific NanoDrop One User Guide 55
Measure Protein A280
Best practices for protein measurements
• Isolate and purify protein samples before measurement to remove impurities. Depending on the sample, impurities could include DNA, RNA and some buffer components. See
Preparing Samples for more information.
Note Extraction reagents that contribute absorbance between 200 nm and 280 nm will affect measurement results if present in samples (even residual amounts).
• Ensure the sample absorbance is within the instrument’s absorbance detection limits.
• Choosing a blank:
For the Protein A280, Protein A205, and Proteins & Labels applications, blank with
the same buffer solution used to resuspend the analyte of interest. The blanking solution should be a similar pH and ionic strength as the analyte solution.
For the Protein BCA, Protein Bradford, and Protein Lowry applications, blank with
deionized water (DI H
For the Protein Pierce 660 application, blank with the reference solution used to
make the standard curve (reference solution should contain none of the standard protein stock). For more information, see Working with standard c u r ves.
2
O).
•Run a blanking cycle to assess the absorbance contribution of your buffer solution. If the buffer exhibits strong absorbance at or near the analysis wavelength (typically 280 nm or 205 nm), you may need to choose a different buffer or application, such as a colorimetric assay (for example, BCA or Pierce 660). See Choosing and Measuring a Blank for more information.
Note Buffers such as Triton X, RIPA, and NDSB contribute significant absorbance and are not compatible with direct A280 or A205 measurements.
• For micro-volume measurements:
Ensure pedestal surfaces are properly cleaned and conditioned. (Proteins tend to stick
to pedestal surfaces.)
Gently (but thoroughly) vortex samples before taking a measurement. Avoid
introducing bubbles when mixing and pipetting.
Follow best practices for micro-volume measurements.
Use a 2 μL sample volume. See Recommended Sample Volumes for more
information.
C
• For cuvette measurements (NanoDrop One
instruments only), use compatible cuvettes
and follow best practices for cuvette measurements.
56 NanoDrop One User Guide Thermo Scientific
Related Topics
• Best practices for protein measurements
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
• Prepare Samples and Blanks
• Basic Instrument Operations

Protein A280 Reported Results

Protein A280 measurement screen
Measure Protein A280
For each measured sample, this application shows the absorbance spectrum and a summary of the results. Here is an example:
Thermo Scientific NanoDrop One User Guide 57
Measure Protein A280
Tap row to select sample and update spectrum; tap more rows to overlay up to five spectra. Press and hold sample row to view measurement details.
Drag tab down/up to see
more/less sample data
Swipe screen left to view table with more measurement results
Pinch and zoom to adjust axes; double-tap to reset
Tap to end experiment and export data
UV spectrum
Tap to select unitMenu of options;
tap to open
Sample name; tap to edit
Protein concentration
Absorbance at 280 nm
Purity ratio
Note Micro-volume absorbance measurements and measurements taken with nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
Protein A280 reported values
The initial screen that appears after each measurement (see previous image) shows a summary of the reported values. To view all reported values, press and hold the sample row. Here is an example:
58 NanoDrop One User Guide Thermo Scientific
Measure Protein A280
Sample name;
tap to edit
Sampling method
Application
Date/time measured
Protein conc.
Absorbance at 280 nm
Sample type
Baseline correction wavelength
Baseline correction absorbance
Purity ratio

Settings for Protein A280 Measurements

Related Topics
• Basic Instrument Operations
• Protein A280 Calculations
To show the Protein A280 settings, from the Protein A280 measurement screen, tap >
Protein A280 Setup.
Protein A280 settings
The Protein A280 application provides a variety of sample type options for purified protein analysis.
Thermo Scientific NanoDrop One User Guide 59
Measure Protein A280
Each sample type applies a unique extinction coefficient to the protein calculations. If the extinction coefficient of the sample is known, choose the option and enter the value. Otherwise, calculate the extinction coefficient or choose the option that best matches the sample solution. If you only need a rough estimate of protein concentration and the sample extinction coefficient is unknown, select the 1 Abs=1 mg/mL sample type option.
Tip Ideally, the extinction coefficient should be determined empirically using a solution of the study protein at a known concentration using the same buffer.
+ MW (molar) or 1% (mass)
Setting Available Options
Sample type
a
1 Abs = 1 mg/mL General reference Recommended when extinction coefficient is unknown
BSA 6.67 Calculates BSA (Bovine Serum Albumin) protein
IgG 13.7 Suitable for most mammalian antibodies (i.e.,
Mass Ext. Coefficient (L/gm-cm)
Description
and rough estimate of protein concentration is acceptable for a solution with no other interfering substances. Assumes 0.1% (1 mg/mL) protein solution produces 1.0A at 280 nm (where pathlength is 10 mm), i.e.,
1% = 10.
concentration using mass extinction coefficient (
6.67 L/gm-cm at 280 nm for 1% (i.e., 10 mg/mL) BSA solution. Assuming MW is 66,400 daltons (Da), molar extinction coefficient at 280 nm for BSA is approximately 43,824 M
immunoglobulin G or IgG). Calculates protein concentration using mass extinction coefficient (
13.7 L/gm-cm at 280 nm for 1% (i.e., 10 mg/mL) IgG solution. Assuming MW is 150,000 Da, molar extinction coefficient at 280 nm for IgG is approximately 210,000 M
-1cm-1
-1cm-1
.
.
) of
) of
Lysozyme 26.4 Calculates lysozyme protein concentration using mass
extinction coefficient ( for 1% (i.e., 10 mg/mL) lysozyme solution. Assumes molar extinction coefficient for egg white lysozyme ranges between 36,000 M
60 NanoDrop One User Guide Thermo Scientific
) of 26.4 L/gm-cm at 280 nm
-1cm-1
and 39,000 M-1cm-1.
Measure Protein A280
Setting Available Options
Other protein (
+ MW)
Other protein (
1%)
Baseline Correction
On or off
Enter baseline correction wavelength in nm or use default value (340 nm)
Mass Ext. Coefficient (L/gm-cm)
User entered molar extinction coefficient and molecular weight
Description
Assumes protein has known molar extinction coefficient () and molecular weight (MW), where:
(
)*10=(
molar
Enter MW in kiloDaltons (kDa) and molar extinction coefficient (
) in M
percent
)*(MW
-1cm-1
protein
divided by 1000 (i.e.,
)
/1000). For example, for protein with molar extinction
coefficient of 210,000 M
User entered mass extinction coefficient
N/A Corrects for any offset caused by light scattering
Assumes protein has known mass extinction coefficient (). Enter mass extinction coefficient in L/gm-cm for 10 mg/mL (
particulates by subtracting measured absorbance at specified baseline correction wavelength from absorbance values at all wavelengths in sample spectrum. As a result, absorbance of sample spectrum is zero at specified baseline correction wavelength.
Tip: If the sample has a modification that absorbs light at 340 nm, select a different correction wavelength or turn off Baseline Correction.
1%) protein solution.
-1cm-1
, enter 210.
a
To add or edit a custom protein, use Protein Editor.
Protein editor
Use the Protein Editor to add a custom protein to the list of available protein sample types in
Protein A280 Setup and Proteins & Labels Setup.
To access the Protein Editor:
• from the Home screen, tap > Protein Editor
• from the Protein A280 or Proteins & Labels measurement screen, tap >
Settings > Protein Editor
Thermo Scientific NanoDrop One User Guide 61
Measure Protein A280
Tap to add custom protein
Tap to edit selected custom protein
Tap to delete selected custom protein
Custom proteins (will appear in Sample Type list in Protein A280 Setup and Proteins & Labels Setup)
Tap to close Protein Editor
Protein Editor
These operations are available from the Protein Editor:
Add custom protein
in Protein Editor, tap to show New Protein Type box
enter unique Name for new protein (tap field to display keyboard, tap Done key to
close keyboard)
–enter Description for new protein
specify whether to enter Molar Extinction coefficient or Mass Extinction coefficient
for custom protein
–if Mass Extinction coefficient is selected
enter mass extinction coefficient in L/gm-cm for 10 mg/mL (
solution
1%) protein
62 NanoDrop One User Guide Thermo Scientific
Measure Protein A280
Tap a field to show keyboard; to close, tap Done key
–if Molar Extinction is selected
enter molar extinction coefficient (
/1000). For example, for protein with molar extinction coefficient of
210,000 M
enter molecular weight (MW) in kiloDaltons (kDa)
–tap OK to close New Protein Type box
After you choose OK, the new custom protein appears in the Type list in Protein A280 Setup and Proteins & Labels Setup.
Edit custom protein
in Protein Editor, tap to select custom protein
tap to show Edit Protein Type box
edit any entries or settings
–tap OK
-1cm-1
, enter 210
) in M
-1cm-1
divided by 1000 (i.e.,
Thermo Scientific NanoDrop One User Guide 63
Measure Protein A280
Delete custom protein
in Protein Editor, tap to select custom protein
–tap
Note Deleting a custom protein permanently removes the protein and all associated information from the software.
Related Topics
Instrument Settings

Detection Limits for Protein A280 Measurements

Detection limits and reproducibility specifications for purified BSA proteins are provided
here. The BSA lower detection limit and reproducibility values apply to any protein sample
type. The upper detection limits are dependent on the upper absorbance limit of the instrument and the sample’s extinction coefficient.
To calculate upper detection limits for other (non-BSA) protein sample types
To calculate upper detection limits in ng/μL for proteins, use the following equation:
(upper absorbance limit
For example, if the sample’s mass extinction coefficient at 280 nm is 6.67 for a 1% (10 mg/mL) solution, the equation looks like this:
(550 / 6.67) * 10 = 824.6 (or ~825)
instrumen
Related Topics
• Detection Limits for All Applications
/mass extinction coefficient
t
sample
) * 10
64 NanoDrop One User Guide Thermo Scientific

Calculations for Protein A280 Measurements

Measure Protein A280
The Protein A280 application uses the Beer-Lambert
equation to correlate absorbance with concentration.
Solving Beer’s law for concentration yields the equation at the right.
The extinction coefficient of a peptide or protein is related to its tryptophan (W), tyrosin (Y) and cysteine (C) amino acid composition.
Tip: The extinction coefficient is wavelength specific for each protein and can be affected by buffer type, ionic strength and pH.
Beer-Lambert Equation (solved for concentration)
c = A / ( * b)
where:
A = UV absorbance in absorbance units (AU)
= wavelength-dependent molar absorptivity coefficient (or extinction
coefficient) in liter/mol-cm
b = pathlength in cm
c = analyte concentration in moles/liter or molarity (M)
Note: Dividing the measured absorbance of a sample solution by its molar extinction coefficient yields the molar concentration of the sample. See
Published Extinction Coefficients for more information regarding molar
vs. mass concentration values.
Extinction Coefficients for Proteins
At 280 nm, the extinction coefficient is approximated by the weighted sum of the 280 nm molar extinction coefficients of the three constituent amino acids, as described in this equation:
= (nW * 5500) + (nY * 1490) + (nC * 125)
where:
= molar extinction coefficient
n = number of each amino acid residue 5500, 1490 and 125 = amino acid molar absorptivities at 280 nm
Thermo Scientific NanoDrop One User Guide 65
Measure Protein A280
This application offers six options (shown at right) for selecting an appropriate extinction coefficient for each measured sample, to be used in conjunction with Beer’s Law to calculate sample concentration.
If the extinction coefficient of the sample is known, choose the enter the value. Otherwise, calculate the extinction coefficient or choose the option that best matches the sample solution.
Tip: Ideally, the extinction coefficient should be determined empirically using a solution of the study protein at a known concentration using the same buffer.
Most sources report extinction coefficients for proteins measured at or near 280 nm in phosphate or other physiologic buffer. These values provide sufficient accuracy for routine assessments of protein concentration.
+ MW (molar) or 1% (mass) option and
Available Options for Extinction Coefficient
1 Abs = 1 mg/mL, where sample type and/or ext. coefficient is unknown (produces rough estimate of protein concentration)
BSA (Bovine Serum Albumin, 6.67 L/gm-cm)
IgG (any mammalian antibody, 13.7 L/gm-cm)
Lysozyme (egg white lysozyme, 26.4 L/gm-cm)
Other protein (
Other protein (
Note: See Sample Type for details.
Published Extinction Coefficients
Published extinction coefficients for proteins may be reported as:
• wavelength-dependent molar absorptivity (or extinction) coefficient
) with units of M
(
+ MW), user-specified molar ext. coefficient1%), user-specified mass ext. coefficient
-1cm-1
The equation at the right shows the relationship between molar extinction coefficient ( percent extinction coefficient (
1%).
molar
) and
• percent solution extinction coefficient (1%) with units of (g/100 mL) 1 cm cuvette)
• protein absorbance values for 0.1% (i.e., 1 mg/mL) solutions
Tip: Assess published values carefully to ensure unit of measure is applied correctly.
Conversions Between
(
molar
Example: To determine percent solution extinction coefficient (1%) for a protein that has a molar extinction coefficient of 43,824 M molecular weight (MW) of 66,400 daltons (Da), rearrange and solve the above equation as follows:
1% = (
-1cm-1
(i.e., 1% or 1 g/100 mL solution measured in a
and 1%
molar
) * 10 = (1%) * (MW
* 10) / (MW
molar
protein
protein
)
)
-1cm-1
and a
1% = (43,824 * 10) / 66,400 Da)1% = 6.6 g/100 mL
66 NanoDrop One User Guide Thermo Scientific
Measure Protein A280
To determine concentration (c) of a sample in mg/mL, use the equation at the right and a conversion factor of 10.
Tip: The NanoDrop One software includes the conversion factor when reporting protein concentrations.
Calculated protein concentrations are based on the absorbance value at 280 nm, the selected (or entered) extinction coefficient and the sample pathlength. A single-point baseline correction (or analysis correction) may be applied.
Concentration is reported in mass units. Calculators are available on the Internet to convert concentration from mass to molar units based on sample sequence.
Absorbance values at 260 nm and 280 nm are used to calculate purity ratios for the measured protein samples.
Purity ratios are sensitive to the presence of contaminants in the sample, such as residual solvents and reagents typically used during sample purification.
Conversions Between g/100 mL and mg/mL
C
Example: If measured absorbance for a protein sample at 280 nm relative to the reference is 5.8 A, protein concentration can be calculated as:
C
C
C
Measured Values
A280 absorbance
Note: For micro-volume absorbance measurements and measurements
taken with nonstandard (other than 10 mm) cuvettes, the spectra are normalized to a 10 mm pathlength equivalent.
• Protein absorbance values are measured at 280 nm using the normalized spectrum. If Baseline Correction is not selected, this is the reported A280 value and the value used to calculate protein concentration.
•If Baseline Correction is selected, the normalized and baseline-corrected absorbance value at 280 nm is reported and used to calculate protein concentration.
A260 absorbance
in mg/mL = (A / 1%) * 10
protein
= (A / 1%) * 10
protein
= (5.8/6.6 g/100 mL) * 10
protein
= 8.79 mg/mL
protein
• Normalized and baseline-corrected (if selected) absorbance value at 260 nm is also reported.
Sample Pathlength
• For micro-volume measurements, the software selects the optimal pathlength (between 1.0 mm and 0.03 mm) based on sample absorbance at the analysis wavelength.
• For cuvette measurements, pathlength is determined by the cuvette Pathlength setting in the software (see General Settings).
• Displayed spectra and absorbance values are normalized to a 10 mm pathlength equivalent.
Thermo Scientific NanoDrop One User Guide 67
Measure Protein A280
Reported Values
Protein concentration. Reported in selected unit (mg/mL or μg/mL). Calculations are based on Beer-Lambert equation using corrected protein absorbance value.
A260/A280 purity ratio. Ratio of corrected absorbance at 260 nm to corrected absorbance at 280 nm. An A260/A280 purity ratio of ~0.57 is generally accepted as “pure” for proteins.
Note: Although purity ratios are important indicators of sample quality, the best indicator of protein quality is functionality in the downstream application of interest (e.g., real-time PCR).
68 NanoDrop One User Guide Thermo Scientific

Measure Proteins and Labels

Measures the concentration of purified proteins that have been labeled with up to two fluorescent dyes.
Measure Labeled Proteins
Reported Results
Settings
Detection Limits
Calculations

Measure Labeled Protein Samples

Use the Proteins and Labels application to quantify proteins and fluorescent dyes for protein array conjugates, as well as metalloproteins such as hemoglobin, using wavelength ratios. This application reports protein concentration measured at 280 nm, an A269/A280 absorbance ratio, and the concentrations and measured absorbance values of the dyes, allowing detection of dye concentrations as low as 0.2 picomole per microliter. This information is useful for evaluating protein/dye conjugation (degree of labeling) for use in downstream applications.
To measure labeled protein samples
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently damage the quartz fiber optic cables.
Thermo Scientific NanoDrop One User Guide 69
Measure Proteins and Labels
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new laboratory wipe. For more information, see Cleaning the Pedestals.
70 NanoDrop One User Guide Thermo Scientific
To measure a labeled protein sample
Dye absorbance peak used to calculate dye concentration
A280 absorbance peak used to calculate protein concentration
Peptide backbone
Typical sample spectrum measured with Proteins & Labels application
1. From the Home screen, select the Proteins tab and then tap Protein & Labels.
2. Specify the sample type and the type of dye(s) used.
Tip: Select a dye from the pre-defined list or add a custom dye using the Dye/Chromophore Editor.
3. Pipette 1–2 μL of the blanking solution onto the lower pedestal and lower the arm, or insert the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette
light path with the instrument light path.
4. Tap Blank and wait for the measurement to complete.
Measure Proteins and Labels
Tip: If Auto-Blank is On, the blank measurement starts automatically after you lower the arm. (This option is not available for cuvette measurements.)
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the blanking cuvette.
6. Pipette 2 μL sample solution onto the pedestal and lower the arm, or insert the sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if
Cuvette: Tap Measure
When the sample measurement is completed, the spectrum and reported values are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
Auto-Measure is off, lower arm and tap Measure.
Thermo Scientific NanoDrop One User Guide 71
Measure Proteins and Labels
Related Topics
• Best practices for protein measurements
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
• Prepare Samples and Blanks
• Basic Instrument Operations

Proteins & Labels Reported Results

Proteins & Labels measurement screen
For each measured sample, this application shows the absorbance spectrum and a summary of the results. Here is an example:
72 NanoDrop One User Guide Thermo Scientific
Measure Proteins and Labels
Tap row to select sample and update spectrum; tap more rows to overlay up to five spectra. Press and hold sample row to view measurement details.
Drag tab down/up to see
more/less sample data
UV-visible spectrum
Tap to select unitMenu of options;
tap to open
Sample name; tap to edit
Protein concentration
Dye concentration(s)
Swipe screen left to view table with more measurement results
Pinch and zoom to adjust axes; double-tap to reset
Tap to end experiment and export data
Note
• A baseline correction is performed at 750 nm (absorbance value at 750 nm is subtracted from absorbance values at all wavelengths in sample spectrum).
• Micro-volume absorbance measurements and measurements taken with nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
Thermo Scientific NanoDrop One User Guide 73
Measure Proteins and Labels
Proteins & Labels reported values
The initial screen that appears after each measurement (see previous image) shows a summary of the reported values. To view all reported values, press and hold the sample row. Here is an example:
Reported values for Proteins & Labels application
• Sample details (application and sampling method used, i.e., pedestal or cuvette)
•Sample Name
•Creation date
•Protein
•A280
•Sample Type
•Dye 1/Dye 2
• Sloping Dye Correction
•Analysis Correction
Related Topics
• Basic Instrument Operations
• Proteins & Labels calculations

Settings for Proteins and Labels Measurements

To show the Proteins & Labels settings, from the Proteins & Labels measurement screen, tap
> Proteins & Labels Setup.
74 NanoDrop One User Guide Thermo Scientific
Measure Proteins and Labels
Setting Available Options
Sample type
a
1 Abs = 1 mg/mL
BSA
IgG
Lysozyme
Other protein
+ MW)
(
Other protein (
1%)
Analysis Correction
On or off
b
Enter analysis correction wavelength in nm or use default value (340 nm)
Mass Ext. Coefficient (L/gm-cm)
General reference
6.67
Description
Tap he re for detailed description of each available setting.
Each sample type applies a unique extinction coefficient to the protein calculations. If the extinction coefficient of the
13.7
26.4
user-entered molar extinction coefficient/molecular weight
User entered mass extinction coefficient
sample is known, choose the
+ MW (molar) or 1%
(mass) option and enter the value. Otherwise, calculate the extinction coefficient or choose the option that best matches the sample solution. If you only need a rough estimate of protein concentration and the sample extinction coefficient is unknown, select the 1 Abs=1 mg/mL sample type option.
Tip: Ideally, the extinction coefficient should be determined empirically using a solution of the study protein at a known concentration using the same buffer.
N/A Corrects sample absorbance measurement for any offset
caused by light scattering particulates by subtracting absorbance value at specified analysis correction wavelength from absorbance value at analysis wavelength. Corrected value is used to calculate sample concentration.
Dye 1/Dye 2
c
Ty pe
Cy3, 5, 3.5, or 5.5, Alexa Fluor 488, 546, 555, 594, 647, or 660
See
Dye/Chromophore Editor for specific
values for each dye
Dye 1/Dye 2 Unit
picomoles/microliter (pmol/uL), micromoles
not applicable Select unit for reporting dye concentrations.
(uM), or millimoles (mM)
Sloping Dye Correction
a
To add or edit a custom protein, use Protein Editor.
b
Analysis Correction affects calculation for protein concentration only.
c
To add custom dye or edit list of available dyes, use Dye/Chromophore Editor.
d
Sloping Dye Correction affects calculations for dye concentration only.
On or off Corrects dye absorbance measurements for any offset
d
Tip: If the sample has a modification that absorbs light at 340 nm, select a different correction wavelength or turn off Analysis Correction.
Select pre-defined dye used to label sample material, or one that has been added using Dye/Chrom. Editor.
caused by light scattering particulates by subtracting absorbance value of a sloping baseline from 400 nm to 750 nm from absorbance value at dye’s analysis wavelength.
Thermo Scientific NanoDrop One User Guide 75
Measure Proteins and Labels
Related Topics
Instrument Settings
• Protein Editor
• Dye/Chromophore Editor

Detection Limits for Proteins and Labels Measurements

Detection limits and reproducibility specifications for purified BSA proteins and dyes that are pre-defined in the software are provided here. The BSA lower detection limit and reproducibility values apply to any protein sample type. The upper detection limits are dependent on the upper absorbance limit of the instrument and the sample’s extinction coefficient.
To calculate upper detection limits for other (non-BSA) protein sample types
To calculate upper detection limits in ng/μL for proteins, use the following equation:
(upper absorbance limit
For example, if the sample’s mass extinction coefficient at 280 nm is 6.67 for a 1% (10 mg/mL) solution, the equation looks like this:
(550 / 6.67) * 10 = 824.6 (or ~825)
instrumen
Related Topics
• Detection Limits for All Applications
/mass extinction coefficient
t
sample
) * 10
76 NanoDrop One User Guide Thermo Scientific

Calculations for Proteins and Labels Measurements

Measure Proteins and Labels
As with the other protein applications, Proteins & Labels uses the Beer-Lambert equation to correlate absorbance with concentration based on the sample’s extinction coefficient and pathlength.
This application offers six options (shown at right) for selecting an appropriate extinction coefficient for each measured sample, to be used in conjunction with Beer’s Law to calculate sample concentration.
If the extinction coefficient of the sample is known, choose the enter the value. Otherwise, calculate the extinction coefficient or choose the option that best matches the sample solution.
Tip: Ideally, the extinction coefficient should be determined empirically using a solution of the study protein at a known concentration using the same buffer.
Calculated protein concentrations are based on the absorbance value at 280 nm, the selected (or entered) extinction coefficient and the sample pathlength. A single-point baseline correction (or analysis correction) may be applied.
Concentration is reported in mass units. Calculators are available on the Internet to convert concentration from mass to molar units based on sample sequence.
+ MW (molar) or 1% (mass) option and
Available Options for Extinction Coefficient
1 Abs = 1 mg/mL, where sample type and/or ext. coefficient is unknown (produces rough estimate of protein concentration)
BSA (Bovine Serum Albumin, 6.67 L/gm-cm)
IgG (any mammalian antibody, 13.7 L/gm-cm)
Lysozyme (egg white lysozyme, 26.4 L/gm-cm)
Other protein (
Other protein (
+ MW), user-specified molar ext. coefficient1%), user-specified mass ext. coefficient
Note: See Sample Type for details.
Measured Values
A280 absorbance
Note: The absorbance value at 750 nm is subtracted from all wavelengths
in the spectrum. As a result, the absorbance at 750 nm is zero in the displayed spectra. Also, for micro-volume absorbance measurements and measurements taken with nonstandard (other than 10 mm) cuvettes, the spectra are normalized to a 10 mm pathlength equivalent.
• Protein absorbance values are measured at 280 nm using the 750 nm-corrected and normalized spectrum. If Analysis Correction and Dye Correction are not selected, this is the reported A280 value and the value used to calculate protein concentration.
•If Analysis Correction is selected, the 750-corrected, normalized and analysis-corrected absorbance value at 280 nm is reported and used to calculate protein concentration.
• If a Dye is used, the 750-corrected, normalized, analysis-corrected and
dye-corrected absorbance value at 280 nm is reported and used to
calculate protein concentration.
Thermo Scientific NanoDrop One User Guide 77
Measure Proteins and Labels
Dye concentrations are calculated from the absorbance value at the dye’s analysis wavelength, the dye’s extinction coefficient, and the sample pathlength. A sloped-line dye correction may also be used.
Dye absorbance
• Dye absorbance values are measured at specific wavelengths. See
Dye/Chromophore Editor for analysis wavelengths used.
• If Sloping Dye Correction is selected, a linear baseline is drawn between 400 nm and 750 nm and, for each dye, the absorbance value of the sloping baseline is subtracted from the absorbance value at each dye’s analysis wavelength. Baseline-corrected dye absorbance values are reported and used to calculate dye concentrations.
Dye correction
• Pre-defined dyes have known correction values for A260 and A280. See Dye/Chromophore Editor for correction values used.
• A280 dye correction is subtracted from A280 absorbance value used to calculate protein concentration.
Sample Pathlength
• For micro-volume measurements, the software selects the optimal pathlength (between 1.0 mm and 0.03 mm) based on sample absorbance at the analysis wavelength.
Related Topics
• Beer-Lambert Equation
• Protein A280 Calculations
• For cuvette measurements, pathlength is determined by the cuvette Pathlength setting in the software (see General Settings).
• Displayed spectra and absorbance values are normalized to a 10 mm pathlength equivalent.
Reported Values
Protein concentration. Reported in selected unit (mg/mL or μg/mL). Calculations are based on Beer-Lambert equation using corrected protein absorbance value.
Dye1/Dye2 concentration. Reported in pmol/μL. Calculations are based on Beer’s Law equation using (sloping) baseline-corrected dye absorbance value(s).
78 NanoDrop One User Guide Thermo Scientific

Measure Protein A205

Measures the concentration of purified protein populations that absorb at 205 nm.
Measure A205 Proteins
Reported Results
Settings
Detection Limits
Calculations

Measure Protein Concentration at A205

Use the Protein A205 application to quantify purified peptides and other proteins that contain peptide bonds, which exhibit absorbance at 205 nm. This application reports protein concentration and two absorbance values (A205 and A280). A single-point baseline correction can also be used. This application does not require a standard curve.
Note If your samples contain mainly amino acids such as tryptophan or tyrosine, or cys-cys disulfide bonds, use the Protein A280 application instead of Protein A205.
To measure Protein A205 samples
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently damage the quartz fiber optic cables.
Thermo Scientific NanoDrop One User Guide 79
Measure Protein A205
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new laboratory wipe. For more information, see Cleaning the Pedestals.
To measure a Protein A205 sample
1. From the Home screen, select the Proteins tab and then tap Protein A205.
2. Specify a sample type and baseline correction if desired.
3. Pipette 1–2 μL of the blanking solution onto the lower pedestal and lower the arm, or insert the blanking cuvette into the cuvette holder.
Tip: If using a cuvette, make sure to align the cuvette light path with the instrument light path.
4. Tap Blank and wait for the measurement to complete.
Tip: If Auto-Blank is On, the blank measurement starts automatically after you lower the arm. (This option is not available for cuvette measurements.)
5. Lift the arm and clean both pedestals with a new laboratory wipe, or remove the blanking cuvette.
6. Pipette 2 μL sample solution onto the pedestal and lower the arm, or insert the sample cuvette into the cuvette holder.
7. Start the sample measurement:
–Pedestal: If Auto-Measure is On, lower arm; if Auto-Measure is off, lower arm and
tap Measure.
Cuvette: Tap Measure
When the sample measurement is completed, the spectrum and reported values are displayed (see the next section).
8. When you are finished measuring samples, tap End Experiment.
9. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
Related Topics
• Best Practices for Protein Measurements
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
80 NanoDrop One User Guide Thermo Scientific
• Prepare Samples and Blanks
• Basic Instrument Operations
Measure Protein A205
Thermo Scientific NanoDrop One User Guide 81
Measure Protein A205
Tap row to select sample and update spectrum; tap more rows to overlay up to five spectra. Press and hold sample row to view measurement details.
Drag tab down/up to see
more/less sample data
Swipe screen left to view table with more measurement results
Pinch and zoom to adjust axes; double-tap to reset
Tap to end experiment and export data
UV spectrum
Tap to select unitMenu of options;
tap to open
Sample name; tap to edit
Protein concentration
Absorbance at 205 nm
Absorbance at 280 nm

Protein A205 Reported Results

Protein A205 measurement screen
For each measured sample, this application shows the absorbance spectrum and a summary of the results. Here is an example:
Note Micro-volume absorbance measurements and measurements taken with nonstandard cuvettes are normalized to a 10.0 mm pathlength equivalent.
82 NanoDrop One User Guide Thermo Scientific
Measure Protein A205
Sample name;
tap to edit
Sampling method
Application
Date/time measured
Protein conc.
Absorbance at 205 nm
Sample type
Baseline Correction wavelength
Baseline Correction absorbance
Absorbance at 280 nm
Protein A205 reported values
The initial screen that appears after each measurement (see previous image) shows a summary of the reported values. To view all reported values, press and hold the sample row. Here is an example:

Settings for Protein A205 Measurements

Thermo Scientific NanoDrop One User Guide 83
Related Topics
• Basic Instrument Operations
• Protein A205 Calculations
To show the Protein A205 settings, from the Protein A205 measurement screen, tap >
Protein A205 Setup.
Measure Protein A205
Setting Available Options
Sample type 31 31 Assumes 0.1% (1 mg/mL) at 205 nm = 31
Scopes 27 + 120 * (A280/A205) Assumes
Other protein
1%)
(
Baseline Correction
On or off
Enter baseline correction wavelength in nm or use default value (340 nm)
Mass Ext. Coefficient (L/gm-cm)
Description
0.1% (1 mg/mL) at 205 nm = 27 + 120 *
(A280/A205)
User entered mass extinction coefficient
N/A Corrects for any offset caused by light scattering
Assumes protein has known mass extinction coefficient
(). Enter mass extinction coefficient in L/gm-cm for
1mg/mL (
particulates by subtracting measured absorbance at specified baseline correction wavelength from absorbance values at all wavelengths in sample spectrum. As a result, absorbance of sample spectrum is zero at specified baseline correction wavelength.
Tip: If the sample has a modification that absorbs light at 340 nm, select a different correction wavelength or turn off Baseline Correction.
0.1%) protein solution.
Related Topics
Instrument Settings
84 NanoDrop One User Guide Thermo Scientific

Calculations for Protein A205 Measurements

Measure Protein A205
As with the other protein applications, Proteins A205 uses the Beer-Lambert equation to correlate absorbance with concentration based on the sample’s extinction coefficient and pathlength.
This application offers three options (shown at right) for selecting an appropriate extinction coefficient for each measured sample, to be used in conjunction with Beer’s Law to calculate sample concentration.
If the extinction coefficient of the sample is known, choose the Otherwise, calculate the extinction coefficient or choose the option that best matches the sample solution.
Tip: Ideally, the extinction coefficient should be determined empirically using a solution of the study protein at a known concentration using the same buffer.
Calculated protein concentrations are based on the absorbance value at 205 nm, the selected (or entered) extinction coefficient and the sample pathlength. A single-point baseline correction may also be applied.
Concentration is reported in mass units. Calculators are available on the Internet to convert concentration from mass to molar units based on the sample sequence.
1% (mass) option and enter the value.
Available Options for Extinction Coefficient
31, assumes
Scopes, assumes
0.1% (1 mg/mL) at 205 nm = 31
0.1% (1 mg/mL) at 205 nm = 27 + 120 *
(A280/A205)
Other protein, enter mass extinction coefficient in L/gm-cm for
1mg/mL (0.1%) protein solution
Note: See Sample Type for details.
Measured Values
A205 absorbance
Note: For micro-volume absorbance measurements and measurements
taken with nonstandard (other than 10 mm) cuvettes, the spectra are normalized to a 10 mm pathlength equivalent.
• Protein absorbance values are measured at 205 nm using the normalized spectrum. If Baseline Correction is not selected, this is the reported A205 value and the value used to calculate protein concentration.
•If Baseline Correction is selected, the normalized and baseline-corrected absorbance value at 205 nm is reported and used to calculate protein concentration.
A280 absorbance
• Normalized and baseline-corrected (if selected) absorbance value at 280 nm is also reported.
Thermo Scientific NanoDrop One User Guide 85
Measure Protein A205
Sample Pathlength
• For micro-volume measurements, the software selects the optimal pathlength (between 1.0 mm and 0.03 mm) based on sample absorbance at the analysis wavelength.
• For cuvette measurements, pathlength is determined by the cuvette Pathlength setting in the software (see General Settings).
• Displayed spectra and absorbance values are normalized to a 10 mm pathlength equivalent.
Reported Values
Protein concentration. Reported in selected unit (mg/mL or μg/mL). Calculations are based on Beer-Lambert equation using corrected protein absorbance value.
Related Topics
• Beer-Lambert Equation
• Protein A280 Calculations
86 NanoDrop One User Guide Thermo Scientific

Measure Protein BCA

Measures total protein concentration of unpurified protein samples using a bicinchoninic acid colorimetric detection reagent.
Measure Total Protein
Reported Results
Settings
Detection Limits

Measure Total Protein Concentration

The Protein BCA assay uses bicinchoninic acid as a colorimetric detection reagent to determine total protein concentration in unpurified protein samples. This application is useful for measuring dilute protein solutions or proteins in the presence of components that exhibit significant absorbance between 200 nm and 280 nm, which rules out direct protein measurements at 280 nm or 205 nm. This application measures absorbance at 562 nm and uses a standard curve to calculate protein concentration. A single-point baseline correction is applied.
Theory of Protein BCA assay
The Protein BCA assay uses bicinchoninic acid (BCA) as the detection reagent for Cu+1, which is formed when Cu purple reaction product is formed by the chelation of two molecules of BCA with one cuprous ion (Cu 562 nm and baseline-corrected using the absorbance value at 750 nm. Pre-formulated kits of BCA reagent and CuSO
Thermo Scientific NanoDrop One User Guide 87
+1
). The resulting Cu-BCA chelate formed in the presence of protein is measured at
+2
is reduced by certain proteins in an alkaline environment. A
are available from us or a local distributor.
4
Measure Protein BCA
Protein assay kits and protocols
Please refer to the NanoDrop website for up-to-date kits and protocols for the NanoDrop One instruments. Follow the assay kit manufacturer’s recommendations for all standards and samples (unknowns). Ensure each is subjected to the same timing and temperature throughout the assay.
Protein standards for generating a standard curve may also be provided by the kit manufacturer. Since the NanoDrop One pedestals can measure higher protein concentrations than traditional cuvette-based spectrophotometers, you may need to supply your own protein standards at higher concentrations than provided by the manufacturer. For example, additional standards may be required to ensure the standard curve covers the dynamic range of the assay and the expected range of the unknown samples.
Working with standard curves
A standard curve is required for colorimetric protein analysis.
• Each experiment requires a new standard curve.
• Prepare standards and unknown samples the same way. See the kit manufacturer’s guidelines and recommendations.
All reference and standards solutions should be the same buffer used to resuspend
the samples plus the same volume of reagent added to the samples.
First standard is a reference measurement. The reference solution should contain
none of the analyte of interest. (The reference measurement is not the same as a blank measurement. This application requires both.)
Concentration range of the standards must cover the dynamic range of the assay
and the expected range of the unknown samples. Sample analyte concentrations are not extrapolated beyond the concentration of the highest standard.
• Use the application setup screen to enter concentration values for the standards and to specify how standards and samples will be measured (number of replicates, etc.).
Depending on the Curve Type setting, a standard curve can be generated using two
or more standards.
–The software requires one reference measurement and allows up to 7 standards.
Concentration values for standards can be entered in any order but the standards
must be measured in the order in which they were entered; however, best practice dictates that standards be measured from the lowest concentration of the standard analyte stock to the highest.
88 NanoDrop One User Guide Thermo Scientific
Measure Protein BCA
Reference concentration and absorbance value
Standard concentrations and absorbance values
Swipe left one screen
to view standard curve
Menu; tap to open
Press and hold any row to view details
• For all colorimetric assays except Protein Pierce 660, blank the instrument with DI H
O (deionized water). For Protein Pierce 660, blank with the reference solution (see
2
below).
Measure the reference and all standards before you start analyzing samples. (After the first sample has been measured, no additional changes are allowed to the standard curve.)
As you measure the standards, a measurement screen appears, similar to the measurement screens for samples.
Thermo Scientific NanoDrop One User Guide 89
Measure Protein BCA
Swipe left one screen to view data table for standards
Curve type setting
R2 value (1.0 equals perfect fit)
White circles indicate data points for standards
Standard curve
Press and hold any row to view details
Swipe left one screen to see the standard curve as you build it. Here is an example:
2
The R perfect fit; all points lie exactly on the curve).
value indicates how well the standard curve fits the standard data points (1.0 is a
90 NanoDrop One User Guide Thermo Scientific
Measure Protein BCA
Press and hold any row to view details
Tap to delete
this measurement
Swipe left one screen to see the data table for the standards. Here is an example:
Press and hold a row in any of the previous screens to view details about an individual
standard. Here is an example:
Thermo Scientific NanoDrop One User Guide 91
Measure Protein BCA
After the minimum number of standards has been measured for the selected curve type, a message similar to the following appears:
Load more standards: returns to the setup screen where you can add or edit the concentration value for any standard and then measure the standard.
Run samples: continues to sample measurement screen, after which standards can no longer be edited.
• You can add, edit or delete a standard any time before the first sample measurement.
Add standard
from standards measurement screen, tap > [application name] Setup
tap the next empty Concentration field and enter the concentration value for the new
standard
–tap Done
Edit standard
from standards measurement screen, tap > [application name] Setup
tap the Concentration field and edit the concentration value
–tap Done
Delete standard
from standards measurement screen, standard curve screen, or standards data table,
press and hold the row to show Standard Details box
–tap
The standard no longer appears in the table on the measurement screen and its concentration value no longer appears on the setup screen.
Note You can use this method to delete the reference measurement; however, a new reference must be measured immediately afterwards.
92 NanoDrop One User Guide Thermo Scientific
• After the minimum number of standards has been measured for the selected curve type, the message “Invalid Curve” changes to “Valid Curve.” (This occurs even when additional standards have been defined but not yet measured.) If the “Invalid Curve” message remains after all entered standards have been measured, try:
selecting a different curve type
remeasuring standards using the correct standard material
Valid Curve indicator: This is only an indicator that the required minimum number of points has been established for the selected curve type. It does not validate the integrity of the curve. For example, additional standards may be required to cover the expected assay concentration range.
To measure Protein BCA standards and samples
Measure Protein BCA
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently damage the quartz fiber optic cables.
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new laboratory wipe. For more information, see Cleaning the Pedestals.
To measure Protein BCA standards and samples
1. From the Home screen, select the Proteins tab and tap Protein BCA.
2. Specify a curve type and number of replicates for each standard and enter the
concentration of each standard.
Tip: For this assay, we recommend setting Curve Type to “Linear”.
3. Measure blank:
pipette 2 μL DI H
O onto lower pedestal and lower arm, or insert DI H2O
2
blanking cuvette into cuvette holder
Thermo Scientific NanoDrop One User Guide 93
Measure Protein BCA
Tip: If using a cuvette, make sure to align cuvette light path with instrument light path.
–tap Blank and wait for measurement to complete
lift arm and clean both pedestals with new laboratory wipe, or remove cuvette
4. Measure reference standard:
pipette 2 μL reference solution onto pedestal, or insert reference cuvette (reference
solution should contain none of the standard protein stock, see Working With
Standard Curves for details)
lower arm to start measurement (or tap Measure if Auto-Measure is off)
lift arm and clean both pedestals with new wipe, or remove cuvette
if Replicates setting is greater than 1, repeat measurement
5. Measure remaining standards:
pipette 2 μL standard 1 onto pedestal, or insert standard 1 cuvette
lower arm to start measurement (or tap Measure if Auto-Measure is off)
lift arm and clean both pedestals with new wipe, or remove cuvette
if Replicates setting is greater than 1, repeat measurement
repeat substeps above for each additional standard (when specified number of
standards and replicates have been measured, a message asks whether to load more standards or begin measuring samples)
if finished measuring standards, tap Done (swipe left to view standard curve)
6. Measure samples:
pipette 2 μL sample 1 onto pedestal, or insert sample 1 cuvette
lower arm to start measurement (or tap Measure if Auto-Measure is off)
lift arm and clean both pedestals with new wipe, or remove cuvette
if Replicates setting is greater than 1, repeat measurement
7. When you are finished measuring samples, tap End Experiment.
8. Lift the arm and clean both pedestals with a new wipe, or remove the sample cuvette.
Related Topics
• Best practices for protein measurements
• Measure a Micro-Volume Sample
• Measure a Sample Using a Cuvette
94 NanoDrop One User Guide Thermo Scientific
Loading...