Thermo Microtiter shakers digital User guide

Digital Microplate Shaker
88882005 & 88882006
In the United States:
For customer service, call 1-800-766-7000 To fax an order, use 1-800-926-1166 To order online: thermofisher.com
In Canada:
Find out more at thermofisher.com
For Research Use Only. Not for use in diagnostic procedures. © 2019 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
Operating Manual
Revision A . 09 03 2019
Contents
Section 1 Important Information
Disposing of Product
Section 2 Inspection and Installation
Packing List Connections Structure Diagram
Section 3 Overview
Specifications Environmental Conditions Safety Instructions Capacity and Speed
Section 4 Operation
Control Panel Installation Settings Accessory Installation
1 2
3 3 4 4
5 5 5 6 7
8 8 9 9 10
Section 5 Safety Tips and Maintenance
Section 6 Troubleshooting
Section 7 Optional Accessories
Section 8 Warranty Information
11
12
13
15
Section 1 Important Information
Ignoring the following warnings could cause serious injuries or even fatal accidents.
Check the voltage, phase and capacity of power supply on the ID plate before installation. Connect properly.
Power supply must be properly grounded. Abnormal grounded connection causes serious damage. Grounded connection must not be on the water pipe and gas pipe.
Use provided power cord. Power cord: Wall outlet with grounded terminal power cord 250V 10A.
Do not install the product in a place that gas could leak. Do not use in a place that has industrial oil smoke or metallic dust. It causes fire or electric shock. Do not use the machine near to places where explosion could happen due to organic evaporating gases.
Explosive materials: acid, esther, nitro compound.
Inflammable materials: salt peroxides, inorganic peroxide, salt acids.
Check equipment for permissible environmental conditions when using inside of Temperature and Humidity Chamber or Incubator. It can be the cause of fire or trouble by stirrer electricity, electronic, and damage of motor.
Shaker's permissible environmental condition. Temperature 5°C to 40°C, Maximum relative humidity 80%.
Unplug if there is a strange sound, smell and/or smoke from the product. Stop
operating and request the service.
Keep out of the direct sunlight. It may influence product life and proper operation.
Do not use the machine at places where moisture is high and flooding can happen.
Do not assemble, repair, modify on your own. The product may not work well and electric shock is possible with changes in the efficiency of the product. Also this will void the warranty.
Indicates a hazardous situation which, if not avoided, may result in minor or moderate injury.
Do not put heavy things on the power cord. Do not put the machine on the cord. It may take off the wire coating and cause electric shock or fire.
Do not touch it with wet hands and place the main plug correctly. It could cause the electric shock or injuries.
Installing power outlet near instrument may be convenient
Do not install the stirrer near machinery generating high frequency noise. Avoid installation close to high frequency­welding machine, sewing machine, or mass SCR controller.
Do not inject any liquid and inflammable things inside of product.
Do not pour water or put liquid on the top of the product when cleaning. Disconnect the main power immediately and request the service if water may be in the product.
Do not let the product take any strong shock or vibration. It could cause abnormal operation or trouble. It may deteriorate the ability of the product operation and not obtain correct results.
Do not sprinkle insecticide or flammable spray on the product. Use smooth cloths. Cleaning with solvent can cause fire and deformity.
Power off while product cleaning. It may cause electric shock or fire.
Do not drop or allow the machine to fall. It will cause wrong operation and malfunction.
Disposing of Product
Dispose the unit with separating plastic mold, and motor.
1 | Section 1 Important Information
Digital Microplate Shaker 2
Section 2 Inspection and Installation
Before unpacking the unit, first check for damages in the package of the unit.
Then unpack the unit. Check carefully to see if there were damages incurred during transit.
2.1 Packing List
Table -1. Packing List
Description
Digital
Microplate Shaker
Tray+96-
Microplate Clamp×6
Power Adaptor
well
(Installed)
General
US Plug
CN Plug
Catalog Number
88882005
1
1
1
N/A
After unpacking, check that all unit parts and accessories are as listed below. Contact us or the agent from which you purchased the unit if any components were omitted.
Figure
88882006
1
1
N/A
1
2.2 Connections
2.3 Structure Diagram
96-well Microplate Clamp (6)
Foot (4)
Power Cord
Power Adaptor
Main Body
Tray
Housing
Screw M3×6 (12)
Control Panel
EU Plug
UK Plug
Screw Driver
×
M3 6 Screw
(Installed)
3 | InspectionSection 2 and Installation
N/A
N/A
12
1
1
11
12
Digital Microplate Shaker 4
Section 3 Overview
3.1 Specifications
Rotation Speed
Weight
Power Supply
Others
Speed Range......................................................150 to 1200rpm
Orbit Diameter.................................................................2.5mm
Speed Accuracy......................±1% of set speed up to 299rpm
±2% 300 to 1200rpm
Maximum Load……………………………………...........1kg @1200rpm/
Load
4kg @400rpm(accessory included) Maximum Capacity(Centered on tray)……96-well Microplate 4 or 6 depends on different platforms
Timing Range…………..……............................0, 1min. 99h59min.
Time
Size
Overall Dimensions........................................383×260×145mm
Tray Dimensions...............................................275×235×14mm
Packaging Dimensions.................................487×387×264mm
Net Weight.............................................................12.8Kg(28.2lb)
Gross Weight............................................................15Kg(33.1lb)
Requirement................................AC100-240V, 50/60Hz, 72VA
Certification..............................RoHS, WEEE, cCSAus, CE Mark
3.2 Environmental Conditions
Application Environmental Conditions: indoor use
Temperature....................................................................5 to 40
Altitude...........................................................................2,000 m
Humidity...................................................................... 20% to 85%
Storage Environmental Conditions
Temperature....................................................................0 to 60
Altitude...........................................................................2,000 m
Humidity.......................................20% to 90%, non-condensing
3.3 Safety Instructions
Please read the entire instruction manual before operating the Digital Microplate Shaker.
WARNING DO NOT use the Mircopalte Shaker atmosphere or with hazardous materials for which the unit was
~
not designed. Also, the user should be aware that the protection provided by the equipment may be impaired if accessories used are no provided or recommended by the manufacturer, or are used in a manner not specified by the manufacturer. CAUTION ! To avoid electrical shock, completely cut off power to the unit by disconnecting the power cord from the unit or unplug from the wall outlet. Disconnect unit from the power supply prior to maintenance and service. Any spills should be removed promptly. Bio hazard spills should be cleaned using approved liquid promptly. Solvent spills are a fire hazard. Stop the unit immediately, and DO NOT operate until clean up is complete and vapors have dissipated. DO NOT immerse the unit for cleaning. DO NOT operate the unit if it shows
in a hazardous
Digital
signs of electrical or mechanical damage.
Position of Loads
Please place loads symmetrically during operation.
Symmetrical placement
5 | Section 3 Overview
Digital Microplate Shaker 6
3.4 Capacity and Speed
Section 4 Operation
Load Type
96-well Microplate
96-well Microplate
Liquid Capacity
1/2-well deep
2/3-well deep
Fixed Tube Rack
50ר11mm 1.5ml tube
Fixed Tube Rack
40ר18mm 15ml tube
Fixed Tube Rack
21ר30mm 50ml tube
Note: These max. speeds are just for reference. We did the tests on the marble surface. Its friction coefficient is 0.32.
Cat. No.
Max Speed
1200rpm
800rpm
88882117
88882118
88882119
500rpm
450rpm
400rpm
Warning: The rotation speed is inversely proportional to the load. When the instrument is running, it is recommended to adjust the rotation speed from low to high step by step and run the instrument at an appropriate speed to avoid spillage of liquids.
This chapter covers the control panel and its operation.
4.1 Control Panel
. Run indicator: The light is on when the instrument is running and off when the instrument is in standby.
. Speed display window: The window
displays set speed (when the instrument is in standby) or current speed (when the instrument is running).
. Time display window: The window
displays cumulative time (in continuous mode) or remaining time (in timer mode).
SpeedTime
. Time setting buttons: UP/DOWN arrow buttons are used to increase/ decrease the set time of the instrument.
. Speed setting buttons: UP/DOWN arrow buttons are used to increase/ decrease the set speed of the instrument.
. Run/Stop button: Start or stop the instrument.
rpm
Run
Run
Stop
7 | Section 3 Overview7 | Section 3 Overview
Digital Microplate Shaker 8
4.2 Installation
1. Connect all the components according to the figures shown on page 4 of this manual. Use grounded power outlet.
2. Press the power switch "I" side and switch on the instrument.
4.3 Settings
Time Settings
1. Continuous mode Press the “ ” or “ ” arrow button below the Time display window. When the number shown on the Time display window starts flashing, press “ ” arrow button to decrease the time to 00:00 and then release the button. The time setting is finished after the number shown on the Time display window flashed twice.
2. Timer mode Press the “ ” or “ ” arrow button below the Time display window. When the number shown on the Time display window starts flashing, press “ ” or “ ” arrow button to increase or decrease the time value. Release the button when the time shown on the Time display window reaches the set value. The time setting is finished after the number shown on the Time display window flashed twice.
Speed Settings
Press the “ ” or “ ” arrow button
below the Speed display window. When the number shown on the Speed display window starts flashing, press “ ” or “ ” arrow button to increase or decrease the speed value. Release the button when the speed shown on the Speed display window reaches the set value. The speed setting is finished after the number shown on the Speed display window flashed twice. Note: Press the “ ” or “ ” arrow button for a longer time to accelerate the setting.
Run and Stop
Run
Press “ ” button and the instrument
Stop
will start running with the specified settings and the Run indicator light will be on. The Time display window will show the cumulative time (Continuous Mode), or remaining time (Timer Mode) and the Speed display window will show the current speed. Press “ ” button again and the
Run
Stop
instrument will slow down until it stops. The instrument will then be in standby and the two display windows will show the set values. Note: To ensure shaking operation smooth and steady, it may take 1minute for the microprocessor control system to accelerate the tray to the set speed.
Finish Operation After the operation is finished, press the power switch at the back right side of the instrument and put it into the “O” state. Unplug the instrument and store the instrument according to the storage guide.
Alarm System
Exceed the speed limit: After entering the operation mode for 5s, if the actual speed of the instrument is 0 or more than 1200rpm, the instrument will alarm immediately.
Err1: After entering the operation mode for 10s, and the instrument runs steadily (running stability means that the actual speed of the instrument is within the set value of ±10rpm). If the speed exceeds the set speed of ±10rpm, the instrument stops running and buzzes to alarm, Run indicator light goes off, and "Err1" is displayed in the Speed display window.
End of timer: The instrument buzzing alarm, the "End" is displayed in the Speed display window.
When the instrument alarms, press any key, the instrument is back to the standby mode.
Power Recovery
If the power supply is cut off suddenly while the instrument is in operation, the unit will automatically run at the previously set parameter upon power
restoration. The display windows will flash. Press any button to stop flashing.
4.4 Accessory Installation
Loosen the screws of the tray with a
1. cross screwdriver eplace the standard tray with platform
r large (figure 2).
2. Split the Fixed Tube Rack and the
holder (88882117, 88882118,
88882119), put the holder on tray and
(figure 3)
fix it with the 4 sunk screws .
3. Vertically clip the Fixed Tube Rack into the holder and make sure there is no gap between the rack and the holder (figure 4)
.
4. Insert the test tube. Note: When using single test tube, it is recommended not to set the speed too high because the imbalanced load may damage the instrument.
figure 1
figure 3
(figure 1), and
figure 2
figure 4
9 | Section 4 Operation
Digital Microplate Shaker 10
Section 5 Safety Tips and Maintenance
Section 6 Troubleshooting
Safety Tips
1. Use independent power supply.
2. Check if the local power supply voltage is suitable for use.
3. Do not drag the power supply cable when unplugging.
4. Do not use non-specified power cable or damage cable.
5. Service should only be performed by a qualified professional.
6. The power supply must be unplugged under the following situations: (1). When the unit is moved (2). When the electrical cabinet or the moving component is opened (3). When the equipment is malfunctioning (4). When the equipment is not in use
Maintenance
a. This instrument uses brushless DC motor. It is maintenance free and has a long service time, high quality, and low noise level. b. Surface can be cleaned with a mild detergent and water.
Clean Spill
If accidental spillage of liquids caused by mishandling or contained breakage occurs on the surface of the instrument, please shut down the instrument and clean up the liquid immediately. If the liquid has already spilled into the unit, cut off the power supply first and immediately clean up the liquid at the surface of the instrument. Place the instrument in a ventilated and dry environment for 24 hours before reuse. If the instrument is not functioning after drying for 24 hours, please contact the manufacturer. Warning: Disassembling/Assembling without a qualified professional's guidance may cause malfunctioning of the instrument.
Please refer to the following table to troubleshoot if any malfunction occurs.
Error
Power disconnected
Cannot start
instrument,
display window
off
No shaking of the tray
Loud noise
Other
Note:
Err1-Speed alarm If Err1 occurs, please contact Thermo Scientific Customer Service for solutions.
Power switch off
Power adaptor failure
Over-weighted or unbalanced load
Electrical malfunction
Mechanical malfunction
Microplate loose
Tray loose
Cause
If the problem still exists, contact your local sales representative.
Solution
Connect the power
Switch on power
Replace power adaptor
Adjust the weight and position of load,decrease rotation speed
Contact Thermo Scientific
Contact Thermo Scientific
Adjust position of the clamp
Fasten screws
Keep record for maintenance
microplate
11 Section 5 | Safety Tips and Maintenance
Digital Microplate Shaker 12
Section 7 Optional Accessories
Description
Platform Large (For 6 x Microplates)
96-well Microplate clamp
Platform Large Assemble (For 6 x Microplates)
Fixed Tube Rack 50ר11mm 1.5ml tube
Fixed Tube Rack 40ר18mm 15ml tube
Fixed Tube Rack 21ר30mm 50ml tube
General Power Adaptor w/US plug
General Power Adaptor w/AUS, CN plug
Cat. No.
88882114
88882115
88882116
88882117
88882118
88882119
88870126
88870127
Dimensions
292x275x14mm
54x19x19mm (7pcs/pack)
292x275x25mm
197x86x102mm
262x112x140mm
262x112x140mm
125VAC 10A 1.8m
250VAC 10A 1.8m
Max. Qty.
1
7
1
3
2
2
1
1
Figure
Description
Screw GB/T819.1 M4×8 (Platform installation)
Screw GB/T819.1 M5×10 (Tube rack installation)
Screw GB/T9074.4 M3×6 (96-well Microplate clamp installation)
Cat. No.
88882130
88882131
88882132
Dimensions
4ea/pack
4ea/pack
2ea/pack
Max. Qty.
/
/
/
Figure
General Power Adaptor w/EU plug
General Power Adaptor w/UK plug
13 | Section 7 Optional Accessories
88870128
88870129
250VAC 16A 1.8m
250VAC13A 1.8m
1
1
Digital Microplate Shaker 14
Section 8 Warranty
THERMO FISHER SCIENTIFIC STANDARD PRODUCT WARRANTY
The Warranty Period starts two weeks from the date your equipment is shipped from our facility. This allows for shipping time so the warranty will go into effect at approximately the same time your equipment is delivered. The warranty protection extends to any subsequent owner during the first year warranty period.
During the first two (2) years, component parts proven to be non-conforming in materials or workmanship will be repaired or replaced at Thermo's expense, labor included. Installation and calibration are not covered by this warranty agreement. The Technical Services Department must be contacted for warranty determination and direction prior to performance of any repairs. Expendable items, glass, filters and gaskets are excluded from this warranty.
Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period. The Technical Services Department must give prior approval for return of any components or equipment. At Thermo's option, all non-conforming parts must be returned to Thermo Fisher Scientific postage paid and replacement parts are shipped FOB destination.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, WHETHER WRITTEN, ORAL OR IMPLIED. NO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE SHALL APPLY. Thermo shall not be liable for any indirect or consequential damages including, without limitation, damages relating to lost profits or loss of products.
Your local Thermo Sales Office is ready to help with comprehensive site preparation information before your equipment arrives. Printed instruction manuals carefully detail equipment installation, operation and preventive maintenance.
If equipment service is required, please call your Technical Services Department at 1-866-984-3766, option number 2. We're ready to answer your questions on equipment warranty, operation, maintenance, service and special application. Outside the USA, please contact local Thermo Technical Services Department or local distributor for warranty information.
15 | Section 8 Warranty
Digital Microplate Shaker 16