Thermo Maxi Mix II Operations Manual & Parts List

12 /8 /1 0
2
General Specifications ........................................................................................................................................3
D
eclaration of Conformity ..............................................................................................................................3
Environmental Conditions ..............................................................................................................................3
Safety Information ..............................................................................................................................................4
Important Information ......................................................................................................................................4
Warnings ..........................................................................................................................................................4
Introduction..........................................................................................................................................................5
Intended Use ..................................................................................................................................................5
Principles of Operation ..................................................................................................................................5
Installation ..........................................................................................................................................................6
Unpackaging ....................................................................................................................................................6
Site Selection: ..................................................................................................................................................6
Specifications ..................................................................................................................................................6
Operation ............................................................................................................................................................7
Power Switch: ..................................................................................................................................................7
Speed Control: ................................................................................................................................................7
Maintenance and Servicing ................................................................................................................................8
To Replace the Power Switch: ........................................................................................................................8
To Replace Rheostat (Speed Control): ............................................................................................................8
To Replace Microswitch: ..................................................................................................................................9
To Replace Motor:..........................................................................................................................................10
To Replace Damper Springs: ........................................................................................................................11
To Replace Mixing Surface Disc: ..................................................................................................................12
To Install Single Cup Mixing Head: ..............................................................................................................12
Problem Solving Guide ....................................................................................................................................13
Replacement Parts List ....................................................................................................................................14
Ordering Procedures ........................................................................................................................................14
Exploded View ..................................................................................................................................................15
One Year Limited Warranty ..............................................................................................................................16
Table of Contents
Thermo Fisher Scientific Robert-Bosch-Strasse 1,D-63505 Langenselbold,Germany.
4
Important Information
Your Thermo Scientific Maxi Mix II Vortex Mixer has been designed with function, reliability, and safety in mind. It is your responsibility to install it in conformance with local elec­trical codes.
This manual contains important operating and safety infor­mation. You must carefully read and understand the contents of this manual prior to the use of this equipment. For safe operation, please pay attention to the alert signals through­out the manual.
Warnings
To avoid electrical shock, always:
1. Use a properly grounded electrical outlet of cor­rect voltage and current handling capacity.
2. Disconnect from power supply before servicing.
To avoid personal injury:
1. Do not use in the presence of flammable or com­bustible materials; fire or explosion may result. This device contains components which may ignite such materials.
2. Do not turn on mixer unless the speed control is turned to minimum; failure to do so may cause accessories and vessels to be thrown off the mixer.
3. Refer servicing to qualified personnel.
Safety Information
Warning
Warnings alert you to a possibility of per­sonal injury.
Caution
Cautions alert you to a possibility of damage to the equipment.
Note
Notes alert you to pertinent facts and conditions.
Alert Signals
5
Intended Use
The Type M37600 vortex mixers are general laboratory vortex mixers. They are intended for applications where vortexing of single or multiple tubes is required. Vessels with closure tops up to a maximum of 125 ml can also be vortexed.
The unit consists of 1) a mixing surface disc, 2) a three­position power switch and, 3) an adjustable speed control (See Figure 1).
Principles of Operation
The mixer incorporates a permanently lubricated motor to drive the mixing platform. A three-position power switch enables the mixer to be operated continuously or by depression of the silicone rubber top. The speed of the mixer is regulated by an adjustable rheostat. Also, suction cup feet are incorporated to prevent unit movement.
Figure 1: Maxi-Mix II
Three-position power switch
Mixing surface disc
Adjustable speed control
Introduction
6
Unpackaging
Unpack Maxi-Mix II from box. An interchangeable rubber cup for mixing test tubes and a replacement mixing surface disc are included in the box.
Site Selection
Place mixer on a sturdy flat surface.
Specifications
The electrical specifications are located on the specifica­tion plate on the back of the unit. Consult customer service if your electrical service is different than those listed on the specification plate.
Installation
Weight Dimensions Electrical
(lbs.) (Inches)
Model No. Net Ship HLTop Diameter Volts Amps Hz
M37615 676” 7” 3.5” 120 .45 50/60 M37610-33 676” 7” 3.5” 230 .20 50 M37610-33CN 676” 7” 3.5” 230 .20 50 M37614 676” 7” 3.5” 100 .56 50/60
Warning
Use a properly grounded electrical outlet of correct voltage and current handling capacity.
Do not use in the presence of flam­mable or combustible materials; fire or explosion may result. This device contains components which may ignite such materials.
Note
Some misalignment of the motor bearings in this product may have occurred during shipping and han­dling. Prior to first use, the mixer should be run at maximum speed for 10 minutes to realign the motor bearings.
Specifications
7
Power Switch:
T
he power switch is a combination ON/OFF and mode switch to enable TOUCH operation. Move the power switch to either the “FULL” position or the “TOUCH” position. In the “FULL” position, the mixing head will begin oscillating automatically. In the “TOUCH” position, the mixing head will oscillate when it is depressed.
Speed Control:
The speed control is a nonlinear type control. The oscillat­ing speed will increase at a non-uniform rate as the control is turned clockwise.
To operate Maxi-Mix, move power switch to either the “FULL” position or the “TOUCH” position. Turn speed con­trol to the desired setting. Touching the vessel to the disc will initiate the vortex action. Mixing action will intensify as greater pressure is exerted against the mixing surface. To stop the mixing action, remove vessel from disc or move the power switch to the “OFF” position. (See Maintenance and Servicing for installation of the single cup mixing head and the mixing surface disc).
Operation
W
arning
Do not turn the mixer on unless the speed control is turned to minimum; failure to do so may cause acces­sories and vessels to be thrown off the mixer.
Note
There is no absolute limit on the maximum size of vessel which can be used; much depends on tech­nique, the shape of the vessel, the viscosity of the contents, and the amount of liquid. (Vessels require closure tops to prevent contents from being spilled out during mix­ing). As a general rule, do not vortex vessels larger than 250 ml and vol­umes greater than 125 ml.
Note
Unit may need some loading to slow the motor when going from fast to slow.
Note
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equip­ment may be impared.
8
To Replace the Power Switch:
a) Disconnect mixer from power supply.
b) Peel off the mixing surface disc. Hold the mixing
platform to keep from turning and remove the platform from the counterweight assembly.
c) Turn mixer upside down and remove the four
screws securing plastic housing. Turn mixer upright and remove housing.
d) Desolder four wires from switch. Identify or
mark wires disconnected to insure proper placement and connection when reinstalling.
e) Remove defective switch.
f) Install new switch with the number “3” terminal
on top. Resolder the wires as identified or marked in Step (d).
g) Reinstall plastic housing and secure with four
screws.
h) While holding the mixing platform to keep it from
turning, secure the platform to the counter­weight assembly. If the mixing surface is in good condition, reapply it to the mixing platform.
i) Reconnect mixer to power supply.
To Replace Rheostat (Speed Control):
a) Disconnect mixer from power supply.
b) Peel off the mixing surface disc. Hold the mixing
platform to keep from turning and remove the platform from the counterweight assembly.
c) Turn mixer upside down and remove the four
screws securing plastic housing. Also, remove two screws securing dial plate.
d) Turn mixer upright and remove the housing.
Maintenance and Servicing
W
arning
Disconnect from power supply before servicing.
Refer servicing to qualified personnel.
9
e) Desolder the two wires from rheostat. Identify or mark
wires disconnected to insure proper placement and connection when reinstalling.
f) Loosen set screw on knob and remove knob. Remove
lock nut from rheostat and remove defective rheostat.
g) Insert new rheostat into dial plate and secure with lock
nut. Resolder the wires as identified or marked in Step (e).
h) Turn rheostat shaft fully counterclockwise and slide
knob over shaft. Align the pointer on knob with the edge of the graduated band. Secure knob with set screw.
i) Secure dial plate to base.
j) Reinstall plastic housing and secure with four screws.
k) While holding the mixing platform to keep it from turning,
secure the platform to the counterweight assembly. If the mixing surface is in good condition, reapply it to the mixing platform.
l) Reconnect mixer to power supply.
To Replace Microswitch:
a) Disconnect mixer from power supply.
b) Peel off the mixing surface disc. Hold the mixing plat-
form to keep from turning and remove the platform from the counterweight assembly.
c) Turn mixer upside down and remove four screws secur-
ing plastic housing.
d) Turn mixer upright and remove the housing.
e) Remove the four screws holding the motor to the rubber
mounts. Position motor on its side.
f) Desolder the two wires from microswitch and remove
MA
INTENANCE ANDSERVICING
10
microswitch from motor. Save the spring and plunger from defective microswitch. Identify or mark wires dis-
connected to insure proper placement and connec­tion when reinstalling.
g) Insert spring and plunger into the shaft of the new
microswitch. Position microswitch on mounting studs and secure with two screws. Resolder the wires as iden­tified or marked in Step (f).
h) Reposition motor and secure with four screws.
i) Reinstall plastic housing and secure with four screws.
j) While holding the mixing platform to keep it from turn-
ing, secure the platform to the counterweight assembly. If the mixing surface is in good condition, reapply it to the mixing platform.
k) Reconnect mixer to power supply.
To Replace Motor:
a) Disconnect mixer from power supply.
b) Peel off the mixing surface disc. Hold the mixing plat-
form to keep from turning and remove the platform from the counterweight assembly.
c) Turn mixer upside down and remove four screws secur-
ing plastic housing.
d) Turn mixer upright and remove the housing.
e) Desolder the two wires from motor. Identity or mark
wires disconnected to insure proper placement and connection when reinstalling.
f) Remove the counterweight assembly, brackets and
damper springs from top of motor.
g) Remove the motor and then the microswitch assembly
from bottom of motor.
h) Reinstall the components removed in Steps (f) and (g)
to the new motor.
MA
INTENANCE ANDSERVICING
11
i) Resolder the wires as identified or marked in Step (e).
j) Reinstall plastic housing and secure with four screws.
k) While holding the mixing platform to keep it from turn-
ing, secure the platform to the counterweight assem­bly. If the mixing surface is in good condition, reapply it to the mixing platform.
l) Reconnect mixer to power supply.
To Replace Damper Springs:
a) Disconnect mixer from power supply.
b) Peel off the mixing surface disc. Hold the mixing plat-
form to keep from turning and remove the platform from the counterweight assembly.
c) Turn mixer upside down and remove four screws
securing plastic housing.
d) Turn mixer upright and remove housing.
e) Remove defective damper springs.
f) Install new damper springs.
g) Reinstall plastic housing and secure with four screws.
h) While holding the mixing platform to keep it from turn-
ing, secure the platform to the counterweight assem­bly. If the mixing surface is in good condition, reapply it to the mixing platform.
i) Reconnect mixer to power supply.
To Replace Mixing Surface Disc:
a) Disconnect mixer from power supply.
b) Peel off old mixing surface disc and clean the bonding
area.
MA
INTENANCE ANDSERVICING
12
c) Remove the paper backing from the new mix-
ing surface disc and apply.
d) Reconnect mixer to power supply.
To Install Single Cup Mixing
Head:
a) Disconnect mixer from power supply.
b) Peel off mixing surface disc and while holding
the mixing platform to keep from turning, remove mixing platform.
c) While holding the rubber mixing cup to keep
from turning, install the single cup mixing head with the hold down screw.
d) Reconnect mixer to power supply.
MA
INTENANCE ANDSERVICING
Wiring Diagram
The Problem Solving tips section is intended to aid in defining and correcting possible service problems. When using the guide, select the problem category that resembles the malfunction, then proceed to the possi­ble causes category and take necessary corrective action.
13
Problem
Possible Causes
Corrective Action
Mixer doesnʼt operate when power switch is in the “Full On” position.
Check mixer connections to power supply.
Replace power switch.
Replace speed control.
Replace motor.
Mixer doesnʼt operate when power switch is in the “Touch On” position and the disc is depressed.
Not connected to power sup­ply.
Defective microswitch.
Speed control at minimum setting and too much pres­sure on disc.
Defective speed control.
Defective motor.
Check mixer connections to power supply.
Replace microswitch.
Turn speed control to higher setting.
Replace speed control.
Replace motor.
Mixer doesnʼt operate when power switch is in either “ON” position.
Check mixer connections to power supply.
Replace power switch.
Replace speed control.
Replace motor.
Speed control doesnʼt regu­late speed.
Internal chattering noise.
Defective speed control.
Going from fast to slow
Replace speed control.
Load the unit to slow the motor.
Defective spring damper links.
Replace spring damper links.
Not connected to power supply.
Defective power switch.
Defective speed control.
Defective motor.
Not connected to power supply.
Defective power switch.
Defective speed control.
Defective motor.
Problem Solving
Thermo Fisher Scientific Robert-Bosch-Strasse 1,D-63505 Langenselbold,Germany.
943-2006 or 800-926-0505
800-
15
Exploded View
Exploded View
800-926-0505.
800-
800-926-0505.
800-943-2006 or 800- 92 6- 05 05 .
North America: USA/Canada +1 86 6 98 4 37 66 ( 86 6- 9- TH ER MO ) ww w.t he rm o. co m Europe: Austria +43 1 801 40 0, Belgi um + 32 2 4 82 3 0 30 , Fr an ce + 33 2 2 80 3 21 80 , Ge rm an y national toll free 08001-536 376 , Germany international +49 6184 9 0 69 40 , It al y +3 9 02 0 2 95 05 9 43 4- 25 4- 37 5, N et he rl ands +31 76 571 4440, Nordic/Baltic countries +358 9 329 1 00 , Russia/CIS +7 (812) 703 42 15, Spain/Portuga l +3 4 93 2 23 0 9 18 , Sw it ze rl an d +4 1 44 4 54 1 2 12 , UK / Ireland +44 870 609 9203 Asia: China +86 21 6865 4588 or +86 10 8 41 9 35 88 , In di a to ll f re e 18 00 2 2 83 74 , In di a +9 1 22 6 716 2200, Japan +81 45 453 9220, Other Asian countries +852 2885 461 3 Co un tr ie s no t li st ed : +4 9 61 84 9 0 69 40 o r +3 3 2 28 03 2 18 0
Loading...