Thermo Heraeus Pico 17, Heraeus Pico 21 User manual

Page 1
Thermo Scientific
Heraeus Pico 17 / 21
Instruction Manual
50133388-2 September 2012
Visit us online to register your warranty.
www.thermoscientific.com/labwarranty
Page 2
Please reference this page for the most up-to-date information on the
following*
• web site addresses
• contact information
• copyright and trademark information
*All subsequent pages in this manual may have incorrect web site addresses and contact information.
©2013 Thermo Fisher Scientific Inc. All rights reserved. HERAEUS is a registered trademark of
Heraeus Holding GmbH licensed to Thermo Fisher Scientific. Delrin, TEFLON, and Viton are
registered trademarks of DuPont. Noryl is a registered trademark of SABIC. POLYCLEAR is a registered trademark of Hongye CO., Ltd. Hypaque is a registered trademark of Amersham Health As. RULON A and Tygon are registered trademarks of Saint-Gobain Performance Plastics. Alconox is a registered trademark of Alconox. Ficoll is a registered trademark of GE Healthcare. Haemo-Sol is a registered trademark of Haemo-Sol. Triton X-100 is a registered trademark of Sigma-Aldrich Co. LLC. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Manufacturer: Thermo Fisher Scientific Robert-Bosch-Straße 1 D - 63505 Langenselbold Germany
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is c o mplete, accurate or error-free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that migh t result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents
Page 3
Page 4
WEEE Conformity
This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96. It is marked by the following symbol:
Thermo Fisher Scientific has entered into agreements with recycling and disposal companies in all EU Member States for the recycling and disposal of this device.
Page 5
T
Table of Contents
Preface .......................................................................................................................................iii
Scope of Supply ....................................................................................................................... iii
Intended Use ..............................................................................................................................iv
Accident Prevention .................................................................................................................iv
Precautions .................................................................................................................................v
Chapter 1 Introduction and Description ............................................................................................... 1-1
Characteristics of the Heraeus Pico 17 / 21 .............................................................. 1-2
Technical Data ........................................................................................................ 1-2
Directives, Standards and Guidelines ...................................................................................... 1-4
Functions and Features ............................................................................................................ 1-4
Mains Supply ........................................................................................................... 1-5
Rotor Selection ........................................................................................................ 1-5
Chapter 2 Before Use ............................................................................................................................... 2-1
Before Setting up ..................................................................................................... 2-2
Transporting the Centrifuge .................................................................................... 2-2
Location .................................................................................................................. 2-2
Leveling the Centrifuge ............................................................................................ 2-3
Mains Connection ................................................................................................... 2-3
Storage ..................................................................................................................... 2-4
Chapter 3 Control Panel .......................................................................................................................... 3-1
Control Panel .......................................................................................................... 3-2
Keys ......................................................................................................................... 3-2
Chapter 4 Operation ................................................................................................................................. 4-1
Switch on Centrifuge ............................................................................................... 4-2
Lid Opening ............................................................................................................ 4-2
Close Door .............................................................................................................. 4-2
Rotor Installation ..................................................................................................... 4-2
Entering Parameters ................................................................................................. 4-4
Centrifugation ......................................................................................................... 4-6
Short-term Centrifugation ....................................................................................... 4-7
Removing the Rotor ................................................................................................ 4-8
Aerosol-tight Rotors ................................................................................................. 4-8
Audible Alarm ......................................................................................................... 4-8
Turning off the Centrifuge ...................................................................................... 4-9
Chapter 5 Maintenance and Care ......................................................................................................... 5-1
Thermo Scientific Heraeus Pico 17 / 21 i
Page 6
Inhalt
Cleaning Intervals .................................................................................................... 5-2
Cleaning .................................................................................................................. 5-2
Disinfection ............................................................................................................. 5-3
Decontamination ..................................................................................................... 5-4
Autoclaving .............................................................................................................. 5-5
Service of Thermo Fisher Scientific .......................................................................... 5-5
Shipping and Deposing of Accessories ..................................................................... 5-6
Chapter 6 Troubleshooting ..................................................................................................................... 6-1
Mechanical Emergency Door Release ...................................................................... 6-2
Troubleshooting by User ......................................................................................... 6-3
When to contact a Service Technician ..................................................................... 6-6
Chemical Compatibility Chart ...............................................................................A-1
Contact Information ................................................................................................B-1
ii Heraeus Pico 17 / 21 Thermo Scientific
Page 7
P
Preface
Before starting to use the centrifuge, read through this instruction manual carefully and follow the instructions.
The information contained in this instruction manual is the property of Thermo Fisher Scientific; it is forbidden to copy or pass on this information without explicit approval.
Failure to follow the instructions and safety information in this instruction manual will result in the expiration of the sellers warranty.
Scope of Supply
Article Number Quantity Check
Centrifuge Heraeus Pico 17 / 21 1
Power supply cable 1
Rotor 1
20360104 Square box wrench 1
50134706 CD with Manual 1
If any parts are missing, please contact your nearest Thermo Fisher Scientific representative.
This symbol refers to general hazards. WARNING means that injuries or material damage or contamination could occur. CAUTION means that material damage could occur.
This symbol refers to biological hazards.
Observe the information contained in the instruction manual to keep yourself and your environment safe.
This symbol refers to hot surfaces. Observe the information contained in the instruction manual to keep yourself and your environment safe.
Thermo Scientific Heraeus Pico 17 / 21 iii
Page 8
P Intended Use
g
cm
3
---------- -
Intended Use
This centrifuge is a laboratory product designed to separate components by generation of a
relative centrifugal force. It separates human samples (e.g. blood, urine and other body fluids) collected in appropriate containers, either alone or after addition of reagents or other additives.
As general-purpose centrifuge, it is designed to also run other containers filled with chemicals,
environmental samples and other non-human body samples.
Together with the hematocrit rotor 75003473 the centrifuge can be used as an in-vitro
diagnostics device to define the hematocrit value by centrifugation. Apply DIN 58933 to define the hematocrit value.
The centrifuge is to be used for separating materials of different density or particle size
suspended in a liquid.
Maximum sample density at maximum speed: 1.2
Accident Prevention
WARNING If a hazardous situation occurs, turn off the power supply to the centrifuge
and leave the area immediately.
WARNING In case of rotor failure the centrifuge can be damaged. The coolant can escape. Ventilate the room well and leave it. Inform customer service.
WARNING Plug the centrifuge only into sockets which have been properly grounded.
Prerequisite for the safe operation of the Heraeus Pico 17 / 21 is a work environment in compliance with standards, directives and trade association safety regulations and proper instruction of the user.
The safety regulations contain the following basic recommendations:
Maintain a radius of at least 30 cm around the centrifuge.
Implementation of special measures which ensure that no one can approach the
centrifuge for longer than absolutely necessary while it is running.
The mains plug must be freely accessible at all times. Pull out the power supply plug or
disconnect the power supply in an emergency.
iv Heraeus Pico 17 / 21 Thermo Scientific
Page 9
Precautions
P Precautions
In order to ensure safe operation of the Heraeus Fresco 17 / 21, the following general safety regulations must be followed:
The centrifuge should be operated by trained specialists only.
The centrifuge is to be used for its intended use only.
While handling centrifuge, rotor, and samples you must wear laboratory clothing (e.g.
gloves).
Do not move the centrifuge while it is running.
Do not lean on the centrifuge.
Use only rotors and accessories for this centrifuge which have been approved by Thermo
Fisher Scientific. Exceptions to this rule are commercially available glass or plastic centrifuge tubes, provided they have been approved for the speed or the RCF value of the rotor.
Do not use rotors which show any signs of corrosion and/or cracks.
Do not touch the mechanical components of the rotor and do not make any changes to
the mechanical components.
Use only with rotors which have been properly installed. Follow the instructions in
section “Rotor Installation” on page 4-2.
Use only with rotors which have been loaded properly. Follow the instructions given in
the rotor manual.
Never overload the rotor. Follow the instructions given in the rotor manual.
Never open the lid until the rotor has come to a complete stop and this has been
confirmed in the display.
The lid emergency release may be used in emergencies only to recover the samples from
the centrifuge, e.g. during a power failure (see section “Mechanical Emergency Door
Release” on page 6-2).
Never use the centrifuge if parts of its cover panels are damaged or missing.
Do not touch the electronic components of the centrifuge or alter any electronic or
mechanical components.
Please observe the safety instructions.
Thermo Scientific Heraeus Pico 17 / 21 v
Page 10
P Precautions
Remaining risk: Improper use can cause damages, contamination, and inju­rieswith fatal consequences.
Remaining risk: Improper use can cause damages, contamination, and inju­rieswith fatal consequences.
Please pay particular attention to the following aspects:
Location: well-ventilated environment, set-up on a level and rigid surface with adequate
load-bearing capacity.
Rotor installation: make sure the rotor is locked properly into place before operating the
centrifuge.
Especially when working with corrosive samples (salt solutions, acids, bases), the
accessory parts and vessel have to be cleaned carefully.
Always balance the samples.
Centrifuging hazardous substances:
Do not centrifuge explosive or flammable materials or substances which could react
violently with one another.
The centrifuge is neither inert nor protected against explosion. Never use the centrifuge
in an explosion-prone environment.
Do not centrifuge inflammable substances.
Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms
without suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the Bio-safety Manual" of the World Health Organization (WHO)), aerosol-tight biological seals have to be used.
For materials in a higher risk group, extra safety measures have to be taken.
If toxins or pathogenic substances have gotten into the centrifuge or its parts,
appropriate disinfection measures have to be taken (see “Disinfection” on page 5-3).
Highly corrosive substances which can cause material damage and impair the
mechanical stability of the rotor, should only be centrifuged in corresponding protective tubes.
vi Heraeus Pico 17 / 21 Thermo Scientific
Page 11
1
Introduction and Description
Contents
“Characteristics of the Heraeus Pico 17 / 21” on page 1-2
“Technical Data” on page 1-2
“Directives, Standards and Guidelines” on page 1-4
“Functions and Features” on page 1-4
“Mains Supply” on page 1-5
“Rotor Selection” on page 1-5
Thermo Scientific Heraeus Pico 17 / 21 1-1
Page 12
1 Introduction and Description
Characteristics of the Heraeus Pico 17 / 21
Characteristics of the Heraeus Pico 17 / 21
Several rotors with a wide range of tubes can be used.
The set speed is reached within seconds. The maintenance-free induction motor ensures quiet and low-vibration operation even at high speeds, and guarantees a very long lifetime.
The user-friendly control panel makes it easy to pre-set the speed, RCF value, run time, and temperature. You can choose between the display of speed and RCF or the entry mode.
These settings can be changed even while the centrifuge is running.
With the help of the key , you can also centrifuge a sample for just a few seconds, if called for.
The Heraeus Pico 17 / 21 is equipped with various safety features:
The housing is made of impact-resistant plastic and the interior of armor steel.
The door is equipped with a view port and a lock.
The centrifuge door can only be opened while the centrifuge is switched on and the rotor has come
to a complete stop. The centrifuge cannot be started until the door has been closed properly.
Door emergency release: For emergencies only, e.g. during power failures (see “Mechanical
Emergency Door Release” on page 6-2)
Technical Data
The technical data of the Heraeus Pico 17 / 21 is listed in the following table.
Table 1-1. Technical data Heraeus Pico 17
Feature Value
Environmental Conditions -Use in interior Spaces
Permissible Ambient Temperature during Operation
Permissible Ambient Temperature during Storage and Transport
-Altitudes of up to 2,000 m above Sea Level
-Max. relative Humidity 80% up to 31°C; decreasing linearly up to 50% relative Humidity at 40°C.
+5°C to +40°C +5°C to +40°C +5°C to +40°C
-10°C to +50°C -10°C to +50°C -10°C to +50°C
Overvoltage Category II II II
Pollution Degree 2 2 2
Heat Dissipation Heraeus Pico 17 Heraeus Pico 17 Heraeus Pico 17
230 V 120 V 100 V
614 BTU/h 614 BTU/h 580 BTU/h
IP 20 20 20
Run Time unlimited unlimited unlimited
1-2 Heraeus Pico 17 / 21 Thermo Scientific
Page 13
Table 1- 1. Technical data Heraeus Pico 17
Feature Value
1 Introduction and Description
Technical Data
Max Speed n
Min Speed n
max
min
Maximum RCF-Value at n
max
13 300 rpm (depending on the Rotor)
13 300 rpm (depending on the Rotor)
13 300 rpm (depending on the Rotor)
300 rpm 300 rpm 300 rpm
17,000 x g 17,000 x g 17,000 x g
Maximum Kinetic Energy <1.90 kNm <1.90 kNm <1.90 kNm
Noise Level at maximum Speed < 56 dB (A) < 56 dB (A) < 56 dB (A)
Dimensions
Height Width Depth
230 mm 240 mm 350 mm
230 mm 240 mm 350 mm
230 mm 240 mm 350 mm
Weight with emptyRotor 10.5 kg 10.5 kg 10.5 kg
Table 1- 2. Technical data Heraeus Pico 21
Feature Value
Environmental Conditions -Use in interior Spaces
-Altitudes of up to 2,000 m above Sea Level
-max. relative humidity 80% up to 31°C; decreasing linearly up to 50% relative humidity at 40°C.
Permissible Ambient Temperature during
+5°C to +40°C +5°C to +40°C +5°C to +40°C
Operation
Permissible Ambient Temperature during
-10°C to +50°C -10°C to +50°C -10°C to +50°C
Storage and Transport
Overvoltage Category II II II
Pollution Degree 2 2 2
Heat Dissipation Heraeus Pico 21 Heraeus Pico 21 Heraeus Pico 21
230 V 120 V 100 V
785 BTU/h 751 BTU/h 785 BTU/h
IP 20 20 20
Run Time unlimited unlimited unlimited
Max Speed n
Min Speed n
max
min
Maximum RCF-Value at n
max
14 800 rpm (depending on the Rotor)
14 800 rpm (depending on the Rotor)
14 800 rpm (depending on the Rotor)
300 rpm 300 rpm 300 rpm
21,100 x g 21,100 x g 21,100 x g
Maximum Kinetic Energy <2.35 kNm <2.35 kNm <2.35 kNm
Noise level at maximum Speed < 56 dB (A) < 56 dB (A) < 56 dB (A)
Dimensions
Thermo Scientific Heraeus Pico 17 / 21 1-3
Page 14
1 Introduction and Description
Directives, Standards and Guidelines
Table 1-2. Technical data Heraeus Pico 21
Feature Value
Height Width Depth
Weight with empty Rotor 10.5 kg 10.5 kg 10.5 kg
230 mm 240 mm 350 mm
Directives, Standards and Guidelines
Table 1-3. Directives, Standards and Guidelines
Tension / Frequency
230V 50Hz/60Hz,
100V 50Hz cCSAus: UL Std. No. 61010-1 (2
Marking
CE:
98/79/EC In-vitro-Diagnostika (IvD)
2006/42/EC Machine Directive:
2004/108/EC EMC Directive
2006/95/EC Low Voltage Directive:
Produced and inspected according to the following standards and guidelines
EN 61010-1, 2
EN 61010-2-020, 2
EN 61010-2-101
EN 61326-1
EN 61326-2-6
EN 55011B
EN 61000-6-2
EN ISO 13485
/CSA-C22.2 No. 61010-1, -04nd Edition
CAN/CSA-C22.2 No. 61010-2-020-09 Part 2-020:
IEC61010-2-020 (2
(Pollution degree 2, Overvoltage category II)
IEC 61010-2-101
nd
Edition
nd
Edition
nd
nd
Edition)
230 mm 240 mm 350 mm
Edition)
230 mm 240 mm 350 mm
Functions and Features
The following table gives an overview of the important functional and performance characteristics of the Heraeus Pico 17 / 21.
Table 1-4. Functions and Features
Component / Function Description / Features
Structure / Housing Sheet Metal with attached Plastic Housing and Steel Chamber
Rotor Chamber Plastic
Up to 48 ml of spilled Liquid is retained in the Chamber and cannot enter the Instrument.
Drive Induction Drive without Carbon Brushes
Keys and Display Easy-to-clean Keypad and Display Surface
Control Microprocessor-controlled
Internal Memory The most recent Data is saved
1-4 Heraeus Pico 17 / 21 Thermo Scientific
Page 15
Table 1- 4. Functions and Features
Component / Function Description / Features
Functions RCF, Temperature, and pre-temp Selection
Door Lock Automatic Door Closing and Locking starting from an initial hold Position
Mains Supply
The following table contains an overview of the electrical connection data for the Heraeus Pico 17 / 21. This data is to be taken into consideration when selecting the mains connection socket.
Table 1-5 . Electrical connection data
1 Introduction and Description
Mains Supply
Cat.
75002410 PICO 17 230 V 50 / 60 Hz 1.4 A 180 W 4 A, 2-pole circuit
75002411 PICO 17 120 V 60 Hz 2.6 A 180 W 8 A, 2-pole circuit
75002412 PICO 17 100 V 60 Hz 2.9 A 170 W 8 A, 2-pole circuit
75002415 PICO 21 230 V 50 / 60 Hz 1.7 A 230 W 4 A, 2-pole circuit
75002416 PICO 21 120 V 60 Hz 3.4 A 220 W 8 A, 2-pole circuit
75002417 PICO 21 100 V 60 Hz 3.9 A 230 W 8 A, 2-pole circuit
Mains Voltage
Frequency
Rated Current
Power Consumption
Equipment Fuse
breaker
breaker
breaker
breaker
breaker
breaker
Rotor Selection
The Heraeus Pico 17 / 21 is is supplied with a rotor.
Various rotors are available to choose from.
Building Fuse
16 AT
15 AT
16 AT
16 AT
15 AT
16 AT
Microliter 24 x 2 7500 3424
Microliter 36 x 0.5 7500 3436
Dual Rotor 18 x 2 / 0.5 7500 3418
PCR Rotor 4x8 7500 3440
PCR Rotor 8x8 7500 3489
Hematocrit Rotor 7500 3473
The technical data of the rotors and the corresponding adapters and reduction sleeves for various commercially available containers can be found in the corresponding rotor operating manuals.
For more information visit our website at: http://www.thermo.com
Thermo Scientific Heraeus Pico 17 / 21 1-5
Page 16
2
Before Use
Contents
“Before Setting up” on page 2-2
“Transporting the Centrifuge” on page 2-2
“Location” on page 2-2
“Leveling the Centrifuge” on page 2-3
“Mains Connection” on page 2-3
“Storage” on page 2-4
Thermo Scientific Heraeus Pico 17 / 21 2-1
Page 17
2 Before Use
Before Setting up
Before Setting up
1. Check the centrifuge and the packaging for any shipping damage. Inform the shipping company and Thermo Fisher Scientific immediately if any damage is discovered.
2. Remove the packaging.
3. Remove the rotor transport protection. The cardboard is suppose to prevent damage due to impact.
4. Check the order for completeness (see “Scope of Supply” on page iii). If the order is incomplete, please contact Thermo Fisher Scientific.
Transporting the Centrifuge
WARNING Always lift the centrifuge on both sides. Never lift the centrifuge by its front
or the back panel. Transport the centrifuge if possible in its packaging and with the rotor transport protection.
Location
Due to its weight (see “Technical Data” on page 1-2), the centrifuge should be carried by several
people.
Always lift the centrifuge at both sides on the bottom plate.
Figure 2-1.
The centrifuge can be damaged by impacts.
The centrifuge should only be operated indoors.
Lifting the centrifuge at both sides.
The set-up location must fulfill the following requirements:
A safety zone of at least 30 cm must be maintained around the centrifuge.
People and hazardous substances must be kept out of the safety zone while centrifuging.
The supporting structure must be stable and free of resonance, for example a level laboratory
bench.
The supporting structure must be suitable for horizontal setup of the centrifuge.
2-2 Heraeus Pico 17 / 21 Thermo Scientific
Page 18
The location must be free of grease and dust.
Main Power Switch
Mains plug
The centrifuge should not be exposed to heat and strong sunlight.
WARNING UV rays reduce the stability of plastics. Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
The set-up location must be well-ventilated at all times.
Leveling the Centrifuge
The horizontal leveling of the centrifuge must be checked every time after moving it to a different location. The supporting structure must be suitable for horizontal setup of the centrifuge.
CAUTION If the centrifuge is not level, imbalances can occur and the centrifuge can be damaged. Do not place anything under the centrifuge to level.
2 Before Use
Leveling the Centrifuge
Mains Connection
Figure 2-2. Mains Connection
1. Turn off the power supply switch on the back (press "0").
2. Plug the centrifuge into grounded electrical sockets only.
Note The centrifuge must be connected directly to the socket. Extension cords and multiple
sockets are not allowed.
3. Check whether the cable complies with the safety standards of your country.
Thermo Scientific Heraeus Pico 17 / 21 2-3
Page 19
2 Before Use
Storage
Storage
4. Make sure that the voltage and frequency correspond to the figures on the rating plate. Establish the connection to the power supply with the connecting cable.
Before storing the centrifuge and the accessories it must be cleaned and if necessary disinfected and
decontaminated.
Store the centrifuge in a clean, dust-free location.
Be sure to place the centrifuge on its feet.
Avoid direct sunlight.
2-4 Heraeus Pico 17 / 21 Thermo Scientific
Page 20
3
Control Panel
Contents
“Control Panel” on page 3-2
“Keys” on page 3-2
Thermo Scientific Heraeus Pico 17 / 21 3-1
Page 21
3 Control Panel
PULSE
OPEN
STOP
START
Run Time
Speed / RCF Value
Toggle k e y
Control Panel
Control Panel
In the control panel you find the keys and displays of the centrifuge. The main power switch is located on the left side. All parameters can be selected and changed during operation.
Keys
The keys allow user intervention for controlling the operating mode as follows:
Key Display contents
Start
Stop
Open door
Pulse
Normal start of the centrifuge.
End run manually.
Automatic release (possible only when device is switched on); Emergency release (see “Mechanical Emergency Door Release” on page 6-2).
By pressing the key the centrifuge starts immediately and accelerates up to the end speed. Releasing the key initiates a stopping process at the highest braking curves.
By pressing the key up and down you change the value in thedisplay. If you hold the key pressed, the display changes continuously at first slowly and after a few seconds at an accelerated pace to the higher or lower values. By pressing the key briefly, you increase or decrease the speed in one step. When pressing both arrows the cursor moves to the left.
Changing the display mode
3-2 Heraeus Pico 17 / 21 Thermo Scientific
Use the key to change the display mode. (speed / RCF value, sample / chamber temperature, run time counter from start or preset speed on)
Page 22
4
Operation
Contents
“Switch on Centrifuge” on page 4-2
“Lid Opening” on page 4-2
“Close Door” on page 4-2
“Rotor Installation” on page 4-2
“Entering Parameters” on page 4-4
“Centrifugation” on page 4-6
“Short-term Centrifugation” on page 4-7
“Removing the Rotor” on page 4-8
“Aerosol-tight Rotors” on page 4-8
“Audible Alarm” on page 4-8
“Turning off the Centrifuge” on page 4-9
Thermo Scientific Heraeus Pico 17 / 21 4-1
Page 23
4 Operation
Switch on Centrifuge
Switch on Centrifuge
1. Switch on the centrifuge. The centrifuge shows the actual value in the display. Speed and run time show 0. The display shows the current temperature of the sample.
Lid Opening
1. Press the key. The display shows the following message:
Note Use the emergency release only for malfunctions and power failures (see “Mechanical
Emergency Door Release” on page 6-2).
Close Door
1. Close the door by pressing down on it lightly in the middle or on both sides of it.
Note The door should audibly click into place.
CAUTION Do not slam the door.
Rotor Installation
The approved rotors for the Heraeus Pico 17 / 21 are listed in section “Rotor Selection” on page 1-5. Use only the rotors and accessories from this list in the centrifuge.
CAUTION Unapproved or incorrectly combined accessories can cause serious damage to the centrifuge.
Proceed as follows:
1. Open the door of the centrifuge and if necessary remove any dust, foreign objects or residue from the chamber. Thread and o-ring must be clean and undamaged.
4-2 Heraeus Pico 17 / 21 Thermo Scientific
Page 24
4 Operation
Rotor Installation
CAUTION Do not install the rotor when the temperature difference between shaft and rotor lock is >20°C. Otherwise the rotor might jam during the installation. A jammed rotor can lead to damages of centrifuge and rotor.
2. Hold the rotor over the centrifuge spindle. The two bars in the labeling on the upper side of the rotor (a) must be above the retaining pin (b) of the centrifuge spindle.
a
b
3. Let the rotor slide slowly down the centrifuge spindle.
CAUTION Do not force the rotor onto the centrifuge spindle. If the rotor is very light, then it may be necessary to press it onto the centrifuge spindle with a bit of pressure.
4. Thread the fastening tool into the centrifuge spindle clockwise. Hold the rotor with the other hand into position.
5. Close the rotor.
6. Check if the rotor is properly installed by lifting it slightly on the handle.
Thermo Scientific Heraeus Pico 17 / 21 4-3
Page 25
4 Operation
Speed display
RCF display
Entering Parameters
WARNING Check for any damage to the rotor: Damaged rotors must not be used. Keep the centrifuge spindle area of the rotor clear of objects.
CAUTION Check that the rotor is properly locked on the centrifuge spindle before each use by pulling it at its handle. Fasten the rotor if necessary.
Be sure to check all sealings before starting any aerosol-tight applications.
See the information in the rotor instruction manual.
7. Close the centrifuge door.
Entering Parameters
Note Due to limited display digits there is a need to round the values. The direct comparison
between the two values speed and RCF is therefore restricted.
Select Speed or RCF-Value
1. Press the key under the left display. When the lower indicator is lit the display shows the speed. Press the key to toggle between the two modes.
2. When the upper indicator is lit the display shows the RCF value. Press the key to toggle between the two modes.
4-4 Heraeus Pico 17 / 21 Thermo Scientific
Page 26
4 Operation
RCF 11 18,
n
1000
-------------

2
r××=
Entering Parameters
Pre-selecting Speed
1. Enter the desired value by pressing the key repeatedly, until the desired value shows. You can adjust the speed in 100 rpm increments.
2. Press the key to confirm the pre-selected value. If you do not press any key, the display flashes for a few seconds The new preselected value is now stored and the display shows the actual value.
Note The centrifuge speed can be set to a minimum of 300rpm.
The maximum speed depends on the centrifuge variant.
Pre-selecting RCF-Value
1. Enter the desired value by pressing the key repeatedly, until the desired value shows. You can adjust the RCF pre-selected value in steps of 100 xg.
2. Press the key to confirm the pre-selected value. If you do not press any key, the display flashes for a few seconds The new preselected value is now stored and the display shows the actual value.
Note The RCF value can be set to a minimum of 100 xg.
The maximum speed depends on the centrifuge variant. The displayed RCF value is always corresponding to the maximum of centrifuge radius of the micro liter rotor 24 x 2 (75003424). Refer to the “Explanation of RCF-Value” on page 4-5 and the rotor manual for further information.
Explanation of RCF-Value
The relative centrifugal force (RCF) is given as a multiple of the force of gravity g. It is a unitless numerical value which is used to compare the separation or sedimentation capacity of various centrifuges, since it is independent of the type of device. Only the centrifuging radius and the speed come into play in it:
r = centrifuging radius in cm
n = rotational speed in rpm
The maximum RCF value is related to the maximum radius of the tube opening.
Thermo Scientific Heraeus Pico 17 / 21 4-5
Remember that this value is reduced depending on the tubes and adapters used.
This can be accounted for in the calculation above if required.
Page 27
4 Operation
Centrifugation
Run Time Preselection
Note You can select a run time between 1 and 99 min or continuous operation.
1. Enter the desired value by pressing the key below the display in the middle repeatedly, until the desired value shows. You can adjust the run time in steps of 1 min.
2. Press the key to confirm the pre-selected value. If you do not press any key, the display flashes for a few seconds The new preselected value is now stored and the display shows the actual value.
Continuous Operation
1. Press the key until hd is shown.
2. During continuous operation, the centrifuge will continue running until you stop it manually with the key.
CAUTION Please note that the lifetime of particularly plastic rotor tubes is limited.Continuous operation (extended use) may cause damage to them.
Changing the settings during the run
You can change the settings during the run as follows:
1. Press one of the two The current value will switch into the pre-selecting value mode.
keys in the control panel.
2. Enter the new value as described above.
3. Press the key. The value is taken over immediately.
Centrifugation
Once the rotor has been properly installed, the main switch has been turned on and the door has been closed, you can start centrifuging.
4-6 Heraeus Pico 17 / 21 Thermo Scientific
Page 28
4 Operation
Short-term Centrifugation
Starting Centrifuge Program
Press the key on the control panel. The centrifuge accelerates to the preset speed with the time display active.
The run display begins to count down from the pre-selected value. If the remaining run time is less then 1 minute, the remaining time is given in seconds.
The circulating indicators in the left display represent the spinning rotor.
In continuous operation hd the time display counts up. The displayed run time is first in seconds. After one minute the displays changes every minute.
Stopping the Centrifugation Program
With preset Run Time
If the run time has been pre-selected, and all you have to do, is to wait until the centrifuge terminates the run automatically.
As soon as the speed drops to zero, the message END will appear in the display. By pressing the key, you can open the door and remove the centrifuge material.
You can also stop the centrifuging program manually at any time by pressing the key.
Continuous Operation
If you selected continuous operation (see “Continuous operation” on page 4-6), you will have to stop the centrifuge manually.
1. Press the key on the control panel. The centrifuge will be decelerated at the designated rate.
2. Press the key to open the centrifuge door and the remove the samples when the message END appears in the display.
Short-term Centrifugation
For short-term centrifuging, the Heraeus Pico 17 / 21 has a PULSE- function.
By holding down the key, spinning will start and continue until the key is let go.
The centrifuge accelerates and brakes at maximum power. The pre-selected value is ignored.
Note The centrifuge accelerates to the maximum speed.
Check carefully whether you have to maintain a certain speed for your application.
The displayed run time is first in seconds. After one minute the displays changes every minute.
After an short-term centrifugations the set values are restored.
Thermo Scientific Heraeus Pico 17 / 21 4-7
Page 29
4 Operation
Removing the Rotor
Removing the Rotor
To remove the rotor, proceed as follows:
1. Open the centrifuge door.
2. Unscrew the rotor with the allen wrench.
3. Grab the rotor in the middle. Pull the rotor directly upwards and remove it from the centrifuge spindle. Make sure not to tilt the rotor while doing this.
WARNING Be careful, when you change a rotor after a run. Centrifuge spindle and motor bearing assembly can be hot (>55°C).
Aerosol-tight Rotors
When using an aerosol tight lid the rotor can be removed with the lid closed. This is to protect you and the samples.
Audible Alarm
Error
Accompanying all error messages, a warning signal is given out.
Press any key to silence the warning signal.
End of Run
By default there is an acoustic signal at the end of any centrifugation run. To switch off this signal proceed as follows:
Press the key when you turn on the centrifuge. The display shows:
or
4. Press the key under the display in the middle. The acoustic signal is turned on or off.
5. Press the key to confirm the pre-selected value.
4-8 Heraeus Pico 17 / 21 Thermo Scientific
Page 30
Turning off the Centrifuge
1. To turn off the centrifuge put the mains switch to "0".
Note The centrifuge is equipped with a special switch for balancing potential voltage discrepancies
in the power grid. After pressing the mains switch the display therefore may still flash up to 10 seconds.
4 Operation
Turning off the Centrifuge
Thermo Scientific Heraeus Pico 17 / 21 4-9
Page 31
5
Maintenance and Care
Contents
“Cleaning Intervals” on page 5-2
“Cleaning” on page 5-2
“Disinfection” on page 5-3
“Decontamination” on page 5-4
“Autoclaving” on page 5-5
“Service of Thermo Fisher Scientific” on page 5-5
Thermo Scientific Heraeus Pico 17 / 21 5-1
Page 32
5 Maintenance and Care
Cleaning Intervals
Cleaning Intervals
For the sake of personal, environmental, and material protection, it is your duty to clean and if necessary disinfect the centrifuge on a regular basis.
Maintenance Recommended Interval
Clean Rotor Chamber Daily or when polluted
Clean Rotor Daily or when polluted
Accessories Daily or when polluted
Cabinet Once per Month
Filter Unit Every three Months
Ventilation Holes Every six Months
CAUTION Refrain from using any other cleaning or decontamination procedure than those recommended here, if you are not entirely sure that the intended procedure is safe for the equipment. Use only approved cleansers. If in doubt, contact Thermo Fisher Scientific.
Cleaning
When cleaning centrifuge mind the following:
Use warm water with a neutral solvent.
Never use caustic cleaning agents such as soap suds, phosphoric acid, bleaching solutions or
scrubbing powder.
Rinse the cavities out thoroughly.
Use a soft brush without metal bristles to remove stubborn residue.
Afterwards rinse with distilled water.
Place the rotors on a plastic grate with their cavities pointing down.
If drying boxes are used, the temperature must never exceed 50°C, since higher temperatures could
damage the material and shorten the lifetime of the parts.
Use only disinfectants with a pH of 6-8.
Dry aluminum parts off with a soft cloth.
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (7000
9824). Also treat the cavities with oil.
Store the aluminum parts at room temperature or in a cold-storage room with the cavities pointing
down.
5-2 Heraeus Pico 17 / 21 Thermo Scientific
Page 33
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment.
Clean centrifuge and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Release the rotor.
5. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
6. Remove the centrifuge tubes and adapters.
7. Use a neutral cleaning agent with a pH value between 6 and 8 for cleaning.
5 Maintenance and Care
Disinfection
8. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (7000
Disinfection
Disinfect the centrifuge immediately whenever infectious material has spilled during centrifugation.
maximum temperature of 50°C.
9824). Also treat the cavities with oil..
CAUTION When cleaning, do not allow liquids, especially organic solvents, to get on the drive shaft or the bearings of the centrifuge. Organic solvents break down the grease in the motor bearing. The drive shaft could freeze up.
WARNING Infectious material can get into the centrifuge when a tube breaks or as a result of spills. Keep in mind the risk of infection when touching the rotor and take all necessary precautions. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately. Take other precautions if need be.
The rotor chamber and the rotor should be treated preferably with a neutral disinfectant.
CAUTION Before using any cleaning or decontamination methods except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment. Observe the safety precautions and handling instructions for the cleaning agents used.
Contact the Service Department of Thermo Fisher Scientific for questions regarding the use of other disinfectants.
Disinfect the rotor and accessories as follows:
Thermo Scientific Heraeus Pico 17 / 21 5-3
Page 34
5 Maintenance and Care
Decontamination
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Release the rotor.
5. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
6. Remove the centrifuge tubes and adapters and dispose of them or disinfect them.
7. Treat the rotor and accessories according to the instructions for the disinfectant. Adhere strictly to
8. Be sure the disinfectant can drain off the rotor.
9. Rinse the rotor and accessories thoroughly with water.
10. Dispose of the disinfectant according to the applicable guidelines.
11. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
the given application times.
maximum temperature of 50°C.
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (7000
9824). Also treat the cavities with oil.
Decontamination
Decontaminate the centrifuge immediately whenever radioactive material has spilled during centrifugation.
WARNING Infectious material can get into the centrifuge when a tube breaks or as a result of spills. Keep in mind the risk of infection when touching the rotor and take all necessary precautions. In case of contamination, make sure that others are not put at risk. Decontaminate the affected parts immediately. Take other precautions if need be.
CAUTION Before using any cleaning or decontamination methods except those recommended by the manufacturer, users should check with the manufacturer that the proposed method will not damage the equipment.
For general radioactive decontamination use a solution of equal parts of 70% ethanol, 10% SDS and water.
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Release the rotor.
5. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
6. Remove the centrifuge tubes and adapters and dispose of them or disinfect them.
5-4 Heraeus Pico 17 / 21 Thermo Scientific
Page 35
7. Rinse the rotor first with ethanol and then with de-ionized water.
Adhere strictly to the given application times.
8. Be sure the decontamination solution can drain off the rotor.
9. Rinse the rotor and accessories thoroughly with water.
10. Dispose of the decontamination solution according to the applicable guidelines.
11. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (7000
Autoclaving
1. Before autoclaving clean rotor and accessories as described above.
5 Maintenance and Care
Autoclaving
maximum temperature of 50°C.
9824). Also treat the cavities with oil.
2. Place the rotor on a flat surface.
Rotors and adapter can be autoclaved at 121°C.
The maximum permissible autoclave cycle is 20 minutes at 121°C.
Note No chemical additives are permitted in the steam.
CAUTION Never exceed the permitted temperature and duration when autoclaving.
If the rotor shows signs of corrosion or wear, it must be replaced.
Service of Thermo Fisher Scientific
Thermo Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an authorized service technician. The service technician checks the following:
the electrical equipment
the suitability of the set-up site
the door lock and the safety system
the rotor
the fixation of the rotor and the drive shaft
Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are performed for free during the warranty period and afterwards for a charge.
This is only valid if the centrifuge has only been maintained by a Thermo Fisher Scientific service technician.
Thermo Scientific Heraeus Pico 17 / 21 5-5
Page 36
5 Maintenance and Care
Shipping and Deposing of Accessories
Shipping and Deposing of Accessories
WARNING Before shipping or deposing centrifuges and accessories you have ton
cleanand if necessary disinfect or decontaminate everything. Before storing the centrifuge and the accessories it must be cleaned and if necessary disinfected and decontaminated.
5-6 Heraeus Pico 17 / 21 Thermo Scientific
Page 37
6
Troubleshooting
Contents
“Mechanical Emergency Door Release” on page 6-2
“Troubleshooting by User” on page 6-3
“When to contact a Service Technician” on page 6-6
Thermo Scientific Heraeus Pico 17 / 21 6-1
Page 38
6 Troubleshooting
Emergency Door Release
Mechanical Emergency Door Release
Mechanical Emergency Door Release
During a power failure, you will not be able to open the centrifuge door with the regular electric lid release. A mechanical override is provided to allow sample recovery in the case of an emergency. However, this should be used only in emergencies and after the rotor has come to a complete stop.
WARNING The rotor can still be spinning at high speed. If touched, it can cause serious injuries.
Always wait a few minutes until the rotor has come to a stop without braking. The brake does not work when there is no current. The braking process lasts much longer than usual.
Proceed as follows:
1. Make sure the rotor has stopped (view port in the door).
WARNING Never use your hand or any tools to brake the rotor.
2. Pull out the power supply plug.
3. Insert a 3 inch long wire (e.g. a staple) into the hole above the control panel.
4. Press the centrifuge door down gently. Push the wire further into the hole until you hear and feelthe door latch unlocking.
5. Remove the wire from the hole and open the centrifuge door. Now you can remove your samples.
Figure 6-1.
Reconnect the centrifuge once the power has been restored. Switch on the centrifuge.
6-2 Heraeus Pico 17 / 21 Thermo Scientific
Emergency Door Release
Page 39
Troubleshooting by User
If problems occur other than those listed in this table, the authorized customer service representative must be contacted.
Failure message Problem with centrifuge Possible causes and cures
6 Troubleshooting
Troubleshooting by User
Display remains dark. The drive stops. The centrifuge decelerates
without braking. The centrifuge door cannot be opened.
Display fails briefly The drive stops. The centrifuge decelerates
without braking.
The centrifuge door cannot be opened.
Pressing key has no effect. Centrifuge door is not correctly engaged or warped.
Exceptionally running noise Imbalance
No mains connection.
1. Is the centrifuge turned on?
2. Check the mains connection.
If the display remains dark, inform a service technician.
Mains connection interrupted for some seconds.
1. Turn off mains switch.
2. Check whether the mains power cord is connected properly.
3. Restart the centrifuge.Restart the centrifuge.
1. Check if mains connection is working and the instrument is switched on (display is lit).
2. If this is unsuccessful, you may open the centrifuge doorusing the mechanical emergency door release (“Mechanical
Emergency Door Release” on page 6-2)
1. Stop the centrifuge. Press the key or unplug mains powercord.
2. Wait until the centrifuge comes to a complete stop.
3. Check whether the rotor is properly loaded.
Check whether a broken tube, damage to the rotor or motor isresponsible for the run noise. If the running noise remains exceptionally, inform a service technician.
Display oP appears although lid is closed.
Lid Rotor stops with deceleration to standstill. Centrifuge door was opened manually during the run. Close
bAL Rotor stops with deceleration to standstill. Imbalance switch releases.
Thermo Scientific Heraeus Pico 17 / 21 6-3
Centrifuge does not start. Centrifuge door not properly closed. Open the centrifuge door
and repeat locking procedure. If the message appears again, inform a service technician.
centrifuge door immediately.
Rotor stops with deceleration to standstill.
For further centrifugation, you have to switch the instrument off and switch it on again
1. Open the centrifuge by pressing the key.
2. Check whether the rotor is properly loaded.
3. Check whether a broken tube or damaged rotor released the
imbalance switch.
If the message appears again, inform a service technician.
Page 40
6 Troubleshooting
Troubleshooting by User
Failure message Problem with centrifuge Possible causes and cures
E-01 - E-13 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
E-14 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
E-15 - E-16 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
E-22 - E-23 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
E-24 The centrifuge cannot be operated. The
door cannot be opened.
Internal program error
Switch the instrument off and on again.
If the message appears again, inform a service technician.
Over-temperature in the centrifuge chamber.
Switch the centrifuge off and turn it on again after approx. one minute.
If the message appears again, inform a service technician.
Temperature measurement error
Switch the instrument off and on again.
If the message appears again, inform a service technician.
Error in speed entry
Switch the instrument off and on again.
The display shows BR and a countdown from 100 - 0.
If the message appears again, inform a service technician.
Wrong status information from the door latch.
1. Switch the instrument off and on again.
2. After re-switching on, the display shows Lid FAiL.
3. If the centrifuge door has been already opened, the displayshows CLOSE Lid. Thereupon close the door.
4. The centrifuge tries to open the door to switch for starting thenormal operation mode.
If the message appears again, inform a service technician.
E-29 Motor does not start. Motor or rotor blocked
1. Switch the instrument off and on again using the mains switch.
2. Door Opening
3. Check whether the rotor can turn freely.
If the message appears again, inform a service technician.
E-31 Rotor stops without deceleration to
standstill or does not start.
Overtemperature in the motor.
1. Turn instrument off and unplug mains power cord.
2. Check and clean the venting slots if necessary and respectively the filter unit of the cooled centrifuge.
3. After approx. 60 minutes you can restart the instrument.
Observe the maximum permissible environmental temperature.
If the message appears again, inform a service technician.
6-4 Heraeus Pico 17 / 21 Thermo Scientific
Page 41
Failure message Problem with centrifuge Possible causes and cures
E-33 Rotor stops with deceleration to standstill. Overpressure in the refrigeration unit
1. Wait 15 minutes before switching off the centrifuge. The fan will operate at maximum power in order to reduce the overpressure.
2. Check the draft at the back of the centrifuge if you cannot here the fan.
If you do not feel any draft:
1. Turn instrument off and unplug mains power cord.
2. Check and clean the venting slots if necessary and respectively the filter unit of the cooled centrifuge.
3. After approx. 30 minutes you can restart the instrument.
Observe the maximum permissible environmental temperature.
If the message appears again, inform a service technician.
6 Troubleshooting
Troubleshooting by User
E-36 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
Over-current or over-voltage, the brake resistance the hot.
After approx. 60 minutes you can restart the instrument.
Switch the instrument off and on again.
If the message appears again, inform a service technician.
E-39 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
Error in speed entry.
Switch the instrument off and on again.
If the message appears again, inform a service technician.
E-41 - E-56 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
Internal program error
Switch the instrument off and on again.
If the message appears again, inform a service technician.
E-60 Rotor stops with deceleration to standstill. Under-temperature in the centrifuge chamber.
1. Stop the centrifugation run.
2. Open the centrifuge door and defrost the chamber. Never touch the chamber directly with your hands – you mayfreeze up.
3. After approx. 60 minutes you can restart the instrument.
Observe the maximum permissible ambient temperature.
4. If a strong ice sheet is present in the internal chamber, besure to remove all condensate after defrosting.
If the message appears again, inform a service technician.
E-63 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
Temperature measurement error
Switch the instrument off and on again.
If the message appears again, inform a service technician.
E-64 The cooling system could not be started. Power supply undervoltage
Check the power supply and contact you supplier if necessary. The message LINE appears in the display.
E-65 Rotor stops with deceleration to standstill.
The centrifuge cannot be operated.
Temperature measurement error
Switch the instrument off and on again.
If the message appears again, inform a service technician.
Thermo Scientific Heraeus Pico 17 / 21 6-5
Page 42
6 Troubleshooting
When to contact a Service Technician
When to contact a Service Technician
If you need to contact a service technician, please provide the order no. and the serial no. of your centrifuge. This information can be found on the back near the inlet for the power supply cable.
To identify the software version, proceed as follows:
1. Hold down the key and then switch on the centrifuge. In the display all segments are lit.
Subsequently, the following readings will be displayed for 5 seconds each:
Software number SOFT 058 3_
Software version _01
NV-RAM number EEPRO 462 1_
NV-RAM version _01
Cycle counter CYCLE 001 25
The values shown above are just examples.
Software 0583 Version 01
NV-RAM 4521 Version 01
cycles completed 125
2. Communicate the software version to the service technician.
6-6 Heraeus Pico 17 / 21 Thermo Scientific
Page 43
A
Chemical Compatibility Chart
CHEMICAL
2-mercaptoethanol
Acetaldehyde
Acetone
Acetonitrile
Alconox
Allyl Alcohol
Aluminum Chloride
Formic Acid (100 %)
Ammonium Acetate
Ammonium Carbonate
Ammonium Hydroxide (10 %)
Ammonium Hydroxide (28 %)
Ammonium Hydroxide (conc.)
Ammonium Phosphate
Ammonium Sulfate
Amyl Alcohol
Aniline
Sodium Hydroxide (<1 %)
Sodium Hydroxide (10 %)
Barium Salts
Benzene
Benzyl Alcohol
Boric Acid
Cesium Acetate
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINUM
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
SSU- SMS- SUSSUSS- SSSSUSSSSSS
S-UU---M-U---MUUUMM- MSU- S- U
MSUUSUMSSUUSUSUUUSSUUSMMSUU
SSU- SMS- SSUSUMUU- SMUUSSSSUU
UUS- SSS- SSSSSSMSSSSSSSSSSSU
---U--S----S-SSMSSS-MS--S--
UUSSSSUSSSSMSSSS- SSSSSMUUSS
-SMU--U----U-SMUUSS-US-US-U
SSU- SSS- SSSSSSSU- SSSSSSSSSS
MSUSSSSSSSSSSSUU- SSSSSSMSSS
UUSUSSMSSSSS- SUMSSSSSSSSSMS
UUSUSUMSSSSSUSUMSSSSSSSSSMS
UUUUSUMS- S- SUSUUSSS- MSSSS- U
U- S- SSSSSSSS- SSM- SSSSSSMSSS
UMS- SSUSSSSSSSSS- SSSSSSUSSU
S- MU- - SS- M- S- MSSSSM---U-S-M
SSUUSUSMSUUUUUUU- SMUUSSSSUS
U- MSSS- - SMSS- SMMSSSSSSMSS- U
U- MU- - U- MMSSUSUUSSSSSSMSS- U
MUS- SSSSSSSSSSSM- SSSSSSMSSS
SSUUSUMUSUUSUUUMUMUUUSUUS US
S- UU- - MM- M- SUUUUUUU- MSM- S- S
USSMSSUSSSSSSSSSUSSSSSSSSSS
M- S- SSS- SSSS- SS- - SSSSSSMSSS
, POLYCLEAR,CLEARCRIMP
1
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
VITON
Thermo Scientific Heraeus Pico 17 / 21 A-1
Page 44
A Chemical Compatibility Chart
CHEMICAL
Cesium Bromide
Cesium Chloride
Cesium Formate
Cesium Iodide
Cesium Sulfate
Chloroform
Chromic Acid (10 %)
MATERIAL
, POLYCLEAR,CLEARCRIMP
ALUMINUM
ANODIC COATING for ALUMINUM
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
1
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
VITON
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSSUSSS- SSSSSSS- - SSSSSSMSSS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
UUUUSSMUSUUMUMUUUMMUUSUUUMS
U- UUSUU- SSSUSSMUMSSUMSMUSSS
Chromic Acid (50 %)
Cresol Mixture
Cyclohexane
Deoxycholate
Distilled Water
Dextran
Diethyl Ether
Diethyl Ketone
Diethylpyrocarbonate
Dimethylsulfoxide
Dioxane
Ferric Chloride
Acetic Acid (Glacial)
Acetic Acid (5%)
Acetic Acid (60 %)
Ethyl Acetate
Ethyl Alcohol (50 %)
Ethyl Alcohol (95 %)
Ethylene Dichloride
Ethylene Glycol
Ethylene Oxide Vapor
Ficoll-Hypaque
Hydrofluoric Acid (10 %)
Hydrofluoric Acid (50 %)
Hydrochloric Acid (conc.)
U-UU-UU---SUUSMUMSSUMS-UM-S
SSU---S-SUUUUUU- - UU- USSSSUS
SSS- SSSUSUSSUUUMSMUMMSUMMUS
SSS- SSS- SSSSSSS- - SSSSSSSSSS
SSSSSSSSSSSSSSSSSSSSSSSSSSS
MSSSSSS- SSSSSSSSSSSSSSSMSSS
SSUUSSSUSUUSUUUUUUUUUSSSSMU
S- UU- - M- SU- S- MUUUMM- US- - SUU
SSU- SSS- SSUSUSU- - SSSMSSSSSS
SSUUSSS- SUSSUSUU- SSUUSSSSUU
MSUUSSMMSUUSUMUU- MMMUSSSSUU
UUS---MS-M-S-S---SS---MUS-S
SSUUSSUMSUSUUUUUMSUMUSUUS- U
SSMSSSMSSSSSMSSSSSSSMSSMSSM
S S U U S S U - S MS U U MU S MS MS MS MU S MU
MMUUSSMMSS USUMUU- SS UUSMMSUU
SSSSSSMSSSSSUSUSSSSSSSSMSMU
SSSUSSMSSSSSUSU- SSSMSSSUSMU
S- UU- - SM- UUSUUUUUUU- USU- S- S
SSSSSSSSSSSS- SUSSSSSSSSMSMS
S- U- - U- - SU- S- SM- - SSSUSUSSSU
MSS- SSS- SSSS- SS- SSSSSSSMSSS
UUUM- - U- - UUS- SMUSSSSMSUUU- -
UUUU- - U- - UUUUSUUUSS MMSUUU- M
UUUU- UUM- UMUUMUUU- S- USUUU- -
A-2 Heraeus Pico 17 / 21 Thermo Scientific
Page 45
CHEMICAL
Formaldehyde (40 %)
Glutaraldehyde
Glycerol
Guanidine Hydrochloride
Haemo-Sol
Hexane
A Chemical Compatibility Chart
, POLYCLEAR,CLEARCRIMP
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINUM
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
1
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
MMMSSSSMSSSSMSSSUSSMSSSMSMU
SSSS- - S- SSSSSSS- - SSS- - SSS- -
MSS- SSSSSSSSSSSS- SSSSSSSSSS
UUS- SSS- SSSSSSS- - SSSSSSUSSS
SSS---S-SSSSSSS- - SSSSSSSSSS
SSS- SSS- SSUSUMUSSUSSMSUSSUS
VITON
Isobutyl Alcohol
Isopropyl Alcohol
Iodoacetic Acid
Potassium Bromide
Potassium Carbonate
Potassium Chloride
Potassium Hydroxide (5 %)
Potassium Hydroxide (conc.)
Potassium Permanganate
Calcium Chloride
Calcium Hypochlorite
Kerosene
Sodium Chloride (10 %)
Sodium Chloride (sat'd)
Carbon Tetrachloride
Aqua Regia
Solution 555 (20 %)
Magnesium Chloride
Mercaptoacetic Acid
Methyl Alcohol
Methylene Chloride
Methyl Ethyl Ketone
Metrizamide
Lactic Acid (100 %)
Lactic Acid (20 %)
N-Butyl Alcohol
- - MU- - SS- U- SUSSMSSS- SSS- S- S
MMMU SSSSSUSSUSUMSSSSSSSMMMS
SSM- SSS- SMSSMSS- MSSSSSMSSMM
USS- SSS- SSSSSSSSSSS- SSSMSSS
MUSSSSS- SSSSSSUSSSSSSSSSSSS
USS- SSSSSSSSSSS- SSSSSSSUSSS
UUSSSSM- SSSS- SUSSSSSSSMUMSU
UUMU- - M- MSS- UMUUUSM- MU- UU- U
SSS- SSS- SSSUSSSM- SMSUSSMSUS
MUSSSSSSSSSSSSMS- SSSSSSMSSS
M- U- SMMS- M- S- SMS- SSSMSMUS- S
SSS- SSSUSMUSUMMS- MMMSSUSSUS
S- SSSSSS- - - SSSSS- SSSS- SSM- S
U-SUSSS----SSSSS- SS- S- SSM- S
U U MS S U MU S U U S U MU S S MMS MMMMU S S
U-UU--U-----UUUUUUU-----S-M
SSS---S-SSSSSSS- - SSS- SSSSSS
MSS- SSSSSSSSSSSSSSSSSSSMSSS
USU- SMS- SMSUUUU- SUUSMSUSSSS
SSSUSSMSSSSSUSUMSSSSSSSMSMU
UUUUMSSUSUUSUUUUUMUUUSSMUSU
SSUUSSMSSUUSUSUUUSSUUSSSSUU
MSS- SSS- SSSS- SS- - SSSSSSMSSS
--S------MSU-SSSMSS- MSMSS- S
--SS-----MSM-SSSSSSSMSMSS- S
S- SU- - S- - SM- USMSSSSMMSM- S- S
Thermo Scientific Heraeus Pico 17 / 21 A-3
Page 46
A Chemical Compatibility Chart
CHEMICAL
N-Butyl Phthalate
N, N-Dimethylformamide
Sodium Borate
Sodium Bromide
Sodium Carbonate (2 %)
Sodium Dodecyl Sulfate
Sodium Hypochlorite (5 %)
Sodium Iodide
Sodium Nitrate
Sodium Sulfate
Sodium Sulfide
Sodium Sulfite
Nickel Salts
Oils (Petroleum)
Oils (Other)
Oleic Acid
Oxalic Acid
Perchloric Acid (10 %)
Perchloric Acid (70 %)
Phenol (5 %)
Phenol (50 %)
Phosphoric Acid (10%)
Phosphoric Acid (conc.)
Physiologic Media (Serum, Urine)
Picric Acid
Pyridine (50 %)
Rubidium Bromide
MATERIAL
, POLYCLEAR,CLEARCRIMP
ALUMINUM
ANODIC COATING for ALUMINUM
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
1
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
VITON
SSU- SSS- SUUSUUUM- UUSUSMMSUS
SSSUSMS- SSUSUSUU- SSUUSMSSSU
MSSSSSSSSSSUSSSS- SSSSSSMSSS
USS- SSS- SSSSSSSS- SSSSSSMSSS
MUSSSSSSSSSSSSUSSSSSSSSSSSS
SSS- SSS- SSSSSSS- SSSSSSSSSSS
UUMSSMUSSMSSSMSSSSMSSSMUSMS
MSS- SSS- SSSSSSS- - SSSSSSMSSS
SSS- SSSSSSSSSSSS- SSSSSUSSSS
USS- SSSSSSSSSSSSSSSSSSSMSSS
S-SS---S---SSSUU- - S- - - SSM- S
SSS- SSSSMSSSSSSM- SSSSSSSSSS
USSSSS- SSS- - SSSS- SSSSSSMSSS
SSS---SUSSSSUUMSMUUSSSUSSSS
S-S---SMSSSSUSSSSUSSSS- SSMS
S- USSSUUSUSSMSSSSSSSSSMUSMM
UUMSSSUSSSSSUSUSSSSSSSSUMSS
U- U- SUU- SMM- - MUMSMM- MSU- S- S
UUU- - UU- SUMUUMUUUMMUMSUUSUS
U S U - S MM- S U MU U S U MS MS U U S U MMMS
USU- SUM- SUMUUUUUSUMUUSUUUMS
UUMSSSUSSSSU- SSSSSSSSSUMUSS
U U MM- - U S - MS U U MMS S S MS MS U MU - S
MSSS- - S- SSSSSSSSSSSSSSSSSSS
SSU- SMSSSMSUSSSUSSSSUSUMSMS
USUUSUU- USSUUMUU- USMUSSUUUU
MSS- SSS- SSSSSSS- - SSSSSSMSSS
Rubidium Chloride
Sucrose
Sucrose, Alkaline
Sulfosalicylic Acid
Nitric Acid (10 %)
MSS- SSS- SSSSSSS- - SSSSSSMSSS
MSS- SSSSSSSSSSSSSSSSSSSSSSS
MSS- SSS- SSSSSSUSSSSSSSSMSSS
UUSSSSS- SSSUSSS- SSS- SSSUSSS
USUSSUU- SUSU- SSSSSSSSSMSSSS
A-4 Heraeus Pico 17 / 21 Thermo Scientific
Page 47
CHEMICAL
Nitric Acid (50 %)
Nitric Acid (95 %)
Hydrochloric Acid (10 %)
Hydrochloric Acid (50 %)
Sulfuric Acid (10 %)
Sulfuric Acid (50 %)
Sulfuric Acid (conc.)
Stearic Acid
Tetrahydrofuran
Toluene
Trichloroacetic Acid
Trichloroethane
Trichloroethylene
Trisodium Phosphate
Tris Buffer (neutral pH)
Triton X-100
Urea
Hydrogen Peroxide (10 %)
Hydrogen Peroxide (3 %)
Xylene
Zinc Chloride
Zinc Sulfate
Citric Acid (10 %)
1
Polyethyleneterephthalate
A Chemical Compatibility Chart
, POLYCLEAR,CLEARCRIMP
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINUM
BUNA N
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
1
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
USUMSUU- SUSUUMMUMMMSSSUSSMS
U- UU- UU- - UUUUMUUUUMUUSUSS- S
UUMSSSU- SSSUUSUSSSSSSSSUMSS
UUUUSUU- SMSUUMUUSSSSMSMUUMM
MUUSSUU- SSMUSSSSSSSSSSUUUSS
MUUUSUU- SSMUUSUUMSSSSSUUUMS
MUUU- UUM- - MUUSUUUMSUMSUUU- S
S-S---SMSSSS- SSSSSSSSSMMSSS
SSUUSUUMSUUSUUU- MUUUUSUSSUU
SSUUSSMUSUUSUUUSUMUUUSUSUUM
UUU- SSUMSUSUUSM- MSSUUS UUUMU
S-U---MU-U-SUUUUUUUUUSU- S- S
--UU---U-U-SUUUUUUUUUSU- U- S
---S--M------S--SSS--S--S-S
USSSSSS- SSSSSSSSSSSSSSSSSSS
SSS- SSS- SSSSSSSSSSSSSSSSSSS
S- USSSS----SSSMSSSS- SSSMS- S
UUMSSUU- SSSUSSSMUSSSSSSMSUS
SMSSS- S- SSSSSSSSMSSSSSSSSSS
SSUSSSMUSUUUUUUMUMUUUSUMSUS
UUSSSSUSSSSSSSSSSSSSSSSUSSS
USS- SSSSSSSSSSSSSSSSSSSSSSS
MSSMSSMSSSSSSSSSMSSSSSSSSSS
VITON
Thermo Scientific Heraeus Pico 17 / 21 A-5
Page 48
A Chemical Compatibility Chart
Key
S Satisfactory
M Moderate attack, may be satisfactory for use in centrifuge depending on length of exposure, speed involved, etc.
Suggest testing under actual conditions of use.
U Unsatisfactory, not recommended.
-- Performance unknown; suggest testing, using sample to avoid loss of valuable material.
Chemical resistance data is included only as a guide to product use. No organized chemical resistance data exists for materials under the stress of centrifugation. When in doubt we recommend pretesting sample lots.
A-6 Heraeus Pico 17 / 21 Thermo Scientific
Page 49
B
Contact Information
United States / Canada 866-9-THERMO
Austria +43 1 801 400
Belgium +32 53 73 42 41
Germany 08001 536 376
France +33 2 2803 2180
Italy +39 02 95059 448
Netherlands +31 76 579 55 55
Nordic / Baltic Countries / CIS +358 9 329 10200
Russia +7 (812) 703 42 15
Spain / Portugal +34 932 23 09 18
Switzerland +41 44 454 12 12
+1 866 984 3766
+49 6184 90 6940
UK / Ireland +44 870 609 9203
China +86 21 6865 4588
+86 10 8419 3588
India 1800 22 8374
+91 22 6716 2200
Japan +81 45 453 9220
Other Asian Countries +852 2885 4613
Australia +61 39757 4300
New Zealand +64 9 980 6700
Other Countries +49 6184 90 6940
Thermo Scientific Heraeus Pico 17 / 21 B-1
Page 50
S
Index
A
Accident Prevention .......................................... iv
Aerosol-tight Rotors ........................................ 4-8
Audible Alarm .................................................4-8
Autoclaving ..................................................... 5-5
B
Before Setting up .............................................2-2
Before Use ....................................................... 2-1
C
Care ................................................................ 5-1
Centrifugation ................................................. 4-6
Characteristics ................................................. 1-2
Chemical Compatibility Chart ....................... A-1
Cleaning ..........................................................5-2
Cleaning Intervals ........................................... 5-2
Close Door ...................................................... 4-2
Contact Information .......................................B-1
Continuous Operation ....................................4-6
Control Panel ..................................................3-2
D
Decontamination ............................................ 5-4
Deposing ......................................................... 5-6
Directives ........................................................1-4
Disinfection .................................................... 5-3
G
Guidelines .......................................................1-4
I
Intended Use ..................................................... iv
K
Keys ................................................................ 3-2
L
Leveling ........................................................... 2-3
Lid Opening ....................................................4-2
Location ..........................................................2-2
M
Mains Connection ...........................................2-3
Mains Supply .................................................. 1-5
Maintenance ...................................................5-1
O
Operation ........................................................ 4-1
P
Precautions ......................................................... v
Preface ............................................................... iii
E
Emergency Door Release .................................6-2
End of Run - Alarm ........................................ 4-8
Entering Parameters ........................................ 4-4
Error - Alarm ..................................................4-8
Explanation of RCF-Value ..............................4-5
RCF-Value ......................................................4-4
Removing the Rotor ........................................4-8
Rotor Installation ............................................ 4-2
Rotor Selection ................................................1-5
S
R
F
Features ........................................................... 1-4
Functions ........................................................1-4
Thermo Scientific Heraeus Pico 17 / 21 S-1
Scope of Supply ................................................. iii
Service .............................................................5-5
Service Technician ..........................................6-6
Page 51
Index:
Shipping ..........................................................5-6
Short-term Centrifugation ..............................4-7
Speed .............................................................. 4-4
Standards ........................................................ 1-4
Storage ............................................................ 2-4
Switch on ........................................................4-2
T
Technical Data ................................................ 1-2
Transporting ................................................... 2-2
Troubleshooting .......................................6-1, 6-3
Turning off .....................................................4-9
S-2 Heraeus Pico 17 / 21 Thermo Scientific
Page 52
Loading...