Heraeus Holding GmbH licensed to Thermo Fisher Scientific. Delrin, TEFLON, and Viton are
registered trademarks of DuPont. Noryl is a registered trademark of SABIC. POLYCLEAR is a
registered trademark of Hongye CO., Ltd. Hypaque is a registered trademark of Amersham Health As.
RULON A and Tygon are registered trademarks of Saint-Gobain Performance Plastics. Alconox is a
registered trademark of Alconox. Ficoll is a registered trademark of GE Healthcare. Haemo-Sol is a
registered trademark of Haemo-Sol. Triton X-100 is a registered trademark of Sigma-Aldrich Co. LLC.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use
in the product operation. This document is copyright protected and any reproduction of the whole or
any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher
Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document
supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is c o mplete,
accurate or error-free and assumes no responsibility and will not be liable for any errors,
omissions, damage or loss that migh t result from any use of this document, even if the
information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a
purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which
Terms and Conditions of Sale shall govern all conflicting information between the two documents
ThermoFisher
SCIENTIFIC
KON FORM ITATSE RKLARU
DECLARATION
Name
für die Zusammenstellung der relevanten technischen
Name
representative
Anschrift
und
and address of the
des
to compile the
Herstellers
manufacturer and of the
CONFORMITY
OF
und
relevant technical documentation:
NG
des Bevollmächtigten
authorized
Thermo Electron LED GmbH
Zweigniederlassu n
g
Osterode
Am Kalkberg
37520 Osterode
Germany
Hiermit erklären wir,
Herewith
Besch rei bunq /descri otion
Marke /
we declare, that the
brand
dass die
Modellbezeichnunq / model name
Modell Nr. / model no.
Die Schutzziele nach Anhang 1 der Maschinenrichtlinie
2006/9 s/EG we rden ei nge h alte n.
protection goals
The
met.
Angewandte harmonisierte Normen/
Harmonized standards used:
EN 61010-1:2004
EN 61010-2-020: 2006
EN 61010-2-101:2002
EN 61326-1: 2006
EN 61326-2-6: 2006
EN 5501 18:2007
Osterode. den O9.O7 .2012
,'+r--d.aqrz
Andreas Karl,
Dr.
Director R&D
Datum, Name, Funktion und Unterschrift
Ort,
Place. date. name,
function
in
auch
r
Übereinstimmung
e I e
ktrom
of
n
ag
eti sc
with
Annex 1 for the directives
id
and
mit allen einschlägigen Bestimmungen
h
V
rt räg I i c h ke
e
e
all relevant terms
machinery 2006/42/EC
/
i
t.
of directives for
2006/42/EG
Richtlinie
der
electromagnetic compatibility
und der Niederspannungsrichtlinie
low voltage 2006/95/EC
and
are
WEEE Conformity
This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment
(WEEE) Directive 2002/96. It is marked by the following symbol:
Thermo Fisher Scientific has entered into agreements with recycling and disposal companies in all EU
Member States for the recycling and disposal of this device.
Mechanical Emergency Door Release ...................................................................... 6-2
Troubleshooting by User ......................................................................................... 6-3
When to contact Service Technician ........................................................................ 6-5
Chemical Compatibility Chart ...............................................................................A-1
Contact Information ................................................................................................B-1
iiHeraeus Fresco 17 / 21Thermo Scientific
P
Preface
Before starting to use the centrifuge, read through this instruction manual carefully and follow the
instructions.
The information contained in this instruction manual is the property of Thermo Fisher Scientific; it is
forbidden to copy or pass on this information without explicit approval.
Failure to follow the instructions and safety information in this instruction manual will result in the
expiration of the sellers warranty.
Scope of Supply
Article numberQuantityCheck
Centrifuge Heraeus Fresco 17 / 211
Power Supply Cable1
Rotor1
20360104Allen Wrench1
50133376CD with Instruction Manual1
If any parts are missing, please contact your nearest Thermo Fisher Scientific representative.
This symbol refers to general hazards.
WARNING means that injuries or material damage or contamination could occur.
CAUTION means that material damage could occur.
This symbol refers to biological hazards.
Observe the information contained in the instruction manual to keep yourself and your
environment safe.
This symbol refers to hot surfaces
Observe the information contained in the instruction manual to keep yourself and your
environment safe.
Thermo ScientificHeraeus Fresco 17 / 21iii
P
g
cm
3
---------- -
Remaining risk: Inproper use can cause damages, contamincation, and injuries with fatal consequences.
Intended Use
Intended Use
• This centrifuge is a laboratory product designed to separate components by generation of Relative
Centrifugal Force. It separates human samples (e.g. blood, urine and other body fluids) collected in
appropriate containers, either alone or after addition of reagents or other additives.
• As general purpose centrifuge, it is designed to also run other containers filled with chemicals,
environmental samples and other non-human body samples.
• Together with the hematocrit rotor 75003473 the centrifuge can be used as an in-vitro diagnostics
device to determine the haematicrit value by using centrifugation. Apply DIN 58933 to determine
the hematocit value.
• The centrifuge is to be used for separating materials of different density or particle size suspended
in a liquid.
Maximum sample density at maximum speed: 1.2
Accident Prevention
WARNING If a hazardious situation occurs, turn off the power supply to the centrifuge
and leave the area immediately.
WARNING In case of a rotor failure the centrifuge can be damanged. The cooling agent
can escape. Ventilate the room well and leave it. Inform Customer Service.
Prerequisite for the safe operation of the Heraeus Fresco 17 / 21 is a work environment in
compliance with standards, directives and trade association safety regulations and proper
instruction of the user.
The safety regulations contain the following basic recommendations:
• Maintain a radius of at least 30 cm around the centrifuge.
• Implementation of special measures which ensure that no one can approach the
centrifuge for longer than absolutely necessary while it is running.
• The mains plug must be freely accessible at all times. Pull out the power supply plug or
disconnect the power supply in an emergency.
ivHeraeus Fresco 17 / 21Thermo Scientific
Precautions
P
Precautions
In order to ensure safe operation of the Heraeus Fresco 17 / 21, the following general safety
regulations must be followed:
• The centrifuge should be operated by trained specialists only.
• The centrifuge is to be used for its intended use only.
• While handling centrifuge, rotor, and samples you must wear laboratory clothing (e.g.
gloves).
• Plug the centrifuge only into sockets which have been properly grounded.
• Do not move the centrifuge while it is running.
• Do not lean on the centrifuge.
• Use only rotors and accessories for this centrifuge which have been approved by Thermo
Fisher Scientific. Exceptions to this rule are commercially available glass or plastic
centrifuge tubes, provided they have been approved for the speed or the RCF value of
the rotor.
• Do not use rotors which show any signs of corrosion and/or cracks.
• Do not touch the mechanical components of the rotor and do not make any changes to
the mechanical components.
• Use only with rotors which have been properly installed. Follow the instructions in
section “Rotor Installation” on page 4-2.
• Use only with rotors which have been loaded properly. Follow the instructions given in
the rotor manual.
• Never overload the rotor. Follow the instructions given in the rotor manual.
• Never open the lid until the rotor has come to a complete stop and this has been
confirmed in the display.
• The lid emergency release may be used in emergencies only to recover the samples from
the centrifuge, e.g. during a power failure (see section “Mechanical Emergency Door
Release” on page 6-2).
• Never use the centrifuge if parts of its cover panels are damaged or missing.
• Do not touch the electronic components of the centrifuge or alter any electronic or
mechanical components.
• Please observe the safety instructions.
Thermo ScientificHeraeus Fresco 17 / 21v
P
Remaining risk: Inproper use can cause damages, contamincation, and injuries with fatal consequences.
Remaining risk: Inproper use can cause damages, contamincation, and injuries with fatal consequences.
Precautions
Please pay particular attention to the following aspects:
• Location: well-ventilated environment, set-up on a level and rigid surface with adequate
load-bearing capacity.
• Rotor installation:Check that the rotor is locked properly into place before operating the
centrifuge.
• Especially when working with corrosive samples (salt solutions, acids, bases), the
accessory parts and rotor chamber have to be cleaned carefully.
• Always balance the samples.
Centrifuging hazardous substances:
• Do not centrifuge explosive or flammable materials or substances which could react
violently with one another.
• The centrifuge is neither inert nor protected against explosion. Never use the centrifuge
in an explosion-prone environment.
• Do not centrifuge inflammable substances.
• Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms
without suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the
Biosafety Manual" of the World Health Organization (WHO)), aerosol-tight biological
seals have to be used.
For materials in a higher risk group, extra safety measures have to be taken.
• If toxins or pathogenic substances have gotten into the centrifuge or its parts,
appropriate disinfection measures have to be taken (see “Disinfection” on page 5-4).
• Highly corrosive substances which can cause material damage and impair the
mechanical stability of the rotor, should only be centrifuged in corresponding protective
tubes.
viHeraeus Fresco 17 / 21Thermo Scientific
1
Introduction and Description
Contents
• “Characteristics of the Heraeus Fresco 17 / 21” on page 1-2
• “Technical Data” on page 1-3
• “Directives, Standards and Guidelines” on page 1-4
• “Functions and Features” on page 1-5
• “Mains Supply” on page 1-5
• “Rotor Selection” on page 1-6
Thermo ScientificHeraeus Fresco 17 / 211-1
1
Introduction and Description
Characteristics of the Heraeus Fresco 17 / 21
Characteristics of the Heraeus Fresco 17 / 21
You can use various rotors with commercially available containers.
The set speed is reached within seconds.The maintenance-free induction motor ensures quiet and
low-vibration operation even at high speeds, and guarantees a very long lifetime.
The user-friendly control panel makes it easy to pre-set the speed, RCF value, running time, and
temperature. You can choose between the display of speed and RCF or the entry mode.
These settings can be changed even while the centrifuge is running.
With the help of the key you can also centrifuge a sample for just a few seconds, if called for.
The Heraeus Fresco 17 / 21 is equipped with various safety features:
• The housing is made of high-impact resistant plastic and the interior of armour steel.
• The lid is equipped with a view port and a lock.
• The lid of the centrifuge can only be opened while the centrifuge is switched on and the rotor has
come to a complete stop. The centrifuge cannot be started until the lid has been closed properly.
• An electronic imbalance recognition in ventilated centrifuges.
• For emergencies only, e.g. during power failures (see “Mechanical Emergency Door Release” on
page 6-2)
1-2Heraeus Fresco 17 / 21Thermo Scientific
Technical Data
The technical data of the Heraeus Fresco 17 / 21 is listed in the following table.
Table 1- 1. Technical Data Heraeus Fresco 17
FeatureValue
Environmental Conditions-Use in interior Spaces
1
Introduction and Description
Technical Data
-Altitudes of up to 2,000 m above Sea Level
-Max. relative Humidity 80% up to 31°C; Decreasing linearly up to
50% relative Humidity at 40°C
Noise Level at maximum Speed < 50 dB (A)< 50 dB (A)< 50 dB (A)
Dimensions
Hight
Width
Depth
230 mm
240 mm
350 mm
Weight without Rotor10.5 kg10.5 kg10.5 kg
Directives, Standards and Guidelines
Table 1-3. Directives, Standards and Guidelines
Tension /
Frequency
230V
50Hz/60Hz,
Marked
EC:
98/79/EC In-vitro-Diagnostica (IvD)
2006/42/EC Machine Directive
2004/108/EC EMC Directive
2006/95/EC Low Voltage Directive
100V 50HzcCSAus:UL Std. No. UL 61010-1, 2
Produced and inspected according to the following standards
and guidelines
EN 61010-1, 2
EN 61010-2-020, 2
nd
Edition
nd
Edition
EN 61010-2-101
EN 61326-1
EN 61326-2-6
EN 55011B
EN 61000-6-2
EN ISO 13485
CAN/CSA-C22.2 No. 61010-1-04
CAN/CSA-C22.2 No. 61010-2-020-09 Part 2-020:
IEC 61010-2-020 2
nd
Edition
(Pollution degree 2, Overvoltage category II)
IEC 61010-2-101
14 800 rpm
(Depending on
Rotor)
230 mm
240 mm
350 mm
nd
Edition
14 800 rpm
(Depending on
Rotor)
230 mm
240 mm
350 mm
1-4Heraeus Fresco 17 / 21Thermo Scientific
1
Introduction and Description
Functions and Features
Functions and Features
The following table contains an overview of the electrical connection data for the Heraeus Fresco 17 /
21.
Table 1-4 . Functions and Features
Component / FunctionDescription / Features
Structure / HousingSheet steel chassis with attached plastic housing and steel
reinforcement
Rotor ChamberPlastic
Up to 48 ml of spilled liquids are retained in the chamber and cannot
enter the instrument.
DriveInduction drive without carbon brushes
Keys and DisplayEasy-to-clean keypad and display surface
Controls Microprocessor-controlled
Internal MemoryMost recent data is saved
FunctionsRCF-selection, temperature control, pretemp
Imbalance RecognitionElectronic in ventilated centrifuges
Door LockAutomatic door closing and locking starting from an initial hold
position
Mains Supply
The following table contains an overview of the electrical connection data for the der Heraeus Fresco
17 / 21.This data is to be taken into consideration when selecting the mains connection socket.
Table 1-5 . Electrical connection data
Cat.
75002420FRESCO 17230 V 50 / 60 Hz1.9A330 W4A, 2-pole circuit breaker16 AT
75002421FRESCO 17120 V 60 Hz2.2 A330 W8A, 2-pole circuit breaker15 AT
75002422FRESCO 17100 V 60 Hz2.4 A330 W8A, 2-pole circuit breaker16 AT
75002425Fresco 21230 V 50 / 60 Hz2.2 A370 W4A, 2-pole circuit breaker16 AT
75002421Fresco 26120 V 60 Hz4.4 A370 W8A, 2-pole circuit breaker15 AT
75002422Fresco 27100 V 60 Hz5.1 A360 W8A, 2-pole circuit breaker16 AT
Mains
voltage
Frequency
Rated
current
Power
consumption
Equipment fuse Building fuse
Thermo ScientificHeraeus Fresco 17 / 211-5
1
Introduction and Description
Rotor Selection
Rotor Selection
The Heraeus Fresco 17 / 21 is supplied with a rotor
Various rotors are available to choose from.
Microliter 24 x 27500 3424
Microliter 36 x 0.57500 3436
Dual Rotor 18x2 / 0.5 ml7500 3418
PCR Rotor 4x87500 3440
PCR Rotor 8x87500 3489
Hematocrit Rotor7500 3473
The technical data of the rotors and the corresponding adapters and reduction sleeves for various
commercially available containers can be found in the corresponding rotor operating manuals.
For more information visit our website at: www.thermoscientific.com/centrifuge.
1-6Heraeus Fresco 17 / 21Thermo Scientific
2
Before Use
Contents
• “Before Setting up” on page 2-2
• “Transporting the Centrifuge” on page 2-2
• “Location” on page 2-2
• “Aligning the Centrifuge” on page 2-3
• “Mains Connection” on page 2-3
• “Storage” on page 2-4
Thermo ScientificHeraeus Fresco 17 / 212-1