Thermo Heraeus Fresco 17, Heraeus Fresco 21 User manual

Thermo Scientific
Heraeus Fresco 17 / 21
Instruction Manual
50133376-2 August 2012
Visit us online to register your warranty.
www.thermoscientific.com/labwarranty
Please reference this page for the most up-to-date information on the
following*
• web site addresses
• contact information
• copyright and trademark information
*All subsequent pages in this manual may have incorrect web site addresses and contact information.
©2013 Thermo Fisher Scientific Inc. All rights reserved. HERAEUS is a registered trademark of
Heraeus Holding GmbH licensed to Thermo Fisher Scientific. Delrin, TEFLON, and Viton are
registered trademarks of DuPont. Noryl is a registered trademark of SABIC. POLYCLEAR is a registered trademark of Hongye CO., Ltd. Hypaque is a registered trademark of Amersham Health As. RULON A and Tygon are registered trademarks of Saint-Gobain Performance Plastics. Alconox is a registered trademark of Alconox. Ficoll is a registered trademark of GE Healthcare. Haemo-Sol is a registered trademark of Haemo-Sol. Triton X-100 is a registered trademark of Sigma-Aldrich Co. LLC. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Manufacturer: Thermo Fisher Scientific Robert-Bosch-Straße 1 D - 63505 Langenselbold Germany
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the product operation. This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this document is for reference purposes only. System configurations and specifications in this document supersede all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is c o mplete, accurate or error-free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that migh t result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall govern all conflicting information between the two documents
ThermoFisher
SCIENTIFIC
KON FORM ITATSE RKLARU
DECLARATION
Name
für die Zusammenstellung der relevanten technischen
Name
representative
Anschrift
und
and address of the
des
to compile the
Herstellers
manufacturer and of the
CONFORMITY
OF
und
relevant technical documentation:
NG
des Bevollmächtigten
authorized
Thermo Electron LED GmbH Zweigniederlassu n
g
Osterode
Am Kalkberg
37520 Osterode
Germany
Hiermit erklären wir, Herewith
Besch rei bunq /descri otion
Marke /
we declare, that the
brand
dass die
Modellbezeichnunq / model name Modell Nr. / model no.
Gültig ab Equipmentnr.
Valid
from equipment no.
mit
allen einschlägigen Bestimmungen der
Übereinstimmung ist.
in
is
conformity with all relevant terms of directive
nachstehend beschriebene
machinery
described
Richtlinie über ln-vitro-Diagnostika
below
: Labor-Zentrifuqe / centrifuqe : Heraeus : FRESCO 17, FRESCO 21 : 75002420 75002423 75002425
: 41385539
for in vitro diagnostic medical
Unterlagen:
Maschine
(€
98/79/EG
devices 98n9/EC.
in
Die Maschine ist 2004n
0 8/EG ü be
The machinery is in accordance 2004/108/EC.
Die Schutzziele nach Anhang 1 der Maschinenrichtlinie
2006/9 s/EG we rden ei nge h alte n.
protection goals
The
met.
Angewandte harmonisierte Normen/
Harmonized standards used:
EN 61010-1:2004 EN 61010-2-020: 2006
EN 61010-2-101:2002 EN 61326-1: 2006 EN 61326-2-6: 2006 EN 5501 18:2007
Osterode. den O9.O7 .2012
,'+r--d.aqrz
Andreas Karl,
Dr.
Director R&D
Datum, Name, Funktion und Unterschrift
Ort, Place. date. name,
function
in
auch
r
Übereinstimmung
e I e
ktrom
of
n
ag
eti sc
with
Annex 1 for the directives
id
and
mit allen einschlägigen Bestimmungen
h
V
rt räg I i c h ke
e
e
all relevant terms
machinery 2006/42/EC
/
i
t.
of directives for
2006/42/EG
Richtlinie
der
electromagnetic compatibility
und der Niederspannungsrichtlinie
low voltage 2006/95/EC
and
are
WEEE Conformity
This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96. It is marked by the following symbol:
Thermo Fisher Scientific has entered into agreements with recycling and disposal companies in all EU Member States for the recycling and disposal of this device.
T
Table of Contents
Preface .......................................................................................................................................iii
Scope of Supply ....................................................................................................................... iii
Intended Use ..............................................................................................................................iv
Accident Prevention .................................................................................................................iv
Precautions .................................................................................................................................v
Chapter 1 Introduction and Description ...............................................................................................1-1
Characteristics of the Heraeus Fresco 17 / 21 ........................................................... 1-2
Technical Data ........................................................................................................ 1-3
Directives, Standards and Guidelines ...................................................................................... 1-4
Functions and Features ............................................................................................................ 1-5
Mains Supply ........................................................................................................... 1-5
Rotor Selection ........................................................................................................ 1-6
Chapter 2 Before Use ...............................................................................................................................2-1
Before Setting up ..................................................................................................... 2-2
Transporting the Centrifuge .................................................................................... 2-2
Location .................................................................................................................. 2-2
Aligning the Centrifuge ........................................................................................... 2-3
Mains Connection ................................................................................................... 2-3
Storage ..................................................................................................................... 2-4
Chapter 3 Control Panel ..........................................................................................................................3-1
Control Panel .......................................................................................................... 3-2
Keys ......................................................................................................................... 3-2
Chapter 4 Operation .................................................................................................................................4-1
Switch on Centrifuge ............................................................................................... 4-2
Door Opening ......................................................................................................... 4-2
Close Door .............................................................................................................. 4-2
Rotor Installation ..................................................................................................... 4-2
Entering Parameters ................................................................................................. 4-4
Centrifugation ......................................................................................................... 4-8
Short-term Centrifugation ....................................................................................... 4-8
Removing the Rotor ................................................................................................ 4-9
Aerosol-tight Rotors ................................................................................................. 4-9
Audible Alarm ......................................................................................................... 4-9
Turning off the Centrifuge .................................................................................... 4-10
Chapter 5 Maintenance and Care .........................................................................................................5-1
Cleaning Intervals .................................................................................................... 5-2
Cleaning .................................................................................................................. 5-2
Disinfection ............................................................................................................. 5-4
Thermo Scientific Heraeus Fresco 17 / 21 i
Contents
Decontamination ..................................................................................................... 5-5
Autoclaving .............................................................................................................. 5-5
Service of Thermo Fisher Scientific .......................................................................... 5-6
Shipping and Deposing of Accessories ..................................................................... 5-6
Chapter 6 Troubleshooting ....................................................................................................................6-1
Mechanical Emergency Door Release ...................................................................... 6-2
Troubleshooting by User ......................................................................................... 6-3
When to contact Service Technician ........................................................................ 6-5
Chemical Compatibility Chart ...............................................................................A-1
Contact Information ................................................................................................B-1
ii Heraeus Fresco 17 / 21 Thermo Scientific
P
Preface
Before starting to use the centrifuge, read through this instruction manual carefully and follow the instructions.
The information contained in this instruction manual is the property of Thermo Fisher Scientific; it is forbidden to copy or pass on this information without explicit approval.
Failure to follow the instructions and safety information in this instruction manual will result in the expiration of the sellers warranty.
Scope of Supply
Article number Quantity Check
Centrifuge Heraeus Fresco 17 / 21 1
Power Supply Cable 1
Rotor 1
20360104 Allen Wrench 1
50133376 CD with Instruction Manual 1
If any parts are missing, please contact your nearest Thermo Fisher Scientific representative.
This symbol refers to general hazards. WARNING means that injuries or material damage or contamination could occur. CAUTION means that material damage could occur.
This symbol refers to biological hazards. Observe the information contained in the instruction manual to keep yourself and your environment safe.
This symbol refers to hot surfaces Observe the information contained in the instruction manual to keep yourself and your environment safe.
Thermo Scientific Heraeus Fresco 17 / 21 iii
P
g
cm
3
---------- -
Remaining risk: Inproper use can cause damages, contamincation, and inju­ries with fatal consequences.
Intended Use
Intended Use
This centrifuge is a laboratory product designed to separate components by generation of Relative
Centrifugal Force. It separates human samples (e.g. blood, urine and other body fluids) collected in appropriate containers, either alone or after addition of reagents or other additives.
As general purpose centrifuge, it is designed to also run other containers filled with chemicals,
environmental samples and other non-human body samples.
Together with the hematocrit rotor 75003473 the centrifuge can be used as an in-vitro diagnostics
device to determine the haematicrit value by using centrifugation. Apply DIN 58933 to determine the hematocit value.
The centrifuge is to be used for separating materials of different density or particle size suspended
in a liquid.
Maximum sample density at maximum speed: 1.2
Accident Prevention
WARNING If a hazardious situation occurs, turn off the power supply to the centrifuge
and leave the area immediately.
WARNING In case of a rotor failure the centrifuge can be damanged. The cooling agent can escape. Ventilate the room well and leave it. Inform Customer Service.
Prerequisite for the safe operation of the Heraeus Fresco 17 / 21 is a work environment in compliance with standards, directives and trade association safety regulations and proper instruction of the user.
The safety regulations contain the following basic recommendations:
Maintain a radius of at least 30 cm around the centrifuge.
Implementation of special measures which ensure that no one can approach the
centrifuge for longer than absolutely necessary while it is running.
The mains plug must be freely accessible at all times. Pull out the power supply plug or
disconnect the power supply in an emergency.
iv Heraeus Fresco 17 / 21 Thermo Scientific
Precautions
P
Precautions
In order to ensure safe operation of the Heraeus Fresco 17 / 21, the following general safety regulations must be followed:
The centrifuge should be operated by trained specialists only.
The centrifuge is to be used for its intended use only.
While handling centrifuge, rotor, and samples you must wear laboratory clothing (e.g.
gloves).
Plug the centrifuge only into sockets which have been properly grounded.
Do not move the centrifuge while it is running.
Do not lean on the centrifuge.
Use only rotors and accessories for this centrifuge which have been approved by Thermo
Fisher Scientific. Exceptions to this rule are commercially available glass or plastic centrifuge tubes, provided they have been approved for the speed or the RCF value of the rotor.
Do not use rotors which show any signs of corrosion and/or cracks.
Do not touch the mechanical components of the rotor and do not make any changes to
the mechanical components.
Use only with rotors which have been properly installed. Follow the instructions in
section “Rotor Installation” on page 4-2.
Use only with rotors which have been loaded properly. Follow the instructions given in
the rotor manual.
Never overload the rotor. Follow the instructions given in the rotor manual.
Never open the lid until the rotor has come to a complete stop and this has been
confirmed in the display.
The lid emergency release may be used in emergencies only to recover the samples from
the centrifuge, e.g. during a power failure (see section “Mechanical Emergency Door
Release” on page 6-2).
Never use the centrifuge if parts of its cover panels are damaged or missing.
Do not touch the electronic components of the centrifuge or alter any electronic or
mechanical components.
Please observe the safety instructions.
Thermo Scientific Heraeus Fresco 17 / 21 v
P
Remaining risk: Inproper use can cause damages, contamincation, and inju­ries with fatal consequences.
Remaining risk: Inproper use can cause damages, contamincation, and inju­ries with fatal consequences.
Precautions
Please pay particular attention to the following aspects:
Location: well-ventilated environment, set-up on a level and rigid surface with adequate
load-bearing capacity.
Rotor installation:Check that the rotor is locked properly into place before operating the
centrifuge.
Especially when working with corrosive samples (salt solutions, acids, bases), the
accessory parts and rotor chamber have to be cleaned carefully.
Always balance the samples.
Centrifuging hazardous substances:
Do not centrifuge explosive or flammable materials or substances which could react
violently with one another.
The centrifuge is neither inert nor protected against explosion. Never use the centrifuge
in an explosion-prone environment.
Do not centrifuge inflammable substances.
Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms
without suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the Biosafety Manual" of the World Health Organization (WHO)), aerosol-tight biological seals have to be used.
For materials in a higher risk group, extra safety measures have to be taken.
If toxins or pathogenic substances have gotten into the centrifuge or its parts,
appropriate disinfection measures have to be taken (see “Disinfection” on page 5-4).
Highly corrosive substances which can cause material damage and impair the
mechanical stability of the rotor, should only be centrifuged in corresponding protective tubes.
vi Heraeus Fresco 17 / 21 Thermo Scientific
1
Introduction and Description
Contents
“Characteristics of the Heraeus Fresco 17 / 21” on page 1-2
“Technical Data” on page 1-3
“Directives, Standards and Guidelines” on page 1-4
“Functions and Features” on page 1-5
“Mains Supply” on page 1-5
“Rotor Selection” on page 1-6
Thermo Scientific Heraeus Fresco 17 / 21 1-1
1
Introduction and Description
Characteristics of the Heraeus Fresco 17 / 21
Characteristics of the Heraeus Fresco 17 / 21
You can use various rotors with commercially available containers.
The set speed is reached within seconds. The maintenance-free induction motor ensures quiet and low-vibration operation even at high speeds, and guarantees a very long lifetime.
The user-friendly control panel makes it easy to pre-set the speed, RCF value, running time, and temperature. You can choose between the display of speed and RCF or the entry mode.
These settings can be changed even while the centrifuge is running.
With the help of the key you can also centrifuge a sample for just a few seconds, if called for.
The Heraeus Fresco 17 / 21 is equipped with various safety features:
The housing is made of high-impact resistant plastic and the interior of armour steel.
The lid is equipped with a view port and a lock.
The lid of the centrifuge can only be opened while the centrifuge is switched on and the rotor has
come to a complete stop. The centrifuge cannot be started until the lid has been closed properly.
An electronic imbalance recognition in ventilated centrifuges.
For emergencies only, e.g. during power failures (see “Mechanical Emergency Door Release” on
page 6-2)
1-2 Heraeus Fresco 17 / 21 Thermo Scientific
Technical Data
The technical data of the Heraeus Fresco 17 / 21 is listed in the following table.
Table 1- 1. Technical Data Heraeus Fresco 17
Feature Value
Environmental Conditions -Use in interior Spaces
1
Introduction and Description
Technical Data
-Altitudes of up to 2,000 m above Sea Level
-Max. relative Humidity 80% up to 31°C; Decreasing linearly up to 50% relative Humidity at 40°C
Permissible Ambient Temperature during
+5°C bis +40°C +5°C bis +40°C +5°C bis +40°C
Operation
Permissible Ambient Temperature during
+2°C bis +50°C +2°C bis +50°C +2°C bis +50°C
Storage and Shipping
Overvoltage Category II II II
Pollution Degree 2 2 2
Heat Dissipation Heraeus Fresco 17 Heraeus Fresco 17 Heraeus Fresco 17
230 V 120 V 100 V
614 BTU/h 614 BTU/h 580 BTU/h
IP 20 20 20
Run Time unlimited unlimited unlimited
Max Speed n
Min Speed n
max
min
Maximum RCF Value at n
max
13 300 rpm (Depending on Rotor)
13 300 rpm (Depending on Rotor)
13 300 rpm (Depending on Rotor)
300 rpm 300 rpm 300 rpm
17 000 x g 17 000 x g 17 000 x g
Maximum Kinetic Energy <1,90 kNm <1,90 kNm <1,90 kNm
Noise Level at maximum Speed < 56 dB (A) < 56 dB (A) < 56 dB (A)
Dimensions
Hight Width Depth
230 mm 240 mm 350 mm
230 mm 240 mm 350 mm
230 mm 240 mm 350 mm
Weight without Rotor 10.5 kg 10.5 kg 10.5 kg
Table 1- 2. Technical Data Heraeus Fresco 21
Feature Value
Environmental Conditions -Use in interior Spaces
-Altitudes of up to 2,000 m above Sea Level
-Max. relative Humidity 80% up to 31°C; Decreasing linearly up to 50% relative Humidity at 40°C
Permissible Ambient Temperature during Operation
Permissible Ambient Temperature during Storage and Shipping
Thermo Scientific Heraeus Fresco 17 / 21 1-3
+5°C bis +40°C +5°C bis +40°C +5°C bis +40°C
+2°C bis +50°C +2°C bis +50°C +2°C bis +50°C
1
Introduction and Description
Directives, Standards and Guidelines
Table 1-2. Technical Data Heraeus Fresco 21
Feature Value
Overvoltage Category II II II
Pollution Degree 2 2 2
Heat Dissipation Heraeus Fresco 21 Heraeus Fresco 21 Heraeus Fresco 21
IP 20 20 20
Run Time unlimited unlimited unlimited
230 V 120 V 100 V
785 BTU/h 751 BTU/h 785 BTU/h
Max Speed n
max
14 800 rpm (Depending on Rotor)
Min Speed n
min
Maximum RCF Value at n
max
300 rpm 300 rpm 300 rpm
21 100 x g 21 100 x g 21 100 x g
Maximum Kinetic Energy <2,35 kNm <2,35 kNm <2,35 kNm
Noise Level at maximum Speed < 50 dB (A) < 50 dB (A) < 50 dB (A)
Dimensions
Hight Width Depth
230 mm 240 mm 350 mm
Weight without Rotor 10.5 kg 10.5 kg 10.5 kg
Directives, Standards and Guidelines
Table 1-3. Directives, Standards and Guidelines
Tension / Frequency
230V 50Hz/60Hz,
Marked
EC:
98/79/EC In-vitro-Diagnostica (IvD)
2006/42/EC Machine Directive
2004/108/EC EMC Directive
2006/95/EC Low Voltage Directive
100V 50Hz cCSAus: UL Std. No. UL 61010-1, 2
Produced and inspected according to the following standards and guidelines
EN 61010-1, 2
EN 61010-2-020, 2
nd
Edition
nd
Edition
EN 61010-2-101
EN 61326-1
EN 61326-2-6
EN 55011B
EN 61000-6-2
EN ISO 13485
CAN/CSA-C22.2 No. 61010-1-04
CAN/CSA-C22.2 No. 61010-2-020-09 Part 2-020:
IEC 61010-2-020 2
nd
Edition
(Pollution degree 2, Overvoltage category II)
IEC 61010-2-101
14 800 rpm (Depending on Rotor)
230 mm 240 mm 350 mm
nd
Edition
14 800 rpm (Depending on Rotor)
230 mm 240 mm 350 mm
1-4 Heraeus Fresco 17 / 21 Thermo Scientific
1
Introduction and Description
Functions and Features
Functions and Features
The following table contains an overview of the electrical connection data for the Heraeus Fresco 17 /
21.
Table 1-4 . Functions and Features
Component / Function Description / Features
Structure / Housing Sheet steel chassis with attached plastic housing and steel
reinforcement
Rotor Chamber Plastic
Up to 48 ml of spilled liquids are retained in the chamber and cannot enter the instrument.
Drive Induction drive without carbon brushes
Keys and Display Easy-to-clean keypad and display surface
Controls Microprocessor-controlled
Internal Memory Most recent data is saved
Functions RCF-selection, temperature control, pretemp
Imbalance Recognition Electronic in ventilated centrifuges
Door Lock Automatic door closing and locking starting from an initial hold
position
Mains Supply
The following table contains an overview of the electrical connection data for the der Heraeus Fresco 17 / 21.This data is to be taken into consideration when selecting the mains connection socket.
Table 1-5 . Electrical connection data
Cat.
75002420 FRESCO 17 230 V 50 / 60 Hz 1.9A 330 W 4A, 2-pole circuit breaker 16 AT
75002421 FRESCO 17 120 V 60 Hz 2.2 A 330 W 8A, 2-pole circuit breaker 15 AT
75002422 FRESCO 17 100 V 60 Hz 2.4 A 330 W 8A, 2-pole circuit breaker 16 AT
75002425 Fresco 21 230 V 50 / 60 Hz 2.2 A 370 W 4A, 2-pole circuit breaker 16 AT
75002421 Fresco 26 120 V 60 Hz 4.4 A 370 W 8A, 2-pole circuit breaker 15 AT
75002422 Fresco 27 100 V 60 Hz 5.1 A 360 W 8A, 2-pole circuit breaker 16 AT
Mains voltage
Frequency
Rated current
Power consumption
Equipment fuse Building fuse
Thermo Scientific Heraeus Fresco 17 / 21 1-5
1
Introduction and Description
Rotor Selection
Rotor Selection
The Heraeus Fresco 17 / 21 is supplied with a rotor
Various rotors are available to choose from.
Microliter 24 x 2 7500 3424
Microliter 36 x 0.5 7500 3436
Dual Rotor 18x2 / 0.5 ml 7500 3418
PCR Rotor 4x8 7500 3440
PCR Rotor 8x8 7500 3489
Hematocrit Rotor 7500 3473
The technical data of the rotors and the corresponding adapters and reduction sleeves for various commercially available containers can be found in the corresponding rotor operating manuals.
For more information visit our website at: www.thermoscientific.com/centrifuge.
1-6 Heraeus Fresco 17 / 21 Thermo Scientific
2
Before Use
Contents
“Before Setting up” on page 2-2
“Transporting the Centrifuge” on page 2-2
“Location” on page 2-2
“Aligning the Centrifuge” on page 2-3
“Mains Connection” on page 2-3
“Storage” on page 2-4
Thermo Scientific Heraeus Fresco 17 / 21 2-1