HERAEUS is a registered trademark of Heraeus Holding GmbH licensed to Thermo Fisher Scientific. Delrin,
TEFLON, and Viton are registered trademarks of DuPont. Noryl is a registered trademark of SABIC.
POLYCLEAR is a registered trademark of Hongye CO., Ltd. Hypaque is a registered trademark of Amersham
Health As. RULON A and Tygon are registered trademarks of Saint-Gobain Performance Plastics. Alconox is a
registered trademark of Alconox. Ficoll is a registered trademark of GE Healthcare. Haemo-Sol is a registered
trademark of Haemo-Sol. Triton X-100 is a registered trademark of Sigma-Aldrich Co. LLC
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
Manufacturer:
Thermo Electron LED GmbH
Robert-Bosch-Straße 1
D - 63505 Langenselbold
Germany
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or
errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss
that might result from any use of this document, even if the information in the document is followed
properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Original instruction manual 50139071-a printed in February 2013.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment
(WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European
Union (EU) Member State, and these companies should dispose of or recycle this product. See
www.thermo.com/rohsweee
Directives and the recyclers in your country.
for further information on Thermo Fisher Scientific’s compliance with these
Before starting to use the rotor, read through these instruction manual carefully and follow the
instructions.
Failure to follow the instructions and safety information in this instruction manual will result in the
expiration of the seller’s warranty.
Items Supplied
Hematocrit Rotor1
Anti-corrosion Oil1
Linear scale1
QuantityCheck
1
Preface
Instruction manual1
If any parts are missing, please contact the nearest Thermo Fisher Scientific representative.
Intended Use
This rotor is used in combination with the according centrifuge as a laboratory product designed to
separate components by generation of Relative Centrifugal Force. It separates human samples (e.g.
blood, urine and other body fluids) collected in appropriate containers, either alone or after addition of
reagents or other additives.
The rotor in the centrifuge is designed to run other containers filled with chemicals, environmental
samples and other non-human body samples.
If the rotor is used in a manner not specified by the manufacturer, the protection provided by the
equipment may be impaired.
Together with this hematocrit rotor the centrifuge becomes an in-vitro diagnostics device to define the
hematocrit value by centrifugation.
Apply DIN 58933 to define the hematocrit value. To obtain reproducible results, standardization is
required both in interest of patients and blood donors and to guarantee the comparability of research
results. The hematocrit value allows you to calculate the rate of erythrocytes in the blood. This applies
only if the Hematocrit Rotor is used with approved accessories and items listed in the items supplied.
This rotor should be operated by trained specialists only.
Thermo ScientificHematocrit Rotor7
1
g
cm
3
---------- -
Precautions
Precautions
In order to ensure safe operation of the Hematocrit Rotor, the following general safety regulations
must be followed:
• Do not remove the magnet at the rotor bottom
• Do not use rotors which show any signs of corrosion and/or cracks. Do not touch the electronic
• Use only with rotors which have been loaded properly.
• Never overload the rotor.
• Use only accessories which have been approved by Thermo Fisher Scientific. Exceptions to this
• Observe the safety instructions.
components of the rotor and do not make any changes to the electronic or mechanical
components.
rule are commercially available glass or plastic centrifuge tubes, provided they have been approved
for the speed or the RCF value of the rotor.
Pay particular attention to the following aspects:
• Rotor installation: Check that the rotor is locked properly into place before operating the
centrifuge.
• Always balance the samples.
Maximum sample density at maximum speed: 1,2
This symbol refers to general hazards.
CAUTION means that material damage could occur.
WARNING means that injuries or material damage or contamination could occur.
This symbol refers to biological hazards.
Observe the information contained in the instruction manual to keep yourself and your
environment safe.
8Hematocrit RotorThermo Scientific
Rotor Specifications
Contents
• “Thermo Scientific Heraeus Megafuge 8 Centrifuge” on page 10
• “Thermo Scientific Sorvall ST 8 Centrifuge” on page 11
• “Thermo Scientific SL 8 Centrifuge” on page 13
2
Rotor Specifications
Thermo ScientificHematocrit Rotor9
2
Rotor Specifications
Thermo Scientific Heraeus Megafuge 8 Centrifuge
Thermo Scientific Heraeus Megafuge 8 Centrifuge
Table 1. Heraeus 230V 50Hz/60Hz
CentrifugeHeraeus® Megafuge® 8
Catalog #75007211
Weight empty [kg]1.0
Maximum permissible load [g]24 x 0.2
Maximum speed n
Maximum RCF value at n
[rpm]13300
max
16810
max
Max. / min. radius [cm]8.5 / 2.0
Pitch angle [°]90
Accel. / braking time [s]14 / 24
Sample heating at n
temperature of 23 °C, running time 60 minutes
Aerosol-tight
1
[° C] referred to ambient
max
16
No
Permissible temperature range autoclavable °C134
1
Tested by HPA, Porton-down, UK
Table 2. Heraeus 120V 60Hz
CentrifugeHeraeus® Megafuge® 8
Catalog #75007210
Weight empty [kg]1.0
Maximum permissible load [g]24 x 0.2
Maximum speed n
Maximum RCF value at n
[rpm]13300
max
16810
max
Max. / min. radius [cm]8.5 / 2.0
Pitch angle [°]90
Accel. / braking time [s]14 / 24
Sample heating at n
temperature of 23 °C, running time 60 minutes
Aerosol-tight
1
[° C] referred to ambient
max
16
No
Permissible temperature range autoclavable °C134
1
Tested by HPA, Porton-down, UK
10Hematocrit RotorThermo Scientific
Thermo Scientific Sorvall ST 8 Centrifuge
Table 3.Sorvall 230V 50/60Hz
CentrifugeSorvall® ST 8
Catalog #75007201
Weight empty [kg]1.0
Maximum permissible load [g]24 x 0.2
2
Rotor Specifications
Thermo Scientific Sorvall ST 8 Centrifuge
Maximum speed n
Maximum RCF value at n
[rpm]13300
max
16810
max
Max. / min. radius [cm]8.5 / 2.0
Pitch angle [°]90
Accel. / braking time [s]14 / 24
Sample heating at n
temperature of 23 °C, running time 60 minutes
Aerosol-tight
1
[° C] referred to ambient
max
16
No
Permissible temperature range autoclavable °C134
1
Tested by HPA, Porton-down, UK
Table 4.Sorvall 120V 60Hz
CentrifugeSorvall® ST 8
Catalog #75007200
Weight empty [kg]1.0
Maximum permissible load [g]24 x 0.2
Maximum speed n
Maximum RCF value at n
[rpm]13300
max
16810
max
Max. / min. radius [cm]8.5 / 2.0
Pitch angle [°]90
Accel. / braking time [s]14 / 24
Sample heating at n
temperature of 23 °C, running time 60 minutes
Aerosol-tight
1
[° C] referred to ambient
max
16
No
Permissible temperature range autoclavable °C134
1
Tested by HPA, Porton-down, UK
Thermo ScientificHematocrit Rotor11
2
Rotor Specifications
Thermo Scientific Sorvall ST 8 Centrifuge
Table 5. Sorvall 100V 50/60Hz
CentrifugeSorvall® ST 8
Catalog #75007202
Weight empty [kg]1.0
Maximum permissible load [g]24 x 0.2
Maximum speed n
Maximum RCF value at n
[rpm]13300
max
16810
max
Max. / min. radius [cm]8.5 / 2.0
Pitch angle [°]90
Accel. / braking time [s]14 / 24
Sample heating at n
temperature of 23 °C, running time 60 minutes
Aerosol-tight
1
[° C] referred to ambient
max
16
No
Permissible temperature range autoclavable °C134
1
Tested by HPA, Porton-down, UK
12Hematocrit RotorThermo Scientific
Thermo Scientific SL 8 Centrifuge
Table 6.Thermo Scientific 230V 50/60Hz
CentrifugeSL 8
Catalog #75007221
Weight empty [kg]1.0
Maximum permissible load [g]24 x 0.2
2
Rotor Specifications
Thermo Scientific SL 8 Centrifuge
Maximum speed n
Maximum RCF value at n
[rpm]13300
max
16810
max
Max. / min. radius [cm]8.5 / 2.0
Pitch angle [°]90
Accel. / braking time [s]14 / 24
Sample heating at n
temperature of 23 °C, running time 60 minutes
Aerosol-tight
1
[° C] referred to ambient
max
16
No
Permissible temperature range autoclavable °C134
1
Tested by HPA, Porton-down, UK
Table 7.Thermo Scientific120V 60Hz
CentrifugeSL 8
Catalog #75007220
Weight empty [kg]1.0
Maximum permissible load [g]24 x 0.2
Maximum speed n
Maximum RCF value at n
Max. / min. radius [cm]8.5 / 2.0
[rpm]13300
max
16810
max
Pitch angle [°]90
Accel. / braking time [s]14 / 24
Sample heating at n
temperature of 23 °C, running time 60 minutes
Aerosol-tight
1
[° C] referred to ambient
max
16
No
Permissible temperature range autoclavable °C134
1
Tested by HPA, Porton-down, UK
Thermo ScientificHematocrit Rotor13
3
Accessories
Accessories
ArticleNo.
Capillaries76000923
Sealing Kit75000964
Rotor Seal Replacement75003437
Replacement Linear Scale Reading Graph76000938
Thermo ScientificHematocrit Rotor15
4
Thermo Scientific Auto-Lock Rotor Exchange
Thermo Scientific Auto-Lock Rotor Exchange
Contents
• “Rotor Installation” on page 18
• “Removing the Rotor” on page 19
Thermo ScientificHematocrit Rotor17
4
O-ring
Auto-Lock
Thermo Scientific Auto-Lock Rotor Exchange
Rotor Installation
Rotor Installation
CAUTION Unapproved or incorrectly combined accessories can cause serious damage to
the centrifuge.
This rotor is equipped with an Auto-Lock™-system.
This system is used to automatically lock the rotor to the centrifuge spindle. The rotor does not have to
be bolted onto the centrifuge spindle.
Proceed as follows:
1. Open the lid of the centrifuge and if necessary remove any dust, foreign objects or residue from the
chamber.
Auto-Lock
and O-ring must be clean and undamaged.
Figure 1. Auto-Lock
2. Hold the rotor over the centrifuge spindle and let it slide slowly down the centrifuge spindle.
The rotor clicks automatically into place.
CAUTION Do not force the rotor onto the centrifuge spindle.
If the rotor is very light, then it may be necessary to press it onto the centrifuge spindle
with a small amount of pressure.
3. Check if the rotor is properly installed by lifting it slightly on the handle. If the rotor can be pulled
up, then it must be reclamped to the centrifuge spindle.
WARNING If the rotor cannot be properly locked in place after several attempts, then the
Auto-Lock is defective and you are not permitted to operate the rotor.
Check for any damage to the rotor. Damaged rotors must not be used.
Keep the hub area clear of objects.
CAUTION Check that the rotor is properly locked on the centrifuge spindle before each
use by pulling it at its handle.
18Hematocrit RotorThermo Scientific
WARNING Be sure to check all seals before starting any aerosol-tight applications.
4. Close the centrifuge door.
Removing the Rotor
To remove the rotor, proceed as follows:
1. Open the centrifuge door.
2. Grab the rotor handle with both hands and press against the green Auto-Lock button. At the same
time, pull the rotor directly upwards with both hands and remove it from the centrifuge spindle.
Make sure not to jam the rotor while doing this.
4
Thermo Scientific Auto-Lock Rotor Exchange
Removing the Rotor
Figure 2. Handling of Auto-Lock
Thermo ScientificHematocrit Rotor19
4
Thermo Scientific Auto-Lock Rotor Exchange
Removing the Rotor
20Hematocrit RotorThermo Scientific
Rotor Loading
Contents
• “Before a Run” on page 22
• “Filling the Hematocrit Capillaries” on page 22
• “Rotor Temperature Range” on page 22
• “Duration” on page 23
• “Rotor Loading” on page 23
• “Maximum Loading” on page 24
• “Using the Reading Graph” on page 24
5
Rotor Loading
• “Rotor Life-time” on page 25
Thermo ScientificHematocrit Rotor21
5
Rotor Loading
Before a Run
Before a Run
1. Please read and observe the safety instructions contained in these operating instructions and in the
instructions for use.
2. Check the rotor and all accessory parts for damages such as cracks, scratches or traces of corrosion.
3. Check the rotor chamber, the centrifuge spindle and the Auto-Lock.
4. Check the rotor suitability using the “Chemical Compatibility Chart” on page 39.
Filling the Hematocrit Capillaries
Note Mind the lifetime of the capillaries.
1. Fill the hematocrit capillary (760000923) with a blood column of approximately 65 mm.
Make sure that the second opening remains dry.
2. Close the dry end of the hematocrit capillary with sealing putty (75000964). To do this, push the
hematocrit capillary vertically into the sealing putty until its rim touches the sealing putty’s plate.
Tilt it slightly and pull the hematocrit capillary carefully out of the putty. Make sure that the capillaries
are well sealed with the sealing putty.
WARNING The rotor cannot be used for aerosol-tight applications. The O-ring on the
rotor lock has the function to secure the rotor lid when screwed on. The external lid lip
cannot be sealed.
Rotor Temperature Range
CAUTION Operate the rotor in a temperature range between -9°C and +40°C only. A
pre-tempering in a freezer below -9°C is not allowed.
Note The rotor can warm up in ventilated centrifuges. Temperature above 45°C the samples can be
damaged. Let the rotor cool down between two runs.
22Hematocrit RotorThermo Scientific
Duration
RCF11 18,
n
1000
-------------
2
r××=
5
Rotor Loading
Duration
The duration depends on the RCF-value. For a clear separation the run should last at least 5 minutes.
The following values are suggested for reproducible results.
Speed rpmDuration in Minutes
133008
Explanation of RCF-Value
The relative centrifugal force (RCF) is given as a multiple of the force of gravity g. It is a unitless
numerical value which is used to compare the separation or sedimentation capacity of various
centrifuges, since it is independent of the type of device. Only the centrifuging radius and the speed
come into play in it:
r = centrifuging radius in cm
n = rotational speed in rpm
The maximum RCF value is related to the maximum radius of the tube opening.
Remember that this value is reduced depending on the tubes and adapters used.
This can be accounted for in the calculation above if required.
Rotor Loading
1. Place the hematocrit capillary with its closed end pointing outwards (towards the rim of the rotor)
horizontally into the hematocrit rotor. The hematocrit capillary must touch the sealing (7500
3437). The sealing protects the sensible capillaries from the hard rotor rim. The capillaries can
break during a run when the sealing is missing.
2. Close the rotor lid by turning it clockwise.
CAUTION Ensure that a balancing hematocrit capillary is placed opposite in order to avoid
imbalance and possible damage.
If not all places are filled with hematocrit capillaries, they must be placed with an opposite.
Thermo ScientificHematocrit Rotor23
CAUTION Damaged parts must be replaced immediately.
The rotor can run at high speeds. The rotor design has sufficient reserve stability even when spinning
at top speed.
The safety system of the centrifuge requires that you do not overload the rotor.
There are two options available for centrifuging samples whose weight, including adapter, exceeds the
maximum permissible load:
• Reduce the fill level.
• Reduce the speed.
Use the table or the formula:
n
adm
n
max
Actual Load (g/cavity)RPM
0.213300
0.310859
0.49405
0.58412
0.67679
0.77109
0.86650
0.96270
1.05948
= admissible speed
= maximum speed
Using the Reading Graph
The reading graph (76000938) is supplied with the rotor. To read the values, proceed as follows:
max
1. Remove the hematocrit capillaries one by one from the rotor.
Note Be careful not to shake up the sample while removing. If the samples have been shaken up
during removal, they can be centrifuge again.
2. Place the bottom end of the hematocrit capillary on the zero line with the blood column and the
upper end of the plasma column on the 100% line of the graph.
The dividing line between erythrocytes and plasma indicates the percentage of packed cells.
Note Be careful to place the hematocrit capillary at a right angle to the zero line.
24Hematocrit RotorThermo Scientific
Rotor Life-time
The rotor has no life-time restrictions. For safety reasons please bear the following in mind:
• UV rays reduce the stability of plastics. Do not subject the centrifuge, rotors and plastic accessories
to direct sunlight.
• If the rotor shows signs of decoloration, deformation, wear or imbalance it must be replaced.
5
Rotor Loading
Rotor Life-time
Thermo ScientificHematocrit Rotor25
Maintenance and Care
Contents
• “Cleaning Intervals” on page 28
• “Cleaning” on page 29
• “Disinfection” on page 31
• “Decontamination” on page 32
• “Au to cl av ing ” on page 33
• “Thermo Fisher Scientific Service” on page 33
6
Maintenance and Care
Thermo ScientificHematocrit Rotor27
6
Touching Ends
Maintenance and Care
Cleaning Intervals
Cleaning Intervals
For the sake of personal, environmental, and material protection, you have to clean and if necessary
disinfect the centrifuge on a regular basis.
MaintenanceRecommended Interval
Clean rotor chamberDaily or when polluted
Clean rotorDaily or when polluted
AccessoriesDaily or when polluted
CAUTION Refrain from using any other cleaning or decontamination procedure than
those recommended here, if you are not entirely sure that the intended procedure is safe
for the equipment.
Use only approved cleansers.
If in doubt, contact Thermo Fisher Scientific.
Replace Sealing
1. Remove the old sealing.
2. Form the new sealing into a ring without twisting it.
3. Put the touching ends into the rotor groove so that they are placed between two capillary slots.
Press the cushion ring completely into the groove, avoiding creases or waves.
28Hematocrit RotorThermo Scientific
How To Deal With Broken Hematocrit Capillary Tubes
How To Deal With Broken Hematocrit Capillary Tubes
WARNING Infectious material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
Broken capillary tubes have sharp edges and are a risk of injury
1. Remove the lid of the Hematocrit Rotor carefully.
2. Remove the larger pieces of the capillary tubes with tweezers.
3. Remove the rotor.
4. Remove the sealing slowly and carefully with the tweezers.
5. Clean and disinfect the rotor as described below.
6. Place a new sealing.
6
Maintenance and Care
Cleaning
When cleaning the centrifuge:
• Use warm water with a neutral solvent.
• Never use caustic cleaning agents such as soap suds, phosphoric acid, bleaching solutions or
scrubbing powder.
• Rinse the cavities out thoroughly.
• Use a soft brush without metal bristles to remove stubborn residue.
• Afterwards rinse with distilled water.
• Place the rotors on a plastic grate with their cavities pointing down.
• If drying boxes are used, the temperature must never exceed 50 °C, since higher temperatures
could damage the material and shorten the lifetime of the parts.
• Use only disinfectants with a pH of 6-8.
• Dry aluminum parts off with a soft cloth.
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (part no.
70009824). Also treat the cavities with oil.
• Store the aluminum parts at room temperature or in a cold-storage room with the cavities pointing
down.
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
Thermo ScientificHematocrit Rotor29
6
Maintenance and Care
Cleaning
Clean centrifuge and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
5. Remove the centrifuge tubes and adapters.
6. Use a neutral cleaning agent with a pH 6-8 for cleaning.
7. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50 °C.
8. Clean the housing of the centrifuge as needed.
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (part no.
70009824). Also treat the cavities with oil.
• Tread the bolt of the swing out rotor with bolt grease (part no. 75003786).
CAUTION When cleaning, do not allow liquids, especially organic solvents, to get on the
drive shaft, the bearings, the Auto-Lock or the locks.
Organic solvents break down the grease in the motor bearing. The drive shaft could freeze
up.
After some applications their might be ice in the rotor chamber. Let the ice melt and drain it off. Clean
the rotor chamber as described above.
30Hematocrit RotorThermo Scientific
Disinfection
Disinfect the centrifuge immediately whenever infectious material has spilled during centrifugation.
Use a sprayer whenever possible so that all surfaces are covered evenly.
The rotor chamber and the rotor should be treated preferably with a neutral disinfectant.
6
Maintenance and Care
Disinfection
WARNING Infectious material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if need be.
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
Observe the safety precautions and handling instructions for the cleaning agents used.
Contact the Service Department of Thermo Fisher Scientific for questions regarding the use of other
disinfectants.
Disinfect the rotor and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
5. Remove the centrifuge tubes and adapters and dispose of them or disinfect them.
6. Treat the rotor and accessories according to the instructions for the disinfectant (spray or soak in
solution). Adhere strictly to the given application times.
7. Be sure the disinfectant can drain off the rotor.
8. Rinse the rotor and rotor lid thoroughly with water and then rub down.
9. Dispose of the disinfectant according to the applicable guidelines.
10. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50 °C.
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (part no.
70009824). Also treat the cavities with oil.
• Tread the bolt of the swing out rotor with bolt grease (part no. 75003786).
Thermo ScientificHematocrit Rotor31
6
Maintenance and Care
Decontamination
Decontamination
Decontaminate the centrifuge immediately whenever radioactive material has spilled during
centrifugation.
WARNING Radioactive material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
neccessary precautions.
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if needed.
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
For general radioactive decontamination use a solution of equal parts of 70% ethanol, 10% SDS and
water.
Disinfect the rotor and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Grasp the rotor with both hands and lift it vertically off the centrifuge spindle.
5. Remove the centrifuge tubes and adapters and dispose of them or disinfect them.
6. Rinse the rotor first with ethanol and then with de-ionized water.
• Adhere strictly to the given application times.
7. Be sure the decontamination solution can drain off the rotor.
8. Rinse the rotor and accessories thoroughly with water.
9. Dispose of the decontamination solution according to the applicable guidelines.
10. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50 °C.
• After cleaning, treat the entire surface of aluminum parts with corrosion protection oil (part no.
70009824). Also treat the cavities with oil.
• Tread the bolt of the swing out rotor with bolt grease (part no. 75003786).
32Hematocrit RotorThermo Scientific
Autoclaving
1. Before autoclaving clean rotor and accessories.
2. Place the rotor on a flat surface.
Clean the rotor before autoclaving and rinse it with distilled water. Remove all accessories (tubes,
adapters) from the rotor. Place the rotor on a flat surface.
Note No chemical additives are permitted in the steam.
6
Maintenance and Care
• Rotors and adapter can be autoclaved at 121 °C.
• The maximum permissible autoclave cycle is 20 minutes at 121 °C.
CAUTION Never exceed the permitted temperature and duration when autoclaving.
If the rotor shows signs of corrosion or wear, it must be replaced.
WARNING Dangerous materials through leaky seals. The O-ring of a sealed bucket will
be damaged through heat exposure. Remove it before autoclaving a sealed bucket. Use a
new O-ring.
Autoclaving
Thermo Fisher Scientific Service
Thermo Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an
authorized service technician. The service technicians check the following:
• the electrical equipment
• the suitability of the set-up site
• the lid lock and the safety system
•the rotor
• the fixation of the rotor and the drive shaft
Thermo Fisher Scientific offers inspection and service contracts for this work.
Thermo ScientificHematocrit Rotor33
RCF-Values
A
RCF-Values
Speed (rpm)R
3002.08.529
4002.08.5415
5002.08.5624
6002.08.5834
7002.08.51147
8002.08.51461
9002.08.51877
10002.08.52295
11002.08.527115
12002.08.532137
13002.08.538161
14002.08.544186
15002.08.550214
16002.08.557243
17002.08.565275
18002.08.572308
min
R
max
RCF R
min
RCF R
max
19002.08.581343
20002.08.589380
21002.08.599419
22002.08.5108460
23002.08.5118503
24002.08.5129547
25002.08.5140594
26002.08.5151642
27002.08.5163693
28002.08.5175745
29002.08.5188799
30002.08.5201855
31002.08.5215913
32002.08.5229973
33002.08.52441035
34002.08.52581099
Thermo ScientificHematocrit Rotor35
A
RCF-Values
Speed (rpm)R
min
R
max
RCF R
min
RCF R
35002.08.52741164
36002.08.52901232
37002.08.53061301
38002.08.53231372
39002.08.53401445
40002.08.53581520
41002.08.53761597
42002.08.53941676
43002.08.54131757
44002.08.54331840
45002.08.54531924
46002.08.54732011
47002.08.54942099
48002.08.55152189
49002.08.55372282
50002.08.55592376
max
51002.08.55822472
52002.08.56052570
53002.08.56282669
54002.08.56522771
55002.08.56762875
56002.08.57012980
57002.08.57263088
58002.08.57523197
59002.08.57783308
60002.08.58053421
61002.08.58323536
62002.08.58603653
63002.08.58873772
64002.08.59163892
65002.08.59454015
66002.08.59744140
67002.08.510044266
68002.08.510344394
69002.08.510654524
70002.08.510964656
71002.08.511274790
36Hematocrit RotorThermo Scientific
A
RCF-Values
Speed (rpm)R
min
R
max
RCF R
min
RCF R
72002.08.511594926
73002.08.511925064
74002.08.512245204
75002.08.512585345
76002.08.512925489
77002.08.513265634
78002.08.513605782
79002.08.513955931
80002.08.514316082
81002.08.514676235
82002.08.515036390
83002.08.515406547
84002.08.515786705
85002.08.516166866
86002.08.516547028
87002.08.516927193
max
88002.08.517327359
89002.08.517717527
90002.08.518117697
91002.08.518527869
92002.08.518938043
93002.08.519348219
94002.08.519768397
95002.08.520188576
96002.08.520618758
97002.08.521048941
98002.08.521479127
99002.08.521929314
100002.08.522369503
101002.08.522819694
102002.08.523269887
103002.08.5237210082
104002.08.5241810278
105002.08.5246510477
106002.08.5251210678
107002.08.5256010880
108002.08.5260811084
Thermo ScientificHematocrit Rotor37
A
RCF-Values
Speed (rpm)R
min
R
max
RCF R
min
RCF R
109002.08.5265711291
110002.08.5270611499
111002.08.5275511709
112002.08.5280511921
113002.08.5285512134
114002.08.5290612350
115002.08.5295712568
116002.08.5300912787
117002.08.5306113009
118002.08.5311313232
119002.08.5316613457
120002.08.5322013684
121002.08.5327413913
122002.08.5332814144
123002.08.5338314377
124002.08.5343814612
max
125002.08.5349414848
126002.08.5355015087
127002.08.5360615327
128002.08.5366315570
129002.08.5372115814
130002.08.5377916060
131002.08.5383716308
132002.08.5389616558
133002.08.5395516810
38Hematocrit RotorThermo Scientific
Chemical Compatibility Chart
®
®
CCCLEARCRIMP
B
Chemical Compatibility Chart
CHEMICAL
MATERIAL
ALUMINUM
ANODIC COATING for ALUMINIUM
BUNA N
,CLEARCRIMP
®
, POLYCLEAR
1
CELLULOSE ACETATE BUTYRATE
POLYURETHANE ROTOR PAINT
COMPOSITE Carbon Fiber/Epoxy
DELRIN®ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL®NYLON
PET
POLYALLOMER
POLYCARBONATE
POLYESTER, GLASS THERMOSET
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
®
, TEFLON
®
RULON A
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
®
TYGON®VITON
2-mercaptoethanolSSU- SMS- SUSSUSS- SSSSUSSSSSS
AcetaldehydeS -U U ---M-U---MU U U MM- MS U - S - U