Thermo Conductivity tester Control 330 digital User guide

Page 1
Operating Instructions
TKA Control 330 Conductivity meter
  Art.-no.: 04.1805   Art.-no.: 04.1805-SV   Art.-no.: 04.1805-SVE   Art.-no.: 04.1806   Art.-no.: 04.1806-SV   Art.-no.: 04.1806-SVE   Art.-no.: 04.1807   Art.-no.: 04.1807-SV   Art.-no.: 04.1807-SVE
These instructions must be read prior to assembly and first use!
Page 2
Conductivity meter
TKA Control 330
EC-Declaration of Conformity
in accordance with the EEC low-voltage directive 2006/95/EC accordance with
appendix III B
Hereby we explain that those corresponds to below designated products in this conception and
design as well as in circulation the execution the fundamental safety and health requirements of
the Community directive low-voltage brought by us. In case of a change of the product not co-
ordinated with us this explanation loses its validity.
Manufacturer: TKA Wasseraufbereitungssysteme GmbH „Part of Thermo Fisher Scientific“ Stockland 3 D-56412 Niederelbert
Description of the electrical equipment:
function: Conductivity meter
type: Control 330
article number: 04.1805, 04.1806, 04.1807
04.1805-SV, 04.1806-SV, 04.1807-SV
04.1805-SVE, 04.1806-SVE, 04.1807-SVE
The agreement with further valid guidelines/regulations following for the product is explained:
EMC Directive (2004/108/EC)
Reference to the harmonised standards:
DIN EN 55014-1, Electromagnetic compatibility - Emission DIN EN 55014-2, Electromagnetic compatibility - Immunity DIN EN 60204-1, Electrical equipment of machines DIN EN 61000-3-2, Electromagnetic compatibility: Limits DIN EN 61000-3-3, Electromagnetic compatibility: Limits
Year of the CE characteristic assignment: 10
Niederelbert, 18th February 2010
Detlef Opp, Head of Technical documentation
Signature
2
Page 3
Conductivity meter
TKA Control 330
1. Contents
Page
1. Contents ................................................................................................. 3
2. Extent of delivery.................................................................................... 4
2.1 Completeness and intactness............................................................................................ 4
2.2 Parts lists ........................................................................................................................... 5
3. Assembly and starting for the first time .................................................. 8
3.1 Set-up diagram - Cartridges with R ¾“ connectors (04.1805) ......................................... 10
3.2 Set-up diagram - Stainless steel cartridges with quick-connects (04.1805-SVE)............ 11
3.3 Set-up diagram - Plastic cartridges with quick-connects (04.1805-SV)........................... 12
4. Operation.............................................................................................. 13
4.1 Menu................................................................................................................................ 14
4.2 Setting the limiting conductivity value .............................................................................. 14
4.3 Setting the limiting value for temperature ........................................................................ 15
4.4 Communication................................................................................................................ 15
5. Replacement parts and accessories ..................................................... 16
6. Technical Specifications....................................................................... 17
3
Page 4
Conductivity meter
TKA Control 330
2. Extent of delivery
2.1 Completeness and intactness
Please read these operating instructions attentively before you bring your new conductivity meter into operation.
The instructions clearly and simply lead you through the way to handle the instrument.
We wish you every success
in your work with
Conductivity meter
TKA-Control 330
The individual parts of the TKA Control 330 conductivity meter were assembled and packed
with the greatest possible care.
Nevertheless, please check the delivery for completeness and intactness before starting on assembly.
The correct content of the delivery is given in the parts lists that follow.
Should you have reason to complain, please contact:
TKA Wasseraufbereitungssysteme GmbH Stockland 3 D-56412 Niederelbert
Telephone: +49 (26 02) 1 06 99-0 Telefax: +49 (26 02) 1 06 99-50 E-mail: info@tka.de Internet: www@tka.de
4
Page 5
Conductivity meter
TKA Control 330
2.2 Parts lists
Art. no.: 04.1805
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded. For connection to a stainless steel cartridge with R 3/4" connector
Conductivity meter of type Control 330 incl. wall mount  Plug transformer 90 - 264 V, 50-60 Hz / 12 V DC.  Measuring cell incl. cable and T-piece connector
for connection to a water deionizing cartridge
Feedwater hose with R ¾“ connecting spouts  Purified water hose with R ¾“ connecting spouts  1 x Gasket, R ¾“
Art. no.: 04.1806
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded and potential-free contact. For connection to a stainless steel cartridge with R 3/4" connector
Conductivity meter of type Control 330 incl. wall mount  Plug transformer 230 V, 50-60 Hz / 12 V DC incl. potential-free contact.  Measuring cell incl. cable and T-piece connector
for connection to a water deionizing cartridge
Feedwater hose with R ¾“ connecting spouts  Purified water hose with R ¾“ connecting spouts  1 x Gasket, R ¾“
Art. no.: 04.1807
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded and including a plug transformer with switchbox for connection to a solenoid valve. For connection to a stainless steel cartridge with R 3/4" connector
Conductivity meter of type Control 330 incl. wall mount  Chem. nickel-plated solenoid valve  Plug transformer 230 V, 50-60 Hz / 12 V DC incl. potential-free
contact and switchbox for connection to a solenoid valve
Measuring cell incl. cable and T-piece connector for connection to a water deionizing
cartridge
Feedwater hose with R ¾“ connecting spouts  Purified water hose with R ¾“ connecting spouts  2 x Gasket, R ¾“
5
Page 6
Conductivity meter
TKA Control 330
Art. no.: 04.1805-SV
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded. For connection to a plastic cartridge with quick-connect connectors
Conductivity meter of type Control 330 incl. wall mount  Plug transformer 90 - 264 V, 50-60 Hz / 12 V DC.  Measuring cell incl. cable  Feedwater hose with R ¾“connector + quick-connect connector
for connection to a water deionizing cartridge
Purified water hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
2 x Gasket for R 3/4" connector
Art. no.: 04.1806-SV
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded and potential-free contact. For connection to a plastic cartridge with quick-connect connectors
Conductivity meter of type Control 330 incl. wall mount  Plug transformer 230 V, 50-60 Hz / 12 V DC incl. potential-free contact  Measuring cell incl. cable  Feedwater hose with R ¾“connector + quick-connect connector
for connection to a water deionizing cartridge
Purified water hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
2 x Gasket for R 3/4" connector
Art. no.: 04.1807-SV
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded and including a plug transformer with switchbox for connection to a solenoid valve. For connection to a plastic cartridge with quick-connect connectors
Conductivity meter of type Control 330 incl. wall mount  Chem. nickel-plated solenoid valve  Plug transformer 230 V, 50-60 Hz / 12 V DC incl. potential-free
contact and switchbox for connection to a solenoid valve
Measuring cell incl. cable  Feedwater hose with R ¾“connector + quick-connect connector
for connection to a water deionizing cartridge
Purified water hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
2 x Gasket for R 3/4" connectors
6
Page 7
Conductivity meter
TKA Control 330
Art. no.: 04.1805-SVE
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded. For connection to a stainless steel cartridge with quick-connect connectors
Conductivity meter Control 330 incl. wall mount  Plug transformer 90 - 264 V, 50-60 Hz / 12 V DC.  Measuring cell incl. cable  Feedwater hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
Purified water hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
Art. no.: 04.1806-SVE
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded and potential-free contact. For connection to a stainless steel cartridge with quick-connect connectors
Conductivity meter Control 330 incl. wall mount  Plug transformer 230 V, 50-60 Hz / 12 V DC incl. potential-free contact  Measuring cell incl. cable  Feedwater hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
Purified water hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
Art. no.: 04.1807-SVE
Conductivity meter of type Control 330 with digital 3½-digit LCD measured value display in S/cm, temperature measurement in °C, automatic temperature compensation, limiting conductivity limit switch settable up to 30 S/cm, limiting temperature switch settable up to 40°C, acoustic and optical signals when a limiting value is exceeded and including plug transformer with switchbox for connection to a solenoid valve. For connection to a stainless steel cartridge with quick-connect connectors
Conductivity meter Control 330 incl. wall mount  Chem. nickel-plated solenoid valve  Plug transformer 230 V, 50-60 Hz / 12 V DC incl. potential-free
contact and switchbox for connection to the solenoid valve
Measuring cell incl. cable  Feedwater hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
Purified water hose with R ¾“ connector + quick-connect connector
for connection to a water deionizing cartridge
1 x Gasket, R ¾“
7
Page 8
Conductivity meter
TKA Control 330
3. Assembly and starting for the first time
1. Screw the feedwater hose (Part 2) to the cartridge inlet. With cartridges with quick­connects, plug the hose on so that there is an audible click.
2. Make contact to the water tap: Connect the R ¾“ connecting spout on the other end of the feedwater hose to a water tap that can be turned off.
With instrument versions of article numbers: 04.1807(-SV and -SVE) - Control 330 with solenoid valve, plug transformer and integrated switchbox for connection to the solenoid valve: First connect the solenoid valve to the water tap with the R¾“ connecting spout. Following this, connect the R¾“ male thread on the other end of the hose to the solenoid valve.
3. Hold the wall mount at the selected position, mark where to bore, insert plugs and fix tightly with the screws supplied.
4. Stick the smooth Velcro strip to the holder and the rough Velcro strip to the instrument.
5. Hang the conductivity meter (Part 5) on.
6. Cartridges with ¾“ connector: Fix the measuring cell (Part 4) to the cartridge outlet with the R¾“ coupling nut, then screw the purified water hose (Part 3) to the ¾“ male thread of the measuring cell. Lead the other end of the purified water hose to the user or the storage tank.
Cartridges with quick-connects: Fit the quick-connect connector of the purified water hose (with the measuring cell) onto the outlet of the cartridge so that there is an audible clicking into position. Lead the other end of the purified water hose to the user or the storage tank.
7. Connect the cable from the measuring cell (Part 4) to the lower right socket of Control 330 (Part 5).
8. Connect the plug transformer to a suitable electrical socket (see Technical (see Specifications) and to the lower left socket of Control 300.
Safety precaution
Ensure that the cable connection does not allow liquid to run into the inside of the Control 300 conductivity meter!
8
Page 9
Conductivity meter TKA Control 330
9. Set the display mode required, the limiting conductivity value and, if appropriate, the limiting temperature value.
10. Open the water tap until the water deionizing cartridge has become filled and air-vent it with the venting screw. Purified water can now be taken from the end of the purified water hose. (Switching of decimals are automatically)
Important note: an exact conductivity value can only be read off during withdrawal.
After long downtimes (because of holidays, for example), re-ionization effects can cause the conductivity to increase and even reach the limiting value. Please then rinse the system for as long as it takes for a relevant conductivity value to be shown in the display.
9
Page 10
Conductivity meter
TKA Control 330
3.1 Set-up diagram - Cartridges with R ¾“ connectors (04.1805)
O
Feedwater connector, R¾“ male thread
1
Feedwater hose, 1.5 m long; R¾“ coupling nut
2
Purified water hose, 1.5 m long; R¾“ coupling nut
3
Measuring cell and T-piece connector
4
TKA Control 330 Conductivity meter incl. wall mount
5
Plug transformer for Art.-no. 04.1805
6
Solenoid valve, chem. nickel-plated, with ¾“-connectors for Art.-no. 04.1807
7
TKA Control 330 Conductivity meter with solenoid valve, plug
8
transformer and switching contact for solenoid valve TKA Control 330 Conductivity meter with potential-free contact, for Art.-no. 04.1806
9
Fits mixed-bed ion exchangers with R ¾“ male thread of type:
10
Type in mm Height
DI 1500 240 410 DI 2000 240 490 DI 2800 240 600 DI 4000 240 700 DI 6000 240 1155 DI 11000 363 857 DI 15000 363 1095
10
for Art.-no. 04.1807
Art.-no.: 02.1500 Art.-no.: 02.2000 Art.-no.: 02.2800 Art.-no.: 02.4000 Art.-no.: 02.6000 Art.-no.: 02.11000 Art.-no.: 02.15000
Page 11
Conductivity meter
TKA Control 330
3.2 Set-up diagram - Stainless steel cartridges with quick-connects (04.1805-SVE)
O
Feedwater connector, R¾“ male thread
1
Feedwater hose, 1.5 m long; R¾“ coupling nut and quick-
2
connect connector Purified water hose, 1,5 m long; R¾“ coupling nut and quick-
3
connect connector Measuring cell incl. connecting cable
4
TKA Control 330 Conductivity meter incl. wall mount
5
Plug transformer for Art.-no.: 04.1805-SVE
6
Solenoid valve, chem. nickel-plated with ¾“ connectors for Art.-no.: 04.1807-SVE
7
TKA Control 330 Conductivity meter with solenoid valve,
8
plug transformer and switching contact for Solenoid valve TKA Control 330 Conductivity meter
9
with potential-free contact Fits mixed-bed ion exchangers with quick-connects
10
Type in mm Height
DI 1500 240 410 DI 2000 240 490 DI 2800 240 600 DI 4000 240 700 DI 6000 240 1155 DI 11000 363 857 DI 15000 363 1095 Quick-connect adapter Art.-no.: 03.1503
11
for Art.-no.: 04.1807-SVE
for Art.-no.: 04.1806-SVE
Art.-no.: 02.1500 Art.-no.: 02.2000 Art.-no.: 02.2800 Art.-no.: 02.4000 Art.-no.: 02.6000 Art.-no.: 02.11000 Art.-no.: 02.15000
11
Page 12
Conductivity meter
TKA Control 330
3.3 Set-up diagram - Plastic cartridges with quick-connects (04.1805-SV)
O
Feedwater connector, R¾“ male thread
1
Feedwater hose 1,5 m long; R¾“ coupling nut and quick-connect
2
connector Purified water hose, 1,5 m long; R¾“ coupling nut and quick-
3
connect connector Measuring cell incl. connecting cable
4
TKA Control 330 Conductivity meter incl. wall mount
5
Plug transformer for Art.-no.: 04.1805-SV
6
Solenoid valve, chem. nickel-plated, with ¾“ connectors for Art.-no.: 04.1807-SV
7
TKA Control 330 Conductivity meter with solenoid valve,
8
plug transformer and switching contact for solenoid valve TKA Control 330 Conductivity meter
9
with potential-free contact Fits fit mixed-bed ion exchangers with quick-connects
10
Type in mm Height
DI 800 P 205 410 DI 1900 P 205 610
12
for Art.-no.: 04.1807-SV
for Art.-no.: 04.1806-SV
Art.-no.: 01.1606-SV Art.-no.: 01.1614-SV
Page 13
Conductivity meter
TKA Control 330
4. Operation
O
When the instrument is switched on, all segments of the 7-segment display are shown for 3 seconds. During this time, the complete measurement cycle is run through and tested.
In the basic setting, the display shows the conductivity in µS/cm. The display is automatically scaled in the range from 0.055 to 199 µS/cm.
The upper rows of LED's show the measurement mode at which the instrument is currently in. The appropriate LED lights up green to show the measurement unit (µS/cm, Mxcm or °C).
The value 200 is shown in the display when the measurement range is exceeded.
Should there be an interruption between the instrument and the sensors, then the red "Status Sens." LED lights up and "---" is shown in the display instead of a measured value. A buzzer is additionally activated in the case of an interruption in the connection to the measurement sensor.
Quit-key:
The acoustic signal (buzzer) that is given when a fault occurs can be switched off with the Quit/UP-key. The acoustic signal is then no longer active until the limiting value is gone below. If a solenoid valve is attached, this is again opened.
13
Page 14
Conductivity meter
TKA Control 330
4.1 Menu
The menu consists of three sub-prompts; switching to the dual-mode and selection of the measurement unit for the conductivity measurement, the limiting value setting for the conductivity and the limiting value setting for the temperature.
Mono/ Dual-measuring mode:
Mono-measuring mode: In this mode only the conductivity is displayed.
The temperature can be read as long as the Enter-key is pressed.
Dual-measuring mode: In this mode the temperature and the conductivity are alternately displayed every 2 seconds. the LEDs for the measurement unit switch correspondingly with the display.
When the menu-key is pressed once, the measurement mode just set is shown. The up-
and down-keys allow switching between the permanent conductivity display and the
alternating conductivity/temperature display. The display shows "c" for conductivity or
"ct" for conductivity/temperature.
At the same time a choice can be made between the measurement units µS/cm and
Mxcm in this menu. The unit chosen must be confirmed by pressing the Enter-key, and
that chosen is shown by the corresponding LED.
The following combinations are so possible:
ct / MΩxcm
c / Mxcm
up/down
ct / µS/cm
c / µS/cm (default)
4.2 Setting the limiting conductivity value
Setting range: 0.06 - 30 µS/cm
Basic setting: off
The limiting conductivity value can be set when the menu-key is pressed twice (the red
“Cond." LED lights up). The setting is made with the up- and down-keys. The particular
position that is to be changed can be selected with the Enter-key and flashes in the
display. When the wanted value is reached, then this must again be confirmed with the
Enter-key.
Should a value of 30.1 µS/cm be set, then the word OFF appears in the display and the
evaluation of the limiting value is switched off.
When the limiting conductivity value (Limit value Conductivity) is exceeded, then the red
"Cond." LED lights up and the exceeding of it is additionally signalled by a buzzer.
14
Page 15
Conductivity meter TKA Control 330
Only the buzzer is switched off by the quit-key, the "Cond." LED is not switched off until the limiting value is gone below or the mains plug is unplugged.
Important note: The limiting value can only be entered in µS/cm.
4.3 Setting the limiting value for temperature
Setting range: 10 - 40°C Basic setting: OFF
The limiting temperature value can be set when the menu-key is pressed three times (the red “Temp." LED lights up). The setting is made with the up- and down-keys. The particular position that is to be changed can be selected with the Enter-key and flashes in the display. When the wanted value is reached, then this must again be confirmed with the Enter­key. Should a value of 41°C be set, then the word OFF appears in the display and the evaluation of the limiting value is switched off. When the limiting temperature value (Limit value Temperature) is exceeded, then the red "Cond." LED lights up and the exceeding of it is additionally signalled by a buzzer.
4.4 Communication
The 04.1806 and 04.1807 versions have a potential-free contact available via the plug transformer, article-no. 26.0034, as well as an RS 232 Interface. The interface runs at a transmission rate of 9600 bits/sec., 8 data bits, 1 stop bit and no parity.
The conductivity in µS/cm and the temperature in °C are given out on the serial interface separated by a comma. The conductivity is automatically scaled to three significant figures. Output takes place once per hour. When in operation, a press on the down-key triggers a measured value print-out on the serial interface.
The SUB-D socket has the following assignment: PIN 2: TXD PIN 3: RXD PIN 5: GND
15
Page 16
Conductivity meter TKA Control 330
Pin assignment of the DIN-socket of the
potential-free contact:
5. Replacement parts and accessories
Designation Article no.
Plug transformer 90 - 264 V, 50-60 Hz / 12 V DC:
Plug transformer 230 V, 50 Hz / 12 V DC with pot. free contact and switchbox:
Adapter plug for the plug transformer, for EU/US sockets:
Measuring cell incl. connecting cable:
Connecting cable for Control 330 with measuring cell and plug transformer:
PVC connecting hose:
Gasket, R 3/4":
Solenoid valve:
Printer for measured values print-out:
16.0235
26.0034
16.0317
16.0229
16.0310
18.0045
21.5008
03.1503
09.2207
16
Page 17
Conductivity meter
TKA Control 330
6. Technical Specifications
Conductivity measurement error: +/- 0,2% MBE
Temperature measurement error: +/- 0.5°C
Conductivity measurement range: 0.001 – 199 µS/cm (Scaling takes place automatically)
Temperature measurement range: 0.1 – 99.9°C
Measurement unit display: via LED
Temperature compensation: automatic
Setting range of the limiting conductivity value switch: 0.06- 200 µS/cm (continuously settable)
Setting range of the limiting temperature value switch: 10- 40 °C (continuously settable)
Exceeding of the limiting value: Exceeding is signalled when by the red LED and the buzzer. Both limiting values can be switched off
Interface: RS 232 (only article-no. 04.1806 + 04.1807)
Electrical connection: Plug transformer for article-no. 04.1805 90 - 264 V, 50-60 Hz; 6 W
Plug transformer for article-no. 04.1806 +
04.1807 with switchbox and pot. free contact 230 V, 50-60 Hz; 1.5 W
Measuring instrument voltage: 12 Volt DC Protection class: IP 54 Housing dimensions in mm: W 75 x D 30 x H 130 Positioning: Wall mounting
Should service be required, please always contact our service department:
TKA Wasseraufbereitungssysteme GmbH „Part of Thermo Fisher Scientific“ Stockland 3 D-56412 Niederelbert
Telephone: +49 (26 02) 1 06 99-0 Telefax: +49 (26 02) 1 06 99-50 E-mail: info@tka.de Internet: www.tka.de
17