Testo 535 Operating Manual

testo 535
Instruction manual
en
2 line display
Hold/Max/Min
MEAN button
START/STOP
ON/OFF
button
Screw-off
battery
compartment
PRINT button
The conformity certificate confirms that this product fulfills the guidelines in accor­dance with 2004/108/EEC.
Please read prior to operation
- Never measure on live parts
- Observe sensor measuring ranges!
Overheating may destroy probes.
- Observe storage and transport temperature as
well as the max. operating temperature (e.g. protect measuring instrument from direct sunlight).
- Inexpert handling or use of force cancel your
warranty claims!
- If there is a change in the ambient temperature
(change of location, e.g. inner/outer temperature) the measuring instrument will need an adaptation phase of several minutes.
- The CO2 probe contains optical components. Please treat the probe like your camera. Strong vibrationa may chenge the manufac­turer´s calibration. Check the readings in fresh air: 350 to 450 ppm. It is recommended that the instrument is sent to a Testo service point for adjustment purpose.
- The CO2 probe is permanently attached to the measuring instrument. Probes can only be changed at Testo service points.
- Ensure the probe does not come into frequent contact with water otherwise the long-term stability will be affected.
- If the level of high frequency is high, there may be deviation from the required
Operation
Switch on instrument. After the instrument is switched on a short
the sensor has a warm-up phase of approx. 30 s. The instrument is then ready to operate. The CO2 concentration in the sensor needs approx. 1 minute to adapt to ambient conditions. This adaptation time can be shortened by swinging the probe gently back and forth. If the probe is subjected to condensation, this can lead to uncreased mesured values.
display and function test follows and
Mains connection (mains unit no. 0554.0088) recommended for long-term measuring.
The probe should be held as far as possible from your body so as to eliminate the influences caused by the CO2 in your breath.
Auto-Off function
If the Auto-Off function is activated the instrument switches itself off auto­matically after 10 minutes of inactivity (except in the case of HOLD) and timed mean calculation.
Changing the battery
If“Bat“ appears in the display, the battery will last approx. 1 hour out of a total lifetime of 4 hours, if referring to an alkali-manganese
off automatically if there is not enough power. Unscrew the battery compartment, remove the used battery and put in new battery, type 9V (IEC 6 F 22). (Observe polarisation). User settings remain intact if the instrument is not longer than 3 minutes without power.
battery. The instrument will switch
Instructions for disposal:
Only empty batteries should be disposed. The batteries should be placed in plastic bags to prevent short-circuits.
Technical data - testo 535
Sensor: 2 channel infrared
absorption principle Measuring range: 0 to 9,999 ppm CO2 (0 to 0,999 vol.% CO2)
Resolution: 1 ppm or 0.001 vol.% Ambient
temperature: 0 to +50°C Storage temp.: -20 to +50 °C
Battery lifetime: >4h (9 V block, Al-Mn) Switchover: ppm / vol. % Display: LCD (11 mm digit height) Housing: ABS Dimensions: 190 x 57 x 42 mm Weight: Approx. 300 g Ambient humidity: 0 to 99 %RH
not subjected to condensation
Warranty
testo 535 meas. instr. ...............2 years
Probe .........................................1 year
2
* Temp. coefficient: ± (1 ppm + 0,4 % of meas.) / °C
Changing the settings - Configuration
START/STOP” button: keep pressed when switching on.
The following settings can be changed:
Date
Auto OFF
not activated
Auto ON
Auto-OFF
function
activated
Auto-OFF
function
BAT
Battery in
instrument
ACCU
Rechargeable battery in inst-
rument
01:01 1998
01:01 1998
Next position
Set value
Pressure compensation
Please note that the CO2 reading depends on the absolute air pressure. This effect is compensated in the instrument. To do this, enter the correct air pressure (See Changing the settings - Configuration). The air pressure depends on the height of the location above mean sea level (See the “Barometric height formula” table on the next page) and weather influences.
Example:
You are located at a height of 800 m above sea level where you have a mean annual air pressure of 920 hPa. According to the barometer (1003 hPa) and the height correction of the barometer (to 1013 hPa) the mean annual air pressure has to be reduced by 10 (910 hPa air pressure).
Time
00:00
00:00
Next position
Set value
Air pressure
ppm
Vol%
1013
hPa
Next position
Meas-
1013
hPa
uring
mode
Set value
Memory function for current, maximum and minimum values
The following displays appear when this button is pressed again and again:
1st press: HOLD - Readings are frozen in display 2nd press: MAX - Displays saved max. values 3rd press: MIN - Displays saved min. values 4th press: MEAN - Multi-point mean calculation 5th press: MEAN - Timed mean calculation 6th press: MEAN- Return to measuring mode
Important note:
The min/max values are measured from the time the instrument is switched on. The values are reset if the instrument is switched off and then on again.
Printout
A printout of the reading with date and time can be activated by pressing the PRINT button. The max, min and mean values also appear on the printout of a multi-point or timed mean cal­culation. If a reading was frozen at a certain point in the display via the HOLD button, this rea­ding appears in the printout with the time and date when the reading was frozen.
3
Loading...
+ 5 hidden pages