8.Care and maintenance ......................................61
9.Questions and answers......................................62
10. Technical data ....................................................63
11. Accessories / spare parts ..................................65
Page 2
General notes34
General notes
This chapter provides important advice on using this documentation.
The documentation contains information that must be applied if the product is
to be used safely and efficiently.
Please read this documentation through carefully and familiarise yourself with
the operation of the product before putting it to use. Keep this document to
hand so that you can refer to it when necessary.
Identification
Representation MeaningComments
Warning advice: Warning!Read warning advice carefully and take the precautionary measures
indicated! Serious physical injury could occur if you do not take the
precautionary measures indicated.
Warning advice: Caution!Read warning advice carefully and take the precautionary measures
NoteOffers helpful tips and information.
±, 1, 2ObjectiveDenotes the objective that is to be achieved via the steps described.
ConditionA condition that must be met if an action is to be carried out as
i, 1, 2, ...StepCarry out steps. Where steps are numbered, you must always follow
TextDisplay textText appears on the instrument display.
Button
-ResultDenotes the result of a previous step.
ºCross-referenceRefers to more extensive or detailed information.
Control buttonPress the button.
Function buttonPress the button.
indicated! Slight physical injury or damage to equipment could occur
if you do not take the precautionary measures indicated.
Where steps are numbered, you must always follow the order given!
described.
the order given!
Page 3
General notes 35
Short form
This document uses a short form for describing operating steps (e.g. calling up
a function).
Example: Calling up the “Instrument data” function
Short form: Device ¬¬Inst.data ¬.
1) (2) (3) (4)
(
OKOK
Steps required:
1 Press / to select the Device function.
2 Confirm selection with .
3 Press / to select the Inst.data function.
4 Confirm selection with .
OK
OK
de
enfresitptsvnl????
Page 4
1. Safety instructions36
1.Safety instructions
This chapter gives general rules which must be followed and observed if the
product is to be handled safely.
Avoid personal injury/damage to equipment
i Do not use the measuring instrument and probes to measure on or near live
parts.
i Never store the measuring instrument/measuring cells together with solvents
and do not use any desiccants.
Product safety/preserving warranty claims
i Operate the measuring instrument only within the parameters specified in the
Technical data.
i Always use the measuring instrument properly and for its intended purpose.
Do not use force.
i Do not expose handles and feed lines to temperatures in excess of 70 °C
unless they are expressly permitted for higher temperatures. Temperatures
given on probes relate only to the measuring range of the sensors.
i Open the instrument only when this is expressly described in the
documentation for maintenance or repair purposes.
Carry out only the maintenance and repair work that is described in the
documentation. Follow the prescribed steps when doing so. For safety
reasons, use only original spare parts from Testo.
Ensure correct disposal
i Take faulty rechargeable batteries/spent batteries to the collection points
provided for them.
i Send the product back to Testo at the end of its useful life. We will ensure
that it is disposed of in an environmentally friendly manner.
Page 5
1. Safety instructions 37
Instruments with radio module 915.00MHz FSK
Warning: Changes or modifications not expressly approved by the party responsible for compliance could void
the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part
15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the following measures:
· Reorient or relocate the receiving antenna.
· Increase the separation between the equipment and receiver.
· Connect the equipment into an outlet on a circuit different from that to which the receiver is needed.
· Consult the dealer or an experienced radio/TV technician for help.
Operation is subject to the following two conditions:
· this device may not cause harmful interference, and
· this device must accept any interference received, including interference that may cause undesired operation.
2.Intended purpose
This chapter gives the areas of application for which the product is intended.
Use the product only for those applications for which it was designed. Ask
Testo if you are in any doubt.
testo 435 is a compact, multifunction measuring instrument for measuring
temperature, humidity and flow rate.
The product was designed for the following tasks/applications:
· Measuring room climate
· Regulating and inspecting ventilation and air-conditioning installations
· Measuring the pressure dew point in compressed air systems
· Assessing the quality of room air with the help of the IAQ probe
The product should
· Areas at risk of explosion.
· Diagnostic measurements for medical purposes.
not
be used in the following areas:
de
enfresitptsvnl????
Page 6
3. Product description38
3.Product description
This chapter provides an overview of the components of the product and their
functions.
3.1Display and control elements
Overview
Infrared, USB interface
Display (light can be activated)
Control buttons
Rear: Battery and radio module compartment, holding
magnets
Strong magnets
Damage tto oother iinstruments!
i
Keep a safe distance from products which
could be damaged by magnetism (e.g.
monitors, computers, pacemakers, credit
cards).
Probe socket(s)
Button functions
ButtonFunctions
Function button (3x): The function depends on the button assignment
at the time
Change display of the 1
In configuration mode: Increase value, select option
Change display of the 2
In configuration mode: Decrease value, select option
Print data
435-1/-3 only: If the Cyclical Printing function is activated, the
programmed measuring program is started.
Switch instrument on, switch display light on/off;
switch instrument off (press and hold)
st
reading line
nd
reading line
Page 7
3. Product description 39
Function buttons (Function dependant on profile and setting)
ButtonFunctions
Open (main) menu
OK
ESC
ACTHold
/Hold value/display current measurement value
Reset
Mean
Measp
Start
End
Save
Turb
Area
P=0
Enter confirmation
Cancel
Reset max./min. values to current measurement value
Open menu item “Multi-point mean calculation“
Open menu item “Measuring program“ (435-2/-4 only)
435>
Start test series (435-2/-4 only)
End test series (435-2/-4 only), End Cyclical Print (435-1/ -3 only)
Save values (435-2/-4 only)
Activate “Turb“ test series (435-2/-4 only with attached turbulence probe)
Open menu item “Area”
Zero internal pressure sensor (435-3/-4 only)
Important displays
DisplayMeaning
Battery capacity (only for operation by battery/rechargeable battery):
· 4 segments in the battery symbol are lit: Instrument battery is fully charged
· No segments in the battery symbol are lit: Instrument battery is almost spent
(flashing) Print function: Data are sent to the printer
If a measurement channel is a radio channel, the radio symbol lights up as well as the
measurement channel no.
de
enfresitptsvnl????
Page 8
3. Product description40
3.2Interfaces
Infrared interface
Measurement data can be sent to a Testo printer via the infrared interface on
the head of the instrument.
USB interface
The mains unit (accessory part) can be connected to the head of the
instrument via the USB interface to power the instrument.
Instruments with a memory: Measurement/instrument data can be exchanged
with a PC via the USB interface.
Probe socket(s)
Plug-in measuring probes can be connected via the probe socket(s) on the
base of the instrument. The instrument is a HighPower device, possibly an
additional USB-Hub is required!
Radio module (accessory part)
Radio probes may only be used in countries in which they have been Type
Approved (see application information of the radio probe).
Up to three radio probes can be connected via the radio module.
3.3Voltage supply
Voltage is supplied via three mignon batteries (included in delivery) or
rechargeable batteries or through a mains unit (accessory part). It is not
possible to charge rechargeable batteries in the instrument.
When operating the instrument with the mains unit, insert batteries in order
to avoid switching off the instrument in case of a power interruption.
Page 9
4. Commissioning 41
4.Commissioning
This chapter describes the steps required to commission the product.
Inserting bbatteries/rechargeable bbatteries aand aa rradio mmodule ((accessory
²
part):
1 Undo the two screws on the rear of the instrument and lift off the battery
compartment cover.
2 Insert batteries/rechargeable batteries (3x mignon) into the battery
compartment. Observe the polarity!
3 Push the radio module (accessory part) into the radio module
compartment until it engages in place. Note the guide groove.
4 Replace the battery compartment cover, press down and secure by
tightening the two screws.
de
enfresitptsvnl????
Page 10
5. Operation42
5.Operation
This chapter describes the steps that are executed frequently when using the
product.
5.1Connecting a probe
Plug-in probes
Plug-in probes must be connected before the measuring instrument is
switched on so that they are recognised by the instrument.
i Insert the connector of the probe into the probe socket of the measuring
instrument.
Radio probes
Radio probes may only be used in countries in which they have been Type
Approved (see application information of the radio probe).
A radio module (accessory part) is required for the use of radio probes. The
radio module must be connected before the measuring instrument is switched
on so that it is recognised by the measuring instrument.
Each radio probe has a probe ID (identification number) which must be set in
configuration mode.
º See chapter PROBE, p. 47.
5.2Switching on/off
Switching tthe iinstrument oon:
²
i Press .
- Measurement view is opened: The current reading is displayed, or
---- lights up if no reading is available.
Instruments with a memory: The activated location is displayed
(topmost line).
-oor-
The instrument is switched on for the first time, a reset was carried out or
the power supply was interrupted for a lengthy period of time:
- The Language function is opened.
º See the chapter
LANGUAGE, p. 50.
Page 11
Switching tthe iinstrument ooff:
²
i Press and hold (for approx. 2s) until the display goes out.
5.3Display light
Switching tthe ddisplay llight oon/off:
²
The instrument is switched on.
i Press .
5. Operation 43
de
enfresitptsvnl????
Page 12
6. Setting the instrument44
6.Setting the instrument
This chapter describes the steps that are required in order to adapt the
measuring instrument for specific measuring tasks.
6.1Configuration menu
The basic settings for the measuring instrument are performed in the
configuration menu.
Opening tthe cconfiguration mmenu:
²
The instrument is in measurement view.
i Press and hold (approx. 2s) until config. is displayed.
Press to go one menu level back. To leave the configuration menu,
press several times until the instrument changes to measurement view.
6.1.1Profile
ESC
ESC
The instrument has predefined measurement profiles that are tailored to
specific areas of application.
The profile setting influences the following points in measurement mode:
· Assignment of the function buttons
· Number of predefined functions
· Structure of the main menu
All functions are available in the standard profile. In the application-specific
measurement profiles, the available functions are reduced to only those that are
needed to ensure speedier access.
Setting aa pprofile:
²
The configuration menu is open, config. is displayed.
1 Profile ¬.
2 Select the desired profile with / and confirm with.
OK
OK
Page 13
6. Setting the instrument 45
6.1.2Units
Predefined systems and individual setting options:
The configuration menu is open, config. is displayed.
1 Units ¬.
2 Press / ISO/US (to set the system) or a parameter (to set
individually) and confirm with.
3 Set the system of units or the desired unit with / and confirm
with.
OK
OK
OK
6.1.3Device
Instrument data
Displaying iinstrument ddata:
²
The configuration menu is open, config. is displayed.
1 Device ¬¬Inst.data ¬.
- The firmware version and the serial number of the instrument are
displayed.
OKOK
de
enfresitptsvnl????
Date/Time
Setting tthe ddate/time:
²
The configuration menu is open, config. is displayed.
1 Device ¬¬date/time ¬.
2 Use / to set the value for year and confirm with .
3 Set the other values as described in step 2.
OKOK
OK
Page 14
6. Setting the instrument46
Battery type
To ensure that the battery capacity is displayed correctly, the battery type used
must be set.
Setting tthe bbattery ttype:
²
The configuration menu is open, config. is displayed.
1 Device ¬¬Bat-type ¬.
2 Press / Battery or ReBa and confirm with.
OKOK
OK
Auto OFF
If Auto OFF is switched on, the instrument switches itself off automatically after
10min if no button is pressed. Exception: Cyclical printing (instruments without
a memory) or a measuring program (instruments with a memory) is active.
Switching AAuto OOFF oon/off:
²
The configuration menu is open, config. is displayed.
1 Device ¬¬Auto OFF ¬.
2 Press / to select On or Off and confirm with.
OKOK
OK
Reset
When a reset is carried out, the instrument is reset to the default settings, all
settings/data are deleted. Exception: Language, Date/Time.
Resetting:
²
The configuration menu is open, config. is displayed.
1 Device ¬¬reset ¬.
2 Reset with or cancel the reset with .
OKOK
ESCOK
Setting min. / max. printing function
If pr MinMaxAuto is activated, minimum and maximum values are also printed
with the measurement readings.
Switching ooff ppr MinMax:
²
The configuration menu is open, Config. is displayed.
1 Device ¬¬pr MinMax ¬.
2 Choose On or Off with / and confirm with.
OKOK
OK
Page 15
6. Setting the instrument 47
K-factor
The querying of the K-factor (correction factor for areas) when entering the
parameter „Area“.can be switched on/off.
º See chapter PARAMETER/AREA, p. 56.
Switching KK-ffactor oon/off:
²
The configuration menu is open, config. is displayed.
1 Device ¬¬K-factor ¬.
2 With / , select On or Off and confirm with.
OKOK
OK
Num holes
The querying of the number of holes (number of measurement points) in the
calculation of a multi-point mean value can be switched on/off. The number of
holes is required to allocate the readings to the number of measurement points
for later evaluation via the PC software.
º See Chapter M
Switching nnumber oof hholes oon/off:
²
The configuration menu is open, config. is displayed.
1 Device ¬¬Num holes ¬.
2 With / , select On or Off and confirm with.
EAN, p. 54 and
MEASURING, p. 58.
OKOK
OK
de
enfresitptsvnl????
6.1.4Probe
RadioC
Radio probes may only be used in countries in which they have been Type
Approved (see application information of the radio probe).
A radio module (accessory part) is required for the use of radio probes. The
instrument can establish a connection with a maximum of three radio probes.
Each radio probe has a probe ID (RF ID). This consists of the last 3 digits of the
serial no. and the position of the slide switch (H or L) in the radio probe.
Setting uup aa rradio pprobe:
²
A radio module (accessory part) is inserted in the instrument.
º See chapter COMMISSIONING, p. 41.
The configuration menu is open, config. is displayed.
The radio probe is switched on and the transfer rate is set to 2 readings
per second (see the advice on using the radio probe).
Page 16
6. Setting the instrument48
1 Probe ¬¬RadioC ¬.
2 Press / to select the desired channel no. for the radio probe (P. 1 ,
P. 2 or P. 3 ) and confirm with .
OKOK
OK
- The instrument searches for switched-on radio probes in the receiving
range.
- The probe IDs of the radio probes found are displayed.
If no radio probes were found, this may be because of the following:
· The radio probe is not switched on or the battery of the radio probe is
spent.
· The radio probe is outside the range of the measuring instrument.
· Sources of interference are influencing the radio transmission (e.g.
reinforced concrete, metal objects, walls or other barriers between
transmitter and receiver, other transmitters of the same frequency,
strong electromagnetic fields).
i If necessary, rectify the possible causes of the fault in radio
transmission.
Alternatively, the probe ID can also be entered manually.
MAN
i¬ Press / to enter the probe ID.
3 Press / to select the probe that is to be assigned to the chosen
channel no.
4 Assign the radio probe to the chosen channel no. with or exit the
function with , without changing the probe configuration.
ESC
OK
Humidity probe calibration (435-2/-4 only)
This function is only available if a humidity probe is plugged in.
The calibration values can be reset to the default settings (Reset). A 2-point
calibration can be performed.
Resetting tthe ccalibration vvalues:
²
The configuration menu is open, config. is displayed.
1 Probe ¬¬Calibr. ¬.
2 Press / to select Reset and confirm by pressing twice.
OKOK
OK
- The calibration values are reset to the default settings.
Page 17
6. Setting the instrument 49
Calibrating:
²
The configuration menu is open, config. is displayed.
1 Probe ¬¬Calibr. ¬.
OKOK
2 Press / to select calibration point P1 or P2 and confirm by
pressing twice.
OK
3 Put the humidity probe into the reference medium and wait for the
equalisation period to elapse.
- The current humidity reading and the calibration point (nominal value) are
displayed.
4 Start the calibration menu with .
5 Save calibration with or cancel calibration with .
OK
ESCOK
P internal (435-3/-4 only)
The internal pressure sensor can be switched on/off.
The configuration menu is open, config. is displayed.
1 Probe ¬¬P intern ¬.
2 Press / On or Off and confirm with.
OKOK
OK
de
enfresitptsvnl????
Te-Type
The probe characteristic curves stored in the instrument can be set for the
probe type used.
Setting pprobe ttype:
²
The configuration menu is open, Config. is displayed.
1 Probe ¬¬Te-Type ¬.
2 Select the desired probe type with / and confirm with.
OKOK
OK
Adjustment
The function is only available when an absolute pressure probe is attached.
The display value for the measurement of abolute pressure.
Carrying oout aadjustment:
²
The configuration menu is open, config. is displayed.
1 Probe ¬¬Adjustm. ¬.
2 With / , set the adjusted value and confirm with.
OKOK
OK
Page 18
6. Setting the instrument50
Pressure
The function is only available when an absolute pressure probe is attached.
You can set whether the absolute air pressure (measured with an absolute
pressure probe) or the barometric air pressure (calculated from the measured
absolute pressure and the input of the altitude above sea level).
º For the input of abs alt. for the calculation of the barometric air pressure, see
chapter PARAMETER/AREA, p. 56.
Setting mmeasurement pparameter:
²
The configuration menu is open, config. is displayed.
1 Probe ¬¬Pres. ¬.
2 With / , select the desired measurement parameter and confirm
OK
with.
OKOK
6.1.5Language
Setting tthe llanguage:
²
The configuration menu is open, config. is displayed.
1 Language ¬.
2 Select the desired language with / and confirm with.
OK
OK
6.2Main menu
Settings by which the measuring instrument can be adapted to the particular
measuring task are performed in the main menu.
The instrument has predefined measurement profiles that are tailored to
specific areas of application.
º See the chapter PROFILE, p. 44.
The profile setting influences the number of available functions and the
structure of the main menu.
The method described in this chapter for calling up the functions in the main
menu relates to the Standard profile setting. If a different profile is set, the
method for calling up individual functions may change or the function may
not be available in that particular profile. Some functions are only available
when a probe is connected or a wireless probe is switched on and
registered.
The instrument is in measurement view.
i Press .
- Menu is displayed.
Press to go one menu level back. To leave the main menu, press
ESC
ESC
several times until the instrument changes to measurement view.
6.2.1Memory (435-2/-4 only)
Info
The free memory space is displayed
Location
The active location can be changed. Up to 99 locations can be created. The
numerical location designations (01-99) can be changed into any text (max.
10 characters) using the PC software.
Changing aan aactive llocation:
²
The main menu is open, Menu is displayed.
1 Memory ¬¬Location ¬.
2 Press / to select the location to be activated and confirm with
OK
.
OKOK
Protocol
Saved measurement protocols can be printed out on a Testo printer (accessory
part) via the infrared interface.
Printing aa mmeasurement pprotocol:
²
The main menu is open, Menu is displayed.
1 Memory ¬¬Protocol ¬.
2 Press / to select the measurement protocol that is to be printed.
3 Press to start printing out the measurement protocol.
OKOK
Delete
The entire memory with all measurement protocols can be cleared.
Page 21
6. Setting the instrument 53
Clearing tthe mmemory:
²
The main menu is open, Menu is displayed.
1 Memory ¬¬Delete ¬.
2 Press to clear the entire memory.
OK
OKOK
6.2.2Measuring program (435-2/-4 only)
A measuring program can be programmed and activated/deactivated:
DesignationDescription
OffMeasuring program switched off: Readings can be stored manually
AUTOAutomatic measuring program: The measuring cycle (min. 1s) and the number of
TurbAutomatic measuring program for measuring turbulence (only if a turbulence probe
Deactivating aa mmeasuring pprogram:
²
The main menu is open, Menu is displayed.
1 Meas.Prog ¬.
2 Press / to select Off and confirm with .
- The instrument returns to measurement view.
readings can be set freely.
is available and plugged in): The measuring cycle (1/5s) and duration (180s) are
preset.
1 Meas.Prog ¬.
2 Press / to select AUTO and confirm with .
OK
OK
The measuring cycle is set in the order: hours /minutes/seconds.
3 Press / to set the measuring cycle in hours and confirm with
OK
.
4 Perform the setting for minutes and seconds as described in step 3.
5 Press / to set the number of readings and confirm with .
- The instrument returns to measurement view.
OK
Page 22
6. Setting the instrument54
Activating tthe TTURB mmeasuring pprogram:
²
The measuring program for measuring turbulence is only available if a
turbulence probe is plugged in.
The main menu is open, Menu is displayed.
1 Meas-Prog ¬.
2 Press / to select Turb and confirm with .
OK
OK
- The instrument returns to measurement view.
6.2.3Mean
The menu item Mean value calculation is only available in the instrument
testo 435-2/-4. In the instrument testo 435-1/-3, the function Mean value
calculation is called up with the function button .
Mean
For carring out Mean value calculation see chapter Measuring, p. 58.
6.2.4P = 0 (435-3/-4 only)
The internal pressure sensor can be zeroed.
The measurement values can be falsified by a change in the position of the
measuring instrument. After zeroing, the position of the measuring instrument must not be changed. Carry out zeroing before every measurement in
order to compensate faulty positioning or long-term zero-point drift. Zeroing
is only possible in a range of 0...25% of the measuring range.
Zeroing tthe iinternal ppressure ssensor:
²
The main menu is open, Menu is displayed.
i P=0¬.
OK
Page 23
6. Setting the instrument 55
6.2.5Calculation
If calculation is switched on, additional parameters with calculated values can
be displayed from the readings of
additional measurement channels in measurement view.
Particular measurement channels must be available in order to perform the
calculation.
Additional calculation parameters need to be set for some calculated variables.
º See the chapter PARAMETER, p. 56.
The following variables can be calculated:
· Flow velocity (435-3/-4 only)
· Volumetric flow rate
· Dew point (below 0°Ctd/32°Ftd frost point temperatures are displayed)
· Psychrometric temperature
· Enthalpy (Heating/cooling performance of assemblies)
The heat transfer coeffiient (alpha) required for the calculation of the U-value can
be set.
It is also possible to calculate the difference between two measurement
channels (Delta). This is only possible if the selected measurement channels
have the same unit.
Activating/deactivating rreading ccalculation:
²
The main menu is open, Menu is displayed.
1 Calc. ¬.
2 Press / to select the variable that is to be activated/deactivated
and confirm with.
3 Press / to select On (= activated) or Off (= deactivated) and
confirm with.
OK
OK
OK
one
probe. These are then displayed as
de
enfresitptsvnl????
Activating ddifferential ccalculation ((delta):
²
The instrument is in measurement view.
Differential calculation is performed with the parameters that are shown on
the display.
1 Press and to select the measurement channels from which the
difference is to be calculated.
2 Press to open the main menu.
3 calc. ¬.
OK
Page 24
6. Setting the instrument56
4 Press / to select Delta and confirm with .
OK
6.2.6Parameter/Area
Some calculated variables relate to particular reference values (ambient
conditions or factors for certain probes). These can be entered by means of
calculation parameters.
Parameters used for calculating variables:
ParametersCalculated measuring variables
Temp. (reference temperature)Flow velocity, volumetric flow rate (for measurement with a
(435-3/-4 only)pitot tube)
Humid. (reference humidity)Flow velocity, volumetric flow rate (for measurement with a
(435-3/-4 only)pitot tube)
Pres. (reference pressure)Flow velocity, volumetric flow rate (for measurement with a
Area (cross-section area)Volumetric flow rate
P-factor (pitot tube factor)Flow velocity and volumetric flow rate (for measurement with a
(435-3/-4 only)pitot tube)
abs alt.Barometric air pressure
Five areas can be stored for the “Area” parameter. Three shapes are defined
in the default setting (one rectangles: edge length a and b, one circle:
diameter d, any shaped area: cross-section area A). The dimensions of the
areas can be adapted in the instrument. It is possible to reassign the shapes
using the PC software (435-2/-4 only).
When K-factor activated (see chapter
DEVICE, p. 45): An offset factor is stored
for each area. If parts of an area are covered (e.g. grilles on ventilation
openings), this can be calculated out using the offset factor. What must be
indicated is the free part of the area (20% covered --> 80% free area -->
offset factor 0.8).
For measurements made at outlets and volume flow regulators with defined
differential pressure measurement points, a component-specific correction
factor (k-Vol) prescribed by the manufacturer can be entered instead of the
area input.
For measurements at ventilation outlets with a funnel the parameter Funnel
must be activated. The funnel set (order no. 0563 4170) consists of a funnel
for measurements at plate outlets (200 x 200mm) and a funnel for
measurements at ventilation outlets (330 x 330mm) in combination with the
testo 435 and with the 100mm vane probe 0635 9435.
de
enfresitptsvnl????
The main menu is open, Menu is displayed.
1 testo 435-2/-4 only: Parameter ¬.
2 Press / to select Area and confirm with.
3 Press / to select the shape by which the area is to be described
and confirm the input with.
4 Press / to set the parameter(s) and confirm each one with .
OK
OK
OK
OK
- The settings are applied and the last shape to be set is activated.
6.2.7Cyclical Print (435-1/-3 only)
The Cyclical Print function can be activated/deactivated. A measuring program
for cyclical printing can be programmed. This enables readings (up to 999) to
be printed in a defined measuring cycle (min. 1min). The readings are sent to a
Testo printer.
Page 26
6. Setting the instrument58
Activating ccyclical pprinting/programming aa mmeasuring pprogram:
²
The main menu is open, Menu is displayed.
1 cyc.Print ¬.
2 Press / to select Off (deactivated) or On (activated) and confirm
with . The measuring cycle is set in the order: minutes/hours.
3 Press / to set the measuring cycle in minutes and confirm with
OK
.
4 Perform the setting for hours as described in step 3.
5 Press / to set the number of readings and confirm with .
OK
OK
OK
- The instrument returns to measurement view.
- The measurement series is programmed and cyclical print can be
started with .
7.Measuring
This chapter describes the steps that are required to perform measurements
with the product.
Particular probes must be plugged in or switched on and registered (radio
probes) according to the variable that is to be measured.
Some probes require a warming-up phase until they are ready to measure.
For some variables additional calculation parameters have to be set if correct
measurements are to be obtained.
º See PARAMETER, p. 56.
For the calculation of the U-value, please refer to the documentation included
with the U-value temperature probe (0614 1635).
Required for the calculation of the heating/cooling performance of assemblies:
· A vane probe (for determining the volume flow) must be connected.
· 2 wireless humidity probes (for determining enthalpy at the input and output
of the assembly and for calculating tightness).
· The parameter pressure (for the calculation of tightness) must be entered.
· The calculation factor enthalpy must be activated. The enthalpy is integrated
in the calculation of performance, an enthalpy value cannot be displayed..
· The wireless humidity probe allocated to radio channel 1 must be positioned
next to the vane probe, as its readings are automatically used to calculate
the mass flow.
Page 27
7. Measuring 59
When measuring ambient CO, observe:
· The ambient CO probe must be situated in fresh air (CO-free) during the
zeroing phase.
· Cigarette smoke influences the measurement by more than 50ppm. The
breathed air of a smoker influences the reading by approx. 5ppm.
· The flow impact of the gas influences the measurement accuracy. Frontal
impact on the sensor leads to higher measurement values. The best
measurement results are acheived with a slight movement of the probe back
and forth.
Taking aa mmeasurement:
²
The instrument is in measurement view.
The measuring program AUTO or TURB is not activated (435-2/-4 only).
i Put the probe in position and take the readings.
- The following are displayed in consecutive order:
· Available measurement channels
· Maximum value of the variable in the upper display line
· Minimum value of the variable in the upper display line
· Lower measurement line not shown
Resetting mmax./min. vvalues:
²
The minimum or maximum values of all measurement channels are reset.
1 Press several times until the maximum or minimum value is displayed.
2 Reset the max./min. values with .
Reset
de
enfresitptsvnl????
Holding rreadings:
²
i Press .
i Press to change back to displaying the actual reading.
Saving rreadings ((435-22/-44 oonly):
²
i Press .
Hold
Act
Save
- A measurement protocol with the readings of all available measurement
channels is created for the active location.
Page 28
7. Measuring60
Timed mmean ccalculation:
²
The mean is formed as a moving mean value and individual values are not
displayed.
1 435-1/-3: Press , 435-2/-4: ¬ Mean ¬.
2 Timed ¬.
3 Press to start mean calculation.
Press to stop mean calculation.
Multi-ppoint mmean ccalculation:
²
OK
Start
End
OKMEAN
The mean is formed as a moving mean value.
1 435-1/-3: Press , 435-2/-4: ¬ Mean ¬.
2 Multi-poi ¬.
3 Press to include readings.
Press to stop mean calculation.
OK
Pick
End
OKMEAN
Only testo 435-2/-4 in the profile Duct measurement and with hole query
activated (see chapter
4 With / , enter number of holes and confirm with .
Running tthe AAUTO oor TTURB mmeasuring pprogram ((435-22/-44 oonly):
²
The instrument is in measurement view and the AUTO or TURB measuring
DEVICE, p. 45):
OK
program is activated.
1 Start the measuring program with .
Start
- The measuring program starts. The readings are recorded.
- The measuring program continues to run until cancelled with or
End
until the end criterion is met (number of readings is reached or time has
expired when measuring turbulence ).
- The readings are saved in a protocol.
Cyclical pprinting ((435-11/-33 oonly):
²
The instrument is in measurement view and Cyclical Print is activated.
i Start cyclical printing with .
- The measuring program starts. The readings are transmitted to the
Testo printer.
- Measurement continues to run until cancelled with or until the
End
end criterion is met (number of readings is reached).
Page 29
8. Care and maintenance 61
8.Care and maintenance
This chapter describes the steps that help to maintain the functionality of the
product and extend its service life.
Cleaning tthe hhousing:
±
i Clean the housing with a moist cloth (soap suds) if it is dirty. Do not use
aggressive cleaning agents or solvents!
Changing tthe bbattery/rechargeable bbattery:
±
Instrument is switched off.
1 Undo the two screws on the rear of the instrument and lift off the battery
compartment cover.
2 Remove spent batteries/rechargeable batteries and insert new
batteries/rechargeable batteries (3x mignon) into the battery
compartment. Observe the polarity!
3 Replace the battery compartment cover and tighten the two screws.
de
enfresitptsvnl????
Page 30
9. Quenstions and answers62
9.Questions and answers
This chapter gives answers to frequently asked questions.
QuestionPossible causesPossible solution
lights up· Instrument battery is almost spent.· Replace instrument battery.
Instrument switches ·
off automatically.· Residual capacity of the battery is too low. · Replace battery.
-----· Probe is not plugged in.· Switch instrument off, connect probe
Display:
uuuuu· Permitted measuring range· Keep to permitted measuring range.
Display:
ooooo· Permitted measuring range· Keep to permitted measuring range.
Display:
Instrument settings are· Power supply was interrupted for a · Re-enter instrument settings.
no longer correctlong time.
If we are unable to answer your question, please contact your dealer or Testo
Customer Service. Contact details can be found on the guarantee card or on
the Internet under
Auto Off function is switched on.· Switch function off.
· Radio contact with radio probe is· Switch radio probe on, if necessary
interrupted.register radio probe again.
Memory435-2/-4 only: max. 99 locations, up to 10000 readings (depending on number of
Battery life160h (typical for vane measurement)
Power supply3x mignon battery (included in delivery)/rechargeable battery or mains unit
Housing materialABS/TPE/metal
Protection classIP65
Dimensions225 x 74 x 46mm
Operating temperature range -20...+50°C
Storage temperature-30...+70°C
Measuring rate2/s
EC Directive2004/108/EEC
WarrantyInstrument: 2 years
435-3/-4 only: 2x pressure nipple
locations, protocols, channels)
(accessory part)
Page 33
11. Accessories/spare parts 65
11.Accessories/spare parts
This chapter gives important accessory and spare parts for the product.
NamePart no.
Probes
Water-proof immersion/penetration probe,TC type K0602 1293
Water-proof surface probe with widened measurement tip for smooth surfaces,TC type K0602 1993
Robust affordable air probe, TC type K0602 1793
Vane probe, 1000mm diameter0635 9435
Vane probe, 60mm diameter, telescopic to max. 910mm0635 9335
Vane probe, 16mm diameter, telescopic to max. 890mm0635 9335
Humidity/temperature probe, 12mm diameter (testo 435-2/-4 only)0636 9735
Handle for humidity/temperature probe, for connection to the measuring instrument,
including probe cable, for measures/calibrates of humidity probehead (testo 435-2/-4 only)0430 9735
Absolute pressure probe 2000hPa0638 1835
Pressure dew point probe for measurements in compressed air systems (testo 435-2/-4 only)0636 9835
Hot wire probe for m/s and °C, probehead 7.5mm diameter, telescopic to max. 820mm0635 1025
IAQ probe to assess Indoor air quality, CO“,humidity, temperature and
absolute pressure measurement0632 1535
Thermal flow probe with integrated temperature and humidity measurment,
12mm diameter,with telescope max. 745mm0635 1535
Comfort probe for degree of turbulence measurement, with telescopic handle and stand,
fulfills DIN EN 13779 requirements (testo 435-2/-4 only)0628 0109
Lux probe, probe for measuring luminous intensity (testo 435-2/-4 only)0635 0545
Ambient CO probe0632 1235
Miscellaneous
Plug-in mains unit, 5VDC, 500mA with European plug0554 0447
Funnel set sonsisting of funnel for plate outlets and funnel for vents.0563 4170
External recharger incl. 4 Ni-MH rechargeable batteries with built-in, international plug,