Testo 420 Operating Manual

Page 1
testo 420 · Flow Hood
Instruction manual
Page 2
2
Page 3

1 Contents

1 Contents
1 Contents ................................................................................................... 3
2 Safety and the environment .................................................................... 4
2.1. About this document ........................................................................ 4
2.2. Ensure safety................................................................................... 5
2.3. Protecting the environment .............................................................. 5
3 Specifications .......................................................................................... 5
3.1. Use .................................................................................................. 6
3.2. Technical data ................................................................................. 6
3.2.1. Bluetooth module ................................................................................................ 6
3.2.2. General technical data ........................................................................................ 7
4 Product description ............................................................................... 10
4.1. Overview ....................................................................................... 10
4.1.1. Measurement setup ..........................................................................................10
4.1.2. Overview of testo 420 .......................................................................................11
5 First steps .............................................................................................. 12
6 Using the product .................................................................................. 14
6.1. Switching Bluetooth® on and off.................................................... 14
6.2. Settings for the measurement........................................................ 15
6.2.1. Damping (Gliding Average) ..............................................................................15
6.2.2. Hood Calibration ...............................................................................................15
6.2.3. Zeroing Interval (Automatic Zeroing) ................................................................16
6.3. Volume flow measurement setup .................................................. 16
6.4. Measuring ...................................................................................... 19
6.4.1. Volume flow measurement ...............................................................................19
6.4.2. Pitot tube measurement ....................................................................................19
6.4.3. Differential pressure measurement...................................................................19
6.5. Saving ........................................................................................... 20
6.6. Transferring measurement data to the PC .................................... 21
7 Maintaining the product ........................................................................ 22
7.1. Cleaning the instrument ................................................................. 22
8 Tips and assistance............................................................................... 22
8.1. Questions and answers ................................................................. 22
8.2. Accessories and spare parts ......................................................... 23
9 EC Declaration of Conformity ............................................................... 24
3
Page 4

2 Safety and the environment

Representation
Explanation
Menu
2 Safety and the environment

2.1. About this document

Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay particular attention to the safety instructions and warning advice in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.

Symbols and writing standards

Warning advice, risk level according to the signal word:
Warning! Serious physical injury may occur. Caution! Slight physical injury or damage to
the equipment may occur. > Implement the specified precautionary
measures.
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Elements of the instrument, the instrument display or the program interface.
[OK]
Control keys of the instrument or buttons of the program interface.
... | ... Functions/paths within a menu.
“...” Example entries
4
Page 5

2.2. Ensure safety

> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
> Do not perform contact measurements on non-insulated, live
parts.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the prescribed steps exactly. Use only original spare parts from Testo.
> Dangers may also arise from the systems being measured or
the measuring environment: Note the safety regulations valid in

2.3. Protecting the environment

your area when performing the measurements.
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal.

3 Specifications

3 Specifications
The testo 420 is used for volume flow measurements (main application), Pitot tube measurements and pressure measurements for air conditioning and ventilation systems. Thanks to its interchangeable flow hoods, the testo 420 can be used for air inlets and outlets of various sizes.
With the additional App (Android/iOS), the readings can be conveniently displayed on a tablet or smartphone, and in addition a measurement can be started, stopped and saved via the App.
5
Page 6
3 Specifications

3.1. Use

3.2. Technical data

The use of the wireless module is subject to the regulations and stipulations of the respective country of use, and the module may only be used in countries for which a country certification has been granted. The user and every owner has the obligation to adhere to these regulations and prerequisites for use, and acknowledges that the re­sale, export, import etc. in particular in countries without wireless permits, is his responsibility.

3.2.1. Bluetooth module

Feature Values
Bluetooth Range >20 m (free field)
Bluetooth type LSD Science & Technology Co., Ltd
L Series BLE module (08 May 2013) based on TI CC254X chip
Qualified Design ID B016552
Bluetooth radio class
Bluetooth company 10274
Class 3
Certification
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE), Estonia (EE), Finland (FI), France (FR), Greece (GR), Ireland (IE), Italy (IT), Croatia (HR), Latvia (LV), Lithuania (LT), Luxembourg (LU), Malta (MT), Netherlands (NL), Austria (AT), Poland (PL), Portugal (PT), Romania (RO), Sweden (SE), Slovakia (SK), Slovenia (SI), Spain (ES), Czech Republic (CZ), Hungary (HU), United Kingdom (GB), Republic of Cyprus (CY).
EFTA countries
Norway
Other countries
USA, Turkey, India and Hong Kong
6
Page 7
Information from the FCC (Federal Communications Commission) This instrument complies with part 15 of the FCC Rules. Its
commissioning is subject to the following two conditions: (1) This instrument may not cause harmful interference and (2) this instrument must be able to accept interference, even if this could have undesired effects on the operation.
Changes The FCC demands that the user be informed that any changes or
modifications to the instrument that are not explicitly approved by Testo AG may void the user's right to use this instrument.

3.2.2. General technical data

Feature Values
Measurement parameters
Measuring cycle 1/sec
Interfaces Probe interface Mini DIN
Measuring ranges Temperature: -20 to +60 °C / -4 to 140 °F
Temperature: °C / °F Humidity: %RH / %rF / td°C / WB°C Flow velocity : m/s / ft/min Volume flow: m³/h / cfm / l/s Pressure (absolute pressure): hPa / mbar / kPa Pressure (differential pressure): Pa / hPa /
mbar / mmH
Micro USB
Humidity: 0 to 100% RH
Flow velocity: 0 to 14 m/s / 0 to 2750 ft/min
Volume flow: 40 to 4000 m³/h / 25 to 2300 cfm / 11 to 1100 l/s
Pressure (absolute pressure): 700 to 1100 hPa
Pressure (differential pressure): ­120 to +120 Pa
O / inH2O
2
3 Specifications
7
Page 8
3 Specifications
since
Feature Values
Resolution Temperature: 0.1 °C / 0.1°F
Humidity: 0.1% RH Flow velocity: 0.01 m/s Volume flow: 1 m³/h / 1 cfm Pressure (absolute pressure): 0.1 hPa /
0.1 mbar / 0.01 kPa Pressure (differential pressure): 0.001 Pa /
0.00001 hPa / 0.00001 mbar / 0.0001 mm H2O / 0.000001 inH2O
Accuracy (nominal temperatur e 22°C/71.6°F)
Temperature: ±0.5 °C (0 to +70 °C) / ±0.8 °C (-20 to 0 °C)
Humidity: ±1.8% RH +3% of m.v. at +22 °C (5 to 80% RH) (longer high humidity application can result in a temporary sensor drift)
Flow velocity: no accuracy specification calculated variable
Volume flow 1: ±3% of m.v. +12 m³/h at +22 °C, 1013 hPa (85 to 3500 m³/h)
±3% +7 cfm (50 to 2,100 cfm) Error absolute pressure compensation:
±0.04% of m.v. / hPa deviating from 1013 hPa
Pressure (absolute pressure): ± 3 hPa
Pressure (differential pressure): ±2% of m.v. + 0.5 Pa (at 22°C, 1013 hPa)
Error absolute pressure compensation: ±0.04% of m.v. / hPa deviating from 1013 hPa
1
All accuracy specifications apply under laboratory conditions or with necessary compensation (correction factor) with the standard hood 610x610 mm. Minimum outlet size of 360x360 mm.
8
Page 9
3 Specifications
Feature Values
Temperature coefficient
Humidity: ± 0.03% RH / K (deviating from 22°C, in the range 0 to 60 °C)
Volume flow: ± 0.02% of m.v. / K (deviating from 22 °C, in the range 0 to 60 °C)
Pressure (absolute pressure): ± 0.02% of m.v. / K (deviating from 22 °C, in the range 0 to 60 °C)
Pressure (differential pressure): ± 0.02% of m.v. / K (deviating from 22 °C, in the range 0 to 60 °C)
Response time t90 Temperature: approx. 45 s
Humidity: approx. 15 s Flow velocity: approx. 1 s Volume flow: approx. 1 s Pressure (absolute pressure): approx. 1 s Pressure (differential pressure): approx. 1 s
Operating and ambient conditions
Storage temperature: -20 to +60 °C / -4 to 140 °F
Operating temperature: -5 to +50 °C/+23 ­+122°F
Humidity: 0 to 100% RH Pressure range: 800 to 1100 hPa
Housing / measurement setup
Measuring instrument housing material: ABS Body material: PP Standard hood material: Nylon Measuring instrument dimensions:
150x85x35 mm Body dimensions: 490x970x610 mm Dimensions of measurement setup with
standard hood: 610x970x610 mm Weight of entire measurement setup
approx. 2900 g
Power supply 4 x 1.5 V rechargeable/non-rechargeable
batteries Type AA / alkaline manganese, mignon Battery lifetime: approx. 40h (eroing interval
10 seconds, display illumination off, Bluetooth off)
9
Page 10

4 Product description

Feature Values
Display Type: Dot matrix
Dimensions: 3.5 inches
Directives, standards and tests
Warranty Duration: 2 years
EU guideline: 2014/30/EU
Warranty terms: see website www.testo.com/warranty
4 Product description

4.1. Overview

4.1.1. Measurement setup

1 Volume flow hood (standard hood 610x610 mm) 2 Actuator for manual measurement 3 Measuring instrument testo 420 4 Measurement base with differential pressure cross 5 Integrated flow straightener
10
Page 11

4.1.2. Overview of testo 420

1 Battery compartment, on rear of the instrument 2 Display 3 Control keys 4 Probe socket Mini-DIN (only for use on the measurement base) 5 Micro-USB port 6 Connection for pressure measurement
Instrument status icons:
Icon Meaning
Battery capacity
/ /
Actual
Standard
Bluetooth
Measuring mode: Pressure measurement, Pitot tube, volume
flow (air pressure from above into the hood / suction outlet)
Actual volume flow: The current ambient conditions are used to
calculate the volume flow. The actual barometric pressure is measured with the internal sensor. In applications with volume flow hood, the temperature is measured by the integrated temperature/humidity sensor, with Pitot tube measurement the actual temperature needs to be input manually.
Standard volume flow: The standard settings for temperature and
barometric pressure (21 °C / 1013 hPa) are used to calculate the volume flow.
4 Product description
11
Page 12

5 First steps

instruction manual or you should ask your
Icon Meaning
K-factor
Factor by which the current reading is multiplied. Depends on which outlet measurements are being taken at.
Pitot Tube factor
(P-factor)
The Pitot tube factor for Pitot tubes is generally the same and must be entered:
Pitot tubes from Testo: 1.00
Pitot tubes from other manufacturers: the
Pitot tube factor can be found in the
supplier.
Control keys
Ke y Function
Menu
[►, ■]
[ESC]
[►, ▲, ▼,◄]
[ ]
5 First steps

Inserting batteries/rechargeable batteries

1. Open the battery compartment.
2. Insert the batteries or rechargeable batteries (scope of delivery includes 4x 1.5V Type AA/ LR6).
3. Close the battery compartment.
When not in use for a long period of time, take the batteries/rechargeable batteries out.
Holds / starts / stops a measurement
Switches to the previous view / to the measurement view
Saves the measured values
Navigation within the menu
Confirms a selection
Switch instrument on/off (press and hold down)
Switch illumination on/off (press briefly)
12
Page 13
5 First steps
►, ▲, ▼,◄
measurement (nur für

Implementing settings

1. Press to access the menu.
2. Select the menu item you require using ►, ▲, ▼, ◄.
Key functions
Display Explanation
Change parameter, select unit
Confirm input

adjustable parameters

1. Menu Level 2. Menu Level 3. Menu Level
Application Flow Hood K-factor
Actual/Standard
Pitot tube Duct
Pitot tube factor
Pitot Temperature
Actual/Standard
Pressure only --
Measuring Program Single Measurment --
Continous Measurement
2
Continous/Punctual
--
measuring duration
3
2
Maximum of 15 minutes
3
Maximum of 25 points and 1 minute per point.
13
Staurohr)
Memory New Folder --
T420 folder --
Display Volume Flow on/off
Diff.Pressure on/off
Temperature on/off
Page 14

6 Using the product

Information about compatibility can be found in the relevant
1. Menu Level 2. Menu Level 3. Menu Level
Device Settings Language English/German/Italia
Reset Device -- --
Velocity on/off
Humidity on/off
Abs.Pressure on/off
n/French/Spanish
Backlight Auto Off On/Off
Auto Off On/Of
Bluetooth On/Off
Date&Time Date Format
Time Format
Set Date&Time
Gliding average 5 – 20 sec
Hood adjustment Supply Air
Exhaust Air
Zeroing int 1-20 sec
6 Using the product

6.1. Switching Bluetooth® on and off

In order to be able to establish a connection via Bluetooth, you need a tablet or smartphone with the Testo App testo 420 already installed on it.
You can get the App for iOS instruments in the App Store or for Android instruments in the Play Store.
App Store.
14
Page 15
6 Using the product
Display
Explanation
Measurements can be performed and saved with App into the instruments memory. Measuring mode and save function on the instrument itself is not available during Bluetooth communication is established with a mobile device.

Switching on Bluetooth

1. Hold down for 3 seconds.
- Once the Bluetooth icon is shown on the display, Bluetooth is switched on.
- If no connection is established, Bluetooth switches off after 10 mins.
or
1. Press -> Device Settings -> Bluetooth, ► and using ▲/
-> select Off. Confirm with .
- Once the Bluetooth icon is shown on the display, Bluetooth is switched on.
- If no connection is established, Bluetooth switches off after 10 mins.

6.2. Settings for the measurement

6.2.1. Damping (Gliding Average)

6.2.2. Hood Calibration

15
flashes
is permanently
There is no Bluetooth connection, or a potential connection is being searched for.
There is a Bluetooth connection.
displayed
is not displayed
Bluetooth is disabled.
If the readings fluctuate widely, it is advisable to damp the readings. The time range for the damping can be set manually between 5-20 seconds.
1. Press , then select Device settings and Gliding average.
- The damping can be set between 5-20 seconds.
This input is provided for the filing of calibration data by the relevant calibration laboratory. The flow hood-specific adjustment data can
Page 16
6 Using the product
be input manually for supply air and exhaust air and has a direct effect on the measurement results. Input option of 0.001-9.999.
1. Press , then select Device settings and Hood adjustment.
- Calibration of the hood can be set for supply air and exhaust air.

6.2.3. Zeroing Interval (Automatic Zeroing)

The pressure sensor carries out automatic zeroing at regular intervals. These intervals can be set via the automatic zeroing.
1. Press , then select Device settings and Zeroing int.

6.3. Volume flow measurement setup

- The zeroing interval can be set between 1-20 seconds.

Standard hood (610x610 mm, scope of delivery; 360x360, accessories)

16
Page 17
6 Using the product
1. Pull the lower end of the hood over the measurement base.
2. Attach the hood at two corners using the snap fasteners.
3. Tighten the closure.
4. Push the support rods through the hood, along the markings and into the funnel in the measurement base.
5. Push the support rods on the top of the hood into the brackets.
- The hood is installed.

Large hood (1220x610 and 1220x305, accessories)

1. Install the aluminium frame and stretch the fabric hood over the framework, so that the elastic band lies in the recess of the frame. Make sure that the elastic band fits correctly, especially at the corners.
2. Pull the lower end of the hood over the measurement base.
3. Attach the hood at two corners using the snap fasteners.
4. Tighten the closure.
17
Page 18
6 Using the product
5. Push the support rods through the hood, along the markings
6. Push the support rods on the top of the hood into the brackets.
- The hood is installed.

Attaching the measuring instrument

and into the funnel in the measurement base.
1. Push the testo 420 completely into the instrument holder, pay
attention to the catch on the right-hand and left-hand side in the bracket.
18
Page 19

6.4. Measuring

6.4.1. Volume flow measurement

The volume flow hood is fitted.
1. Switch on the instrument.
2. In the instrument settings, set the application volume flow hood and also the required measurement program: Single measurement or Continuous measurement.
3. Press ►, ■ on the testo 420, or the trigger on the measurement base to hold or start and stop the measurement.
4. Press to save the measurement data. Unsaved measurement data is lost once you carry out the next measurement.
- The target folder and file name are displayed, confirm with to save the measurement data under this name and in the selected folder.

6.4.2. Pitot tube measurement

1. Remove the testo 420 from the measurement base.
2. Connect the hoses to the testo 420 and to the Pitot tube.
3. Press -> Application -> Pitot tube and there set the duct geometry, the Pitot tube factor and the temperature, and choose between actual and standard.
4. Select the required measurement program.
5. Carry out the measurement.
6. Press to save the measurement data. Unsaved measurement data is lost once you carry out the next measurement.
- The target folder and file name are displayed, confirm with to save the measurement data under this name and in the selected folder.
6 Using the product

6.4.3. Differential pressure measurement

1. Remove the testo 420 from the measurement base.
2. Connect the hoses to the testo 420 at + and -.
3. Press -> Application -> Pressure only.
4. Carry out the measurement.
19
Page 20
6 Using the product
5. Press to save the measurement data. Unsaved
measurement data is lost once you carry out the next measurement.
- The target folder and file name are displayed, confirm with
to save the measurement data under this name and in the selected folder.

6.5. Saving

A maximum of 99 measurements can be saved in one folder.
> -> Memory ->
- The folder overview is shown on the display. Create a new
folder via New Folder.

Open folder

> Using the arrow keys, navigate to the required folder and press
.
- The selected folder is opened and the individual files are
displayed.

Delete folder

1. Using the arrow keys, navigate to the required folder and press
.
2. Select the menu item Delete Folder and confirm with .
- The folder that you want to delete is shown on the display.
3. Confirm once again with in order to delete the folder,
or cancel by pressing Esc.
20

Set as Logging Folder

This setting establishes which folder is to be specified as the standard storage location for saving the measurements.
The folder which is set as the standard storage location is marked in black.
1. Using the arrow keys, navigate to the required folder and press
.
2. Select the menu item Set as Logging Folder and confirm with
.
- The selected folder is set as the standard storage location.
- The stored location can be changed during the saving process.
Page 21
6 Using the product
the total volume flow of all measurements can be displayed

Total Volume Flow

If data from single measurements is saved within a folder,
using this function.
1. Using the arrow keys, navigate to the required folder and press
.
2. Select the menu item Total Volume Flow and confirm with .
- The single measurements and the total volume flow are displayed.
3. Press .
- The result of the total volume flow is saved.

6.6. Transferring measurement data to the PC

No Bluetooth connection is possible while there is a connection to the PC. The existing Bluetooth connection is aborted.
1. Connect the testo 420 to your PC using the USB cable.
- The testo 420 switches on automatically, a window appears on the PC, select Open folder here. The folders and files saved on your testo 420 are displayed. The files are available in the file format *.txt.
21
Page 22

8 Tips and assistance

Can not
available in Bluetooth
instrument is connected to a Tablet

7 Maintaining the product

7.1. Cleaning the instrument

Do not use any aggressive cleaning agents or solvents! Mild household cleaning agents or soap suds may be used.
> If the housing of the instrument is dirty, clean it with a damp
8 Tips and assistance

8.1. Questions and answers

cloth.
Question Possible causes/solution
For selected parameters, no values are shown on
No temperature/humidity probe is
connected, for instance. the instrument display (-----)
Warning message
turn on! when selecting
certain parameters in the Display menu.
This parameter is not available for the application currently selected.
Four parameters are already displayed. Disable the display of one parameter to enable another one.
[►, ■] on the
Bluetooth connection is active, the
instrument does not work.
Warning message
Function not
or Smartphone via Bluetooth and the App is active.
Fully close the App or end the Bluetooth connection.
mode.
22
Page 23

8.2. Accessories and spare parts

free, for differential
Description Article no.
Flow hood 360x360 mm 0554 4200
Flow hood 305x1220 mm 0554 4201
Flow hood 610x1220 mm 0554 4202
Fabric cover for the hood 610x610 0400 4200
Aluminium frame for the hood 610x610 0440 4204
Tripod, extendable to 4 m 0554 4209
Connection hose, silicone, length 5 m, maximum load capacity 700 hPa (mbar)
Connection hose, silicone­pressure measurement, length 5 m, maximum load capacity 700 hPa (mbar)
Pitot tube, length 500 mm, Ø 7 mm, stainless steel, for measuring flow velocity (Connection hose required)
Pitot tube, length 350 mm, Ø 7 mm, stainless steel, for measuring flow velocity (Connection hose required)
Pitot tube, 1,000 mm long, stainless steel, measures flow velocity (Connection hose required)
Connection hose 0554 0453
Tension rod 0440 4201
8 Tips and assistance
0554 0440
0554 0453
0635 2045
0635 2145
0635 2345
For a complete list of all accessories and spare parts, please refer to the product catalogues and brochures or visit our website www.testo.com
If you have any questions, please contact your dealer or Testo Customer Service. The contact details can be found on the back of this document or on the Internet at www.testo.com/service- contact..
23
Page 24

9 EC Declaration of Conformity

9 EC Declaration of Conformity
24
Page 25
Page 26
0970 4200 en 01 V01.00
Loading...