7.10. Checking/replacing the particle filter .............................................. 62
8 Tips and assistance ............................................................................... 63
8.1. Questions and answers ................................................................. 63
8.2. Accessories and spare parts ......................................................... 64
8.3. Updating the instrument software .................................................. 68
5
Page 6
2 Safety and the environment
2 Safety and the environment
2.1. About this document
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Warnings
Always pay attention to information that is marked by the following
warnings with warning pictograms. Implement the specified
precautionary measures.
Representation Explanation
WARNING
CAUTION
NOTICE
Indicates potential serious injuries
indicates potential minor injuries
indicates circumstances that may lead to
damage to the products
6
Page 7
2 Safety and the environment
Symbols and writing standards
Represen-
Explanation
tation
Note: Basic or further information.
1. ...
Action: more steps, the sequence must be followed.
2. ...
> ... Action: a step or an optional step.
- ... Result of an action.
[OK] Control keys of the instrument or buttons of the
2.2.Ensure safety
program interface.
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
> Do not perform contact measurements on non-insulated, live
parts.
> The testo 320 is not suitable for long-term measurements and
should not be used as a safety (alarm) instrument.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
Testo.
> Any further or additional work must only be carried out by
authorised personnel. Testo will otherwise refuse to accept
responsibility for the proper functioning of the measuring
instrument after repair and for the validity of certifications.
> Only use the device in closed, dry rooms and protect it from rain
and moisture.
> Temperatures given on probes/sensors relate only to the
measuring range of the sensors. Do not expose handles and
feed lines to any temperatures in excess of 70 °C unless they
are expressly permitted for higher temperatures.
7
Page 8
2 Safety and the environment
> The testo 320 must be checked before commissioning for any
visible damage. Do not commission the testo 320 if there are
signs of damage on the housing, mains unit or supply lines.
Electrical risk.
> The objects to be measured or the measurement environment
may also pose risks: Note the safety regulations valid in your
area when performing the measurements.
For products with Bluetooth® (optional)
Changes or modifications that have been made without the explicit
consent of the responsible approval authority, may cause the
retraction of the type approval.
Data transfer may be disturbed by equipment that uses the same
ISM-band, e.g. WLAN, microwave ovens, ZigBee.
The use of radio communication links is not permitted, among
others, in aeroplanes and hospitals. For this reason the following
points must be ensured before entering:
> Switch off the device:
> Isolate the device from any external power sources (mains
2.3. Protecting the environment
cable, external rechargeable batteries, ...).
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
8
Page 9
3 Specifications
3.1. Use
The testo 320 is a handheld measuring device for the professional
flue gas analysis of combustion plants:
• Small combustion plants (burning oil, gas, wood, coal)
• Low-temperature and condensing boilers
• Gas water heaters
These systems can be adjusted using the testo 320 and checked
for compliance with the applicable limit values.
The following tasks can also be carried out with the testo 320:
• Regulating the O2, CO and CO2 values in combustion plants
for the purpose of ensuring optimal operation.
• Draught measurement.
• Measuring and regulating the gas flow pressure in gas water
heaters.
• Measuring and optimising the flow and return temperatures of
heating systems.
• Ambient CO measurement (only possible with additional
ambient CO probe 0632 3331).
• Ambient CO measurement (only possible with additional
ambient CO probe 0632 1240).
• Detection of CH4 (methane) and C3H8 (propane) (only possible
with additional gas leak probe 0632 3370).
The Bluetooth
is type approved.
®
option may only be operated in countries in which it
3 Specifications
9
Page 10
3 Specifications
3.2. Technical data
3.2.1. Examinations and licenses
As declared in the certificate of conformity, this product complies
with Directive 2004/108/EC.
This product is TÜV-tested in compliance with 1. BImSchV. The
sensors 0393 0105 (CO, H2-compensated), 0393 0003 (O2),
temperature and pressure are TÜV-tested in accordance with EN
50379 part 2.
The measuring cell 0393 0053 (CO, not H2-compensated) is TÜVtested as per EN 50379 part 3.
This product is EMC-tested as per DIN EN 61326-1.
For official measurements in accordance with 1. BImSchV
(chimney sweeps), the measuring instrument must be checked
every six months by a technical testing body of the Guild of Master
Chimney Sweeps or another testing body recognised by the
authorities.
3.2.2. Bluetooth® module (option)
• Bluetooth® type: BlueGiga WT11
• Bluetooth
• Bluetooth
• Bluetooth
®
product note: WT11
®
identification: B011198
®
company: 10274
10
Certification
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE),
Estonia (EE), Finland (FI), France (FR), Greece (GR), Ireland (IE),
Italy (IT), Latvia (LV), Lithuania (LT), Luxembourg (LU), Malta (MT),
Netherlands (NL), Austria (AT), Poland (PL), Portugal (PT),
Romania (RO), Sweden (SE), Slovakia (SK), Slovene (SI), Spain
(ES), Czech Republic (CZ), Hungary (HU), United Kingdom (GB),
Republic of Cyprus (CY).
Iceland, Liechtenstein, Norway and Switzerland.
Turkey, El Salvador, Ecuador
Page 11
3 Specifications
Information of the FCC (Federal Communications
Commission)
This device fulfils part 15 of the FCC-guidelines. Commissioning is
subject to the two following conditions: (1) This device must not
generate any dangerous interferences and (2) this device must be
able to receive interferences, even if these could have undesired
effect on the operation.
Changes
The FCC demands that the user is to be informed that with any
changes and modifications to the device, which have not been
explicitly approved by testo AG, the right of the user to use this
device will become null and void.
11
Page 12
3 Specifications
3.2.3.Declaration of Conformity
12
Page 13
3.2.4.Measurement ranges and resolution
Measurement
parameter
O2 0 to 21 Vol.% 0.1 vol.%
CO 0...4000 ppm 1 ppm
CO, H2-comp.
COlow, H2-comp. 0 to 500 ppm 0.1 ppm
Draught1 -9.99 to 40.00 hPa 0.01 hPa
Fine draught1
∆P (only with gas
pressure set 0554
1203)
Fine pressure1
(only with gas
pressure set 0554
1203)
temperature -40 to 1200°C 0.1°C (-40.0 to 999.9°C)
Efficiency 0 to 120 % 0.1 %
Flue gas loss 0 to 99.9 % 0.1 %
Measuring range Resolution
0 to 8000 ppm 1 ppm
-9.999 hPa to
+40.000 hPa
0 to 300 hPa 0.1 hPa
0 to 300 hPa 0.01 hPa
0.001 hPa
1°C (from 1000°C)
3 Specifications
1
Depending on the country version
13
Page 14
3 Specifications
3.2.5.Accuracy and response time
Measurement
parameter
O2 ±0.2 vol.% < 20 s
CO
CO, H2-comp.
COlow, H2-comp. ±2 ppm (0 to 39.9 ppm)
Draught1
Fine draught1
∆P (only with gas
pressure set 0554
1203)
Accuracy
±20 ppm (0 to 400 ppm)
±5% of meas. val. (401 to 2000
ppm)
±10% of meas. val. (2001 to
4000 ppm)
±10 ppm or ±10% of meas. val.
(0 to 200 ppm)
±20 ppm or ±5% of meas. val.
(201 to 2000 ppm)
±10% of meas. val. (2001 to
8000 ppm)
±5% of meas. val. (rest of range)
±0.02 ppm or ±5% of meas. val.
(-0.50 to 0.60 hPa)
± 0.03 hPa (0.61 to 3.00 hPa)
±1.5% of meas. val. (3.01 to
40.00 hPa)
±0.02 ppm or ±5% of meas. val.
(-0.50 to 0.60 hPa)
± 0.03 hPa (0.61 to 3.00 hPa)
±1.5% of meas. val. (3.01 to
40.00 hPa)
± 0.5 hPa (0.0 to 50.0 hPa)
±1% of meas. val.
(50.1 to 100.0 hPa)
±1.5% of meas. val. (rest of
range)
Response
time (t
< 60 s
2
< 40 s
2
< 40 s
2
-
2
-
)
90
2
higher value is valid
14
Page 15
3 Specifications
Measurement
Accuracy
parameter
Fine pressure1
(only with gas
pressure set 0554
1203)
± 0.5 hPa (0.0 to 50.0 hPa)
±1% of meas. val.
(50.1 to 100.0 hPa)
±1.5% of meas. val. (rest of
range)
temperature ± 0.5°C (0.0 to 100.0°C)
±0.5% of meas. val. (rest of
range)
Efficiency - Flue gas loss - -
3.2.6.Other instrument data
Feature Values
Storage and
-20 to 50°C
transport
temperature
Operating
-5 to 45°C
temperature
Ambient humidity
Power supply
0…90 % rH, not condensing
Rech. batt.: 3.7 V / 2.4 Ah
Mains unit: 5.0 V/1000 mA
Protection class IP40
Weight 573 g
Dimensions 240 x 85 x 65 mm
Memory 500 measured values
Display Graphic colour display, 240 x 320 pixels
Gas leak testing
probe
Optimum rech. batt.
storage conditions
visual indication (LED)
audible indication by buzzer
Charge level: capacity at 50-80% ambient
temperature: 10-20°C
Battery charge time Approx. 5-6 h with mains unit supplied
Rechargeable
battery life
Approx. 6 h (pump on, 20°C ambient
temperature)
Data transfer IrDA, USB, Bluetooth® (option)
Response
time (t
)
90
depending on
the probe
15
Page 16
3 Specifications
Feature Values
Bluetooth® (option) Range < 10 m
Warranty Measuring instrument: 24 months
O2 sensor: 24 months
CO sensor: 24 months
CO sensor with H2 comp.: 24 months
CO/H2 low sensor (TCHL): 24 months
Flue gas probe: 24 months
Thermocouple: 12 months
Rech. batt.: 12 months
Terms of warranty
Terms of warranty: see website
www.testo.com/warranty
16
Page 17
4 Product description
4.1. Measuring instrument
4.1.1. Front view
4 Product description
1 Display
2 Function keys
17
3 Keypad
Page 18
4 Product description
4.1.2.Keypad
Button Functions
[]
[OK]
Example
[▲]
[▼]
[esc]
[]
[]
4.1.3.Display
Switch measuring instrument on / off
Function key (orange, 3x), relevant function is shown on
the display
Scroll up, increase value, navigate
Scroll down, reduce value, navigate
Back, cancel function
Open main menu
Transmit data to the Testo protocol printer.
18
1 Status bar (dark grey background):
• Warning symbol (only if there is an instrument error,
display of error in instrument diagnosis menu), otherwise:
Instrument designation.
• Symbol (only if data is stored in the temporary memory).
• Display of date and time.
• Indication of Bluetooth
®
status, power supply and remaining
rechargeable battery capacity:
Icon Feature
blue symbol = Bluetooth
grey symbol = Bluetooth
®
®
on,
off
Page 19
Icon Feature
Battery operation
Display of remaining rechargeable battery
capacity by colour and fill level of the battery icon
(green = 5-100%, red = < 5%)
Mains operation
Display of remaining rechargeable battery
capacity: see above
2 Info field of register tabs: Indication of selected folder/location,
chosen fuel, chosen measurement type.
3 Selection field for functions (selected function appears against a
white background, unavailable functions are identified by grey
font) or display of measured values.
4.1.4. Instrument connections
4 Function display for function keys.
4 Product description
1 Probe socket
2 Gas outlet
3 Probe socket
4 Micro USB socket (battery charging, data transfer)
1 Attachment for carrying strap
2 Condensate trap
3 Magnetic holder
CAUTION
> Keep a safe distance from products which could be damaged
by magnets (e.g. monitors, computers, pacemakers, credit
cards).
4 Service lid
21
Page 22
4 Product description
4.1.7.Components
22
1 Rechargeable battery
2 Measured gas pump
3 Slot for O2 sensor
4 Slot for CO sensor, COlow sensor or CO, H2-compensated
sensor
Page 23
4.2. Compact flue gas probe
1 Removable filter chamber with window and particle filter
2 Probe handle
3 Connector plug for measuring instrument
4 Connecting cable
4.3. Modular flue gas probe
4 Product description
1 Removable filter chamber with window and particle filter
2 Lock release
3 Probe module
4 Connector plug for measuring instrument
5 Probe handle
6 Connecting cable
23
Page 24
5 First steps
5 First steps
5.1. Commissioning
The measuring instrument is supplied with a rechargeable battery
already fitted.
> Charge the rechargeable battery fully before using the
5.2. Getting to know the product
measuring instrument, see Charging the battery, page 58.
5.2.1. Mains operation
If the mains unit is connected, the measuring instrument is
automatically powered from the unit.
1. Connect the mains unit instrument plug to the instrument's
micro USB socket.
5.2.2. Connecting probes
2. Connect the mains plug of the mains unit to a mains socket.
Probe detection at the flue gas socket is carried out
continuously. New probes are recognised automatically.
Connect a probe to the probe socket before switching on
the measuring instrument or start sensor detection
manually after changing the probe: [Options] → Sensor
detection.
24
Connecting flue gas probes/gas pressure
adapters/temperature adapters
> Insert the connector plug into the flue gas socket and lock by
slightly turning it clockwise (bayonet lock).
There must be no more than one extension lead (0554
1201) between measuring instrument and flue gas probe.
Page 25
Connecting other sensors
> Insert the connector plug of the probe into the probe socket.
5.2.3. Switching on
> Press [].
- The start screen is displayed (duration: approx 15 s).
- During commissioning, when the instrument is switched on, the
Country version menu is displayed.
Set the country version:
1. Select the country version: [▲], [▼] → [OK].
2. Confirm confirmation request: Yes → [OK]
- The system is restarted.
- If the voltage supply was interrupted for an extended period:
The menu Date / Time is opened.
- The gas sensors are zeroed.
- There is an instrument error: The Error diagnosis is displayed.
- The Measurement options menu is displayed.
5 First steps
5.2.4. Calling up the function
1. Select function: [▲], [▼].
- The selected function appears in a frame.
2. Confirm selection: [OK].
- The selected function is opened.
25
Page 26
5 First steps
5.2.5. Entering values
Some functions require values (numbers, units, characters) to be
entered. Depending on the selected function, the values are
entered either via a list field or an input editor.
List field
1. Select the value to be changed (numerical value, unit): [▲],
[▼], [◄], [►] (depending on the selected function).
2. Press [Edit].
3. Set value: [▲], [▼], [◄], [►] (depending on the selected
function).
4. Confirm the entry: [OK].
5. Repeat steps 1 and 4 as required.
6. Save the entry: [Finished].
26
Page 27
Input editor
5 First steps
1. Select the value (character) to be changed: [▲], [▼], [◄], [►].
2. Apply value: [OK].
Options:
> Toggle between upper/lower case:
select Ι← ABC→&$/ →Ι: [▲], [▼] → [ABC→&$/].
> Position the cursor in the text:
select Ι← ABC→&$/ →Ι: [▲], [▼] → [Ι←] or [→Ι].
> Delete character before or after the cursor:
select ← next → [▲], [▼] → [←] or [→].
3. Repeat steps 1 and 2 as required.
4. Save the entry: Select ← next → [▲], [▼]
5.2.6. Printing/saving data
Data is printed out via the key []. Data is saved via the Options
menu. The Options menu is accessed via the left function key and
is available in many different menus.
To assign the right function key with the Save function or see
Assigning the right function key, page 32,
Only measured values that have a display field assigned in the
measurement view are saved/printed out.
The measurement data can be printed out parallel to the
saving process, while a measurement program is running.
To be able to transfer data to a record printer via infrared or
Bluetooth interface, the printer used must be enabled, see Printer,
page 36.
→ [Next].
27
Page 28
5 First steps
5.2.7. Saving data to the clipboard (temporary memory)
Using the clipboard, measurement results from various
measurement types can be combined to produce a common
record, which can then be printed out (see above). Data is saved to
the clipboard via the Options menu and the Clipboard command.
If there is data in the clipboard, the status bar shows the symbol .
If there is data in the clipboard and the Print command is triggered,
all data in the clipboard will be printed out.
5.2.8. Confirming an error message
If an error occurs, an error message is shown on the display.
> Confirming an error message: [OK].
Errors that have occurred but have not yet been rectified are
indicated by a warning symbol () in the header.
Error messages that have not yet been cancelled can be displayed
in the Error diagnosis menu, see Instrument diagnosis, page 31.
5.2.9. Switching off
Unsaved measured values are lost if the flue gas analyser
is switched off.
> Press [].
- The pump may start and the sensors may be rinsed until the
switch-off thresholds (O2 > 20%, other measurement
parameters < 50 ppm) are reached. The maximum rinsing
period is 3 minutes.
- The measuring instrument switches off.
5.3. Address/Location
All measuring values can be saved under the currently active
location. Measuring values that have not been saved are lost when
the measuring instrument is switched off!
Addresses and locations can be created, edited, copied and
enabled. Addresses and locations (incl. protocols) can be deleted.
Call up function:
> [] → Address/Location → [OK].
There are various options for opening address.
1. Edit search setting: [Edit].
2. Select search setting: [▲], [▼] → [OK].
Possible settings:
28
Page 29
5 First steps
• Show all: All address/location are displayed.
• Search: A search text only brings up address/location that
contain characteristics of the search text.
• Filter: Individual letters or numbers can be selected. All data
beginning with the relevant letter/number is displayed.
The initial letter is the determining factor for the filter
function, and this can only be selected individually. The
search function can also be used to find a series of several
letters within the address!
3. Carry out search according to search setting: [Search]
Show all
1. Select address: [▲], [▼].
2. Show details: [Details].
3. Enable a location: select the location → [OK].
- The location is activated.
> Open measurements menu: press [OK] again.
Search
1. Edit search criteria: [►] → [Edit].
2. Select search criteria: [▲], [▼] → [OK].
Possible options:
• Contact person
• Address
• Town/city
• Postcode
• Street
- The selected criterion is displayed.
3. Call up entry field for search text: [►] or [▼]
> Enter search text → [Finished]
Do not use the special character * as a placeholder.
Filter
1. Edit search criteria: [Edit].
2. Select search criteria: [▲], [▼] → [OK].
Possible options:
• Contact person
• Address
• Town/city
29
Page 30
5 First steps
• Postcode
• Street
- The selected criterion is displayed.
3. Enable tab: [▼]
4. Select the required tab: [▲], [▼] and sometimes [◄], [►]→
[Filter].
- The search result for the relevant letter or number is displayed.
Create a new measuring location:
A location is always created under an address.
1. Select the address in which the location is to be created.
2. [Options] → New/Location → [OK].
3. Enter values or make settings.
4. Finalise the entry: [Finished].
Other location options:
> [Options] → Edit location: make changes to an existing
location.
> [Options] → Copy location: make a copy of an existing
location in the same address.
> [Options] → Delete location: delete an existing location.
Create new address:
1. [Options] → New address → [OK].
2. Enter values or make settings.
3. Finalise the entry: [Finished].
30
Other address options:
• Edit address: make changes to an existing folder.
• Copy address: make a copy of an existing folder.
• Delete address: delete an existing folder, including the
locations created therein.
Delete All addresses: delete all existing folders, including the
•
locations created in them.
Page 31
5.4. Measurement records
Call up function:
> [] → Measurement records → [OK].
There are various options for opening records. see
Address/Location, page 28.
Displaying a record:
1. Select the required record from the detailed view.
2. [Data].
Options:
> [Options] → Delete Record: delete the selected record.
> [Options] → Number of lines: change the number of
measured values per display page.
> [Options] → Delete all Records: delete all saved records for a
location.
5.5. Instrument diagnosis
Important operating values and instrument data are displayed. A
gas path check can be carried out. The status of the sensors and
any instrument errors not yet rectified can be displayed.
Call up function:
> [] → Instrument diagnosis → [OK].
- The status of the sensor is indicated by a traffic light.
A sensor is able to recover. The sensor status indication
may therefore change from yellow to green or from red to
yellow.
Displaying instrument information
> Device information → [OK].
- Information is displayed.
31
Page 32
6 Using the product
6 Using the product
6.1. Performing settings
6.1.1. Assigning the right function key
The right function key can have a function from the Options menu
assigned to it. The menu Options is accessed via the left function
key and is available in many different menus. This assignment is
only valid for the currently opened menu / the opened function.
✓ A menu / function is opened in which the Options menu is
6.1.2. Instrument settings
displayed on the left function key.
1. Press [Options] .
2. Select option: [▲], [▼].
Depending on the menu / function from which the Options menu
was opened, the following functions are available.
3. Assign the selected function to the right function key. Press
It is assumed that the contents of the chapter First steps
(see First steps, page 24) are known.
Calling up a function:
> [] → Instrument Settings.
see First steps, page 24
6.1.2.1. Measurement view
The parameters/units and the display (number of measured values
shown per display page) can be set.
The settings are only valid for the currently chosen measurement
type, which is indicated by the symbol in the info field.
Total overview of selectable measurement parameters and units
(available selection depends on the set country version and
selected measurement type):
Display Measurement parameter
FT
AT
32
Flue gas temperature
Combustion air temperature
Page 33
Display Measurement parameter
GT
O2
CO2
qA+
Instrument temperature
Oxygen
Carbon dioxide
Flue gas loss with due consideration of the
calorific value range
η+
Efficiency with due consideration of the calorific
value range
Performing settings
> Set parameter: [◄], [►]→ [OK].
6.1.2.7. Printer
The headers (lines 1-3) and the footers for the printout can be set.
The printer that is used can be activated.
Calling up the function:
> [] → Instrument Settings → [OK] → Printer → [OK]
Activating the printer:
The printer 0554 0543 can only be selected after the
Bluetooth
Bluetooth
1. Select Printer → [OK].
2. Select the printer: [▲], [▼] → [OK].
- The printer is activated and the menu Printer is opened.
®
-interface has been activated, see
®
, page 36.
Setting the print text:
1. Print text → [OK].
2. Select function: [▲], [▼]→[Edit].
3. Enter values → [Next].
4. Save the entry: [Finished].
see Bluetooth
6.1.2.8. Bluetooth®
This menu is only available if the instrument is equipped with
Bluetooth. The Bluetooth module can be switched on / off.The relay
36
can now be tested.
Calling up the function:
> [] → Instrument Settings → [OK] → Bluetooth → [Edit].
Making settings:
> Set parameter →[OK].
®
, page 36
Page 37
6.1.2.9. Language
The menu language can be set. The number of available
languages depends on the activated country version, see Country
version, page 37.
Calling up the function:
> [] → Instrument Settings → [OK] → Language → [OK]
Activating the language:
> Select the language →[OK].
see Country version, page 37
6.1.2.10. Country version
Changing the country version may alter the basis for calculation
and therefore also the displayed measurement parameters, fuels,
fuel parameters and calculation formulas.
The selection of the country version influences the menu languages
that can be enabled.
For information concerning the assignment table, the basis for
calculation and the country version, see www.testo.com/downloadcenter (registration required).
Calling up the function:
> [] → Instrument Settings → [OK] → Country Version →
[OK]
This action can be password protected. A password is
specified in the menu Password Protection, see
Password protection, page 38.
Possibly:
> Enter the password: [Enter]→ Enter password → [Next] →
[OK].
6 Using the product
Setting the country version:
1. Select the country version: [▲], [▼] → [OK].
2. Confirm the confirmation request: Yes → [OK]
- The system is restarted.
37
see Password protection, page 38
Page 38
6 Using the product
6.1.2.11. Password protection
The password protection is only valid for functions identified by the
following symbol: or .
Password protection can be activated / deactivated, the password
can be changed.
To deactivate the password protection change the password to
0000 (factory setting).
Calling up the function:
> [] → Instrument Settings → [OK] → Password Protection
→[OK]
Possibly:
> Enter the currently valid password:
[Enter] → Enter password → [Next] → [OK].
Changing the password:
1. [Edit].
2. Enter the new password → [Next].
3. [Edit].
4. Enter the new password again to confirm → [Next].
5. Save changes: [Finished].
6.1.3. Sensor settings
6.1.3.1. O2 reference
The O2 reference value can be set.
The O2 reference value setting may be password protected, see
Password protection, page 38.
Call up function:
> [] → Sensor settings → O2 reference → [Edit].
Possibly:
> Enter the password: [Enter]→ Enter password → [Next] →
[OK].
reference:
2
38
Setting the O
> Set value → [OK].
Page 39
6.1.3.2. Sensor protection
Protection limits can be set to protect the sensors against overload.
Sensor protection switch-off is available for the CO sensor.
Sensor protection is activated if the threshold is exceeded.
To disable sensor protection, the threshold values must be set to 0
ppm.
Call up function:
> [] →Sensor settings→Sensor protection→[OK].
Setting sensor protection thresholds:
1. Select parameter: [Edit].
2. Set value → [OK].
3. Save changes: [Finished].
6.1.3.3. Recalibration/adjustment
The CO sensor can be recalibrated and adjusted.
For recalibration/adjustment, Testo recommends using calibration
adapter 0554 1205 or sending the instrument off to Testo Customer
Service.
If obviously unrealistic measured values are displayed, the
sensors should be checked (calibrated) and, if required,
adjusted.
Adjustments made with low gas concentrations can lead to
accuracy deviations in the upper measuring ranges.
when handling test gas.
> Use test gases in well ventilated rooms only.
1. Connect the calibration adapter to the flue gas socket.
2. Enable CO measurement parameter: [OK].
3. [Edit] → Enter the test gas concentration (nominal value).
4. Attach the connecting line of the test gas bottle to the
calibration adapter.
5. Apply test gas to the sensor.
6. Start recalibration: [Start].
7. Apply the target value once the actual value is stable
(adjustment): [OK].
-or-
Cancel (no adjustment): [esc].
8. Save changes: [Finished].
6.1.4. Fuels
The fuel can be selected. The fuel-specific coefficients and limits
can be set.
Call up function:
> [] → Fuels → [OK].
In order to maintain the measuring accuracy of the
instrument, the correct fuel must be selected or configured.
Correct representation of measuring results is only assured
if the threshold values for the ideal range of the
corresponding measurement task have been set correctly.
The pre-set threshold values are typical values for the
selected system type and the chosen type of fuel.
40
Activating fuels:
> Select the fuel → [OK].
- The fuel is activated and the main menu is opened.
Page 41
Setting coefficients:
1. Select the fuel → [Coeff.].
2. Select the coefficients: [Edit].
Possibly:
> Enter the password: [Enter]→ Enter password → [Next] →
[OK].
3. Set values → [OK].
4. Save changes: [Finished].
Setting limits:
1. Select limit → [Edit].
2. Set values → [OK].
6.2. Measuring
3. Save changes: [Finished].
6.2.1. Preparing for measurement
The First steps chapter (see First steps, page 24) must
have been read.
6.2.1.1. Testing for leaks
The entire measurement system (probe, condensate trap, hoses
and connections) must be tested for leaks before each
measurement to avoid incorrect measurements due to the
infiltration of external air. Testing is carried out while the pump is
running and may be performed by attaching a compressed balloon
pump. The measurement system is leak-tight if the balloon pump is
not filled with air.
6 Using the product
6.2.1.2.Zeroing phases
Measuring the combustion air temperature
If no combustion air temperature probe is connected, during the
zeroing phase, the measured temperature of the flue gas probe is
41
taken as the combustion air temperature.
The flue gas probe should not be in the flue gas duct during
the zeroing phase.
All dependent parameters are calculated using this value. This
method of measuring combustion air temperature is sufficient for
systems dependent on ambient air.
If a temperature probe is connected, the combustion air
temperature is measured continuously via this probe.
Page 42
6 Using the product
Gas zeroing
When the instrument is switched on, the measurement menu is
opened and the gas sensors are zeroed.
The flue gas probe must be in fresh air during the zeroing
phase!
Draught/pressure zeroing
The pressure sensors are zeroed when a pressure measuring
function is called up.
The flue gas probe must be in the fresh air during the
zeroing phase / the instrument must not be pressurised
during zeroing.
6.2.1.3.Using the modular flue gas probe
Checking the thermocouple
The thermocouple of the flue gas probe must not lie against the
probe cage.
> Check before use. Bend the thermocouple back if necessary.
Aligning the flue gas probe
The flue gas must be able to flow freely past the thermocouple.
> Align the probe by turning it as required.
42
Page 43
The tip of the probe must be in the centre of the flue gas flow.
> Align the flue gas probe in the flue gas duct so that the tip is in
the core current (area of the highest flue gas temperature).
6.2.1.4. Measurement view
Only those parameters and units that are enabled in the measured
value display appear in the measured value display, in the saved
measurement records and on the record printouts.
> Before carrying out measurements, set up the measured value
display in such a way that the required parameters and units
are enabled, see Measurement view, page 32.
6.2.1.5. Setting the location and fuel
Before carrying out measurements, the location and fuel must be
correctly selected, see Address/Location, page 28 und see Fuels,
page 40.
6 Using the product
43
Page 44
6 Using the product
6.2.2. Flue gas
To achieve usable measurement results, the measurement
period of a flue gas measurement should be approx. 3 min
and the measuring instrument should display stable
measuring values.
If a separate measurement of CO undiluted has not yet
been carried out, this value is calculated using the
measured values of the flue gas probe and is updated
continuously.
If CO undiluted and/or a draught measurement has already
been carried out separately, the value obtained is applied.
- The measured values are displayed.
2. End measurement: [].
Options
> [Options] → Clipboard: data is saved to the clipboard.
> [Options] → Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] → Save: the measured values are saved in a record.
> [Options] → Fluegas matrix: the measured values are
displayed as a flue gas matrix, see below.
> [Options] → Number of lines: change the number of
measured values per display page.
> [Options] → Recalibrate: the gas sensors are set to zero.
> [Options] → Measurement view: (This function is not available
during a measurement): the measured value display menu is
opened.
44
Page 45
3.
6 Using the product
Showing the flue gas matrix
This function is only available if the measurement parameter CO
has been activated in the measured value display.
Call up function:
✓ The flue gas function is open.
> [Options] → Fluegas matrix.
Options
> [Options] → Clipboard: data is saved to the clipboard.
> [Options] → Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] → Save: the measured values are saved in a record.
> [Options] → Show numeric value: data is displayed as
numerical values.
> [Options] → System type: (This function is not available
during a measurement) Set the system type to be able to
configure the ideal zone (green) of the flue gas matrix, using the
limits pre-configured for each system type.
> [Options] → Reset graphic: the displayed graphical values are
deleted.
> [Options] → Thresholds: (This function is not available during
a measurement) Enter limits to be able to configure the ideal
zone (green) of the flue gas matrix.
> [Options] → CO + O2 or CO + CO2: choose which parameter
should be assigned to the x-axis of the display matrix (O2 or
CO2).
> [Options] → Measurement view: (This function is not available
during a measurement) Open the measured value display
menu.
45
Page 46
6 Using the product
6.2.3. Draught measurement
Call up function:
✓ A flue gas probe must be connected.
This function is only available when the Germany country version
is selected.
A qA average value measurement can be carried out. For this
purpose the average is determined continuously over a period of 30
s, the measuring cycle takes 1 s. The average values actually valid
at the corresponding time of recording are displayed.
Call up function:
✓ A flue gas probe and a combustion air temperature probe must
- The calculated gas flow rate and the gas burner capacity (in
kW) are displayed.
Options:
> [Options] → Clipboard: data is saved to the clipboard.
> [Options] → Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] → Save: the measured values are saved in a record.
> [Options] → Enter Gas Flow: set the gas flow value.
> [Options] → Change unit: The unit for the gas flow can be
changed (m3 > l or l > m3).
Page 53
6.2.13. Oil flow
The function is only available if the chosen fuel is an oil.
Calling up the function:
> [] → Measurements → [OK] → Oil Flow → [OK].
Performing the measurement:
1. Select the parameters Oil Flow (of the oil nozzle) and Oil
Pressure (no effect on calculation): [▲], [▼] → [Edit].
2. Enter values. [▲], [▼] and partly [◄], [►]→ [OK].
- The calculated oil burner capacity (in kW) is displayed.
Options:
> [Options] → Clipboard: Data are saved to the clipboard
> [Options] → Delete clipboard: Any data saved to the clipboard
is deleted.
> [Options] → Save: The readings are saved in a protocol.
> [Options] → Edit Unit: The unit for the oil flow can be changed
(kg/h > gal/h or gal/h > kg/h).
6.2.14. Ambient CO
This function is not available when the Spain country version is
selected.
✓ An ambient CO probe (recommended) or a flue gas probe must
be connected.
Cigarette smoke influences the measurement by more than
50 ppm. The breath of a smoker influences the
measurement by about 5 ppm.
When using an ambient CO probe, note that:
The direction of flow of the gas has an effect on the
accuracy of measurement. Frontal flow onto the probe
leads to higher measured values. The best measurement
results are achieved when the probe is moved gently
backwards and forwards.
When using the ambient CO probe and the flue gas probe,
note that:
the probe must be in the fresh air (CO-free) during the
zeroing phase!
Call up function:
> [] → Measurement options → [OK] → CO ambient →
[OK].
6 Using the product
53
Page 54
6 Using the product
Carrying out the measurement:
1. Start measurement: [].
- The measurement starts and the measured value is displayed
graphically (trend display).
- An audible alarm signal is triggered when the alarm limit is
reached.
2. End measurement: [].
3. Confirm the message: [OK].
Options:
> [Options] → Clipboard: data is saved to the clipboard.
> [Options] → Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] → Save: the measured values are saved in a record.
> [Options] → Set alarm limits: the alarm limits menu is opened.
6.2.15. CO2 ambient
This function is not available when the Spain country version is
selected.
✓ An ambient CO2 probe (0632 1240) must be connected.
In order to obtain correct measured values, it is imperative
to enter the prevailing absolute pressure. This can be
entered directly (Pressure absolute), or it is automatically
calculated when entering Altitude and barometric pressure
(Barometric pressure).
2. Enter values. [▲], [▼] and in some cases [◄], [►] → [OK].
3. Start measurement: [].
4. End measurement: [].
- The ambient CO2 value is displayed.
Options:
> [Options] → Clipboard: data is saved to the clipboard.
> [Options] → Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] → Save: the measured values are saved in a record.
Page 55
> [Options] → Alarm limit: the alarm limits menu is opened.
> [Options] → Edit: values for adjustable parameters can be
edited.
> [Options] → Measurement view: (This function is not available
during a measurement) The measured value display menu is
opened.
6.2.16. Leak detection
This function is not available when the Spain country version is
selected.
During gas leak detection, no measurement is carried out, but gas
detection is performed.
✓ A gas leak probe (0632 3330) must be connected.
You must also refer to the documentation that comes with
the gas leak probe.
Call up function:
> [] → Measurement options → [OK] → Gas leak detection
→[OK].
Carrying out detection:
> Set the gas type to be detected and perform gas leak testing by
following the instructions described in the documentation of the
gas leak probe.
1. Start detection: [].
- The gas concentration is displayed, and if it exceeds the limit,
an alarm signal will sound.
6 Using the product
Options:
> [Options] → Alarm limit: (this function is not available during a
measurement) Adjusting alarm limits.
> [Options] → Alarm signal: (this function is not available during
a measurement) Disable/enable alarm signal.
> [Options] → Zeroing probe: perform zeroing.
6.3. Transferring data
2. End detection: [].
6.3.1. Report printer
To be able to transmit data via infrared or Bluetooth interface to a
Testo report printer, the printer to be used must have been
activated, see Printer, page 36.
55
Page 56
7 Maintaining the product
Printing out data takes place via [Print] or []. The function is only
available if a printout is possible.
Representation of the available characters per line is
limited on the printout. Thus, the unit mg/KWh is displayed
on the printout as mg/k.
6.3.2. PC / Pocket PC
Data transfer to a PC can take place via USB, IrDA or Bluetooth®.
Data transfer to a Pocket PC can take place via IrDA or Bluetooth
You must also refer to the documentation that comes with the
7 Maintaining the product
software.
7.1. Cleaning the measuring instrument
> If the housing of the measuring instrument is dirty, clean it with
a damp cloth. Do not use any aggressive cleaning agents or
solvents! Mild household cleaning agents and soap suds may
be used.
7.2. Replacing the rechargeable battery
✓ The measuring instrument must not be connected to a mains
socket via the mains unit. The instrument must be switched off.
Change the rechargeable battery within 2 minutes so that
instrument settings (e.g. date/time) are not lost.
1. Place the measuring instrument on its front.
®
.
56
2. Unscrew, lift up and remove the service cover.
Page 57
7 Maintaining the product
3. Press down on the holder lightly with your fingers. Pull the
retaining clip in the direction of the arrow until the catch is
released. Remove holder.
4. Unplug the plug-in connection from the slot.
5. Carefully pull the retaining clips (1, 2) outwards and push
rechargeable battery up and out of the holder (3).
6. Insert the new rechargeable battery in the holder. Make sure
that the plug-in connection cable is routed out of the holder at
the side. Press the rechargeable battery into the holder until the
retaining clips click into place.
7. Plug the new rechargeable battery plug-in connection into the
slot.
8. Insert the holder into the guide rail and slide it in the direction of
the arrow until the holder clicks into place.
9. Refit and close the service cover.
57
Page 58
7 Maintaining the product
7.3. Charging the battery
The rechargeable battery can only be charged at an ambient
temperature of ±0 to +35°C. If the rechargeable battery has been
completely discharged, the charging time at room temperature with
the testo mains unit is approx. 6 h.
Charging in the measuring instrument
1. Connect the mains unit instrument plug to the instrument's
micro USB socket.
2. Connect the mains plug of the mains unit to a mains socket.
- The charging process will start. The charging process will stop
automatically when the battery is fully charged.
Battery care
> Do not fully exhaust rechargeable batteries.
> Store rechargeable batteries only in charged condition and at
low temperatures, but not below 0°C (best storage conditions
with a charge level of 50-80%, at an ambient temperature of 1020°C, recharge completely before use).
7.4. Replacing sensors
✓ The measuring instrument must be switched off.
1. Place the measuring instrument on its front.
2. Unscrew, lift up and remove the service cover.
3. Disconnect the hose connections from the faulty sensor/bridge.
4. Remove the faulty sensor/bridge from the slot.
5. Install new sensor / new bridge in the slot.
58
Page 59
7 Maintaining the product
6. Push the hose connections on to the sensor/bridge.
7. Refit and close the service cover.
After replacing an O2 sensor, wait for an equalisation
period of 60 min to elapse before using the instrument
again.
When retrofitting a sensor, the associated measurement
parameter and unit must be enabled, see Sensor settings,
page 38.
7.5. Recalibrating/adjusting sensors
See Recalibration/adjustment, page 39.
7.6. Cleaning the modular flue gas probe
✓ Disconnect the flue gas probe from the measuring instrument
prior to cleaning.
1. Release the probe catch by pressing the key on the probe
handle and remove the probe module.
2. Blow compressed air through the flue gas ducts in probe
module and probe handle (see illustration). Do not use a brush!
3. Fit a new probe module on the probe handle and engage it in
59
place.
Page 60
7 Maintaining the product
7.7. Replacing the probe module
✓ Disconnect the flue gas probe from the measuring instrument
prior to cleaning.
1. Press the key on the top of the probe handle and remove the
probe module.
2. Plug in the new probe module and engage it in place.
7.8. Changing the thermocouple
7.8.1. Modular flue gas probe
1. Release the probe catch by pressing the key on the probe
handle and remove the probe module.
2. Remove the thermocouple plug-in head from the socket using a
screwdriver and pull the thermocouple out of the probe shaft.
3. Insert the new thermocouple into the probe shaft until the
connection head clicks into place.
4. Fit a new probe module on the handle and engage in place.
7.8.2. Compact flue gas probe
1. Loosen and remove halfshell handles.
2. Turn sealing cap clockwise as far as it will go and remove.
3. Release and remove the halfshell element.
4. Detach adapter and hose, push thermocouple out of its holder
and disconnect cable from thermocouple
5. Connect cables to the new thermocouple (white -, green +) and
push thermocouple back into the holder.
60
Page 61
6. Connect adapter and hose to thermocouple. Adjust cables and
hoses, and refit halfshell elements.
7. Refit sealing cap and turn anti-clockwise (as far as it will go,
observe markings).
8. Replace halfshell handles and secure with screws.
7.9. Condensate container
The fill level of the condensate container can be monitored via the
markings on the condensate trap.
7 Maintaining the product
Draining the condensate container
The condensate consists of a weak mix of acids. Avoid skin
contact. Make sure that the condensate does not run over
the housing.
CAUTION
Damage to the sensors and the flue gas pump due to
condensate entering the gas path!
> Do not empty the condensate container while the flue gas
pump is in operation.
1. Open the condensate outlet on the condensate trap: pull out the
plug to the stop.
61
Page 62
7 Maintaining the product
2. Let the condensate run out into a sink.
3. Wipe off any drops still on the condensate outlet with a cloth
and close the condensate outlet.
The condensate outlet must be completely closed
(marking), otherwise measuring errors could be caused by
infiltrated air.
7.10.Checking/replacing the particle filter
Checking the particle filter:
> Check the particle filter of the modular and compact flue gas
probe regularly for contamination: check visually by looking
through the window of the filter chamber.
Replace the filter if there are signs of contamination.
62
Replacing the particle filter:
The filter chamber may contain condensate.
1. Open the filter chamber: Turn gently anti-clockwise.
2. Remove the filter cartridge and replace it with a new one (0554
3385).
3. Attach the filter chamber and lock it Turn gently clockwise.
Page 63
8 Tips and assistance
8.1. Questions and answers
Question Possible causes/solution
Rechargeable battery low > Switch to mains operation.
Measuring instrument
switches off automatically
or cannot be switched on
Batteries/rechargeable batteries
empty.
> Charge rechargeable battery or
switch to mains operation.
Error message: Cell protection enabled
The shutdown limit of the CO sensor
was exceeded.
> Remove the probe from the stack.
Error message: Printing
not possible
• With printer 0554 0543: Bluetooth
interface not enabled.
• Wrong printer enabled.
• Printer switched off.
• Printer outside wireless
transmission range.
> Enable Bluetooth interface, see
Bluetooth
> Enable printer to be used, see
Printer, page 36.
> Switch on printer.
> Move printer into wireless
transmission range.
If we could not answer your question, please contact your dealer or
Testo Customer Service. For contact details, see back of this
document or the website www.testo.com/service-contact.
0.5mm, probe shaft diameter: 6 mm
Probe shaft module 300 mm, 1000°C, thermocouple
0554 8764
1.0 mm, probe shaft diameter: 6 mm
Probe shaft module 700mm, 1000°C, thermocouple
0554 8765
1.0 mm, probe shaft diameter: 6 mm
Spare thermocouple for module 0554 9760,
0430 9760
0554 9762
Spare thermocouple for module 0554 9761,
0430 9761
0554 9763
Spare thermocouple for module 0554 8764 0430 8764
Spare thermocouple for module 0554 8765 0430 8765
Cone, 8mm, steel 0554 3330
Cone, 6mm, steel 0554 3329
Multi-hole probe shaft, length 300 mm, Ø 8 mm, for
0554 5762
CO averaging
Multi-hole probe shaft, length 180 mm, Ø 8 mm, for
0554 5763
CO averaging
Flexible probe shaft module 0554 9764
Hose extension 2.8 m, extension line probe -
0554 1202
instrument
Particle filter for modular flue gas probe, 10 pieces 0554 3385
Probe modules/accessories for compact flue gas probes
Description Article no.
Spare thermocouple for 0600 9740 0430 0383
Spare thermocouple for 0600 9741 0430 0382
Particle filter for compact flue gas probe, 10 pieces 0554 0040
65
Page 66
8 Tips and assistance
Temperature probe
Description Article no.
Combustion air temperature probe, 300mm 0600 9791
Combustion air temperature probe, 190mm 0600 9787
Combustion air temperature probe, 60mm 0600 9797
Surface probe (angled) 0604 0994
Fast reaction surface sensor 0604 0194
Miniature ambient air sensor 0600 3692
Other probes
Description Article no.
O2 annular gap probe 0632 1260
Gas leak probe 0632 3330
Ambient CO probe 0632 3331
CO2 ambient probe (without connecting cable) 0632 1240
Connecting cable for CO2 ambient probe, 1.5 m 0430 0143
Gas pressure set: Draught path adapter, silicone
0554 1203
hose 4 mm / 6 mm, reducing cones
Smoke tester incl. oil, soot plates, for measuring
0554 0307
soot in flue gas
66
Spare sensors
Description Article no.
O2 sensor 0393 0003
CO sensor 0393 0053
CO sensor H2-compensated 0393 0105
COlow sensor 0393 0103
Cases
Basic system case with double floor for instrument,
0516 3331
probes and other accessories
Basic system case for instrument, probes and
0516 3330
accessories
Page 67
8 Tips and assistance
System case with tool pouch without contents, click-
0516 0329
fastened to basic system case
Universal system case without compartments, click-
0516 0331
fastened to basic system case
Instrument case (leather) with compartments for
0516 0303
instruments and accessories
Other accessories
Description Article no.
Mains unit incl. micro USB cable 0554 1105
Spare rechargeable battery 0515 0046
Readout adapter for automatic furnaces 0554 1206
Connecting cable instrument / PC 0449 0047
Easyheat (PC configuration software) 0554 3332
Draught set 0554 3150
External micro pressure probe 0638 0330
Set of capillary hoses 0554 1215
Adhesive kits (x50) for printouts, paper barcode
0554 0116
labels etc.
Instrument cleaner (100 ml) 0554 1207
Straight Pitot tube 0635 2050
ISO Calibration Certificate Flue Gas 0520 0003
For a complete list of all accessories and spare parts, please refer
to the product catalogues and brochures or look up our website
67
www.testo.com
Page 68
8 Tips and assistance
8.3. Updating the instrument software
Under www.testo.com/download-center you can download the
current instrument software (Firmware) for testo 320 (registration
required).
> Unplug the micro USB cable and switch off the testo 320.
1. Hold down [▲].
2. Plug in the micro USB cable, continue holding down [▲].
- The display shows Firmware update along the bottom edge.
3. Release [▲].
- Your PC recognises the testo 320 as a removable medium.
4. Copy the new file (ap320rel.bin) to the detected removable
medium.
- On the display the status bar progresses from left to right. This
process may take a few minutes.
- Once the instrument software (firmware) has been updated, the
system will automatically reboot and is ready for use.
5. Disconnect the connecting cable from the testo 320.
68
Page 69
8 Tips and assistance
69
Page 70
8 Tips and assistance
70
Page 71
Page 72
0970 3200 en 02 V01.00 en
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.