Testo 320 Operating Manual

Page 1
testo 320 · Flue gas analyzer
Instruction manual
Page 2
2
Page 3
1 Contents
1 Contents
1 Contents ................................................................................................... 3
2 Safety and the environment .................................................................... 6
2.1. About this document ........................................................................ 6
2.2. Ensure safety ................................................................................... 7
2.3. Protecting the environment .............................................................. 8
3 Specifications .......................................................................................... 9
3.1. Use .................................................................................................. 9
3.2. Technical data ............................................................................... 10
3.2.1. Examinations and licenses ............................................................................ 10
3.2.2. Bluetooth® module (option) ............................................................................ 10
3.2.3. Declaration of Conformity .............................................................................. 12
3.2.4. Measurement ranges and resolution .............................................................. 13
3.2.5. Accuracy and response time .......................................................................... 14
3.2.6. Other instrument data .................................................................................... 15
4 Product description ............................................................................... 17
4.1. Measuring instrument .................................................................... 17
4.1.1. Front view ...................................................................................................... 17
4.1.2. Keypad .......................................................................................................... 18
4.1.3. Display ........................................................................................................... 18
4.1.4. Instrument connections .................................................................................. 19
4.1.5. Condensate outlet and interfaces .................................................................. 20
4.1.6. Rear view ....................................................................................................... 21
4.1.7. Components .................................................................................................. 22
4.2. Compact flue gas probe ................................................................ 23
4.3. Modular flue gas probe .................................................................. 23
5 First steps .............................................................................................. 24
5.1. Commissioning .............................................................................. 24
5.2. Getting to know the product ........................................................... 24
5.2.1. Mains operation ............................................................................................. 24
5.2.2. Connecting probes ......................................................................................... 24
5.2.3. Switching on .................................................................................................. 25
5.2.4. Calling up the function ................................................................................... 25
5.2.5. Entering values .............................................................................................. 26
5.2.6. Printing/saving data ....................................................................................... 27
5.2.7. Saving data to the clipboard (temporary memory).......................................... 28
5.2.8. Confirming an error message ......................................................................... 28
5.2.9. Switching off .................................................................................................. 28
5.3. Address/Location ........................................................................... 28
5.4. Measurement records .................................................................... 31
5.5. Instrument diagnosis ..................................................................... 31
Page 4
1 Contents
6
Using the product ................................................................................. 32
6.1. Performing settings ....................................................................... 32
6.1.1. Assigning the right function key ..................................................................... 32
6.1.2. Instrument settings ........................................................................................ 32
6.1.2.1. Measurement view .......................................................................... 32
6.1.2.2. Alarm limits ...................................................................................... 34
6.1.2.3. Units ................................................................................................ 35
6.1.2.4. Date / time ....................................................................................... 35
6.1.2.5. Energy management ....................................................................... 35
6.1.2.6. Display brightness ........................................................................... 36
6.1.2.7. Printer ............................................................................................. 36
6.1.2.8. Bluetooth® ....................................................................................... 36
6.1.2.9. Language ........................................................................................ 37
6.1.2.10. Country version ............................................................................... 37
6.1.2.11. Password protection ........................................................................ 38
6.1.3. Sensor settings .............................................................................................. 38
6.1.3.1. O2 reference .................................................................................... 38
6.1.3.2. Sensor protection ............................................................................ 39
6.1.3.3. Recalibration/adjustment ................................................................. 39
6.1.4. Fuels .............................................................................................................. 40
6.2. Measuring ..................................................................................... 41
6.2.1. Preparing for measurement ........................................................................... 41
6.2.1.1. Testing for leaks .............................................................................. 41
6.2.1.2. Zeroing phases ................................................................................ 41
6.2.1.3. Using the modular flue gas probe .................................................... 42
6.2.1.4. Measurement view .......................................................................... 43
6.2.1.5. Setting the location and fuel ............................................................ 43
6.2.2. Flue gas ......................................................................................................... 44
6.2.3. Draught measurement ................................................................................... 46
6.2.4. External micro pressure probe ....................................................................... 47
6.2.5. Average ......................................................................................................... 47
6.2.6. BImSchV ....................................................................................................... 48
6.2.7. CO undiluted .................................................................................................. 49
6.2.8. Smoke number/HCT ...................................................................................... 49
6.2.9. Differential pressure ....................................................................................... 50
6.2.10. Differential temperature ................................................................................. 51
6.2.11. O2 air ............................................................................................................. 52
6.2.12. Gas flow rate ................................................................................................. 52
6.2.13. Oil flow ........................................................................................................... 53
6.2.14. Ambient CO ................................................................................................... 53
6.2.15. CO2 ambient ................................................................................................. 54
6.2.16. Leak detection ............................................................................................... 55
6.3. Transferring data ........................................................................... 55
6.3.1. Report printer ................................................................................................. 55
6.3.2. PC / Pocket PC .............................................................................................. 56
4
Page 5
1 Contents
7
Maintaining the product ........................................................................ 56
7.1. Cleaning the measuring instrument ............................................... 56
7.2. Replacing the rechargeable battery ............................................... 56
7.3. Charging the battery ...................................................................... 58
7.4. Replacing sensors ......................................................................... 58
7.5. Recalibrating/adjusting sensors ..................................................... 59
7.6. Cleaning the modular flue gas probe ............................................. 59
7.7. Replacing the probe module .......................................................... 60
7.8. Changing the thermocouple........................................................... 60
7.8.1. Modular flue gas probe .................................................................................. 60
7.8.2. Compact flue gas probe ................................................................................. 60
7.9. Condensate container ................................................................... 61
7.10. Checking/replacing the particle filter .............................................. 62
8 Tips and assistance ............................................................................... 63
8.1. Questions and answers ................................................................. 63
8.2. Accessories and spare parts ......................................................... 64
8.3. Updating the instrument software .................................................. 68
Page 6
2 Safety and the environment
2 Safety and the environment
2.1. About this document
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay particular attention to the safety instructions and warning advice in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Warnings
Always pay attention to information that is marked by the following warnings with warning pictograms. Implement the specified precautionary measures.
Representation Explanation
WARNING
CAUTION
NOTICE
Indicates potential serious injuries
indicates potential minor injuries
indicates circumstances that may lead to damage to the products
6
Page 7
2 Safety and the environment
Symbols and writing standards
Represen-
Explanation
tation
Note: Basic or further information.
1. ...
Action: more steps, the sequence must be followed.
2. ...
> ... Action: a step or an optional step.
- ... Result of an action.
[OK] Control keys of the instrument or buttons of the
2.2. Ensure safety
program interface.
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not use any force.
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
> Do not perform contact measurements on non-insulated, live
parts.
> The testo 320 is not suitable for long-term measurements and
should not be used as a safety (alarm) instrument.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the prescribed steps exactly. Use only original spare parts from Testo.
> Any further or additional work must only be carried out by
authorised personnel. Testo will otherwise refuse to accept responsibility for the proper functioning of the measuring instrument after repair and for the validity of certifications.
> Only use the device in closed, dry rooms and protect it from rain
and moisture.
> Temperatures given on probes/sensors relate only to the
measuring range of the sensors. Do not expose handles and feed lines to any temperatures in excess of 70 °C unless they are expressly permitted for higher temperatures.
Page 8
2 Safety and the environment
> The testo 320 must be checked before commissioning for any
visible damage. Do not commission the testo 320 if there are signs of damage on the housing, mains unit or supply lines. Electrical risk.
> The objects to be measured or the measurement environment
may also pose risks: Note the safety regulations valid in your area when performing the measurements.
For products with Bluetooth® (optional)
Changes or modifications that have been made without the explicit consent of the responsible approval authority, may cause the retraction of the type approval.
Data transfer may be disturbed by equipment that uses the same ISM-band, e.g. WLAN, microwave ovens, ZigBee.
The use of radio communication links is not permitted, among others, in aeroplanes and hospitals. For this reason the following points must be ensured before entering:
> Switch off the device: > Isolate the device from any external power sources (mains
2.3. Protecting the environment
cable, external rechargeable batteries, ...).
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal.
8
Page 9
3 Specifications
3.1. Use
The testo 320 is a handheld measuring device for the professional flue gas analysis of combustion plants:
Small combustion plants (burning oil, gas, wood, coal)
Low-temperature and condensing boilers
Gas water heaters These systems can be adjusted using the testo 320 and checked
for compliance with the applicable limit values. The following tasks can also be carried out with the testo 320:
Regulating the O2, CO and CO2 values in combustion plants for the purpose of ensuring optimal operation.
Draught measurement.
Measuring and regulating the gas flow pressure in gas water heaters.
Measuring and optimising the flow and return temperatures of heating systems.
Ambient CO measurement (only possible with additional ambient CO probe 0632 3331).
Ambient CO measurement (only possible with additional ambient CO probe 0632 1240).
Detection of CH4 (methane) and C3H8 (propane) (only possible with additional gas leak probe 0632 3370).
The Bluetooth is type approved.
®
option may only be operated in countries in which it
3 Specifications
Page 10
3 Specifications
3.2. Technical data
3.2.1. Examinations and licenses
As declared in the certificate of conformity, this product complies with Directive 2004/108/EC.
This product is TÜV-tested in compliance with 1. BImSchV. The sensors 0393 0105 (CO, H2-compensated), 0393 0003 (O2), temperature and pressure are TÜV-tested in accordance with EN 50379 part 2.
The measuring cell 0393 0053 (CO, not H2-compensated) is TÜV­tested as per EN 50379 part 3.
This product is EMC-tested as per DIN EN 61326-1. For official measurements in accordance with 1. BImSchV
(chimney sweeps), the measuring instrument must be checked every six months by a technical testing body of the Guild of Master Chimney Sweeps or another testing body recognised by the authorities.
3.2.2. Bluetooth® module (option)
• Bluetooth® type: BlueGiga WT11
• Bluetooth
• Bluetooth
• Bluetooth
®
product note: WT11
®
identification: B011198
®
company: 10274
10
Certification
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE), Estonia (EE), Finland (FI), France (FR), Greece (GR), Ireland (IE), Italy (IT), Latvia (LV), Lithuania (LT), Luxembourg (LU), Malta (MT), Netherlands (NL), Austria (AT), Poland (PL), Portugal (PT), Romania (RO), Sweden (SE), Slovakia (SK), Slovene (SI), Spain (ES), Czech Republic (CZ), Hungary (HU), United Kingdom (GB), Republic of Cyprus (CY).
Iceland, Liechtenstein, Norway and Switzerland. Turkey, El Salvador, Ecuador
Page 11
3 Specifications
Information of the FCC (Federal Communications Commission)
This device fulfils part 15 of the FCC-guidelines. Commissioning is subject to the two following conditions: (1) This device must not generate any dangerous interferences and (2) this device must be able to receive interferences, even if these could have undesired effect on the operation.
Changes The FCC demands that the user is to be informed that with any
changes and modifications to the device, which have not been explicitly approved by testo AG, the right of the user to use this device will become null and void.
Page 12
3 Specifications
3.2.3. Declaration of Conformity
12
Page 13
3.2.4. Measurement ranges and resolution
Measurement parameter
O2 0 to 21 Vol.% 0.1 vol.% CO 0...4000 ppm 1 ppm CO, H2-comp. COlow, H2-comp. 0 to 500 ppm 0.1 ppm Draught1 -9.99 to 40.00 hPa 0.01 hPa Fine draught1
P (only with gas pressure set 0554
1203)
Fine pressure1 (only with gas pressure set 0554
1203)
temperature -40 to 1200°C 0.1°C (-40.0 to 999.9°C)
Efficiency 0 to 120 % 0.1 % Flue gas loss 0 to 99.9 % 0.1 %
Measuring range Resolution
0 to 8000 ppm 1 ppm
-9.999 hPa to +40.000 hPa
0 to 300 hPa 0.1 hPa
0 to 300 hPa 0.01 hPa
0.001 hPa
1°C (from 1000°C)
3 Specifications
1
Depending on the country version
Page 14
3 Specifications
3.2.5. Accuracy and response time
Measurement parameter
O2 ±0.2 vol.% < 20 s CO
CO, H2-comp.
COlow, H2-comp. ±2 ppm (0 to 39.9 ppm)
Draught1
Fine draught1
P (only with gas pressure set 0554
1203)
Accuracy
±20 ppm (0 to 400 ppm) ±5% of meas. val. (401 to 2000
ppm) ±10% of meas. val. (2001 to
4000 ppm) ±10 ppm or ±10% of meas. val.
(0 to 200 ppm) ±20 ppm or ±5% of meas. val.
(201 to 2000 ppm) ±10% of meas. val. (2001 to
8000 ppm)
±5% of meas. val. (rest of range) ±0.02 ppm or ±5% of meas. val.
(-0.50 to 0.60 hPa) ± 0.03 hPa (0.61 to 3.00 hPa) ±1.5% of meas. val. (3.01 to
40.00 hPa) ±0.02 ppm or ±5% of meas. val.
(-0.50 to 0.60 hPa) ± 0.03 hPa (0.61 to 3.00 hPa) ±1.5% of meas. val. (3.01 to
40.00 hPa) ± 0.5 hPa (0.0 to 50.0 hPa)
±1% of meas. val. (50.1 to 100.0 hPa)
±1.5% of meas. val. (rest of range)
Response time (t
< 60 s
2
< 40 s
2
< 40 s
2
-
2
-
)
90
2
higher value is valid
14
Page 15
3 Specifications
Measurement
Accuracy
parameter
Fine pressure1 (only with gas pressure set 0554
1203)
± 0.5 hPa (0.0 to 50.0 hPa) ±1% of meas. val. (50.1 to 100.0 hPa) ±1.5% of meas. val. (rest of
range)
temperature ± 0.5°C (0.0 to 100.0°C)
±0.5% of meas. val. (rest of
range) Efficiency - ­Flue gas loss - -
3.2.6. Other instrument data
Feature Values
Storage and
-20 to 50°C transport temperature
Operating
-5 to 45°C temperature
Ambient humidity
Power supply
0…90 % rH, not condensing
Rech. batt.: 3.7 V / 2.4 Ah
Mains unit: 5.0 V/1000 mA Protection class IP40 Weight 573 g Dimensions 240 x 85 x 65 mm Memory 500 measured values Display Graphic colour display, 240 x 320 pixels Gas leak testing
probe
Optimum rech. batt. storage conditions
visual indication (LED)
audible indication by buzzer
Charge level: capacity at 50-80% ambient
temperature: 10-20°C Battery charge time Approx. 5-6 h with mains unit supplied Rechargeable
battery life
Approx. 6 h (pump on, 20°C ambient
temperature) Data transfer IrDA, USB, Bluetooth® (option)
Response time (t
)
90
depending on the probe
Page 16
3 Specifications
Feature Values
Bluetooth® (option) Range < 10 m Warranty Measuring instrument: 24 months
O2 sensor: 24 months CO sensor: 24 months CO sensor with H2 comp.: 24 months CO/H2 low sensor (TCHL): 24 months Flue gas probe: 24 months Thermocouple: 12 months Rech. batt.: 12 months
Terms of warranty
Terms of warranty: see website www.testo.com/warranty
16
Page 17
4 Product description
4.1. Measuring instrument
4.1.1. Front view
4 Product description
1 Display 2 Function keys
3 Keypad
Page 18
4 Product description
4.1.2. Keypad
Button Functions
[]
[OK]
Example
[] []
[esc]
[]
[]
4.1.3. Display
Switch measuring instrument on / off
Function key (orange, 3x), relevant function is shown on the display
Scroll up, increase value, navigate Scroll down, reduce value, navigate Back, cancel function Open main menu
Transmit data to the Testo protocol printer.
18
1 Status bar (dark grey background):
• Warning symbol (only if there is an instrument error, display of error in instrument diagnosis menu), otherwise: Instrument designation.
• Symbol (only if data is stored in the temporary memory).
• Display of date and time.
• Indication of Bluetooth
®
status, power supply and remaining
rechargeable battery capacity:
Icon Feature
blue symbol = Bluetooth grey symbol = Bluetooth
®
®
on,
off
Page 19
Icon Feature
Battery operation Display of remaining rechargeable battery
capacity by colour and fill level of the battery icon (green = 5-100%, red = < 5%)
Mains operation Display of remaining rechargeable battery
capacity: see above
2 Info field of register tabs: Indication of selected folder/location,
chosen fuel, chosen measurement type.
3 Selection field for functions (selected function appears against a
white background, unavailable functions are identified by grey font) or display of measured values.
4.1.4. Instrument connections
4 Function display for function keys.
4 Product description
1 Probe socket 2 Gas outlet 3 Probe socket 4 Micro USB socket (battery charging, data transfer)
Page 20
4 Product description
4.1.5. Condensate outlet and interfaces
1 Infrared interface (IrDA) 2 Bluetooth interface (option) 3 Condensate outlet
20
Page 21
4.1.6. Rear view
4 Product description
1 Attachment for carrying strap 2 Condensate trap 3 Magnetic holder
CAUTION
> Keep a safe distance from products which could be damaged
by magnets (e.g. monitors, computers, pacemakers, credit cards).
4 Service lid
Page 22
4 Product description
4.1.7. Components
22
1 Rechargeable battery 2 Measured gas pump 3 Slot for O2 sensor 4 Slot for CO sensor, COlow sensor or CO, H2-compensated
sensor
Page 23
4.2. Compact flue gas probe
1 Removable filter chamber with window and particle filter 2 Probe handle 3 Connector plug for measuring instrument 4 Connecting cable
4.3. Modular flue gas probe
4 Product description
1 Removable filter chamber with window and particle filter 2 Lock release 3 Probe module 4 Connector plug for measuring instrument 5 Probe handle 6 Connecting cable
Page 24
5 First steps
5 First steps
5.1. Commissioning
The measuring instrument is supplied with a rechargeable battery
already fitted.
> Charge the rechargeable battery fully before using the
5.2. Getting to know the product
measuring instrument, see Charging the battery, page 58.
5.2.1. Mains operation
If the mains unit is connected, the measuring instrument is automatically powered from the unit.
1. Connect the mains unit instrument plug to the instrument's micro USB socket.
5.2.2. Connecting probes
2. Connect the mains plug of the mains unit to a mains socket.
Probe detection at the flue gas socket is carried out continuously. New probes are recognised automatically.
Connect a probe to the probe socket before switching on the measuring instrument or start sensor detection manually after changing the probe: [Options] Sensor
detection.
24
Connecting flue gas probes/gas pressure adapters/temperature adapters
> Insert the connector plug into the flue gas socket and lock by
slightly turning it clockwise (bayonet lock).
There must be no more than one extension lead (0554
1201) between measuring instrument and flue gas probe.
Page 25
Connecting other sensors
> Insert the connector plug of the probe into the probe socket.
5.2.3. Switching on
> Press [].
- The start screen is displayed (duration: approx 15 s).
- During commissioning, when the instrument is switched on, the
Country version menu is displayed.
Set the country version:
1. Select the country version: [], [] [OK].
2. Confirm confirmation request: Yes [OK]
- The system is restarted.
- If the voltage supply was interrupted for an extended period: The menu Date / Time is opened.
- The gas sensors are zeroed.
- There is an instrument error: The Error diagnosis is displayed.
- The Measurement options menu is displayed.
5 First steps
5.2.4. Calling up the function
1. Select function: [], [].
- The selected function appears in a frame.
2. Confirm selection: [OK].
- The selected function is opened.
Page 26
5 First steps
5.2.5. Entering values
Some functions require values (numbers, units, characters) to be entered. Depending on the selected function, the values are entered either via a list field or an input editor.
List field
1. Select the value to be changed (numerical value, unit): [],
[], [], [] (depending on the selected function).
2. Press [Edit].
3. Set value: [], [], [], [] (depending on the selected
function).
4. Confirm the entry: [OK].
5. Repeat steps 1 and 4 as required.
6. Save the entry: [Finished].
26
Page 27
Input editor
5 First steps
1. Select the value (character) to be changed: [], [], [], [].
2. Apply value: [OK]. Options: > Toggle between upper/lower case:
select Ι← ABC&$/ →Ι: [], [] [ABC&$/].
> Position the cursor in the text:
select Ι← ABC&$/ →Ι: [], [] [Ι←] or [→Ι].
> Delete character before or after the cursor:
select next [], [] [] or [].
3. Repeat steps 1 and 2 as required.
4. Save the entry: Select next [], []
5.2.6. Printing/saving data
Data is printed out via the key []. Data is saved via the Options menu. The Options menu is accessed via the left function key and is available in many different menus.
To assign the right function key with the Save function or see Assigning the right function key, page 32,
Only measured values that have a display field assigned in the measurement view are saved/printed out.
The measurement data can be printed out parallel to the saving process, while a measurement program is running.
To be able to transfer data to a record printer via infrared or Bluetooth interface, the printer used must be enabled, see Printer, page 36.
[Next].
Page 28
5 First steps
5.2.7. Saving data to the clipboard (temporary memory)
Using the clipboard, measurement results from various measurement types can be combined to produce a common record, which can then be printed out (see above). Data is saved to the clipboard via the Options menu and the Clipboard command.
If there is data in the clipboard, the status bar shows the symbol . If there is data in the clipboard and the Print command is triggered,
all data in the clipboard will be printed out.
5.2.8. Confirming an error message
If an error occurs, an error message is shown on the display. > Confirming an error message: [OK]. Errors that have occurred but have not yet been rectified are indicated by a warning symbol ( ) in the header. Error messages that have not yet been cancelled can be displayed
in the Error diagnosis menu, see Instrument diagnosis, page 31.
5.2.9. Switching off
Unsaved measured values are lost if the flue gas analyser is switched off.
> Press [].
- The pump may start and the sensors may be rinsed until the
switch-off thresholds (O2 > 20%, other measurement parameters < 50 ppm) are reached. The maximum rinsing period is 3 minutes.
- The measuring instrument switches off.
5.3. Address/Location
All measuring values can be saved under the currently active location. Measuring values that have not been saved are lost when the measuring instrument is switched off!
Addresses and locations can be created, edited, copied and enabled. Addresses and locations (incl. protocols) can be deleted.
Call up function: > [] Address/Location [OK].
There are various options for opening address.
1. Edit search setting: [Edit].
2. Select search setting: [], [] [OK].
Possible settings:
28
Page 29
5 First steps
Show all: All address/location are displayed.
Search: A search text only brings up address/location that
contain characteristics of the search text.
Filter: Individual letters or numbers can be selected. All data
beginning with the relevant letter/number is displayed.
The initial letter is the determining factor for the filter function, and this can only be selected individually. The
search function can also be used to find a series of several
letters within the address!
3. Carry out search according to search setting: [Search]
Show all
1. Select address: [], [].
2. Show details: [Details].
3. Enable a location: select the location [OK].
- The location is activated.
> Open measurements menu: press [OK] again.
Search
1. Edit search criteria: [] [Edit].
2. Select search criteria: [], [] [OK]. Possible options:
• Contact person
• Address
• Town/city
• Postcode
• Street
- The selected criterion is displayed.
3. Call up entry field for search text: [] or []
> Enter search text [Finished]
Do not use the special character * as a placeholder.
Filter
1. Edit search criteria: [Edit].
2. Select search criteria: [], [] [OK]. Possible options:
• Contact person
• Address
• Town/city
Page 30
5 First steps
• Postcode
• Street
- The selected criterion is displayed.
3. Enable tab: []
4. Select the required tab: [], [] and sometimes [], []
[Filter].
- The search result for the relevant letter or number is displayed.
Create a new measuring location: A location is always created under an address.
1. Select the address in which the location is to be created.
2. [Options] New/Location [OK].
3. Enter values or make settings.
4. Finalise the entry: [Finished].
Other location options:
> [Options] Edit location: make changes to an existing
location.
> [Options] Copy location: make a copy of an existing
location in the same address.
> [Options] Delete location: delete an existing location.
Create new address:
1. [Options] New address [OK].
2. Enter values or make settings.
3. Finalise the entry: [Finished].
30
Other address options:
Edit address: make changes to an existing folder.
Copy address: make a copy of an existing folder.
Delete address: delete an existing folder, including the
locations created therein.
Delete All addresses: delete all existing folders, including the
locations created in them.
Page 31
5.4. Measurement records
Call up function: > [] Measurement records [OK].
There are various options for opening records. see Address/Location, page 28.
Displaying a record:
1. Select the required record from the detailed view.
2. [Data].
Options:
> [Options] Delete Record: delete the selected record. > [Options] Number of lines: change the number of
measured values per display page.
> [Options] Delete all Records: delete all saved records for a
location.
5.5. Instrument diagnosis
Important operating values and instrument data are displayed. A gas path check can be carried out. The status of the sensors and any instrument errors not yet rectified can be displayed.
Call up function: > [] Instrument diagnosis [OK].
5 First steps
Displaying instrument errors:
> Error diagnosis[OK].
- Unrectified errors are displayed. > Display next/previous error: [], [].
Displaying sensor diagnosis:
1. Sensor diagnosis [OK].
2. Select sensor. [], [].
- The status of the sensor is indicated by a traffic light.
A sensor is able to recover. The sensor status indication may therefore change from yellow to green or from red to yellow.
Displaying instrument information
> Device information [OK].
- Information is displayed.
Page 32
6 Using the product
6 Using the product
6.1. Performing settings
6.1.1. Assigning the right function key
The right function key can have a function from the Options menu assigned to it. The menu Options is accessed via the left function key and is available in many different menus. This assignment is only valid for the currently opened menu / the opened function.
A menu / function is opened in which the Options menu is
6.1.2. Instrument settings
displayed on the left function key.
1. Press [Options] .
2. Select option: [], []. Depending on the menu / function from which the Options menu
was opened, the following functions are available.
3. Assign the selected function to the right function key. Press
It is assumed that the contents of the chapter First steps (see First steps, page 24) are known.
Calling up a function: > [] Instrument Settings.
see First steps, page 24
6.1.2.1. Measurement view
The parameters/units and the display (number of measured values shown per display page) can be set.
The settings are only valid for the currently chosen measurement type, which is indicated by the symbol in the info field.
Total overview of selectable measurement parameters and units (available selection depends on the set country version and selected measurement type):
Display Measurement parameter
FT
AT
32
Flue gas temperature
Combustion air temperature
Page 33
Display Measurement parameter
GT
O2
CO2
qA+
Instrument temperature
Oxygen
Carbon dioxide
Flue gas loss with due consideration of the calorific value range
η+
Efficiency with due consideration of the calorific value range
CO
COunv
λ
COumg
CO2um
O2ref
Draught
P
E draught
E-P
Carbon monoxide
Carbon monoxide undiluted
Air ratio
Ambient carbon monoxide
Ambient carbon dioxide
Oxygen reference
Draught measurement
Differential pressure measurement
External draught (external micro pressure probe)
External differential pressure (external micro pressure probe)
cCO
ExAir
qA
Carbon dioxide reduction
Air surplus
Flue gas loss without due consideration of the calorific value range
η
Efficiency without consideration of the heat value range
Dew Pt
Nett
GI
ET
Flue gas dew point temperature
Differential temperature
Toxin index
qA+ - qA
6 Using the product
Page 34
6 Using the product
Calling up the function:
> [] Device settings [OK] Measurement view
[OK]
Changing the parameter/unit in a line:
1. Select the line: [], [] [Edit]
2. Select the parameter: [], [] [OK]
3. Select the unit: [], [] [OK]
4. Save changes: [OK]
Options:
> [Options] Number of lines: change the number of
measured values per display page.
> [Options] Blank line: insert a blank line in front of the
selected line. > [Options] Delete line: delete the selected line. > [Options] Factory setting: reset the measured value
display to the factory settings.
6.1.2.2. Alarm limits
Alarm limits can be set for several display parameters. An audible alarm signal is triggered when the alarm limit is reached.
Calling up the function: > [] Instrument Settings [OK] Alarm Limits [OK]
34
Switching alarm signals on / off, changing alarm limits:
1. Select function or parameter: [], [] [Edit].
2. Set parameter: [], [] and partly [], [][OK].
3. Save changes: [Finished]. > Reset the enabled value to the factory setting: [Standard].
Page 35
6.1.2.3. Units
The units used for parameters in configuration menus can be set. Call up function:
> [] Device settings [OK] Units [OK].
6 Using the product
Adjustable units
Parameter Unit
Altitude m, ft Pressure mbar, hPa
1. Select the line: [], [][Edit].
2. Select the unit to be changed: [], [] [OK].
3. Confirm the entry: [Finished].
6.1.2.4. Date / time
Date, time mode and time can be set. Calling up the function:
> [] Instrument Settings [OK] Date/Time [OK]
Setting date/time:
1. Select parameter: [], [], [] [Edit].
2. Set parameter: [], [] and partly [], [][OK].
3. Save changes: [Save].
6.1.2.5. Energy management
Automatic instrument shutdown (Auto-Off) and switching off of the display light in battery operation can be set.
Calling up the function: > [] Instrument Settings [OK] Energy Management
[OK]
Making settings:
1. Select function or parameter: [], [] [Edit].
2. Set parameter: [], [] and partly [], [][OK].
3. Save changes: [Finished].
Page 36
6 Using the product
6.1.2.6. Display brightness
The intensity of the display illumination can be set. Calling up the function:
> [] Instrument Settings [OK] Display Brightness
[OK]
Performing settings > Set parameter: [], [][OK].
6.1.2.7. Printer
The headers (lines 1-3) and the footers for the printout can be set. The printer that is used can be activated.
Calling up the function: > [] Instrument Settings [OK] Printer [OK]
Activating the printer:
The printer 0554 0543 can only be selected after the Bluetooth Bluetooth
1. Select Printer [OK].
2. Select the printer: [], [] [OK].
- The printer is activated and the menu Printer is opened.
®
-interface has been activated, see
®
, page 36.
Setting the print text:
1. Print text [OK].
2. Select function: [], [] [Edit].
3. Enter values [Next].
4. Save the entry: [Finished].
see Bluetooth
6.1.2.8. Bluetooth®
This menu is only available if the instrument is equipped with Bluetooth. The Bluetooth module can be switched on / off.The relay
36
can now be tested.
Calling up the function: > [] Instrument Settings [OK] Bluetooth [Edit].
Making settings:
> Set parameter [OK].
®
, page 36
Page 37
6.1.2.9. Language
The menu language can be set. The number of available languages depends on the activated country version, see Country version, page 37.
Calling up the function: > [] Instrument Settings [OK] Language [OK]
Activating the language:
> Select the language [OK].
see Country version, page 37
6.1.2.10. Country version
Changing the country version may alter the basis for calculation and therefore also the displayed measurement parameters, fuels, fuel parameters and calculation formulas.
The selection of the country version influences the menu languages that can be enabled.
For information concerning the assignment table, the basis for calculation and the country version, see www.testo.com/download­center (registration required).
Calling up the function: > [] Instrument Settings [OK] Country Version
[OK]
This action can be password protected. A password is
specified in the menu Password Protection, see
Password protection, page 38.
Possibly: > Enter the password: [Enter] Enter password [Next]
[OK].
6 Using the product
Setting the country version:
1. Select the country version: [], [] [OK].
2. Confirm the confirmation request: Yes [OK]
- The system is restarted.
see Password protection, page 38
Page 38
6 Using the product
6.1.2.11. Password protection
The password protection is only valid for functions identified by the following symbol: or . Password protection can be activated / deactivated, the password
can be changed. To deactivate the password protection change the password to
0000 (factory setting).
Calling up the function: > [] Instrument Settings [OK] Password Protection
[OK] Possibly: > Enter the currently valid password:
[Enter] Enter password [Next] [OK].
Changing the password:
1. [Edit].
2. Enter the new password [Next].
3. [Edit].
4. Enter the new password again to confirm [Next].
5. Save changes: [Finished].
6.1.3. Sensor settings
6.1.3.1. O2 reference
The O2 reference value can be set. The O2 reference value setting may be password protected, see
Password protection, page 38. Call up function:
> [] Sensor settings O2 reference [Edit]. Possibly: > Enter the password: [Enter] Enter password [Next]
[OK].
reference:
2
38
Setting the O
> Set value [OK].
Page 39
6.1.3.2. Sensor protection
Protection limits can be set to protect the sensors against overload. Sensor protection switch-off is available for the CO sensor.
Sensor protection is activated if the threshold is exceeded. To disable sensor protection, the threshold values must be set to 0
ppm. Call up function:
> [] Sensor settings Sensor protection [OK].
Setting sensor protection thresholds:
1. Select parameter: [Edit].
2. Set value [OK].
3. Save changes: [Finished].
6.1.3.3. Recalibration/adjustment
The CO sensor can be recalibrated and adjusted. For recalibration/adjustment, Testo recommends using calibration adapter 0554 1205 or sending the instrument off to Testo Customer Service.
If obviously unrealistic measured values are displayed, the sensors should be checked (calibrated) and, if required, adjusted.
Adjustments made with low gas concentrations can lead to accuracy deviations in the upper measuring ranges.
Call up function: > [] Sensor settings Recalibration [OK].
Possibly: > Enter the password: [Enter] Enter password [Next]
[OK].
- Gas zeroing (30 s).
6 Using the product
Page 40
6 Using the product
Performing recalibration/adjustment:
WARNING
Dangerous gases
Danger of poisoning!
> Observe safety regulations/accident prevention regulations
when handling test gas. > Use test gases in well ventilated rooms only.
1. Connect the calibration adapter to the flue gas socket.
2. Enable CO measurement parameter: [OK].
3. [Edit] → Enter the test gas concentration (nominal value).
4. Attach the connecting line of the test gas bottle to the
calibration adapter.
5. Apply test gas to the sensor.
6. Start recalibration: [Start].
7. Apply the target value once the actual value is stable
(adjustment): [OK].
-or-
Cancel (no adjustment): [esc].
8. Save changes: [Finished].
6.1.4. Fuels
The fuel can be selected. The fuel-specific coefficients and limits can be set.
Call up function: > [] Fuels [OK].
In order to maintain the measuring accuracy of the instrument, the correct fuel must be selected or configured.
Correct representation of measuring results is only assured if the threshold values for the ideal range of the corresponding measurement task have been set correctly.
The pre-set threshold values are typical values for the selected system type and the chosen type of fuel.
40
Activating fuels:
> Select the fuel [OK].
- The fuel is activated and the main menu is opened.
Page 41
Setting coefficients:
1. Select the fuel [Coeff.].
2. Select the coefficients: [Edit]. Possibly: > Enter the password: [Enter] Enter password [Next]
[OK].
3. Set values [OK].
4. Save changes: [Finished].
Setting limits:
1. Select limit [Edit].
2. Set values [OK].
6.2. Measuring
3. Save changes: [Finished].
6.2.1. Preparing for measurement
The First steps chapter (see First steps, page 24) must have been read.
6.2.1.1. Testing for leaks
The entire measurement system (probe, condensate trap, hoses and connections) must be tested for leaks before each measurement to avoid incorrect measurements due to the infiltration of external air. Testing is carried out while the pump is running and may be performed by attaching a compressed balloon pump. The measurement system is leak-tight if the balloon pump is not filled with air.
6 Using the product
6.2.1.2. Zeroing phases
Measuring the combustion air temperature
If no combustion air temperature probe is connected, during the zeroing phase, the measured temperature of the flue gas probe is
taken as the combustion air temperature.
The flue gas probe should not be in the flue gas duct during the zeroing phase.
All dependent parameters are calculated using this value. This method of measuring combustion air temperature is sufficient for systems dependent on ambient air.
If a temperature probe is connected, the combustion air temperature is measured continuously via this probe.
Page 42
6 Using the product
Gas zeroing
When the instrument is switched on, the measurement menu is opened and the gas sensors are zeroed.
The flue gas probe must be in fresh air during the zeroing phase!
Draught/pressure zeroing
The pressure sensors are zeroed when a pressure measuring function is called up.
The flue gas probe must be in the fresh air during the zeroing phase / the instrument must not be pressurised during zeroing.
6.2.1.3. Using the modular flue gas probe
Checking the thermocouple
The thermocouple of the flue gas probe must not lie against the probe cage.
> Check before use. Bend the thermocouple back if necessary.
Aligning the flue gas probe
The flue gas must be able to flow freely past the thermocouple. > Align the probe by turning it as required.
42
Page 43
The tip of the probe must be in the centre of the flue gas flow. > Align the flue gas probe in the flue gas duct so that the tip is in
the core current (area of the highest flue gas temperature).
6.2.1.4. Measurement view
Only those parameters and units that are enabled in the measured value display appear in the measured value display, in the saved measurement records and on the record printouts.
> Before carrying out measurements, set up the measured value
display in such a way that the required parameters and units are enabled, see Measurement view, page 32.
6.2.1.5. Setting the location and fuel
Before carrying out measurements, the location and fuel must be correctly selected, see Address/Location, page 28 und see Fuels, page 40.
6 Using the product
Page 44
6 Using the product
6.2.2. Flue gas
To achieve usable measurement results, the measurement period of a flue gas measurement should be approx. 3 min and the measuring instrument should display stable measuring values.
Call up function:
1. [] Measurement options [OK]Flue Gas [OK].
2. Select the fuel [OK].
Carrying out the measurement:
1. Start measurement: [].
If a separate measurement of CO undiluted has not yet been carried out, this value is calculated using the measured values of the flue gas probe and is updated continuously.
If CO undiluted and/or a draught measurement has already been carried out separately, the value obtained is applied.
- The measured values are displayed.
2. End measurement: [].
Options
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted. > [Options] Save: the measured values are saved in a record. > [Options] Fluegas matrix: the measured values are
displayed as a flue gas matrix, see below. > [Options] Number of lines: change the number of
measured values per display page. > [Options] Recalibrate: the gas sensors are set to zero. > [Options] Measurement view: (This function is not available
during a measurement): the measured value display menu is
opened.
44
Page 45
3.
6 Using the product
Showing the flue gas matrix
This function is only available if the measurement parameter CO has been activated in the measured value display.
Call up function: The flue gas function is open. > [Options] Fluegas matrix.
Options
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] Save: the measured values are saved in a record. > [Options] Show numeric value: data is displayed as
numerical values.
> [Options] System type: (This function is not available
during a measurement) Set the system type to be able to configure the ideal zone (green) of the flue gas matrix, using the limits pre-configured for each system type.
> [Options] Reset graphic: the displayed graphical values are
deleted.
> [Options] Thresholds: (This function is not available during
a measurement) Enter limits to be able to configure the ideal zone (green) of the flue gas matrix.
> [Options] CO + O2 or CO + CO2: choose which parameter
should be assigned to the x-axis of the display matrix (O2 or CO2).
> [Options] Measurement view: (This function is not available
during a measurement) Open the measured value display menu.
Page 46
6 Using the product
6.2.3. Draught measurement
Call up function: A flue gas probe must be connected.
1. [] Measurement options [OK]Draught [OK].
Carrying out the measurement:
During the zeroing phase, the flue gas probe must be outside the flue gas duct.
Do not measure for longer than 5 min, as a drift of the pressure sensor means that the measured values may be outside the tolerance limits.
1. Start measurement: [].
- Draught zeroing is carried out.
2. Position the flue gas probe in the hot spot (area of the highest
flue gas temperature).
The display showing the maximum measured flue gas
temperature (AT max) helps when positioning the probe.
- The measured value is displayed.
3. End measurement [].
4.
46
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted. > [Options] Save: the measured values are saved in a record. > [Options] Measurement view: (This function is not available
during a measurement): the measured value display menu is
opened.
Page 47
6.2.4. External micro pressure probe
The following measurements can be performed using the external micro pressure probe (0638 0330):
Ext-Draught
Ext-Delta-P Single meas.
Ext-Delta Program
Ext 4Pa-Measurement (only available if Germany country version is selected)
Heating Check (only available if Germany country version is selected)
5.
See instruction manual for external micro pressure probe.
6.2.5. Average
This function is only available when the Italy country version is selected.
Call up function: A flue gas probe or a multi-hole probe (0554 5762) is
connected.
> [] Measurement options [OK]Average [OK].
6 Using the product
Options:
> [Options] Recalibrate: the gas sensors are set to zero. > [Options] Addresse/Location: the Address/Location folder
is opened.
> [Options] Fuels: select fuel. > [Options] Sensor detection: once the probe has been
changed, start sensor detection manually.
To calculate the average, a series of 3 measurements are carried out.
Averaging:
1. Position the flue gas probe in the centre of flow (area of the highest flue gas temperature).
2. Start measurements
> First measurement: []. > Second and third measurement: [OK]
- The set measurement parameters, measurement period and measured values are displayed.
Page 48
6 Using the product
- A signal is sounded after 2 min (recommended measurement
period)
3. End measurements: [].
- Once the series of measurements has been carried out, the
record for averaging is displayed.
> If necessary, scroll through the record: [], []
4. [Next]
5. Enter checks:
> Select criterion: [], []. > Change value: [Edit] [], [] [OK].
6. End check: [Close]
- The record is saved.
7.
6.2.6. BImSchV
This function is only available when the Germany country version is selected.
A qA average value measurement can be carried out. For this purpose the average is determined continuously over a period of 30 s, the measuring cycle takes 1 s. The average values actually valid at the corresponding time of recording are displayed.
Call up function: A flue gas probe and a combustion air temperature probe must
be connected.
> [] Measurement options [OK]BImSchV [OK]. > Select the fuel [OK].
48
Carrying out the measurement:
1. Start the measurement series: []
Wait for the balancing time, until O 20%.
shows a value below
2
2. [Next].
- The qA measured values (O
, AT, VT) are determined (30 s).
2
- The measurement stops automatically.
- The measured values are displayed and saved automatically in
a record.
3. End measurement: [Close]
or End measurement and call up draught measurement function:
[Draught - Measuring].
Page 49
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] Save: the measured values are saved in a record. > [Options] Address/Location: the Address/Location folder is
4.
opened.
6.2.7. CO undiluted
Call up function: A multi-hole probe (0554 5762) should be connected.
> [] Measurement options [OK]CO undiluted
[OK].
Carrying out the measurement:
1. Start measurement: []
- The measured value is displayed.
2. End measurement: []
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] Save: the measured values are saved in a record.
6 Using the product
6.2.8. Smoke number/HCT
Calling up the function:
> [] Measurement options [OK]Smoke
number/HCT[OK].
The parameters Smoke No. and Oil depos. are only available for oil fuels.
Determine smoke tester no./smoke nos./oil depos. with the smoke pump and enter manually:
1. Select parameter [Edit].
2. Enter data or values [Next] or [OK].
Page 50
6 Using the product
Determine smoke tester no./smoke nos./oil depos. with the smoke tester testo 308 and transmit wirelessly:
- The testo 308 must be in data transfer mode ( lights up). > [Options] t308.
- The values recorded by the smoke tester are transferred to the
testo 330.
Entering the heat carrier temperature:
> Heat carrier. [Edit] enter value [OK].
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted. > [Options] Save: the measured values are saved in a record. > [Options] Reset values: The entered values are deleted.
6.2.9. Differential pressure
The gas pressure set (0554 1203) must be connected. Call up function: > [] Measurement options [OK]Differential
pressure [OK].
50
Carrying out the measurement:
WARNING
Dangerous mixture of gases
Danger of explosion!
> Make sure there are no leaks between the sampling point and
the measuring instrument. > Do not smoke or use naked flames during measurement.
Do not measure for longer than 5 min, as the drift of the pressure sensor could mean that the measured values are outside the tolerance limits.
1. Connect the gas pressure set to the probe socket.
2. Start measurement: [].
- Pressure zeroing is carried out (system must be unpressurised).
Page 51
6 Using the product
3. Pressurise the system.
- The measured value is displayed
4. End measurement: [].
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] Save: the measured values are saved in a record. > [Options] Measurement view: (This function is not available
during a measurement): the measured value display menu is opened.
6.2.10. Differential temperature
The differential temperature set (0554 1208) must be
connected.
Call up function: > [] Measurement options [OK]Differential
temperature [OK].
Carrying out the measurement:
1. Start measurement: [].
- The measured values and the calculated differential temperature (T1 - T2) are displayed.
2. End measurement: [].
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted.
> [Options] Save: the measured values are saved in a record. > [Options] Measurement view: (This function is not available
during a measurement): the measured value display menu is
opened.
Page 52
6 Using the product
6.2.11. O2 air
An O2 dual wall clearance probe (0632 1260) must be
connected. Call up function: > [] Measurement options [OK]O2air [OK].
Carrying out the measurement:
1. Start measurement: [].
- The measured value is displayed.
2. End measurement: [].
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted. > [Options] Save: the measured values are saved in a record.
6.2.12. Gas flow rate
The function is only available if the chosen fuel is a gas. Call up function:
> [] Measurement options [OK]Flowrate [OK].
52
Carrying out the measurement:
1. Make a note of the gas meter value.
2. Start measurement: [].
3. When the adjusted gas flow is reached: [].
- The calculated gas flow rate and the gas burner capacity (in
kW) are displayed.
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted. > [Options] Save: the measured values are saved in a record. > [Options] Enter Gas Flow: set the gas flow value. > [Options] Change unit: The unit for the gas flow can be
changed (m3 > l or l > m3).
Page 53
6.2.13. Oil flow
The function is only available if the chosen fuel is an oil. Calling up the function:
> [] Measurements [OK]Oil Flow [OK].
Performing the measurement:
1. Select the parameters Oil Flow (of the oil nozzle) and Oil
Pressure (no effect on calculation): [], [] [Edit].
2. Enter values. [], [] and partly [], [][OK].
- The calculated oil burner capacity (in kW) is displayed.
Options:
> [Options] Clipboard: Data are saved to the clipboard > [Options] Delete clipboard: Any data saved to the clipboard
is deleted.
> [Options] Save: The readings are saved in a protocol. > [Options] Edit Unit: The unit for the oil flow can be changed
(kg/h > gal/h or gal/h > kg/h).
6.2.14. Ambient CO
This function is not available when the Spain country version is selected.
An ambient CO probe (recommended) or a flue gas probe must
be connected.
Cigarette smoke influences the measurement by more than 50 ppm. The breath of a smoker influences the measurement by about 5 ppm.
When using an ambient CO probe, note that: The direction of flow of the gas has an effect on the accuracy of measurement. Frontal flow onto the probe leads to higher measured values. The best measurement results are achieved when the probe is moved gently backwards and forwards.
When using the ambient CO probe and the flue gas probe, note that: the probe must be in the fresh air (CO-free) during the zeroing phase!
Call up function: > [] Measurement options [OK]CO ambient
[OK].
6 Using the product
Page 54
6 Using the product
Carrying out the measurement:
1. Start measurement: [].
- The measurement starts and the measured value is displayed
graphically (trend display).
- An audible alarm signal is triggered when the alarm limit is
reached.
2. End measurement: [].
3. Confirm the message: [OK].
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted. > [Options] Save: the measured values are saved in a record. > [Options] Set alarm limits: the alarm limits menu is opened.
6.2.15. CO2 ambient
This function is not available when the Spain country version is selected.
An ambient CO2 probe (0632 1240) must be connected.
In order to obtain correct measured values, it is imperative to enter the prevailing absolute pressure. This can be entered directly (Pressure absolute), or it is automatically calculated when entering Altitude and barometric pressure (Barometric pressure).
Call up function: > [] Measurement options [OK]AmbCO2 [OK].
54
Carrying out the measurement:
1. Select parameter [Edit].
2. Enter values. [], [] and in some cases [], [] [OK].
3. Start measurement: [].
4. End measurement: [].
- The ambient CO2 value is displayed.
Options:
> [Options] Clipboard: data is saved to the clipboard. > [Options] Delete clipboard: any data saved to the clipboard
is deleted. > [Options] Save: the measured values are saved in a record.
Page 55
> [Options] Alarm limit: the alarm limits menu is opened. > [Options] Edit: values for adjustable parameters can be
edited.
> [Options] Measurement view: (This function is not available
during a measurement) The measured value display menu is opened.
6.2.16. Leak detection
This function is not available when the Spain country version is selected.
During gas leak detection, no measurement is carried out, but gas detection is performed.
A gas leak probe (0632 3330) must be connected.
You must also refer to the documentation that comes with the gas leak probe.
Call up function: > [] Measurement options [OK]Gas leak detection
[OK].
Carrying out detection:
> Set the gas type to be detected and perform gas leak testing by
following the instructions described in the documentation of the gas leak probe.
1. Start detection: [].
- The gas concentration is displayed, and if it exceeds the limit, an alarm signal will sound.
6 Using the product
Options:
> [Options] Alarm limit: (this function is not available during a
measurement) Adjusting alarm limits.
> [Options] Alarm signal: (this function is not available during
a measurement) Disable/enable alarm signal.
> [Options] Zeroing probe: perform zeroing.
6.3. Transferring data
2. End detection: [].
6.3.1. Report printer
To be able to transmit data via infrared or Bluetooth interface to a Testo report printer, the printer to be used must have been activated, see Printer, page 36.
Page 56
7 Maintaining the product
Printing out data takes place via [Print] or []. The function is only available if a printout is possible.
Representation of the available characters per line is limited on the printout. Thus, the unit mg/KWh is displayed on the printout as mg/k.
6.3.2. PC / Pocket PC
Data transfer to a PC can take place via USB, IrDA or Bluetooth®. Data transfer to a Pocket PC can take place via IrDA or Bluetooth You must also refer to the documentation that comes with the
7 Maintaining the product
software.
7.1. Cleaning the measuring instrument
> If the housing of the measuring instrument is dirty, clean it with
a damp cloth. Do not use any aggressive cleaning agents or solvents! Mild household cleaning agents and soap suds may be used.
7.2. Replacing the rechargeable battery
The measuring instrument must not be connected to a mains
socket via the mains unit. The instrument must be switched off.
Change the rechargeable battery within 2 minutes so that instrument settings (e.g. date/time) are not lost.
1. Place the measuring instrument on its front.
®
.
56
2. Unscrew, lift up and remove the service cover.
Page 57
7 Maintaining the product
3. Press down on the holder lightly with your fingers. Pull the retaining clip in the direction of the arrow until the catch is released. Remove holder.
4. Unplug the plug-in connection from the slot.
5. Carefully pull the retaining clips (1, 2) outwards and push rechargeable battery up and out of the holder (3).
6. Insert the new rechargeable battery in the holder. Make sure that the plug-in connection cable is routed out of the holder at the side. Press the rechargeable battery into the holder until the retaining clips click into place.
7. Plug the new rechargeable battery plug-in connection into the slot.
8. Insert the holder into the guide rail and slide it in the direction of the arrow until the holder clicks into place.
9. Refit and close the service cover.
Page 58
7 Maintaining the product
7.3. Charging the battery
The rechargeable battery can only be charged at an ambient temperature of ±0 to +35°C. If the rechargeable battery has been completely discharged, the charging time at room temperature with the testo mains unit is approx. 6 h.
Charging in the measuring instrument
1. Connect the mains unit instrument plug to the instrument's
micro USB socket.
2. Connect the mains plug of the mains unit to a mains socket.
- The charging process will start. The charging process will stop
automatically when the battery is fully charged.
Battery care
> Do not fully exhaust rechargeable batteries. > Store rechargeable batteries only in charged condition and at
low temperatures, but not below 0°C (best storage conditions with a charge level of 50-80%, at an ambient temperature of 10­20°C, recharge completely before use).
7.4. Replacing sensors
The measuring instrument must be switched off.
1. Place the measuring instrument on its front.
2. Unscrew, lift up and remove the service cover.
3. Disconnect the hose connections from the faulty sensor/bridge.
4. Remove the faulty sensor/bridge from the slot.
5. Install new sensor / new bridge in the slot.
58
Page 59
7 Maintaining the product
6. Push the hose connections on to the sensor/bridge.
7. Refit and close the service cover.
After replacing an O2 sensor, wait for an equalisation period of 60 min to elapse before using the instrument again.
When retrofitting a sensor, the associated measurement parameter and unit must be enabled, see Sensor settings, page 38.
7.5. Recalibrating/adjusting sensors
See Recalibration/adjustment, page 39.
7.6. Cleaning the modular flue gas probe
Disconnect the flue gas probe from the measuring instrument
prior to cleaning.
1. Release the probe catch by pressing the key on the probe handle and remove the probe module.
2. Blow compressed air through the flue gas ducts in probe module and probe handle (see illustration). Do not use a brush!
3. Fit a new probe module on the probe handle and engage it in
place.
Page 60
7 Maintaining the product
7.7. Replacing the probe module
Disconnect the flue gas probe from the measuring instrument
prior to cleaning.
1. Press the key on the top of the probe handle and remove the
probe module.
2. Plug in the new probe module and engage it in place.
7.8. Changing the thermocouple
7.8.1. Modular flue gas probe
1. Release the probe catch by pressing the key on the probe
handle and remove the probe module.
2. Remove the thermocouple plug-in head from the socket using a
screwdriver and pull the thermocouple out of the probe shaft.
3. Insert the new thermocouple into the probe shaft until the
connection head clicks into place.
4. Fit a new probe module on the handle and engage in place.
7.8.2. Compact flue gas probe
1. Loosen and remove halfshell handles.
2. Turn sealing cap clockwise as far as it will go and remove.
3. Release and remove the halfshell element.
4. Detach adapter and hose, push thermocouple out of its holder
and disconnect cable from thermocouple
5. Connect cables to the new thermocouple (white -, green +) and
push thermocouple back into the holder.
60
Page 61
6. Connect adapter and hose to thermocouple. Adjust cables and hoses, and refit halfshell elements.
7. Refit sealing cap and turn anti-clockwise (as far as it will go, observe markings).
8. Replace halfshell handles and secure with screws.
7.9. Condensate container
The fill level of the condensate container can be monitored via the markings on the condensate trap.
7 Maintaining the product
Draining the condensate container
The condensate consists of a weak mix of acids. Avoid skin contact. Make sure that the condensate does not run over the housing.
CAUTION
Damage to the sensors and the flue gas pump due to condensate entering the gas path!
> Do not empty the condensate container while the flue gas pump is in operation.
1. Open the condensate outlet on the condensate trap: pull out the plug to the stop.
Page 62
7 Maintaining the product
2. Let the condensate run out into a sink.
3. Wipe off any drops still on the condensate outlet with a cloth
and close the condensate outlet.
The condensate outlet must be completely closed (marking), otherwise measuring errors could be caused by infiltrated air.
7.10. Checking/replacing the particle filter
Checking the particle filter:
> Check the particle filter of the modular and compact flue gas
probe regularly for contamination: check visually by looking through the window of the filter chamber.
Replace the filter if there are signs of contamination.
62
Replacing the particle filter:
The filter chamber may contain condensate.
1. Open the filter chamber: Turn gently anti-clockwise.
2. Remove the filter cartridge and replace it with a new one (0554
3385).
3. Attach the filter chamber and lock it Turn gently clockwise.
Page 63
8 Tips and assistance
8.1. Questions and answers
Question Possible causes/solution
Rechargeable battery low > Switch to mains operation. Measuring instrument
switches off automatically or cannot be switched on
Batteries/rechargeable batteries
empty.
> Charge rechargeable battery or
switch to mains operation.
Error message: Cell protection enabled
The shutdown limit of the CO sensor was exceeded.
> Remove the probe from the stack.
Error message: Printing
not possible
• With printer 0554 0543: Bluetooth interface not enabled.
• Wrong printer enabled.
• Printer switched off.
• Printer outside wireless transmission range.
> Enable Bluetooth interface, see
Bluetooth
> Enable printer to be used, see
Printer, page 36.
> Switch on printer. > Move printer into wireless
transmission range.
If we could not answer your question, please contact your dealer or Testo Customer Service. For contact details, see back of this document or the website www.testo.com/service-contact.
8 Tips and assistance
®
, page 36.
Page 64
8 Tips and assistance
8.2. Accessories and spare parts
Printer
Description Article no.
Infrared high-speed printer 0554 0549 Bluetooth® printer, incl. rechargeable battery and
0554 0553
charging adapter Spare thermal paper for printer (6 rolls) 0554 0568
Modular flue gas probes
Description Article no.
Modular flue gas probe 180mm, 500 °C,
0600 9760
thermocouple 0.5 mm, probe shaft diameter: 8 mm Modular flue gas probe 300 mm, 500 °C,
0600 9761
thermocouple 0.5 mm, probe shaft diameter: 8 mm Modular flue gas probe 180mm, 500 °C,
0600 9762
thermocouple 0.5 mm, probe shaft diameter: 6 mm Modular flue gas probe 300 mm, 500 °C,
0600 9763
thermocouple 0.5 mm, probe shaft diameter: 6 mm Flexible modular flue gas probe, length 330 mm,
0600 9764 Tmax. 180 °C, short-term 200 °C, bending radius max. 90° for measurements at difficult to access locations
Compact flue gas probes
Description Article no.
Compact flue gas probe 180 mm, 500°C,
0600 9740 thermocouple 1.0 mm, probe shaft diameter: 6 mm incl. cone
Compact flue gas probe 300mm, 500°C,
0600 9741 thermocouple 1.0 mm, probe shaft diameter: 6 mm incl. cone
Flexible compact flue gas probe, length 330 mm,
0600 9742 Tmax. 180 °C, short-term 200 °C, bending radius max. 90° for measurements at difficult to access locations
64
Page 65
8 Tips and assistance
Probe modules/accessories for modular flue gas probes
Description Article no.
Probe shaft module 180mm, 500°C, thermocouple
0554 9760
0.5mm, probe shaft diameter: 8 mm Probe shaft module 300mm, 500°C, thermocouple
0554 9761
0.5mm, probe shaft diameter: 8 mm Probe shaft module 180mm, 500°C, thermocouple
0554 9762
0.5mm, probe shaft diameter: 6 mm Probe shaft module 300mm, 500°C, thermocouple
0554 9763
0.5mm, probe shaft diameter: 6 mm Probe shaft module 300 mm, 1000°C, thermocouple
0554 8764
1.0 mm, probe shaft diameter: 6 mm Probe shaft module 700mm, 1000°C, thermocouple
0554 8765
1.0 mm, probe shaft diameter: 6 mm Spare thermocouple for module 0554 9760,
0430 9760
0554 9762 Spare thermocouple for module 0554 9761,
0430 9761
0554 9763 Spare thermocouple for module 0554 8764 0430 8764 Spare thermocouple for module 0554 8765 0430 8765 Cone, 8mm, steel 0554 3330 Cone, 6mm, steel 0554 3329 Multi-hole probe shaft, length 300 mm, Ø 8 mm, for
0554 5762
CO averaging Multi-hole probe shaft, length 180 mm, Ø 8 mm, for
0554 5763
CO averaging Flexible probe shaft module 0554 9764 Hose extension 2.8 m, extension line probe -
0554 1202
instrument Particle filter for modular flue gas probe, 10 pieces 0554 3385
Probe modules/accessories for compact flue gas probes
Description Article no.
Spare thermocouple for 0600 9740 0430 0383 Spare thermocouple for 0600 9741 0430 0382 Particle filter for compact flue gas probe, 10 pieces 0554 0040
Page 66
8 Tips and assistance
Temperature probe
Description Article no.
Combustion air temperature probe, 300mm 0600 9791 Combustion air temperature probe, 190mm 0600 9787 Combustion air temperature probe, 60mm 0600 9797 Surface probe (angled) 0604 0994 Fast reaction surface sensor 0604 0194 Miniature ambient air sensor 0600 3692
Other probes
Description Article no.
O2 annular gap probe 0632 1260 Gas leak probe 0632 3330 Ambient CO probe 0632 3331 CO2 ambient probe (without connecting cable) 0632 1240 Connecting cable for CO2 ambient probe, 1.5 m 0430 0143 Gas pressure set: Draught path adapter, silicone
0554 1203 hose 4 mm / 6 mm, reducing cones
Smoke tester incl. oil, soot plates, for measuring
0554 0307 soot in flue gas
66
Spare sensors
Description Article no.
O2 sensor 0393 0003 CO sensor 0393 0053 CO sensor H2-compensated 0393 0105 COlow sensor 0393 0103
Cases
Basic system case with double floor for instrument,
0516 3331 probes and other accessories
Basic system case for instrument, probes and
0516 3330 accessories
Page 67
8 Tips and assistance
System case with tool pouch without contents, click-
0516 0329
fastened to basic system case Universal system case without compartments, click-
0516 0331
fastened to basic system case Instrument case (leather) with compartments for
0516 0303
instruments and accessories
Other accessories
Description Article no.
Mains unit incl. micro USB cable 0554 1105 Spare rechargeable battery 0515 0046 Readout adapter for automatic furnaces 0554 1206 Connecting cable instrument / PC 0449 0047 Easyheat (PC configuration software) 0554 3332 Draught set 0554 3150 External micro pressure probe 0638 0330 Set of capillary hoses 0554 1215 Adhesive kits (x50) for printouts, paper barcode
0554 0116
labels etc. Instrument cleaner (100 ml) 0554 1207 Straight Pitot tube 0635 2050 ISO Calibration Certificate Flue Gas 0520 0003
For a complete list of all accessories and spare parts, please refer to the product catalogues and brochures or look up our website
www.testo.com
Page 68
8 Tips and assistance
8.3. Updating the instrument software
Under www.testo.com/download-center you can download the current instrument software (Firmware) for testo 320 (registration required).
> Unplug the micro USB cable and switch off the testo 320.
1. Hold down [].
2. Plug in the micro USB cable, continue holding down [].
- The display shows Firmware update along the bottom edge.
3. Release [].
- Your PC recognises the testo 320 as a removable medium.
4. Copy the new file (ap320rel.bin) to the detected removable
medium.
- On the display the status bar progresses from left to right. This
process may take a few minutes.
- Once the instrument software (firmware) has been updated, the
system will automatically reboot and is ready for use.
5. Disconnect the connecting cable from the testo 320.
68
Page 69
8 Tips and assistance
Page 70
8 Tips and assistance
70
Page 71
Page 72
0970 3200 en 02 V01.00 en
Loading...