TESTO 312, 312-2, 312-3 Instruction Manual

Page 1
testo 312 - 2/3
Instruction manual
www. .com
information@itm.com1.800.561.8187
Page 2
Introduction2
Initial operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Diagram of instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operating the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Measurement example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Switching on instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Differential pressure measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Fine pressure/Draught measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connection options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Pre-test / Main test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Service ability test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Setting alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setting date/time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting timer for printout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Setting number of printouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Changing the units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Changing rechargeable battery or battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Ordering data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Pressure transducer (Accessories). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Dear Customer
You have made the right decision by purchasing testo 312.
testo 312 is designed to adjust and maintain gas heating systems.
Thousands of customers buy our high standard products every year. There are at least 7 good reasons for doing so.
1) Cost-performance ratio. Reliable quality at a fair price.
2) Extended warranty times of up to 3 years - depending on instrument!
3) We have the ideal solutions for your measuring tasks based on our expert experience gained over 40 years.
4) Our high quality standard is confirmed by the ISO 9001 certificate.
5) Of course, our instruments carry the CE symbol required by the EU.
6) Calibration certificates for all relevant parameters. Seminars, advice and calibration on location.
7) Our after-sales service. Contact us for more information.
Contents
Introduction
2
Instrument conforms with 2004/108/EEC
www. .com
information@itm.com1.800.561.8187
Page 3
No reading in display? Send instrument for service.
Function test
for both pressure nipples
Warnings
Safety instructions
Initial operation
3
Capacity display
Voltage >7.2 V Life of rechargeable battery: approx. 5 h Life of battery: approx. 15 h (at an ambient temperature of 20°C)
Flashing symbol, voltage: <7.2 V Life of rechargeable battery: approx. 15 min. Life of battery: approx. 45 min. If the rechargeable battery voltage drops below
6.7 V automatic switch-off follows as protection against total discharge.
Low Bat
++
Function check/Leak check
Prior the commencing any testing the complete measuring system (intrument, probe and hoses) should be checked for leaks
Temperature influence on the leak test
During any leak tests the ambient temperature must remain constant to obtain valid data.
Instrument should only be used by trained personnel.
Always adhere to any system safety instructions and never use the equipment in an explosive environment.
Under normal operating conditions gas will not escape from the instrument when connected to the gas pipe / system under test.
The instrument is not designed to operate while wet, or in an environment of condensing humidity
The instrument should only be used in the conditions and for the purposes for which it was constructed. Please take particular note of the safety instructions and the technical data.
The instrument should only be used in the operation and measurement ranges specified in the technical data.
www. .com
information@itm.com1.800.561.8187
Page 4
Initial operation
4
Symbol for alarm
On/Off
Print
Battery capacity
Start
measurement/
Freeze readings
Print button
Cancel/Restore
Audible alarm On/Off button
Alarm levels
Parameters
Parameters
Measuring units
Measuring units
Date/Time
On/Off button
OK button
Scroll buttons
Time
Pressure
connection
Freeze readings
Parameters
Readings
Diagram of instrument
www. .com
information@itm.com1.800.561.8187
Page 5
Initial operation
5
Operating the instrument
Keypad
- Scrolling buttons
Using the arrow buttons / you can scroll between the parameters, digits in date/time and the alarm limits can be set.
- Selection button
In the date/time and alarm limits menu, access to the variable parameters is via the arrow button pointing to the right . The units can also be changed. The parameters are set using the up and down buttons.
- Printing
All of the saved readings can be printed on the printer by pressing Print .
- I/O button
The instrument is switched on or off via the I/O button
. The display switches off automatically after 5 seconds once the instrument is switched off. This can be prevented by pressing any button. The instrument then goes to the Δp measurement menu. The readings in the display are usually lost when the instrument is switched off.
- Cancel button
The Cancel button is used to cancel a selected process or to leave a sub-menu.
- OK button
The changed parameters are saved by pressing .
- START/HOLD button
Start measurement by pressing . The displayed
readings are frozen by pressing again
.
- Alarm on/off button
Switches audible signal on or off.
www. .com
information@itm.com1.800.561.8187
Page 6
Self-test
Press button. Displays all segments for approx. 3 seconds.
Battery capacity / Instrument version (2 digit)
Displays battery/rechargeable battery voltage and instrument version for approx. 3 seconds.
Instrument version (3 digit)
If the button is pressed within 3 seconds the version number is shown in 3 digits. You then go to the measurement menu.
Instrument switches off
6
9.0
08:38
--------
1.0
Self-test
Measurement menu
Display for
battery
capacity /
instrument version
9.0
1.00
Instrument
version
Connect hose.
3 s
3 s
5 s
3 s
Measurement example: testo 312-2/-3
Switching on instrument
www. .com
information@itm.com1.800.561.8187
Page 7
7
08:38
2
09:15
13.0
08:38
-----
Measurement
menu
Switch on instrument
Wait on calibration phase (5 s)
Save reading. HOLD appears in display.
The saved readings can be printed.
Position pressure measurement hose. The reading appears in the display.
Start measurement
- Segment test (3 s)
- Voltage display (3 s)
(see page 6)
8
7
09:15
13.0
6
5
3
2
4
9
Measurement example: testo 312-2/-3
Differential pressure measurement
Connect pressure
measuring
hose
testo 312-2: Range: 0 to 200 hPa. Resolution: 0.1 hPa testo 312-3: Range: 0 to 6000 hPa. Resolution: 1 hPa
1
www. .com
information@itm.com1.800.561.8187
Page 8
8
5
152
--
08:38
-----
Measurement
menu
08:38
-----
Measurement
menu
Measurement menu selection with arrow button.
Switch on instrument.
- Segment test (3 s)
- Voltage display (3 s)
(See page 6)
- Function check
(See page 7)
3
2
1
4
09:25
2
09:25
- 0.24
Wait on calibration phase (5 s)
Save reading. HOLD appears in the display
The saved readings can be printed.
Position probe in flue gas duct. The reading appears in the display.
Start measurement.
9
8
7
6
10
Measurement example: testo 312-2
Fine pressure / Draught measurement
Connect probe.
09:25
13.0
testo 312-2: ΔP range 0 to 40 hPa. Resolution: 0.01 hPa testo 312-3: ΔP range 0 to 300 hPa. Resolution: 0.1 hPa
www. .com
information@itm.com1.800.561.8187
Page 9
9
Testing
pump
Balloon
pump
Compressor Inert gas
Connection hose
High pressure stepped plugs
Conical
test plugs
Single-pipe
counter cap
JunctionMini valve
Connect pressure transducer directly to instrument. Do not
use extensions. Apply
pressure.
Remove pressure from pipe
before detaching pressure
transducer from instrument.
Connection options
Measurement example: testo 312-2/-3
Suitable for:
- Pre-test
- Main test
- Testing service ability
- Checking pressure in water pipelines
Observe maximum pressure
range.
Positive pressure destroys the
instrument.
Pressure
drop set
Pressure transducer (for liquid substances); Refer to instructions on page 22.
www. .com
information@itm.com1.800.561.8187
Page 10
3
2
1
4
5
Insert single­pipe counter cap and tighten.
Example
Manual printout triggers an automatic printout after 10 minutes.
6
Pre-test / Main test
Measurement example: testo 312-2/-3
10
Switch on instrument.
Start measurement with START/HOLD button.
Ensure connection is correct.
Connect pressure drop set and
testo 312.
Compare printouts OK/not OK
Build up test pressure
Carry out function test! Gas should not get into the instrument when carrying out measurements on gas pipes.
Pressure drop
The complete measuring system (instrument, probe, hoses and connec-
tions) should be checked for a drop in
pressure e.g. using pressure drop test
set by attaching the single-valve stop.
Preliminary test
Pipes are checked, without fittings, with a test pressure of 1 bar. This pressure should not fall during the 10-minute duration of the test.
Main test
Pipes are checked, with fittings, with a test pressure of 110 mbar. This pressure should not fall during the 10-minute duration of the test. The measuring instrument should be able to recognise a drop in pressure of 0.1 mbar.
www. .com
information@itm.com1.800.561.8187
Page 11
11
Start measurement with START/HOLD button.
3
2
1
4
5
Insert single­pipe counter cap and tighten.
Printout of initial pressure
6
10
7
Switch on instrument.
Ensure connection is correct.
Connect pressure drop set and testo 312.
Freeze reading with START/HOLD
Audible signal after 1 min
8
9
Prints reading
Determines Δp
Build up test pressure.
Use slide rule to determine what measure to take
Measurement example: testo 312-2/-3
Service ability test
45
Carry out function test! Gas should not get into the instrument when carrying out measurements on gas pipes.
Pressure drop
The complete measuring system (instrument, probe, hoses and
connections) should be checked for a
drop in pressure e.g. using pressure
drop test set by attaching the single-
valve stop.
Checking if fit for use
The pipe is pumped with air up to the respective test pressure and the drop in pressure in one minute is measured.
www. .com
information@itm.com1.800.561.8187
Page 12
12
08:54
-----
Measurement
menu
Switch on
instrument.
- Segment test (3 s)
- Voltage display (3 s)
(See page 6)
3
2
1
4
5
-----
--
Measurement menu selection via arrow button.
08:54
-----
08:51
100.0
Increase or decrease alarm limits via arrow buttons.
Go automatically to measurement menu.
Differential pressure
Draught
08:54
- 0.04
6
8
7
Note:
If the alarm limit has a value of 0.00 or 0.0, the alarm of the corresponding alarm limit is switched off. The dP or draught alarm is inactivated following initialisation. Alarm symbols do not appear on the display. The alarm is deactivated by the manufacturer i.e. the alarm limit is set at 0.00 or 0.0. Press the alarm button once the alarm limit is changed. The symbol is visible.
After the measurement, the flashing is a sign that an alarm limit is set and was reached.
0 to 200 0 to 6000
0 to -40 0 to 300
testo 312-2
testo 312-3
Alarm setting range
Setting alarm limits
Differential pressure (testo 312-2/-3),
draught (testo 312-2)
Δ
P
www. .com
information@itm.com1.800.561.8187
Page 13
13
Setting date/time
08:34
-----
Measurement
menu
Switch on
instrument
- Segment test (3 s)
- Voltage display (3 s)
(See page 6)
3
2
1
5
4
6
Press and buttons simultaneously.
Increase or decrease digit with arrow buttons. Return to measurement menu via OK
Increase or decrease digit with arrow buttons. Press OK button to return to measurement menu
Increase or decrease digit with arrow buttons. Press OK button to return to measurement menu
Setting the hour
08:35
-----
7
Press button.
Setting minutes
08:35
-----
Press button
Setting the day
02.06
-----
www. .com
information@itm.com1.800.561.8187
Page 14
14
Save selection with OK button. Return to measurement menu selection.
Note
You can leave the menu by pressing ESC but the changes are not carried out.
8
Press button
Note:
The date display can be changed from Day/Month (EUR) to Month/Day (US) by pressing
.
Setting the month
02:06
-----
9
10
Press button.
Setting the year
1999
-----
Setting date/time
Increase or decrease digit with arrow buttons. Press OK button to return to measurement menu.
Increase or decrease digit with arrow buttons. Press OK button to return to measurement menu.
www. .com
information@itm.com1.800.561.8187
Page 15
15
08:34
-----
Measurement
menu
Switch on
instrument
- Segment test (3 sec.)
- Voltage display (3 sec.)
(See page 6)
3
2
1
5
4
Press button 2 sec.
Increase or reduce value using arrow buttons.
7
Alarm limit
display
8
Press button.
9
Press button. Return to measurement menu.
08:35
205
Timer
setting
6
08:35
00
Press button.
Increase or reduce value using arrow buttons.
Timer
setting
08:35
05
Setting timer for printout
www. .com
information@itm.com1.800.561.8187
Page 16
16
7
Timer display
8
Press button
Alarm limit display
9
Press button
10
Press button. Return to measurement menu.
Display timer
6
08:35
15
Press button
Increase or reduce number using arrow buttons.
Set number of
printouts
08:35
2
Setting number of printouts
08:34
-----
Measurement
menu
Switch on
instrument
- Segment test (3 sec.)
- Voltage display (3 sec.)
(See page 6)
3
2
1
5
4
Press button 2 sec.
www. .com
information@itm.com1.800.561.8187
Page 17
17
Changing the units
1
2
3
4
Selection saved. Return to measurement menu.
You can leave the menu by pressing but the changes are not made.
Press button
-
Measurement menu
Measurement menu selection with arrow button.
Select units by pressing :
Differential pressure measurement
hPa, mbar, mmH
2
O, inchH2O
Draught measurement hPa, mbar, mmH2O, inchH2O (testo 312-2 only)
08:57
- 20.4
www. .com
information@itm.com1.800.561.8187
Page 18
Phase Symbol Cause
General Low Bat is flashing Power to instrument is too low. If the
battery voltage drops below 6.7 volt, the instrument switches off automatically.
T in top line is Non-permitted ambient temperature flashing Adapt to ambient temperature. The message Send instrument for service. Exxx appears in the bottom line
-- -- -- -- Pressure sensor is overloaded.
Error messages
18
Remove empty/defect rechargeable batteries or empty batteries from the battery compartment and replace with new ones.
Remove rechargeable batteries/batteries if the instrument is not used for a longer period of time. If the battery cells leak in the instrument, return instrument to factory to be cleaned and checked. Only the rechargeable batteries/batteries specified in the technical data should be used.
Observe correct polarisation of rechargeable batteries or
batteries.
Instrument should only be used if the battery
compartment cover is closed.
Switch off instrument and disconnect from all
measuring circuits before changing the rechargeable
battery or battery.
Date and time are saved when the instrument is switched
off.
Maintenance
Changing the rechargeable battery or battery
1
2
3
4
old
new
If we were unable to answer your question, please contact your distributor or Testo Customer Service. For contact data, see back of this document or web page www.testo.com/service-contact
www. .com
information@itm.com1.800.561.8187
Page 19
19
testo 312
Technical data
testo 312-2
Measuring range* ±40 hPa ±200 hPa Accuracy** <3.00 hPa ±0.03 hPa ±0.5 hPa (0 to 50 hPa)
>3.00 hPa ±1.5% of m.v.±1,5% of m.v. (>50 hPa) Resolution ±0.01 hPa ±0.1 hPa Alarm limit -0.01 to -40 hPa 0.1 to 200 hPa Adjustment step 0.01 hPa 0.1 hPa Alarm display Audible and optical Max. overload 1 bar
Differential pressure measurement
Draught (testo 312-2 only)
testo 312-2
Measuring range* ±40 hPa Accuracy** <3.00 hPa ±0.03 hPa
>3.00 hPa ±1.5% of m.v. Resolution ±0.01 hPa Alarm limit -0.01 to -40 hPa Adjustment step 0.01 hPa Alarm display Audible and optical Max. overload 1 bar
testo 312-3
Measuring range* ±300 hPa ±6000 hPa Accuracy** ±0.5 hPa <50 hPa ±4 hPa <400 hPa
±1.5% of m.v. 50 hPa ±2% of m.v. 400-2000 hPa
±4% of m.v. >2000 hPa Resolution ±0.1 hPa ±1 hPa Alarm limit 0.1 to 300 hPa 1 to 6000 hPa Adjustment step 0.1 hPa 1 hPa Alarm display Acoustic and optical Max. overload 8 bar
Δ
P
General Technical data for testo 312-2 and testo 312-3
Storage temperature: -20 to +60 °C Operation temperature: +5 to +45 °C Dimensions: 215mm x 68mm x 47mm Weight: Approx. 400 g
* Measuring instrument is temperature-compensated ** Accuracies do not apply in connection with pressure transducers
0554.3159 and 0554.3168
Warranty
Instrument: 2 years Probes: 1 year Accessories 6 months Printer 1 year (excluding printing mechanism)
www. .com
information@itm.com1.800.561.8187
Page 20
20
testo 312
Ordering data
Description
Part no.
Instrument
testo 312-2 compact pressure gauge with
instruction manual and battery 0632.0313
testo 312-3 compact pressure gauge with instruction manual and battery 0632.0314
Accessories
Testo printer, prints measurement data with location,
date and time 0554.0547
Spare rolls for printer 0554.0569 Rechargeable battery for printer (4 off) 0515.3120 Rechargeable battery 9V for instrument 0515.0025 Recharger for 9 V rechargeable battery 0554.0025 Recharger for printer 0554.0110 Pressure set with flue probe 0554.3150 TopSafe, indestructible protection case 0516.0443 Case 0516.0191 Service case 0516.3120 Pressure drop test set, 200 mbar 0554.3153 Test pump for pressure drop test set 0554.3157 Single-pipe counter cap 0554.3156 Two-valve junction 0554.3161 Single-valve stop 0554.3162 Connection hose LW6 0554.3158 Conical test plug 16-32mm 0554.3151 Conical test plug 24-44mm 0554.3155 Conical test plug 35-65mm 0554.3152 High pressure stepped plug 3/8” and 3/4” 0554.3163 High pressure stepped plug 1/2” and 1” 0554.3164 Complete high pressure set with case 0554.3160 Pressure transducer for liquid substances, 1 to 6000 mbar 0554.3159 Pressure transducer for liquid substances, 1 to 1000 mbar 0554.3168 System case 0554.3165 Leak detection spray 0554.3166 Slide rule 0554.3169 Test system set 0563.0314
www. .com
information@itm.com1.800.561.8187
Page 21
Normal use
The pressure transducer protects the testo 312 measuring instrument from moisture and high temperatures: measures vapour and water pressure.
Measuring
Do not use an extension cable between the pressure transducer and the instrument.
1 Connect connection A of the pressure transducer
directly to the instrument (+).
2 Switch on instrument. 3 Press “Start” and wait for calibration phase (5s). 4 Connect connection P of the pressure transducer
directly to the control pressure pipe.
Observe maximum pressure range of instrument and the pressure transducer used!
5 Apply pressure to pipe. 6 Take readings. 7 Press “Hold”. 8 Remove pressure from pipe. 9 Disconnect pressure transducer from instrument.
Technical data
Connection size: ..........1/8” plug-in connection
Size
(Height/Diameter): ........33mm/64.8mm
Weight: ........................175g
Housing material:..........Aluminium
Pressure range: ............0-1bar (0554.3168)
....................................0-6bar (0554.3159)
Pressure loss: ..............0 to 6% of reading
Diaphragm: ..................Temperature resistant to 120°C,
....................................oil resisting
Overload:......................2bar (0554.3168)
....................................8bar (0554.3159)
Pressure transducer (Accessories)
Instructions
Connection
A
Connection P
21
www. .com
information@itm.com1.800.561.8187
Page 22
testo AG
Postfach 11 40, 79849 Lenzkirch Testo-Straße 1, 79853 Lenzkirch
Telefon: (07653) 681-0 Fax: (07653) 681-100
E-Mail: info@testo.de Internet: http://www.testo.com
0973 3124 en 02 V01.04 en_DE
www. .com
information@itm.com1.800.561.8187
Loading...