TESTO 300 XXL Instruction Manual

testo 300 XXL
Instruction manual
en
3
Contents
Page
Preface ............................................................................................................................7
Principal safety information ..............................................................................................8
System description 9
Control unit ..............................................................................................................9
Device operation ....................................................................................................10
Touchscreen (option) ..............................................................................................11
Overview of the measuring system..........................................................................12
Menu overview 14
Control unit with analyser box..........................................................................14
Control unit......................................................................................................15
Commissioning ..............................................................................................................16
Function menu and function bar..............................................................................16
Status display of the LED (analyser box)..................................................................17
Power supply ..........................................................................................................18
Power supply, analyser box/control unit ..........................................................18
Mains operation ......................................................................................18
Rechargeable battery operation ..............................................................18
Power supply, control unit................................................................................18
Mains operation ......................................................................................18
Rechargeable battery operation with battery pack ..................................18
Battery operation ....................................................................................18
Rechargeable battery operation with standard batteries..........................18
Charging batteries ..................................................................................................19
Charging batteries, analyser box/control unit ..................................................19
Charging battery pack, control unit..................................................................19
Example measurement ..................................................................................................20
Correct use of the flue gas probe............................................................................20
Preparing for measurements with the control unit and analyser box........................21
Core current search with the control unit and analyser box ....................................22
Flue gas measurement at burners with control unit and analyser box ....................23
Conducting a CO measurement in gas burners
with the control unit and the analyser box ..............................................................24
Draught measurement with the control unit and analyser box ................................25
In compliance with the conformity certificate, the units fulfil the guidelines of DIN EN 61010-1 safety regulations for electrical devices, control, regulating
and laboratory equipment of guideline 73/23/EEC.
In compliance with the conformity certificate, the unit fulfils the guidelines of
2004/108/EEC.
Page
Differential pressure measurement with the control unit ..........................................26
Differential pressure measurement with the control unit ..........................................27
CO measurement with the control unit ....................................................................28
Gas leak probe with control unit..............................................................................29
Humidity measurement with the control unit............................................................30
Velocity measurement with the control unit..............................................................31
Volume measurement in a duct ......................................................................32
Mean value ......................................................................................................33
Volume flow measurement ..............................................................................34
Velocity measurement with the pitot tube ........................................................35
CO
2
measurement with the control unit ..................................................................36
Current/voltage measurement with the control unit ................................................37
Printing ..........................................................................................................................38
Printing the measurement results ............................................................................38
Main menu Memory (analyser box and control unit) ......................................................39
Overview ................................................................................................................39
Saving without selecting a location..........................................................................40
Creating a new location ..........................................................................................41
Selecting saved location..........................................................................................42
Using an existing location as format for a new location ..........................................43
Changing existing location ......................................................................................44
Deleting existing location ........................................................................................45
Barcode pen overview ............................................................................................46
Reading a location with the barcode pen ................................................................47
Reading a location from the measurement menu ....................................................47
Reading several locations with barcode pen from the measurement menu ............47
Displaying saved readings ......................................................................................48
Mean calculation ....................................................................................................49
Printing mean calculation result ..............................................................................50
Deleting entire memory............................................................................................51
Printing saved readings of a location ......................................................................52
Free memory? ........................................................................................................53
Main menu Probe for control unit ..................................................................................54
Scaling ....................................................................................................................54
Resetting ................................................................................................................55
Info..........................................................................................................................55
Main menu Input for control unit ....................................................................................56
Parameters..............................................................................................................56
Temperature ....................................................................................................56
Humidity ........................................................................................................57
Pressure ..........................................................................................................58
Density ............................................................................................................59
Contents
4
5
Contents
Page
Pitot tube factor ..............................................................................................60
Cross section ..................................................................................................61
Offset factor ....................................................................................................62
Info ..................................................................................................................62
Alarm limits..............................................................................................................63
Timer ......................................................................................................................63
Main menu Device for control unit ................................................................................64
Changing date/time ................................................................................................64
Setting the date ..............................................................................................64
Setting the time ..............................................................................................66
Auto off ..................................................................................................................66
Printer ....................................................................................................................67
Setting the contrast ........................................................................................67
Entering print text............................................................................................67
Light........................................................................................................................68
Diagnostic ..............................................................................................................68
Units ......................................................................................................................69
Main menu Service for control unit ................................................................................70
Operating values ....................................................................................................70
Resetting factory settings ........................................................................................70
Address ..................................................................................................................70
Device data ............................................................................................................70
Language ................................................................................................................71
Main menu Periphery (with touchscreen only) ................................................................71
Solid fuel measurement option for analyser box ............................................................72
CO measuring ranges in different dilution stages ....................................................73
Main menu Sensors for analyser box and control unit ..................................................73
Recalibration ..........................................................................................................74
NO
2
addition ..........................................................................................................75
Print cal. data..........................................................................................................75
Main menu Input for analyser box and control unit ........................................................76
Spot number index / WTT / oil derivatives ..............................................................76
Fuel ........................................................................................................................77
O
2
rel/CO2max/coefficient ........................................................................................77
Parameters..............................................................................................................79
Pressure ..........................................................................................................79
Pitot tube factor ..............................................................................................80
Cross section ..................................................................................................81
Offset factor ....................................................................................................82
Info ..................................................................................................................83
Dew point AT ..........................................................................................................84
Main menu Device for analyser box and control unit......................................................85
View (display sequence) ..........................................................................................85
Diagnostic ..............................................................................................................86
6
Contents
Page
Main menu Service for analyser box and control unit ....................................................87
Operating values ....................................................................................................87
Switching off CO/NO ..............................................................................................88
Device data ............................................................................................................89
Bus address............................................................................................................89
Maintenance ..................................................................................................................90
Inserting new paper roll in printer ............................................................................90
Changing the battery pack ......................................................................................91
Control unit......................................................................................................91
Analyser box....................................................................................................91
Filter change ..........................................................................................................92
Condensation trap ..................................................................................................92
Flue gas probe ........................................................................................................93
Cleaning the flue gas probe ............................................................................93
Cleaning the flue gas probe with gas path closed............................................93
Changing a defective thermocouple ................................................................93
Changing O
2
, CO and NO measuring cells..............................................................94
Installation of CO/NO2 measuring cell ....................................................................94
Installation of NO measuring cell ............................................................................94
Maintenance of the gas pumps ..............................................................................95
Opening the analyser box ................................................................................95
Cleaning the main gas pump and the purging pump ......................................95
Cleaning the main gas pump with optional gas preparation ............................96
Changing the pump cassette of the condensation pump
with optional gas preparation ..................................................................................97
Upgrades ................................................................................................................98
Calculation information ..................................................................................................99
Error messages ..........................................................................................................100
Technical data..............................................................................................................101
Ordering data ..............................................................................................................106
Notes
7
Preface
Dear Testo customer,
Your decision to buy the testo 300 XXL was an excellent choice.
The testo 300 XXL is designed for flue gas analysis and all measurements in heating systems.
Thousands of customers buy our high-quality products every year. There are at least 7 good reasons for this:
1) Our price/performance ratio is good. Reliable quality at a fair price.
2) Substantially extended warranty periods up to 2 years – depending on the model.
3) We provide an optimum solution to your measuring tasks with over 40 years of professional experience.
4) Our high quality standards are confirmed by the ISO 9001 certificate.
5) Of course, our devices also bear the CE label required by the EU.
6) Calibration certificates for all relevant measurements. Seminars, consulting and calibration on site.
7) We don’t leave you on your own after your purchase.
Use the device only in the operating and measuring ranges specified in “Technical data”.
Use the device only under the conditions and for the purpose for which it was designed. For this, particularly observe the safety information and the technical data.
To prevent damage to the instrumentation and items under test, suitably trained persons should only use the equipment.
Observe all local Health and Safety regulations before operating the test equipment.
Do not make measurements on live components.
Observe the warnings on the control unit and analyser box!
Magnets of the control unit
The magnets of the control unit can damage EC cards, telephone cards etc.
Operation with mains adapter
Use only the original mains adapter to operate the device. Replace damaged mains cable only with an original mains cable of the manufacturer or agent (i.e. by authorised service).
Check for leaks
The entire measuring system (probe, condensation trap, filter, hoses and connections) must be checked for leaks before each measurement, e.g. by pushing on a collapsed rubber bladder. Measuring errors can be caused if external air infiltrates.
Expose the gas outlet
When measuring, ensure that the gas outlets are exposed to allow the gas to escape unobstructed. If this is not the case, the measurement results can be falsified.
Empty the condensation trap
The condensation trap must only be emptied when the pump is switched off (the measuring cells are otherwise in danger of damage)!
Disposing of batteries
The end user is legally required to dispose of all used batteries in accordance with their country’s regulations.
Disposing of measuring cells
Storing the analyser in rooms containing solvents, vapours or gases will damage the measurement cells e.g. cleaning fluids, polishes, paints, cooking operations. Protect the device from direct sunlight and store at temperatures <45°C.
Storing the device
Storing the analyser in rooms containing solvents, vapours or gases will damage the measurement cells e.g. cleaning fluids, polishes, paints, cooking operations. Protect the device from direct sunlight and store at temperatures <45°C.
Charge the battery before use
Before making the first measurement or after several days of disuse, fully charge the internal battery (see menu Service Operating values). Charge the battery at least every 4 weeks during longer periods of disuse to prevent damagingly low discharging.
Changing the configuration
Switch off the device to change the configuration (e.g. probe change), as the device reads the probe-specific values when it is switched on.
PC cable connected
When the PC cable is connected, it is impossible to issue a print command.
Using probes with plug head
Ensure the good contact of probes with plug heads. Fully tighten the union nut on the probe handle.
Opening the device
If the device is opened, misused or subjected to force, the warranty will be voided.
Because the pressure sensor heats up, it is recommended to wait on the “Warm up” for 15 minutes to achieve the best possible accuracy.
8
Principal safety information
System description
Control unit
9
On/off button
Function keys
Function bar
Touchscreen (option)
on the rear:
Contacts to analyser box
Reading display
Device active display
Touchscreen pen holder
Printer
Reading window
Display lighting
Mains adapter connection
Pressure connection
Probe connection
Cancel/back
PC connection
Main menu display
Status display
Location
Acknowledgement
Cursor keys
Connection of the spiral BUS
cable to the analyser box
10
System description
Device operation
The analyser measurement box is controlled, read and programmed using the control unit.
The operation of the testo 300 XXL is designed so that the key functions are visible at all times on the display.
The keypad
The function keys allow rapid changes of the functions of the device and the overall control of the measurement. The significance of the function keys is indicated in the function bar. Not all function keys are used in every menu. Arrows in the function bar indicate further accessible functions in the menu. When the right or left cursor key is pressed, these additional functions are displayed and can be activated with the function key.
With the up /down cursor keys, you can browse between the reading windows (in the measurement menu) or select a menu item in a list. When you have reached the last window/menu item, the device automatically loops back to the first window or the first menu item.
With the Browse key, you can branch from the measurement menu to the main menu or from a submenu to the measurement menu.
The I/O button switches the device on and off. With the Cancel key, you can terminate
selected procedures or a chosen selection and exit submenus. When you exit a submenu, you will always move one menu window backwards until you reach the main menu.
With the Acknowledge key, the selection or the execution of the selected function is accepted. If the key is pressed in the measurement menu, the menu “Device selection list” appears. The menu provides information on which device is in use. 001=Control unit active, 002=Analyser box active etc. If an analyser box is connected, this is indicated automatically in the window when the device is started.
With the Lighting key , the display lighting can be switched on and off or the automatic timer can be started.
The display
The display constantly shows all important information.
In the Reading/Menu display area, 6 readings per window are indicated in the measurement menu. Only three readings are indicated in larger lettering when the zoom function is used (by function key).
The function bar indicates the significance of the individual function keys . Their effects can change according to the menu. The description of the function keys changes accordingly on the display.
The function bar can be assigned as desired to all functions of the function menu. When the key
is pressed together with a function key, the function menu opens. The entries in the function menu change according to the connected components of the measuring system.
The Operation display provides a graphic de- scription of the current mode of the device, e.g. whether the pump of the device is running and if the device is running in mains operation.
The Location box constantly shows whether and which location hasn’t been selected and is active. If no location has been selected, “NONAME” appears on the display.
In the measurement menu, the Reading window on the right indicates the measured value window in which you are currently working, e.g. 2/5 indicates that you are working in the second of 5 possible pages.
Errors are indicated on the display as soon as they occur. The “!” symbol in the status display often indicates the occurrence of errors.
The touchscreen function (option) allows direct access to the functions indicated on the display using a special pen. A brief touch is sufficient to activate the function.
11
System description
Touchscreen (option)
Touch a position on the display with the special pen. The main menu appears. Each menu item can be accessed and activated with the special pen. The example shows the individual steps to set the lighting status of the control unit.
Measurement menu
Measurement menu
Main menu
Change “Device” settings
Change “Device” settings
or
Adjustment menu
Adjust “Light”
Adjust “Light”
Set “On/Off”
Set automatic
PPeerriipphheerryy
PPeerriipphheerryy
12
System description
Overview of the measuring system
Measuring cell heater protects against destruction by condensation
Gas connection
DP
measurement
Spiral BUS cable
connection
Flue gas probe
connection
Probe
connection
110/230 V
50/60 Hz
Carrying
handle
Control unit
catch
Contact strips to the control unit
LED status display
Gas path filter
Gas outlet 2
Fresh air inlet filter Fresh air inlet
Diluting air filter Gas outlet 1
Condensation tank
13
System description
Overview of the measuring system
0
12.1
1.27
92.4
180.1
22.2
BrSt
l
AT °C
VT °C
h %
002
04/06
Noname
Hold
Max
Min
Zoom
Volume flow rate measurement with funnel
Pressure measurement
Temperature measuremen t
CO safety measurement
Gas leak measurement
CO
2/humidity
measurement
Current measurement Voltage measurement
Cable connection control unit/spiral analyser box cable
Control unit
4 user-defined function keys
Mains connection/ fast battery charge
Integrated differential pressure measurement
Integrated printer
Multi-probe input
RS 232 interface
Display
14
MEMORY
SENSORS
INPUT
DEVICE
SERVICE
MMaaiinn mmeennuu
Op. values Switch-off Device data Bus address
Display sequence
Units
View
Diagnostic
Spot number/HCT
Fuel O2rel/CO2max Parameter
Dewpoint/ambient
Smoke number 1 Smoke number 2 Smoke number 3 Heat carrier temp. Oil derivative Info
O2 rel. CO2max. coeff.
Temperature VL
Humidity VL Dew point VL
Pres. Pitot tube factor Cross-section Offset factor Info
Recal. NO2 addition Print cal. data
CO
NO
CO
NO
Read out Mean value Fuel meas. Delete memory Free memory?
System description
Menu overview Control unit with analyser box
from height
Absolute
Pres. Metres above m.s.l. Diff. pressure
Duration No. values Info Save Delete
Duration Dilution Info Save Delete
Yes
No
15
MEMORY
SENSORS
INPUT
DEVICE
SERVICE
PERIPHERY*
MMaaiinn mmeennuu
Op. values Reset factory Address Device data Lang.
Change date Auto Off Printer Light Diagnostic Units
Touchscreen cal.
Set date
°C, °F
m/s, fpm
ppm, %
Off, td°C, g/m3,
g/kg, J/g
m3/h, cfm, m3/m,
m
3
/s, l/s, M3/h
hPa, inW, mbar,
Pa, bar, psi
Parameter Alarm limits Timer
Temp. Humid. Pres. Density Pitot tube factor Cross-section Offset factor Info
CO limit CO2 limit dP limit 1 dP limit 2 Acoustic alarm
Duration No./values Info Save Delete
Circle Square Rectangle
Area
from height absolute
Pressure
Metres above m.s.l.
Diff. pressure
contrast
Print text
On/off
Automatic
Temp. Humid. Velocity Vol.-Flow Pressure Gas
yes no
Probe socket 1
min input max input min output max output Unit Decimal place Info Save Delete
Scaling Reset Info
Read out Mean Delete memory Free memory?
System description
Menu overview
Control unit
Off
5 Min.
10 Min.
....
*with touchscreen option only
Function menu and function bar
Commissioning
16
The function bar can be assigned as desired to all functions of the function menu. Pressing the key with a function key opens the function menu. The entries in the function menu change according to the connected components of the measuring system.
F
UNCTION MENU ANALYSER BOX/CONTROL UNIT
Pump starts. Flue gas measurement runs.
Doubles the character size.
Saves readings.
Differential temperature measurement with 2 probe inputs of the analyser box.
Graphic display of the temperature during core current search.
Fine draught with 0.001 hPa resolution.
Starts solid fuel measurement.
Prints readings.
Printer paper line feed.
Re-zeros the measuring cell.
Fresh air purging of the measuring cells.
Starts CO undiluted measurement.
F
UNCTION MENU, CONTROL UNIT
Doubles the character size.
Freezes the current readings.
Shows the highest value during a measurement.
Shows the lowest value during a measurement.
Calculates mean value.
Additional display for volume flow rate measurement.
Displays in metres per second.
Zeros external pressure probe.
Zeros CO probe.
Starts solid fuel measurement.
Analyser box is detected without restarting the device.
PPSSttaarrtt
uuCCOO oonn
SSttaarrtt
MMeemm..
DDeellttaa TT
CCoorrccuurr
DDrraagghhtt
PPrriinntt
ZZeerroo
GGaass
ZZoooomm
ZZoooomm
HHoolldd
MMiinn
MMaaxx
VVooll
MMeeaann
mm//ss
hhPPaa==00
ppppmm==00
SSeeaarrcchh
SSttaarrtt
LLFF PPrr
17
Commissioning
Function menu and function bar
LED 1
LED 2
LED 3
DDeellttaa TT
MMeemm..
ddPP 11
ddPP 22
PPrriinntt
LLFF PPrr
DDiiaagg..
EESSCC
OOKK
TTeesstt
CChhaannggee
EEnndd
IInnffoo
aa  AA
<<
Status display of the LED
Saves readings.
T1 + T2, displays differential temperature.
Differential pressure 0 - 80 mbar
Differential pressure 0 - 1000 mbar
Prints readings.
Printer paper line feed
Direct display of all error messages
F
UNCTION BAR
Cursor up (optional with touchscreen)
Cursor down (optional with touchscreen)
Menu exit (optional with touchscreen)
Confirms input (optional with touchscreen).
Set values come into effect
Change
Prints test page.
Accepts settings and closes window.
Upper/lower case switch
Back + deletes 1 character.
Space
Overview of settings
Sets standard values.
Current value comes into effect
CCuurrrr..
LED1 (Power):
Mains operation Green/Permanent Battery operation (batt. full) Green/Flashing Battery operation (batt. empty) Red/Flashing Battery recharging, Off mode Off
LED2 (Status):
Measuring Green/Permanent Fresh air/Zeroing Green/Flashing Defect Red/Flashing
LED3 (Battery recharging):
Battery recharging (fast charge) Green, flashing Batt. full, compensation charge Green, permanent
18
Commissioning
Power supply
115/230 V
Mains adapter
0554.1084
or
or
Plug mains cable into analyser box. The power supply for the control unit is ensured when
- control unit is connected to the analyser box by the contact strips
or
- control unit is connected to the analyser box by the spiral BUS cable
The power supply for the control unit in rechargeable battery operation is ensured when:
- control unit is connected to the analyser box by the contact strips
or
- control unit is connected to the analyser box by the spiral BUS cable
Observe battery capacity.
Ensure correct polarity.
There should always be rechargeable batteries/batteries in
the control unit.
Power supply, analyser box/control unit
Control unit power supply
MAINS OPERATION
Connect mains adapter to the control unit.
MAINS OPERATION
RECHARGEABLE BATTERY OPERATION
RECHARGEABLE BATTERY OPERATION with battery pack
(art. no. 0554.0097)
RECHARGEABLE BATTERY OPERATION WITH STANDARD
BATTERIES
BATTERY OPERATION
19
Battery charging
Commissioning
Charging batteries, analyser box/control unit
Charging battery pack, control unit
The battery packs must be installed in the analyser box and the control unit for charging.
Plug mains cable into analyser box.
Note
The units must be switched off. Charging is impossible during operation.
The charging of the battery pack for the control unit is ensured when:
- control unit is connected to the analyser box by the contact strips
or
- control unit is connected to the analyser box by the spiral BUS cable.
Charge level displays:
- Analysis box: Recharging is in progress: Recharging battery LED flashes green, Recharging is complete: Recharging battery LED lights up green
- Control unit: Shows charge level in display.
The battery pack must be installed in the control unit for charging.
Standard batteries cannot be charged.
Plug mains adapter into the control unit.
Note
The control unit must be switched off. Charging is impossible during operation.
Charge level display:
- Charge level is shown in display.
115/230 V
Mains adapter
0554.1084
The end user is legally required to dispose of all used batteries in
in accordance with their country’s regulations.
CChhaarrggiinngg CCoonnttrrooll UUnniitt
CChhaarrggiinngg CCoonnttrrooll UUnniitt
20
Example measurement
Correct use of the flue gas probe
The measurement of the flue gas temperature is made by the thermocouple at the tip of the flue gas probe. The probe tube protects the thermocouple, but has openings in to allow the stream of flue gas to envelop the thermocouple.
The
temperature
measurement is made during the zeroing phase. The flue gas probe can be adjusted in the flue at this time (Note: the AT probe must be connected). This makes it possible to conduct a core current search during the zeroing phase. The temperature measured by the flue gas probe is interpreted by the testo 300 XXL as the combustion air temperature unless a AT probe is connected (Note: the probe must be removed from the flue). When the zeroing phase has been completed, this value is saved as the combustion air temperature value.
All dependent parameters are calculated by this value.
incorrect
Flue gas stream
Thermocouple
incorrect
correct
correct
The thermocouple must always be exposed to the stream
of flue gas to allow an exact measurement of the flue
gas temperature and thereby a precise determination of the flue gas losses. It must not be screened by a web of the probe pipe.
The tip of the thermocouple must not touch the guard. If
necessary, bend the tip of the thermocouple accordingly.
Example measurement
21
Preparing for measurement
with the control unit and analyser box
Control unit + analyser box + flue gas probe
1 Connect flue gas probe to analyser box.
2 Switch on .
3 Zeroing phase
Any connected CO probe and the gas measuring cells of the analyser box are zeroed during the zeroing phase.
4 Fuel selection
The menu “Fuel selection” appears during the zeroing phase. Select the desired fuel with the keys or and confirm with .
5. Back to the zeroing phase.
Note
After zeroing phase back to the measurement menu with the key.
10 sec
Flue gas probe
Analyser box
Initialisation
Zeroing phase
Select the
fuel
Connect the flue gas probe before switching on the control unit and the analyser box.
Note
The analyser box does not need to be activated in the device selection menu. It is detected automatically when the measuring unit starts.
LLooccaattiioonn 33
22
Example measurement
Starting core current search:
1 Connect flue gas probe to analyser box.
2 Place flue gas probe in flue pipe.
3 Activate the function menu with the key and one of
the function keys. Select function key with
and confirm with .
4 Start core current search with .
5 Determine core current temperature.
6 Terminate core current search with .
Note
A core current search is also possible during the zeroing phase.
Corcur
Corcur
Core current search with the control unit and analyser box
Measurement menu
Measurement menu
Core current menu
Select core current
10 sec
Initialisation
Device zeroing
Automatic fuel display
Zeroing phase
3 x
1 sec
Function menu
CC oo rr ccuu rr
LLooccaattiioonn 33
LLooccaattiioonn 33
Preparing for measurement
1 Specify new location and select (see page 41):
2 The four function keys of the control unit are freely
configurable with the functions of the function menu.
3 Activate the function menu with the key and one of
the function keys. Select function key with
and confirm with .
4 Start measurement with .
5 Position the flue gas probe in the core current:
Seek flue gas temperature AT max, then fasten the probe with the cone in this position.
6 When the readings no longer change, stop the pump.
Place flue gas probe in fresh air. The readings are retained until the pump is next started.
PStart
PStart
Example measurement
Flue gas measurement at burners
with the control unit and the analyser box
23
Back to the measurement menu
If the flue gas probe is positioned vertically or at an angle
of up to 45°, falling drops of condensation can cause a
sudden drop in the temperature.
Remedy: Readjust flue gas probe.
Measurement menu
Location
Measurement menu
Function menu
Measurement menu
Measurement menu
1 sec
Conducting a CO measurement in gas burners (flue gas path test) with the control unit and analyser box
1 Activate the function menu with with the key and
one of the function keys. Select function key with and confirm with .
2 Start measurement with .
3 Measure undiluted CO with .
Display of the function key changes to .
4 Finish undiluted CO measurement with .
Undiluted CO is saved.
5 Finish measurement with .
PStop
uCO off
uCO off
uCO on
PStart
uCO on
uCO off
uCO on
24
Example measurement
Measurement menu
Function menu
Measurement menu
CO limit for the rejection of a system (in relation to
undiluted flue gas): > 1000 ppm uCO
Draught measurement
with the control unit and the analyser box
25
Preparing for measurement
Use the flue gas probe for the draught measurement. The draught measurement can be conducted before or after a flue gas measurement.
Start measurement
1 Activate the function menu with the key and one of
the function keys. Select function key with
and confirm with .
2 Start measurement with .
3 Place probe in flue.
4 Display of the current reading.
Draught
Draught
Example measurement
A draught measurement is only possible with the pump
switched off. The + pressure inlet of the device must be
open!
Do not change between battery and mains operation during the draught measurement (voltage fluctuations influence the result of the measurement)!
If you conduct the draught measurement after a flue gas measurement, a head of pressure exists in the hose after the pump has stopped, which must be dispelled. This occurs within approx. 30 seconds.
Remove any traces of condensation in the flue gas probe (shake out the probe with the tip towards the floor).
Sensor zeros automatically, probe must be outside
the flue.
Negative sign indicates vacuum, positive sign indicates pressure in the flue.
Select “Draught”
Measurement menu
Flue gas probe outside flue
Place probe in flue
Stable reading?
Autom. zeroing
Result
Measurement menu
Function menu
1 sec
DDrrgghhtt
Preparing for measurement
Selection of a suitable measuring range. Two measuring ranges P1 (-80 to +80 hPa) and P2 (-1000 to +1000 hPa) are available in the function menu:
Start measurement 1 Device selection: control unit
2 Activate the function menu with the key and one of
the function keys. Select the function key (low range) or (high range) with and confirm with .
3 Zero sensor with function key .
4 Start measurement with function key , stop
measurement with function key or the measure­ment is terminated automatically when the time has expired (Input menu - Timer).
5 Display and printing of the current reading.
TStop
TStart
dP1
dP2
dP1
26
Differential pressure measurement with the control unit
Example measurement
Do not use in the main duct!
Ensure that there are no leaks between the sampling point and the device (escaping gas and atmospheric oxygen can create an explosive mixture)!
Do not smoke or use naked flames during the measurement!
Observe the rated pressure of the connections!
Do not exceed the measuring range!
No gas enters the device because an air cushion forms between the sensor and the gas inlet. The measuring unit is therefore not protected against explosions.
Measurement menu
Measurement menu
Measurement menu
Function menu
Select “dP1” and/or
“dP2”
Automatic sensor
zeroing
Start measurement
Stop measurement
Print
1 sec
10 sec
dP1
TT SStt aarrtt
TT SStt oopp
01:00 min
00:27 min
27
Measurement with control unit + temperature probe
For pure temperature measurement, e.g. supply / return temperature
The testo 300 XXL control unit has a multi-probe input.
1 Activate the function menu with the key and one of
the function keys. Select the function key with and confirm with .
2 Measure T1: Press function key .
3 Measure T2: Press function key .
4 Display of the measured differential temperature value
Measurement with control unit + analyser box + 2 temperature probes
Device selection: flue gas measuring box
1 Configure function key with (see above)
2 display of the current value of Temp.1 (probe
connection), Temp.2 (flue gas probe connection) and the differential temperature.
DeltaT
DeltaT
T2
T1
DeltaT
Differential temperature measurement
with the control unit
Display
Temp. 1
Display
Temp. 2
Before switching on the testo 300 XXL, connect the temperature probe!
Example measurement
T1 T2
Measurement menu
Differential temperature
Measurement menu
Measurement menu
Function menu
Select “Delta T”
1 sec
DDeellttaa TT
TT 11
TT 22
4499..88
1155..44
3344..44
Measurement with: control unit + CO probe
1 - Initialisation - Zeroing of the CO probe (60 sec)
After zeroing, the CO probe is ready for measuring.
2 Open cap.
Note
The direction of flow of the gas to the probe influences the measuring accuracy! Optimum measurement results are achieved when the probe is moved gently backwards and forwards. Frontal flow onto the probe leads to excessive readings!
3 Close cap after the measurement.
28
CO measurement with the control unit
Example measurement
Connect probe before switching on the control unit!
Cap must be closed during zeroing (measuring error
otherwise possible).
Open the cap only during the measurement, then replace.
Warning message when adjusted CO value is exceeded
(change limit in Main menu “Input - Alarm limits Æ- CO
limit”).
10 sec
Initialisation
CO probe zeroing
60 sec
Measurement menu
11 0000
Explosive and flammable gases, particularly natural gas and propane, are detected by the gas leak probe even in low concentrations in the air. For example, suitable for finding leaks in gas pipes, vessels and equipment. 1 Connect the probe before switching on the
testo 300 XXL.
2 After the device has been switched on, the warming up
phase of the leak detection probe is automatically started (approx. 30 sec). Green, yellow and red LED, steady tone.
3 Ready for use
Green LED flashes steadily, steady tone stops.
4 Conduct leak detection:
• When gas escapes, acoustic signal (ticking) is audible, becoming faster with increasing concentration.
• When the 1st switching limit (> 200 ppm) is exceeded, yellow LED flashes.
• When the 2nd switching limit (> 1%) is exceeded, red LED flashes, steady tone. No readings of the gas leak probe are indicated in the display.
The power supply of the gas leak probe is provided by the control unit.
29
Service tip: The trigger limit is factory adjusted to 200 ppm CH4 gas.
The 1st alarm limit can be finely adjusted in a range from 0...1000 ppm (perform in fresh air only).
Example measurement
Gas leak probe
with control unit
Do not use the gas leak probe and testo 300 XXL in
enclosed rooms or facilities in which gases have
accumulated into an explosive mixture.
The use of other electrical equipment is also prohibited.
It must be ensured that the gas concentration does not exceed 20% LEL.
Duration of measurement with the gas leak probe with fully
charged battery, maximum 2 hours.
Remove the gas leak probe from the device immediately after the measurement to prevent heavy battery discharge.
Never leave the gas leak probe connected when charging the batteries.
red LED
yellow LED
green LED
Trimmer potentiometer for fine adjustment of the 1st alarm limit.
30
Example measurement
Humidity measurement with the control unit
Measurement with: combination humidity probe, e.g. Part no. 0636.9740
1 Connect combination humidity probe to the probe input
of the control unit.
2 Switch on the control unit.
3 Current readings of relative humidity, temperature and
dew point are displayed.
Changing units:
- Press key
- In Main menu - Device - Units – Change unit
A detailed description is contained in chapter Main menu, Device for control unit - Units.
10 sec
Initialisation
Zeroing
Measurement menu
Loading...
+ 90 hidden pages