Handling instructions
Pressure probe
0638.1445 (±10 hPa), 0638.1545 (±100 hPa)
0638.1645 (2 bar abs)
Please read before using instrument!
Observe measuring range (max. excessive pressure), too high pressure values can dest-
Please read the appropriate instruction manual before using the instrument!
In order to obtain reproducable measurements with differential pressure probes and to maintain the accuracy ±3Pa with the 10 hPa probe, the following should be observed:
- The probe must be adjusted to the ambient temperature before the start of the measurement!
- Zero calibration must be carried out before beginning a measurement, the position of the pressure probe should not be changed during the measurements that follow!
- The pressure probe can be positioned as required.
- Avoid changes in temperature during the measurement!
- Observe + and - inputs!
roy the sensor!
The pressure probe works with a differential pressure
Description
according to the principle of a strain gauge.
Connecting the probes
The pressure inlets in the differential pressure probe are marked with + and -. The higher pressure should be applied to the + inlet to achieve the best possible accurary level. Higher pressure at the + inlet produces a positive pressure indication, while higher pressure at the - inlet produces a negative pressure indication.
Application
Set the connected probe to its position of use before beginning a measurement. We recommend the use of a magnetic holding fixutre (see ordering data). Zero calibrati­on should be carred out before the measurement in order to guarantee reproducable measu­rements. The position of the probe should not be altered during the measurement that follows! Zero calibration is not necessary for measurements with the absolute pres­sure probe (0638.1645). The pressure probes are calibra­ted in the factory, calibration is carried out when the pres­sure probe is lying flat (normal position). We therefore recommend that the probe remains in this position during measurements. Make sure that the pressure probe is not subject to vibra­tions during the measurement (vibrations effect the results of the measurement).
Handling instructions
Technical data
Probe Meas.range Zero t.c. value Accuracy Max.
during long-term meas. * overload
0638.1345 ± 100 Pa ± (0.3 Pa + 0.5% of m.v.) 100 hPa
0638.1445 ±10 hPa type 0.3% of f.v./°C ± 3 Pa 50 hPa in range >10°C to 50°C
type 1% of f.v./°C
in range 0°C to 10°C
0638.1545 ± 100 hPa type 0.1% of f.v./°C ± 10 Pa (0 to 20 hPa) 200 hPa
± 0.5% of m.v.
(20 to 100 hPa)
0638.1645 2 bar type 0,05% of f.v./°C ± 5 hPa 3200 hPa
in range 0°C to 20°C
* unimportend if constant tempered and
reset before each measurement.
Operating temperature...............................0 to +50 °C
Storage and transport temp. ..................-20 to +70 °C
Care and maintenance
The pressure probe is maintenance-free and is not subject to a particular maintenance interval.
Clean the housing with a damp cloth.
Strong detergents or solvents should
never be used for cleaning the measuring
probe! Make sure that no liquid gets into the connec-
tion nipple!
0638.1445
0638.1645
Pressure sensor ±100 Pa ............................0638.1345
Pressure sensor ±10 hPa ............................0638.1445
Pressure sensor ±100 hPa ..........................0638.1545
Pressure probe 2bar abs.............................0638.1645
Magnetic holding fixture...............................0554.0225
testo
testo
Silicone hose ...............................................0554.0440
Pressure probe ............................................24 months
If the instrument is opened, improperly handled or if force is applied, no warranty will be granted!