• The instruction manual is an integral part of the instrument.
• Pay particular attention to the safety instructions and warning advice in
order to prevent injury and damage to the product.
• Keep this documentation to hand so that you can refer to it when
necessary.
• Hand this instruction manual on to any subsequent users of the product.
• Only operate the product properly, for its intended purpose, and within the
parameters specified in the technical data.
• Do not apply any force.
• Do not operate the instrument if there are signs of damage on the housing,
mains unit or connected cables.
• Dangers may also arise from objects to be measured or the measuring
environment. Always comply with the locally valid safety regulations when
carrying out measurements.
• Do not store the product together with solvents.
• Do not use any desiccants.
• Only perform maintenance and repair work on this instrument that is
described in this documentation. Follow the prescribed steps exactly when
doing the work.
• Use only original spare parts from Testo.
• Maintenance work that is not described in this documentation must only be
carried out by trained service engineers.
• Temperature information given on probes/sensors relates only to the
measuring range of the sensor technology. Do not expose handles and feed
lines to temperatures in excess of 70 °C (158 °F), unless they are expressly
authorized for use at higher temperatures.
• Do not carry out any contact measurements on uninsulated, live parts.
• Transport and store the instrument only in the associated packaging, in
order to prevent any damage to the sensor.
• Improper use of batteries may cause the batteries to be destroyed, or lead
to injury due to current surges, fire or escaping chemicals.
• Only use the batteries supplied in accordance with the instructions in the
instruction manual.
• Do not short-circuit the batteries.
• Do not take the batteries apart and do not modify them.
• Do not expose the batteries to heavy impacts, water, fire or temperatures in
excess of 60 °C.
• Do not store the batteries in the proximity of metal objects.
• Do not use any leaky or damaged batteries.
• Take the rechargeable battery out of the instrument immediately if it is not
functioning properly or if it shows signs of superheating. Rechargeable
battery may be hot!
• In the event of contact with battery acid: rinse affected areas thoroughly
with water, and if necessary consult a doctor.
• When not in use for a long period of time, take the rechargeable battery out
of the instrument to prevent exhaustive discharge.
Always pay attention to any information marked with the following warning
notices along with warning pictograms. Implement the specified precautionary
measures!
• Dispose of faulty rechargeable batteries and spent batteries in accordance
with the valid legal specifications.
• At the end of its useful life, deliver the product to the separate collection
point for electric and electronic devices (observe local regulations) or return
the product to Testo for disposal.
Please find the current country approvals in the attached
document.
2 Authorisations and certification
The testo 440 is used for measuring climate-related parameters. The testo 440
is particularly suitable for comfort level measurements for the workplace
evaluation and disturbance measurements in and at air conditioning systems.
The instrument is only to be used by qualified expert personnel.
Then product must not be used in potentially explosive atmospheres!
The testo 440 is supplied with inserted batteries. The batteries are
secured by a battery safety strip.
Remove the battery safety strip.
The testo 440 is ready for use.
4 Operation
For initial commissioning or after a factory reset, the
menu is opened automatically. In normal mode, the last menu used is
displayed after switching it on.
Adjustment time for humidity probe: at least 30 minutes.
Humidity Adjustment
1
Select the relevant reference point of 11.3 or 75.3% RH with .
2
Select the probe to be adjusted.
3
Adjust
Adjustment successful!
Humidity Adjustment
For humidity adjustment, the measurement parameter of the connected probe
is adjusted to the reference value at the two standard adjustment points 11.3%
RH and 75.3% RH; any deviations between the measuring value and the
nominal value are minimised across the entire measuring range.
The Testo calibration set provides a reference value to calculate the offset for a
humidity adjustment.
Humidity adjustment is possible with the following probes:
Order number Name
0636 9771 High-precision temperature-humidity probe with
0636 9775 Robust temperature-humidity probe for temperatures up
testo 440 is switched on and the corresponding probe is connected.
Probe is already exposed to reference conditions (e.g. salt pot) for an
adequate adjustment time.
See Section 3.3.8
Bluetooth
®
22
You are in the
Use to select
and confirm with .
menu.
An information window displays the remaining adjustment time and
the adjustment is carried out.
4.3.9 Resetting the instrument or probe to default
settings
You are in the Settings menu.
4
General
5
6
Use to select Reset Device or Reset Probe.
7
8
Confirm
9
4.4 Managing saved measurement data
Memory [Memory]
Press or to exit the menu.
After performance of a probe reset, the probe uses
set by the factory.
Use to select
Press or on the navigation key.
.
Press .
Use to select
.
Press or on the navigation key.
When the instrument is restarted, it will be reset to default settings.
In the
menu item, all measurements from the application
menu are saved with the time and date.
Measurements are saved in the last folder set up. If there is no folder available,
one will automatically be set up. Every day, a new folder is automatically set up
on the measuring instrument.
You can add other details about the project to the CSV file.
The window with available file folders opens.
Click on the required folder.
The window with available files opens.
Drag the file from the folder to the
If your file format is not displayed correctly, this is probably because
the language version of your operating system and your device are
different.
In this case, please open Excel and the relevant measurement data
document on the testo 440 from within Excel. A corresponding Excel
wizard enables modification of the file format.
is shown top left on the display, as soon as the probe and testo
is displayed.
440 are connected to each other.
A connection to compatible Bluetooth® probes is automatically
searched for and established. Only one probe can be
testo 440 by Bluetooth®.
Please observe the relevant instructions for the probe when
commissioning it.
The LED on the handle flashes yellow. The LED flashes green while
the connection is being established.
Press the button on the probe handle a minimum of 3 seconds to
switch the probe off.
LED status Description
Flashing yellow Probe is switched on and searching
®
for the Bluetooth
connection.
Flashing green The probe is switched on and
connected via Bluetooth
®
to the testo
In the
, both current and minimum and maximum
measuring values can be read and saved. All compatible probes can be used.
For a list of all compatible probes see Section 2.6.
The
of following probes can be connected at the same time:
Depending on the connected probe, parameters can be configured for the
measurement, for example, visibility of individual values or units.
Press to open the
.
Individual values being hidden does not affect the application, only the Basic
View and long-term measurement. The configured units, on the other hand, are
also taken up in the application menus.
For the instrument model testo 440 dP, zeroing can also be
undertaken in the Basic View.
The accuracy specification applies immediately after zeroing of the
sensor. The sensor is affected by changes in position or attachment to
magnetic surfaces. Therefore, zero the sensor only in the instrument
end position.
The testo 440 has permanently installed application menus. These enable the
user to carry out convenient configuration and implementation of specific
measuring tasks.
Available application menus are enabled as soon as a probe is
connected. Unavailable application menus are greyed out on the
display. More than one probe has to be connected to make some
application menus accessible.
The units for the measuring values depend on the ISO/US setting and
configuration in the
Use this application to measure the volume flow at the outlet or in a ventilation
system duct. There are different options for this. These mainly differ in the
measuring range and require corresponding probes:
• Thermal probes (including temperature measurement and possibly humidity
measurement) for low flow velocities
• 16 mm wheel measuring head (including temperature measurement) for
average flow velocities
• Pitot tube for measurements at high velocities and in heavily contaminated
flows with a high particle content
This application menu is accessed by one of the following probes:
.
[Volume Flow]
Order number Name
0635 1032 Hot wire probe including temperature sensor, fixed
0635 1571 Hot wire probe with Bluetooth®, including temperature
0635 1572 Hot wire probe with Bluetooth®, including temperature
and humidity sensor
0635 1026 Hot wire probe (Ø 7.5 mm), including temperature
0635 1051 Hot wire probe (Ø 3 mm), including temperature
0635 9571 Vane probe (Ø 16 mm) with Bluetooth®, including
Use to select the required parameter and make settings.
3
Carrying out the measurement
A suitable probe is connected to the testo 440.
1
Remove the protective cap from the probe head.
2
Move the probe into the flow.
3
4
5
Carry out the measurement and save the measuring values.
4.5.6 Funnel volume flow application
- Volume Flow Unit: m3/h, cfm, l/s, m3/s
- Correction Factor: 1% to 200%
In a multi-point measurement, a mean value is calculated from the
individual measuring values.
In a timed average measurement, a mean value is calculated over a
certain period.
Several points can be measured. This generates a timed
average/multi-point total mean value.
Press or to exit the menu.
For measurements in flows with a known direction, the arrow mark on
the probe head must point in the direction of flow.
Align the probe with the assumed flow axis.
Read off the readings.
Low flow velocities may result in greater measurement uncertainty
during temperature and humidity measurement.
[Funnel
Volume Flow]
A volume flow rate funnel is required to determine the flow rate on ventilation
systems. The measurement can be carried out with a compatible vane probe in
connected. If two flow probes are connected, a cable probe is used for
conjunction with a funnel set. Alternatively, a thermal anemometer can also be
used in conjunction with a funnel.
The funnels differ in size. When selecting the funnel, make sure that the
opening of the funnel covers the grille completely and tightly.
The application menu is enabled by one of the following probes:
Order number Name
0635 1032 Hot wire probe including temperature sensor, fixed
cable
0635 1571 Hot wire probe with Bluetooth®, including temperature
0635 1572 Hot wire probe including temperature and humidity
0635 9571 Vane probe (Ø 16 mm) with Bluetooth®, including
0635 9572 Vane probe (Ø 16 mm) including temperature sensor,
0635 9371 High-precision vane probe (Ø 100 mm) with
®
Bluetooth
, including temperature sensor
0635 9372 High-precision vane probe (Ø 100 mm) including
0635 9431 Vane probe (Ø 100 mm) with Bluetooth®, including
0635 9432 Vane probe (Ø 100 mm) including temperature sensor,
0560 1405 testo 405 i - thermal anemometer operated by
A maximum of 1 Bluetooth® probe and one cable probe can be
For Pitot tubes from other manufacturers, refer to the instruction
In a multi-point measurement, a mean value is calculated from the
individual measuring values.
In a timed average measurement, a mean value is calculated over a
certain period.
Several points can be measured. This generates a timed
average/multi-point total mean value.
Press or to exit the menu.
Observe the minimum clearances to points of discontinuity:
- From points of discontinuity upstream of the flow, observe a
clearance of at least six times the hydraulic diameter Dh = 4A/U (A:
channel cross-section, U: channel circumference).
- From points of discontinuity downstream of the flow, observe a
clearance of at least twice the hydraulic diameter Dh = 4A/U (A:
channel cross-section, U: channel circumference).
Volume Flow]
The testo 440 can determine the volume flow by measuring the reference
resistance and input of the K-factor. This allows the testo 440 to remain
connected to the air outlet during adjustment work and the changes in volume
flow can be read directly on the display.
The application menu can only be accessed with testo 440 dP or used
in combination with a testo 510i with Bluetooth
®
.
This process for determining volume flow can always be used when there are
appropriate specifications available from the component manufacturer. In line
with these specifications, the differential pressure is measured at a position
specified by the manufacturer or supplier. The volume flow is determined from
the differential pressure via a component-specific K-factor using the following
mathematical equation:
Use to select the required parameter and make settings.
3
The maximum measurement period depends on the condition of the
testo recommends using an external power supply for each micro USB
Press to configure the measurement.
You can make the following settings:
- Measuring interval: in s
- Measurement duration: in hrs and min
Press or to exit the menu.
battery, how much free memory there is and the probe used. This is
displayed during configuration.
In the long-term mode application [Logger Mode], the measurement
(from firmware version 1.0.4) is always saved automatically at the end
of the configured measurement period.
in the case of particularly long measurements. It is then possible to
record significantly longer measurement series.
0554 1105 - USB mains unit including cable
* The accuracy specification applies immediately after zeroing of the sensor. The sensor is affected by
changes in position or attachment to magnetic surfaces. Therefore, zero the sensor only in the
instrument end position.
Probe connections testo 440 testo 440 dP
Type K thermocouple
±0.4 °C (-40 to -25.1 °C)
±0.3 °C (-25 to +74.9 °C)
±0.4 °C (+75 to +99.9 °C)
±0.5 °C for cold junction
±0.05 hPa (0 to +1.00
hPa)
±0.2 hPa + 1.5% of m.v.
1x
Testo Universal
1x
Connector (TUC) for
connection of cabled
probes with