Specifications:
Capacity 14 Cu. Ft. Capacity (12 Ref/1.5 Freezer)
Temperature Range: Refrigerator Preset 41°F/5°C
Temperature Range: Freezer Preset -4°F/-20°C
Exterior (w x d x h) 25¾” x 22¾” x 84”
Interior (w x d x h) see drawings (pg. 3)
Temperature Control
Refrigerator: Microprocessor Temperature Controller.
Audible Visual Temperature Alarms With Dry Contacts
Freezer: Data Logger Acts As Digital Temperature Display.
Exterior Finish
White powder coated exterior
Legs
Leveling Legs
Door
Two Exterior Doors (Glass Ref/Solid Freezer) Keyed Door
Locks. Magnetic door gasket for positive seal. Heavy Duty
Door Handles
Shelves
Three Adjustable Refrigerator/Two Fixed Freezer
Insulation
High density urethane foam cabinet and door insulation
Ports
Two Probe Access Ports (3/8") - One for Refrigerator/One
for Freezer
Refrigerant
R134a
Compressor
1/6 HP Freezer, 1/5 HP Refrigerator (Requires 2 Outlets)
Defrost
Manual Defrost Freezer / Auto Defrost Refrigerator
Voltage: Domestic
115V/60Hz
Amps
1.1 Amp Freezer, 4.1 Amps Refrigerator
Shipping Weight
319 lbs.
Warranty
(Domestic)
Two (2) year parts and labor warranty
Five (5) year compressor warranty
Comes with pharmacy refrigerator/freezer toolkit, temperature logs and CDC
recommended labels. Requires two power outlets.
Two data loggers come with two year NIST traceable certificates calibrated to ISO
17025 standards. Certified by an ILAC accredited laboratory. *Unit designed to follow
CDC guidelines.
Specifications subject to change without notice
1-866-923-PURE • Fax 856-361-2772 • Contact: sales@tempurescientic.com • Website: www.tempurescientic.com
Optimal Operating Conditions
Optimal operation of this unit requires it to be located in a climate controlled facility that maintains temperatures at or below 75°F with 55% relative humidity