Additional User Information
IMPORTANT: Read instructions carefully before use.
Type BF Applied Part
User should contact Blue Spark Technologies, Inc. for assistance in using TempTraq when
needed, or to report unexpected operation or events.
WARNING: No modification of this equipment is allowed.
WARNING: Do not modify this equipment without authorization of the
manufacturer.
WARNING: If this equipment is modified, appropriate inspection and testing must
be conducted to ensure continued safe use of equipment.
Warranty
TempTraq warrants each new TempTraq patch against defects in materials or workmanship
until the expiration date of the product, and agrees to replace any defective product without
charge.
IMPORTANT:
This warranty does not cover damage resulting from accident, misuse or abuse, or lack of
reasonable care. No responsibility is assumed for any special, incidental or consequential
damages. To obtain warranty replacement, simply call TempTraq Customer Care at 1 (440)
249-5440 for return instructions, or email info@temptraq.com.
NOTE:
No other warranty, written or verbal is authorized by TempTraq. This warranty gives you
specific legal rights and you may also have other rights which vary from state to state. Some
states do not allow the exclusion or limitation of incidental or consequential damages, so the
above exclusion and limitations may not apply to you.
TempTraq Connect Troubleshooting Guide
(Applies to Models TT-100 and TT-200 only)
FCC Statement and Legal Notices
This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions (1) this device may not cause
harmful interference, and (2) this device must accept any interference received including interference that may cause undesired
operation.
FCC WARNING: any changes or modifications not expressly approved by the party responsible for compliance could void the user’s
authority to operate this equipment under FCC regulations.
NOTE: this equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the
FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be
determined by turning the equipment o and on, the user is encouraged to try to correct the interference by one or more of the
following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit dierent from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
P/N: TT-100-201 Rev 2
© 2017 Blue Spark Technologies, Inc.
TempTraq® and other related marks are owned by Blue Spark Technologies, Inc. All other trademarks are property of their
respective owners.
Protected by U.S. Patent Nos. 8,029,927; 8,268,475; 8,722,235. Other foreign patents pending.
Specifications
Model Numbers (Patch
Operating Life)
TT-100 (24 hrs.), TT-200 (48 hrs.), TT-300 (72 hrs.)
Patch Dimensions Length: 3.90 in (99.2mm), Height: 1.94 in (49.2mm), Thickness: 0.10 in
(2.5mm)
Patch Weight 0.16 ounces (4.5 grams)
Patch Environmental Conditions
• Transport and Storage
• Use
-4 to 122 °F (-20 to 50 °C) / 15 - 95% RH (non-condensing)
61 to 104 °F (16 to 40 °C) / 15 - 95% RH (non-condensing)
Temperature Measurement
Location
Axillary (under the arm) - TempTraq measures under-the-arm "axillary"
temperature and this is converted to an oral temperature which is
displayed in the TempTraq app.
Temperature Range 87.0 °F - 109.3 °F (30.6 °C - 42.9 °C). Temperatures below this range will
be displayed in the app as “Lo”. Temperatures above this range will be
displayed in the app as “Hi”.
Temperature Indications Less than 100.4 °F (38.0 °C) – GREEN
Equal to or greater than 100.4 °F (38.0 °C) – ORANGE
Equal to or greater than the user defined alert temperature – RED
Accuracy Conforms with ASTM E1112-00 (2011); +/- 0.1 °C or +/- 0.2 °F over the
temperature range of the device
Display Resolution 0.1 °F (0.1 °C)
Response Time Once applied to the underarm, the patch requires up to 8 minutes to
reach a stable reading
Transmission Distance Patch will transmit up to 40 feet (12 meters) when applied to the
underarm
Compatible Smart Devices Smartphones and tablets with Bluetooth® 4.0 running Apple® operating
system iOS 8.0 or later or Android™ operating system 4.3 or later. Visit
temptraq.com for a list of compatible devices.
Standards ASTM E1112-00 (2011) standard for electronic thermometers
IEC 60601-1 Medical Design Standards, 3rd Edition
IEC 60601-1-2 Electromagnetic Compatibility
Bluetooth® Version 4.0
Radio Regulations FCC Part 15
Industry Canada License exempt RSS standard(s)
ISO/IEC 60601-1-2 Electromagnetic Compatibility
CISPR 11 Group 1, Class B Radiated Emissions within 30 – 1,000 MHz
Water Ingress Protection IP24 - Protected from damage due to water splashing on the product.
This thermometer conforms to all requirements established in ASTM standard E1112-00. Full
responsibility for conformance of this product to the specification is assumed by Blue Spark
Technologies, Inc., 806 Sharon Drive, Suite G, Westlake, Ohio 44145
Problem Troubleshooting Hints
Unable to access Connect
within the app
• User must be logged into Connect to access features of
Connect. Select “Log In” in the Hamburger menu to log into
Connect.
Unable to view a shared
patch
• Check to ensure that the latest version of the app is installed
on the device
• Check to ensure that you are logged into Connect on your
phone
Unable to change the
“Temperature Alert” setting
or “Name” when viewing a
shared patch
• Only the user monitoring the patch locally and sharing the
patch can change the “Alert Temperature” or “Name”
Not receiving alerts when
viewing a shared patch
• Only the user monitoring the patch locally and sharing the
patch will receive alerts on their phone
A shared patch is no longer
visible in the app
• Check with the person who shared the patch to ensure that
they are still sharing the patch with you
• Check to ensure that you are still logged into Connect
No longer receiving
temperature data updates
for a shared patch. App
displays “No temperature
data received in last 60
minutes….”
• The person that shared the patch may no longer be
receiving updates from the patch. This could occur if they
are too far from the patch or have closed the TempTraq app.
• Check to ensure that your phone is connected to the internet
via cellular or WiFi
TempTraq Connect
(Applies to Models TT-100 and TT-200 only)
TempTraq Connect enables you to invite anyone anywhere in the world to view temperature
data from a patch that you are monitoring on your smart device.
• Sign into TempTraq Connect within the TempTraq app.
• Select the patch that you wish to share.
• From the Patch Detail screen select the TempTraq Connect cloud icon.
• Enter the e-mail address for the user that you wish to share the patch with.
• The TempTraq Connect screen for that patch will show who you have shared the patch
with. You can share the patch with multiple Connect users, and you can add and delete
users from this list.
When someone invites you to view a patch using TempTraq Connect you may view the data
available for that patch.
• Sign into Connect within the TempTraq app to view a patch that has been shared with
you.
• Patch Name, Current Temperature, Temperature History, Temperature Alert Setting, and
Notes will be available to view, but cannot be changed.
• When viewing a shared patch, no alerts are generated on your device for that patch.
Alerts only occur on the device that is within 40 ft. of the patch and that is receiving data
from the patch via Bluetooth 4.0.
• You may delete the patch from the Patch List screen.
• If you log out of Connect, shared patches will no longer be visible on your phone.
Performance of this device may be degraded should one or more of the following occur:
• Operation outside of the manufacturer’s stated temperature and humidity range
• Storage outside of the manufacturer’s stated temperature and humidity range
• Mechanical shock
The device or its components are not intended to be serviced, replaced, or repaired.
Device is powered by a 3V non-rechargeable battery. Not intended to be recharged,
serviced, replaced, or repaired.
The device is single use only. Do not reuse.
May be safely applied to self or used on others.
May be discarded in common household waste.
Need More?
VISIT TEMPTRAQ.COM
CALL 1 (440) 249-5440
EMAIL info@temptraq.com
Blue Spark Technologies, Inc.
806 Sharon Drive, Suite G
Westlake, OH 44145