Teledyne LeCroy T3PM1006 User manual

T3PM1006
Digital Power Meter
User Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
The information in this manual was correct at the time of printing. However, Teledyne LeCroy continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
TABLE OF CONTENTS
Table of Contents
SAFETY INSTRUCTIONS .................................................. 5
GETTING STARTED ........................................................ 10
Characteristics ............................................................ 11
Appearance ................................................................. 14
Set Up ........................................................................ 23
BASIC SETTING ............................................................. 26
Setting up measurement range ................................... 27
Setting up measurement status .................................. 31
Setting up System status ............................................ 41
MEASUREMENT AND OTHER FUNCTIONS .................... 46
Measurement function ................................................ 47
Other functions .......................................................... 50
Integration measurement function.............................. 52
REMOTE CONTROL ....................................................... 59
Configure Remote Control Interface ............................ 60
Return to Local Control ............................................... 63
COMMAND OVERVIEW ................................................. 64
Command Syntax ........................................................ 64
Command List ............................................................ 68
APPENDIX .................................................................... 102
Specifications ........................................................... 103
Dimensions .............................................................. 108
Power measurement ................................................. 109
Introduction to IEC-62301 ......................................... 111
EUP Directive Lot6 specifications ............................. 112
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Connection Guide ..................................................... 113
CERTIFICATIONS .......................................................... 115
EMC Compliance ...................................................... 115
Safety Compliance .................................................... 117
Environmental Compliance ....................................... 118
RESTRICTION OF HAZARDOUS SUBSTANCES
(RoHS) ..................................................................... 118
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the T3PM1006 or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
1 O
On (Supply) Off (Supply)
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the instrument.
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
General Guideline
CAUTION
Make sure that the voltage input level does not
exceed DC848V/AC600V.
Make sure the current input level does not
exceed 20A.
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that can
lead to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Do not block or obstruct the cooling fan vent
opening.
Do not perform measurement at the source of a
low-voltage installation or at building installations (Note below).
Do not disassemble the instrument unless you
are qualified as service personnel.
Make sure that the COM terminal to earth is
limited to 300Vpk.
Remove all test leads before disconnecting the
mains power cord from the socket.
If the equipment is used in a manner not
specified by the manufacturer, the protection provided by the equipment may be impaired.
The device should be placed in a way that the
plug connected to it can be removed easily.
Safety Guidelines
Find Quality Products Online at: sales@GlobalTestSupply.com
6
www.GlobalTestSupply.com
SAFETY INSTRUCTIONS
(Note) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The T3PM1006 falls under category II 300V.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Power Supply
WARNING
AC Input voltage: 100-240 VAC 50/60Hz The power supply voltage should not fluctuate
more than 10%.
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid electrical shock.
Cleaning the Instrument
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Temperature: 0°C to 40°C Humidity: < 30°C: < 80%RH(non-condensing);
30°C~40°C:<70%RH(non-condensing); >40°C: <50%RH (non-condensing)
Altitude: <2000m
(Note) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The T3PM1006 falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -40°C to 70°C Humidity: <90%RH(non-condensing)
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the unit in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Characteristics ............................................................ 11
Accessories .............................................................................13
Package Contents ...................................................................13
Appearance ................................................................. 14
Front Panel ..............................................................................14
Display Overview ....................................................................18
Rear Panel ...............................................................................21
Set Up ......................................................................... 23
Tilting the Stand......................................................................23
Power Up .................................................................................24
Connect the wires to the T3PM1006 ......................................25
GETTING STARTED
This chapter describes the T3PM1006, including accessories, package contents, its main features and front / rear panel introduction.
Find Quality Products Online at: sales@GlobalTestSupply.com
10
www.GlobalTestSupply.com
GETTING STARTED
Operation
Press the buttons on the front panel to easily
turn on the T3PM1006 measurement function. All settings and measurements results are displayed on the TFT-LCD screen panel for easy use of each function.
Standard display mode: 2 main measurement
results and 6 secondary measurement results are displayed in this screen.
Simple display mode: 4 major measurement
results are displayed in this screen.
Performance
6 selectable voltage ranges available from 15V to
300V with 0.1% of reading + 0.1% of range.
12 selectable current ranges available from 5mA
to 20A with 0.1% of reading + 0.1% of range.
It can even measure the voltage of abnormal
wave of CF 3. The half-range CF is up to 6.
Characteristics
The T3PM1006 is a high-precision, programmable power meter for measuring low power devices such as switching power supplies, transformers, power supplies, and adapters. It is equipped with a color TFT-LCD screen which is very convenient for reading the measurement results. The T3PM1006 has become a reliable power measurement instruments because of its simple operation, excellent performance and automatic measurement interface.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
It can even measure the current of abnormal
wave of CF 3. The half-range CF is up to 6.
Test terminals in the front panel. Total harmonic distortion measurement.
Features
Full five-digit measurement. Voltage measurement range: 15V ~ 600V or
automatic switching
Current measurement range: 5mA ~ 20A or
automatic switching
Maximum accuracy of 0.1% of reading + 0.1% of
range
2 main measurement readings and 6 minor
measurement readings are displayed in the screen of standard display mode.
4 main measurement readings are displayed in
the screen of simple display mode.
Added stand-alone display of total harmonic
distortion measurement function (13 steps)
Test bandwidth of voltage and current: DC ~
6kHz.
Added W-h power time integrator function Selectable boot settings (Previous / Default)
Interface
Standard interface: USB / RS232 / LAN
Application
It can be applied to production test such as
power supplies, transformers, motors, electrical equipment and other equipment with low standby power.
It can be applied to power measurement
conforms to IEC 62301
It can be applied to assess the power
consumption of product design.
Find Quality Products Online at: sales@GlobalTestSupply.com
12
www.GlobalTestSupply.com
GETTING STARTED
Standard Accessories
Power Cords x 3
Test leads: 2x red, 2x black
Opening the box
Contents (single unit)
Main unit Test leads (red x2,
black x2)
Power cords x 3 Quick Start Guide
Accessories
Package Contents
Check the contents before using the instrument.
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
V -
SetupMode
Enter
StopStart Reset
MAX Hold
Key Lock
ESCLocal
POWER
INTEGRATOR
Range I - Range
CURRENT VOLTAGE
I
AC 10A MAX
I
V
V
AC 600V
MAX
CAT II 300V
Hold
Current input Voltage input
Power
key
ESC
key
Enter
key
Arrow
keys
Function
keys
LCD display
Power Switch
POWER
Turns On or Off the main power. For the power up sequence, see page 24.
Current, Voltage Terminals
CURRENT VOLTAGE
I
AC 10A
MAX
I
V
V
AC 600V
MAX
CAT II 300V
Current input: I+ and I– terminals; Voltage input: V+ and V­terminals.
Appearance
Front Panel
14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
GETTING STARTED
Note
If the measurement power supply has positive and negative electrode, please connect + to the positive electrode of power supply and - to the negative electrode of power supply.
Warning
The maximum measurable current and voltage are 600 V and 10A for voltage and current terminals of the front panel of the T3PM1006. Do not input exceeded voltage and current, otherwise it will burn the device.
Function keys
V - Range
Enter
V-Range key, up/down arrow keys and Enter key can be used together to select a voltage range or auto range measurement mode. See page
27.
MAX HoldRange I -Range
Enter
I-Range key, up/down arrow keys and Enter key can be used together to select a current range or auto range measurement mode. See page
27.
15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
MAX Hold
Press this button to display the maximum measurement reading. See page 50.
Mode
Press this key to select measure mode (DC/AC/AC+DC). See page
51.
Setup
Press this key to enter the measurement settings menu. See page 31.
Hold
Press this key to switch window and stop refreshing. See page 51.
Enter
StopStart Reset
INTEGRATOR
Use the left and right arrow keys to select Integrator mode, and press Enter button to enter the time integrator function. See page
52.
Key Lock
Local
Press this key to toggle to key lock. In Remote control mode, press this button to switch to local mode. See page 51.
Confirm button
Enter
This button is used to enter the menu, confirm the settings and switch between the standard display mode and simple display mode (no function table and display icon). See page 51.
16
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
GETTING STARTED
Cancel (Exit) button
ESC
Press this button to cancel the current setting. The cursor returns to the default position or return to the previous menu according to the situation. See page 51.
Arrow Keys
These four arrow keys are used to edit the parameters, browse the menu system and select the parameter range.
17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
FilterPTCT
Max. Hold
KeyLock
Average Range
Display Hold
Peak Voltage Peak Current
Remote Error
Remote CF
ModeVoltage Range
Current Range
Secondary Menus
Main
Measurement
Display
Minor
Measurement
Display
Item
Status icon
Description
Voltage Range
V_Range 300V
Voltage measurement range. Example here range is 300V.
V_Auto means that Auto Range is turned on.
Current Range
I_Auto 50mA
Current measurement range. Example here range is 50mA.
I_Auto means that Auto Range is turned on.
Mode
AC+DC
Measurement mode (AC, DC, AC+DC)
Remote
RMT
Remote control mode (on/off)
Crest Factor
CF3
Crest Factor (3/6)
Filter
Filter
Voltage and current filters (on/off)
PT Ratio State
PT
External voltage magnification (on/off)
Display Overview
Find Quality Products Online at: sales@GlobalTestSupply.com
18
www.GlobalTestSupply.com
GETTING STARTED
CT Ratio State
CT
External current magnification (on/off)
Maximum Hold
Max. Hold
Retain and display the maximum measurement reading.
Keyboard Lock
KeyLock
Lock Key button
Average
Avg-1
Average number of sampling (1/2/4/8/16/32/64)
Display Hold
Hold
Retain and display the current measurement reading.
Peak Voltage
P.V
The voltage exceeds the measurement range
Peak Current
P.I
The current exceeds the measurement range
Remote Error
Err-XXX
An error occurs in remote command
Standard Display Mode
Display the measurement result of 2 major and 6 minor measurement parameters
Simple Display Mode
Display the measurement result of 4 major measurement parameters
Secondary menus
Display secondary function menu
Enlarge
This function key is used to switch display of measurement result from 2 major plus 6 minor to 4 major ones.
Integrator
This function key is used to set up integrator measurement parameters and execute integrator measurement function.
Parameter
This function key is used to set up measurement parameters.
19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
System
This function key is used to enter the system setting and system configuration screens.
20
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
GETTING STARTED
I I
20A
MAX
100 240V
25VA MAX
AC 50 / 60Hz
V V
CAT II 300V
RS232
LAN
GPIB
SER.NO. LABEL
REMOVE INPUTS BEFORE OPENING.
TO AVOID SHOCK,
WARNING
600V
MAX
CURRENT VOLTAGE
AC power input
USB B
port
RS232
port
LAN port
Voltage
input
Current
input
Power Cord Socket
100 240V
25VA MAX
AC 50 / 60Hz
Accepts the power cord. AC 100~240V ±10%, 50/60Hz
RS232
RS232
RS232 port. This port is used for remote control.
USB Device Port
Type B USB port. This port is used for remote control.
LAN Port
LAN
LAN Port.
Rear Panel
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
21
T3PM1006 User Manual
Rear Voltage/Current input terminal
I I
20A
MAX
V V
CAT II 300V
600V
MAX
CURRENT VOLTAGE
Rear Voltage/Current input terminal is used to connect the main measurement signals.
Warning
Do not use damaged device. Before using the
equipment, check its housing first to ensure there are no cracks. Do not operate this device in an environment containing explosive gases, steam or dust.
The maximum measurable current and voltage are
600 V and 20A for voltage and current terminals of the rear panel of the T3PM1006. Do not input exceeded voltage and current, otherwise it will burn the device.
Always use the supplied cable for connection. Before connecting the device, observe all the safety
symbols marked on the device.
Turn off the power to the device and the application
system before connecting I/O terminals.
Do not install replacement parts on the device or
perform any unauthorized modifications.
Do not use this device if the removable cover is
removed or loosened.
Do not connect any cables and terminals before
performing self-test.
Use only the power adapter supplied by the
manufacturer to avoid accidental injury.
Do not use this device for life support systems or
any other equipment that has safety requirements.
22
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
GETTING STARTED
From the base of the handle, gently pull the handle out sideways and then rotate it to one of the following positions.
Horizontal position
Tilt stand position
Carry position
Set Up
Tilting the Stand
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
23
T3PM1006 User Manual
Steps
1. Ensure the AC voltage is 100~ 240V.
2. Connect the power
cord to the AC voltage input.
Note
Make sure the ground connector on the power cord is connected to a safety ground. This will influence the measurement accuracy.
3. Push to turn on the main
power switch on the front panel.
4. The display turns on and shows the last
function that was used before the power was reset.
Power Up
24
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
GETTING STARTED
Background
Two separate wires are used to connect the T3PM1006, so voltage and current measurement are isolated and dont interfere with each other.
Connection diagram
CURRENT VOLTAGE
I
AC 10A MAX
I
V
V
AC 600V
MAX
CAT II 300V
I I
20A
MAX
V V
CAT II 300V
600V
MAX
CURRENT VOLTAGE
Note
The terminals on the front and rear panels cant be used as input terminal at the same time.
Description
V +
The positive voltage input (+)
V -
The negative voltage input (-)
I +
The positive current input (+), 10A for input on the front panel, 20A for input on the rear panel.
I -
The negative current input (-), 10A for input on the front panel, 20A for input on the rear panel.
GND
Provide reference grounding.
Connect the wires to the T3PM1006
25
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Setting up measurement range ................................... 27
Auto Range ..............................................................................29
Setting up measurement status .................................. 31
Setting up synchronization source .........................................31
Setting up filter .......................................................................32
Setting up crest factor ............................................................33
Setting up auto-zero function .................................................34
Setting up average value .........................................................35
Setting up method of calculating harmonics .........................36
Setting up the PT ratio status .................................................37
Setting up the CT ratio status.................................................38
Setting up the voltage and current skipping configuration ...39
Setting up System status ............................................ 41
System configuration setting screen ......................................41
Setting up power on status ....................................................41
Setting up brightness .............................................................42
Setting up key sound ..............................................................43
Setting up interface .................................................................44
BASIC SETTING
Find Quality Products Online at: sales@GlobalTestSupply.com
26
www.GlobalTestSupply.com
BASIC SETTING
Set voltage range
1. Press V-Range button.
V - Range
2. Use up and down arrow keys to
select the desired range.
3. Press Enter button to confirm your
selection.
Enter
Available range
Crest Factor is 3:
AUTO, 15V, 30V, 60V, 150V, 300V, 600V
Crest Factor is 6:
AUTO, 7.5V, 15V, 30V, 75V, 150V, 300V
Set current range
1. Press I-Range button.
MAX HoldRange I -Range
Setting up measurement range
To get accurate measurement results, set an appropriate measurement range before performing measurements.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
27
T3PM1006 User Manual
2. Use up and down arrow keys to
select the desired range.
3. Press Enter button to confirm your
selection.
Enter
Available range
Crest Factor is 3:
AUTO, 5mA, 10mA, 20mA, 50mA, 100mA, 200mA, 0.5A, 1A, 2A, 5A,10A, 20A
Crest Factor is 6:
AUTO, 2.5mA, 5mA, 10mA, 25mA, 50mA, 100mA, 250mA, 0.5A, 1A, 2.5A, 5A, 10A
Note
When the measurement range is set manually, if the range status icon lights in green, the measured value meets the setting range. On the contrary, if the range status icon lights in red, the measured value doesnt meet the best setting range. In this case, it is better to switch to another range to get more accurate measurement results.
Note
The P.I status icon lights in red when the current measurement circuit detects that the measured value exceeds setting range by 3 folds (CF is set to 3) or 6 folds (CF is set to 6).
Find Quality Products Online at: sales@GlobalTestSupply.com
28
www.GlobalTestSupply.com
BASIC SETTING
Note
The P. V status icon lights in red when the voltage measurement circuit detects that the measured value exceeds setting range by 3 folds (CF is set to 3) or 6 folds (CF is set to 6).
Range is shift up
The range is shifted up when either of the following conditions is met.
Vrms or Irms exceeds the current setting range
by 110%.
The Vpk or Ipk value of the input signal exceeds
the current setting range by 330% at CF 3.
The Vpk or Ipk value of the input signal exceeds
the current setting range by 660% at CF 6.
Range is shift down
The range is shifted down when all of the following conditions are met.
Vrms or Irms is equal to or less than the 60% of
previous range.
The Vpk or Ipk value of the input signal is less
than the next setting range by 300% at CF 3.
The Vpk or Ipk value of the input signal is less
than the next setting range by 600% at CF 6.
Auto Range
The range is automatically switched according to the voltage and current of input signal.
29
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Example
Irms exceeds the current setting range by 110%, so range is shifted to 20mA
Irms is less than or equal to 60% of the previous setting range, so range is shifted down to 10mA.
Find Quality Products Online at: sales@GlobalTestSupply.com
30
www.GlobalTestSupply.com
BASIC SETTING
Steps
1. Press Setup button.
Setup
2. Press Enter button.
Enter
3. Press down arrow key.
4. Press Enter button to enter Sync
Source item. Use up and down arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
V:
Select the voltage of signals as synchronization source.
I:
Select the current of signals as synchronization source.
OFF:
Select the entire interval of data updating period as synchronization source.
Default value
V
Setting up measurement status
Setting up synchronization source
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
31
T3PM1006 User Manual
Steps
1. Press Setup button.
Setup
2. Press Enter button.
Enter
3. Press down arrow key twice.
x2
4. Press Enter button to enter Filter
item. Use up and down arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
On:
Turn on the line filter function and Filter status icon on the display lights up in green.
Off:
Turn off the line filter function. Line filter cutoff frequency is 500Hz
Default value
Off
Setting up filter
32
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BASIC SETTING
Steps
1. Press Setup button.
Setup
2. Press Enter button.
Enter
3. Press down arrow key three times.
x3
4. Press Enter button to enter Crest
Factor item. Use up and down
arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
3:
Crest Factor is three.
6:
Crest Factor is six.
Default value
3
Setting up crest factor
33
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Steps
1. Press Setup button.
Setup
2. Press Enter button.
Enter
3. Press down arrow key four times.
x4
4. Press Enter button to enter Auto
Zero item. Use up and down arrow
keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
On:
Auto-zero function is activated once per hour or when range is switched
Off:
Auto-zero function is only activated once when the range is switched. The auto-zero function is turned off when the integrator function is executed
Default value
Off
Setting up auto-zero function
34
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BASIC SETTING
Steps
1. Press Setup button.
Setup
2. Press Enter button.
Enter
3. Press down arrow key five times.
x5
4. Press Enter button to enter Average
item. Use up and down arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
1, 2, 4, 6, 8, 16, 32 and 64: The measurement time is synchronized with the average value that you set. The larger the average value is, the longer the measurement time is. When the average value is set to 1, the measurement time is about 0.1 seconds. The larger the number is, the longer the measurement time is, and so forth.
Default value
2
Setting up average value
35
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Steps
1. Press Setup button.
Setup
2. Press Enter button.
Enter
3. Press down arrow key six times.
x6
4. Press Enter button to enter
Harmonics item. Use up and down
arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
IEC:
Calculate the ratio of harmonic quantity of the 2nd through the 13th harmonic to the 1st harmonic.
CSA:
Calculate the ratio of harmonic quantity of the 2nd through the 13th harmonic to the 1st through the 13th harmonic.
Off:
Turn off the harmonic calculation function.
Default value
Off
Setting up method of calculating harmonics
36
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BASIC SETTING
Steps
1. Press Setup button.
Setup
2. Press Enter button.
Enter
3. Press down arrow key seven times.
x7
4. Press Enter button to enter PT
Ratio Status item. Use up and
down arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
On:
Turn on the PT ratio calculation function and PT status icon on the display lights up in green. The setting range is from "1" to "9999.999".
Off:
Turn off the PT ratio calculation function.
Default option
Off
Setting up the PT ratio status
37
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Steps
1. Press Setup button.
Setup
2. Press Enter button.
Enter
3. Press down arrow key eight times.
x8
4. Press Enter button to enter CT
Ratio Status item. Use up and
down arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
On:
Turn on the CT ratio calculation function and CT status icon on the display lights up in green. The setting range is from "1" to "9999.999".
Off:
Turn off the CT ratio calculation function.
Default option
Off
Setting up the CT ratio status
38
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Steps
1. Press Setup button.
Setup
2. Press right arrow key to enter
SETUP1 tab.
3. Press Enter button.
Enter
4. Press down arrow key to enter the
V/I-Range Skipping Config.
5. Press Enter button.
Enter
6. Press up and down arrow keys to
turn On the V/I-Range Skipping Config.
BASIC SETTING
Setting up the voltage and current skipping configuration
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
39
T3PM1006 User Manual
7. Press Enter button to confirm
setting.
Enter
8. Press up, down, right and left
arrow keys to move cursor to V- Range and I-Range fields where options are available for user to enable or disable for measurement.
9. When, for example, disabling 15V
option, move cursor to 15V followed by pressing Enter button and using up and down arrow keys to select Off. Press Enter key to confirm setting in the end.
Enter
Enter
Option
On
When the option(s) is On, the selected range(s) will be applied to measurement.
Off
It is able is skip certain measurement range(s) that are not used by turning Off. By doing so, It can reduce measured data loss which happens while ranges are switched.
Default value
On
40
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BASIC SETTING
Steps
1. Use left and right arrow keys on
the front panel to select System function key.
2. Press Enter button to Enter
SYSTEM INFORMATION setting screen.
Enter
3. Press right arrow key to select
Config key.
4. Press Enter button to enter
SYSTEM CONFIG setting screen.
Enter
Background
Continue the following setting from SYSTEM CONFIG setting screen
Steps
1. Press down arrow key.
Setting up System status
System configuration setting screen
Setting up power on status
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
41
T3PM1006 User Manual
2. Press Enter button to enter Power
On Status Setup item. Use up and
down keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
Previous:
The status of device on powering on is set to the status before the last shutdown.
Default:
The status of device on powering on is set to the factory default status.
Default value
Default
Background
Continue the following setting from SYSTEM CONFIG setting screen
Steps
1. Press down arrow key twice.
x2
Setting up brightness
42
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BASIC SETTING
2. Press Enter button to enter
Brightness item. Use up and down
keys to select a number and then press Enter button again to confirm your selection.
Enter
Enter
Option
1 to 9
The display is the darkest when set to
1. On the contrary, the brightest when set to 9.
Default option
7
Background
Continue the following setting from SYSTEM CONFIG setting screen
Steps
1. Press down arrow key three times.
x3
Setting up key sound
43
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
2. Press Enter button to enter Key
Sound item. Use up and down
arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
On:
A short sound is heard from the speaker of device when pressing the keys on the front panel.
Off:
No sound from the speaker of device when pressing the keys on the front panel.
Default option
Off
Background
Continue the following setting from SYSTEM CONFIG setting screen
Steps
1. Press down arrow key four times.
x4
Setting up interface
44
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
BASIC SETTING
2. Press Enter button to enter I/O
Model item. Use up and down
arrow keys to select the desired option and then press Enter button again to confirm your selection.
Enter
Enter
Option
RS232:
If interface is set to RS232, the Baud Rate can be selected from the following options.
1200, 2400, 4800, 9600, 19200, 38400, 57600 or 115200
For details about configuring RS 232 interface, please see page 60.
USB:
For details about configuring USB interface, please see page 60.
GPIB:
Disabled
LAN:
If interface is set to LAN, the IP model can be selected “Manual” or “DHCP”.
Note that the socket port is fixed in “23”
for the unit. For details about configuring LAN interface, please see page 61.
Default value
RS232, 9600
45
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Measurement function ................................................ 47
Introduction to measurement parameters ............................47
Setting measurement parameters ..........................................48
Other functions .......................................................... 50
Introduction to other functions ..............................................50
Integration measurement function .............................. 52
Setting up Integrator measurement .......................................52
Introduction to integrator parameters ...................................55
Using the integrator function .................................................57
MEASUREMENT AND
OTHER FUNCTIONS
46
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MEASUREMENT AND OTHER FUNCTIONS
Parameter name
Display icon
Voltage
Vdc (DC voltage), Vrms (AC voltage)
Current
Idc (DC current), Irms (AC current)
Active Power
P
Apparent Power
VA
Reactive power
VAR
Power Factor
PF
Phase Angle
DEG
Frequency
IHz and VHz
Voltage Peak
V+pk and V-pk
Current Peak
I+pk and I-pk
Measurement function
The T3PM1006 provides a wide range of basic electricity and power measurement functions. It is equipped with different accurate measurement parameters for accurately measuring the voltage, current, power, DC/AC/AC + DC, power factor, harmonics, frequency, etc. The input impedance of the device is 2.4MΩ, the maximum input voltage is 600Vrms. There are 2 sets of internal resistance (Shunt), 500mΩ and 5mΩ respectively. The maximum input current is 20Arms. The device will issue a warning sound when the input voltage and current exceed 700 Vrms or 25Arms.
Introduction to measurement parameters
47
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Active Power Peak
P+pk and P-pk
Total Harmonic Distortion
THDI and THDV
Crest factor
CFV, CFI
Steps
1. Use left and right arrow keys on
the front panel to select Parameter function key.
2. Press Enter button. A measurement
parameter will be highlighted in green.
Enter
3. Press Enter button to confirm
setting or use up and down arrow keys to select other desired measurement parameter.
4. You can use same method as
shown in the last step to set other measurement parameters in this screen.
Enter
Enter
Setting measurement parameters
Please follow the steps blow to set the measurement parameters
48
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MEASUREMENT AND OTHER FUNCTIONS
Switching display mode
5. In standard display mode, you
simply press the Enter button to switch display mode to simple.
Enter
6. Press ESC button to return back to
original display mode.
ESC
49
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Max. Hold KeyLock Display HoldMode
Function name
Button
Description
MAX Hold
MAX Hold
When the MAX Hold button is pressed, the MAX Hold status icon will light in red in the LCD display to indicate that this function is activated. To deactivate this function, press this button again.
If the MAX Hold function is activated, the display value on the display is updated only when the current measured value is greater than the previous measured value. The maximum display value is retained on the display.
Other functions
Introduction to other functions
50
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MEASUREMENT AND OTHER FUNCTIONS
Mode
Mode
Press this button to select measurement mode. There are 3 measurement modes.
AC+DC: Displays all the
components of the measurement signal
DC: Displays the DC part of the
measurement signal.
AC: Displays the AC part of the
measurement signal.
Hold
Hold
When the Hold button is pressed, the Hold status icon will light in red in the LCD display to indicate that this function is activated. To deactivate this function, press this button again.
When the Hold function is activated, the displayed value on the LCD display is not updated and the range is locked. Measurement is performed in the background.
Local/ KeyLock
Key Lock
Local
Dual function key. When Remote mode is activated, press this button to deactivate Remote mode and switch to Local mode. When Remote mode is not activated, this button is used as lock key of keypad.
Enter
Enter
This button is used to select function or confirm selection.
ESC
ESC
This button is used to exit current screen or return to main measurement screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
51
T3PM1006 User Manual
Steps
1. Use left and right arrow keys on
the front panel to select Integrator function key.
2. Press Enter button to enter the
integrator measurement screen.
Enter
3. Press right arrow key to select Set
key.
Integration measurement function
Setting up Integrator measurement
Find Quality Products Online at: sales@GlobalTestSupply.com
52
www.GlobalTestSupply.com
MEASUREMENT AND OTHER FUNCTIONS
Select integrator measurement mode
4. Press Enter button to enter
integrator measurement setting screen.
Enter
5. Press Enter button to enter Mode
item. Use up and down arrow keys to toggle between Manual and Standard mode. Press Enter button again to confirm your selection.
If you select Manual mode, the Set time is disabled and displayed in gray.
Enter
Enter
Find Quality Products Online at: sales@GlobalTestSupply.com
53
www.GlobalTestSupply.com
T3PM1006 User Manual
If you select standard mode, you need to set integrator measurement time before using integrator function. It can be set from 1 second to 9999 hours, 59 minutes and 59 seconds.
6. Press down arrow key to select
Function item in the integrator measurement setting screen.
Select integrator measurement function
7. Press Enter button to enter
Function item. Use up and down
arrow keys to toggle between Ampere Hours and Watt Hours. Press Enter button again to confirm your selection.
Enter
Enter
Find Quality Products Online at: sales@GlobalTestSupply.com
54
www.GlobalTestSupply.com
MEASUREMENT AND OTHER FUNCTIONS
Parameter name
Description
Mode
Manual Standard
Function
Watt Hours
WP: Total power WP+: Positive total power WP-: Negative total power
Ampere Hours
q: Total mAh q+: Positive total mAh q-: Negative total mAh
Set time
It indicates the time of integrator measurement to be set. It can be set from 1 second to 9999 hours, 59 minutes and 59 seconds.
Test time
It indicates that elapsed time of integrator measurement.
Introduction to integrator parameters
55
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
State
Running
Integrator measurement is in progress.
Stop
Integrator measurement has been stopped manually.
Timeout
The time for running integrator measurement is up.
Reset
The integrator measurement status is cleared.
56
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Manual mode
1. In manual mode, you can directly
press the Start button in the front panel to start integrator function.
Start
INTEGRATOR
2. To stop integration function, press
the Stop button in the front panel.
Stop
3. Press the Reset button in the front
panel to clear integrator.
Reset
MEASUREMENT AND OTHER FUNCTIONS
Using the integrator function
57
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Standard mode
1. Set integrator measurement time before using
integrator function.
2. Other steps are the same as running in manual
mode.
When integrator is running, the test time will increase until the setting integrator measurement time.
Note
In the integration process, select the Measure
key and press Enter button to return main measurement screen. Select Integrator key and press Enter button to switch back to integration measurement screen.
In the integration process, you cant change
measurement range and enter system to set measurement parameters.
In the integration process, if the voltage or
current measurement value is exceeded, the measured value will display in red.
Find Quality Products Online at: sales@GlobalTestSupply.com
58
www.GlobalTestSupply.com
REMOTE CONTROL
Configure Remote Control Interface ............................ 60
USB Interface ......................................................................... 60
Configure USB Interface ........................................................ 60
Configure RS232 Interface ..................................................... 60
Configure LAN Interface ........................................................ 61
Return to Local Control ............................................... 63
REMOTE CONTROL
This chapter describes basic configuration of IEEE488.2 based remote control. For a command list, refer to the Command Overview chapter on page 64.
59
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
USB Configuration
PC connector
Type A, host
T3PM1006 connector
Rear panel Type B, slave
Speed
1.1/2.0 (full speed/high speed)
USB Class
CDC (Communications device class)
Hardware flow control
Off
Data Bits
8
Stop bit
1
RS232 Configuration
Selectable Baud rate
1200, 2400,4800, 9600, 19200, 38400, 57600, 115200
Parity
None
Hardware flow control
Off
Data Bits
8
Stop bit
1
Configure Remote Control Interface
USB Interface
The USB device port on the rear panel is used for remote control. The USB port is configured as CDC interface.
When configured to CDC, the USB port on the T3PM1006 will appear as a virtual COM port to a connected PC. Any terminal program that can communicate via a serial port can be used for remote control. Before the T3PM1006 can be used for remote control using the CDC USB class, install the appropriate CDC USB driver available for download on the Teledyne LeCroy website.
Configure USB Interface
Configure RS232 Interface
60
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
REMOTE CONTROL
RS232 Pin Assignments
Pin 2: RxD Pin 3: TxD Pin 5: GND Pin 1, 4, 6 ~ 9: No Connection
1
2345
6789
PC Connection
Use a Null Modem connection as shown in the diagram below.
T3PM1006 PC
RxDPin2 RxD Pin2
GNDPin5 GND Pin5
TxD Pin3
TxDPin3
Background
Continue the following setting from SYSTEM CONFIG setting screen
Steps
1. Press down arrow key four times.
x4
2. Press Enter button to enter I/O
Model item. Use up and down
arrow keys to select LAN option and then press Enter button again to confirm your selection.
Enter
Enter
Configure LAN Interface
61
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
3. Select a desired IP Model.
Option
Manual
Set up IP Address, Subnet mask and Gateway manually.
DHCP
DHCP server automatically assigns IP Address, Subnet mask and Gateway.
62
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
REMOTE CONTROL
Background
When the unit is in remote control mode, the RMT icon above the main display can be seen. When this icon is not displayed, it indicates that the unit is in local control mode.
Procedure
1. Press the LOCAL key when in remote mode.
2. The unit will go back into local mode and the
RMT icon will turn off.
Return to Local Control
63
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Compatible Standard
IEEE488.2
Partial compatibility
SCPI, 1994
Partial compatibility
Command Structure
SCPI (Standard Commands for Programmable Instruments) commands follow a tree-like structure, organized into nodes. Each level of the command tree is a node. Each keyword in a SCPI command represents each node in the command tree. Each keyword (node) of a SCPI command is separated by a colon (:). For example, the diagram below shows an SCPI sub-structure and a command example.
DCAC
DC AC
INPut:MODE DC
:MODE
:INPut
COMMAND OVERVIEW
The Command overview chapter lists all programming commands in functional order as well as alphabetical order. The command syntax section shows you the basic syntax rules you have to apply when using commands.
Command Syntax
64
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
Command Types
There are a number of different instrument commands and queries. A command sends instructions or data to the unit and a query receives data or status information from the unit.
Command types
Simple
A single command with/without a parameter
Example
:INPut:MODE DC
Query
A query is a simple or compound command followed by a question mark (?). A parameter (data) is returned.
Example
:INPut:CFACtor?
Command Forms
Commands and queries have two different forms, long and short. The command syntax is written with the short form of the command in capitals and the remainder (long form) in lower case. The commands can be written either in capitals or lower-case, just so long as the short or long forms are complete. An incomplete command will not be recognized. Below are examples of correctly written commands.
Long form
:INPut:SYNChronize VOLTage
:COMMunicate:HEADer ON
Short form
:INP:SYNC VOLT :COMM:HEAD ON
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
65
T3PM1006 User Manual
Square Brackets
Commands that contain square brackets indicate that the contents are optional. The function of the command is the same with or without the square bracketed items, as shown below. For example, for the query:
[:INPut]:FILTer? Both :INPut:FILTer? and :FILTer? are valid forms.
Command Format
300
1
2 3
:INPut:VOLTage:RANGe
1. Command header
2. Space
3. Parameter 1
Common Input Parameters
Type
Description
Example
<Boolean>
boolean logic
0, 1
<NR1>
integers
0, 1, 2, 3
<NR2>
decimal numbers
0.1, 3.14, 8.5
<NR3>
floating point with exponent
4.5e-1, 8.25e+1
<NRf>
any of NR1, 2, 3
1, 1.5, 4.5e-1
[MIN] (Optional parameter)
For commands, this will set the setting to the lowest value. This parameter can be used in place of any numerical parameter where indicated.
For queries, it will return the lowest possible value allowed for the particular setting.
66
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
[MAX] (Optional parameter)
For commands, this will set the setting to the highest value. This parameter can be used in place of any numerical parameter where indicated.
For queries, it will return the highest possible value allowed for the particular setting.
Message Terminator (EOL)
Remote Command
Marks the end of a command line. The following messages are in accordance with IEEE488.2 standard.
CR+LF
The most common EOL character is CR+LF
Message Separator
EOL or ; (semicolon)
Command Separator
67
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
SCPI Commands
*CLS ............................................................................. 70
*IDN ............................................................................ 70
*ESE ............................................................................. 70
*ESR ............................................................................. 71
*OPC ............................................................................ 71
*RST ............................................................................. 71
*SRE ............................................................................. 72
*STB ............................................................................. 72
COMMunciate Commands
:COMMunicate:HEADer ............................................. 73
:COMMunicate:REMote .............................................. 73
:COMMunicate:VERBose ............................................ 74
DISPlay Commands
:DISPlay[:NORMal]:ITEM<x> ...................................... 75
:DISPlay:INTegrate:ITEM<x> ...................................... 76
:DISPlay:PAGE ............................................................. 77
HARMonics Command
:HARMonics:THD ....................................................... 78
HOLD Command
:HOLD .......................................................................... 79
INPut Commands
[:INPut]:CFACtor .......................................................... 80
[:INPut]:MODE ............................................................ 80
[:INPut]:VOLTage:RANGe ........................................... 81
[:INPut]:VOLTage:AUTO ............................................. 81
[:INPut]:VOLTage:CONFig .......................................... 82
[:INPut]:CURRent:RANGe ........................................... 82
[:INPut]:CURRent:AUTO ............................................. 83
[:INPut]:CURRent:CONFig .......................................... 83
[:INPut]:RCONfig ......................................................... 84
[:INPut]:SCALing:{VT/PT|CT}:STATe .......................... 84
[:INPut]:SCALing:{VT/PT|CT}:RATio ........................... 85
[:INPut]:SYNChronize .................................................. 85
[:INPut]:FILTer ............................................................. 85
Command List
Find Quality Products Online at: sales@GlobalTestSupply.com
68
www.GlobalTestSupply.com
COMMAND OVERVIEW
[:INPut]:ZERO ..............................................................86
INTegrate commands
:INTegrate:MODE ........................................................87
:INTegrate:FUNCtion ..................................................87
:INTegrate:TIMer .........................................................88
:INTegrate:STARt .........................................................88
:INTegrate:STOP ..........................................................88
:INTegrate:RESet .........................................................88
:INTegrate:STATe ........................................................89
MEASure commands
:MEASure:AVERaging:COUNt ....................................90
:MEASure:MHOLd ......................................................90
NUMeric commands
:NUMeric[:NORMal]:VALue ........................................91
:NUMeric[:NORMal]:NUMBer ....................................92
:NUMeric[:NORMal]:ITEM<x> ....................................92
:NUMeric[:NORMal]:PRESet .......................................94
:NUMeric[:NORMal]:CLEar .........................................96
:NUMeric[:NORMal]:DELete .......................................97
:NUMeric[:NORMal]:HEADer .....................................97
SYSTem commands
:SYSTem:MODel ..........................................................98
:SYSTem:SERial ...........................................................98
:SYSTem:VERSion ........................................................98
:SYSTem:KLOCk ..........................................................99
:SYSTem:BRIGhtness ..................................................99
:SYSTem:KEY:BEEPer ................................................100
STATus command
:STATus:ERRor ...........................................................101
69
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
*CLS ............................................................................. 70
*IDN ............................................................................ 70
*ESE ............................................................................. 70
*ESR ............................................................................. 71
*OPC ............................................................................ 71
*RST ............................................................................. 71
*SRE ............................................................................. 72
*STB ............................................................................. 72
*CLS
Description
Clears the Event Status register (Output Queue, Operation Event Status, Standard Event Status).
Syntax
*CLS
*IDN
Description
Returns the manufacturer, model number, serial number, and system version of the instrument.
Query Syntax
*IDN?
Return parameter
<String>
Example
*IDN?
->TELEDYNE,T3PM1006, GXXXXXXXX,V1.00
*ESE
Description
Sets or returns the ESER (Event Status Enable Register) contents.
Syntax
Query Syntax
*ESE <NR1>
*ESE?
Set
Query
Set
Query
SCPI Commands
70
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
Parameter/
Return parameter
<NR1>
0~255
Example
*ESE 65 Set the ESER to 01000001
*ESE?
->130 ESER=10000010
*ESR
Description
Returns SESR (Standard Event Status Register).
Query Syntax
*ESR?
Return parameter
<NR1>
0~255
Example
*ESR?
->198 SESR=11000110
*OPC
Description
Sets or returns the operation complete bit (bit0) in SERS (Standard Event Status Register) when all pending operations are completed.
Syntax
Query Syntax
*OPC
*OPC?
Return parameter
<NR1>0
Operation isnt completed
<NR1>1
Operation is completed
Example
*OPC?
Returns 1.
*RST
Description
Initializes the settings
Syntax
*RST
Query
Set
Query
Set
71
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
*SRE
Description
Sets or returns SRER (Service Request Enable Register)
Syntax
*SRE <NR1>
Query Syntax
*SRE?
Parameter/ Return parameter
<NR1>
0~255
Example
*SER 7 Set the the SRER to 00000111 *SRE?
->3
SRER=00000011
*STB
Description
Returns the SBR (Status Byte Register) contents.
Query Syntax
*STB?
Return parameter
<NR1>
0~255
Example
*STB 8
->81 SESR=01010001
Set
Query
Query
72
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
:COMMunicate:HEADer ..............................................73
:COMMunicate:REMote ..............................................73
:COMMunicate:VERBose ............................................74
:COMMunicate:HEADer
Description
Sets or returns whether headers are attached to query responses
Syntax
:COMMunicate:HEADer <Boolean>|{OFF | ON}
Query Syntax
:COMMunicate:HEADer?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the header function off
1
Turn the header function on
Example
:COMMUNICATE:HEADER ON
:COMMUNICATE:HEADER?
->:COMMUNICATE:HEADER 1
Note
Example of a response with a header
:INPUT:VOLTAGE:RANGE 150.0E+00
Example of a response without a header 150.0E+00
:COMMunicate:REMote
Description
Sets or returns the T3PM1006 to remote or local mode. ON is remote mode.
Syntax
:COMMunicate:REMote <Boolean>|{OFF | ON}
Query Syntax
:COMMunicate:REMote?
Set
Query
Set
Query
COMMunciate Commands
73
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
T3PM1006 User Manual
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the remote function off
1
Turn the remote function on
Example
:COMMUNICATE:REMOTE ON
:COMMUNICATE:REMOTE?
->:COMMUNICATE:REMOTE 1
:COMMunicate:VERBose
Description
Sets or returns whether the response to a query is returned fully spelled out or in its abbreviated form.
Syntax
:COMMunicate:VERBose <Boolean>|{OFF | ON}
Query Syntax
:COMMunicate:VERBose?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the verbose function off
1
Turn the verbose function on
Example
:COMMUNICATE:VERBOSE ON
:COMMUNICATE:VERBOSE?
->:COMMUNICATE:VERBOSE 1
Note
Example of a response fully spelled out
:INPUT:VOLTAGE:RANGE 150.0E+00
Example of a response in abbreviated form
:VOLT:RANG 150.0E+00
Set
Query
74
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
:DISPlay[:NORMal]:ITEM<x> ......................................75
:DISPlay:INTegrate:ITEM<x> ......................................76
:DISPlay:PAGE .............................................................77
:DISPlay[:NORMal]:ITEM<x>
Description
Sets or returns a normal measurement data display item.
Syntax
:DISPlay[:NORMal]:ITEM<x> <Function>
Query Syntax
:DISPlay[:NORMal]:ITEM<x>?
Parameter/
Return parameter
<x>
1 to 8 (display)
<Function>
{U|UPPeak|UMPeak|I|IPPeak|IMPeak |P|PPPeak|PMPeak|S|Q|LAMBda|CFU |CFI|PHI|FU|FI|UTHD|ITHD}
Example
:DISPLAY:NORMAL:ITEM1 U
:DISPLAY:NORMAL:ITEM1?
->:DISPLAY:NORMAL:ITEM1 U
<Function>
Function
T3PM1006 Indicator
U
Voltage U
[V]
UPPeak
Maximum voltage: U+pk
[V+pk]
UMPeak
Minimum voltage: U-pk
[V-pk]
I
Current I
[I]
IPPeak
Maximum current: I+pk
[I+pk]
IMPeak
Minimum current: I-pk
[I-pk]
P
Active power P
[P]
PPPeak
Maximum power: P+pk
[P+pk]
PMPeak
Minimum power: P-pk
[P-pk]
S
Apparent power S
[VA]
Set
Query
DISPlay Commands
75
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Q
Reactive power Q
[VAR]
LAMBda
Power factor λ
[PF]
CFU
Voltage factor λ
[CFV]
CFV
Current factor λ
[CFI]
PHI
Phase difference Φ
[DEG]
FU
Voltage frequency fu
[VHz]
FI
Current frequency fI
[AHz]
UTHD
Total harmonic distortion of voltage Uthd
[THDV]
ITHD
Total harmonic distortion of current Ithd
[THDI]
:DISPlay:INTegrate:ITEM<x>
Description
Sets or returns an Integrate measurement data display item.
Syntax
:DISPlay:INTegrate:ITEM<x> <Function>
Query Syntax
:DISPlay:INTegrate:ITEM<x>?
Parameter/
Return parameter
<x>
1 to 2(display)
<Function>
{WHP|WHM|AHP|AHM|U|I }.
Example
:DISPLAY:INTEGRATE:ITEM1 WHP
:DISPLAY:INTEGRATE:ITEM1?
->:DISPLAY:INTEGRATE:ITEM1 WHP
<Function>
Function
T3PM1006 Indicator
WHP
Positive watt hour WP+
[WP+]
WHM
Positive watt hour WP-
[WP-]
AHP
Positive ampere hour q+
[q+]
AHM
Positive ampere hour q
[q-] U Voltage U
[V]
I
Current I
[I]
Set
Query
76
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
:DISPlay:PAGE
Description
Sets or returns the display page item.
Syntax
:DISPlay:PAGE <Function>
Query Syntax
:DISPlay:PAGE?
Parameter/
Return parameter
<Function>
{MEASurement|ENLArge|INTEgral| SYSTem_INFO|SYSTem_CONFig|SETUp}
Example
:DISPLAY:PAGE MEASUREMENT
:DISPLAY:PAGE?
->:DISPLAY:PAGE MEASUREMENT
Set
Query
77
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
:HARMonics:THD
Description
Sets or returns the equation used to compute the THD (total harmonic distortion).
Syntax
:HARMonics:THD {TOTal|FUNDamental}
Query Syntax
:HARMonics:THD?
Parameter/
Return parameter
TOTal
(CSA)
FUNDamental
(IEC)
Example
:HARMONICS:THD FUNDAMENTAL
:HARMONICS:THD?
->:HARMONICS:THD FUNDAMENTAL
Set
Query
HARMonics Command
78
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
:HOLD
Description
Sets or returns the on/off state of the output hold feature for display, communication, and other types of data.
Syntax
:HOLD <Boolean>|{OFF|ON}
Query Syntax
:HOLD?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the hold function off
1
Turn the hold function on
Example
:HOLD OFF
:HOLD?
->:HOLD 0
Set
Query
HOLD Command
79
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
[:INPut]:CFACtor .......................................................... 80
[:INPut]:MODE ............................................................ 80
[:INPut]:VOLTage:RANGe ........................................... 81
[:INPut]:VOLTage:AUTO ............................................. 81
[:INPut]:VOLTage:CONFig .......................................... 82
[:INPut]:CURRent:RANGe ........................................... 82
[:INPut]:CURRent:AUTO ............................................. 83
[:INPut]:CURRent:CONFig .......................................... 83
[:INPut]:RCONfig ......................................................... 84
[:INPut]:SCALing:{VT/PT|CT}:STATe .......................... 84
[:INPut]:SCALing:{VT/PT|CT}:RATio ........................... 85
[:INPut]:SYNChronize.................................................. 85
[:INPut]:FILTer ............................................................. 85
[:INPut]:ZERO .............................................................. 86
[:INPut]:CFACtor
Description
Sets or returns the crest factor.
Syntax
[:INPut]:CFACtor {<NRf>}
Query Syntax
[:INPut]:CFACtor?
Parameter/ Return parameter
<NR1>
3, 6
Example
:INPUT:CFACTOR 3
:INPUT:CFACTOR?
->:INPUT:CFACTOR 3
[:INPut]:MODE
Description
Sets or returns the voltage and current measurement mode.
Syntax
[:INPut]:MODE {DC|ACDC|AC}
Query Syntax
[:INPut]:MODE?
Set
Query
Set
Query
INPut Commands
80
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
Parameter/ Return parameter
Select the dc measurement mode.
Select the acdc measurement mode.
Select the ac mode.
Example
:INPUT:MODE DC
:INPUT:MODE?
->:INPUT:MODE DC
[:INPut]:VOLTage:RANGe
Description
Sets or returns the voltage range.
Syntax
[:INPut]:VOLTage:RANGe {<Voltage>}
Query Syntax
[:INPut]:VOLTage:RANGe?
Parameter/ Return parameter
<Voltage>
15, 30, 60, 150, 300, 600(V) when the crest factor is set to 3.
7.5, 15, 30, 75, 150, 300(V) when the crest factor is set to 6
Example
:INPUT:VOLTAGE:RANGE 600V
:INPUT:VOLTAGE:RANGE?
->:INPUT:VOLTAGE:RANGE 600.0E+00
[:INPut]:VOLTage:AUTO
Description
Sets or returns the voltage auto range on/off state.
Syntax
[:INPut]:VOLTage:AUTO {<Boolean>}
Query Syntax
[:INPut]:VOLTage:AUTO?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the voltage auto range function off.
1
Turn the voltage auto range function on.
Set
Query
Set
Query
81
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Example
:INPUT:VOLTAGE:AUTO ON
:INPUT:VOLTAGE:AUTO?
->:INPUT:VOLTAGE:AUTO 1
[:INPut]:VOLTage:CONFig
Description
Sets or returns the valid voltage range.
Syntax
[:INPut]:VOLTage:CONFig {ALL|<Voltage>[,Voltage]}
Query Syntax
[:INPut]:VOLTage:CONFig?
Parameter/ Return parameter
ALL
All ranges are valid.
<Voltage>
See(:INPut:VOLTage:RANGe).
Example
:INPUT:VOLTAGE:CONFIG 300,150,30
:INPUT:VOLTAGE:CONFIG?
->:INPUT:VOLTAGE:CONFIG 300.0E+00,150.0E+00,
30.0E+00
[:INPut]:CURRent:RANGe
Description
Sets or returns the current range.
Query
[:INPut]:CURRent:RANGe {<Current>}
Query Syntax
[:INPut]:CURRent:RANGe?
Parameter/ Return parameter
<Current>
5, 10, 20, 50, 100, 200, 500(mA) 1, 2, 5, 10, 20(A) when the crest factor is set to 3.
2.5, 5, 10, 25, 50, 100, 250(mA)
0.5, 1, 2.5, 5, 10(A)when the crest factor is set to 6
Example
:INPUT:CURRENT:RANGE 20A
:INPUT:CURRENT:RANGE?
->:INPUT:CURRENT:RANGE 20.0E+00
Set
Query
Set
Query
82
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
[:INPut]:CURRent:AUTO
Description
Sets or returns the current auto range on/off state.
Syntax
[:INPut]:CURRent:AUTO {<Boolean>}
Query Syntax
[:INPut]:CURRent:AUTO?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the current auto range function off.
1
Turn the current auto range function on.
Example
:INPUT:CURRENT:AUTO ON
:INPUT:CURRENT:AUTO?
->:INPUT:CURRENT:AUTO 1
[:INPut]:CURRent:CONFig
Description
Sets or returns the valid current range.
Syntax
[:INPut]:CURRent:CONFig {ALL|<Current>[,Current]}
Query Syntax
[:INPut]:CURRent:CONFig?
Parameter/ Return parameter
ALL
All ranges are valid.
<Current>
See(:INPut:CURRent:RANGe).
Example
:INPUT:CURRENT:CONFIG 20,10,1
:INPUT:CURRENT:CONFIG?
->:INPUT:CURRENT:CONFIG 20.0E+00,10.0E+00,
1.0E+00
Set
Query
Set
Query
83
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
[:INPut]:RCONfig
Description
Sets or returns the on/off state of the range configuration (valid range selection) feature.
Syntax
[:INPut]:RCONfig {<Boolean>|OFF|ON}
Query Syntax
[:INPut]:RCONfig?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the range configuration feature off.
1
Turn the range configuration feature on.
Example
:INPUT:RCONFIG ON
:INPUT:RCONFIG?
->:INPUT:RCONFIG 1
[:INPut]:SCALing:{VT/PT|CT}:STATe
Description
Sets or returns the scaling vt/pt,ct on/off state.
Syntax
[:INPut]:SCALing:{VT/PT|CT}:STATe {<Boolean>}
Query Syntax
[:INPut]:SCALing:{VT/PT|CT}:STATe?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the scaling vt/pt, ct function off.
1
Turn the scaling vt/pt, ct function on.
Example
:INPUT:SCALING:VT:STATE ON
:INPUT:SCALING:VT:STATE?
->:INPUT:SCALING:VT:STATE 1
Set
Query
Set
Query
84
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
[:INPut]:SCALing:{VT/PT|CT}:RATio
Description
Collectively Sets or returns the vt/pt ratio or ct ratio.
Syntax
[:INPut]:SCALing:{VT/PT|CT|}:RATio {<NRf>}
Query Syntax
[:INPut]:SCALing:{VT/PT|CT}: RATio?
Parameter/ Return parameter
<NRf>
1.000 to 9999.999
Example
:INPUT:SCALING:VT:RATIO 1
:INPUT:SCALING:VT:RATIO?
->:INPUT:SCALING:VT:RATIO 1
[:INPut]:SYNChronize
Description
Sets or returns the synchronization source.
Syntax
[:INPut]:SYNChronize {VOLTage|CURRent|OFF}
Query Syntax
[:INPut]:SYNChronize?
Parameter/ Return parameter
Select the voltage synchronization source.
Select the current synchronization source.
Select the off synchronization source.
Example
:INPUT:SYNCHRONIZE VOLTAGE
:INPUT:SYNCHRONIZE?
->:INPUT:SYNCHRONIZE VOLTAGE
[:INPut]:FILTer
Description
Sets or returns the filter state.
Syntax
[:INPut]:FILTer {<Boolean>}
Query Syntax
[:INPut]:FILTer?
Set
Query
Set
Query
Set
Query
85
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the filter function off.
1
Turn the filter function on.
Example
:INPUT:FILTER OFF
:INPUT:FILTER?
->:INPUT:FILTER 0
[:INPut]:ZERO
Description
Sets or returns the zero state.
Syntax
[:INPut]:ZERO {<Boolean>}
Query Syntax
[:INPut]:ZERO?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the zero function off.
1
Turn the zero function on.
Example
:INPUT:ZERO OFF
:INPUT:ZERO?
->:INPUT:ZERO 0
Set
Query
86
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
:INTegrate:MODE ........................................................87
:INTegrate:FUNCtion ..................................................87
:INTegrate:TIMer .........................................................88
:INTegrate:STARt .........................................................88
:INTegrate:STOP ..........................................................88
:INTegrate:RESet .........................................................88
:INTegrate:STATe ........................................................89
:INTegrate:MODE
Description
Sets or returns the integration mode.
Syntax
:INTegrate:MODE {MANUal | STANdard}
Query Syntax
:INTegrate:MODE?
Parameter/ Return parameter
MANUal
Continuous integration mode.
STANdard
Standard integration mode.
Example
:INTEGRATE:MODE MANUAL
:INTEGRATE:MODE?
->:INTEGRATE:MODE MANUAL
:INTegrate:FUNCtion
Description
Sets or returns the integration function.
Syntax
:INTegrate:FUNCtion {WATT | AMPEre}
Query Syntax
:INTegrate: FUNCtion?
Parameter/ Return parameter
Select the integration function watt.
Select the integration function ampere.
Example
:INTEGRATE:FUNCTION WATT
:INTEGRATE:FUNCTION?
->:INTEGRATE:FUNCTION WATT
Set
Query
Set
Query
INTegrate Commands
87
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
:INTegrate:TIMer
Description
Sets or returns the integration timer value.
Syntax
:INTEGrate:TIMer {<NRf>,<NRf>,<NRf>}
Query Syntax
:INTEGrate:TIMer?
Parameter/ Return parameter
{<NRf>,<NRf>,<NRf>}
0,0,0 to 9999,59,59
First <NRf>
0 to 9999 (hours)
Second <NRf>
0 to 59 (minutes)
Third <NRf>
0 to 59 (seconds)
Example
:INTEGRATE:TIMER 1,0,0
:INTEGRATE:TIMER?
->:INTEGRATE:TIMER 1,0,0
:INTegrate:STARt
Description
Starts integration.
Syntax
:INTegrate:STARt
Example
:INTEGRATE:START
:INTegrate:STOP
Description
Stops integration.
Syntax
:INTegrate:STOP
Example
:INTEGRATE:STOP
:INTegrate:RESet
Description
Resets the integrated value.
Syntax
:INTegrate:RESet
Example
:INTEGRATE:RESET
Set
Query
Set
Set
Set
88
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
:INTegrate:STATe
Description
Queries the integration status.
Syntax
:INTegrate:STATe?
Example
:INTEGRATE:STATE?
->RESET
Response
Overflow
Integration overflows.
RESET
Integration resets.
RUNNING
Integration is in progress.
STOP
Integration stops.
TIMEUP
Integration stops due to integration timeout.
Query
89
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
:MEASure:AVERaging:COUNt .................................... 90
:MEASure:MHOLd ...................................................... 90
:MEASure:AVERaging:COUNt
Description
Sets or returns the averaging coefficient.
Syntax
:MEASure:AVERaging:COUNt {<NRf>}
Query Syntax
:MEASure:AVERaging:COUNt?
Parameter/ Return parameter
<NRf>
1, 2, 4, 8, 16, 32, 64
Example
:MEASURE:AVERAGING:COUNT 8
:MEASURE:AVERAGING:COUNT?
->:MEASURE:AVERAGING:COUNT 8
:MEASure:MHOLd
Description
Sets the MAX hold on/off state.
Syntax
:MEASure:MHOLd {<Boolean>}
Query Syntax
MEASure:MHOLd?
Parameter
<Boolean>0
OFF
<Boolean>1
ON
Return parameter
0
Turn the MAX hold function off.
1
Turn the MAX hold function on.
Example
:MEASURE:MHOLD ON
:MEASURE:MHOLD?
->:MEASURE:MHOLD 1
Set
Query
Set
Query
MEASure Commands
90
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
:NUMeric[:NORMal]:VALue ........................................91
:NUMeric[:NORMal]:NUMBer ....................................92
:NUMeric[:NORMal]:ITEM<x> ....................................92
:NUMeric[:NORMal]:PRESet .......................................94
:NUMeric[:NORMal]:CLEar .........................................96
:NUMeric[:NORMal]:DELete .......................................97
:NUMeric[:NORMal]:HEADer .....................................97
:NUMeric[:NORMal]:VALue
Description
Returns the numeric data.
Syntax
:NUMeric[:NORMal]:VALue?
Example
:NUMERIC:NORMAL:VALUE?
->
103.79E+00,1.0143E+00,105.27E+00,..(omitted)..,50.0 01E+00
Numeric Data Format
Measurement values U, I, P, PPPeak, PMPeak, S,
Q, LAMBda, CFU, CFI, FU, FI, UTHD and ITHD
Integrated values WH, WHP, WHM, AH, AHP and
AHM. ASCII: <NR3> format. Example: [-]12.345E+00
Measurement values UPPeak, UMPeak, IPPeak
and IMPeak. ASCII: <NR3> format. Example: [-]12.34E+00
Measurement values (PHI)
ASCII: <NR3> = 0~9.9 format. Example:[-]9.9E+00 ASCII: <NR3> = 10~99.9 format. Example:[-]99.9E+00 ASCII: <NR3> = 100~999.9 format. Example:[-]999.9E+000
Query
NUMeric Commands
91
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
Elapsed integration time (TIME)
ASCII: <NR1> format in units of seconds. Example: 3600 for 1 hour (1:00:00).
No items (-----)
ASCII: NAN (Not A Number)
Error Data
Data does not exist (the display shows “-----”)
ASCII: NAN (Not A Number)
:NUMeric[:NORMal]:NUMBer
Description
Sets or returns the specified numeric data output item function.
Syntax
:NUMeric[:NORMal]:ITEM<x> {<Function>}ALL}
Query Syntax
:NUMeric[:NORMal]:NUMBer?
Parameter/ Return parameter
<NRf>
1 to 34(ALL)
Example
:NUMERIC:NORMAL:NUMBER 10
:NUMERIC:NORMAL:NUMBER
->:NUMERIC:NORMAL:NUMBER 10
Note
If the parameter is omitted from
the :NUMeric[:NORMal]:VALue? command, the numeric data items from 1 to the specified value are output in order.
By default, the number of numeric data items is set
to 3.
:NUMeric[:NORMal]:ITEM<x>
Description
Sets or returns the specified numeric data output item function.
Syntax
:NUMeric[:NORMal]:ITEM<x> {<Function>}
Query Syntax
:NUMeric[:NORMal]:ITEM<x>?
Set
Query
Set
Query
92
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
Parameter/ Return parameter
<x>
1 to 34 (item number)
<Function> {U|UPPeak|UMPeak|I|IPPeak|IMPeak
|P|PPPeak|PMPeak|S|Q|LAMBda|CFU |CFI|PHI|FU|FI|UTHD|ITHD|WH |WHP|WHM|AH|AHP|AHM|TIME |URANge|IRANge}
Example
:NUMERIC:NORMAL:ITEM1 U
:NUMERIC:NORMAL:ITEM1?
->:NUMERIC:NORMAL:ITEM1 U
<Function>
Function
T3PM1006 Indicator
U
Voltage U
[V]
UPPeak
Maximum voltage: U+pk
[V+pk]
UMPeak
Minimum voltage: U-pk
[V-pk]
I
Current I
[I]
IPPeak
Maximum current: I+pk
[I+pk]
IMPeak
Minimum current: I-pk
[I-pk]
P
Active power P
[P]
PPPeak
Maximum power: P+pk
[P+pk]
PMPeak
Minimum power: P-pk
[P-pk]
S
Apparent power S
[VA]
Q
Reactive power Q
[VAR]
LAMBda
Power factor λ
[PF]
CFU
Voltage factor λ
[CFV]
CFV
Current factor λ
[CFI]
PHI
Phase difference Φ
[DEG]
FU
Voltage frequency fu
[VHz]
FI
Current frequency fI
[AHz]
UTHD
Total harmonic distortion of voltage Uthd
[THDV]
93
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
ITHD
Total harmonic distortion of current Ithd
[THDI] WH
Watt hour WP
[WP]
WHP
Positive watt hour WP+
[WP+]
WHM
Positive watt hour WP-
[WP-]
AH
Ampere hour q
[q]
AHP
Positive ampere hour q+
[q+]
AHM
Positive ampere hour q
[q-]
TIME
Integration time
URANge
Voltage range
IRANge
Current range
:NUMeric[:NORMal]:PRESet
Description
Presets the numeric data output item pattern.
Syntax
:NUMeric[:NORMal]:PRESet {<NRf>}
Parameter/ Return parameter
<NRf>
1 to 4 Example
:NUMERIC:NORMAL:PRESET 1
Patterns 1
ITEM<x>
<Function>
1
U
2 I 3
P
Patterns 2
ITEM<x>
<Function>
1
U
2 I 3
P
4 S 5
Q
6
LAMBda
Set
94
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
7
PHI
8
FU 9 FI
Patterns 3
ITEM<x>
<Function>
1
U
2 I 3
P
4
S
5
Q
6
LAMBda
7
PHI
8
FU 9 FI
10
UPPeak
11
UMPeak
12
IPPeak
13
IMPeak
14
PPPeak
15
PMPeak
Patterns 4
ITEM<x>
<Function>
1
U
2 I 3 P 4
S
5
Q
6
LAMBda
7
PHI
8
FU 9 FI
95
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
10
UPPeak
11
UMPeak
12
IPPeak
13
IMPeak
14
TIME
15
WH
16
WHP
17
WHM
18
AH
19
AHP
20
AHM
21
PPPeaK
22
PMPeaK
23
CFU
24
CFI
25
UTHD
26
ITHD
27
URANge
28
IRANge
:NUMeric[:NORMal]:CLEar
Description
Clears numeric data output items (sets the items to -----“).
Syntax
:NUMeric[:NORMal]:CLEar {ALL|<NRf>[,<NRf>]}
Parameter
First <NRf>
1 to 34 (the number of the first item to clear)
Second <NRf>
1 to 34 (the number of the last item to clear)
Example
:NUMERIC:NORMAL:CLEAR ALL
Set
96
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
Note
If the 2nd <NRf> is omitted, the output item specified by the first and all following output items (up to number 34) are cleared.
:NUMeric[:NORMal]:DELete
Description
Deletes numeric data output items.
Syntax
:NUMeric[:NORMal]:DELete {ALL|<NRf>[,<NRf>]}
Parameter
First <NRf>
1 to 34 (the number of the first item to delete)
Second <NRf>
1 to 34 (the number of the last item to delete)
Example
:NUMERIC:NORMAL:DELETE 1 (Deletes ITEM1 and shifts ITEM2 and subsequent items forward).
:NUMERIC:NORMAL:DELETE 1,3 (Deletes ITEM1 to ITEM3 and shifts ITEM4 and subsequent items forward).:INTEGRATE:RESET
Note
When output items are deleted, subsequent items
shift forward to fill the empty positions. Empty positions at the end are set to -----.
If the second <NRf> is omitted, only the output
item specified by the first number is deleted.
:NUMeric[:NORMal]:HEADer
Description
Returns the numeric data header.
Syntax
:NUMeric[:NORMal]:HEADer?
Example
The data names of the items from 1 to the number specified by the :
NUMeric[:NORMal]:NUMBer command are output in order.
:NUMERIC:NORMAL:NUMBER 3
:NUMERIC:NORMAL:HEADER?
-> Urms,Irms,P
Set
Query
97
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
:SYSTem:MODel.......................................................... 98
:SYSTem:SERial ........................................................... 98
:SYSTem:VERSion........................................................ 98
:SYSTem:KLOCk .......................................................... 99
:SYSTem:BRIGhtness .................................................. 99
:SYSTem:KEY:BEEPer ................................................ 100
:SYSTem:MODel
Description
Returns the model code.
Syntax
:SYSTem:MODel?
Example
:SYSTEM:MODEL?
->:SYSTEM:MODEL "T3PM1006"
:SYSTem:SERial
Description
Returns the serial number.
Syntax
:SYSTem:SERial?
Example
:SYSTEM:SERIAL?
->:SYSTEM:SERIAL "123456789A"
Note
Returns the No. item string of the system Information menu.
:SYSTem:VERSion
Description
Returns the firmware version.
Syntax
:SYSTem:VERsion?
Example
:SYSTEM:VERSION?
->"V1.00"
Query
Query
Query
SYSTem Commands
98
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
COMMAND OVERVIEW
Note
Returns the Ver. item string of the system Information menu.
:SYSTem:KLOCk
Description
Sets or returns the on/off state of the key protection.
Syntax
:SYSTem:KLOCk {<Boolean>}
Query Syntax
:SYSTem:KLOCk?
Parameter
<Boolean> 0
OFF
<Boolean> 1
ON
Return parameter
0
Turn the key protection function off
1
Turn the key protection function on.
Example
:SYSTEM:KLOCK OFF
:SYSTEM:KLOCK?
->:SYSTEM:KLOCK 0
:SYSTem:BRIGhtness
Description
Sets or returns the brightness level.
Syntax
:SYSTem:BRIGhtness {<NRf>}
Query Syntax
:SYSTem:BRIGhtness?
Parameter/
Return parameter
<NRf>
1to 9
Example
:SYSTEM:BRIGHTNESS 7
:SYSTEM:BRIGHTNESS?
->:SYSTEM:BRIGHTNESS 7
Set
Query
Set
Query
99
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3PM1006 User Manual
:SYSTem:KEY:BEEPer
Description
Sets or returns the keyclick beeper state.
Syntax
:SYSTem:KEY:BEEPer {<Boolean>}
Query Syntax
:SYSTem:COMMunicate:LAN:CONFigure?
Parameter
<Boolean> 0
OFF
<Boolean> 1
ON
Return parameter
0
Turn the keyclick beeper function off.
1
Turn the keyclick beeper function on.
Example
:SYSTEM:KEY:BEEPER OFF
:SYSTEM:KEY:BEEPER?
->:SYSTEM:KEY:BEEPER 0
Set
Query
100
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com