When you notice any of the unusual conditions listed below, immediately
terminate operation and disconnect the power cable.
Please contact the TELEDYNE service center representative for repair of the
instrument. If you continue to operate without repairing the instrument, there
is a potential fire or shock hazard for operators.
The instrument is operating abnormally if:
The instrument emits abnormal noise, smell, smoke, or a spark-like light
during the operation.
The instrument generates high temperature or electrical shock during
operation.
The power cable, plug, or receptacle on the instrument is damaged.
Foreign substances or liquid has fallen into the instrument.
The following general safety precautions must be observed during all phases
of operation, service, and repair of this instrument. Failure to comply with
these precautions or with specific WARNINGS elsewhere in this manual may
impair the protection provided by the equipment. In addition it violates safety
standards of design, manufacture, and intended use of the instrument.
TELEDYNE assumes no liability for the
customer's failure to comply with these
requirements.
To avoid electric shock hazards, the instrument
chassis and cabinet must be connected to a safety
earth ground by the supplied power cable with
earth blade.
Do not operate the instrument in the presence of
inflammable gasses or fumes. Operation of any
electrical instrument in such an environment
constitutes a definite safety hazard.
Operating personnel must not remove instrument
covers. Component replacement and internal
adjustments must be made by qualified
maintenance personnel. Do not replace
components with the power cable connected.
Under certain conditions, dangerous voltages may
exist even with the power cable is removed. To
avoid injuries, always disconnect the power and
discharge the circuits before touching them.
Do not attempt internal service or adjustment
unless another person, capable of rendering first
aid and resuscitation, is present.
Because of the danger of introducing additional
hazards, do not install substitute parts or perform
unauthorized modifications to the instrument.
Return the instrument to a TELEDYNE
authorized service center for service and repair to
ensure that safety features are maintained.
If an abnormality is detected, contact the company and
transport the meter to your nearest TELEDYNE
service office. Save the packing box, shock-absorbing
material, and packaged items as you received them
you need to mail the instrument to an authorized
TELEDYNE distributor or service center.
Set up the T3LCR Series where the following
environmental requirements are satisfied.
Operating Environment
Ensure that the operating environment meets the
following requirements:
Temperature:
0ºC to 50ºC
Temperature range
at calibration:
temperature when performing
calibration)
Humidity: <70% at wet bulb temperature
-condensation)
0 to 2000m
To prevent electrical shock, disconnect the T3
power cable from the receptacle before cleaning. Use a
dry cloth or a cloth slightly dipped in water to clean the
casing. Do not attempt to clean the T3LCR Series
internally.
components, materials, and semiconductor devices over a
wide range of frequencies (10 Hz to 300 kHz) and test
With its built
This chapter contains general information about T3LCR Series .The
information is organized as follows.
Introduction
Main Specifications
Feature overview
Thank you for purchasing an T3LCR Series LCR meter.
The TELEDYNE T3LCR Series is a general-purpose
LCR meter for incoming inspection of components,
quality control, and laboratory use.
The T3LCR Series is used for evaluating LCR
signal levels (10.00mV to 2.00V & 100.0uA to 20.00mA).
-in comparator, the T3LCR Series can
output comparison/decision results for sorting
components into a maximum of ten bins. Furthermore,
by using the handler interface, the T3LCR Series can be
easily combined with a component handler, and a system
controller to fully automate component testing, sorting,
and quality-control data processing.
The list measure function permits entry of up to 10
frequencies or test signal levels points to be automatically
measured.
port of the LCR meter and if the meter detects that the
Ready. Press <Enter> to save screen. on the lower part
of the LCD screen.
If the keypad is not needed to enter a numeric value, a
and saved to the
the screen shot is saved successful
be shown in the message area at the bottom of the
screen. The location and name of the screenshot will also
be shown, for example:
F:\T3LCR1300\Screen\ScreenShot001.bmp.
This chapter describes names and functions of the front panel, rear panel,
and screen display and provides the basic procedures for operating the T3
Series.
measurements and avoids mutual inductance, interference
T3LCR Series is ready to be used as soon as the powerÂup sequence has completed. However, to achieve the
specification accuracy, first warm up the instrument for 30
minutes.
The T3LCR Series uses a four terminal measurement
configuration that provides easy, accurate and stable
from measurement signals, noise and other factors
inherent with other types of connections.
The T3LCR Series simultaneously measures four
components of complex impedance (parameters) in a
measurement cycle. These include a primary parameter,
secondary parameter and two monitor parameters.
The T3LCR has an internal data buffer that record up to
10000 measurement readings. These readings can be
saved to an external USB drive in a .csv file format.
These readings can then be opened on a PC using
software such as Windows Excel. When the LOG
function is active, the display is fixed to the [MEAS
DISPLAY] screen. The [FUNC], [FREQ] and [LEVEL]
settings are also fixed and cannot be changed.
Step 1. Press the [Measure] key.
Step 2. Use the left/right cursor keys to select the
LOG status field.
Step 3. Use the [START LOG] soft-keys to start a
new log.
Soft-key Function
START LOG To start a new measurement log.
Step 4. After a new log has started, the measurement
internal buffer. The options in the following
table are available during this recording
process:
saves the logs into the buffer. An
exclamation mark will be added after
the record counts (in the [LOG] status
field) to remind you that there are
recorded readings in the buffer that
have not yet been processed.
buffer to an external USB flash drive.
The internal buffer is cleared after this
operation.
BUFFER
SAVE & STOP Stops the recording and saves the
recorded readings in the internal
buffer to an external USB flash drive.
The internal buffer will be cleared
after the save operation.
If an external USB flash drive is not
plugged in, the meter will not stop the
ongoing recording.
buffer.
Step 5.
be added after the number of recorded
counts(in the [LOG] status field). The
following two operations are available in this
situation:
Soft-key Function
SAVE TO USB Saves the recorded readings in the
buffer to an external USB flash
drive. The internal buffer is cleared
after this operation.
those fixed trimming points. Please refer to the 12.1
Before saving
to the external USB flash drive, remember to plug the
external USB drive into the USB port on the front panel.
The recorded readings will be saved in a subdirectory
named MEAS DATA, whose parent directory will have
the same name as the T3LCR model being used. See the
example in below:
Example:
F:\T3LCR1300\MEAS DATE\ MEAS0000.CSV
Up to 9999 recording files can be saved, with their file
names from 0001.csv-9999.csv.
The size of the internal buffer can be adjusted in
setting field. The maximum data buffer size is 10000.
the [Measure] key and the [OPEN
SHORT] soft-key, the [ OPEN SHORT ] page appears.
In this page, the OPEN/SHORT/SPOT correction for
correcting the stray admittance and residual impedances
can be performed.
The OPEN and SHORT function performs open or
short correction over a range of fixed trimming points.
The correction for all other frequencies are interpolated
from
Specification page to see what the trimming frequency
points are for each T3LCR model.
The SPOT function will perform open, or short
correction at a user-
In the [ OPEN SHORT] page, you can configure each of
the following controls with the cursor placed in the
corresponding field:
The T3LCR OPEN correction capability cancels
errors due to the stray admittance (G, B) in parallel with
the device under test (Refer to Figure 4-3).
The OPEN correction data is taken at all preset
frequency points, independent of the test frequency you
set. Refer to the tables on page 130 for the trimming
frequency point lists for each model.
Step 1. Press the [Measure] key.
Step 2. Press the [OPEN SHORT] soft-key.
Step 3. Use the cursor key to select the [OPEN] field.
OFF Disables open correction.
MEAS OPEN Starts open correction (AC).
DCR OPEN Starts open correction (DC).
Step 4. Press the [MEAS OPEN] or [DCR OPEN]
soft-key. A dialog message, -circuit the
will appear.
Step 5. Connect the test fixture to the BNC terminals.
Do not connect the DUT to the fixture.
Step 6. Press the [OK] soft-key. The T3LCR will then
measure the open admittances (capacitance
and conductance) at the all test frequency
points.
LCR OPEN
sage is
shown on the display.
When the measurement has finished,
measurement, you can press the [ABORT]
soft-key to abort the OPEN correction.
Step 7. After the OPEN measurement has finished,
the meter will enable OPEN correction
automatically. You can, however, still choose
to manually enable or disable OPEN
correction by depressing the ON or OFF softÂkeys.
The short correction feature of the T3LCR Series
compensates for any residual impedances (R,X) that may
exist outside of calibration, such as the impedance of the
cables and the DUT connection points.
Step 1. Press the [Measure] key.
Step 2. Press the [OPEN SHORT] soft-key.
Step 3. Use the cursor key to select the [SHORT] field.
Soft-key Function
Step 4. Press the [MEAS SHORT] or [DCR SHORT]
soft-key, a dialog message, -circuit the
, will appear.
Step 5. Connect the test fixture to the BNC terminals
and short-circuit the test terminals.
Step 6. Press the [OK] soft-key. The T3LCR
the short impedances (inductances and
resistance) at the all test frequency points.
LCR SHORT
og message is
shown on the display.
When the measurement has finished,
measurement, you can press the [ABORT]
soft-key to abort the SHORT correction.
Step 7. After the SHORT measurement has finished,
the meter will enable SHORT correction
automatically. You can, however, still choose
to manually enable or disable SHORT
correction by depressing the ON or OFF softÂkeys.
The spot correction function involves performing
open/short correction at userÂYou can specify 1 frequency point.
Step 1. Press the [Measure] key.
Step 3. Press the [OPEN SHORT] soft-key.
Step 3. Use the cursor key to select the [SPOT] field.
Step 4. Enter the frequency using the numeric entry
keys.
Soft-key Function
ON Enable this point.
OFF Disables this point.
CURRENT FREQ Use current frequency
MEAS OPEN Starts open correction.
MEAS SHORT Starts short correction
Step 4. Press the [MEAS OPEN] soft-key. A dialog
message, -circuit the test
appear.
Step 5. Connect the test fixture to the BNC terminals.
Do not connect the DUT to the test fixture.
Step 6. Press the [OK] soft-key. The T3LCR
the open admittances (capacitance and
conductance) at this frequency point.
will
be shown on the display.
38
When the measurement has finished,
, will be displayed.
During the measurement, you can press the
[ABORT] soft-key to abort open correction.
Step 2. Connect the test fixture to the BNC terminals
and short-circuit the test terminals.
Step 3. Press the [OK] soft-key. The T3LCR
the short impedances (inductances and
resistance) at this frequency point.
mwill
be shown on the display.
When the measurement has finished,
on the
screen
the [ABORT] soft-key to abort short
correction.
Press the [Measure] key and press the [LIST SETUP]
soft-key to open the [LIST SETUP] page. The List
Measure feature can perform automatic sweep
measurements by sweeping the frequency or signal level
through a maximum 10 list points.
Before using the List Measure feature, you have to
configure the list setup.
In the [LIST SETUP] page, you can configure each of
the following list measurement controls with the cursor
placed in the corresponding field.
Select the primary and the secondary measurement here.
The [LIST MEAS] page will execute a list of up to 10
swept frequency or amplitude tests.
When [MODE] is set to SEQ and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute each test step on the list in sequence until the
last step is executed. The meter will then wait for the
trigger button to be depressed before repeating the
process.
When [MODE] is set to STEP and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute the first test step. The meter will then wait for
the trigger button to be depressed before executing the
next step. This process is repeated for each step on the
list.
Soft-key Function
SEQ Sequence measure all step by one
trigger.
STEP Measure one step by one trigger.
There are 4 available trigger modes in the [LIST MEAS]
page.
The list parameter used for the list measurement can be
the measurement frequency or the signal level [V/A].
Use the list point field to specify the list measurement
parameter.
Step 1. Press the [Measure] key.
Step 2. Press the [LIST SETUP] soft-key.
Step 3. Use the cursor key to select FREQ or VOLT
or CURR field.
Step 4. Use the soft-keys to select list measurement
The List measurement feature supports up to 10 list
points as well as measurement limit values. Each list
point can be turned on or off.
Step 1. Press the [Measure] key.
Step 2. Press the [LIST SETUP] soft-key.
Step 3. Use the cursor key to select one of the list
point field (1 through 10)
Step 4. Use the soft-keys to turn on/off the current
point.
Step 5. Enter the list point value (maybe the
frequency value or the signal level
voltage/current value). When you enter the
frequency value, the soft-key labels change to
unit labels (Hz, kHz). When you enter the
enter units. When you enter the signal level
current value, the soft-key labels change to
unit labels (uA, mA).
Step 6. Using the cursor keys, select the LMT field.
42
Step 7. Configure the limit parameter by pressing the
appropriate soft-key.
Soft-key Function
PRIMARY A Uses the primary parameter as the
limit parameter.
2NDARY B Uses the secondary parameter as the
limit parameter.
Step 8. Using the cursor keys, select the LOWER
field.
Step 9. Enter the lower limit value.
Step 10. Using the cursor keys, select the UPPER field
The [LIST MEAS] page will appear when you press the
[Measure] key and then the [LIST MEAS] soft-key.
On the [LIST MEAS] page, the list points are swept and
the measurement results are compared to the limits.
During a sweep, an asterisk mark (*) will appear on the
left side of the list point currently being measured.
This page provides the following information:
Trigger mode [TRIG].
Test mode [MODE].
Range mode [RANGE].
Measurement recording [LOG].
These conditions can be set from [MEAS DISPLAY]
page and [SETUP] page.
Select the trigger mode for LIST MEAS here; manual
trigger is generally used for LIST MEAS.
When entering the LIST MEAS function, the default
trigger mode is the manual trigger mode.
You can use the cursor button to highlight the TRIG
source field and select the other available trigger sources.
When entering LIST MEAS for the f
set Trig Mode to MAN.
Trig Mode Function
INT Internal Trigger. All ten list points are
swept continuous.
MAN
is triggered by [Trig]
key, the list points are swept one by one.
instrument is triggered by the
handler trigger pin, the list points are
swept one by one.
BUS BUS Trigger. Each time the instrument is
triggered by SCPI command, the list
points are swept one by one.
The [LIST MEAS] page will execute a list of up to 10
swept frequency or amplitude tests.
When [MODE] is set to SEQ and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute each test step on the list in sequence until the
last step is executed. The meter will then wait for the
trigger button to be depressed before repeating the
process.
When [MODE] is set to STEP and [TRIG] is set to
MAN, the [LIST MEAS] function will automatically
execute the first test step. The meter will then wait for
the trigger button to be depressed before executing the
next step. This process is repeated for each step on the
list.
There are 4 available trigger modes in the [LIST MEAS]
page.
source.
MAN Trigger by pressing the trigger button.
EXT Trigger by sending a signal to the
trigger pin on the handler interface.
BUS Trigger by the RS-232/USB port.
Displays the selected measurement range for each
measurement step; this field is here only for display, its
you need to change the
measurement range; it can be done in the
[MEAS DISPLAY] or [MEAS SETUP] function.
The T3LCR has an internal data buffer that record up to
10000 measurement readings. These readings can be
saved to an external USB drive in a .csv file format.
These readings can then be opened on a PC using
software such as Windows Excel. When the LOG
function is active, the display is fixed to the [MEAS
DISPLAY] screen. The [FUNC], [FREQ] and [LEVEL]
settings are also fixed and cannot be changed.
Step 1. Press the [LIST MEAS] soft-key.
Step 2. Use the cursor keys to select the [LOG] status
field.
Step 3. Use the [START LOG] soft-keys to start a
new log.
Step 4. After a new log has started, the measurement
readi
internal buffer. The options in the following
table are available during this recording
process:
Before saving the recorded readings in the internal buffer
Soft-key Function
SAVE & STOP Stops the recording and saves the
recorded readings in the internal
buffer to an external USB flash drive.
The internal buffer will be cleared
after the save operation.
If an external USB flash drive is not
plugged in, the meter will not stop the
ongoing recording.
STOP
buffer.
Step 5. When th
be added after the number of recorded
counts(in the [LOG] status field). The
following two operations are available in this
situation:
SAVE TO USB Saves the recorded readings in the
buffer to an external USB flash drive.
operation.
CLEAR
Clears the internal buffer.
BUFFER
to the external USB flash drive, remember to plug the
external USB drive into the USB port on the front panel.
The recorded readings will be saved in a subdirectory
named LIST DATA, whose parent directory will have
the same name as the T3LCR model being used. See the
example in below:
Example: F:\T3LCR1300\LIST DATE\LIST0000.CSV
Up to 9999 recording files can be saved, with their file
names from 0001.csv-9999.csv.
The size of the internal buffer can be adjusted in
setting field. The maximum data buffer size is 10000.
Only four measurement values will be shown in this
simplified display mode: The primary measurement
parameters, the secondary measurement parameters and
two monitored parameters such as , %, Iac or Vac,
etc.
Press the ENLARGE soft-key in
to enter the [ENLARGE DISPLAY] mode (simplified
display mode); on the contrary, press the NORMAL
softÂ[ENLARGE DISPLAY] mode and return to the regular
[MEAS DISPLAY] mode.
Two biggest readings you see in the upper part of the
[ENLARGE DISPLAY] mode are the primary and the
secondary measurement parameters while the two other
smaller readings in the lower part of the display are the
monitored parameters you chose to be displayed. The
PASS/FAIL result of the Direct Comparison function
(see 4.5.2 Direct Comparison Function) for the DUT
currently being measured will be displayed in the bottom
right corner of the enlarged display.
The Direct Comparison function allows you to see
acceptable tolerance range without the complexity of the
BIN sorting function.
The Direct Comparison function can be turned on in the
[BIN SETUP] page. When the number of bins is set to 1
in the [BIN SETUP] page, the Direct Comparison
function will be activated for the
[ENLARGE DISPLAY] mode. To activate Direct
-UP]
page. Please refer to the Total Number of Bins [# BINS]
chapter on page 61 -.
Step 1. Press the [Setup] key.
Step 2. Use the cursor key to select [MON 1] or
[MON 2] field.
Step 3. Use the soft-keys to select DC bias voltage.
Relative deviation value
r Phase radian
d Phase angle
R Resistance (=Rs)
X Reactance
G Conductance
B Sustenance
Y Absolute value of admittance
This setting decides how long the meter will wait to
commence a measurement after the trigger signal is
received. Settable range: 0ms to 60s.
The automatic level control (ALC) feature adjusts the
voltage across the DUT or the current through the DUT
to match the voltage/current level setting. Using this
feature, you can try to ensure a constant signal level
(voltage or current) is applied to the DUT.
Press [Setup] key and press the [BIN SETUP] soft-key to
open [BIN SETUP] page.
This page allows you to configure the T3LCR Series
built-in comparator. The built-in comparator can sort
DUTs into a maximum of
and OUT) using up to nine sets of primary parameter
limits along with one set of secondary parameter limits.
In addition, DUTs that fall within the primary parameter
limits but are out of the secondary parameter limits can
be sorted into an auxiliary bin.
To take full advantage of the comparator, T3LCR Series
is equipped with a handler interface for use in
conjunction with the comparator. All 10 bin signals can
output to your PLC via the handler interface.
In the [BIN SETUP] page, you can configure each of the
following controls with the cursor placed in the
corresponding field.
The T3LCR Series simultaneously measures four
components of the complex impedance (parameters) in a
measurement cycle. These include primary parameter,
secondary parameter and two monitor parameters
refer to the 4.1 [MEAS DISPLAY] Page section on 23 f
details.
T3LCR Series built-in comparator can sort DUTs into a
maximum of 10 bins (BIN1 to BIN9 and OUT), using a
maximum of nine pairs of primary parameter limits and
one pair of secondary parameter limits.
A DUT can be sorted into the auxiliary BIN (AUX)
when the primary parameter is within the limits but the
secondary parameter is not.
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [COMP] field.
Step 4. Use the soft-keys to turn on/off the
comparator.
Soft-key Function
OFF Turns OFF the COMP function
ON Turns ON the COMP function
After AUX is turned on, DUTs that do not fall within
the primary parameter limit values are sorted as OUT. In
addition, DUTs that fall within the primary parameter
limits but are out of the secondary parameter limits are
sorted into the auxiliary (AUX) bin.
Procedure for turning ON/OFF the auxiliary bin [AUX]
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [AUX] field.
Step 4. Use the soft-keys to turn on/off the auxiliary
bin.
Soft-key Function
ON Turns ON the auxiliary bin
OFF Turns OFF the auxiliary bin
You can specify the primary parameter limit values in
one of the following 3 ways:
Tolerance mode [%][ ]
In tolerance mode, the comparison limit values are
based on the deviations from the specified nominal
value, which can be specified in the [NOM] field.
You configure the tolerance mode limit values as a
deviation percentage (%) or absolute ( ) parameter
value.
Sequential mode [SEQ]
In sequential mode, the comparison limit values are
based on the absolute value of the measurement.
When you configure these limit values, you have to
first define the minimum value and then the
maximum value.
Nominal Value [NOM]
Sets the nominal value (reference value) for the
and PER compare modes. Use the number keys to
enter the value and the unit.
In Sequential mode, the comparison limit values are
based on the absolute value of the measurement. The
nominal value does not need to participate in operation.
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [MODE] field.
Step 4. Use the soft-keys to select comparator mode.
ou must configure the nominal value when you use
tolerance mode as the limit mode for the primary
parameter.
In sequential mode the nominal value does not affect
sorting. In sequential mode
the nominal value.
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [NOM] field.
Step 4. Enter the nominal value using the entry keys.
Use the soft-key to select the unit.
The beep feature behaves differently depending on
which beep mode is in effect.
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [BEEP] field.
Step 4. Use the soft-keys to set beep feature.
result is OK
FAIL Beep when the comparator sorting
result is NG
T3LCR series specify nine bins (1-BINS to 9-BINS).
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [#-BINS] field.
Step 4. Use the soft-keys to select beep tone.
1-BINS Set to 1 bin
2-BINS Set to 2 bins
3-BINS Set to 3 bins
4-BINS Set to 4 bins
5-BINS Set to 5 bins
6-BINS Set to 6 bins
7-BINS Set to 7 bins
8-BINS Set to 8 bins
9-BINS Set to 9 bins
T3LCR Series built-in comparator can sort DUTs into a
maximum 10 levels (Bin1 through Bin9 and OUT) using
up to nine sets of primary parameter limits along with
one set of secondary parameter limits.
Step 1. Press the [Setup] key.
Step 2. Press the [BIN SETUP] soft-key.
Step 3. Use the cursor key to select [BIN 1 LOWER]
field
Step 4. Enter the limit value using the entry keys.
When in the ABS and SEQ mode, please enter
the unit using the soft-keys. When in the PER
mode, the unit is %.
Step 5. Repeat Step 4 using the cursor keys until you
have entered all limit values.
Step 6. Use the cursor key to select [2nd LOWER]
field and enter the lower limit value for the
secondary parameter.
Step 7. Select [2nd UPPER] field to enter the upper
limit value.
the [Setup] key and the [BIN MEAS] soft-key is
pressed, the [BIN MEAS] page appears. The bin sorting
results are displayed in larger characters while the
measurement results are shown in normal characters.
SPEED: These conditions can be set from
[MEAS DISPLAY] page.
Comparator function ON/OFF [COMP].
Auxiliary bin ON/OFF [AUX].
T3LCR built-in comparator can sort DUTs into a
maximum of 10 bins (BIN1 to BIN9 and OUT) using a
maximum of nine pairs of primary parameter limits and
one pair of secondary parameter limits.
A DUT which has the primary parameter within the
limits, but secondary parameter is not, can be sorted into
an auxiliary BIN (AUX).
See Figure 5-3 on page 57 for the Comparator Workflow.
Step 1. Press the [Setup] key.
Step 2. Press the [BIN MEAS] soft-key.
Step 3. Use the cursor key to select [COMP] field.
Step 4. Use the soft-keys to turn on/off the
OFF Turns OFF the COMP function
ON Turns ON the COMP function
After AUX is turned on, DUTs that do not fall within
the primary parameter limit values are sorted as OUT. In
addition, DUTs that fall within the primary parameter
limits but out of the secondary parameter limits are
sorted into the auxiliary (AUX) bin.
Step 1. Press the [Setup] key.
Step 2. Press the [BIN MEAS] soft-key.
Step 3. Use the cursor key to select [AUX] field.
Step 4. Use the soft-keys to turn on/off the auxiliary
The number of DUTs sorted into each bin is counted
while the unit sorts the DUTs into the appropriate bins
using the comparator function.
When the maximum count of 99999999 is reached, the
counting operation stops and the overflow message
--------.
Step 1. Press the [Setup] key.
Step 2. Press the [BIN COUNT] soft-key.
Step 3. Use the cursor key to select [COUNT] field.
Step 4. Use the soft-keys to set counter function.
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select date field.
Step 4. Use the soft-keys to edit date.
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select time field
Switch user mode to administrator mode, you must input
CHANGE
Input password
(less than 9 numbers
).
Soft-key Function
HOUR INCR+ Increases the hour in steps of 1.
HOUR DECR- Decreases the hour in steps of 1.
MINUTE INCR+ Increases the minute in steps of 1.
MINUTE INCR- Decreases the minute in steps
SECOND DECR+ Increases the second in steps of 1.
SECOND DECR-
The T3LCR Series has two accounts, administrator and
user.
Administrator: All functions can be configured by the
administrator except [SYSTEM SERVICE] page.
User: All functions can be configured by user except the
[SYSTEM SERVICE] page and the [FILE] page.
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select ACCOUNT field.
Step 4. Use the soft-keys to change account.
Soft-key Function
a correct password.
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select
[ADMINISTRATOR] field. When the
ACCOUNT field is [USER], you should
change to [ADMINISTRATOR].
Step 4. Use the soft-keys to change the password or
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the [KEY BEEP]
field.
Step 4. Use the soft-keys to set the beep settings.
Before you can control the T3LCR Series remotely via
the built-in RS-232/USB controller, you have to
configure the RS-232 RS-232/USB baud rate.
The T3LCR built-in RS-232/USB interface uses the
SCPI language.
RS-232/USB configuration is as follows:
Data bits: 8-bit
Stop bits: 1-bit
Parity: none
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the [BAUD] field.
Step 4. Use the soft-keys to select the baud rate.
Feed character by default. (LF, Decimal 10, Hex 0x0A, or
1200 Sets the baud rate to 1200.
9600 Sets the baud rate to 9600.
38400 Sets the baud rate to 38400.
57600 Sets the baud rate to 57600.
115200 Sets the baud rate to 115200.
Selection the interface for PC/PLC remote control.
For other setting items about remote control mode,
89.
After the device detects the signal change on RS-232 or
USB interface, the device will immediately communicate
with the host at the setting serial transmission rate, and
the keyboard will be locked as well.
In order to be able to communicate correctly, please
confirm whether the setting serial transmission rate and
able to communicate with host computer correctly.
The command terminator EOL uses the ASCII Line
\
The EOL of the command and the EOL of the
command return data are the same.
The terminator (EOL) can change the three types:
1. LF (Line feed, Hex 0x0A).
2. CR (Carriage Return, Hex 0x0D).
3. CR+LF.
The T3LCR PC software only supports LF types.
LF Set up the EOL to LF(0Ah).
CR Set up the EOL to CR(0Dh).
CR+LF Set up the EOL to CR+LF(0D 0Ah).
If handshaking is turned on, the meter will return the
commands it received from the PC first before returning
the result string of the command.
Example: PC sends idn?
The meter returns:
idn? T3LCR1300 RevC1.0
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the
[HAND SHAKE] field.
Step 4. Use the soft-keys to set the handshake setting.
Soft-key Function
If the error code setting is set to on, the meter will return
error codes if the wrong command or an invalid
command is received to help you to debug your control
program.
If the Result setting it set to Auto, the meter will
automatically send out the measurement results
a test is finished; this setting is convenient especially
when the meter is working with a sorting machine. The
meter will start a test after receiving the trigger signal and
then returns the test result to the sorting machine
without the need to receive
either the sorting machine or the control PC.
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the [RESULT]
field.
Step 4. Use the soft-keys to set.
Soft-key Function
FETCH The meter will not send out the test
result after the test.
after the test.
The data buffer setting sets the maximum number of
record counts that the internal buffers can store; Settable
range 1-10000.
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the
[DATA BUFFER] field.
Step 4. Enter the number of sets using the entry keys
To reset setting values (MEAS SETUP) and offset values
ON
To reset to
default
factory settings
OFF
No reset to factory default
settings
(OPEN SHORT) to factory default, use the DEFAULT
SET setting. The device can be quickly restored to its
factory settings.
Default settings:
(MEAS SETUP):
FUNC: Cp-D
FREQ: 1kHz
LEVEL: 1V
RANGE: AUTO
TRIG: INT
SPEED: SLOW
SRC RES: 100
BIAS: OFF
AUTO LCZ: OFF
DELAY: 0ms
ALC: OFF
AVG: 1
MON1: OFF
MON2: OFF
(OPEN SHORT):
OPEN TEST: ON
SHORT TEST: ON
SPOT: OFF
Step 1. Press the [Measure] or [Setup] key.
Step 2. Press the [SYSTEM] bottom soft-key.
Step 3. Use the cursor key to select the
[DEFAULT SET] field.
Step 4. Use the soft-keys to set.
The Media field is used to select the media source from
an ext
drive. A maximum of up to 10 files can be accessed from
either source.
Step 1. Press the [Setup] key.
Step 2. Press the [FILE] bottom soft-key.
Step 3. Use the cursor key to select the [MEDIA] field.
Step 4. Use the soft-keys to select the source.
MEMORY internal memory.
flash drive.
You can recall file0 or the current file when the
instrument starts up.
Step 1. Press the [Setup] key.
Step 2. Press the [FILE] bottom soft-key.
Step 3. Use the cursor key to select the [AUTO
RECALL] field.
Step 4. Use the soft-keys to select the recall option.
component handler or system controller. This means that
This chapter provides information about the T3 built-in handler
interface. It includes:
Pin Assignment
Circuit Diagram
Timing Chart
The T3LCR built-in handler interface outputs
signals that indicate the end of a measurement cycle, the
result of bin sorting by the comparator. In addition, the
instrument accepts input of external trigger. You can use
these signals to easily integrate the T3LCR Series with a
you can fully automate such tasks as component
inspection, component sorting, and processing of quality
management data for higher manufacturing efficiency.
Active low.
The sorting result is within bin2.
Active low.
The sorting result is within bin3.
Active low.
The sorting result is within bin4.
Active low.
The sorting result is within bin5.
Active low.
Active low.
Active low.
The sorting result is within bin8.
Active low.
The sorting result is within bin9.
Interface power requirements: +12.4V-36VDC, 0.2A(Min).
Output circuit: Built-in pull-up resistors are internally connected to the
collector pin of the output transistors. The output pins are isolated by
photocouplers.
This chapter covers basic measurement procedures as well as basic L, C, and
R measurement theory. It also offers various measurement hints. After the
descriptions of basic measurement procedures, practical measurement
examples are shown using T3LCR Series.
The following flow chart shows the basic procedures
used to measure the impedance of capacitors, inductors,
resistors, and other components. Follow the procedures
to perform impedance measurements while referring to
the items noted to the right side of each step.
This chapter provides the following information to remotely control the
T3LCR Series via the RS-232 or USB:
RS-232 Interface
USB Interface
Install USB Driver
The device uses the RS-232 interface or USB interface to communicate with
the computer to complete all functions. With standard SCPI
commands, users can easily create various acquisition system which are
suitable for themselves.
You can connect a controller (i.e. PC and PLC) to the
RS-232 interface using a TELEDYNE RS-232 DB-9
cable. The serial port uses the transmit (TXD), receive
(RXD) and signal ground (GND) lines of the RS-232
standard. It does not use the hardware handshaking lines
CTS and RTS.
also uses these settings. The RS-232 interface transfers
data using:
Data bits: 8 bits
Stop bits: 1 bit
Parity: none
Baud rate: 1200-115200bps
You can connect a controller (i.e. PC and PLC) to the
USB interface using a type A-B USB cable.
This virtual port can perform the same functions as
RS232 and use the same settings as the RS232 port. It
supports USB2.0 and below version.
The USB device port on the rea
control. The USB port is configured as CDC interface.
When configured to CDC, the USB port on the T3LCR
series will appear as a virtual COM port to a connected
PC. Any terminal program that can communicate via a
serial port can be used for remote control. Before the
T3LCR series can be used for remote control using the
CDC USB class, install the appropriate CDC USB driver.
Make sure the controller you connect to T3
also uses these settings. The USB interface transfers
data using:
PC connector: Type A, host
T3LCR series connector: Rear panel Type B, slave
Speed: 1.1/2.0 (full speed/high speed)
USB Class: CDC (Communications device class)
Hardware flow control: Off
Data bits: 8-bit
Stop bits: 1-bit
Parity: none
The EOI line is asserted by New Line or ASCII Line Feed character.
(Decimal 10, Hex 0x0A, \
The following conventions and definitions are used in this chapter to describe
RS-232 operation.
< > Angular brackets enclose words or characters that are used to symbolize a
program code parameter or an RS-232 command.
[ ] A square bracket indicates that the enclosed items are optional.
\n Command Terminator
The T3LCR commands are divided into two types: Common commands and
SCPI commands.
The common commands are defined in IEEE std. 488.2-1987, and these
commands are common for all devices. The SCPI commands are used to
control all of the T3LCR's functions.
The SCPI commands are tree structured, three levels deep. The highest level
commands are called the subsystem commands in this manual. So the lower
level commands are legal only when the subsystem commands have been
selected.
A colon (:) is used to separate the higher level commands and the lower level
commands.
Semicolon (;) A semicolon does not change the current path but separates two
commands in the same message.
(b) If the 4th character is not a vowel, the short form
Every command and character parameter has at least
two forms, a short form and a long form. In some
cases they will be the same. The short form is
obtained using the following rules.
A) If the long form has four characters or less, the
long form and short form are the same.
B) If the long form has more than 4 characters:
(a) If the 4th character is a vowel, the short form is
the first 3 characters of the long form.
For example:
comparator abbreviated to comp
current abbreviated to curr
range abbreviated to rang
is the first 4 characters.
For example:
resistance abbreviated to res
volume abbreviated to vol
If the long form mnemonic is defined as a phrase
rather than a single word, then the long form
mnemonic is the first character of the first word(s)
followed by the entire last word. The above rules,
when the long form mnemonic is a single word, are
then applied to the resulting long form mnemonic to
obtain the short form.
For example:
PercentTolerance abbreviated to ptol
The commands consist of a command header and parameters. (See the
following.)
Example:
comp:nom 100.0e3
Header Parameter
Headers can be of the long form or the short form. The
long form allows easier understanding of the program
code and the short form allows more efficient use of the
computer.
Parameters may be of two types as follows.
(A) Character Data and String Data Character data
consists of ASCII characters. The abbreviation rules are
the same as the rules for command headers.
(B) Numeric Data
(a) NR1 integer: For example, 1,+123,-123
(b) NR2 fix float: For example, 1.23,+1.23,-1.23
(c) NR3 floating point: For example, 1.23e3,
5.67e-3, 123k, 1.23M, 2.34G,
The available range for numeric data is 9.9E37. When
numeric data is used as a parameter, the suffix multiplier
mnemonics and suffix units (The suffix multiplier must
be used with the suffix unit.) can be used for some
commands as follows.
1E15 (PETA) PE
1E9 (GIGA) G
1E6 (MEGA) MA
1E3 (KILO) K
1E-3 (MILLI) M