Teledyne LeCroy T3LCR1300, T3LCR1002, T3LCR1100 User manual

Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Warning Dangerous
Please contact the TELEDYNE service center representative for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard for operators.
The instrument is operating abnormally if:
The instrument emits abnormal noise, smell, smoke, or a spark-like light
during the operation.
The instrument generates high temperature or electrical shock during
operation.
The power cable, plug, or receptacle on the instrument is damaged.
Foreign substances or liquid has fallen into the instrument.
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment. In addition it violates safety standards of design, manufacture, and intended use of the instrument.
TELEDYNE assumes no liability for the customer's failure to comply with these requirements.
To avoid electric shock hazards, the instrument chassis and cabinet must be connected to a safety earth ground by the supplied power cable with earth blade.
Do not operate the instrument in the presence of inflammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
i
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Safety Summary
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with the power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable is removed. To avoid injuries, always disconnect the power and discharge the circuits before touching them.
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
Because of the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a TELEDYNE authorized service center for service and repair to ensure that safety features are maintained.
ii
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
1.1 Incoming Inspection ........................................................................ 10
1.2 Environmental Requirements ......................................................... 11
1.3 Cleaning ............................................................................................ 11
1.4 How to Remove the Handle ........................................................... 12
2.1 Introduction ..................................................................................... 13
2.2 Main Specifications and Features ................................................... 14
2.2.1 Test Function ............................................................................ 14
2.2.2 Equivalent Circuit ..................................................................... 14
2.2.3 Ranging ...................................................................................... 14
2.2.4 Measurement Speed ................................................................. 14
2.2.5 Trigger Mode ............................................................................. 15
2.2.6 Basic Accuracy .......................................................................... 15
2.2.7 Display Range ........................................................................... 15
2.3 Test Signal ......................................................................................... 15
2.3.1 Test Signal Frequency .............................................................. 15
2.3.2 Test Signal Level ....................................................................... 16
2.3.3 Output impedance .................................................................... 16
2.4 Main Functions................................................................................. 16
2.4.1 Correction Function ................................................................ 16
2.4.2 Comparator Function (Sorting) .............................................. 16
2.4.3 List Sweep ................................................................................. 17
2.5 Measurement Assistance Functions ............................................... 17
2.5.1 Files ............................................................................................ 17
2.5.2 Key Lock ................................................................................... 17
2.5.3 RS 232/USB ............................................................................. 17
2.6 Options ............................................................................................. 17
2.6.1 USB host port ........................................................................... 17
2.6.2 DC Bias Voltage ........................................................................ 19
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Table of Contents
3.1 Front panel ....................................................................................... 20
3.2 Rear Panel ......................................................................................... 21
3.3 Power On/Off ................................................................................. 21
3.3.1 Power On .................................................................................. 21
3.3.2 Power Off ................................................................................. 21
3.4 Warm up Time ................................................................................. 22
3.5 Connect to the Device Under Test (DUT) ................................... 22
4.1 [MEAS DISPLAY] Page ................................................................. 23
4.1.1 Measurement Function [FUNC] ............................................ 24
4.1.2 Impedance Range [RANGE] ................................................. 26
4.1.3 Test Frequency [FREQ] .......................................................... 28
4.1.4 Trigger Mode [TRIG] .............................................................. 29
4.1.5 Test Signal Voltage/Current Level [LEVEL] ...................... 30
4.1.6 Measurement Speed [SPEED] ............................................... 31
4.1.7 Measurement log [LOG] ......................................................... 32
4.2 [OPEN/SHORT] Page ................................................................... 34
4.2.1 Open Correction [OPEN] ...................................................... 35
4.2.2 Short Correction [SHORT] .................................................... 36
4.2.3 SPOT Correction ..................................................................... 37
4.3 [LIST SETUP] Page ........................................................................ 39
4.3.1 Measurement Function [FUNC] ............................................ 40
4.3.2 Test Mode [MODE] ................................................................ 40
4.3.3 List Measurement Parameters ................................................ 41
4.3.4 List Point and Limit Modes .................................................... 42
4.4 [LIST MEAS] Page .......................................................................... 43
4.4.1 Trigger Mode [TRIG] .............................................................. 43
4.4.2 Test Mode [MODE] ................................................................ 44
4.4.3 Range Mode [RANGE] ........................................................... 45
4.4.4 Measurement log [LOG]......................................................... 45
4.5 [ENLARGE DISPLAY ] Page ...................................................... 47
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
4.5.1 Enlarge Display ........................................................................ 47
4.5.2 Direct Comparison Function ................................................. 48
5.1 [MEAS SETUP] Page ...................................................................... 49
5.1.1 Source Output Impedance [SRC RES] ................................. 50
5.1.2 Averaging Factor [AVG] ......................................................... 51
5.1.3 DC Bias Voltage [BIAS] .......................................................... 51
5.1.4 Auto LCZ Function [AUTO LCZ] ....................................... 52
5.1.5 Monitor 1 and Monitor 2 [MON 1][MON 2] ...................... 52
5.1.6 Measurement Delay [DELAY] ............................................... 53
5.1.7 AUTO LEVEL CONTROL [ALC] ...................................... 53
5.1.8 Nominal Value [NOMINAL] ................................................. 54
5.2 [BIN SETUP] Page .......................................................................... 55
5.2.1 Measurement Function [FUNC] ............................................ 56
5.2.2 Comparator Function ON/OFF ........................................... 56
5.2.3 Auxiliary Bin [AUX] ................................................................ 58
5.2.4 Primary Parameter Comparator limit mode [MODE] ........ 58
5.2.5 Nominal value for tolerance mode ........................................ 61
5.2.6 Beep feature .............................................................................. 61
5.2.7 Total Number of Bins [# BINS] ........................................... 61
5.2.8 Lower and Upper Limits ......................................................... 62
5.3 [BIN MEAS] Page ........................................................................... 62
5.3.1 Comparator Function ON/OFF ........................................... 63
5.3.2 Auxiliary Bin [AUX] ................................................................ 64
5.4 [BIN COUNT] Page ....................................................................... 64
5.4.1 Counter Function [COUNT] ................................................. 65
6.1 [SYSTEM CONFIG] Page ............................................................. 66
6.1.1 Setting the system date and time ............................................ 67
6.1.2 Account Setting ........................................................................ 68
6.1.3 KEY BEEP Setting ................................................................. 69
6.1.4 Baud rate ................................................................................... 69
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Table of Contents
6.1.5 Remote control interface......................................................... 70
6.1.6 TERMINATOR....................................................................... 70
6.1.7 Hand Shake ............................................................................... 71
6.1.8 Error Code ................................................................................ 71
6.1.9 Result ......................................................................................... 73
6.1.10 DATA BUFFER ...................................................................... 73
6.1.11 DEFAULT SET ...................................................................... 74
6.2 [SYSTEM INFO] Page ................................................................... 75
7.1 [FILE] Page ....................................................................................... 76
7.1.1 [MEDIA] ................................................................................... 77
7.1.2 Recall a File at Startup [AUTO RECALL] ........................... 77
7.1.3 Auto save data to last file [AUTO SAVE] ............................ 78
7.1.4 File operation ............................................................................ 78
8.1 Pin Assignment ................................................................................ 80
8.2 Connection ....................................................................................... 82
8.3 Timing Chart for Handler Interface .............................................. 83
9.1 Basic Measurement Procedure........................................................ 84
9.2 Example ............................................................................................ 85
10.1 RS 232 Interface .............................................................................. 89
10.2 USB Interface ................................................................................... 90
10.3 Install USB Driver ........................................................................... 92
11.1 Terminator ........................................................................................ 95
11.2 Notation Conventions and Definitions ........................................ 95
11.3 Command Structure ........................................................................ 95
11.4 Header and Parameters ................................................................... 98
11.5 Command Reference ....................................................................... 99
11.6 DISPlay Subsystem ........................................................................ 100
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
11.6.1 DISP:LINE ............................................................................. 100
11.6.2 DISP:PAGE ........................................................................... 100
11.7 FUNCtion Subsystem ................................................................... 102
11.7.1 FUNCtion ............................................................................... 102
11.7.2 FUNCtion:IMPedance:AUTO ............................................. 103
11.7.3 FUNCtion:IMPedance:RANGe ........................................... 103
11.7.4 FUNCtion:DCR:RANGe ..................................................... 103
11.7.5 FUNCtion:RANGe:AUTO .................................................. 104
11.7.6 FUNCtion:MONitor 1 /2..................................................... 104
11.8 LEVel Subsystem ........................................................................... 106
11.8.1 LEVel:VOLTage (=VOLTage[:LEVel]) ............................. 106
11.8.2 LEVel:CURRent (=CURRent[:LEVel]) .............................. 107
11.8.3 LEVel:SRESistance (= VOLTage:SRESistance) ............... 107
11.8.4 LEVel:ALC (=AMPlitude:ALC) .......................................... 108
11.8.5 LEVel:MODe? ....................................................................... 108
11.9 APERture Subsystem .................................................................... 109
11.9.1 APERture:RATE? .................................................................. 109
11.9.2 APERture:AVG? .................................................................... 109
11.10 FETCh Subsystem ......................................................................... 110
11.10.1 FETCh? ........................................................................... 110
11.10.2 FETCh:IMPedance? ...................................................... 111
11.10.3 FETCh:MAIN? .............................................................. 111
11.10.4 FETCh:MONitor1? /2? ................................................ 111
11.10.5 FETCh:MONitor? ......................................................... 111
11.10.6 FETCh:LIST? ................................................................. 112
11.11 COMParator Subsystem ............................................................... 113
11.11.1 COMParator:STATe ...................................................... 113
11.11.2 COMParator:MODE ..................................................... 114
11.11.3 COMParator:AUX ......................................................... 114
11.11.4 COMParator:BINS ........................................................ 114
11.11.5 COMParator:TOLerance:NOMinal............................. 115
11.11.6 COMParator:TOLerance:BIN ...................................... 115
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Table of Contents
11.11.7 COMParator:SLIM ........................................................ 115
11.11.8 COMParator:BEEP ....................................................... 116
11.11.9 COMParator:OPEN ...................................................... 116
11.12 LIST Subsystem ............................................................................. 117
11.12.1 LIST:PARAmeter ........................................................... 117
11.12.2 LIST:STAT ..................................................................... 117
11.12.3 LIST:BAND ................................................................... 118
11.13 CORRection Subsystem ............................................................... 119
11.13.1 CORRection:OPEN ...................................................... 119
11.13.2 CORRection:OPEN:STATe ......................................... 120
11.13.3 CORRection:OPEN:LCR ............................................. 120
11.13.4 CORRection:OPEN:DCR ............................................ 120
11.13.5 CORRection:SHORt ..................................................... 120
11.13.6 CORRection:SHORt:STATe ........................................ 121
11.13.7 CORRection:SHORt:LCR ............................................ 121
11.13.8 CORRection:SHORt:DCR ........................................... 121
11.13.9 CORRection:SPOT:STATe .......................................... 121
11.13.10 CORRection:SPOT:FREQuency ................................. 122
11.13.11 CORRection:SPOT:OPEN .......................................... 122
11.13.12 CORRection:SPOT:SHORt .......................................... 122
11.14 TRIGger Subsystem ...................................................................... 123
11.14.1 TRIGger[:IMMediate] .................................................... 123
11.14.2 TRIGger:SOURce .......................................................... 123
11.14.3 TRIGger:DELAY .......................................................... 123
11.15 BIAS Subsystem ............................................................................. 124
11.16 FILE Subsystem ............................................................................ 125
11.16.1 FILE? ............................................................................... 125
11.16.2 FILE:SAVE .................................................................... 125
11.16.3 FILE:LOAD ................................................................... 125
11.16.4 FILE:DELete .................................................................. 126
11.17 ERRor Subsystem .......................................................................... 126
11.17.1 ERRor? ............................................................................ 126
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
11.18 SYSTEM Subsystem ..................................................................... 126
11.18.1 SYSTem:SHAKehand.................................................... 126
11.18.2 SYSTem:CODE ............................................................. 126
11.18.3 SYSTem:KEYLock ........................................................ 127
11.18.4 SYSTem:RESult.............................................................. 127
11.19 Common Commands .................................................................... 127
11.19.1 *IDN? .............................................................................. 127
11.19.2 *TRG ............................................................................... 127
11.19.3 *SAV ................................................................................ 128
11.19.4 *RCL ................................................................................ 128
12.1 General Specification ..................................................................... 129
12.2 Dimensions ..................................................................................... 133
13.1 Accuracy .......................................................................................... 135
13.1.1 L, C, R |Z| Measurement Accuracy ................................... 135
13.1.2 Accuracy for D ....................................................................... 135
13.1.3 Accuracy for Q ....................................................................... 136
13.1.4 Accuracy for ........................................................................ 136
13.1.5 Accuracy for Rp ..................................................................... 136
13.1.6 Accuracy for Rs ...................................................................... 137
13.2 Correction factors that affect measurement accuracy ............... 137
Certifications ................................................................................................ 141
EMC Compliance .................................................................................. 141
Safety Compliance ................................................................................. 143
Environmental Compliance ................................................................. 143
RESTRICTION OF HAZARDOUS SUBSTANCES (RoHS) ..... 144
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
UNPACKING AND PREPERATION
After you receive the instrument,
This chapter describes how to set up and start the T3LCR Series LCR Meter.
Incoming Inspection
Environmental Requirements
Cleaning
How to Remove the Handle
checks during unpacking according to the following procedure:
Make sure that the packing box or shock-absorbing material used to package the instrument has not been damaged.
Check that all packaged items supplied with the meter have been provided as specified: LCR Test Lead Power Cord D-sub
perform the following
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
sales or
in case
23ºC±5ºC (<1ºC deviation from the
Altitude:
LCR Series
If an abnormality is detected, contact the company and transport the meter to your nearest TELEDYNE service office. Save the packing box, shock-absorbing material, and packaged items as you received them you need to mail the instrument to an authorized TELEDYNE distributor or service center.
Set up the T3LCR Series where the following environmental requirements are satisfied.
Operating Environment
Ensure that the operating environment meets the following requirements:
Temperature:
0ºC to 50ºC
Temperature range at calibration:
temperature when performing calibration)
Humidity: <70% at wet bulb temperature
-condensation)
0 to 2000m
To prevent electrical shock, disconnect the T3 power cable from the receptacle before cleaning. Use a dry cloth or a cloth slightly dipped in water to clean the casing. Do not attempt to clean the T3LCR Series internally.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
UNPACKING AND PREPERATION
A handle kit is attached to the T3LCR Series:
1 Retracted
3 Carrying Position
2 Extended
Remove Handle (Lift the handle perpendicular to the unit while pulling it in the direction of 1.)
12
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
components, materials, and semiconductor devices over a
wide range of frequencies (10 Hz to 300 kHz) and test
With its built
This chapter contains general information about T3LCR Series .The information is organized as follows.
Introduction
Main Specifications
Feature overview
Thank you for purchasing an T3LCR Series LCR meter.
The TELEDYNE T3LCR Series is a general-purpose LCR meter for incoming inspection of components, quality control, and laboratory use.
The T3LCR Series is used for evaluating LCR
signal levels (10.00mV to 2.00V & 100.0uA to 20.00mA).
-in comparator, the T3LCR Series can output comparison/decision results for sorting components into a maximum of ten bins. Furthermore, by using the handler interface, the T3LCR Series can be easily combined with a component handler, and a system controller to fully automate component testing, sorting, and quality-control data processing.
The list measure function permits entry of up to 10 frequencies or test signal levels points to be automatically measured.
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Overview
Lp
)
2
Cp
)
2
Auto, Hold and Nominal range. Total 9 Ranges.
About Nominal ranges: The
Y
specify the averaging factor within the range
Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-Rp, Lp-Q, Ls-Rs, Ls-Q, Rs-Q, Rp-Q, R-X, DCR, Z- - -D, Z-Q
Circuit Dissipation
Conversion
Factor
L
C
Rp
Ls
Rp
Cs
Rs
Rs
D=2 FLp/Rp=1/Q
D=Rs/2 FLs=1/Q
D=1/2 FCpRp=1/Q
D=2 FCsRs=1/Q
Ls=Lp/(1+D2) Rs=RpD2/(1+D
Lp=(1+D2)Ls Rp=(1+D2)Rs/D
Cs=(1+D2)Cp Rs=RpD2/(1+D
Cp=Cs/(1+D2) Rp=Rs(1+D2)/D
Q=Xs/Rs, D=Rs/Xs,
T3LCR Series will automatically select the best range according to the nominal value.
2
2
Fast: 40 readings per second; 25ms/per reading*1
Medium: 10 readings per second; 100ms/per reading*1
Slow: 3 readings per second; 333ms/per reading*1
ou can also
of 1 to 256 in steps of 1 to improve performance.
14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Includes Internal, Manual, External and Bus Trigger.
0.05%@SLOW/MED
0.1%@FAST
Parameter Display Range
L 0.00001uH-9999.99H
C 0.00001pF-9999.99mF
R, X, |Z| -
G, B, |Y| 0.01nS-999.999S
D 0.00001-9.99999
Q 0.00001-99999.9
-179.999°-179.999°
-3.14159-3.14159
DCR -
-999999%-999999%
T3LCR1300: 10Hz-300kHz
T3LCR1100: 10Hz-100kHz
T3LCR1002: 10Hz-2kHz
Frequency Accuracy: ±0.01%
15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Overview
AC:
AUX,
Absolute
10.00mV-2.00V (±10%) CV mode:10.00mV-2.00V(±6%)
100.0uA-20.00mA (±10%) CC mode:100.0uA-20.00mA(±6%) @2VMax
DCR: +2V, 0.066A(Max.), Output impedance fixed
30
30 , 50 and 100
Eliminates measurement errors due to stray parasitic impedance in the test fixtures.
The primary parameter can be sorted into BIN1-BIN9,
OUT and HI/IN/LO for each of the primary
measurement parameters.
The sequential mode or tolerance mode can be selected as the sorting mode.
value, deviation value, and % deviation value
can be used for setup.
Countable from 0 to 999999
16
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
You can select from:
Judge with the first sweep parameter/Judge with the
There is a maximum of 10 points.
Test frequency, test voltage, test current.
The comparator function enables you to set one pair of lower and upper limits for each measurement point.
second parameter/Not used for each pair of limits.
The front panel keys can be locked.
Complies with SCPI.
Up to 10 setup conditions can be written to/read from the built-in non-volatile memory.
Universal serial bus jack, type-A (4 contact positions, contact 1 is on your left); female; for connection to USB memory devices only.
USB disk type: Flash disk only.
Format: FAT / FAT32 / exFAT.
Max memory size: 128GB.
17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Overview
After a USB disk
USB disk by depressing the Enter key. If
has been inserted into the USB host
port of the LCR meter and if the meter detects that the
Ready. Press <Enter> to save screen. on the lower part of the LCD screen.
If the keypad is not needed to enter a numeric value, a
and saved to the the screen shot is saved successful be shown in the message area at the bottom of the screen. The location and name of the screenshot will also be shown, for example: F:\T3LCR1300\Screen\ScreenShot001.bmp.
18
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Accuracy:
0V-±2.5V
±0.5%+0.005V
19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Startup
LCR
2
3
This chapter describes names and functions of the front panel, rear panel, and screen display and provides the basic procedures for operating the T3 Series.
Front panel summary
Rear panel summary
Power On/Off
Connect to Device under test
2
1
No.
1
4
5
6
7 Setup Key
8
12
Description
LCD Display
Soft-key
Number Key
Cursor Key
Power Switch
Measure key
BNC Terminal
11
53
4
6 7
8
910
20
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
will
, the POWER LED will flash twice
9
10
11 USB Disk Port (USB-Host)
12
Trigger Key
ESC Key
System Soft-key
No. 1 2 3 USB Interface(Type B) 4 5
Description Power Cable Receptacle(Outlet) Frame Terminal
RS-232 Interface Handler Interface
Press the power key for at least 1 second. Release power key when the POWER LED turns green.
Press the power key at least 1 second. T3LCR Series
shut down when you release the power key after the
POWERLED turns red.
If the AUTOSAVE option of the File Operation menu is set to ON (see page 76) and the AUTOSAVE function will be executed during the power-off sequence.
21
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Startup
measurements and avoids mutual inductance, interference
T3LCR Series is ready to be used as soon as the power­up sequence has completed. However, to achieve the specification accuracy, first warm up the instrument for 30 minutes.
The T3LCR Series uses a four terminal measurement configuration that provides easy, accurate and stable
from measurement signals, noise and other factors inherent with other types of connections.
L Force L Sense H Sense H Force
- DUT +
+
-
22
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
When the
This section includes the following information:
MEAS DISPLAY page
OPEN/SHORT page
LIST SETUP page
LIST MEAS page
[Measure] key is pressed, the [MEAS DISPLAY] page appears. The following measurement controls can be set.
FUNC Measurement Function
RANGE Impedance range
FREQ Test Frequency
TRIG Trigger Mode
LEVEL Test Signal Level
SPEED Measurement Speed
23
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
a
Cs-Rs Cs-D Cp-Rp Cp-D
Lp-Rp Lp-Q Ls-Rs Ls-Q Rs-Q Rp-Q R-X DCR
Z-r Z-d Z-D Z-Q
Z D Q
Vac Iac
r d R X
The T3LCR Series simultaneously measures four components of complex impedance (parameters) in a measurement cycle. These include a primary parameter, secondary parameter and two monitor parameters.
G B Y
24
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Cs Capacitance value measured using the
Capacitance value measured using the
Rs Equivalent series resistance measured
measured using the parallel equivalent
Z Absolute value of impedance
Parameter Description
series equivalent circuit model
Cp
parallel equivalent circuit model
Ls Inductance value measured using the
series equivalent circuit model
Lp Inductance value measured using the
parallel equivalent circuit model
using the series equivalent circuit model (ESR)
Rp Equivalent parallel resistance
circuit model
Y Absolute value of admittance G Conductance B Susceptance R Resistance (=Rs) X Reactance D Dissipation factor Q Quality factor(=1/D)
Phase radian
Phase angle Vac Test signal Voltage Iac Test signal Current
Absolute deviation value
Relative deviation value
DCR Direct Current Resistance
Step 1. Press the [Measure] key. Step 2. Press the [MEAS DISPLAY] soft-key if
[MEAS DISPLAY] is not already the active
mode. Step 3. Use the cursor key to select [FUNC] field. Step 4. Use the soft-keys to select measurement
function.
25
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
Mode
Function
Advantage
Disadvantage
Auto
T3
LCR Series
The
select a proper
optimum range
To achieve the best measurement result, it is recommended to set the impedance range mode to
When using measurement results may occur if the measured value exceeds the full scale of the selected range.
overview
Range
Hold Range
Nominal Range
sets the optimum
need to select
range impedance range for the impedance of the DUT Measurement No ranging
time is
required
T3LCR Series sets the
need to select
range. No depending on the nominal
ranging time is
required value
measurement time is longer due to the ranging time
You need to
range depending on the value of the DUT. Valid only in the sorting mode
The hold range has nine impedance ranges: 10, 30, 100, 300, 1k, 3k, 10k, 30k and 100k.
The impedance range is selected according to the impedance even if the measurement parameter is capacitance or inductance.
26
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Range
I
mpedance
Effective measurement range
8 10 0- 7 30 - 6
100 -
5 300 -
4 1k - 3 3k -
2 10k -
1 30k
-)
Range
5 -
3 - 2 -33 49
1 32-99
1.485
M 0
900-
100M
0 100k
- (This range is available only
when the test frequency< 20kHz.)
No.
Impedance range
Best measurement range Overload
7 - 6 -
4 - 1.485k
27
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
INCR +
Increments the impedance
Frequency range(F)
Resolution
10.00Hz
F
99.99Hz
0.01Hz
100.0Hz
F
999.9Hz
0.1Hz
1.000kHz
F 9.999kHz
1Hz
10.00kHz
F
99.99kHz
10Hz
When
Step 1. Press the [Measure] key. Step 2. Use the cursor key to select the [RANGE]
field.
Step 3. Use the soft-keys to select the impedance
range mode or impedance range.
Soft-key Function AUTO RANGE Sets to AUTO range HOLD RANGE Sets to HOLD range NOMINAL RANGE Sets to NOMINAL range
range in the HOLD mode
DECR - Decrements the impedance
range in the HOLD mode
T3LCR1300: 10Hz 300kHz
T3LCR1100: 10Hz 100kHz
T3LCR1002: 10Hz 2kHz
100.0kHz F 300.0kHz 100Hz
Step 1. Press the [Measure] key. Step 2. Use the cursor key to select the [FREQ] field. Step 3. Use the soft-keys or numeric entry keys to enter
the test frequency.
data is entered with the numeric keys, the soft-keys
change to unit labels (Hz, kHz). Soft-key Function
INCR + Be used to select between predefined DECR -
measuring frequencies.
28
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
INCR+/DECR
- 10Hz
50Hz
60Hz
100Hz
120Hz
50kHz
100kHz
200kHz
250kHz
300kHz
INCR+/DECR
- 10Hz
50Hz
60Hz
100Hz
120Hz
1kHz
2kHz
10kHz
20kHz
40kHz
50kHz
100kHz
INCR+/DECR
-
10Hz
50Hz
60Hz
100Hz
120Hz
1kHz
2kHz
Trigger
Description
INT
T3LCR Series
continuously repeats the
MAN
T3LCR
Series
performs one cycle of
pulse is input
1kHz 2kHz 10kHz 20kHz 40kHz
T3LCR Series supports four trigger modes: INT (internal), EXT (external), MAN (manual) and BUS (RS­232/USB).
measurement cycle.
measurement each time you press the [Trig] key.
EXT T3LCR Series performs one cycle of
measurement each time a rising to the handler external trigger input pin on the rear panel.
BUS T3LCR Series performs one cycle of
measurement each time it receives a trigger command sent via RS-232/USB.
Step 1. Press the [Measure] key. Step 2. Use the cursor key to select the [TRIG] field. Step 3. Use the soft-keys to select desired trigger
mode.
29
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
INT
Internal Trigger Mode
MAN
Manual Trigger Mode
EXT
External Trigger Mode
BUS
BUS Trigger Mode
be set as the effective value (RMS value) of a sine wave of
When a constant voltage or current level measure
Soft
-
key Function
INCR +
DECR
-
ALC ON
Auto Level Control ON
ALC OFF
Auto Level Control OFF
Current range
Resolution
100.0uA
LEVEL
9
99.9uA
0.1uA
Soft-key Function
The T3 test signal voltage/current level can
ment is performed, an asterisk mark (*) appears at the end of the LEVEL display.
Step 1. Press the [Measure] key Step 2. Use the cursor key to select [LEVEL] field Step 3. Use the soft-keys or numeric entry keys to
enter the test signal voltage/current level.
Refer to Table 4-11
INCR+/DECR-
10.00mV 100.0mV 300.0mV 500.0mV
1.00V 1.50V 2.00V INCR+/DECR-
100.0uA 500.0uA 1.00mA 5.00mA 10.00mA 20.00mA
Voltage range Resolution
10.00mV LEVEL 99.99mV 0.01mV
100.0mV LEVEL 999.9mV 0.1mV
1.00V LEVEL 2.00V 0.01V
1.00mA LEVEL 20.00mA 0.01mA
30
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
SLOW, MED and FAST can be selected for T3LCR Series.
SLOW mode will result in more stable and accurate measurement results.
Step 1. Press the [Measure] key. Step 2. Use the cursor key to select the [SPEED] field. Step 3. Use the soft-keys to set measurement speed.
Soft-key Function
FAST 40 times/sec*1
MED 10 times/sec*1
SLOW 3 times/sec*1
Measurement speed [ms]
Speed
Test Frequency(Hz) SLOW MED FAST
10 1600 1600 1600
20 800 800 800
100 483 160 160
31
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
Soft
-
key Function
STOP
Stop
s the recording and temporarily
1k 342 94 30
2k 336 91 26.5
10k 332 88.5 24.5
100k 332 88.5 24.5
300k 332 88.5 24.5
DCR 333 171 48
The T3LCR has an internal data buffer that record up to 10000 measurement readings. These readings can be saved to an external USB drive in a .csv file format. These readings can then be opened on a PC using software such as Windows Excel. When the LOG function is active, the display is fixed to the [MEAS DISPLAY] screen. The [FUNC], [FREQ] and [LEVEL] settings are also fixed and cannot be changed.
Step 1. Press the [Measure] key. Step 2. Use the left/right cursor keys to select the
LOG status field.
Step 3. Use the [START LOG] soft-keys to start a
new log.
Soft-key Function START LOG To start a new measurement log.
Step 4. After a new log has started, the measurement
internal buffer. The options in the following table are available during this recording process:
saves the logs into the buffer. An exclamation mark will be added after the record counts (in the [LOG] status field) to remind you that there are recorded readings in the buffer that have not yet been processed.
32
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
CLEAR
Clear
s the
internal
buffer.
CLEAR & STOP
Stop recording and clear
s the internal
CLEAR
BUFFER
Clears the internal buffer.
SAVE TO USB Saves the recorded readings in the
buffer to an external USB flash drive. The internal buffer is cleared after this operation.
BUFFER
SAVE & STOP Stops the recording and saves the
recorded readings in the internal buffer to an external USB flash drive. The internal buffer will be cleared after the save operation. If an external USB flash drive is not plugged in, the meter will not stop the ongoing recording.
buffer.
Step 5.
be added after the number of recorded counts(in the [LOG] status field). The following two operations are available in this situation:
Soft-key Function SAVE TO USB Saves the recorded readings in the
buffer to an external USB flash drive. The internal buffer is cleared after this operation.
33
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
the recorded readings in the internal buffer
When you press
those fixed trimming points. Please refer to the 12.1
Before saving to the external USB flash drive, remember to plug the external USB drive into the USB port on the front panel.
The recorded readings will be saved in a subdirectory named MEAS DATA, whose parent directory will have the same name as the T3LCR model being used. See the example in below:
Example: F:\T3LCR1300\MEAS DATE\ MEAS0000.CSV
Up to 9999 recording files can be saved, with their file names from 0001.csv-9999.csv.
The size of the internal buffer can be adjusted in
setting field. The maximum data buffer size is 10000.
the [Measure] key and the [OPEN
SHORT] soft-key, the [ OPEN SHORT ] page appears.
In this page, the OPEN/SHORT/SPOT correction for correcting the stray admittance and residual impedances can be performed.
The OPEN and SHORT function performs open or short correction over a range of fixed trimming points. The correction for all other frequencies are interpolated from Specification page to see what the trimming frequency points are for each T3LCR model.
The SPOT function will perform open, or short correction at a user-
In the [ OPEN SHORT] page, you can configure each of the following controls with the cursor placed in the corresponding field:
Open correction [OPEN]
Short correction [SHORT]
SPOT correction [SPOT]
34
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Soft
-
key Function
ON
Enables open correction.
The T3LCR OPEN correction capability cancels errors due to the stray admittance (G, B) in parallel with the device under test (Refer to Figure 4-3).
The OPEN correction data is taken at all preset frequency points, independent of the test frequency you set. Refer to the tables on page 130 for the trimming frequency point lists for each model.
Step 1. Press the [Measure] key. Step 2. Press the [OPEN SHORT] soft-key. Step 3. Use the cursor key to select the [OPEN] field.
35
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
OFF Disables open correction. MEAS OPEN Starts open correction (AC). DCR OPEN Starts open correction (DC).
Step 4. Press the [MEAS OPEN] or [DCR OPEN]
soft-key. A dialog message, -circuit the
will appear.
Step 5. Connect the test fixture to the BNC terminals.
Do not connect the DUT to the fixture.
Step 6. Press the [OK] soft-key. The T3LCR will then
measure the open admittances (capacitance and conductance) at the all test frequency points.
LCR OPEN
sage is
shown on the display.
When the measurement has finished,
measurement, you can press the [ABORT] soft-key to abort the OPEN correction.
Step 7. After the OPEN measurement has finished,
the meter will enable OPEN correction automatically. You can, however, still choose to manually enable or disable OPEN correction by depressing the ON or OFF soft­keys.
The short correction feature of the T3LCR Series compensates for any residual impedances (R,X) that may exist outside of calibration, such as the impedance of the cables and the DUT connection points.
See Figure 4-4 Residual Impedances.
36
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
ON
Enables short correction.
OFF
Disables short correct
ion. MEAS SHORT
Starts short correction
(AC).
DCR SHORT
Starts short correction
(DC).
measures
specified frequency points.
Step 1. Press the [Measure] key. Step 2. Press the [OPEN SHORT] soft-key. Step 3. Use the cursor key to select the [SHORT] field.
Soft-key Function
Step 4. Press the [MEAS SHORT] or [DCR SHORT]
soft-key, a dialog message, -circuit the
, will appear.
Step 5. Connect the test fixture to the BNC terminals
and short-circuit the test terminals.
Step 6. Press the [OK] soft-key. The T3LCR
the short impedances (inductances and resistance) at the all test frequency points.
LCR SHORT
og message is
shown on the display.
When the measurement has finished,
measurement, you can press the [ABORT] soft-key to abort the SHORT correction.
Step 7. After the SHORT measurement has finished,
the meter will enable SHORT correction automatically. You can, however, still choose to manually enable or disable SHORT correction by depressing the ON or OFF soft­keys.
The spot correction function involves performing open/short correction at user­You can specify 1 frequency point.
37
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
, will
measures
Step 1. Press the [Measure] key. Step 3. Press the [OPEN SHORT] soft-key. Step 3. Use the cursor key to select the [SPOT] field. Step 4. Enter the frequency using the numeric entry
keys.
Soft-key Function ON Enable this point. OFF Disables this point. CURRENT FREQ Use current frequency MEAS OPEN Starts open correction. MEAS SHORT Starts short correction
Step 4. Press the [MEAS OPEN] soft-key. A dialog
message, -circuit the test appear.
Step 5. Connect the test fixture to the BNC terminals.
Do not connect the DUT to the test fixture.
Step 6. Press the [OK] soft-key. The T3LCR
the open admittances (capacitance and conductance) at this frequency point.
will
be shown on the display.
38
When the measurement has finished,
, will be displayed. During the measurement, you can press the [ABORT] soft-key to abort open correction.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
measures
. During the measurement, you can press
Step 1. Press the [MEAS SHORT] soft key, a
dialog message, -circuit the test
, will be displayed.
Step 2. Connect the test fixture to the BNC terminals
and short-circuit the test terminals.
Step 3. Press the [OK] soft-key. The T3LCR
the short impedances (inductances and resistance) at this frequency point.
m will be shown on the display.
When the measurement has finished,
on the screen the [ABORT] soft-key to abort short correction.
Press the [Measure] key and press the [LIST SETUP] soft-key to open the [LIST SETUP] page. The List Measure feature can perform automatic sweep measurements by sweeping the frequency or signal level through a maximum 10 list points.
Before using the List Measure feature, you have to configure the list setup.
In the [LIST SETUP] page, you can configure each of the following list measurement controls with the cursor placed in the corresponding field.
Function mode [FUNC]
Test mode [MODE]
Parameter selection [FREQ[Hz], VOLT[V],
CURR[A]]
Limit parameter selection [LMT]
Lower and upper limits [LOWER] [UPPER]
39
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
Select the primary and the secondary measurement here.
The [LIST MEAS] page will execute a list of up to 10 swept frequency or amplitude tests.
When [MODE] is set to SEQ and [TRIG] is set to MAN, the [LIST MEAS] function will automatically execute each test step on the list in sequence until the last step is executed. The meter will then wait for the trigger button to be depressed before repeating the process.
When [MODE] is set to STEP and [TRIG] is set to MAN, the [LIST MEAS] function will automatically execute the first test step. The meter will then wait for the trigger button to be depressed before executing the next step. This process is repeated for each step on the list.
Soft-key Function SEQ Sequence measure all step by one
trigger.
STEP Measure one step by one trigger.
There are 4 available trigger modes in the [LIST MEAS] page.
40
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
EXT
Trigger by sending a signal to the
BUS
Trigger by the RS
-
232
/USB
port.
VOLT
Uses voltage as the list measurement
CURR
Uses current as the list measurement
Trigger Mode Function INT
source.
MAN Trigger by pressing the trigger
button.
trigger pin on the handler interface.
The list parameter used for the list measurement can be the measurement frequency or the signal level [V/A]. Use the list point field to specify the list measurement parameter.
Step 1. Press the [Measure] key. Step 2. Press the [LIST SETUP] soft-key. Step 3. Use the cursor key to select FREQ or VOLT
or CURR field.
Step 4. Use the soft-keys to select list measurement
parameter.
Soft-key Function FREQ Uses frequency as the list
measurement parameter
parameter
parameter
41
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
Soft
-
key Function
ON
Turn on current point
OFF
Turn off current point
OFF
Turns OFF the limit feature.
The List measurement feature supports up to 10 list points as well as measurement limit values. Each list point can be turned on or off.
Step 1. Press the [Measure] key. Step 2. Press the [LIST SETUP] soft-key. Step 3. Use the cursor key to select one of the list
point field (1 through 10)
Step 4. Use the soft-keys to turn on/off the current
point.
Step 5. Enter the list point value (maybe the
frequency value or the signal level voltage/current value). When you enter the frequency value, the soft-key labels change to unit labels (Hz, kHz). When you enter the
enter units. When you enter the signal level current value, the soft-key labels change to unit labels (uA, mA).
Step 6. Using the cursor keys, select the LMT field.
42
Step 7. Configure the limit parameter by pressing the
appropriate soft-key.
Soft-key Function PRIMARY A Uses the primary parameter as the
limit parameter.
2NDARY B Uses the secondary parameter as the
limit parameter.
Step 8. Using the cursor keys, select the LOWER
field.
Step 9. Enter the lower limit value. Step 10. Using the cursor keys, select the UPPER field
and enter the higher limit value.
Step 11. Repeat Step 4 to Step 10.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
The [LIST MEAS] page will appear when you press the [Measure] key and then the [LIST MEAS] soft-key.
On the [LIST MEAS] page, the list points are swept and the measurement results are compared to the limits. During a sweep, an asterisk mark (*) will appear on the left side of the list point currently being measured.
This page provides the following information:
Trigger mode [TRIG].
Test mode [MODE].
Range mode [RANGE].
Measurement recording [LOG].
These conditions can be set from [MEAS DISPLAY] page and [SETUP] page.
Select the trigger mode for LIST MEAS here; manual trigger is generally used for LIST MEAS.
When entering the LIST MEAS function, the default trigger mode is the manual trigger mode.
You can use the cursor button to highlight the TRIG source field and select the other available trigger sources.
43
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
irst time, the system will
Manual Trigger. Each time the instrument
EXT
Exter
nal Trigger. Each time the
Soft
-
key Function
SEQ
Sequence measure all step by one
STEP
Measure one step by one trigger.
Trigger mode
When entering LIST MEAS for the f set Trig Mode to MAN.
Trig Mode Function INT Internal Trigger. All ten list points are
swept continuous.
MAN
is triggered by [Trig] key, the list points are swept one by one.
instrument is triggered by the handler trigger pin, the list points are swept one by one.
BUS BUS Trigger. Each time the instrument is
triggered by SCPI command, the list points are swept one by one.
The [LIST MEAS] page will execute a list of up to 10 swept frequency or amplitude tests.
When [MODE] is set to SEQ and [TRIG] is set to MAN, the [LIST MEAS] function will automatically execute each test step on the list in sequence until the last step is executed. The meter will then wait for the trigger button to be depressed before repeating the process.
When [MODE] is set to STEP and [TRIG] is set to MAN, the [LIST MEAS] function will automatically execute the first test step. The meter will then wait for the trigger button to be depressed before executing the next step. This process is repeated for each step on the list.
Test Mode
trigger.
44
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Trigger Modes
Function
INT
Soft
-
key Function
START LOG
To start a new measurement log.
There are 4 available trigger modes in the [LIST MEAS] page.
source. MAN Trigger by pressing the trigger button. EXT Trigger by sending a signal to the
trigger pin on the handler interface. BUS Trigger by the RS-232/USB port.
Displays the selected measurement range for each measurement step; this field is here only for display, its
you need to change the measurement range; it can be done in the [MEAS DISPLAY] or [MEAS SETUP] function.
The T3LCR has an internal data buffer that record up to 10000 measurement readings. These readings can be saved to an external USB drive in a .csv file format. These readings can then be opened on a PC using software such as Windows Excel. When the LOG function is active, the display is fixed to the [MEAS DISPLAY] screen. The [FUNC], [FREQ] and [LEVEL] settings are also fixed and cannot be changed.
Step 1. Press the [LIST MEAS] soft-key. Step 2. Use the cursor keys to select the [LOG] status
field.
Step 3. Use the [START LOG] soft-keys to start a
new log.
Step 4. After a new log has started, the measurement
readi internal buffer. The options in the following table are available during this recording process:
45
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
CLEAR &
Stop recording and clears the internal
Soft
-
key Function
The internal buffer is cleared after this
Before saving the recorded readings in the internal buffer
Soft-key Function SAVE & STOP Stops the recording and saves the
recorded readings in the internal buffer to an external USB flash drive. The internal buffer will be cleared after the save operation. If an external USB flash drive is not plugged in, the meter will not stop the ongoing recording.
STOP
buffer.
Step 5. When th
be added after the number of recorded counts(in the [LOG] status field). The following two operations are available in this situation:
SAVE TO USB Saves the recorded readings in the
buffer to an external USB flash drive.
operation.
CLEAR
Clears the internal buffer.
BUFFER
to the external USB flash drive, remember to plug the external USB drive into the USB port on the front panel.
The recorded readings will be saved in a subdirectory named LIST DATA, whose parent directory will have the same name as the T3LCR model being used. See the example in below:
Example: F:\T3LCR1300\LIST DATE\LIST0000.CSV
Up to 9999 recording files can be saved, with their file names from 0001.csv-9999.csv.
The size of the internal buffer can be adjusted in
setting field. The maximum data buffer size is 10000.
46
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
the [MEAS DISPLAY]
key in the [ENLARGE DISPLAY] mode to quit the
whether the currently measured DUT is within the
Only four measurement values will be shown in this simplified display mode: The primary measurement parameters, the secondary measurement parameters and
two monitored parameters such as , %, Iac or Vac, etc.
Press the ENLARGE soft-key in to enter the [ENLARGE DISPLAY] mode (simplified display mode); on the contrary, press the NORMAL soft­[ENLARGE DISPLAY] mode and return to the regular [MEAS DISPLAY] mode.
Two biggest readings you see in the upper part of the [ENLARGE DISPLAY] mode are the primary and the secondary measurement parameters while the two other smaller readings in the lower part of the display are the monitored parameters you chose to be displayed. The PASS/FAIL result of the Direct Comparison function (see 4.5.2 Direct Comparison Function) for the DUT currently being measured will be displayed in the bottom right corner of the enlarged display.
The Direct Comparison function allows you to see
acceptable tolerance range without the complexity of the BIN sorting function.
47
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Measure Key
The Direct Comparison function can be turned on in the
[BIN SETUP] page. When the number of bins is set to 1 in the [BIN SETUP] page, the Direct Comparison function will be activated for the [ENLARGE DISPLAY] mode. To activate Direct
- UP] page. Please refer to the Total Number of Bins [# BINS] chapter on page 61 - .
48
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
This section includes the following information:
MEAS SETUP page
BIN SETUP page
BIN MEAS page
BIN COUNT page
Every time or everywhere you can press the [Setup] key to open the [MEAS SETUP] page.
49
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Setup Key
Select [MON 1]
or
ease
In the [MEAS SETUP] Page, you can configure each of the following measurement controls with the cursor placed in the corresponding field.
Measurement Function [FUNC]
Impedance range [RANGE]
Test Frequency [FREQ]
Trigger Mode [TRIG]
Test Signal Voltage Level [LEVEL]
Measurement Speed [SPEED]
Source Output Impedance [SRC RES]
Averaging Factor [AVG]
DC Bias Voltage [BIAS]
Auto LCZ Function [AUTO LCZ]
Monitor 1 and Monitor 2 Parameter
*1
*1
*1
*1
*1
*1
[MON 2]
Measurement Delay Time [DELAY]
Auto Level Control [ALC]
If you use T3LCR Series to test a smaller inductor, pl
If you need to compare test results with Keysight E4980A,
Step 1. Press the [Setup] key. Step 2. Use the cursor key to select the [SRC RES]
field.
Step 3. Use the soft-keys to set the source output
impedance.
50
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
INCR +
Increments the averag
ing factor in
DC bias voltage onto the DUT
1.5V voltage is applied to the device under test.
Soft-key Function 30 50 100
The averaging feature allows you to obtain moving average values of successive measurement results.
You can specify the averaging factor from integer 1 to integer 256.
Step 1. Press the [Setup] key. Step 2. Use the cursor key to select [AVG] field. Step 3. Use the soft-keys or number keys to enter
averaging factor.
Soft-key Function
steps of 1, 2, 4, 8, 16, 32, 64, 128 and 256.
DECR - Decrements the averaging factor in
steps of 1, 2, 4, 8, 16, 32, 64, 128 and 256.
This function will apply a while LCR meter is performing its AC measurement on the DUT. The settable DC bias voltage range is from -
2.5V-2.5V.
Step 1. Press the [Setup] key. Step 2. Use the cursor key to select [BIAS] field. Step 3. Use the soft-keys to select DC bias voltage.
Soft-key Function OFF Turn off the DC bias voltage 2V 2V voltage is applied to the device under test.
1.5V 1.5V voltage is applied to the device under test.
-1.5V -
-2V -2V voltage is applied to the device under test.
51
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Setup Key
0.01V
DC BIAS
2.50V
0.01V
Soft
-
key Function
OFF
Turn off the Auto LCZ Function
Bias Voltage range Resolution
-2.50V DC BIAS -0.01V 0.01V
Auto LCZ Function can help you to select a proper measurement parameter.
Step 1. Press the [Setup] key Step 2. Use the cursor key to select [AUTO LCZ] field Step 3. Use the soft-keys to turn on/off the Auto
LCZ Function.
ON Turn on the Auto LCZ Function.
When Auto LCZ Function is set to ON,
-
The T3LCR Series can monitor the other two parameters.
52
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Soft
-
key Function
OFF
Turn off
the monitor
Z Absolute value of impedance
D Dissipation factor
Q Quality factor(=1/D)
Vac Test signal Voltage
Iac Test signal Current
Absolute deviation value
Step 1. Press the [Setup] key. Step 2. Use the cursor key to select [MON 1] or
[MON 2] field.
Step 3. Use the soft-keys to select DC bias voltage.
Relative deviation value r Phase radian d Phase angle
R Resistance (=Rs) X Reactance G Conductance B Sustenance Y Absolute value of admittance
This setting decides how long the meter will wait to commence a measurement after the trigger signal is received. Settable range: 0ms to 60s.
The automatic level control (ALC) feature adjusts the voltage across the DUT or the current through the DUT to match the voltage/current level setting. Using this feature, you can try to ensure a constant signal level (voltage or current) is applied to the DUT.
53
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Setup Key
Soft
-
key Function
OFF
Turn off the Auto Level Control Function
ON
Turn on the Auto Level Control Function
The Nominal Value parameter becomes displayed and
.
54
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
10 levels (BIN1 through BIN9
Press [Setup] key and press the [BIN SETUP] soft-key to open [BIN SETUP] page.
This page allows you to configure the T3LCR Series built-in comparator. The built-in comparator can sort DUTs into a maximum of and OUT) using up to nine sets of primary parameter limits along with one set of secondary parameter limits.
In addition, DUTs that fall within the primary parameter limits but are out of the secondary parameter limits can be sorted into an auxiliary bin.
To take full advantage of the comparator, T3LCR Series is equipped with a handler interface for use in conjunction with the comparator. All 10 bin signals can output to your PLC via the handler interface.
In the [BIN SETUP] page, you can configure each of the following controls with the cursor placed in the corresponding field.
Measurement Function [FUNC]
Comparator ON/OFF [COMP]
Auxiliary bin ON/OFF [AUX]
Nominal value [NOM]
Primary parameter comparator limit mode [MODE]
Beep feature [BEEP]
Primary parameter total bins [BINS]
*1
*1
55
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Setup Key
. Please
or
The T3LCR Series simultaneously measures four components of the complex impedance (parameters) in a measurement cycle. These include primary parameter, secondary parameter and two monitor parameters refer to the 4.1 [MEAS DISPLAY] Page section on 23 f details.
T3LCR Series built-in comparator can sort DUTs into a maximum of 10 bins (BIN1 to BIN9 and OUT), using a maximum of nine pairs of primary parameter limits and one pair of secondary parameter limits.
A DUT can be sorted into the auxiliary BIN (AUX) when the primary parameter is within the limits but the secondary parameter is not.
56
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Step 1. Press the [Setup] key. Step 2. Press the [BIN SETUP] soft-key. Step 3. Use the cursor key to select [COMP] field. Step 4. Use the soft-keys to turn on/off the
comparator.
Soft-key Function OFF Turns OFF the COMP function ON Turns ON the COMP function
57
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Setup Key
ABS
After AUX is turned on, DUTs that do not fall within the primary parameter limit values are sorted as OUT. In addition, DUTs that fall within the primary parameter limits but are out of the secondary parameter limits are sorted into the auxiliary (AUX) bin.
Procedure for turning ON/OFF the auxiliary bin [AUX] Step 1. Press the [Setup] key. Step 2. Press the [BIN SETUP] soft-key. Step 3. Use the cursor key to select [AUX] field. Step 4. Use the soft-keys to turn on/off the auxiliary
bin.
Soft-key Function ON Turns ON the auxiliary bin OFF Turns OFF the auxiliary bin
You can specify the primary parameter limit values in one of the following 3 ways:
Tolerance mode [%][ ]
In tolerance mode, the comparison limit values are based on the deviations from the specified nominal value, which can be specified in the [NOM] field. You configure the tolerance mode limit values as a deviation percentage (%) or absolute ( ) parameter value.
Sequential mode [SEQ]
In sequential mode, the comparison limit values are based on the absolute value of the measurement. When you configure these limit values, you have to first define the minimum value and then the maximum value.
Nominal Value [NOM]
Sets the nominal value (reference value) for the and PER compare modes. Use the number keys to enter the value and the unit.
58
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Absolute value ( ) = UNKNOWN value nominal value
Includes the point
Excludes the point
Deviation percentages (%) = Absolute value ( )/nominal value × 100%
Includes the point
Excludes the point
59
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Setup Key
Soft
-
key Function
PER
Deviation percentages
Includes the point
Excludes the point
In Sequential mode, the comparison limit values are based on the absolute value of the measurement. The nominal value does not need to participate in operation.
Step 1. Press the [Setup] key. Step 2. Press the [BIN SETUP] soft-key. Step 3. Use the cursor key to select [MODE] field. Step 4. Use the soft-keys to select comparator mode.
ABS Absolute parameter values
SEQ Sequential mode
60
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Y
, you do not need to configure
Soft
-
key Function
OFF
Turns the beep feature off
PASS
Beep when the comparator sorting
ou must configure the nominal value when you use tolerance mode as the limit mode for the primary parameter.
In sequential mode the nominal value does not affect sorting. In sequential mode the nominal value.
Step 1. Press the [Setup] key. Step 2. Press the [BIN SETUP] soft-key. Step 3. Use the cursor key to select [NOM] field. Step 4. Enter the nominal value using the entry keys.
Use the soft-key to select the unit.
The beep feature behaves differently depending on which beep mode is in effect.
Step 1. Press the [Setup] key. Step 2. Press the [BIN SETUP] soft-key. Step 3. Use the cursor key to select [BEEP] field. Step 4. Use the soft-keys to set beep feature.
result is OK
FAIL Beep when the comparator sorting
result is NG
T3LCR series specify nine bins (1-BINS to 9-BINS).
Step 1. Press the [Setup] key. Step 2. Press the [BIN SETUP] soft-key. Step 3. Use the cursor key to select [#-BINS] field. Step 4. Use the soft-keys to select beep tone.
61
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Setup Key
Soft
-
key Function
When
1-BINS Set to 1 bin 2-BINS Set to 2 bins 3-BINS Set to 3 bins 4-BINS Set to 4 bins 5-BINS Set to 5 bins 6-BINS Set to 6 bins 7-BINS Set to 7 bins 8-BINS Set to 8 bins 9-BINS Set to 9 bins
T3LCR Series built-in comparator can sort DUTs into a maximum 10 levels (Bin1 through Bin9 and OUT) using up to nine sets of primary parameter limits along with one set of secondary parameter limits.
Step 1. Press the [Setup] key. Step 2. Press the [BIN SETUP] soft-key. Step 3. Use the cursor key to select [BIN 1 LOWER]
field Step 4. Enter the limit value using the entry keys. When in the ABS and SEQ mode, please enter
the unit using the soft-keys. When in the PER
mode, the unit is %. Step 5. Repeat Step 4 using the cursor keys until you
have entered all limit values. Step 6. Use the cursor key to select [2nd LOWER]
field and enter the lower limit value for the
secondary parameter. Step 7. Select [2nd UPPER] field to enter the upper
limit value.
the [Setup] key and the [BIN MEAS] soft-key is pressed, the [BIN MEAS] page appears. The bin sorting results are displayed in larger characters while the measurement results are shown in normal characters.
62
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
This page provides the following information:
FUNC, RANGE, FREQ, LEVEL, TRIG, LEVEL,
SPEED: These conditions can be set from [MEAS DISPLAY] page.
Comparator function ON/OFF [COMP].
Auxiliary bin ON/OFF [AUX].
T3LCR built-in comparator can sort DUTs into a maximum of 10 bins (BIN1 to BIN9 and OUT) using a maximum of nine pairs of primary parameter limits and one pair of secondary parameter limits.
A DUT which has the primary parameter within the limits, but secondary parameter is not, can be sorted into an auxiliary BIN (AUX).
See Figure 5-3 on page 57 for the Comparator Workflow.
Step 1. Press the [Setup] key. Step 2. Press the [BIN MEAS] soft-key. Step 3. Use the cursor key to select [COMP] field. Step 4. Use the soft-keys to turn on/off the
comparator.
63
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Setup Key
Soft
-
key Function
Soft
-
key Function
ON
Turns ON the auxiliary bin
OFF
Turns OFF the auxiliary bin
When the [Setup]
OFF Turns OFF the COMP function ON Turns ON the COMP function
After AUX is turned on, DUTs that do not fall within the primary parameter limit values are sorted as OUT. In addition, DUTs that fall within the primary parameter limits but out of the secondary parameter limits are sorted into the auxiliary (AUX) bin.
Step 1. Press the [Setup] key. Step 2. Press the [BIN MEAS] soft-key. Step 3. Use the cursor key to select [AUX] field. Step 4. Use the soft-keys to turn on/off the auxiliary
bin.
key and the [BIN COUNT] soft-key is
pressed, the [BIN COUNT] page appears.
This page provides the following information:
Counter function [COUNT]
64
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Soft
-
key Function
COUNT ON
Turns the counter function on
RESET COUNT
The counter is reset when this key
The number of DUTs sorted into each bin is counted while the unit sorts the DUTs into the appropriate bins using the comparator function.
When the maximum count of 99999999 is reached, the counting operation stops and the overflow message
-------- .
Step 1. Press the [Setup] key. Step 2. Press the [BIN COUNT] soft-key. Step 3. Use the cursor key to select [COUNT] field. Step 4. Use the soft-keys to set counter function.
COUNT OFF Turns the counter function off
pressed.
65
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
System Configurations
When the
This section includes the following information:
SYSTEM CONFIG page
SYSTEM INFO page
SYSTEM SERVICE page
[Measure] or [Setup] key is pressed, followed by the [SYSTEM] bottom soft-key, the [SYSTEM CONFIG] page appears.
The following information can be configured in the [SYSTEM CONFIG] page.
System Information
System date and time configuration [DATE/TIME]
Account settings [ACCOUNT]
Key Beep setting [KEY BEEP]
Baud rate setting [BAUD]
Remote control interface [REMOTE]
[HAND SHAKE]
[ERROR CODE]
[RESULT]
[DATA BUFFER]
[DEFAULT SET]
66
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Soft
-
key Function
YEAR INCR+
Increases the year in steps of 1.
YEAR DECR
- Decreases the year in steps of 1.
MONTH INCR+
Increases the
month in steps of 1.
DAY INCR+
Increases the day in steps of 1.
DAY DECR
- Decreases the day in steps of 1.
T3LCR Series features a built-in 24-hour clock.
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select date field. Step 4. Use the soft-keys to edit date.
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select time field
.
Step 4. Use the soft-keys to edit time.
67
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
System Configurations
of 1.
Decreases the second in steps of 1.
ADMIN
Administrator
USER
User
Switch user mode to administrator mode, you must input
CHANGE
Input password
(less than 9 numbers
).
Soft-key Function HOUR INCR+ Increases the hour in steps of 1. HOUR DECR- Decreases the hour in steps of 1. MINUTE INCR+ Increases the minute in steps of 1. MINUTE INCR- Decreases the minute in steps SECOND DECR+ Increases the second in steps of 1. SECOND DECR-
The T3LCR Series has two accounts, administrator and user.
Administrator: All functions can be configured by the administrator except [SYSTEM SERVICE] page.
User: All functions can be configured by user except the [SYSTEM SERVICE] page and the [FILE] page.
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select ACCOUNT field. Step 4. Use the soft-keys to change account.
Soft-key Function
a correct password.
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select
[ADMINISTRATOR] field. When the
ACCOUNT field is [USER], you should
change to [ADMINISTRATOR]. Step 4. Use the soft-keys to change the password or
delete the password. Soft-key Function
PASSWORD DELETE
The password will be removed.
PASSWORD
68
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Soft
-
key Function
ON
Turn on the key beep feature.
OFF
Turn off the key beep feature.
Key tone settings.
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select the [KEY BEEP]
field. Step 4. Use the soft-keys to set the beep settings.
Before you can control the T3LCR Series remotely via the built-in RS-232/USB controller, you have to configure the RS-232 RS-232/USB baud rate.
The T3LCR built-in RS-232/USB interface uses the SCPI language.
RS-232/USB configuration is as follows:
Data bits: 8-bit
Stop bits: 1-bit
Parity: none
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select the [BAUD] field. Step 4. Use the soft-keys to select the baud rate.
69
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
System Configurations
Soft
-
key Function
Soft
-
key Function
RS232
Select RS
-
232 port as remote control
USB
Select USB port as remote control
Feed character by default. (LF, Decimal 10, Hex 0x0A, or
1200 Sets the baud rate to 1200. 9600 Sets the baud rate to 9600. 38400 Sets the baud rate to 38400. 57600 Sets the baud rate to 57600. 115200 Sets the baud rate to 115200.
Selection the interface for PC/PLC remote control.
For other setting items about remote control mode,
89.
After the device detects the signal change on RS-232 or USB interface, the device will immediately communicate with the host at the setting serial transmission rate, and the keyboard will be locked as well.
In order to be able to communicate correctly, please confirm whether the setting serial transmission rate and
able to communicate with host computer correctly.
The command terminator EOL uses the ASCII Line
\
The EOL of the command and the EOL of the command return data are the same.
The terminator (EOL) can change the three types:
1. LF (Line feed, Hex 0x0A).
2. CR (Carriage Return, Hex 0x0D).
3. CR+LF. The T3LCR PC software only supports LF types.
70
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
Soft
-
key Function
OFF
Turn
off the Hand Shake feature.
ON
Turn o
n the Hand Shake feature.
LF Set up the EOL to LF(0Ah). CR Set up the EOL to CR(0Dh). CR+LF Set up the EOL to CR+LF(0D 0Ah).
If handshaking is turned on, the meter will return the commands it received from the PC first before returning the result string of the command.
Example: PC sends idn?
The meter returns:
idn? T3LCR1300 RevC1.0
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select the
[HAND SHAKE] field. Step 4. Use the soft-keys to set the handshake setting.
Soft-key Function
If the error code setting is set to on, the meter will return error codes if the wrong command or an invalid command is received to help you to debug your control program.
"*E00", //No error
"*E01", //"Bad command",
"*E02", //"Parameter error",
"*E03", //"Missing parameter",
"*E04", //"buffer overrun",
"*E05", //"Syntax error",
"*E06", //Invalid separator",
"*E07", //"Invalid multiplier",
"*E08", //"Numeric data error",
71
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
System Configurations
Soft
-
key Function
"*E09", //"Value too long",
"*E10", //"Invalid command"
"*E11", //"Unknown error"
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select the
[ERROR CODE] field. Step 4. Use the soft-keys to set the error code setting.
OFF Turn off the Error Code feature. ON Turn on the Error Code feature.
72
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
each time
AUTO
The meter will send out the test result
Sets the maximum record counts to be
If the Result setting it set to Auto, the meter will automatically send out the measurement results a test is finished; this setting is convenient especially when the meter is working with a sorting machine. The meter will start a test after receiving the trigger signal and then returns the test result to the sorting machine without the need to receive either the sorting machine or the control PC.
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select the [RESULT]
field. Step 4. Use the soft-keys to set.
Soft-key Function FETCH The meter will not send out the test
result after the test.
after the test.
The data buffer setting sets the maximum number of record counts that the internal buffers can store; Settable range 1-10000.
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select the
[DATA BUFFER] field. Step 4. Enter the number of sets using the entry keys
or use the soft-key to select the maximum
number of sets. Soft-key Function
MAX
10000 directly.
73
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
System Configurations
To reset setting values (MEAS SETUP) and offset values
ON
To reset to
default
factory settings
OFF
No reset to factory default
settings
(OPEN SHORT) to factory default, use the DEFAULT SET setting. The device can be quickly restored to its factory settings.
Default settings:
(MEAS SETUP): FUNC: Cp-D FREQ: 1kHz LEVEL: 1V RANGE: AUTO TRIG: INT SPEED: SLOW SRC RES: 100 BIAS: OFF AUTO LCZ: OFF DELAY: 0ms ALC: OFF AVG: 1 MON1: OFF MON2: OFF
(OPEN SHORT): OPEN TEST: ON SHORT TEST: ON SPOT: OFF
Step 1. Press the [Measure] or [Setup] key. Step 2. Press the [SYSTEM] bottom soft-key. Step 3. Use the cursor key to select the
[DEFAULT SET] field. Step 4. Use the soft-keys to set.
Soft-key Function
74
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
followed by
When the [Measure] or [Setup] key is pressed [SYSTEM] bottom soft-key, and then the [SYSTEM INFO] soft-key, the [SYSTEM INFO] page appears.
There are no configurable options in the [SYSTEM INFO] page.
75
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
File Operation
You can save up to 10 files into the internal non
This chapter provides information on the file operation of the T3LCR Series.
-volatile memory.
When the [Setup] key is pressed followed by the [FILE] bottom soft-key, the [FILE] page appears.
In this page, you can configure each of the following controls with the cursor placed in the corresponding field.
Recall a file when the instrument starts up
[AUTO RECALL].
Automatically save a file when the instrument shuts
down [AUTO SAVE].
76
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
ernal USB flash
Soft
-
key Function
INTERNAL
USB MEMORY
To access files from the external USB
Soft
-
key Function
LA
ST FILE
Last used file will be recalled at next
FILE 0
File 0 will be recalled at next
The Media field is used to select the media source from
an ext drive. A maximum of up to 10 files can be accessed from either source.
Step 1. Press the [Setup] key. Step 2. Press the [FILE] bottom soft-key. Step 3. Use the cursor key to select the [MEDIA] field. Step 4. Use the soft-keys to select the source.
MEMORY internal memory.
flash drive.
You can recall file0 or the current file when the instrument starts up.
Step 1. Press the [Setup] key. Step 2. Press the [FILE] bottom soft-key. Step 3. Use the cursor key to select the [AUTO
RECALL] field.
Step 4. Use the soft-keys to select the recall option.
startup.
instrument startup.
77
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
File Operation
RECALL
Recall configuration data from
ERASE
Erase the selected file
. The user
You can save the modified data into the last used file when the instrument power key is pressed.
Step 1. Press the [Setup] key. Step 2. Press the [FILE] bottom soft-key. Step 3. Use the cursor key to select the [AUTO SAVE]
field.
Step 4. Use the soft-keys to select the save option. Soft-key Function
ON Auto save function will be enabled. OFF Turn off the auto save function.
Step 1. Press the [Setup] key. Step 2. Press the [FILE] bottom soft-key. Step 3. Use the cursor keys to select a file (File name
from 0 to 9).
Step 4. Use the soft-keys to perform an operation. Soft-key Function
SAVE Save user configuration data into
currently selected file.
currently selected file.
configuration data will be reset to the default at the next startup.
MODIFY DES Modify the file description.
78
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
component handler or system controller. This means that
This chapter provides information about the T3 built-in handler interface. It includes:
Pin Assignment
Circuit Diagram
Timing Chart
The T3LCR built-in handler interface outputs signals that indicate the end of a measurement cycle, the result of bin sorting by the comparator. In addition, the instrument accepts input of external trigger. You can use these signals to easily integrate the T3LCR Series with a
you can fully automate such tasks as component inspection, component sorting, and processing of quality management data for higher manufacturing efficiency.
79
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Handler Interface
Pin Pin names
Signal descriptions
The sorting result is within bin1
.
The sortin
g result is within bin6
.
The sorting result is within bin7
.
1 O_BIN_1
2 O_BIN_2
3 O_BIN_3
4 O_BIN_4
5 O_BIN_5
6 O_BIN_6
7 O_BIN_7
8 O_BIN_8
9 O_BIN_9
Active low. The sorting result is within bin2. Active low. The sorting result is within bin3. Active low. The sorting result is within bin4. Active low. The sorting result is within bin5. Active low.
Active low.
Active low. The sorting result is within bin8. Active low. The sorting result is within bin9.
Active low. 10 NC 11 NC 12 NC
80
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
ignal on this pin is available only
Over fail occurs on the primary
Low fail occurs on the primary
The overall test result is a
n over fail
.
A low level on this pin indicates that
Pin Pin names
Signal descriptions
External trigger input,
Active on
a
while a high level on this pin
Pin Pin names
Signal descriptions
16,18
GND
GND pin for external power input
17 VCC
VCC pin
for external power input
13 NC
Over fail occurs on the secondary
14 O_S_OVER
measurement parameter. Active low.
(The s
after AUX is turned on.)
15 O_P_OVER
19 O_P_HI
20 O_P_LO
21 O_NG
22 O_INDEX
23 O_EOM
measurement parameter. Active low.
The logic state of this pin = O_P_HI
OR O_P_LO.
Hi fail occurs on the primary
measurement parameter. Active low.
measurement parameter. Active low.
Active low. The logic state of this pin
= O_S_OVER OR O_P_OVER.
This pin indicates that AD conversion
is still ongoing. It is not yet finished
when it is low.
the test is still ongoing and not yet
finished.
24 I_E_TRIG
25 I_K_LOCK
rising edge. Key pad lock signal. Low on this pin locks the key pad on the front panel unlocks the key pad.
81
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Handler Interface
Interface power requirements: +12.4V-36VDC, 0.2A(Min).
Output circuit: Built-in pull-up resistors are internally connected to the collector pin of the output transistors. The output pins are isolated by photocouplers.
Input pins: Isolated by photocouplers.
Sink Current: 5-50mA.
82
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Examples
This chapter covers basic measurement procedures as well as basic L, C, and R measurement theory. It also offers various measurement hints. After the descriptions of basic measurement procedures, practical measurement examples are shown using T3LCR Series.
The following flow chart shows the basic procedures used to measure the impedance of capacitors, inductors, resistors, and other components. Follow the procedures to perform impedance measurements while referring to the items noted to the right side of each step.
84
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
The basic procedure flow to perform this measurement is
Start
Setup measurement
conditions:
[FUNC]
[LEVEL]
[FREQ]
Connect the test fixture
CORRECTIONS
OPEN TEST
SHORT TEST
Connect DUT
Perform measurement
END
This paragraph describes a practical example of measuring a ceramic capacitor.
the same as the Basic Measurement Procedure described previously. In this example, a ceramic capacitor is measured under the following conditions.
Sample (DUT) Ceramic capacitor Measurement Conditions:
Function: Cs-D
Test Frequency: 1 kHz
Test Signal Level: 1V
85
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Examples
OPEN field
Step 1. Turn the T3LCR Series ON. Step 2. Set up the measurement conditions by filling in
the fields on the MEAS DISPLAY page.
Move to the FUNC field using the cursor
keys and choose Cs-D.
Move to the FREQ field using the cursor
keys and input 1 kHz.
Move to the LEVEL field using the cursor
keys and input 1 V
Step 3. Connect the test fixture to the T3LCR Series. Step 4. To compensate for the test fixture, a user
correction is required.
the unit, as shown
below:
1. Press the [Measure] key and the [OPEN
SHORT] soft-key.
2. Move to the OPEN TEST or SPOT field
using the cursor keys.
3. connect the clips to anything, as
shown here:
4. Press the [MEAS OPEN] soft-key and
then the [OK] soft-key. Wait until the
message is
displayed.
5. Press the [ON] soft-key if the
is set to OFF.
6. Move to the SHORT TEST or SPOT
86
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
field using the cursor keys.
7. Connect the clips to a shorting bar as
shown below:
8. Press the [MEAS SHORT] soft-key and
then the [OK] soft-key. Wait until the
message is
displayed.
9. Press the [ON] soft-key if the SHORT
field is set to OFF.
Step 5. Connect the DUT to the test fixture as shown
below:
Step 6. Press the [Measure] key. Measurements are
performed continuously by the internal trigger, and the measured Cs and D values of
87
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Examples
the capacitors are displayed as shown below:
88
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
This chapter provides the following information to remotely control the T3LCR Series via the RS-232 or USB:
RS-232 Interface
USB Interface
Install USB Driver
The device uses the RS-232 interface or USB interface to communicate with the computer to complete all functions. With standard SCPI commands, users can easily create various acquisition system which are suitable for themselves.
You can connect a controller (i.e. PC and PLC) to the RS-232 interface using a TELEDYNE RS-232 DB-9 cable. The serial port uses the transmit (TXD), receive (RXD) and signal ground (GND) lines of the RS-232 standard. It does not use the hardware handshaking lines CTS and RTS.
89
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Remote Control
LCR Series
r panel is used for remote
Pin 2: RxD Pin 3: TxD Pin 5: GND Pin 1, 4, 6-9: No Connection
T3LCR PC
RxDPin2 RxD Pin2
TxD Pin3TxDPin3
GNDPin5 GND Pin5
Make sure the controller you connect to T3
also uses these settings. The RS-232 interface transfers data using:
Data bits: 8 bits
Stop bits: 1 bit
Parity: none Baud rate: 1200-115200bps
You can connect a controller (i.e. PC and PLC) to the USB interface using a type A-B USB cable.
This virtual port can perform the same functions as RS232 and use the same settings as the RS232 port. It supports USB2.0 and below version.
The USB device port on the rea control. The USB port is configured as CDC interface.
When configured to CDC, the USB port on the T3LCR series will appear as a virtual COM port to a connected PC. Any terminal program that can communicate via a serial port can be used for remote control. Before the T3LCR series can be used for remote control using the
90
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
LCR Series
CDC USB class, install the appropriate CDC USB driver.
Make sure the controller you connect to T3
also uses these settings. The USB interface transfers data using:
PC connector: Type A, host T3LCR series connector: Rear panel Type B, slave Speed: 1.1/2.0 (full speed/high speed) USB Class: CDC (Communications device class) Hardware flow control: Off Data bits: 8-bit Stop bits: 1-bit Parity: none
Baud rate: 1200-115200 bps
91
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Remote Control
the USB
er end to the Type
The USB driver needs to be installed when using port for remote control. The USB interface creates a virtual COM port when connected to a PC.
Configure the interface to USB in SYSTEM>CONFIG>REMOTE item.
Connect the Type A-B USB cable to the rear panel USB B port on the T3LCR. Connect the oth A port on the PC.
Go to the Windows Device Manager.
For Windows 7 go to: Start Menu > Control Panel > Hardware and Sound > Device Manager The T3LCR will appear as an unknown LCR Meter
Right-
select the USB VCP Drivers.
The T3LCR and the converter that it is assigned to will now appear in under the Controller (Universal Serial Bus Controller) node.
92
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
The T3LCR will appear as an unknown USB Serial Port
Right-
select the USB VCP Drivers.
The T3LCR and the COM port that it is assigned to will now appear in under the Ports (COM & LPT) node.
93
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Remote Control
94
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
The EOI line is asserted by New Line or ASCII Line Feed character. (Decimal 10, Hex 0x0A, \
The following conventions and definitions are used in this chapter to describe RS-232 operation.
< > Angular brackets enclose words or characters that are used to symbolize a program code parameter or an RS-232 command.
[ ] A square bracket indicates that the enclosed items are optional.
\n Command Terminator
The T3LCR commands are divided into two types: Common commands and SCPI commands.
The common commands are defined in IEEE std. 488.2-1987, and these commands are common for all devices. The SCPI commands are used to control all of the T3LCR's functions.
The SCPI commands are tree structured, three levels deep. The highest level commands are called the subsystem commands in this manual. So the lower level commands are legal only when the subsystem commands have been selected.
A colon (:) is used to separate the higher level commands and the lower level commands.
Semicolon (;) A semicolon does not change the current path but separates two commands in the same message.
95
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
:
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
T3LCR Series User Manual
(b) If the 4th character is not a vowel, the short form
Every command and character parameter has at least
two forms, a short form and a long form. In some cases they will be the same. The short form is obtained using the following rules. A) If the long form has four characters or less, the long form and short form are the same. B) If the long form has more than 4 characters: (a) If the 4th character is a vowel, the short form is the first 3 characters of the long form. For example: comparator abbreviated to comp current abbreviated to curr range abbreviated to rang
is the first 4 characters. For example: resistance abbreviated to res volume abbreviated to vol
If the long form mnemonic is defined as a phrase
rather than a single word, then the long form mnemonic is the first character of the first word(s) followed by the entire last word. The above rules, when the long form mnemonic is a single word, are then applied to the resulting long form mnemonic to obtain the short form. For example: PercentTolerance abbreviated to ptol
97
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Command Reference
Definition
Mnemonic
1E1
8 (EXA)
EX
1E12 (TERA)
T
1E-6 (MICRO)
U
1E-12 (PICO)
P
1E-18 (ATTO)
A
The commands consist of a command header and parameters. (See the following.)
Example:
comp:nom 100.0e3
Header Parameter
Headers can be of the long form or the short form. The
long form allows easier understanding of the program code and the short form allows more efficient use of the computer.
Parameters may be of two types as follows. (A) Character Data and String Data Character data consists of ASCII characters. The abbreviation rules are the same as the rules for command headers. (B) Numeric Data (a) NR1 integer: For example, 1,+123,-123 (b) NR2 fix float: For example, 1.23,+1.23,-1.23 (c) NR3 floating point: For example, 1.23e3,
5.67e-3, 123k, 1.23M, 2.34G,
The available range for numeric data is 9.9E37. When numeric data is used as a parameter, the suffix multiplier mnemonics and suffix units (The suffix multiplier must be used with the suffix unit.) can be used for some commands as follows.
1E15 (PETA) PE
1E9 (GIGA) G 1E6 (MEGA) MA 1E3 (KILO) K 1E-3 (MILLI) M
1E-9 (NANO) N
1E-15 (PEMTO) F
98
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...