Trademark Information
Teledyne Test Tools is the registered trademark of Teledyne LeCroy.
Declaration
Teledyne LeCroy products are protected by patent law.
Teledyne LeCroy reserves the right to modify or change parts of or all the specifications
or pricing policies at company‘s sole decision.
Information in this publication replaces all previously corresponding material.
Any way of copying, extracting or translating the contents of this manual is not allowed
without the permission of Teledyne LeCroy.
Product Certification
Teledyne Test Tools guarantees this product conforms to the national and industrial
Carefully read the following safety precautions to avoid any personal injury or
damage to the instrument and any products connected to it. To avoid potential
hazards, please use the instrument as specified.
Use Proper Power Cord
Only the power cord designed for the instrument and authorized by local country should
be used.
Ground the Instrument
The instrument is grounded through the protective earth conductor of the power cord. To
avoid electric shock, please make sure the instrument is grounded correctly before
connecting its input or output terminals.
Connect the Signal Wire Correctly
The potential of the signal wire ground is equal to the earth, so do not connect the signal
wire to a high voltage.
Look Over All Terminals‟ Ratings
To avoid fire or electric shock, please look over all ratings and sign instruction of the
instrument. Before connecting the instrument, please read the manual carefully to gain
more information about the ratings.
Use Proper Over-voltage Protection
Make sure that no over-voltage (such as that caused by a thunderstorm) can reach the
Product, to avoid exposure to the danger of electrical shock.
Electrostatic Prevention
Operate in an electrostatic discharge protective area environment to avoid damages
induced by static discharge. Always ground both the internal and external conductors of
the cable to release static before connecting.
Maintain Proper Ventilation
Inadequate ventilation may cause an increase of instrument‘s temperature, which may
eventually damage the instrument. Maintain proper ventilation and inspect the intake and
fan regularly.
Do not touch exposed contacts or components when the power is on.
Use the Proper Fuse
Use only the specified fuse.
Do Not Operate Without Covers
Do not operate the instrument with covers or panels removed.
Do Not Operate With Suspected Failures.
If you suspect damage has occurred to the instrument, have it inspected by qualified
service personnel before further operation. Any maintenance, adjustment or replacement
especially to circuits or accessories must be performed by Teledyne Test Tools authorized
personnel.
Do Not Operate in Wet Conditions.
In order to avoid short circuiting to the interior of the device or electric shock, please do
not operate in a humid environment.
Do Not Operate in an Explosive Atmosphere.
In order to avoid damage to the device or personal injury, it is important to operate the
device away from an explosive atmosphere.
Keep Product Surfaces Clean and Dry.
To avoid the influence of dust and/or moisture in air, please keep the surface of device
clean and dry.
Handling Safety
Please handle with care during transportation to avoid damage to buttons, knobs, screen,
connectors and other parts.
Terms in this Manual. These terms may appear in this manual:
This electronic product is subject to disposal and recycling regulations
that vary by country and region. Many countries prohibit the disposal
of waste electronic equipment in standard waste receptacles. For more
information about proper disposal and recycling of your Teledyne
LeCroy product,
Terms and Symbols
The following symbols appear on the product or in its documentation:
WARNING
WARNING or CAUTION Condition or practice could result in injury or
loss of life, or damage to equipment.
PROTECTIVE (EARTH) TERMINAL
Rated for indoor use only. Do not operate where conductive pollutants
may be present.
T3DSO2000 series digital oscilloscope’s measurement terminals are not rated for
measurements within measurement categories II, III, or IV
WARNING
Measuring terminals on this product are not intended to be connected directly
to mains.
Measurement Category Definitions
Measurement category II is for measurements performed on circuits directly connected to
low voltage installations. Examples are measurements on household appliances, portable
tools and similar equipment.
Measurement category III is for measurements performed in the building installation.
Examples are measurements on distribution boards, circuit-breakers, wiring, including
cables, bus-bars, junction boxes, switches, socket-outlets in the fixed installation, and
equipment for industrial use and some other equipment, for example. Stationary motors
with permanent connection to the fixed installation.
Measurement category IV is for measurements performed at the source of the low-voltage
installation. Examples are electricity meters and measurements on primary over current
protection devices and ripple control units.
To avoid short circuit inside the instrument or electric shock, please do not
operate in humid environment.
Altitude
Operating: less than 3 km
Non-operation: less than 15 km
Installation (Over Voltage) Category
This product is powered by mains conforming to installation (over-voltage) category II.
WARNING
Make sure that no over-voltage (such as that caused by lightning) can reach
the product or else the operator might be exposed to danger of electric shock.
Installation (Over-voltage) Category Definitions
Installation (over-voltage) category II refers to the local power distribution level which is
applicable to equipment connected to the AC line (AC power).
This oscilloscope uses fan-forced air cooling. Please make sure that the air intake and
exhaust areas are free from obstructions and have adequate ventilation. When using the
oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside,
above and behind the instrument for adequate ventilation.
WARNING
Inadequate ventilation may cause temperature increase which would damage
the instrument. So please keep the instrument well ventilated during
operation and inspect the intake and fan regularly.
This manual introduces how to use the T3DSO2000 series digital oscilloscope in detail.
Quick Start
To Set the Vertical System
To Set the Horizontal System
To Set the Sample System
To Trigger the Oscilloscope
Serial Trigger and Decode
Digital Channels
To Save Reference Waveform
To Use the Math Operation
To Make Cursor Measurements
To Make Measurements
To Use the History Function
Display Setting
Save and Recall
System Function Setting
Provides information about preparations before
using the instrument and a brief introduction of
the key instrument features
Introduces the functions of the vertical system of
the oscilloscope
Introduces the functions of the horizontal
system of the oscilloscope
Introduces the functions of the sample system
of the oscilloscope
Introduces the trigger modes, trigger coupling,
trigger holdoff, external trigger and various
trigger types of the oscilloscope
Introduces how to trigger and decode an input
signal
Introduces how to set up and use the digital
channels
Introduces how to save and display reference
waveforms
Introduces the math operation functions of the
oscilloscope
Introduces how to use cursors to make
measurements
Introduces how to use measure functions to
measure the waveform parameters
Introduces the History function and how to use it
Introduces how to set the display of the
oscilloscope
Introduces how to save and recall the
measurement result and the setting of the
oscilloscope
Introduces how to configure the system setup
Introduces how to use the arbitrary waveform
generation function
Introduces the Default setup of the oscilloscope
Introduces how to deal with common problems
IX
This manual uses the T3DSO2304 in examples and the descriptions shown list all the
functions and performances of other models. T3DSO2000 series includes the
following models:
This chapter introduces the preparations when using the oscilloscope for the first time, the
front panel, rear panel and user interface of the oscilloscope,
Keep the damaged shipping container or cushioning material until the contents of the
shipment have been completely checked and the instrument is passed both electrical
and mechanical tests.
The consigner or carrier will be responsible for damages to the instrument resulting
from shipment. Teledyne Test Tools does not provide free maintenance or
replacement if the instrument has been damaged in shipment.
2. Inspect the instrument.
If there is mechanical damage or defect, or failure in the electrical and mechanical
tests, please contact your Teledyne Test Tools distributor.
3. Check the accessories
Please check the accessories according to the packing list. If the accessories are
incomplete or damaged, please contact your Teledyne Test Tools distributor.
Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards
for stable placement of the oscilloscope as well as better operation and observation.
The oscilloscope will accept a 100~240V, 50/60Hz or 100~120V, 400Hz power supply.
Please use the power cord provided with the accessories to connect the instrument to the
power source as shown in the figure below.
When the oscilloscope is energized, press the power key at the lower-left corner of the
front panel to start the oscilloscope. During the start-up process, the oscilloscope
performs a series of self-tests and you can hear the sounds of relays switching. After the
self-test is finished, the welcome screen is displayed.
Teledyne Test Tools provides passive probes for the T3DSO2000 series oscilloscopes.
For detailed technical information of the probes, please refer to the corresponding Probe
User‘s Guide.
Connect the Probe:
1. Connect the BNC terminal of the probe to one of the channel BNC connectors on the
front panel.
2. Connect the probe tip to the circuit point under test and the ground alligator clip of the
probe to the ground terminal of the circuit.
1. Press the Default button on the front panel to restore the instrument to its factory
default configuration.
2. Connect the ground alligator clip of the probe to the “Ground Terminal” under the
probe compensation signal output terminal.
Compensation Signal Output Terminal
Ground Terminal
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
“Compensation Signal Output Terminal” of the probe.
4. Press the Auto Setup softkey.
5. Observe the waveform on the display. In normal condition, the display should be a
square waveform as shown in the figure below:
Figure 5: Function Inspection
6. Use the same method to test the other channels. If the square waveforms actually
shown do not match that in the figure above, please perform “Probe Compensation”
in the next section.
WARNING
To avoid electric shock during the use of probe, please make certain that the
insulated wire of the probe is in good condition and does not touch the
metallic part of the probe when the probe is connected to high voltage source
All oscilloscope probes should be properly compensated before their first use with the
oscilloscope. Non-compensated or inadequate compensated probe may cause
measurement inaccuracy or error. The probe compensation procedures are as follows.
1. Set the switch to 10X on the probe.
2. Perform steps 1, 2, 3 and 4 of “Function Inspection” in the previous section.
3. Check the waveforms displayed and compare them with the following:
Over Perfectly Under
Compensated Compensated Compensated
4. Use a nonmetallic flat-head screwdriver to adjust the low-frequency compensation
adjustment hole on the probe until the waveform matches the “Perfectly Compensated”
waveform above.
1 Horizontal Control 13 Function Menus
2 Auto Setup 14 WaveGen Control
3 Run/Stop 15 Vertical Control for Analog Channels
4 Default Setup 16 WaveGen Output
5Clear Sweeps 17
Probe Compensation/ Ground
Terminal
6 Universal Knob 18 USB Host
7 Trigger Control 19 Digital Inputs
8 Decode Control 20 Analog Inputs
9 Digital Channel Control 21 One- Button shortcut for Save
10 Math Function 22 Function Menu Softkeys
11 Reference Function 23 Power On/Off
12 Vertical Control for Math and Ref
Pull up the handle vertically for easy carrying of the instrument. When you do not need
the handle, press it down.
2. LAN
The instrument can be connected to a user-network via this interface to perform
remote control.
3. USB Device
The T3DSO2000-series support SCPI remote control commands. User can control the
oscilloscope through this interface.
4. Pass/Fail or Trigger Out
This connector outputs either a Pass/Fail pulse when using this mode of operation or it
can be used to output a pulse that corresponds to each oscilloscope trigger operation.
5. External Trigger terminal
6. Safety lock Hole
This hole is designed to accommodate a Kensington style of lock (user-supplied).
7. AC Power Socket
AC power input terminal. The power requirements of this oscilloscope are 100-240 V,
50/60 Hz or 100-120 V, 400 Hz. Use the power cord provided with the instrument to
connect it to AC power.
Quickly enter the roll mode. The timebase range is from 1ns/div to 50s/div.
Press the button to open the Zoom function and press again to close the
function.
Horizontal Position Knob: Sets the horizontal location of the trigger event on the
display. The waveform will move left or right when you turn the knob. Meanwhile, the Delay
value will increase or decrease as the waveform moves. Press the knob to reset the trigger
delay to Zero.
Horizontal Scale Knob: Sets the timebase (horizontal sweep speed). Turn it
clockwise or counterclockwise to reduce or increase the timebase. The displayed waveform
will be expanded or compressed when the timebase changes. Press the knob to enter into
the Zoom mode.
Analog input channels. The four channels are marked by different colors which
are also used to mark both the waveforms on the screen and the channel input
connectors. Press channel button to open the corresponding channel and press
again to turn it off.
Vertical Position Knob: Set the vertical offset of the current waveform. Turn
clockwise to increase the position and turn counterclockwise to decrease. When
changing the offset, the waveform will move up and down and the offset message
at the lower part of the screen will change accordingly. Press the knob to quickly reset the
offset to zero.
Vertical Scale Knob: Sets the vertical scale of the current waveform. Turn
clockwise to decrease the scale value and while turn counterclockwise to increase
it. During the modification, the ‘amplitude‘ of the waveform will enlarge or reduce
and the scale message at the right side of the screen will also change. Press the knob to
quickly switch the vertical scale adjustment modes between “Coarse” and “Fine”.
Press the button to open the decode menu. Decode is an optional function.
T3DSO2000 supports two serial buses shown as 1 and 2 for analog signal
decoding. The protocols include I2C, SPI, UART/RS232, CAN and LIN.
Press the button to open the digital channel function menu (Optional function).
The T3DSO2000 support 16 digital channels and requires the optional leadset.
Press the button to open the Math operation menu in order to select functions
including add, subtract, multiply, divide, FFT, integral, differential, and square
root.
Press the button to enable the stored Reference waveform function. The
reference waveform can be used in comparisons to the current waveform.
T3DSO2000 supports saving up to 4 reference waveforms.
Ref/Math Vertical Position Knob: Sets the vertical offset of Math or Ref waveform.
Turn clockwise to increase the offset or turn counterclockwise to decrease the offset. During
the modification, the waveform will move up and down and the offset message at the lower
part of the screen will change accordingly. Press the knob to quickly reset the offset to zero.
Ref/Math Vertical Scale KnobSets the vertical scale of Math or Ref
waveform. Turn clockwise todecrease the scale or turn counterclockwise to
increase it. During the modification, the ‘amplitude‘ of the waveform will increase or decrease
and the scale value displayed on the right side of the screen will change accordingly. In
addition, the knob can be used to change a digital channel.
Press the button to open the trigger menu. The oscilloscope provides various
trigger types: Edge, Slope, Pulse, Video, Window, Interval, Dropout, Runt,
Pattern and Serial Bus (I2C/SPI/UART/RS232/CAN/LIN).
Press the button to set to Auto trigger mode.
Press the button to set to Normal trigger mode.
Press the button to set to Single trigger mode.
Trigger Level Knob: Sets the trigger level. Turn it clockwise or counterclockwise to
make the level move up or down. In addition, the trigger level value is displayed in the upper
right portion of the display. Press the knob to quickly reset the trigger level to 50% of
waveform.
Press the button to enable the waveform auto setup function. The oscilloscope
will automatically adjust the horizontal time base, vertical scale and trigger
mode according to the input signal to provide a triggered stable display.
Press the button to set the state of the instrument to “RUN” or “STOP”. In the
“RUN” state, the button glows yellow; In “STOP” state, the button glows red.
Press the button to reset the oscilloscope to its default setup. The default
voltage scale and timebase scale are respectively 1V/div and 1μs/div.
This button is a shortcut key for clear function. When measurement statistics
are being displayed, press this button to clear the count and begin a new
statistical count. When the screen persistence feature is turned on, press this
button to clear the persistence.
When the menus are hidden, you can use the waveform intensity knob to adjust
waveform intensity (0% ~ 100%). Turn clockwise to increase the brightness and
counterclockwise to reduce. You can also press Display/Persist→Intensity and use
the knob to adjust the intensity. To adjust grid brightness (0% ~ 100%) or transparency
(20% ~ 80%), please operate in the same way as waveform intensity.
2. Universal Knob
In menu operation, the light below the knob is illuminated. In such cases, the knob can
be used to select between sub-menus under the current menu. Press the knob to select
the current submenu. In addition, it can be used to modify parameters and to input a
file name.
Press the button to open the cursor function. It provides manual and tracking
cursor modes.
Press the button to enter the sampling menu. From here the acquisition mode
can be set (Normal/Peak-Detect/Average/Eres), interpolation mode (Sinx/x or
linear) and memory depth. The XY mode can be selected here as well as the
Sequence function.
Press the button to enter the file save and recall function menu. The storable
file types include Setups, Waveforms, Picture and References.
Press the button to open the WaveGen menu. The T3DSO2000 provides a
built-in function generator (optional). 11 types of Waveforms including Sine,
Square, Ramp, Pulse, DC, Noise, Cardiac, Gauss Pulse, ExpRise, ExpFall
and Arb are supported.
.
Press the button to enter the measurement setting menu. The user can set
measurement type(s), statistics and gating. Up to five parameters can be
selected.
Press the button to enter the display menu and quickly enable the persist
function. User can set the grid, Intensity, brightness, transparency.
Press the button to enter the utility menu to set system functions or parameters,
such as I/O set, sound, and screen language. In addition, some advanced
functions (such as Pass/Fail, Self Cal, install option function and Update) are
also supported.
Press this button to enter the history mode. History mode can record up to
80,000 frames. If the Sequence function is enabled, only the selected frames
are recorded (up to 80,000).
The T3DSO2000 has an on line help function that supplies multi-language help information,
that you can recall to help you operate the oscilloscope when you need.
Press any button for 2 seconds to enter the online help. All sub-menus of every main menu
have help information.
Channel settings information items (numbers 3, 4, 5 and 7).
2. Channel Label/Waveform
Different channels are marked by different colors and the color of the waveform
matches with the color of the channel.
3. Trigger Status
Available trigger status includes Ready, Auto, Stop, Arm, Trig‘d.
4. Horizontal Time Base
Represent the time of each grid on the horizontal axis of the screen. Turning the
HORIZONTAL SCALE Knob changes the horizontal sweep speed within the range of
1ns/div to 50s/div.
5. Horizontal Trigger Position (Delay), see item 6.
Use the Horizontal POSITION Knob to modify the parameter. Turn clockwise or
counterclockwise to make the waveform move right or left, which will respectively cause
the Delay parameter to decrease or increase. Press the knob to automatically reset the
parameter to zero as well as return the waveform‘s horizontal center to the middle of
the screen.
Display the trigger position of the waveform on the screen.
7. Frequency Counter
Display the frequency of current waveform as trigger source.
8. Sampling Rate/ Memory Depth
Display the current sample rate and memory depth of the oscilloscope. Use the
Horizontal scale knob to modify the parameters.
9. Trigger Setup
Trigger Type : Displays the currently selected trigger type and trigger condition
setting. Different labels are displayed when different trigger types are selected. Trigger
source : Displays the trigger source currently selected. Different labels are
displayed when different trigger source are selected and the color of the trigger
parameter area will change accordingly.
Trigger condition : Displays the current trigger condition.
Trigger coupling : Displays coupling mode (DC/AC/LF Reject/HF Reject) of the
current trigger source.
Trigger level : Displays the trigger level value of the current waveform.
Press the knob to reset the parameter to 50% of the waveform.
10. Channel Setup
Probe attenuation factor : Displays currently selected probe attenuation factor. All
factors are: 0.1X/0.2X/0.5X/1X/2X/5X/10X/20X/50X/100X/200X/500X/1000X/2000X
/5000X/10000X.
Input impedance : Displays currently selected impedance (1MΩ/50Ω) of the
channel.
Channel coupling : Displays the selected coupling mode of the current channel.
All modes are: DC/AC/GND.
Vertical Scale : Represents the voltage value of each grid on the vertical axis
of the screen.
11. Trigger Level Label
Displays position of the current channel trigger level. Press down the knob to reset the
level to vertical center of the waveform.
12. I/O status
Indicates that the USB Device (USBTMC) is connected.
Indicates that the USB Host is connected.
Indicates that the LAN port is connected. Indicates it is disconnected.
Display menus of the currently selected function module. Press any menu softkey to
select or adjust the corresponding setting.
To Use the Security Lock
Provisions for a Kensington-style lock are provided on the rear panel of the T3DSO2000
(Lock is not included). To use, align the lock with the lock hole and insert, turn the key
clockwise to lock the instrument and then remove the key from the lock.
T3DSO2000 provides 2/4 analog input channels (CH1, CH2, CH3, CH4) and provides
independent vertical control system for each channel. As the vertical system setting
methods of both channels are the same, this chapter takes CH1 as an example to
introduce the configuration method of the vertical system.
Connect a signal to the CH1 channel connector; and then press the CH1 button in the
vertical control area (VERTICAL) at the front panel to enable CH1.
The channel setting menu is displayed at the bottom of the screen and the channel label
at the right side of the screen. The information displayed in the channel label is related to
the current channel setting.
After the channel is turned on, modify the parameters such as the vertical scale, the
horizontal time base and the trigger mode according to the input signal to make the
waveform display easy to observe and measure.
Note: To turn off the channel, press the channel button twice.
If the amplitude of the input waveform is slightly greater than the full scale under the
current scale and the amplitude would be to low if the next scale is used, fine adjustment
can be used to improve the amplitude of waveform display to view signal details or
perform vertical based measurements.
Press the CH1 button on the front panel; then press the Adjust softkey to select the
desired mode. Turn the VERTICAL Scale Knob to adjust the vertical scale (clockwise to
reduce the scale and counterclockwise to increase).
The scale information in the channel label at the right side of the screen will change
accordingly during the adjustment. The adjustable range of the vertical scale is related to
the probe ratio currently set. By default, the probe attenuation factor is 1X and the
adjustable range of the vertical scale is from 1mv/div to 10 V/div.
Note: Push the VERTICAL Scale Knob to quickly switch between Coarse and Fine
adjustments.
To Adjust the Vertical Position
Turn the VERTICAL Position Knob to adjust the vertical position of the channel
waveform. Turn the knob clockwise to increase the vertical position and the channel
waveform moves up while counterclockwise to reduce the vertical position and the
waveform moves down. Push the knob to set the vertical position of the channel waveform
to zero.
During the adjustment, the vertical position information Volts Pos displays at the bottom of
the screen. The table below shows the range of vertical position according to the volt
scale.
Set the coupling mode to filter out undesired signals. For this example, we will assume the
signal-under-test is a square waveform with DC offset.
lWhen the coupling mode set to DC: The DC and AC components of the
signal-under-test can both pass into the T3DSO2000 channel.
lWhen the coupling mode set to AC: The DC components of the signal-under-test are
blocked.
lWhen the coupling mode set to GND: The DC and AC components of the
signal-under-test are both blocked.
Press the CH1 button on the front panel; then press the Coupling softkey and turn theUniversal Knob to select the desired coupling mode. The default setup is DC.
The current coupling mode is displayed in the channel label at the right side of the screen.
You can also press the Coupling softkey continuously to switch the coupling mode.
To Specify Bandwidth Limit
Set the bandwidth limit to reduce noise. For example, the signal-under-test is a pulse
with high frequency oscillation.
lWhen the bandwidth limit set to full, the high frequency components of the
signal-under-test can pass into the T3DSO2000 channel.
lWhen the bandwidth limit is set to 20MHz, the high frequency components that
exceed 20 MHz are attenuated.
Press the CH1 button the front panel; then press the BW Limit softkey to select Full or
20MHz. The default setup is Full. When bandwidth limit is enabled, the character B will be
displayed in the channel label at the right side of the screen.
Set the probe attenuation factor to match the type of the probe that you are using to
ensure correct vertical readouts.
Press the CH1 button on the front panel; then press the Probe softkey and turn theUniversal Knob to select the desired value and push the knob to confirm. The default
setup is 1X.
The probe attenuation factor is displayed in the channel label at the right side of the
screen. You can also press the Probe softkey continuously to switch the probe
attenuation factor.
Select the amplitude display unit for the current channel. The available units are V and A.
When the unit is changed, the unit displayed in the channel label will change accordingly.
1. Press CH1 button on the front panel to enter the CH1 function menu.
2. Press the Next Page softkey to enter the second page of the CH1 function menu.
3. Press the Unit softkey to select the desired unit V or A.
The default setup is V.
To Invert a Waveform
When Invert is set to On, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed and not the trigger settings.
Inverting a channel also changes the result of any math function selected and
measurement functions.
1. Press CH1 button on the front panel to enter the CH1 function menu.
2. Press the Next Page softkey to enter the second page of the CH1 function menu.
3. Press the Invert softkey to turn on or off the invert display.
Turn the HORIZONTAL Scale Knob on the front panel to adjust the horizontal time base.
Turn clockwise to reduce the horizontal time base and turn counterclockwise to increase.
The time base information at the upper left corner of the screen will change accordingly
during the adjustment. The range of the horizontal scale is from 1ns/div to 50s/div.
The Horizontal Scale Knob works (in the Normal time mode) while acquisitions are
running or when they are stopped. When in run mode, adjusting the horizontal scale knob
changes the sample rate. When stopped, adjusting the horizontal scale knob lets you
zoom into acquired data.
Turn the Position Knob on the front panel to adjust the trigger delay of the waveform.
During the modification, waveforms of all the channels will move left or right and the
trigger delay message at the upper-right corner of the screen changes accordingly. Press
the knob to quickly reset the trigger delay.
Changing the delay time moves the trigger point (solid inverted triangle) horizontally and
indicates how far it is from the time reference point. These reference points are indicated
along the top of the display grid.
All events displayed left of the trigger point happened before the trigger occurred. These
events are called pre-trigger information, and they show events that led up to the trigger
point.
Everything to the right of the trigger point is called post-trigger information. The amount of
delay range (pre-trigger and post-trigger information) available depends on the time/div
selected and memory depth.
The position knob works (in the Normal time mode) while acquisitions are running or when
they are stopped.
In Roll mode the waveform moves slowly across the screen from right to left. It operates
only on time base settings of 50 ms/div and slower. If the current time base setting is
faster than the 50 ms/div limit, it will be set to 50 ms/div when Roll mode is entered.
In Roll mode triggering is not supported. The fixed reference point on the screen is the
right edge of the screen and refers to the current moment in time. Events that have
occurred are scrolled to the left of the reference point. Since there is no trigger, no
pre-trigger or post-trigger information is available.
If you would like to stop the display in Roll mode, press the Run/Stop button. To clear the
display and restart an acquisition in Roll mode, press the Run/Stop button again.
Use Roll mode on low-frequency waveforms to yield a display much like a strip chart
recorder. It allows the waveform to roll across the display.
Zoom is a horizontally expanded version of the normal display. You can use Zoom to
locate and horizontally expand part of the normal window for a more detailed
horizontal analysis of the signals.
Press the HORIZONTAL Scale Knob on the front panel to turn on the zoom function, and
press the button again to turn off the function. When Zoom function is on, the display
divides in half. The top half of the display shows the normal time base window and the
bottom half displays the Zoom time base window.
Normal time base Zoom time base
The area of the normal display that is expanded is outlined with a box and the rest of the
normal display is ghosted. The box shows the portion of the normal sweep that is
expanded in the lower half.
To change the time base for the Zoom window, turn the Horizontal Scale Knob. The
Horizontal Scale Knob controls the size of the box. The Horizontal Position Knob sets
the left-to-right position of the zoom window. The delay value, which is the time displayed
relative to the trigger point is momentarily displayed in the upper-right corner of the display
when the Horizontal Position Knob is turned. Negative delay values indicate that a
portion of the waveform before the trigger event is displayed, and positive values indicate
a portion of the waveform after the trigger event.
To change the time base of the normal window, turn off Zoom; then, turn the HorizontalScale Knob.
Press the Run/Stop or Single button on the front panel to run or stop the acquisition
system of the scope.
lWhen the Run/Stop is green, the oscilloscope is running, that is, acquiring data when
trigger conditions are met. To stop acquiring data, press the Run/Stop button. When
stopped, the last acquired waveform is displayed.
lWhen the Run/Stop button is red, data acquisition is stopped. This is indicated by a
"Stop" displayed in the status line at the top of the display. To start acquiring data,
press Run/Stop.
lTo capture and display a single acquisition (whether the oscilloscope is running or
stopped), press Single. The Single run control lets you view single-shot events
without subsequent waveform data overwriting the display. Use Single when you
want maximum memory depth for pan and zoom.
When Single is pressed the display is cleared, the trigger mode is temporarily set to
Normal (to keep the oscilloscope from auto-triggering immediately), the trigger circuitry is
armed, the Single key is illuminated, and the oscilloscope waits until a user defined trigger
condition occurs before it displays a waveform.
When the oscilloscope triggers, the single acquisition is displayed and the oscilloscope is
stopped (the Run/Stop button is illuminated in red).
To understand the oscilloscope's sampling and acquisition modes, it is helpful to
understand sampling theory, sample rate and oscilloscope bandwidth.
Sampling Theory
The Nyquist sampling theorem states that for a limited bandwidth (band-limited) signal
with maximum frequency f
than twice the maximum frequency f
reconstructed without aliasing.
f
= f
MAX
= Nyquist frequency (fN) = folding frequency
S/2
, the equally spaced sampling frequency fS must be greater
The maximum sample rate of the T3DSO2000 is 2GSa/s. The actual sample rate of
the oscilloscope is determined by the horizontal scale and acquisition depth. Turn
the Horizontal Scale Knob to adjust the sample rate.
The current sample rate is displayed in the information area at the upper-right corner of
the screen.
The influence on the waveform when the sample rate is too low:
1. Waveform Distortion: when the sample rate is too low, some waveform details are
lost and the waveform displayed is rather different from the actual signal.
2. Waveform Aliasing: when the sample rate is lower than twice the actual signal
frequency (Nyquist Frequency), the frequency of the waveform rebuilt from the sample
data is lower than the actual signal frequency. The most common aliasing the jitter on
fast edges.
3. Waveform Leakage: when the sample rate is too low, the waveform rebuilt from the
sample data does not reflect all the actual signal information.
An oscilloscope‘s bandwidth is typically described as the lowest frequency at which input
signal sine waves are attenuated by 3 dB (-30% amplitude error).
At the oscilloscope bandwidth, sampling theory says the required sample rate is fS = 2fBW.
However, the theory assumes there are no frequency components above f
case) and it requires a system with an ideal brick-wall frequency response.
0dB
-3dB
Attenuation
(fBW in this
MAX
fN f
S
Frequency
However, digital signals have frequency components above the fundamental frequency
(square waves are made up of sine waves at the fundamental frequency and an infinite
number of odd harmonics), and typically, for 500 MHz bandwidths and below,
oscilloscopes have a Gaussian frequency response.
0dB
-3dB
Attenuation
Aliased frequency
components
fS/4 fN f
S
Frequency
Limiting oscilloscope bandwidth (fBW) to ¼ the sample rate (fS/4)
reduces frequency components above the Nyquist frequency (fN).
So, in practice, an oscilloscope‘s sample rate should be four or more times its bandwidth:
fS = 4fBW. This way, there is less aliasing, and aliased frequency components have a
greater amount of attenuation.
Memory depth refers to the number of waveform points that the oscilloscope can store in a
single trigger acquisition and it reflects the storage ability of the sample memory. The
T3DSO2000 provides up to 140 Mpts memory depth.
Press the Acquire button on the front panel; press the Mem Depth softkey and then turn
the Universal Knob to select the desired value and push the knob to confirm. Pressing
the Mem Depth softkey continually can also select the desired value.
The actual memory depth is displayed in the information area at the upper-right corner of
the screen.
Since the oscilloscope has one acquisition memory per two channels, when only one
channel is on (per two channels), the maximal memory depth is up to 140Mpts.
The relation of memory depth, sample rate and waveform length fulfills the equation
below:
The T3DSO2000 only supports real-time sampling. In this mode, the oscilloscope samples
and displays waveform within a trigger event. The maximum real-time sample rate of the
T3DSO2000 is 2GSa/s.
Press the RUN/STOP button to stop the acquisition, the oscilloscope will hold the last
display. At this point, you can still use the vertical control and horizontal control to explore
the waveform.
Under real-time sampling, the oscilloscope acquires the discrete sample values of the
waveform being displayed. In general, a waveform comprised of a singular dot per sample
point is very difficult to observe. In order to increase the visibility of the signal, the digital
oscilloscope usually uses the interpolation method to display a waveform.
The interpolation method ‘connects all the sampled points’, to estimate the whole
appearance of the waveform. The oscilloscope can use interpolation to fill in the gaps
between points and accurately reconstruct a waveform.
Press the Acquire button on the front panel to enter the ACQUIRE Function menu; then
press the Interpolation softkey to select Sinx/x or x.
lx: The adjacent sample points are directly connected with a straight line. This method
is confined to rebuilding the edge of signals, such as a square wave.
lSinx/x: Connecting the sampling points via mathematical calculation has more
versatility. Sinx/x interpolation method uses mathematical processing to calculate
results in the actual sample interval. This method involves mthematically calculating
the signal waveform, and allows it to produce a more realistic shape than pure square
wave and pulse. When the sampling rate is 3 to 5 times the bandwidth of the system,
the Sinx/x interpolation method is recommended.
The acquisition mode is used to control how to generate waveform points from sampled
points. The T3DSO2000 provides the following acquisition mode: Normal, Peak Detect,
Average and Eres.
1. Press the Acquire button on the front panel to enter the ACQUIRE function menu;
2. Press the Acquisition softkey; then turn the Universal Knob to select the desired
acquisition mode and push the knob to confirm. The default setup is Normal.
Normal
In this mode, the oscilloscope samples the signal at an equal time interval to rebuild the
waveform. For most of the waveforms, the best display effect can be obtained using this
mode. It is the default acquisition mode.
In this mode, the oscilloscope acquires the maximum and minimum values of the signal
within the sample interval to get the envelope of the signal or the narrow pulse of the
signal that might be lost. In this mode, signal aliasing can be prevented but the noise
displayed would be larger. In this mode, the oscilloscope can display all the pulses with
pulse widths as small as 1ns.
In this mode, the oscilloscope averages the waveforms from multiple acquisitions to
reduce the random (uncorrelated) noise of the input signal and improve the vertical
resolution. The greater the number of averages, the lower the noise will be and the higher
the vertical resolution will be, but the slower will be the response of the displayed
waveform to waveform changes.
The available range of averages is from 4 to 1024 with a default value of 16. When
Average mode is selected, press Averages and turn the universal knob or press the
softkey continually to set the desired average time.
This mode uses a digital filter to reduce the random noise on the input signal and generate
much smoother waveforms on the screen.
Eres can be used on both single-shot and repetitive signals and it does not slow down
waveform update. This mode limits the oscilloscope's real-time bandwidth because it
effectively acts like a low-pass filter.
Press the Acquire button on the front panel; then press the XY softkey to set the XY (On)
or YT (Off) mode. The default setup is YT.
YT
It is the normal / default viewing mode for the oscilloscope. It’s a signal (Y axis) amplitude
verses time (T, X axis) display.
XY
XY mode changes the display from a volt-versus-time display to a volt-versus-volt display.
Channel 1 amplitude is plotted on the X-axis and Channel 2 amplitude is plotted on the
Y-axis, the two channels will be turned on or off together.
XY mode can be used to compare frequency and phase relationships between two signals.
It can also be used with transducers to display strain versus displacement, flow versus
pressure, volts versus current, or voltage versus frequency.
The phase deviation between two signals with the same frequency can be easily
measured via the Lissajous method. The figure below shows the measurement schematic
diagram of the phase deviation.
According to sinθ=A/B or C/D (wherein, θ is the phase deviation angle between the two
channels and the definitions of A, B, C and D are as shown in the figure above), the phase
deviation angle is obtained, that is: θ=±arcsin (A/B) or ±arcsin (C/D)
If the principal axis of the ellipse is within quadrant I and III, the phase deviation angle
obtained should be within quadrant I and IV, namely within (0 to π/2) or (3π/2 to 2π). If the
principal axis of the ellipse is within quadrant II and IV, the phase deviation angle obtained
should be within quadrant II and III, namely within (π/2 to π) or (π to 3π/2).
The Sequence Mode does not display the waveform during the sampling process. This
improves the waveform capture rate by focusing the instrument resources on sampling.
This increases the waveform capture rate to about 500,000 wfm/s. The Sequence Mode
increases the ability to capture events with small probability effectively.
The oscilloscope keeps running and filling data memory in segments for each trigger
event, unit the memory is full, and then display all the waveforms on the screen.
To use the sequence mode, the HORIZONTAL Format must be set to YT.
Do the following steps to use the sequence mode.
1. Press the Acquire button on the front panel to enter the ACQUIRE function menu;
2. Press the Sequence softkey to enter the SEQUENCE function menu.
Figure 20: SEQUENCE Function Menu
3. Press the Segments Set softkey; and then turn the Universal Knob to select the
desired value.
Do the following steps to replay the sequence waveform under history mode:
1. Press the History softkey to enable the HISTORY function.
Figure 21: HISTORY Function Menu
2. Press the List softkey to turn on the list display. The list records the acquisition time of
every frame and shows the frame number that displaying on the screen.
3. Press the Frame No. softkey; and then turn the Universal Knob to select the frame to
To observe and measure elements of an input waveform, it is most useful to have a stable
and reproducible image of that waveform. To accomplish this, many oscilloscopes feature
a trigger model that enables you to specify certain conditions for the input signal (only
trigger on the rising edge, for example). The input signal is evaluated according to these
settings and when a waveform meets this condition, the oscilloscope captures it and it is
displayed on the screen. Digital oscilloscopes can display the input waveform continuously
without a stable trigger present. Only a stable trigger can ensure a stable display of the
waveform. The trigger circuit ensures that every time base sweep or acquisition starts from
the input signal and the user-defined trigger condition, namely every sweep is synchronous
to the acquisition and the waveforms acquired overlap to display a stable waveform.
The following is the schematic diagram of the acquisition memory. As shown in the figure
below, the position of the trigger event is determined by the reference time point and the
delay setting.
Post-trigger Buffer
Acquisition Memory
The trigger setting should be based on the type of signal characteristics of the input signal,
thus a certain amount of knowledge about the signal-under-test is helpful in order to quickly
capture the desired waveform.
T3DSO2000 provides abundant advanced trigger functions which can help you to focus on
the desired waveform details. These trigger types are Edge, Slope, Pulse, Video, Window,
Interval, Dropout, Runt, Pattern and Serial trigger. This chapter will introduce these trigger
function types listed above in detail and explain how to set the trigger conditions in order
to capture the desired waveform.
T3DSO2000 trigger sources include analog channels (CH1, CH2, CH3, CH4), as well as
external trigger inputs EXT, EXT/5 and AC Line.
Press the Setup button on the front panel to enter the TRIGGER function menu; press the
Source softkey and then turn the Universal Knob to select the desired trigger source.
The current trigger source is displayed at the upper right corner of the screen. Select
channel with signal input as trigger source to obtain stable trigger.
Analog channel input:
Signals input from analog channels 1~4 can all be used as the trigger source.
External trigger input:
The external trigger source can be used to connect an external trigger signal to the EXT
TRIG channel when all of the four channels are sampling data. The trigger signal (such as
an external clock or a signal from the circuit to be tested) will be connected to EXT and
EXT/5 trigger source via the [EXT TRIG] connector. EXT/5 trigger source attenuates the
signal by a factor of 5. It extends the trigger level. You can set the trigger level within the
range of the trigger level (between -8 div to +8 div).
AC line:
The trigger signal is obtained from the AC power input of the oscilloscope. This type of
trigger signal can be used to display the relationship between a signal of interest (such as
illuminating device) and power (power supply device). For example, it is often used in the
power industry to achieve a stable trigger when observing the waveform output from the
transformer of a transformer substation.
Note: Select a stable channel waveform as the trigger source to stabilize the display.
T3DSO2000 trigger modes include Auto, Normal and Single. The trigger mode affects the
way in which the oscilloscope searches for the trigger.
After the oscilloscope begins to capture data, the oscilloscope operates by first filling the
pre-trigger buffer. It starts searching for a trigger after the pre-trigger buffer is filled and
continues to flow data through this buffer while it searches for the trigger. While searching
for the trigger, the oscilloscope overflows the pre-trigger buffer and the first data put into
the buffer is first pushed out. This is also known as a First Input First Out, or FIFO buffer).
When/if a match for the given trigger condition is found, the pre-trigger buffer contains the
events that occurred just before the trigger. Then, the oscilloscope fills the post-trigger
buffer and displays the full acquisition memory.
Press the Auto, Normal or the Single buttons on the front panel to select the desired
trigger mode, and the corresponding status light will be lit.
Ø In the Auto trigger mode (the default setting), if the specified trigger conditions are
not found, triggers are forced and acquisitions are made so that signal activity is
displayed on the oscilloscope. The waveforms on the display may not be stable or
repeatable in this case, but you should observe some portion of the waveforms.
The Auto trigger mode is appropriate when:
lChecking DC signals or signals with unknown levels or activity.
lWhen trigger conditions occur often enough that forced triggers are
unnecessary.
Ø In the Normal trigger mode, triggers and acquisitions only occur when the specified
trigger conditions are found. Otherwise, the oscilloscope holds the original waveform
until the next trigger happens.
The Normal trigger mode is appropriate when:
lWe only want to acquire specific events specified by the trigger settings.
lTriggering on an infrequent signal from a serial bus (for example, I2C, SPI, CAN,
LIN, etc.) or another signal that arrives in bursts. The Normal trigger mode lets
you stabilize the display by preventing the oscilloscope from auto-triggering and
overwriting the current waveform with undesired data.
Ø In the Single trigger mode, the oscilloscope waits for a trigger and displays the
waveform when the trigger condition is met and then stops acquiring data.
The Single trigger mode is appropriate in order:
lTo capture single event or a periodic signal.
lTo capture burst or other unusual signals.
The trigger level can be adjusted for a selected analog channel by turning the TriggerLevel Knob.
Push the Trigger Level Knob to set the level to the middle of the waveform. If AC
coupling is used, pushing the Trigger Level knob sets the trigger level to approximately
0 V.
The position of the trigger level for the analog channel is indicated by the trigger level icon
(If the analog channel is on) at the left side of the display. The value of the analog
channel trigger level is displayed in the upper-right corner of the display.
Press the Setup button on the front panel to enter the TRIGGER function menu, and then
press the Coupling softkey and turn the Universal Knob or press the Coupling softkey
continually to select the desired coupling mode.
T3DSO2000 provides 4 trigger coupling modes:
lDC: Allows DC and AC components into the trigger path.
lAC: Blocks all the DC components and attenuates signals lower than 8 Hz. Use AC
coupling to get a stable edge trigger when your waveform has a large DC offset.
lLF Reject: Blocks the DC components and reject the low frequency components below
900 kHz. This setting also removes any unwanted low frequency components from a
trigger waveform, such as power line frequencies, etc., that can interfere with proper
triggering. Use LF Reject coupling to get a stable edge trigger when your waveform
has low frequency noise.
lHF Reject: Rejects the high frequency components higher 500kHz
Note: Trigger coupling effects the trigger setting only. Channel
Trigger holdoff can be used to stabilize the triggering of complex waveforms (such as a
pulse series). Holdoff time is the amount of time that the oscilloscope waits before re-arming
the trigger circuitry. The oscilloscope will not trigger until the holdoff time expires.
Use the holdoff to trigger on repetitive waveforms that have multiple edges (or other
events) between waveform repetitions. You can also use holdoff to trigger on the first
edge of a burst when you know the minimum time between bursts.
For example, to get a stable trigger on the repetitive pulse burst shown below, set the
holdoff time to be >200 ns but <600 ns.
Holdoff
200ns 600ns
Trigger Point
The correct holdoff setting is typically slightly less than one repetition of the waveform. Set
the holdoff to this time to generate a unique trigger point for a repetitive waveform. Only
edge trigger and serial triggers have the holdoff option. The holdoff time of T3DSO2000 is
adjustable from 100ns to 1.5s.
1. Press the Stop button, and then use the Horizontal Position Knob and the
Horizontal Scale Knob to find where the waveform repeats. Measure this time using
cursors; then, set the holdoff.
2. Press the Setup button on the front panel to enter the TRIGGER function menu. The
default trigger type is edge.
3. Press the Holdoff Close softkey; and then turn the Universal Knob to set the desired
holdoff time.
Note: Adjusting the time scale and horizontal position will not affect the holdoff time.
Noise Reject adds additional hysteresis to the trigger circuitry. By increasing the trigger
hysteresis band, you reduce the possibility of triggering on noise. However, this also
decreases the trigger sensitivity so that a slightly larger signal is required to trigger the
oscilloscope.
Press the Setup button on the front panel, and then press the Noise Reject softkey
continually to set the option to On or Off to turn on or off the noise rejection function.
If the signal you are probing is noisy, you can setup the oscilloscope to reduce the noise in
the trigger path and on the displayed waveform. First, stabilize the displayed waveform by
removing the noise from the trigger path. Second, reduce the noise on the displayed
waveform.
1. Connect a signal to the oscilloscope and obtain a stable display.
2. Remove the noise from the trigger path by setting trigger coupling to LF Reject, HF
Reject or turning on Noise Reject.
3. Set the Acquisition option to Average to reduce noise on the displayed waveform.
Edge trigger distinguishes the trigger points by seeking the specified edge (rising, falling,
rising & falling) and trigger level.
Rising Edge Falling Edge Trigger
Level
Trigger Point Trigger Point
1. Press the Setup button on the front panel to enter the Trigger system function menu.
2. Press the Type softkey; turn the Universal Knob to set select Edge and then push
the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1, CH2, CH3, CH4,
EXT, EXT/5 or AC Line as the trigger source.
4. Press the Slope softkey; turn the Universal Knob to select the desired trigger edge
(rising, falling or rising & falling), and then press down the knob to confirm. The current
trigger slope is displayed at the upper right corner of the screen.
5. Turn the Trigger Level Knob to adjust the trigger level to obtain stable trigger.
Figure 24: Edge Trigger
Note: Press the Auto Setup button will set the trigger type to Edge and slope to rising.
The slope trigger looks for a rising or falling transition from one level to another level in the
specified time range.
In T3DSO2000, positive slope time is defined as the time difference between the two
crossing points of trigger level line A and B with the positive edge as shown in the figure
below.
A
The upper limit of trigger level
B The lower limit of trigger level
Positive slope time
1. Press the Setup button on the front panel to enter the TRIGGER function menu.
2. Press the Type softkey; turn the Universal Knob to set select Slop and then push the
knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1~4 as the trigger
source.
4. Press the Slop softkey; turn the Universal Knob to set select the desired trigger edge
(rising or falling), and then push down the knob to confirm. The current trigger slope is
displayed at the upper right corner of the screen.
5. Press the Limit Range softkey; then turn the Universal Knob to select the desired
slope condition, and push down the knob to confirm.
l< (less than a time value): Trigger when the positive or negative slope time of the
input signal is lower than the specified time value.
l> (greater than a time value): Trigger when the positive or negative slope time of
the input signal is greater than the specified time value.
l[--,--](within a range of time value): Trigger when the positive or negative slope
time of the input signal is greater than the specified lower limit of time and lower
than the specified upper limit of time value.
l--][--(outside a range of time value): Trigger when the positive or negative slope
time of the input signal is greater than the specified upper limit of time and lower
than the specified lower limit of time value.
6. Press the Next Page softkey and press of Lower Upper softkey to select Lower or
Upper the trigger level; then turn the Trigger Level Knob to adjust the position. The
trigger level values are displayed at the upper right corner of the screen.
The lower trigger level cannot be larger than the upper trigger level. In the trigger
state message box, L1 indicates the upper trigger level while L2 indicates the lower
trigger level.
Video triggering can be used to capture the complicated waveforms of most standard
analog video signals. The trigger circuitry detects the vertical and horizontal interval of the
waveform and produces triggers based on the video trigger settings you have selected.
The T3DSO2000 supports standard video signals for NTSC (National Television
Standards Committee), PAL (Phase Alternating Line) HDTV (High Definition Television)
and custom video signal triggers.
1. Press the Setup button on the front panel to enter the TRIGGER function menu.
2. Press the Type softkey; then turn the Universal Knob to select Video and push down
the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1~4 as the trigger
source. Turn the Trigger Level Knob does not change the trigger level because the
trigger level is automatically set to the sync pulse.
4. Press the Standard softkey to select the desired video standard. T3DSO2000
The table below takes Of Lines as 800 as an example to explain the relation between
Of Lines, Of Fields, Interlace, Trigger Line and Trigger Field.
Of Lines Of Fields Interlace Trigger Line Trigger Field
800 1 1:1 800 1
800 1,2,4 or 8 2:1 400 1, 1~2, 1~4, 1~8
800 1,2,4 or 8 4:1 200 1, 1~2, 1~4, 1~8
800 1,2,4 or 8 8:1 100 1, 1~2, 1~4, 1~8
5. Press the Sync softkey to select Any or Select trigger mode.
lAny: Trigger on any of the horizontal sync pulses
lSelect: Trigger on the appointed line and field you have set. Press the Line or
Field softkey; then turn the Universal Knob to set the value.
The following table lists the line numbers per field for each video standard:
Standard Field 1 Field 2
NTSC 1 to 262 1 to 263
PAL 1 to 312 1 to 313
HDTV 720P/50, HDTV 720P/60 1 to 750
HDTV 1080P/50, HDTV 1080P/60 1 to 1125
HDTV 1080iP/50, HDTV 1080i/60 1 to 562 1 to 563
The following are exercises to familiarize you with video triggering:
lTo trigger on a specific line of video
lTo use Custom video trigger
To Trigger on a Specific Line of Video
Video triggering requires greater than 1/2 division of sync amplitude with any analog
channel as the trigger source.
The example below set to trigger on field 2, line 124 using the NTSC video standard.
1. Press the Setup button on the front panel to enter the TRIGGER system function
menu.
2. Press the Type softkey; then use the Universal Knob to select Video and push down
the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1 as the trigger source,
and press the knob to confirm.
4. Press the Standard softkey; turn the Universal Knob to select NTSC, and press the
knob to confirm.
5. Press the Sync softkey and set the option to Select; press the Line softkey and then
turn the universal to select 022 and push the knob to confirm; press the Field softkey
and then turn the Universal Knob to select 1 and push the knob to confirm
Custom video trigger supports frame rates of 25Hz, 30Hz, 50Hz and 60Hz, and the line
range is available from 300 to 2000. The steps below show how to set custom trigger.
1. Press the Setup button on the front panel to enter the TRIGGER function menu.
2. Press the Type softkey; then use the Universal Knob to select Video and push down
the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1 as the trigger
source, and push down the knob to confirm.
4. Press the Standard softkey; turn the Universal Knob to select Custom, and push
down the knob to confirm.
5. Press the Setting softkey to enter the custom setting function menu. Press the
Interlace softkey; turn the Universal Knob to select the desired value.
6. Press the Of Field softkey; turn the Universal Knob to select the desired value.
7. Press the Sync softkey to enter the TRIG ON menu to set the line and field.
lPress the Type softkey to select or Any.
lIf the Type option set to Select, press the Line softkey; turn the Universal Knob
to select the desired value. Press the Field softkey; turn the Universal Knob to
select the desired value.
The windows trigger provides a high trigger level and a low trigger level. The instrument
triggers when the input signal passes through the high trigger level or the low trigger level.
There are two kinds of window types; Absolute and Relative. They have different trigger
level adjustment methods. Under Absolute window type, the lower and the upper trigger
levels can be adjusted respectively via the Level knob; under Relative window type, adjust
the Center value to set the window center; adjust the Delta value to set the window range,
the lower and the upper trigger levels always move together.
Horizontal Trigger Position
High Level
Low Level
lIf the lower and the upper trigger levels are both within the waveform amplitude range,
the oscilloscope will trigger on both rising and falling edge.
lIf the upper trigger level is within the waveform amplitude range while the lower
trigger level is out of the waveform amplitude range, the oscilloscope will trigger on
rising edge only.
lIf the lower trigger level is within the waveform amplitude range while the upper
trigger level is out of the waveform amplitude range, the oscilloscope will trigger on
falling edge only.
To set window trigger via Absolute window type:
1. Press the Setup button on the front panel to enter the TRIGGER function menu.
2. Press the Type softkey; then use the Universal Knob to select Window and push
down the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1~4 as the trigger
source.
4. Press the Window Type softkey to select Absolute.
5. Press the Lower Upper softkey to select Lower or Upper trigger level; then turn the
Trigger Level Knob to adjust the position. The trigger level values are displayed at
the upper right corner of the screen.
The Lower trigger level cannot be upper than the upper trigger level. In the trigger
state message box, L1 means the upper trigger level while L2 means the lower trigger
level.
1. Press the Setup button on the front panel to enter the TRIGGER system function
menu.
2. Press the Type softkey; then use the Universal Knob to select Window and push
down the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1~4 as the trigger
source.
4. Press the Window Type softkey to select Relative.
5. Press the Center Delta softkey to select Center or Delta trigger level mode; then turn
the Trigger Level Knob to adjust the position. The Center and Delta values are
displayed at the upper right corner of the screen.
In the trigger state message box, C means Center, the center value of the lower and
upper trigger levels; D means Delta, the difference between the lower (or upper)
trigger level and the trigger level center.
DropOut trigger includes two types: Edge and state.
Edge
Trigger when the time interval (△T), measured from when the rising edge (or falling edge)
of the input signal passes through the trigger level to when the neighboring rising edge (or
falling edge) passes through the trigger level, is greater than the timeout time set, as
shown in the figure below.
Trigger Trigger
State
Trigger when the time interval (△T), measured from when the rising edge (or falling edge)
of the input signal passes through the trigger level to when the neighboring falling edge (or
rising edge) passes through the trigger level, is greater than the timeout time set, as
shown in the figure below.
Trigger Trigger
To set edge DropOut trigger:
1. Press the Setup button on the front panel to enter the TRIGGER system function
menu.
2. Press the Type softkey; then use the Universal Knob to select DropOut and push
down the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1~4 as the trigger
source. The current trigger source is displayed at the upper right corner of the screen.
Select channel with signal input as trigger source to obtain stable trigger.
4. Press the Slope softkey to select rising or falling edge.
5. Press the OverTime Type softkey to select Edge.
6. Press the Time softkey; turn the universal to select the desired value.
The Runt trigger looks for pulses that cross one threshold but not another as shown in the
picture below.
Positive runt pulse
High level
Low level
Negative runt pulse
lA positive runt pulse across through a lower threshold but not an upper threshold.
lA negative runt pulse across through an upper threshold but not a lower threshold.
To trigger on runt pulse:
1. Press the Setup button on the front panel to enter the TRIGGER system function
menu.
2. Press the Type softkey; then turn the Universal Knob to select DropOut and push
down the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1~4 as the trigger
source.
4. Press the Polarity softkey to select Positive or Negative pulse to trigger.
5. Press the LimitRange softkey; turn the Universal Knob to select the desired
condition (<, >, [--,--]or --][--).
6. Press the Time Setting softkey, and then turn the Universal Knob to select the
desired value.
7. Press the Next Page softkey to enter the second page of the TRIGGER system
function menu. Press the Lower Upper softkey to select Lower or Upper trigger level,
and the turn the Universal Knob to set the position.
The Pattern trigger identifies a trigger condition by looking for a specified pattern. The
pattern trigger can be expanded to incorporate delays similar to other triggers. Pattern
durations are evaluated using a timer. The timer starts on the last edge that makes the
pattern ‘true’. Potential triggers occur on the first edge that makes the pattern false,
provided that the time qualifier criterion has been met. The T3DSO2000 provides 4 patterns:
logical AND, OR, NAND and NOR combination of the channels. Each channel can set to
low, high or invalid.
Channel 1 High
Channel 2 Low
AND Waveform
Trigger position 1 Trigger position 2
Do the following steps to set pattern trigger:
1. Press the Setup button on the front panel to enter the TRIGGER function menu.
2. Press the Type softkey; turn the Universal Knob to select Pattern, and then push
down the knob to confirm.
3. Press each channel softkey to select Invalid, High or Low.
lLow sets the pattern to low on the selected channel. A low is a voltage level that
is less than the channel's trigger or threshold level.
lHigh sets the pattern to high on the selected channel. A high is a voltage level
that is greater than the channel's trigger or threshold level.
lInvalid sets the pattern to ‘don't care’ on the selected channel. Any channel set
to ‘don't care’ is ignored and is not used as part of the pattern.
However, if all channels in the pattern are set to Invalid, the oscilloscope will not
trigger.
Adjust the trigger level for the selected analog channel by turning the Trigger
Level knob. Invalid doesn‘t require a set trigger level.
4. Press the Next Page softkey to enter the second page of the pattern trigger menu.
5. Press the Logic softkey and then turn the Universal Knob to select the desired logic
combination AND, OR, NAND or NOR.
6. Press the Limit Range softkey; then turn the Universal Knob to select the desired
option.
l <(Less Than) – when the pattern is present for less than a time value.