PP023 Passive Probe Instruction Manual
© 2015 Teledyne LeCroy, Inc. All rights reserved.
Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal
sales and distribution purposes is strictly prohibited. However, clients are encouraged to
duplicate and distribute Teledyne LeCroy documentation for internal educational purposes.
Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Other product or brand
names are trademarks or requested trademarks of their respective holders. Information in this
publication supersedes all earlier versions. Specifications subject to change without notice.
This electronic product is subject to disposal and recycling regulations
that vary by country and region. Many countries prohibit the disposal of
waste electronic equipment in standard waste receptacles. For more
information about proper disposal and recycling of your Teledyne LeCroy
product, visit teledynelecroy.com/recycle.
925646-00
May, 2015
![](/html/1d/1d5f/1d5f5fef8841128c86ae9ddc0515884932f6fb7785e08ff35da5b5fac48d0db3/bg3.png)
Operator’s Manual
1
Terms and Symbols
The following symbols appear on the product or in its documentation:
Refer to Manual. Condition or practice
could result in injury or loss of life.
PROTECTIVE (EARTH) TERMINAL
Introduction
The Teledyne LeCroy PP023 is a passive high impedance voltage
oscilloscope probe designed and calibrated for use with general purpose
oscilloscopes having an input impedance of 1 MΩ and can be compensated
for use with instruments having an input capacitance between 10 to 20 pF.
It is compatible with oscilloscopes that automatically detect probe
attenuation and adjust their readout accordingly.
![](/html/1d/1d5f/1d5f5fef8841128c86ae9ddc0515884932f6fb7785e08ff35da5b5fac48d0db3/bg4.png)
PP023 Passive Probe
2
Specifications
Model Attn.
Input
R (MΩ) C (pF) (MHz) (ps) (m) (pF)
Attenuation Ratio ..................... 10:1 ± 1%
Max. Input Voltage................... Measurement Category I
1
: 500 Vrms
1500 V transient overvoltage
(see voltage derating curve on p. 3)
Measurement Category II
2
: 400 Vrms
Pollution Degree ...................... Pollution Degree 2
3
Operating Temperature .......... 0° C to +50° C
Humidity .................................... 5% to 95% RH (10° C to 30° C)
5% to 75% RH (30° C to 40° C)
5% to 50% RH (above 40° C)
RH not controlled below 10° C
Operating Altitude .................... 3000 meters maximum
To reduce risk of fire or electric shock, do not exceed the
voltage or category ratings of the probe.
Definitions (per EN 61010-031)
1) Measurement Category I (CAT I) refers to measurements performed on circuits not directly
connected to mains.
2) Measurement Category II (CAT II) refers to measurements performed on circuits directly
connected to the low-voltage installation.
3) Pollution Degree 2 refers to an operating environment where normally only dry, non-conductive
pollution occurs. Temporary conductivity caused by condensation must be expected.