Unauthorized duplication of Teledyne LeCroy documentation materials is strictly prohibited.
Customers are authorized to duplicate and distribute Teledyne LeCroy documentation for internal
training purposes.
Teledyne LeCroy is a trademark of Teledyne LeCroy, Inc. Other product or brand names are
trademarks or requested trademarks of their respective holders. Information in this publication
supersedes all earlier versions. Specifications are subject to change without notice.
safety of any system incorporating this accessory is the responsibility of the system owner.
DANGER, HIGH VOLTAGE. Risk of electric shock or burn.
CAUTION of damage to probe or instrument, or WARNING of hazard to health. Attend to t he
ESD CAUTION. Risk of Electrostatic Discharge (ESD) that can damage the probe or instrument
Laser Safety WARNING.
For use only by trained personnel. Not for use in households or by children.
Use only as specified.
Before use, test the probe with a known source
Do not overload; observe all terminal ratings.
Use only accessories compatible with the probe.
Connect and disconnect properly.
Safety
Follow these instructions to keep the probe operating in a correct and safe condition. Observe
generally accepted safety procedures in addition to the precautions specified here.
Symbols
These symbols appear on the probe and accessories or in this manual to alert you to important
safety considerations.
accompanying information to protec t against personal injury or damage. Do not proc eed until
conditions are fully understood and m et.
if anti-static measures are not taken.
Terms
The terms isolated and electrically floating are used in documentation to indicate measurements
where there is no direct conduction path to earth ground.
General Safety Precautions
The overall safety of any system incorporating this probe is the responsibility of the system owner.
Using the probe and/or the equipment it is connected to in a manner other
than specified may impair the protection mechanisms.
to ensure it is operating correctly.
Do not connect the probe to any circuit that exceeds
the CAT, voltage or current ratings of the oscilloscope terminal, probe or probe-accessory
combination.
Use only accessories that are rated for the
application. Using accessories other than those shipped with the probe may create an electrical
hazard.
Connect the probe to the oscilloscope before connecting to the
Keep the probe head and fiber-optic cable away from the circuits being measured.
Do not excessively bend cables
Do not remove the probe’s casing.
Do not disassemble the probe or remove inside parts.
Use only indoors within the operational environment listed.
Keep product surfaces clean and dry.
Do not use damaged cables.
Do not operate with suspected failures.
Not rated for hand-held use.
Do not connect tips to an energized circuit
Beware when using with high-voltage or high-frequency signals.
WARNING: RF burns or electric shock can occur when using the DL-ISO probes.
circuit being measured. De-energize the circuit under test before connecting probe tips.
Only tips are
intended for contact with electrical sources.
. Avoid tight radius bends, crushing, crimping, twisting or otherwise
stressing cables.
Touching exposed connections may result in electric shock.
personnel.
atmospheres.
Before each use, check cables for damage to the protective
insulation.
Before each use, inspect the probe and accessories. If any
part is damaged, cease operation immediately and secure the probe from inadvertent use.
High Voltage Safety
Do not handle probe tip connectors, cables or any part of the probe
head at any time if there is any possibility >60 Vdc/30 Vrms is present in the circuit. The plastic
case of the probe head and tip shields do not provide safe isolation.
. Always de-energize the circuit under test before
installing or removing tips.
before use to understand how the probe and accessories are derated depending on use. Refer to
the tip derating and RF burn limit graphs.
Refer all service to Teledyne LeCroy
Do not use in wet or explosive
Read the manual thoroughly
Clearance Requirements
The unique common mode voltage range of the probe system allows it to be used in the presence
of high-frequency/high-voltage common mode signals, but it is important to observe all
precautions while using it.
the probe head and tip shields do not provide safe isolation from hazardous common mode
voltages. Do not access the probe while measuring signals in the RF burn risk area.
Maximum safe handling limits for common mode voltages
Maintain 1 m (~40 in) safe clearance of the probe head and tip in all directions when
connected to an energized circuit. Place the oscilloscope and any equipment you need to access at
least 1 m from the circuit and probe head.
between the probe head and tips with respect to earth ground.
The DL-ISO probes contain multiple lasers, exposing which may cause laser burns. Do
not remove any plastic or metal covers or otherwise attempt to disassemble the probe.
Using controls, adjustments or procedures other than those specified in this manual may result in
exposure to hazardous levels of invisible radiation.
Avoid tight radius bends, crushing,
crimping, twisting or otherwise stressing the fiber-optic cable.
Before each use, inspect the fiber-optic cable for damage such as tears or other defects.
Outer sleeve
Inner sleeve
Laser Safety
WARNING:
The probe uses optical fibers within a single cable between the probe head and control module.
Therefore, standard optical fiber handling practices are required.
Minimum cable radius bend is 2 in (5.1 cm); avoid making loops smaller than 4 in (10.2 cm).
inner corrugated plastic sleeve is visible or punctured, cease operation immediately and secure the
probe from inadvertent use.
Temperature, Operating, 10 °C to 35 °C
Control Module
Temperature, Non-operating -20 °C to 70 °C
Relative Humidity 5% to 95% RH (non-condensing)
75% RH above 30 °C
45% RH above 50 °C
Altitude, Operating 3000 m (9842 ft.) maximum
Altitude, Non-Operating 10,000 m (32,807 ft.) maximum
Usage Indoor Use Only
Pollution Degree* 2, an operating environment where normally only dry, non-
conductive pollution occurs. Conductivity caused by
temporary condensation should be expected.
* Per IEC/EN 61010-1:2010.
The soft storage case and packaging materials shipped with the probe may not be
suitable for certain ESD sensitive areas, such as EPAs (ESD Protected Areas). If this is a
concern for your application, please take the necessary steps required for your test
environment.
The DL-ISO probes enable the highest confidence in GaN and SiC device characterization with
highest accuracy, best signal fidelity and comprehensive connectivity for testing servers,
motherboards, mobiles, lighting and building automation, residential inverters, UPS, switch mode
power supplies and motors. With up to 1 GHz bandwidth, 2500 V differential input range and 60 kV
common mode range, they provide an industry-leading 1.5% system accuracy when used with
12-bit High Definition Oscilloscopes (HDOs).
Key Features
• 1 GHz bandwidth
• Ideal for GaN and SiC devices
• 1.5% system accuracy (probe plus oscilloscope)
• 435 ps rise time
• High CMRR—160 dB
• Flexible connectivity options
• Auto Zero without disconnection
Compatibility
DL-ISO probes are compatible with a wide range of Teledyne LeCroy oscilloscopes equipped with
the ProBus interface. To confirm compatibility with your oscilloscope model, visit:
Required Firmware Versions
Correct operation of the DL-ISO probes with MMCX tips, HV square-pin tips, or tip accessories
requires that the oscilloscope be installed with MAUI
The
high-performance electrical amplifier. It is connected to the control module by an integrated fiberoptic cable. The
and proper voltage scaling to the oscilloscope. When connected to a ProBus interface, the DL-ISO
probe becomes an integral part of the measuring instrument, able to be controlled through the
oscilloscope application or front panel.
User Manual
contains a frequency modulating optical transmitter and receiver integrated with a
utilizes the ProBus interface to communicate tip attenuation value
DL-ISO probe head with tip (left) and control module with connector box (right)
External AC to DC Adapter
The probe is powered by an
Connect to standard AC power outlets.
Use only the adapter shipped with the probe. To prevent damage, the probe must
be powered from a supply that provides appropriately isolated and floating output voltage
(i.e., no Earth connection). Extra adapters can be ordered from CUI, Inc., Part Number:
SMI36-24.
Tripod Probe Mount
Non-conductive tripod secures probe and tip above the circuit in a stationary test setup.
Soft Storage Case
Soft case with a custom foam insert for storing probe and tips.
Probe tips are sold separately from the probe itself, but at least one tip is required for proper
functioning. Tips can be stored in a compartment in the soft carrying case.
All probe tips are high-quality coaxial attenuating tips that reject unwanted noise.
Probe tip measurement capability ranges from 2 Vpp to 2500 Vpp. The interface is MMCX or
square pins depending on how large the voltage to be measured is.
MMCX Tip
MMCX tips connect directly to an MMCX connector on customer’s circuit board. HV square-pin tips
connect to differential +ve and -ve signals on circuits. On the tip, you will see a red + and a white indicating which side connects to signal and which to reference.
HV Square Pin Tip
Description Part Number Quantity
40 Vpp MMCX Tip DL-ISO-40V-TIP 1
Each tip communicates identifying information to the probe. The tip rating determines the peak-topeak voltage level that can be measured.
Tip selection is important to optimize for best noise performance. Using a tip with a wider
dynamic voltage range than necessary will result in more noise.
All tips are interchangeable with any DL-ISO probe and do not require specialized calibration with a
specific DL-ISO probe. Simply insert the correctly rated tip into the connector on the probe head.
Electromagnetic Field ImmunityProbe Tip and Probe Head
2.3 mVrms
11.5 mVrms
46.0 mVrms
230.0 mVrms
1.2 Vrms
200 kΩ || 3.6 pF1 MΩ || 2.1 pF1 MΩ || 2.1 pF
7.5 MΩ || 2.0 pF
8 MΩ || 1.5 pF
3.0 mVrms
15.0 mVrms
60.0 mVrms
300.0 mVrms
1.5 Vrms
Up to 25 V/m (80 MHz to 2.7 GHz)
per IEC61000-4-3, criteria A
when operating in standard configuration*
Control Module
Up to 3 V/m (80 MHz to 2.7 GHz)
per IEC61000-4-3, criteria A
3.4 mVrms
17.0 mVrms
68.0 mVrms
340.0 mVrms
1.7 Vrms
* Standard configuration is a test setup with probe head positioned close to DUT and at least 1 m from control module,
oscilloscope near control module, and all laboratory conditions within environmental range listed on page 5.
Fiber-optic Cable Length
(tip to oscilloscope input)
3.375 m (11 ft.)
Control Module
Width 106.1 mm
Length
Height
WARNING:
to be used only for measurements on circuits not directly
connected to the mains.
not rated for CAT II, III or IV
measurements.
WARNING:
lower
Do not exceed the CAT, voltage or current rating of the lowest rated
individual component.
WARNING: The probe is not intended for hand-held use.
Part
Input Dynamic Range
Max. Operating
Voltage* (DC+Peak AC)
Max. Non-destruct
Voltage* (DC+Peak AC)
DL-ISO-ACC-KIT
MMCX to Square-pin Adapter
Tip dependent
2" MMCX to Y-lead Socket
Tip dependent
3" MMCX to Y- lead Solder-in
Tip dependent
2" Socket
Tip dependent
3" Solder-in
Tip dependent
Grabbers
Tip dependent
Dimensions
220.7 mm (not including strain relief)
229.8 mm (not including strain relief)
57.8 mm
Tip-Accessory Derating
Tips are derated according to the following table when used with the accessory listed.
Tips and accessories are classified No Rated Measurement Category per
IEC/EN 61010-2-030:2021 definitions. No Rated Measurement Category applies to circuits
not directly connected to the mains supply. The measurement terminals of the probe as
well as all accessories are
The probe and accessories are
accessory used.
between floating probe components and earth ground when measuring high voltages in a
stationary test setup.
The Measurement Category and voltage rating is the
Always maintain adequate spacing
250 V 300 V
2,500 V 3,300 V
* Accessories derate the same as the tip used. Refer to the Non-destruct Input Voltage (tip derating) graph.
CMRR varies by tip and measurement frequency. The table below shows typical values.
CMRR in decibels (dB) is shown in the plot below. All discontinuities are due to a combination of
noise floor limitations of the measurement device and amplification capabilities of the generators.
Operating the probe requires MAUI firmware version 10.1.x.x or higher. To confirm the
version, choose Utilities > Utilities Setup from the oscilloscope menu bar, then open the Status tab.
Connecting the Probe to the Test Instrument
Connect the probe to the oscilloscope through the ProBus input.
Connect the AC-to-DC adapter to the DC inlet on the probe control module, then connect to AC
power. The colored LED will turn green, indicating functional status.
For optimal accuracy, allow the system to warm up powered on for at least 20 minutes, then
Perform a Precision Gain Calibration prior to taking measurements (see p.21).
Attaching and Removing Tips
Insert the desired tip into the connector on the probe head. Be sure the
arrows on both connectors are aligned.
To remove tips from the probe head, gently pull back on the LEMO plug to
unlock it before pulling the tip from the probe. Do not squeeze the LEMO
plug when removing the tip.
De-energize the circuit under test before connecting tips.
: The DL-ISO uses fiber-optic isolation to permit the probe to float above ground
without damaging the oscilloscope or harming the operator.
serious harm will result if you touch these components while they are floating.
Seat MMCX tips directly into MMCX connectors.
To connect MMCX tips to square-pin headers, attach the MMCX to squarepin adapter to the tip before pressing it onto the pins.
The adapter connects to square-pin headers with mating pins
measuring 25 mil/0.025” or 0.635 mm x 0.635 mm and a pitch of 0.1”
between pins. The order of the three pins in the adapter is ReferenceSignal-Reference. Connect the Reference pin to the lower impedance node
of the circuit.
User Manual
The probe is not rated for hand-held use and
HV square-pin tips connect to differential +ve and -ve signals on circuits.
On the tip, you will see a red + and a white - indicating which side connects
19
to signal and which to reference.
The pitch between pins is 0.2” (5.08 mm). The mating pins should be
0.025” x 0.025” (0.635 mm x 0.635 mm). Alternately, round pins could be
used whose diameter varies from 0.026” to 0.033” (0.66 mm to 0.84 mm).
Most tip accessories have a red and a black interface. Red connects to
signal and black connects to reference.
When the entire DL-ISO, including the attenuating tip, is connected to a Teledyne LeCroy
oscilloscope, the displayed scale factor and measurement values are adjusted to account for the
effective gain of the probe. The oscilloscope will set the proper probe attenuation according to the
information received from the tip, while the oscilloscope probe control functions are activated, and
probe gain and offset can be controlled through the oscilloscope.
The CnDL-ISO dialog appears behind the Cn dialog of the probe input channel. The probe’s tip
model, serial number, bandwidth and input voltage range are all shown on the DL-ISO probe dialog.
The probe dialog can be used to:
• Initiate a
• Turn
• Update probe firmware, when required
without having to physically disconnect the probe
or
DL-ISO probe dialog.
procedure
Probe Volts/Div and Attenuation
The oscilloscope’s scale factor is controlled through the front panel Volts/Div knob or input channel
dialog (Cn). When the probe is connected, the oscilloscope voltage per division settings will be
limited to a minimum sensitivity setting based on the attached tip.
Offset
Offset allows you to remove a DC bias voltage from the input signal while maintaining DC coupling.
This ensures that the probe will never be overdriven while a signal is displayed on screen and
prevents inaccurate measurements.
Offset can also be used to adjust the input voltage range that can be measured. For example, the
40 Vpp tip measures -20 V to +20 V with 0 V offset. If the offset is changed to +10 V, the range of
signal that can be measured now changes to -10 to +30 V. This is especially useful when
measuring gate drives with asymmetrical voltage ranges or viewing the full swing of Vds voltages.
When offset is changed, the Attributes section of the probe dialog provides real-time guidance on
what Input Voltage Range is valid (based on the offset setting and probe tip that is connected).
The Auto Zero procedure corrects for DC offset drifts that naturally occur from thermal effects in
the probe system. DC offset drift is independent of probe tip connected, but it is affected by
oscilloscope gain setting and ambient temperature.
DL-ISO Full Scale Range (FSR) is calculated as:
8 x Vertical Gain Sensitivity
For example, at 100 mV/div, FSR = 100 mV/div x 8 = 800 mV.
DL-ISO DC Offset Drift is calculated as:
DC Offset Drift ≤ [0.05% x FSR x temperature change]
For example, if there is a 10 °C change at 100 mV/div, the DC Offset Drift will be:
[(0.05/100) x 800 mV x 10] = 4 mV
To perform an Auto Zero:
1. De-energize the circuit under test (the probe may remain physically connected).
2. Warm up the system (oscilloscope and probe) until the “warming up” indicator on the
probe dialog disappears:
3. Touch the
button on the DL-ISO dialog.
Precision Gain Calibration
Precision Gain Calibration should be performed prior to each use to ensure that gain accuracy for
the current measurement configuration meets the specifications. Depending on the accuracy
desired and/or change in ambient temperature, it may be necessary to calibrate more often.
Precision Gain Calibration includes an Auto Zero procedure.
To perform a Precision Gain Calibration:
1. De-energize the circuit under test (the probe may remain physically connected).
2. Warm up the system (oscilloscope and probe) until the “warming up” indicator on the
probe dialog disappears.
3. Touch the
Each calibration is valid for the temperature (±5 ºC) at which it was performed. The temperature is
polled to determine whether the probe is still in the valid temperature range.
The probe dialog will indicate when the probe has a valid calibration (green “√ Pass”):
Calibrations are not cached. If the probe was previously calibrated at one temperature then recalibrated at another before returning to the first temperature, the Precision Gain Calibration will
have to be repeated. The probe dialog will show that a calibration is needed by placing a red “X Fail”
next to the Precision Gain Calibration button.
If the probe is disconnected then re-connected, repeat Precision Gain Calibration.
Probe Firmware Update
If an
for proper functioning. With the probe connected to any channel of the oscilloscope, touch the
button to begin the update. The probe will appear to temporarily disconnect (there will be no
DLxx-ISO dialog), and the oscilloscope message bar will say “Updating Firmware”. The update
takes approximately two minutes. When the update is complete, the probe will automatically
reconnect and be usable.
This procedure does not update the oscilloscope firmware, only the probe firmware. The
probe must be connected to the oscilloscope throughout this process. Additionally, the DC power
supply must remain plugged in, and
button appears on the probe dialog, the probe firmware requires an update
High Voltage Amplifier (needed for 200,
1000 and 2500 V tips)
Output amplitude of at least 300 Vpp
(106.06 Vrms)
TEGAM 2340 or similar
Performance Verification
DL-ISO probes shipped from Teledyne LeCroy are tested and adjusted to meet the Gain Accuracy
specification. This procedure can be used to verify the performance of an DL-ISO.
Some of the test equipment used to verify the performance may have environmental limitations
required to meet the accuracy needed for the procedure. Make sure that the ambient conditions
meet the requirements of all the test instruments used in the procedure.
Required Test Equipment
The following equipment (or its equivalent) is required for performance verification of the DL-ISO
probe. As input/output connector types vary on different brands of instruments, additional
adapters or cables may be required.
1. Turn on the oscilloscope and allow it to warm up for at least 20 minutes.
2. Connect the probe connector box to the oscilloscope C1 input. Verify that the C1DL-ISO
tab appears behind the C1 setup dialog. This confirms that the probe is sensed.
3. Fully assemble the probe—tip connected to probe head and control module connected to
power—and allow it to warm up for 20 minutes. The LED on the control module should turn
green.
4. Meanwhile, turn on the other test equipment (i.e., DMM and function generator) and allow
these to warm up for the time recommended by the manufacturer.
5. Make a photocopy of the Test Record (page 28) and fill in the necessary data.
Functional Check
The functional check will verify the basic operation of the probe functions. It is recommended to
perform the functional check prior to any other test procedures.
1. Return to the oscilloscope to the factory default settings:
a. Select
b. Touch the
2. If the update firmware button is present, update the firmware
3. Open the C1DL-ISO dialog and touch
4. Confirm that the message "Performing AutoZero on DL-ISO...." is displayed in the message
bar and that no error messages are displayed.
RMS Noise Check
The sensitivity range and RMS noise is dependent on which lead you have available. RMS Noise
check can be performed for any or all the leads shown below and verify the test result meets the
test limit in the Test Record page.
1. Set C1 Vertical Scale to:
a. DL-ISO-2V-TIP: 100 mV/div
b. DL-ISO-10V-TIP: 500 mV/div
c. DL-ISO-40V-TIP: 2 V/div
d. DL-ISO-200V-TIP: 10 V/div
e. DL-ISO-1000V-TIP: 50 V/div
f. DL-ISO-2500V-TIP: 100V/div
2. Touch Precision Gain Calibration on the C1DL-ISO dialog. Wait while calibration completes.
NOTE:
3. Turn on P1 and set it to measure the standard deviation (sdev) of C1. Turn on statistics.
4. Clear Sweeps, wait 10 seconds, then record the mean value shown for measurement
P1:sdev(C1) on the test record and compare it to the test limit for the corresponding tip
voltage and DL-ISO model number.
Gain Accuracy Check
1. Using a BNC-to-BNC coaxial cable and BNC (f) to SMA (m) adapter, connect the function
generator to the PCF200.
For 200 V, 1000 V and 2500 V tips, connect the output of the function generator to
the input of the amplifier with a BNC to BNC cable. Connect the output of the amplifier to
the PCF200 with a BNC cable and BNC (f) to SMA (m) adaptor.
2. Using a banana plug adapter, BNC-to-BNC coaxial cable and BNC (f) to SMA (f) adapter,
connect the DMM to the PCF200.
3. Touch Precision Gain Calibration on the C1DL-ISO dialog. Wait while calibration completes.
4. Connect the probe tip to the PCF200:
a. For 2 V, 10 V, 40 V and 200 V MMCX tips:
User Manual
i. Connect the MMCX-to-square-pin adapter to the probe tip.
ii. Connect the probe tip to the square pin header on the PCF200.
b. For 1000 V and 2500 V square-pin tips:
i. Connect Red pin to socket adaptor to (+) terminal of the tip and black pin
to socket adaptor to the (-) terminal of the tip.
ii. Connect sockets to the square pin header on the PCF200, Red lead should
connect to the transmission line and black lead should connect to ground.
* The function generator is used for making relative measurements. The output of the generator is measured with a DMM or
oscilloscope. Thus, the generator is not required to be calibrated.
RMS Noise
Gain Accuracy
Permission is granted to photocopy this page to record the results of the Performance Verification procedure. File the
completed record as required by applicable internal quality procedures.
Results recorded under "Test Result" are the actual specification limit check. The test limits are included in each step.
Record other measurements and intermediate calculations that support the limit check under "Intermediate Data".
Before each use, check the fiber-optic cable for tears in the soft outer sleeve. If the corrugated
sleeve beneath is visible at any point, stop using the probe and secure it from inadvertent use.
Cleaning the Probe Body
Clean only the exterior surfaces of the device using a soft cloth or swab dampened with water or
75% isopropyl alcohol solution. Do not use harsh chemicals or abrasive cleansers. Dry the probe
and accessories thoroughly before making any voltage measurements.
The DL-ISO probe is not waterproof. Under no circumstances submerge the
probe in liquid or allow moisture to penetrate it.
Calibration Interval
The recommended calibration interval is one year. A performance verification procedure is included
in this manual. The complete performance verification should be performed as the first step of
annual calibration.
Service Strategy
The DL-ISO probe utilizes fine-pitch surface mount devices. Defective probes must be returned to a
Teledyne LeCroy service facility for diagnosis and exchange.
Refer all servicing to qualified personnel. A defective probe under warranty will
be replaced with a factory refurbished probe.
Probes not under warranty can be exchanged for a factory refurbished probe for a modest fee.
Replacement probes are factory repaired, inspected, and calibrated to the same standards as a
new product. You must return the defective probe in order to receive credit for the probe core.
Contact your local Teledyne LeCroy service center for calibration or other service. If the product
cannot be serviced on location, the service center will give you a Return Material Authorization
(RMA) code and instruct you where to ship the product. All products returned to the factory must
have an RMA.
Return shipments must be prepaid. Teledyne LeCroy cannot accept COD or Collect shipments. We
recommend air freighting. Insure the item you’re returning for at least the replacement cost.
1. Remove all accessories from the probe.
2. Pack the probe in its case. If possible, include all tips. Do not include the manual.
3. Pack the case in its original shipping box, or an equivalent carton with adequate padding to
avoid damage in transit.
4. Mark the outside of the box with the shipping address given to you by Teledyne LeCroy. Be
sure to add the following:
ATTN:<RMA code assigned by Teledyne LeCroy>
FRAGILE
5.
completing your import/export documents.
Extended warranty, calibration and upgrade plans are available for purchase. Contact your
Teledyne LeCroy sales representative to purchase a service plan.
30
contact Teledyne LeCroy Service for instructions on
THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS,
OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. TELEDYNE LECROY SHALL NOT BE
LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN
CONTRACT OR OTHERWISE. THE CUSTOMER IS RESPONSIBLE FOR THE TRANSPORTATION
AND INSURANCE CHARGES FOR THE RETURN OF PRODUCTS TO THE SERVICE FACILITY.
TELEDYNE LECROY WILL RETURN ALL PRODUCTS UNDER WARRANTY WITH TRANSPORT
PREPAID.
The product is warranted for normal use and operation, within specifications, for a period of one
year from shipment. Teledyne LeCroy will either repair or, at our option, replace any product
returned to one of our authorized service centers within this period. However, in order to do this we
must first examine the product and find that it is defective due to workmanship or materials and
not due to misuse, neglect, accident, or abnormal conditions or operation.
Teledyne LeCroy shall not be responsible for any defect, damage, or failure caused by any of the
following: a) attempted repairs or installations by personnel other than Teledyne LeCroy
representatives, or b) improper connection to incompatible equipment, or c) for any damage or
malfunction caused by the use of non-Teledyne LeCroy supplies. Furthermore, Teledyne LeCroy
shall not be obligated to service a product that has been modified or integrated where the
modification or integration increases the task duration or difficulty of servicing the oscilloscope.
Spare and replacement parts, and repairs, all have a 90-day warranty.
Products not made by Teledyne LeCroy are covered solely by the warranty of the original
equipment manufacturer.
IEC/EN 61010-1:2010/A1:2019 - Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use, Part 1: General Requirements
IEC/EN 61010-2-30:2021 - Particular Requirements for Equipment Having Testing or Measuring
Circuits
IEC/EN 60825-1 - Safety of Laser Products, Part 1: Equipment Classification and Requirements Edition 3 (2014)
US 21CFR Part 1010 Performance Standards for Electronic Products: General
US 21CFR Part 1040 Performance Standards for Light-Emitting Products
For the full list of current certifications, see the EC Declaration of Conformity shipped with your
product.
The probe is marked with this symbol to indicate that it complies with the applicable
European Union requirements to Directives 2012/19/EU on Waste Electrical and
Electronic Equipment (WEEE).
For more information about proper disposal and recycling of your Teledyne LeCroy
product,
Unless otherwise specified, all materials and processes are compliant with RoHS Directive
2011/65/EU in its entirety, inclusive of any further amendments or modifications of said Directive.