Teledyne Lecroy CP150, CP500 Operator's Manual

Operator’s Manual
CP150 and CP500 Current Probes
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Operator’s Manual
July, 2017
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes Operator’s Manual
© 2017 Teledyne LeCroy, Inc. All rights reserved.
Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales and distribution purposes is strictly prohibited. However, clients are encouraged to duplicate and distribute Teledyne LeCroy documentation for internal educational purposes.
Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Windows is a registered trademark of Microsoft Corporation. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice.
928617-00 Rev A July, 2017
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
Contents
Safety Instructions ..................................................................................................................... 1
Symbols .............................................................................................................................................. 1
Precautions ........................................................................................................................................ 1
Introduction ................................................................................................................................ 3
Specifications ............................................................................................................................. 3
Warranted Characteristics ................................................................................................................ 3
Nominal Characteristics .................................................................................................................... 4
Typical Characteristics ...................................................................................................................... 4
Environmental Characteristics ......................................................................................................... 4
Physical Characteristics ................................................................................................................... 5
Safety Ratings .................................................................................................................................... 5
CP150 Derating Curve - Maximum Input Current vs. Frequency * ................................................. 6
CP150 Insertion Impedance vs. Frequency * .................................................................................. 6
CP500 Derating Curve - Maximum Input Current vs. Frequency ................................................... 7
CP500 Insertion Impedance vs. Frequency ..................................................................................... 7
Operation .................................................................................................................................... 8
Precautions ........................................................................................................................................ 8
Connecting to the Test Instrument .................................................................................................. 9
Connecting to the Test Circuit .......................................................................................................... 9
Operating with an Oscilloscope ......................................................................................................10
Performance Verification .......................................................................................................... 12
Test Equipment Required ................................................................................................................13
Preliminary Procedure .....................................................................................................................13
Functional Check .............................................................................................................................14
Check LF Accuracy ..........................................................................................................................14
CPxxx Test Record ...........................................................................................................................16
i
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Care and Maintenance .............................................................................................................. 17
Cleaning ............................................................................................................................................17
Calibration Interval ..........................................................................................................................17
Service Strategy ...............................................................................................................................17
Troubleshooting ...............................................................................................................................17
Returning a Product for Service ................................................................................................ 18
Technical Support ..................................................................................................................... 19
Live Support .....................................................................................................................................19
Resources .........................................................................................................................................19
Service Centers ................................................................................................................................19
Warranty ................................................................................................................................... 20
Certifications ............................................................................................................................ 21
EMC Compliance ..............................................................................................................................21
Safety Compliance ...........................................................................................................................22
Environmental Compliance .............................................................................................................23
ii
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
CAUTION, possibility of electric shock.
CAUTION of damage to probe or instrument, or WARNING of hazard to health.
Do not apply around or remove from UNINSULATED HAZARDOUS LIVE Equipment protected by double insulation or reinforced insulation.
CAUTION, hot surface.
Safety Instructions
Follow these instructions to keep the probe operating in a correct and safe condition. Observe generally accepted safety procedures in addition to the precautions specified here. The overall safety of any system incorporating this accessory is the responsibility of the assembler of the system.
Symbols
These symbols appear on the probe body or in documentation to alert you to important safety considerations.
Attend to the accompanying information to protect against personal injury or damage. Do not proceed until conditions are fully understood and met.
conductors which may render electric shock, electric burn, or arc flash.
Precautions
Comply with the following safety precautions to avoid personal injury or damage to your equipment:
Use only as specified. The probe is intended to be used only with compatible Teledyne LeCroy instruments. Using the probe and/or the equipment it is connected to in a manner other than specified may impair the protection mechanisms.
Do not use the probe for measurements on Mains circuits. The probe should only be applied around or removed from insulated limited energy circuit conductors that are not directly connected to the mains.
1
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Do not overload; observe all ratings. To avoid electric shock or fire, do not connect the
current probe to any wire that carries voltages or currents that exceed the ratings of the probe.
Connect and disconnect properly. Connect the probe to the measurement instrument before connecting to the circuit/conductor being measured. Avoid damaging the cable through excessive bending.
Never install or remove the probe on bare conductors which are energized. The transformer core and shield are grounded but not insulated and may contact the conductor when the locking lever is open.
Be careful not to damage the insulation surface when making measurements. Before clamping to the conductor being measured, check that the insulation on the conductor is undamaged, and take care not to damage the insulation when clamping the conductor. Any damage to the insulation could cause an electric shock.
Use only indoors and within the operational environment listed. Do not use in wet or explosive atmospheres.
Do not remove the probe's casing. Touching exposed connections may result in electric shock.
Keep product surfaces clean and dry.
Comply with the maximum input current vs. frequency derating when measuring
current that includes a high frequency component. Using the probe at high frequencies or strong magnetic fields may cause the device to become abnormally hot, resulting in fire, equipment damage or burns.
Do not operate with suspected failures. Before each use, inspect the probe and accessories for any damage such as tears or other defects in the probe body, cable jacket, accessories, etc. If any part is damaged, cease operation immediately and sequester the probe from inadvertent use.
NOTE: Depending on the amplitude and frequency of the current being measured, the sensor head may emit a resonant sound. This sound may also occur during demagnetizing operation, but it does not represent a malfunction (device failure).
2
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
CP150
CP150-6M
CP500
Amplitude Accuracy
±1.0 % of reading at 0-150 Amp ±2.0% of reading at 150-300 Apeak
±1.0 % of reading at 0-500 Amp ±2.0% of reading at 500-700 Apeak
Bandwidth (probe only)
DC to 10 MHz
DC to 5 MHz
DC to 2 MHz
Introduction
The CP150 and CP500 current probes are designed for easy, highly accurate current measurements. The compact probes offer wide bandwidth with over-current protection.
The probes utilize a combination of Hall-effect and transformer technology which enables measurements to be made on DC, AC and impulse currents.
The CP150 has a 10 MHz bandwidth and is designed to measure continuous currents up to 150 Amps. The CP150-6M has a 5 MHz bandwidth and is designed to measure continuous currents up to 150 Amps.
NOTE: All CP150 probes, regardless of cable length, will indicate default probe bandwidth of 10 MHz in the Attributes section of the Probe dialog.
The CP500 has a 2 MHz bandwidth and is designed to measure continuous currents up to 500 Amp.
The probes can be used with any Teledyne LeCroy instrument with a ProBus interface, or by using a ProBus interface adapter. With the ProBus interface, the probe becomes an integral part of the measuring instrument. The bandwidth limit, Auto Zero and Degauss functions are all controlled from the instrument’s touch screen user interface. The interface provides power to the probe, so no external power supply is needed.
Specifications
Specifications are subject to change without notice.
Warranted Characteristics
3
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
CP150
CP150-6M
CP500
Max Continuous Input Current
150 Arms
500 Arms Max Peak Current
300 Apeak non-continuous
700 Apeak non-continuous
Intended Output Load
1 MΩ
1 MΩ
Measured Conductors
Insulated Conductors only
Insulated Conductors only
Sensitivity *
100 mA/div to 100 A/div
100 mA/div to 200 A/div
CP150
CP150-6M
CP500
Output voltage
0.01 V/A
0.01 V/A
Sensitivity Temperature Coefficient
±2% or less (0C to +40C)
±2% or less (0C to +40C) Noise (at 20 MHz BW)
6.0 mArms or less
8.0 mArms or less
Rise Time
≤ 35 ns
70 ns
≤ 175 ns
CP150
CP150-6M
CP500
Operating Temperature
0 C to 40 ⁰C
0 C to 40 ⁰C
Operating RH
≤ 80% (non-condensing)
≤ 80% (non-condensing)
Storage Temperature
-10 C to 50 C
-10 C to 50 C
Storage RH
≤ 80% (non-condensing)
≤ 80% (non-condensing)
Max Altitude
2000 m
2000 m
Effect of External Magnetic Field
Maximum of 150 mA in a DC/60 Hz, 400 A/m magnetic field
Maximum of 800 mA in a DC/60 Hz, 400 A/m magnetic field
Nominal Characteristics
* Values are based on oscilloscopes with 1 mV/div sensitivity. Numbers will be higher on instruments with lower sensitivity.
Typical Characteristics
Environmental Characteristics
4
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
CP150
CP150-6M
CP500
Probe Head Length
175 mm (6.9 inch)
175 mm (6.9 inch) Probe Head Width
27 mm (1.1 inch)
27 mm (1.1 inch)
Probe Head Height
69 mm (2.6 inch)
69 mm (2.7 inch)
Maximum diameter of conductors measured
20 mm (0.79 inch)
20 mm (0.79 inch) Cable Length
2 m (6.5 feet)
6 m (19.6 feet)
6 m (19.6 feet)
Physical Characteristics
Safety Ratings
Probe Designation1 Type D
2
Measurement Category
Safe Voltage Rating
4
Pollution Degree
Definitions (per IEC/EN 61010-2-032:2012)
1. Type D current probe is intended to be applied around or removed from insulated conductors.
2
No rated measurement category
3
Use only on fully insulated conductors
2. No rated measurement category; not intended for measurements on circuits directly connected to the Mains supply. Not rated for measurements within Measurement Categories II, III, or IV.
3. Not rated for measurements on uninsulated conductors.
4. Pollution Degree 2 refers to operating environment where normally only dry, non-conductive pollution occurs. Temporary conductivity caused by condensation must be expected.
NOTE: Because of the marking requirements per safety standard EN 61010-2-032:2012 (Third Edition), the measurement category rating that was on the CPxxx probes originally certified per EN 61010-2-032:2002 (Second Edition) has been removed from the product label. However, this change does not reduce the measurement capability or the level of protection offered by the new probe design.
5
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
CP150 Derating Curve - Maximum Input Current vs. Frequency *
* Does not apply to CP150-6M
CP150 Insertion Impedance vs. Frequency *
* Does not apply to CP150-6M
6
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
CP500 Derating Curve - Maximum Input Current vs. Frequency
CP500 Insertion Impedance vs. Frequency
7
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
WARNING: Comply with the derating characteristics of the probe when CAUTION: Do not touch the probe head until it has had sufficient time to
CAUTION: Do not use this probe for making measurements on Mains
Operation
Precautions
The sensor head is a precision assembly consisting of a molded component with a ferrite core and a Hall effect element. It may be damaged if subjected to sudden changes in temperature, mechanical strain or shock.
The mating surfaces of the sensor are precision ground and should be treated with care. If there is any type of dust or dirt on the mating surfaces of the sensor head, measurements may be impaired.
Accurate measurements may not be possible in locations subject to strong magnetic fields such as transformers and high-current conductors, or in locations subject to strong external electric fields.
measuring current that includes a high frequency component, and never measure any current that exceeds the rated current. Using the probe at high frequencies or in strong magnetic fields may cause the probe head to become abnormally hot, resulting in fire, equipment damage or burns.
cool after disconnecting the probe from the circuit. Using the probe near the current and ambient temperature upper limits for extended lengths of time may cause the probe head surface to become hot to the touch.
Circuits. Using the probe on uninsulated hazardous live conductors may result in electric shock, electric burn, or arc flash.
8
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
CAUTION: Never use this probe on bare conductors. The core and shield are
probe or the circuit under test.
Connecting to the Test Instrument
The probe has been designed for use with the Teledyne LeCroy instruments equipped with the ProBus interface. When you attach the probe output connector to the instrument’s input connector, the instrument will:
Recognize the probe model
Set the input termination to 1 MΩ
Activate the probe control functions in the touch screen user interface.
Connecting to the Test Circuit
The probe has been designed with a movable split core, eliminating the need to break the conductor for the core to slip around the conductor. To connect:
1. Pull back the slider so that the clamp opens.
2. Align the sensor so that the current direction indicator corresponds to the
direction of current flow in the conductor.
3. Close the probe jaws around the conductor until the jaws click shut.
4. Close the slider on the sensor head until the "UNLOCK’ indication disappears
and the "LOCK" indication appears.
5. Verify that the opening lever is firmly locked and the clamp is securely closed.
grounded and any voltage applied to the conductor may cause damage the
9
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Operating with an Oscilloscope
When connected to a Teledyne LeCroy oscilloscope, the displayed scale factor and measurement values will be adjusted to account for the effective gain of the probe.
A/div, Offset, and Coupling, and Bandwidth (BWL) are controlled from the Channel setup (Cx) dialog.
Channel setup dialog with Probe dialog behind it.
The probe’s attributes are shown on the Probe dialog, which appears behind the Channel dialog when a probe is detected. The Probe dialog also contains controls for Auto Zero and Degauss.
Probe dialog.
Probe A/div and Attenuation
The front panel Volts/div knob controls the oscilloscope’s scale factor and the probe’s internal attenuation to give full available dynamic range from 100mA/div to 200 A/div for the CP500 and 100mA/div to 100 A/div for the CP150.
Alternatively, A/div may be controlled from the channel setup dialog (Cx).
Bandwidth Limit
The probe is capable of switching from Full (maximum bandwidth) to 20 MHz bandwidth by changing the Bandwidth setting on the Cx dialog.
10
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
AC Coupling
In general, using offset to adjust a DC current on the screen is the preferred method to measure transient signals in the presence of a larger DC currents. The offset has limits that will cause a signal that is beyond the linear operating range of the probe to go off the screen, preventing measurement errors.
There are times, however, when it is convenient to use AC coupling to remove the DC component of the measured signal from the measurement. Selecting AC uses the scope AC coupling at the probe output to remove any steady state value from the displayed voltage.
NOTE: Since this AC coupling is on the probe output, DC current beyond the linear range of the probe will cause the probe to saturate and make the displayed waveform inaccurate. It is important not to exceed the maximum linear input values when using AC coupling.
Auto Zero
The Probe dialog incorporates an Auto Zero function to remove the DC offset from the current probe. Auto Zero must be invoked by the user. After several minutes of warm­up, or when the probe is exposed to a large shift in ambient temperature, some DC offset drift may occur. Open the Probe dialog and touch AUTO ZERO.
Degauss
If the probe has been magnetized by external magnetic field or by excessive input, the core can be demagnetized by degaussing. The demagnetizing process takes about 5 seconds and should always be performed before taking a measurement.
Without clamping the probe around a conductor, slide the opening lever to close and lock the probe. Open the Probe dialog and touch DEGAUSS.
An Auto Zero is automatically performed as part of the degauss cycle.
11
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Performance Verification
This procedure can be used to verify the warranted characteristics of the probe. The recommended calibration interval for the models CP150 and CP500 Current Probes is one year. The complete performance verification procedure should be performed as the first step of annual calibration. Performance verification can be completed without removing the probe covers or exposing the user to hazardous voltages. Test results can be recorded on a photocopy of the Test Record provided at the end of the manual.
The warranted characteristics of the probe are valid at any temperature within the Environmental Characteristics listed in the Specifications. However, some of the other test equipment used to verify the performance may have environmental limitations required to meet the accuracy needed for the procedure. Make sure that the ambient conditions meet the requirements of all the test instruments used in his procedure.
NOTE: Operation of the probe as described requires software version 4.3.0.0 or higher. To confirm the version installed, choose Utilities > Utilities Setup from the oscilloscope menu bar, then open the Status tab.
12
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
Wide Band Oscilloscope
200 MHz bandwidth
Teledyne LeCroy WaveRunner 6 Zi
Digital Multimeter (2)
DC: 0.1% Accuracy
Agilent Technologies 34401A
Function Generator
50 Hz sine wave output 3 Vrms
Teledyne LeCroy WaveStation 3082
Calibration Fixture
500 turn loop in series with 0.5Ω
Teledyne LeCroy CP150-CF02
Calibration Fixture
ProBus Extension Cable
Teledyne LeCroy PROBUS-CF01
Banana Plug Adapter
Female BNC to Dual Banana Plug
Pomona 1269
Patch Cables (4)
Male Banana to Male Banana, 12"
Pomona B-12-0 (black) BNC Adapter
BNC Male to Dual Banana Jack
Pomona 1296
Test Equipment Required
The following table lists the test equipment and accessories (or their equivalents) which are required for performance verification of the CP150/CP0500 Current Probes. Because the input and output connector types may vary on different brands and models of test instruments, additional adapters or cables may be required.
Description Minimum Requirements Example Equipment
ProBus interface equipped
5½ digit resolution
into 50 Ω
±0.1% resistor with sense terminals
Fluke 8842A-09
Stanford Research Model DS340
Teledyne LeCroy CP500-CF02
Pomona B-12-2 (red)
Preliminary Procedure
1. Connect the probe to the C1 input of the instrument and completely close the
probe slider.
2. Turn on the instrument and allow at least 30 minutes warmup time for the probe
before performing the verification procedure.
3. Turn on the other test equipment and allow these to warm up for the time
recommended by the manufacturer.
4. While the instruments are reaching operating temperature, make a photocopy of
the Performance Verification Test Record and fill in the necessary data.
13
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Functional Check
The functional check will verify the basic operation of the probe functions. It is recommended to perform the functional check prior to the performance verification procedure.
1. Open the C1 setup dialog and confirm that the bandwidth is set to Full (BWL
OFF).
2. Verify that the probe is sensed and the probe dialog (CP150 or CP500 tab)
appears behind the C1 setup dialog.
3. Open the probe dialog and degauss the probe by touching DEGAUSS, then OK.
4. Confirm that the message "Performing Degauss on CPxxx...." is displayed in the
message bar and that no error messages are displayed.
Check LF Accuracy
LF Accuracy Test set up.
14
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
1. Set the Function generator to 50 Hz sine wave, output voltage at 1.7 Vrms for
CP150 or 3 Vrms for CP500 with 50 Ω output.
2. Remove the probe from the instrument and reconnect using the ProBus
extension cable. Connect the BNC male connector of the ProBus extension to DMM #1 using a BNC Female to Dual Banana adapter.
3. Using Banana Patch cords and the BNC to Dual Banana Plug adapter, connect
the ’V Source’ and ’V Return’ terminals of the 500 Turn Calibration Loop to the output of the Function Generator.
4. Connect the Current Sense terminals of the 500 Turn Calibration Loop to the
voltage inputs of DMM #2.
5. Set both DMMs to measure AC Volt.
6. With the probe removed from any signal and the slider returned to the LOCKED
position, degauss the probe by pressing the DEGAUSS button.
7. Open the probe slider and position the probe input around the 500 Turn loop.
Close and LOCK the slider.
8. Adjust the Function generator voltage until the voltage measured at the ’Current
Sense’ terminals (DMM #2) reads 10 mV ±0.01 mV (10 A at the probe head) for CP150 or 50 mV ±0.05 mV (500 A at the probe head) for CP500.
9. Record the voltage measured by DMM #1 on the Test Record.
10. Verify that the measured voltage is between 99 mV and 101 mV for CP150, or
0.495 V and 0.505 V for CP500.
15
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Model:
_________________________________
Serial Number:
_________
Asset/Tracking Number:
_________________________________
Date:
_________________________________
Technician:
_________
Equipment
Model
Serial Number
Calibration Due Date
Digital Multimeter #1
Digital Multimeter #2
Function Generator*
N/A
Step
Description
Intermediate Data
Test Result
10
Probe Output
Limit
99
0.495 - 0.505 V for CP500
________________ V
CPxxx Test Record
________________________
________________________
* The function generator is used for making relative measurements. The output of the generator is measured with a DMM or oscilloscope. Thus, the generator is not required to be calibrated.
s:
- 101 mV for CP150
Permission is granted to photocopy this page and record the results of the performance verification procedure on the copy. File the completed record as required by applicable internal quality procedures.
Results recorded under "Test Result" are the actual specification limit check. The test limits are included in all of these steps. Record other measurements and intermediate calculations that support the limit check under "Intermediate Data".
16
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
CAUTION: The probe case is not sealed and should never be immersed in any
fluid.
Care and Maintenance
Cleaning
The exterior of the probe and cable should be cleaned only using a soft cloth moistened with water or isopropyl alcohol. The use of abrasive agent, strong detergents or other solvents may damage the probe.
Calibration Interval
The recommended calibration interval is one year. Adjustment should only be performed by qualified personnel. (A performance verification procedure is included in this manual.)
Service Strategy
Defective probes must be returned to a Teledyne LeCroy service facility for diagnosis and exchange. A defective probe under warranty will be replaced with a factory refurbished probe. A probe that is not under warranty can be exchanged for a factory refurbished probe. A modest fee is charged for this service. A defective probe must be returned in order to receive credit for the probe core.
Troubleshooting
If the probe is not operating properly the problem may be the way in which it is used. Before assuming the probe is defective, perform the following troubleshooting procedures:
1. Verify the test instrument is running the required firmware (4.3.0.0 or higher).
2. If the waveform is inverted, check that the arrow on the slider is in the direction
of the current flow.
3. If there is no signal, check that the slider is closed and locked.
17
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Returning a Product for Service
Contact your regional Teledyne LeCroy service center for calibration or other service. If the product cannot be serviced on location, the service center will give you a Return Material Authorization (RMA) code and instruct you where to ship the product. All products returned to the factory must have an RMA. Return shipments must be prepaid.
Teledyne LeCroy cannot accept COD or Collect shipments. We recommend air­freighting. Insure the item you’re returning for at least the replacement cost.
1. Remove all accessories from the probe. Do not include the manual.
2. Pack the probe in its case, surrounded by the original packing material (or
equivalent).
3. Label the case with a tag containing:
The RMA
Name and address of the owner
Probe model and serial number
Description of failure or requisite service
4. Package the probe case in a cardboard shipping box with adequate padding to
avoid damage in transit.
5. Mark the outside of the box with the shipping address given to you by Teledyne
LeCroy; be sure to add the following:
ATTN: <RMA code assigned by the Teledyne LeCroy>
FRAGILE
6. Insure the item for the replacement cost of the probe.
7. If returning a probe to a different country:
Mark the shipment as a “Return of US manufactured goods for warranty repair/recalibration.”
If there is a cost for the service, list the cost in the value column and the original purchase price “For insurance purposes only.”
Be very specific as to the reason for shipment. Duties may have to be paid on the value of the service.
18
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Warranty
Teledyne LeCroy warrants this oscilloscope accessory for normal use and operation within specification for a period of one year from the date of shipment. Spare parts, replacement parts and repairs are warranted for 90 days.
In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center. However, this will be done only if the product is determined by Teledyne LeCroy’s examination to be defective due to workmanship or materials, and the defect is not caused by misuse, neglect, accident, abnormal conditions of operation, or damage resulting from attempted repair or modifications by a non­authorized service facility.
The customer will be responsible for the transportation and insurance charges for the return of products to the service facility. Teledyne LeCroy will return all products under warranty with transportation charges prepaid.
This warranty replaces all other warranties, expressed or implied, including but not limited to any implied warranty of merchantability, fitness or adequacy for any particular purposes or use. Teledyne LeCroy shall not be liable for any special, incidental, or consequential damages, whether in contract or otherwise.
20
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
Certifications
Teledyne LeCroy certifies compliance to the following standards as of the date of publication. As standards evolve, these may no longer be current. Please see the Declaration of Conformity certificate shipped with your product.
EMC Compliance
EC Declaration of Conformity - EMC
The probes meet the intent of EC Directive 2014/30/EU for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities:
EN 61326-1:2013 EMC requirements for electrical equipment for measurement, control, and laboratory use.
ELECTROMAGNETIC EMISSIONS:
EN 55011/A1:2010 Radiated and Conducted Emissions Group 1 Class A
LECTROMAGNETIC IMMUNITY:
E
EN 61000-4-2:2009 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV vertical/horizontal coupling planes
1
2 3
4
EN 61000-4-3/A2:2010 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz - 2 GHz; 1 V/m, 2 GHz - 2.7 GHz
EN 61000-4-8:2010 Power Frequency Magnetic Field, 3 A/m, 50 Hz; 3 A/m, 60 Hz
1 To ensure compliance with the applicable EMC standards, use high quality shielded interface cables. 2 This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic
interference. 3 Emissions which exceed the levels required by this standard may occur when the probe is connected to a test object. 4 Meets Performance Criteria “B” limits of the respective standard: during the disturbance, product undergoes a
temporary degradation or loss of function or performance which is self-recoverable.
21
www. .com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Australia & New Zealand Declaration of Conformity - EMC
The probes comply with the EMC provision of the Radio Communications Act per the following standards, in accordance with requirements imposed by the Australian Communication and Media Authority (ACMA):
AS/NZS CISPR 11:2011 Radiated and Conducted Emissions, Group 1, Class A.
Safety Compliance
EC Declaration of Conformity – Low Voltage
The probes meet the intent of EC Directive 2014/35/EU for Product Safety. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities:
IEC/EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements
IEC/EN 61010-2-032:2012 Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-032: Particular Requirements for Hand-Held and Hand Manipulated Current Sensors for Electrical Test and Measurement.
22
www. .com
information@itm.com1.800.561.8187
Operator’s Manual
Environmental Compliance
End-Of-Life Handling
The probe is marked with this symbol to indicate that it complies with the applicable European Union requirements to Directives 2012/19/EU 2013/56/EU on Waste Electrical and Electronic Equipment (WEEE) and Batteries.
The probe is subject to disposal and recycling regulations that vary by country and region. Many countries prohibit the disposal of waste electronic
equipment in standard waste receptacles.
Restriction of Hazardous Substances (RoHS)
The product and its accessories conform to the 2011/65/EU RoHS2 Directive.
and
23
www. .com
information@itm.com1.800.561.8187
Loading...