Unauthorized duplication of Teledyne LeCroy documentation materials other than for
internal sales and distribution purposes is strictly prohibited. However, clients are
encouraged to duplicate and distribute Teledyne LeCroy documentation for internal
educational purposes.
Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Windows is a
registered trademark of Microsoft Corporation. Other product or brand names are
trademarks or requested trademarks of their respective holders. Information in this
publication supersedes all earlier versions. Specifications are subject to change
without notice.
CAUTION of damage to probe or instrument, or WARNING of hazard to health.
Do not apply around or remove from UNINSULATED HAZARDOUS LIVE
Equipment protected by double insulation or reinforced insulation.
CAUTION, hot surface.
Safety Instructions
Follow these instructions to keep the probe operating in a correct and safe condition.
Observe generally accepted safety procedures in addition to the precautions specified
here. The overall safety of any system incorporating this accessory is the responsibility of the assembler of the system.
Symbols
These symbols appear on the probe body or in documentation to alert you to important
safety considerations.
Attend to the accompanying information to protect against personal injury or
damage. Do not proceed until conditions are fully understood and met.
conductors which may render electric shock, electric burn, or arc flash.
Precautions
Comply with the following safety precautions to avoid personal injury or
damage to your equipment:
Use only as specified. The probe is intended to be used only with compatible Teledyne
LeCroy instruments. Using the probe and/or the equipment it is connected to in a
manner other than specified may impair the protection mechanisms.
Do not use the probe for measurements on Mains circuits. The probe should only be
applied around or removed from insulated limited energy circuit conductors that are
not directly connected to the mains.
1
www..com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
Do not overload; observe all ratings. To avoid electric shock or fire, do not connect the
current probe to any wire that carries voltages or currents that exceed the ratings of
the probe.
Connect and disconnect properly. Connect the probe to the measurement instrument
before connecting to the circuit/conductor being measured. Avoid damaging the cable
through excessive bending.
Never install or remove the probe on bare conductors which are energized. The
transformer core and shield are grounded but not insulated and may contact the
conductor when the locking lever is open.
Be careful not to damage the insulation surface when making measurements. Before
clamping to the conductor being measured, check that the insulation on the conductor
is undamaged, and take care not to damage the insulation when clamping the
conductor. Any damage to the insulation could cause an electric shock.
Use only indoors andwithin the operational environment listed. Do not use in wet or
explosive atmospheres.
Do not remove the probe's casing. Touching exposed connections may result in electric
shock.
Keep product surfaces clean and dry.
Comply with the maximum input current vs. frequency derating when measuring
current that includes a high frequency component. Using the probe at high frequencies
or strong magnetic fields may cause the device to become abnormally hot, resulting in
fire, equipment damage or burns.
Do not operate with suspected failures. Before each use, inspect the probe and
accessories for any damage such as tears or other defects in the probe body, cable
jacket, accessories, etc. If any part is damaged, cease operation immediately and
sequester the probe from inadvertent use.
NOTE: Depending on the amplitude and frequency of the current being measured, the
sensor head may emit a resonant sound. This sound may also occur during
demagnetizing operation, but it does not represent a malfunction (device failure).
2
www..com
information@itm.com1.800.561.8187
Operator’s Manual
CP150
CP150-6M
CP500
Amplitude Accuracy
±1.0 % of reading at 0-150 Amp
±2.0% of reading at 150-300 Apeak
±1.0 % of reading at 0-500 Amp
±2.0% of reading at 500-700 Apeak
Bandwidth (probe only)
DC to 10 MHz
DC to 5 MHz
DC to 2 MHz
Introduction
The CP150 and CP500 current probes are designed for easy, highly accurate current
measurements. The compact probes offer wide bandwidth with over-current protection.
The probes utilize a combination of Hall-effect and transformer technology which
enables measurements to be made on DC, AC and impulse currents.
The CP150 has a 10 MHz bandwidth and is designed to measure continuous currents
up to 150 Amps. The CP150-6M has a 5 MHz bandwidth and is designed to measure
continuous currents up to 150 Amps.
NOTE: All CP150 probes, regardless of cable length, will indicate default probe
bandwidth of 10 MHz in the Attributes section of the Probe dialog.
The CP500 has a 2 MHz bandwidth and is designed to measure continuous currents up
to 500 Amp.
The probes can be used with any Teledyne LeCroy instrument with a ProBus interface,
or by using a ProBus interface adapter. With the ProBus interface, the probe becomes
an integral part of the measuring instrument. The bandwidth limit, Auto Zero and
Degauss functions are all controlled from the instrument’s touch screen user interface.
The interface provides power to the probe, so no external power supply is needed.
Specifications
Specifications are subject to change without notice.
Warranted Characteristics
3
www..com
information@itm.com1.800.561.8187
CP150 and CP500 Current Probes
CP150
CP150-6M
CP500
Max Continuous Input
Current
150 Arms
500 Arms
Max Peak Current
300 Apeak non-continuous
700 Apeak non-continuous
Intended Output Load
1 MΩ
1 MΩ
Measured Conductors
Insulated Conductors only
Insulated Conductors only
Sensitivity *
100 mA/div to 100 A/div
100 mA/div to 200 A/div
CP150
CP150-6M
CP500
Output voltage
0.01 V/A
0.01 V/A
Sensitivity Temperature
Coefficient
±2% or less (0⁰C to +40⁰C)
±2% or less (0⁰C to +40⁰C)
Noise (at 20 MHz BW)
6.0 mArms or less
8.0 mArms or less
Rise Time
≤ 35 ns
≤ 70 ns
≤ 175 ns
CP150
CP150-6M
CP500
Operating Temperature
0 ⁰C to 40 ⁰C
0 ⁰C to 40 ⁰C
Operating RH
≤ 80% (non-condensing)
≤ 80% (non-condensing)
Storage Temperature
-10 ⁰C to 50 ⁰C
-10 ⁰C to 50 ⁰C
Storage RH
≤ 80% (non-condensing)
≤ 80% (non-condensing)
Max Altitude
2000 m
2000 m
Effect of External
Magnetic Field
Maximum of 150 mA in a DC/60 Hz,
400 A/m magnetic field
Maximum of 800 mA in a DC/60 Hz,
400 A/m magnetic field
Nominal Characteristics
* Values are based on oscilloscopes with 1 mV/div sensitivity. Numbers will be higher on instruments with
lower sensitivity.
Typical Characteristics
Environmental Characteristics
4
www..com
information@itm.com1.800.561.8187
Loading...
+ 18 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.