Unauthorized duplication of Teledyne LeCroy documentation materials other than for
internal sales and distribution purposes is strictly prohibited. Clients are permitted to
duplicate and distribute Teledyne LeCroy documentation for their own internal training
purposes.
Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Windows is a
registered trademark of Microsoft Corporation. Other product or brand names are
trademarks or requested trademarks of their respective holders. Information in this
publication supersedes all earlier versions. Specifications are subject to change without
notice.
Figure 12, ÷10 Effective Gain Indicator .............................................................................40
Figure 13, Common Mode and Differential Mode Range .................................................47
Tables
Table 1, Dynamic ranges and input capacitance at different attenuator settings ............12
Table 2, Recommended settings to obtain desired equivalent Volts/Division .................14
Table 3, Performance Verification Test Equipment ..........................................................24
Table 4, Adjustment Test Equipment ................................................................................35
ii
www..com
information@itm.com1.800.561.8187
Operator’s Manual
HIGH VOLTAGE, risk of electric shock.
CAUTION of Potential damage to probe or instrument it is connected to, or
il conditions are fully
ELECTROSTATIC DISCHARGE (ESD) HAZARD. The probe is susceptible to damage if
DOUBLE INSULATION
PROTECTIVE (EARTH) TERMINAL
Safety Instructions
This section contains instructions that must be observed to keep this oscilloscope
accessory operating in a correct and safe condition. You are required to follow generally
accepted safety procedures in addition to the precautions specified in this section. The
overall safety of any system incorporating this accessory is the responsibility of the
assembler of the system.
Symbols
These symbols may appear on the probe body or in this manual to alert you to important
safety considerations.
WARNING of potential bodily injury. Attend to the accompanying information to
protect against personal injury or damage. Do not proceed unt
understood and met.
anti-static measures are not taken.
Precautions
To avoid personal injury, or damage to the probe or test instrument, comply with the
following safety precautions.
Use product only as specified.
Connect and disconnect properly. Connect probe to the measurement instrument before
connecting the test leads to a circuit/signal being tested.
Use only accessories compatible with the probe. Use only accessories that are rated for
the application. Ensure connections between probe input leads and probe accessories are
secure before connecting them to a voltage source.
Do not overload. To avoid electric shock, do not apply any potential that exceeds the
maximum rating of the probe and/or the probe accessory, whichever is less. Observe all
terminal ratings of the instrument before connecting the probe.
Be careful not to damage the insulation surface when making measurements.
1
www..com
information@itm.com1.800.561.8187
AP033 Active Differential Probe
Use only within operational environment listed. Do not use in wet or explosive
atmospheres. Keep product surfaces clean and dry. Use indoors only.
Handle with care. Probe accessory tips are sharp. They can puncture skin or cause other
bodily injury if not handled properly.
Keep fingers behind the finger guard of the probe accessories.
Do not operate with suspected failures. Before each use, inspect the probe and
accessories for any damage such as tears or other defects in the probe body, cable jacket,
accessories, etc. If any part is damaged, cease operation immediately and sequester the
probe from inadvertent use.
Operating Environment
Only use the product within this operating environment:
Temperature: 0° to 50° C
Humidity: Maximum relative humidity 90 % for temperatures up to 31° C decreasing
linearly to 50% relative humidity at 50° C
Altitude: Up to 10,000 ft (3,048 m)
2
www..com
information@itm.com1.800.561.8187
Operator’s Manual
Overview
Description
The AP033 is a wide band differential active probe. The probe features low noise, low input
capacitance, high common mode rejection, and FET-buffered inputs in the probe head. User
selectable attenuation and offset give the probe flexibility to measure a large range of signal
amplitudes. Plug-on attenuator and AC coupling accessories further extend the application
range. Interconnect accessories included allow connection to surface mount and throughhole components with minimal signal degradation. The input receptacles in the probe head
are compatible with standard 0.025 in. (0.635 mm) square pins, providing a convenient lowcost method of creating device characterization test fixtures.
The probe is powered directly from a Teledyne LeCroy oscilloscope through the ProBus®
interface. The ProBus interface also allows local control of the probe through the
oscilloscope user interface and remote control through the interface buses, (GPIB, RS-
232). The optional ADPPS power supply allows the AP033 to be used with other
instruments such as spectrum analyzers, network analyzers, and oscilloscopes without
ProBus interface.
Applications
The AP033 is ideal for acquiring high speed differential signals such as those found in disk
drive read channels, differential LAN, video, etc. It can also be used with spectrum
analyzers to acquire signals in some RF systems (for example, balanced IF mixers in hand
held cellular telephones). The high impedance characteristics of both inputs allow the
probe to be used as a FET probe to make single-ended measurements in digital systems
without introducing a ground loop, as a conventional FET probe would.
Standard Accessories
Hard Case
÷10 Plug-on Attenuator
Plug-on AC Coupler
Probe Connection Accessory Kit:
Flex Lead Set (1)
Mini Clip, 0.8 mm (3)
Mini Clip, 0.5 mm (2)
Ground Lead (1)
Offset Pins, Round (4)
Square Pin Header Strip (1)
Manual, AP033 Active Differential Probe Operator’s Manual
3
www..com
information@itm.com1.800.561.8187
AP033 Active Differential Probe
Teledyne LeCroy
Part Number
PK033
Accessory Kit, AP033 without AC Coupler and Attenuator
AP03X-FLEX-LEAD
Flex Lead
405400003
Offset Pin, package of 10
AP033-ATTN
Plug-On Attenuator, /10, AP033
AP03X-AC-COUPLER
Plug-On AC Coupler
ADPPS
Power Supply Adapter for Non-ProBus Oscilloscopes
Optional Accessories
Description
Oscilloscope Software Compatibility
Teledyne LeCroy LC series oscilloscopes must have legacy software version 8.1.0 or higher
loaded. All X-Stream oscilloscopes are compatible.
To verify the software version installed on the oscilloscope, either:
•Press the SHOW STATUS button on the front panel (where available), then select
System.
•Choose Help > About or Support > About from the oscilloscope menu bar.
NOTE: When the probe is used with earlier versions of legacy software, Offset can only be
controlled through the buttons on the probe body. Also, the scale factor will be displayed
incorrectly in some modes.
To upgrade the software in your oscilloscope, go to:
4
www..com
information@itm.com1.800.561.8187
Operator’s Manual
Specifications
Nominal Characteristics
Nominal characteristics describe parameters and attributes that have are guaranteed by
design, but do not have associated tolerances.
(1X Gain): ±400 mV(÷1 Attenuation)
±4 V (÷10 Attenuation)
±40 V (÷100 Attenuation)
Common Mode Range: ±4.2 V (÷1 Attenuation)
±42 V (÷10 Attenuation)
±42 V (÷100 Attenuation)
2
Maximum Input Voltage:
±42 V either input from ground
Warranted Electrical Characteristics
Warranted characteristics are parameters with guaranteed performance. Unless
otherwise noted, tests are provided in the Performance Verification Procedure for all
warranted specifications.
3
LF Gain Accuracy: 2% into 50 Ω load
Common Mode Rejection Ratio:
4
, measured at 1 kHz with 0 V offset
(Probe head grounded, DC Coupled, ÷1 attenuation without external attenuator):
1. From combination of gain, internal and external attenuation.
2. Use external plug-on ÷10 attenuator for ÷100.
3. Output impedance 50 Ω, intended to drive 50 Ω. Add uncertainty of termination impedance to accuracy.
4. Teledyne LeCroy measures CMRR with a fixture that connects the probe tip ground to the signal source
ground. This method is necessary to obtain a reproducible CMRR measurement. Not grounding the probe
tip can actually improve CMRR by allowing some of the common mode signal to be impressed across the
entire length of the probe cable instead of from probe tip to probe ground, however, this CMRR
improvement depends on proximity to probe cable ground, and is therefore non-reproducible.
5
www..com
information@itm.com1.800.561.8187
AP033 Active Differential Probe
Typical Electrical Characteristics
Typical characteristics are parameters with no guaranteed performance. Tests for typical
characteristics are not provided in the Performance Verification Procedure.
Output Impedance: 50 Ω nominal, intended to drive 50 Ω
Harmonic Distortion:
(3rd order distortion): -52 dB below fundamental (200 mV
(3rd order intercept): +15 dBm (at 100 MHz measured at output)
AC Coupling LF Cutoff: 1.6 Hz (-3dB)
6
output at 100 MHz)
p-p
www..com
information@itm.com1.800.561.8187
Figure 1, Typical CMRR Graphs
General Characteristics
Temperature:0° to 50° C Operating
-40° to 75° C Storage
Operator’s Manual
Input Connectors:Compatible with 0.025” (0.635 mm) square pins
0.036” (0.91 mm) maximum diameter (for round pins)
Power Requirements:Powered from oscilloscope through ProBus interface or with
ADPPS power supply.
Dimensions
Control Housing:Length:3.625 in (9.2 cm)
Width: 1.50 in (3.8 cm)
Height: 1.00 in (2.5 cm)
Head:Length: 4.0 in (10.2 cm) w/o Attenuator or AC Coupler Width: 0.89 in (2.25 cm)
Height: 0.625 in (1.6 cm)
Cable:Length: 42 in (106.7 cm) Diameter: 0.275 in (7.0 mm)
Weight:Probe only: 6.4 oz (0.18 kg) Shipping: 2 lbs, 8.4 oz (1.15 kg)
7
www..com
information@itm.com1.800.561.8187
AP033 Active Differential Probe
CAUTION: The input circuits in the AP033 incorporate components that protect
Operation
the probe from damage resulting from electrostatic discharge (ESD). Keep in
mind that this is an active probe, and it should be handled carefully to avoid
damage. When using the AP033, you are advised to take precautions against
potential instrument damage due to ESD.
Connecting the Probe to the Test Instrument
Attach the probe output connector to the oscilloscope input connector. When used with
a Teledyne LeCroy oscilloscope equipped with ProBus inputs, the oscilloscope will
recognize the probe, set the input termination to 50 Ω, and activate the probe control
functions on the oscilloscope user interface. These appear on the AP033 dialog behind
the input channel setup dialog.
Figure 2, channel setup dialog with AP033 probe dialog behind it.
To use the AP033 Active Differential Probe with instrumentation not equipped with a
ProBus interface, it is necessary to use the ADPPS Power Supply. Attach the ADPPS
connector to the probe output connector. The output connector of the ADPPS is a
standard male BNC that can be directly connected to another instrument. If necessary,
the output of the ADPPS can be interconnected with a 50 Ω coaxial cable. To minimize the
effect of skin loss, this cable should be 1 m or less in length. The AP033 is designed to
drive a 50 Ω load. The gain will be uncalibrated if the output is not correctly terminated. If
you are using the probe with an instrument with a high input impedance, place a 50
instrument input before attaching the ADPPS..
Connecting the Probe to the Test Circuit
At the probe tip, two inputs and a ground connection are available for connecting the
probe to a circuit under test. For accurate measurements, both the + and – inputs must
always be connected to the test circuit. The ground connection is optional. Positive
voltages applied to the + input relative to the – input will deflect the oscilloscope trace
toward the top of the screen.
Positive voltages applied to the + input relative to the – input will deflect the oscilloscope
trace toward the top of the screen.
8
www..com
information@itm.com1.800.561.8187
Operator’s Manual
To maintain the high performance capability of the probe, user care in connecting the
probe to the test circuit. Increasing the parasitic capacitance or inductance in the input
paths may introduce a “ring,” or slow the rise time of fast signals. To minimize these
effects, use the shortest length possible when connecting the probe to the circuit under
test. Input leads that form a large loop area (even shielded coaxial cables) will pick up any
radiated magnetic field that passes through the loop, and may induce noise in the probe
inputs. Because this signal will appear as a differential mode signal, the probe’s common
mode rejection will not remove it. You can greatly reduce this effect by using short
interconnection leads, and twisting them together to minimize the loop area.
High common mode rejection requires precise matching of the relative gain or
attenuation in the + and – input signal paths. Mismatches in additional parasitic
capacitance, inductance, delay, and a source impedance difference between the + and –
signal paths will lower the common mode rejection ratio. Therefore, it is desirable to use
the same length and type of wire and connectors for both input connections. When
possible, try to connect the inputs to points in the circuit with approximately the same
source impedance.
If AC coupling is desired, install the AC coupling accessory on the probe tip before
connecting it to the test circuit. The low-frequency cutoff (–3 dB point) of the AC coupler
is approximately 1.6 Hz.
If the voltage in the test circuit requires more than ÷10 attentuation, add the external ÷10
attenuator to the probe tip. If using both the external attenuator and AC coupler, install
the attenuator on the probe tip first, then install the AC coupler on the attenuator input.
In addition to being compatible with the included lead set, the probe input connectors
will mate with standard 0.025 in. (0.635 mm) square pins in any rotational orientation. To
avoid damaging the input connectors, do not attempt to insert connectors or wire larger
than 0.036 in. (0.91 mm) in diameter. Avoid rotating square pins after they are inserted
into the input connectors.
The included accessories simplify the task of connecting the probe to the test circuit:
•Use the small (0.5 mm) mini clips with the flexible lead set when connecting to
fine-pitch surface mount IC leads.
•Use the larger (0.8 mm) mini clips to connect to through-hole leaded
components.
•Use the offset round pins for hand-held probing applications. Reposition the pins
by rotating them to obtain the required spacing.
9
www..com
information@itm.com1.800.561.8187
AP033 Active Differential Probe
CAUTION: Do not use the attenuator encoding receptacle (unmarked socket
near the
will not provide adequate grounding and may result in an incorrect effective gain
indication.
M
M M
Probe Input Loading
Attaching any probe to a test circuit will add some loading. In most applications, the high
impedance of the AP033 Active Differential Probe inputs imparts an insignificant load to
the test circuit. However at very high frequencies, the capacitive reactance of the probe’s
input capacitance may load the circuit enough to affect measurement accuracy. The
equivalent model of the probe input circuits is shown below:
Figure 3, AP033 Equivalent Input Model
Grounding the Probe
The single lead along with one of the larger (0.8 mm) mini clips can be used to ground the
probe to the test circuit. Insert the pin end of the lead into the receptacle marked:
In many cases it is not necessary to ground the probe to the circuit under test. However,
if the test circuit is isolated from earth ground, it is usually necessary to connect the
probe ground to a point in the circuit. Grounding test circuits that are referenced to earth
ground may improve the fidelity of high frequency components in the waveforms. The
potential for improvement with grounding will vary depending on the common mode
source impedance. However, connecting the probe ground to a circuit that is referenced
to earth ground can create a ground loop that may add noise to low amplitude signals.
The rejection of high frequency common mode signals is improved when the probe head
is ungrounded.
The best recommendation for connecting or not connecting the probe ground is to try
both configurations and select the one that performs the best.
10
– input) to ground the probe. Connection to the encoding receptacle
www..com
information@itm.com1.800.561.8187
Operator’s Manual
NOTE: The AP033 Active Differential Probe transmits the measured signal differentially
through the probe cable. This essentially eliminates signal degradation from ground loop
effects within the probe. However, creating a ground loop may introduce signal
distortions in the test circuit itself, or in any coaxial cable between the ADPPS power
supply and the test instrument.
Selecting the Proper Range
The AP033 has two independent controls that set the common mode range and
equivalent volts/division. The probe gain can be set to x1 or x10. The gain control (GAIN)
only affects the differential mode range of the probe. A separate input attenuation
control (ATTEN) sets the probe input attenuator to either ÷1 or ÷10. See Figure 4.
Figure 4, AP033 Block Diagram
The attenuator is located at the probe input and, therefore, affects both the differential
mode and common mode ranges. (Refer to the Reference Section for definitions of
Common Mode and Differential Mode.) By using the plug-on attenuator, you can extend
the attenuation range to ÷100. The maximum ranges are given in Table 1.
11
www..com
information@itm.com1.800.561.8187
AP033 Active Differential Probe
Table 1, Dynamic ranges and input capacitance at different attenuator settings
Common Mode Range
Differential Mode Range with x1 Gain*
Differential Mode Range with x10 Gain*
Input Capacitance, each side to Ground
Input Capacitance each side to Ground
with AC Coupler
*Offset moves the center point of this range.
÷1 Attenuation ÷10 Attenuation ÷100 Attenuation
± 4.2 V ± 42 V ± 42 V
± 400 mV ± 4 V ± 40 V
± 40 mV ± 400 mV ± 4 V
6.0 pF 3.0 pF 1.9 pF
7.3 pF 4.3 pF 3.2 pF
When you are using a differential probe or amplifier, be careful not to exceed the
common mode range. Because the common mode signal is rejected by the probe, and
not displayed, changes in the amplitude of the common mode component are not
apparent. Exceeding the common mode range may introduce distortion into the AP033
output. To reduce the possibility of errors caused by exceeding the common mode range,
the probe monitors the input voltage. If the common mode range is exceeded when the
÷1 input attenuator is selected, the probe will automatically switch to the ÷10 attenuator. If the voltage on either input exceeds 55 volts, both the ÷1 and ÷10 attenuator lights will alternately flash to alert the user to the over-range condition.
There are two combinations which result in x1 effective gain. A ÷10 attenuation with x10
gain results in higher common mode range and lower input capacitance, but it increases
the noise referred to the input. Conversely, ÷1 attenuation with x1 gain reduces the
noise at the expense of less common mode range and greater input capacitance.
Circuitry in the probe decodes the effective gain of the probe based on the settings of the
gain, internal attenuation and the presence of the external attenuator. The resulting
effective gain is displayed on the probe front panel.
12
www..com
information@itm.com1.800.561.8187
Loading...
+ 38 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.