HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us
free from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. The Class S-2 Micro-Fuel Cell is warranted for 6 months from the date of shipment from Teledyne. Components manufactured
by others bear the warranty of their manufacturer. This warranty does not cover defects
caused by wear, accident, misuse, neglect or repairs other than those performed by
Teledyne or an authorized service center. We assume no liability for direct or indirect
damages of any kind and the purchaser by the acceptance of the equipment will assume
all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument is intended to be used as a tool to gather valuable data. The
information provided by the instrument may assist the user in eliminating potential
hazards caused by the process that the instrument is intended to monitor; however, it is
essential that all personnel involved in the use of the instrument or its interface with
the process being measured be properly trained in the process itself, as well as all
instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and can be misused. In particular, any
alarm or control system installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy must be provided by the user or specifically requested of Teledyne when
the order is placed.
The purchaser must be aware of the hazardous conditions inherent in the process(es)
he uses. He is responsible for training his personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments (TAI), the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the
manufacturer or his agents is to be construed as a warranty of adequate safety control
under the user’s process conditions.
Spare Parts List................................................................... 15
Drawing List ........................................................................ 16
Material Safety Data Sheet ................................................. 17
Teledyne Analytical Instruments
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
Introduction
The Teledyne Analytical Instruments Model TURBO2 is a portable
trace oxygen analyzer which does not need an external power source to
operate, and can be reliably calibrated without the use of additional calibration gases.
The instrument provides for trace oxygen analysis in measuring ranges
of 0-100, 0-1,000 and 0-10,000 PPM (full scale), plus a special calibration
range that encompasses the known oxygen concentration of atmospheric air
(209,000 PPM).
2
Sample oxygen is read from an extremely accurate integral meter (0.5%
linearity) whose range of measurement is determined by the position of the
range selector switch. The 100-division meter scale and the multiplying
factor indicated by the position of the range switch determine the full scale
oxygen sensitivity of the instrument.
Sample gas is introduced and vented via a pair of quick disconnect
fittings that feature integral shutoff valves which automatically close when
the mating male fitting is withdrawn. The fittings are an integral part of the
measuring cell manifold so that internal sample passage volume is at an
absolute minimum. Sample flow control, although not critical (0.1 to 10
liters/min.), is either accomplished with accessory equipment or by the
addition of a pump and flowmeter (TURBO
2
P).
Method of Analysis
The sample oxygen is measured by a unique electrochemical transducer
which functions as a fuel cell; in this instance, the fuel is oxygen. Oxygen
diffusing into the cell reacts chemically to produce an electrical current that is
proportional to the oxygen concentration in the gas phase immediately
adjacent to the transducer’s sensing surface. The Micro-fuel Cell is a Class
S-2, designed for a quick recovery time from air to low PPM levels of
oxygen.
Teledyne Analytical Instruments
1
TURBO
2
Portable Trace Oxygen Analyzer
The linear, but minute, signal produced by the transducer from trace
oxygen is amplified by a two stage amplifier. The dual stages of amplification provide enough gain to drive the 0-100 microampere meter and thermistor controlled network utilized to compensate for the positive temperature
coefficient of the transducer.
Features
Micro-Fuel Cell:The Micro-fuel Cell (U.S. Pat. #s 3,767,552 and
3,668,101) is a sealed electrochemical transducer with no electrolyte to
change or electrodes to clean. When the cell reaches the end of its useful life,
it is merely thrown away and replaced, as one would replace a worn-out
flashlight battery. The life of the cell is warranted by TAI (see Cell Warranty, page 11).
Reliable Calibration:The unique qualities of the Micro-fuel Cell
allows the user to calibrate the instrument with the most economical, reliable,
abundant, standardization gas there is—atmospheric air.
The special CAL range of the instrument features a mark that coincides
with the 209,000 PPM oxygen concentration of air. By drawing air through
the instrument (see Calibration, page 6) reliable calibration can be achieved.
After the electronics have been properly zeroed (a one-time factory
operation), the instrument cannot produce an output indication in the absence
of oxygen; therefore, there is no need for a zero standardization gas.
Integral Power Supply:For pump option 4, the differential power
requirement (+ and – 3.6 VDC) of the instrument amplifier is furnished by
two internally mounted 750 milliampere•hour NiCad batteries. Fully
charged, these batteries will provide enough power to operate the instrument
continuously for a period of about 45 days with the pump option OFF. With
the pump option ON, the batteries will provide continuous operation for
about 8 hours. With the pump option OFF, an overnight charge of 14–16
hours on a one-month duty cycle should keep the original batteries supplied
usable for many years. With the pump option ON, charge 14–16 hours on
an 8-hour duty cycle.
An integral charging circuit and a detachable power cord are provided
so that the batteries may be recharged from any 50 or 60 cycle, 105 to 125
volt (90 to 110 volt optional), outlet.
2
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
The instrument is designed to either sample or have its batteries recharged. Both operations cannot be carried out simultaneously. TAI has
deliberately interlocked the circuitry so that both operations cannot be carried
out at the same time.
Only when the selector switch is placed in the OFF position will the
neon lamp on the back plate of the Model TURBO
light up to indicate
2
power to the battery charging circuit.
To determine the state of the rechargeable batteries, turn the Range
Selector Switch counterclockwise to the battery test position and hold there
(the switch is spring loaded). Watch the meter indicator: if it does not stay
within the battery limits, the batteries need to be recharged. Release the
Range Selector Switch and it will automatically return to the OFF position.
2
Accuracy and Response:The Model TURBO
provides monitoring
2
accuracies of ±2% of full scale or ±5 PPM, whichever is greater, at constant
temperature. A ±5% of reading accuracy is achievable throughout the operating temperature range of 30 to 125° F.
With a sample flowrate of 150 cc/min., 90% response is achieved in 5–
10 seconds in the 0–10, 0–100, 0–1000, and CAL range switch positions.
Compact Packaging:The instrument is housed in 10¹/
16 " W × 7½ "
H × 6¾ " D aluminum case that is equipped with a carrying handle and foot
pads. When in use, the analyzer should be placed in an upright position on a
level surface (off-level positioning will detract from meter accuracy).
See Fig.1.
The cell block and amplifier PC board are housed inside the analyzer.
The back panel of the analyzer contains the AC power receptacle, the amber
charge light, fuse (½A: 110V; ¼A: 220V), and quick-connect fittings for
bringing the sample into and out of the sensor assembly.
Pump:The TURBO
P has an integral pump used to help draw
2
See Figs. 2 and 3.
sample through the analyzer.
Teledyne Analytical Instruments
3
TURBO
2
Portable Trace Oxygen Analyzer
4
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
Supporting Equipment and Services
Sampling Equipment
The customer must provide a means of controlling the pressure and
flowrate of the sample gas. For positive pressure applications, TAI suggests
a simple throttle valve installed in the sample line between the sample point
and the analyzer. The flowrate should be limited to between 0.1 and 10
liters/min.
2
IMPORTANT:Use pressure regulators with a metallic diaphragm ONLY.
Regulators with organic or plastic diaphragms are permeable to oxygen
and, if used in the sampling system, will lead to high oxygen readings.
For atmospheric pressure sampling, the TURBO
P has a pump and
2
flow control valve downstream of the analyzer to draw (rather than push) the
sample through the instrument.
TAI supplies three (3) male disconnect fittings with the instrument: one
for installation of the customer’s sample line, one to be used to open the vent
fitting of the instrument, and one (equipped with a plastic tube) for drawing
air through the unit for calibration purposes.
Do not place any restriction on the line between the sample point
and the analyzer—as a partial vacuum would then be drawn on the cell.
Since the cell is a partial pressure sensitive device, any oxygen readings
taken under these conditions would be erroneous. In addition, vacuums in
excess of one-third of an atmosphere may damage the cell.
Teledyne Analytical Instruments
5
TURBO
2
Portable Trace Oxygen Analyzer
Power Service
A source of single phase, 105 to 125 VAC (90 to 110 VAC optional),
50 or 60 Hz power, capable of delivering a maximum of ¼ A of current will
be periodically required to recharge the instrument’s battery power supply.
An 8-foot, UL approved, 3-wire detachable power cord is provided with the
instrument and should be stored in a safe place when not in use. The
TURBO
can be optionally furnished with 220 VAC, 50 or 60 Hz charging
2
power or 100 VAC, 50 or 60 Hz charging power.
6
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
Operation
WARNING: The TURBO2 is NOT intrinsically safe and must not be used in or
brought into hazardous areas.
The TURBO2 is supplied completely assembled and ready for instant
use. The Micro-fuel Cell is in place within the manifold, and prior to shipment the manifold was purged with an inert gas to eliminate all but traces of
oxygen from the internal sampling system. The integral shut-off valves in the
quick disconnect sample fittings, if not disturbed, will maintain this inert
atmosphere within the manifold indefinitely. This can be demonstrated by
placing the Range Selector Switch in the 0-1000 PPM position.
2
When the range selector is moved from the OFF position, power to the
instrument circuitry is established. The meter will instantly respond to the
residual oxygen within the integral sample passages.
It is impossible to achieve a perfect seal of the internal sample system,
and what the meter is indicating is the diffusion/consumption balance point
of the internal sample system and the Micro-fuel cell.
This balance point, with a properly calibrated instrument, is always
within the limits of the 0-100 range. If the reading climbs off the limits of this
scale, a leak in the manifold assembly is indicated.
To extend cell life and minimize the time required to make the next
analysis, the instrument should always be purged with sample or an
inert gas prior to being taken out of service for standby or storage.
Teledyne Analytical Instruments
7
TURBO
2
Portable Trace Oxygen Analyzer
Calibration
The inherently constant output of the cell during its useful life precludes
a definitive calibration cycle. TAI feels that the interval between calibrations
should be dictated by your application. If the instrument is being used to
certify the oxygen content of a product for delivery, then, a calibration prior
to certification would be in order. If not, a biweekly or monthly calibration is
adequate. However, the rapid response of the S-2 cell allows more frequent
air calibrations than possible with earlier type instruments.
NOT E: Do not calibrate the instrument unless there is a trace oxygen gas
readily available for purging immediately following calibration.
To calibrate the instrument:
1) Stand the instrument upright on a level surface, and with the
Range Selector Switch in the OFF position, check the alignment
of the meter pointer with the zero mark on the scale. Use the
mirror to eliminate parallax, and adjust the screw on the face of
the meter, if necessary, until the pointer and zero mark are in
precise coincidence.
2) Move the Range Selector Switch to the CAL position.
3) Install the plastic tube equipped male disconnect fitting in either
of the analyzer’s sample ports, and a blank disconnect fitting in
the other port (direction of sample flow is of no importance). A
pump is recommended on the plastic tube. Pump the tube until
the meter reading is stable.
WARNING: DO NOT ATTEMPT TO DRAW SAMPLE THROUGH THE ANALYZER
ORALLY.
The sensor electrolyte is caustic. Do not let it come in contact with
skin. If it does, immediately flush affected area with water. Consult the Emergency First Aid procedures in the Material Safety
Data Sheet at the end of the manual. Do not attempt to open or
repair the sensor. Leaking or exhausted sensors should be disposed of in accordance with local regulations. Consult the Material Safety Data Sheet at the end of the manual.
4) Unlock and adjust the SPAN potentiometer until the meter
pointer is in coincidence with the CAL mark on the meter scale.
Be sure to relock the control after the adjustment is made.
5) Immediately after step 4, disconnect the tubing-equipped
calibration fitting, and plug in either the sample or a source of
inert gas.
8
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
If the instrument is to be used for sampling after the calibration procedure has been completed, follow the decreasing oxygen reading by positioning the Range Selector Switch so that the meter gives the best possible
resolution of the oxygen. Do not attempt to actually take a reading until
the meter indication stabilizes.
If, on the other hand, the instrument is not to be used immediately after
calibration, and a low PPM oxygen gas is being employed as a purge, allow
the manifold to be purged overnight, and then disconnect both male fittings.
Always disconnect the source fitting first and then the vent fitting
immediately after.
Positive Pressure Sampling
When connecting the instrument to a positive pressure sample source,
always proceed as follows:
2
1) Before making any connections to the instrument, establish a
flowrate in the sample line of from 0.1 to 10 liters/min. Allow the
sample to vent to atmosphere long enough to purge the line free
of air.
2) Install the vent fitting first, and then the sample source fitting. Be
prepared to make the connections in rapid order, so that
atmospheric diffusion time through the vent fitting is held to a
minimum.
When disconnecting the instrument, reverse the procedure: source
fitting first, and then vent fitting.
The objective of the connection–disconnection procedure is to obviate
the possibility of pressurizing the manifold. If a flowing sample is connected
to the manifold without the vent fitting in place, the pressure in the manifold
will rise and equal the sample pressure immediately, possibly causing a
manifold leak.
Teledyne Analytical Instruments
9
TURBO
2
Portable Trace Oxygen Analyzer
Atmospheric Pressure Sampling
The TURBO
P is equipped with a sample pump and flowmeter to
2
draw sample through the analyzer in the case of sample at atmospheric or
slightly less than atmospheric pressure.
The pump power switch is located on the back panel of the analyzer.
During calibration and normal operation, set the flowmeter to midrange, and
turn the pump on during step 3 of the calibration procedure.
10
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
Maintenance
Recharging the Battery
TAI suggests that an overnight recharge be performed every 4 weeks of
continuous use. To recharge the batteries:
1. Move the Range Selector Switch to OFF and connect the power
cord to a convenience outlet. The amber charge lamp (back of
case) will light during charging. The integral charging circuit will
automatically energize and regulate the battery charging current
when the Range Selector Switch is in the OFF position and the
AC cord is plugged into the power line.
2
CAUTION: Do not turn the Range Selector Switch to either BATT TEST or any
of the operating range positions while the unit is connected to AC
power. Doing so may cause failure of the integrated circuit.
2. When recharging is completed, unplug the unit from the AC
outlet. Turn the Range Selector Switch to the operating position
and then to the BATT TEST position. Meter reads 60%-80% fs.
NOT E : The BATT TEST position will not give a reliable indication of the
battery charge immediately after a charge cycle. Allow the unit to run
for a while before testing the batteries.
If the instrument is stored with the Range Selector Switch in the OFF
position (charge cord disconnected), the period of time between charge
periods is extended from one month to four months. However, do not leave
it longer than this time period.
Teledyne Analytical Instruments
11
TURBO
2
Portable Trace Oxygen Analyzer
Routine Maintenance
Beyond adhering to a battery recharge schedule, no routine maintenance is required, as there are no moving parts in the instrument other than
the meter movement. The Micro-fuel Cell is a sealed, modular component
that should be replaced only when expired or faulty.
Cell Replacement
The characteristics of the Micro-fuel Cell are similar to those of a
mercury battery in that both provide an almost constant output through their
useful life, and then fall off sharply towards zero at the end. If the sample
being analyzed has a low oxygen concentration, cell failure will probably be
indicated by the inability to properly calibrate the analyzer. You will find that
very little adjustment of the SPAN potentiometer will be required to keep the
analyzer calibrated properly during the duration of a given cell’s useful life.
If many turn adjustments clockwise are required to calibrate the instrument,
or calibration cannot be achieved within the range of the control, the cell
should be replaced.
To offset the possibility of not having a replacement cell available when
it is needed, TAI recommends that a spare cell be purchased shortly after the
instrument is placed in service, and each time the cell is replaced thereafter.
The spare cell should be carefully stored in an area that is not subject to
large variations in ambient temperature (75°F nominal), and in such a way as
to obviate any possibility of incurring damage. Do not open the cell pack-age until the cell is to be actually used. If the cell package is punctured and
air permitted to enter, the cell will immediately start to react to the presence
of oxygen.
To replace the Micro-fuel Cell:
1. Loosen the 2 screws (designated by arrows) on the back panel
(see outline dwg. C-45254 or C-57975).
See Figure 2.
2. Remove the screw located on each side of the top cover and
gently slide the cover up and off.
3. Remove the 3 wing nuts on top of the cell block and remove the
cell cap. Carefully remove the old cell from the cell cap and
dispose of it in accordance with local regulations (see the
Material Safety Data Sheet at the end of the manual).
See Figure 3.
12
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
4. Remove the new cell from its package, and carefully remove the
shorting wire. Do not touch the silver/gold colored sensing
surface of the cell as it is covered with a delicate Teflon
membrane that can be ruptured in handling.
5. The cell has three pins that fit into the cap one way only. Slide
the pins into the cell cap.
6. Replace the cell cap and cover.
7. After the cell has been installed, purge the instrument with an
inert gas (or the sample), and then proceed as directed in the
calibration procedure.
Cell Warranty
2
The Class S-2 cell used in the Model TURBO
is warranted for 6
2
months of service from date of shipment.
The Model TURBO
should not be used in applications where CO2 is
2
a major component in the sample. Concentrations of 1,000 PPM or less will
not effect cell performance.
If a cell was working satisfactorily, but ceases to function before the
warranty period expires, the customer will receive credit, on a prorated basis,
toward the purchase of a new cell.
Customer having warranty claims must return the cell in question to the
factory for evaluation. If it is determined that failure is due to faulty workmanship or material, the cell will be replaced at no cost to the customer.
CAUTION: Evidence of damage due to tampering or mishandling will render
the cell warranty null and void.
Transduction and Temperature Compensation
The Micro-fuel Cell used in the TURBO
has an inherent positive
2
temperature coefficient, the effects of which have been minimized through
the implementation of a thermistor compensation circuit.
Teledyne Analytical Instruments
13
TURBO
2
Portable Trace Oxygen Analyzer
Pump Removal/Installation(TURBO2P only)
Removal
1. Loosen the cover plate mounting screws and remove the instrument
coverplate.
2. Disconnect the pump motor wires from the circuit terminals.
3. Tag the pump inlet and outlet tubings and then disconnect both from
the pump assembly.
4. Loosen and remove the four screws securing the pump assembly to
the instrument chassis and remove the pump.
Installation
1. With the instrument cover plate removed, set the pump assembly in
place.
2. Insert and tighten the four screws to secure the pump assembly to
the instrument chassis.
3. Connect the tagged pump inlet and outlet tubings to the pump
assembly.
4. Connect the pump motor wires to the circuit terminals.
5. Replace the instrument cover plate and then insert and tighten the
cover plate mounting screws.
14
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
Specifications
Ranges:0-100 PPM oxygen
0-1,000 PPM oxygen
0-10,000 PPM oxygen
Sensitivity:0.5% of full scale
Accuracy:±2% of full scale at constant temperature;
±5% of reading or ±1 PPM (whichever is
greater) over the operating temperature
range.
Response Time:90% of full scale in 5-10 seconds at 68°F
(20°C)
Operating Temperature:+32°F to +125°F (0°C to +52°C)
Micro-fuel Cell Class:S-2
Micro-fuel Cell Warranty:6 months
Recommended Span Gas:Atmospheric air
Operating Power:Two rechargeable NiCad batteries
Battery Recharge Circuit:110 VAC, 60 Hz
100/220 VAC, 50/60 Hz optional
Wetted Parts:Nylon sensor housing and nickel-plated
Mini-Micro-Fuel Cells, all classes
Super Cells, all classes except T–5F
Electrochemical Oxygen Sensors, all classes.
Manufacturer:Manufacturer:
Manufacturer:Teledyne Analytical Instruments
Manufacturer:Manufacturer:
Address:Address:
Address:16830 Chestnut Street, City of Industry, CA
Address:Address:
91749
Phone:Phone:
Phone:(626) 961-9221
Phone:Phone:
Technical Support:Technical Support:
Technical Support:(626) 934-1673
Technical Support:Technical Support:
Environment, HealthEnvironment, Health
Environment, Health
Environment, HealthEnvironment, Health
and Safety:and Safety:
and Safety:(626) 934-1592
and Safety:and Safety:
Date Prepared :Date Prepared :
Date Prepared :08/08/91
Date Prepared :Date Prepared :
2
Section II – Physical and Chemical Data
Chemical and Common Names:Potassium Hydoxide (KOH), 15% (w/v)
Lead (Pb), pure
CAS Number:KOH 1310–58–3
Pb 7439–92–1
KOH KOH
KOH (15%)
KOH KOH
Melting Point/Range:–10 to 0 °C328 °C
Boiling Point/Range:100 to 115 °C1744 °C
Specific Gravity:1.09 @ 20 °C11.34
pH:>14N/A
Solubility in Water:Completely solubleInsoluble
Percent Volatiles by Volume:NoneN/A
Appearance and Odor:Colorless, odorless solutionGrey
Pb Pb
Pb (pure)
Pb Pb
metal,
odorless
Teledyne Analytical Instruments
17
TURBO
2
Portable Trace Oxygen Analyzer
Section III – Physical Hazards
Potential for fire and explosion:Potential for fire and explosion:
Potential for fire and explosion: The electrolyte in the Micro-Fuel Cells is
Potential for fire and explosion:Potential for fire and explosion:
not flammable. There are no fire or explosion hazards associated with Micro-Fuel Cells.
Potential for reactivity: The sensors are stable under normal conditions of
use. Avoid contact between the sensor electrolyte and strong acids.
Section IV – Health Hazard Data
Primary route of entry:Primary route of entry:
Primary route of entry:Ingestion, eye/skin contact
Primary route of entry:Primary route of entry:
Exposure limits:Exposure limits:
Exposure limits:OSHA PEL:.05 mg/cu.m. (Pb)
Exposure limits:Exposure limits:
ACGIH TLV:2 mg/cu.m. (KOH)
Effects of overexposureEffects of overexposure
Effects of overexposure
Effects of overexposureEffects of overexposure
Ingestion:Ingestion:
Ingestion:The electrolyte could be harmful or fatal if
Ingestion:Ingestion:
swallowed.
Oral LD50 (RAT) = 3650 mg/kg
Eye:Eye:
Eye:The electrolyte is corrosive; eye contact
Eye:Eye:
could result in permanent loss of vision.
Dermal:Dermal:
Dermal:The electrolyte is corrosive; skin contact
Dermal:Dermal:
could result in a chemical burn.
Inhalation:Inhalation:
Inhalation:Liquid inhalation is unlikely.
Inhalation:Inhalation:
Signs/symptoms of exposure:Signs/symptoms of exposure:
Signs/symptoms of exposure:Contact with skin or eyes will cause a
Signs/symptoms of exposure:Signs/symptoms of exposure:
burning sensation and/or feel soapy or
slippery to touch.
Medical conditionsMedical conditions
Medical conditions
Medical conditionsMedical conditions
aggravated by exposure:aggravated by exposure:
aggravated by exposure:None
aggravated by exposure:aggravated by exposure:
Carcinogenicity:Carcinogenicity:
Carcinogenicity:NTP Annual Report on Carcinogens: Not
Carcinogenicity:Carcinogenicity:
listed
LARC Monographs: Not listed
OSHA: Not listed
18
Other health hazards:Other health hazards:
Other health hazards:Lead is listed as a chemical known to the
Other health hazards:Other health hazards:
State of California to cause birth defects
or other reproductive harm.
Teledyne Analytical Instruments
Portable Trace Oxygen AnalyzerTURBO
Section V – Emergency and First Aid Procedures
Eye Contact:Eye Contact:
Eye Contact:Flush eyes with water for at least 15 minutes and get im-
Eye Contact:Eye Contact:
mediate medical attention.
Skin Contact:Skin Contact:
Skin Contact:Wash affected area with plenty of water and remove con-
Skin Contact:Skin Contact:
taminated clothing. If burning persists, seek medical
attention.
Ingestion:Ingestion:
Ingestion:Give plenty of cold water. Do not induce vomiting. Seek
Ingestion:Ingestion:
medical attention. Do not administer liquids to an unconscious person.
Inhalation:Inhalation:
Inhalation:Liquid inhalation is unlikely.
Inhalation:Inhalation:
Section VI – Handling Information
NOTE: The oxygen sensors are sealed, and under normal circumstances,
the contents of the sensors do not present a health hazard. The
following information is given as a guide in the event that a cell
leaks.
2
Protective clothing:Protective clothing:
Protective clothing:Rubber gloves, chemical splash goggles.
Protective clothing:Protective clothing:
Cleanup procedures:Cleanup procedures:
Cleanup procedures:Wipe down the area several times with a wet paper
Cleanup procedures:Cleanup procedures:
towel. Use a fresh towel each time.
Protective measuresProtective measures
Protective measures
Protective measuresProtective measures
during cell replacement:during cell replacement:
during cell replacement:Before opening the bag containing the sensor cell,
during cell replacement:during cell replacement:
check the sensor cell for leakage. If the sensor
cell leaks, do not open the bag. If there is liquid
around the cell while in the instrument, put on
gloves and eye protection before removing the
cell.
Disposal:Disposal:
Disposal:Should be in accordance with all applicable state,
Disposal:Disposal:
local and federal regulations.
NOTE: The above information is derived from the MSDS provided by the
manufacturer. The information is believed to be correct but does
not purport to be all inclusive and shall be used only as a guide.
Teledyne Analytical Instruments shall not be held liable for any
damage resulting from handling or from contact with the above
product.
Teledyne Analytical Instruments
19
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.