Teledyne RS-232 User Manual

RS-232 Multidrop Manual Addendum
02179 Rev. B Revised 17 January 2003
1
TELEDYNE API RS232 MULTI-DROP ADAPTOR
The Teledyne API RS232 Multi-drop Adaptors #01427 (A-series) and #04528 (E-series) provides the user with a simple method of communicating via asynchronous RS232 with multiple Teledyne API instruments using a single host system (computer, modem, DAS or terminal).
1. Dual DB 9 pin connectors - connected in parallel for ease in interconnecting with additional adaptors in network. Will not interrupt communication with other instruments on the network should an instrument be off-line, lose power, or fail. Network connectors may be connected in any order.
2. Standard connectors are DB-9S 9 pin female. Optional dual DB-9P male (#01428-01) or one each male & female (#01428-02) cable assembly can be replaced in the field at any time.
3. Configuration switch DTE/DCE the host system without the necessity of special wiring cables or “Null MODEM” connectors in most installations.
4. Impedance switch TERM/HI-Z the “HI-Z” position, impedance is approximately 300KΩ. Only one adaptor in the network should be set to the “TERM” position.
NOTE: If multi-drop network includes non-Teledyne API equipment with multi-drop capability, refer to those manufacturers multi-drop specifications to determine if the “TERM” switch on the Teledyne API multi-drop adaptor should be set and also if the network impedance complies with EIA/TIA-232E, V.28 load specification of 3kΩ to 7kΩ.
- changes the signal lines at the adaptor to the analyzer as dictated by
- inserts a 5KΩ termination when set to the “TERM” position. When set to
Software Setup for RS-232 Multi-Drop (A-series analyzers)
Multi-drop RS-232 support is implemented by extending the standard RS-232 commands to support an ID number. If no ID number is specified, each analyzer will accept the command, as it always has, so that the ID feature is transparent. However, analyzers with identical ID numbers will be in conflict and only the first analyzer in the chain will respond.
For example, if the analyzer ID number is 100, then the following commands will be processed as indicated:
Typical Command Analyzer Response t list Lists test measurements. t 100 list Lists test measurements. t 123 list Analyzer ignore this command, since ID is wrong.
For all commands, the ID number is the second argument, and is optional. Use the “?” command to obtain a complete list of commands.
In order to use the multi-drop features in a true multi-drop environment, where multiple analyzers share the same RS-232 channel, you will need to make the following changes:
2
Using a terminal program, issue the command “v rs232_mode=xx”, where xx is defined below:
Model xx M100 35 M200 35 M100A 43 M200A 43 M300 43 M400 43 M700 43
The software uses the RTS signal from the SBC40 to control the multi-drop interface hardware. The software turns this signal on (logic high) prior to transmitting data, and turns it off within 1 second after the last character has been fully transmitted.
Software Setup for RS-232 Multi-Drop (E-series analyzers)
Every analyzer on the multi-drop chain has to have a unique ID number. Default ID numbers are the analyzer model number:
M100E – ID 100 M200E – ID 200 M300E – ID 300 M400E – ID 400
However, if more than one analyzer of the same model is used, the ID numbers have to be changed accordingly.
To enable multi-drop in the software, follow this command sequence (this example assumes that COM1 is used for multi-drop use):
SETUP – MORE – COMM – COM1 – EDIT mode and select NEXT until COM1 MULTIDROP MODE appears and then select ON, press ENTER, then 4x EXIT. The COM1 Mode should now show a number that is 32 higher than before. If you wish to operate an E-series analyzer in series with older models, keep in mind that the baud rate for all
COM ports needs to match. A-series analyzers support a maximum baud rate of 19200 baud, whereas E-series analyzers support up to 115 000 baud. Multi-drop performance may degrade at speeds higher than 19200 baud, particularly if cheap, unshielded cables are used. With short, shielded, good quality cables, the maximum baud rate should work.
Hardware Setup for RS-232 Multi-Drop (E-series analyzers)
Whereas A-series analyzers draw their power from within the analyzer chassis, E-series analyzers can supply 5V DC power directly from the rear panel. The E-series multi-drop option includes a small cable converter with open leads to be connected to the screw terminals of the STATUS output connector on the rear panel. Connect the red wire to the screw terminal labeled + on the STATUS connector and connect the black wire to the terminal labeled G on the same connector. Please refer to Figure 4 for proper connection of a multi-drop option (part # 04528).
CAUTION: Do not reverse the wiring, it will destroy the multi-drop circuitry!
3