Teledyne Mixchek User Manual

MIXCHEK
OPERATING INSTRUCTIONS FOR
MIXCHEK
HELIUM/OXYGEN
MIXTURE ANALYZER
P/N M75030
Rev 2
ECO # 05-0138
This instrument is designed to monitor the helium and oxygen concentrations in a gas mixing process. When processing gas mixtures intended for breathing, the instrument must be calibrated and functioning properly. Suffocation or other fatal respiratory complications can result from improperly calibrated or malfunctioning equipment.
Oxygen accelerates combustion. Do not use this instrument around open flames and combustible gases, liquids or solids.
Only authorized personnel should conduct maintenance and/or servicing. Before conducting any maintenance or servicing, consult with authorized supervisor/manager.
Teledyne Analytical Instruments i
MIXCHEK Manual Rev 2 7/25/05 P/N M75030 ECO 05-0138
DANGER
MIXCHEK
Copyright © 2003 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage, which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its use r, an d serves as a tool b y whic h valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the process conditions and the potential for misuse. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that all devices and instrumentation are calibrated, maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s process conditions.
Teledyne Analytical Instruments ii
MIXCHEK
Specific Model Information
The instrument for which this manual was supplied may incorporate one or more options not supplied in the standard instrument. Commonly available options are listed below.
CE CONFORMITY: This Teledyne Analytical instruments
MIXCHEK Analyzer meets or exceeds all requirements of the Commonwealth of Europe (CE) for Radio Frequency Interference and Electromagnetic Interference (RFI/EMI) protection.
MIXCHEK MCWHE: Wall-mountable instrument with
view window for helium analysis.
MIXCHEK MCWHEO2: Wall-mountable instrument with
view window for helium and oxygen analysis.
MIXCHEK MCPHE: Battery operated portable instrument
for helium analysis.
MIXCHEK MCPHEO2: Battery operated portable
instrument for helium and oxygen analysis.
Optional AC to DC Adapter: Power instrument directly
from 115VAC US plug source.
Optional International Power Adapter: Powers instrument
directly from an AC power source 100-240 VAC 50/60 Hz CE approved. Includes US and 3 international AC plugs.
Teledyne Analytical Instruments iii
MIXCHEK
Safety Messages
Your safety and the safety of others are very important. We have provided many important safety messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions for
details on the specific danger. These cautions warn of specific procedures which if not followed could cause bodily Injury and/or damage the instrument.
CAUTION: HOT SURFACE WARNING: This warning is specific to
heated components within the instrument. Failure to heed the warning could result in serious burns to skin and underlying tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous voltages appear
within this instrument. This warning is specific to an electrical hazard existing at or nearby the component or procedure under discussion. Failure to heed this warning could result in injury and/or death from electrocution.
TECHNICIAN SYMBOL: All operations marked with this symbol are to
be performed by qualified maintenance personnel only.
NOTE: Additional information and comments regarding a
No
Symbol
CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE
specific component or procedure are highlighted in the form of a note.
PURPOSE AND IN THE MANNER DESCRIBED IN THIS MANUAL.
Teledyne Analytical Instruments iv
MIXCHEK
CAUTION: THE MIXCHEK ANALYZER ELECTRONICS ARE NOT
WATER RESISTANT. DO NOT ALLOW WATER INSIDE THE ENCLOSURE.
IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES.
This manual provides information designed to guide you through the installation, calibration and operation of your new analyzer. Please read this manual and keep it available.
Occasionally, some instruments are customized for a particular application or features and/or options added per customer requests. Please check the front of this manual for any additional information in the form of an Addendum which discusses specific information, procedures, cautions and warnings that may be peculiar to your instrument.
Manuals do get lost. Additional manuals can be obtained from Teledyne at the address given in the Appendix. Some of our manuals are available in electronic form via the internet. Please visit our website at: www.teledyne-ai.com.
Teledyne Analytical Instruments v
MIXCHEK
Table of Contents
Safety Messages ..........................................................................iv
Table of Contents.........................................................................vi
List of Figures.............................................................................viii
Introduction .................................................................................11
1.1 Overview 11
1.2 Features 13
1.3 Options 13
1.4 Applications 13
Setup ............................................................................................15
2.1 Unpacking the Analyzer 15
2.2 Mounting the Analyzer (Wall-Mount Unit) 15
2.3 Installing the Oxygen Sensor 16
2.4 AC Power Connections 18
2.5 Battery Test (Portable Model Only) 20
2.6 Battery Installation (Portable Model Only) 20
2.7 Calibration 21
Operation .....................................................................................25
3.1 Operation/Start-up 25
3.1.1 Powering Up 25
3.1.2 Setting the O2 Span 25
3.1.3 Attaching the Sample Gas 25
Maintenance.................................................................................27
4.1 Micro-Fuel Cell Replacement 27
4.1.1 Storing and Handling Replacement Cells 27
4.2 Battery Replacement 27
Teledyne Analytical Instruments vi
MIXCHEK
Appendix......................................................................................29
A.1 Specifications 29 A.2 Spare Parts List 31 A.3 Helium/Air Chart 32 A.4 Material Safety Data Sheet 36
Index.............................................................................................37
Teledyne Analytical Instruments vii
MIXCHEK
List of Figures
Figure 1-1: Wall-mountable MIXCHEK..........................................12
Figure 1-2: MIXCHEK—Portable Configuration.............................12
Figure 2-1: Mounting Dimensions..................................................16
Figure 2-2: Removing the Front Panel .........................................17
Figure 2-3: Installing the Sensor ...................................................18
Figure 2-4: Power Adapter Port (Wall Mount)................................19
Figure 2-5: Power Adapter Port (Portable Model) .........................19
Figure 2-6: Battery Installation.......................................................21
Figure 2-7: Typical Sample System with Calibration.....................24
Figure 3-1: Typical Setup Flow Diagram .......................................26
Figure 4-1: Battery Disposition......................................................28
Teledyne Analytical Instruments viii
MIXCHEK
DANGER
COMBUSTIBLE GAS USAGE
This is a general-purpose instrument designed for use in a non-hazardous area. It is the customer's responsibility to ensure safety especially when oxygen is being analyzed. Oxygen readily promotes and accelerates combustion. Combustible gases should not be present in the immediate area nor should this instrument be used near an open flame or ignition source.
WARNING
Helium and nitrogen, while inert, can pose a suffocation risk. The customer must eliminate the possibility of displacing atmospheric air with helium or nitrogen at or near the instrument during the mixing or analysis process.
The customer must also ensure that the principles of operating this equipment are well understood by the user. Misuse of this product in any manner, tampering with its components, or unauthorized substitution of any component may adversely affect the safety of this instrument.
Since the use of this instrument is beyond the control of Teledyne, no responsibility by Teledyne, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed.
Teledyne Analytical Instruments ix
MIXCHEK
BLANK
Teledyne Analytical Instruments x
MIXCHEK Introduction
Introduction
The Teledyne MIXCHEK is a dedicated instrument designed to accurately monitor the composition of a gas during mixing or cylinder filling operations. A LCD mounted on the front panel of the instrument displays the percentage of helium and oxygen during the mixing operation.
There are two basic configurations available both of which offer oxygen analysis as an option. The wall-mountable unit is housed in an enclosure equipped with a window for viewing the helium and oxygen displays. It is powered by a plug-in 6V DC adapter which uses a conventional 120VAC source (100-240 VAC 50/60 Hz optional).
The portable instrument is housed in a rugged high-impact plastic watertight enclosure. It has a molded-in handle for easily transporting the instrument to the mixing or process area. Both models, the standard wall­mount and the portable instrument, use corrosive-resistant materials where applicable for enhanced durability in salt air or spray environments. The wall-mount model is shown in Figure 1-1 and the portable instrument is shown in Figure 1-2.
NOTE: If your unit is equipped with a Helium only analysis please ignore all references to the oxygen sections of this manual.
1.1 Overview
The MIXCHEK analyzes the helium/oxygen concentrations in a gas mixture using a dedicated two-channel sensor network sharing a common sample system. The gas to be monitored is drawn into the analyzer and passed to the sensor network. A high output micro-fuel cell oxygen sensor is used to determine the oxygen concentration of the gas mixture. The output of the micro-fuel cell is fed to the display and read out as percent oxygen.
The sample gas is then directed to a thermal conductivity sensor. This sensor provides a signal that is proportional to the difference in conductivities between the components of the gas mixture. A microprocessor compares the sensor output to data contained in an onboard chip and converts the raw data into a linearized signal that is sent to the helium LCD display.
Teledyne Analytical Instruments 11
Introduction MIXCHEK
Accuracy is maintained to ± 1% across the entire 0-100% analysis range (He or O
) of the instrument at constant temperature (25°C) at
2
constant pressure (atmospheric).
Figure 1-1: Wall-mountable MIXCHEK
Figure 1-2: MIXCHEK—Portable Configuration
Teledyne Analytical Instruments 12