HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us
free from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Teledyne Analytical Instruments
Photometric Analyzer
Table of Contents
Part I: Control Unit ................................Part I: 1-1
Part II: Analysis Unit ............................Part II: 4-1
The Teledyne Analytical Instruments Model 5000B Control Unit,
together with a 5000B Analysis Unit, is versatile microprocessor-based
instrument.
Part I, of this manual covers the Model 5000B General Purpose NEMA
4 Bulkhead Mount Control Unit. (The Analysis Unit is covered in Part II of
this manual.) The Control Unit is for indoor/outdoor use in a nonhazardous
environment only. The Control or Analysis Unit in the folded optics design
can accomodate a variety of hazardous environments with either full explosion proof housings or Z, X, Y or Cenelec Purging functions.
1.2Typical Applications
A few typical applications of the Model 5000B are:
Hydroxyl Value
Hydrocarbons
Hydrogen Chloride
Hydrogen Fluoride
Hydroxyl Value
Ketones
Olefins
Oximes
Epoxides
Methylene
Ketones
1-4: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
1.3Main Features of the Analyzer
The Model 5000B Photometric Analyzer is sophisticated yet simple to
use. The main features of the analyzer include:
•A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•High resolution, accurate readings of concentration from low
ppm levels through to 100%. Large, bright, meter readout.
•Versatile analysis over a wide range of applications.
•Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•Three user definable output ranges (from 0-1 ppm through
0-100 %) allow best match to users process and equipment.
•Calibration range for convenient zeroing or spanning.
•Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.
•Two adjustable concentration alarms and a system failure alarm.
•Extensive self-diagnostic testing, at startup and on demand, with
continuous power-supply monitoring.
•RS-232 serial digital port for use with a computer or other digital
communication device.
•Analog outputs for concentration and range identification.
(0-1 V dc standard, and isolated 4–20 mA dc)
•Superior accuracy.
•Internal calibration-Manual or Automatic (optional).
Teledyne Analytical Instruments
Part I: 1-5
1 Introduction Model 5000B
1.4Operator Interface
All controls and displays on the standard 5000B are accessible from
outside the housing. The instrument has two simple operator controls. The
operator has constant feedback from the instrument through an alphanumeric
display, and a digital LED meter. The displays and controls are described
briefly here and in greater detail in chapter 3. See Figure 1-1.
1.4.1UP/DOWN Switch
The UP/DOWN switch is used to select between any subfunctions
displayed on the VFD screen such as in the main menue, the system menue,
the Alarm menue, etc. When modifiable values are displayed on the VFD,
the UP/DOWN switch can be used to increment or decrement the values.
1.4.2ESCAPE/ENTER Switch
The ESCAPE/ENTER switch is used to input the data, to enter a
function, or to exit a function displayed in the alphanumeric display:
•EscapeMoves VFD display back to the previous screen in a
series. If none remains, returns to Analyze mode
screen.
•EnterWithin a menue: the funtion selected is entered
moving onto the next screen in a series.
With Value selected: Enters the value into the
analyzer asdata. Advances cursor on VFD to the
next operation.
In the Analyze mode: it calls the main menue.
Functionscalled out by the main menue:
-SystemThis function is a menu that calls a
numberof functions
that regulate the analyzer
operation.
-SpanThis function spans the instrument.
1-6: Part I
-ZeroThis function zeros the instrument.
-AlarmsThis functions sets the alarm
preferences.
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
-RangeThis function selects whether analyzer is
autoranging or locked on one range.
-Standby Places the analyzer in a sleep mode.
WARNING:
The power cable must be disconnected to fully remove
power from the instrument.
Figure 1-1: Model 5000B Controls, Indicators, and Connectors
Digital Meter Display: The meter display is a Light Emitting Diode
LED device that produces large, bright, 7-segment numbers that are legible
in any lighting. It is accurate across all analysis ranges. The 5000B models
produce continuous readout from 0-10,000 ppm and then switch to
continuous percent readout from 1-100 %.
Teledyne Analytical Instruments
Part I: 1-7
1 Introduction Model 5000B
Alphanumeric Interface Screen: The backlit VFD screen is an easy-
to-use interface between operator and analyzer. It displays values, options,
and messages for immediate feedback to the operator.
1.5Control Section Interface Panel
The Control Section interface panel, shown in Figure 1-2, contains the
electrical terminal blocks for external inputs and outputs. The input/output
functions are described briefly here and in detail in the Installation chapter of
this manual.
•Power ConnectionAC power source, 115VAC, 50/60 Hz
•Analog Outputs0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20
mA dc range ID.
•Alarm Connections2 concentration alarms and 1 system
alarm.
•RS-232 PortSerial digital concentration signal output
and control input.
•Remote BenchProvides all electrical interconnect to the
Analysis Section.
Remote Span/ZeroDigital inputs allow external control of
analyzer calibration.
•Calibration ContactTo notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•Range ID ContactsFour separate, dedicated, range relay
contacts.
•Network I/OSerial digital communications for local
network access. For future expansion.
Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 5000B is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
1-8: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
Figure 1-2: Model 5000B Interface Panel
Teledyne Analytical Instruments
Part I: 1-9
1 Introduction Model 5000B
1-10: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
Installation
Installation of Model 5000B Analyzers includes:
1. Unpacking, mounting, and interconnecting the Control/Analysis
Section
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
This chapter covers installation of the Control Section. (Installation of
the Analysis Section is covered in Part II of this manual.)
2.1Unpacking the Control/Analysis Unit
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the Control/Analysis
Unit and inspect it for damage. Immediately report any damage to the shipping agent. Figure 2-2: Required Front Door Clearance
Allow clearance for the door to open in a 90-degree arc of radius 15.5
inches. See Figure 2-2.
Figure 2-2: Required Front Door
Clearance
1
5
.
5
”
2.2Electrical Connections
Figure 2-3 shows the Control/Analysis Unit interface panel. Connections for power, communications, and both digital and analog signal outputs
are described in the following paragraphs. Wire size and maximum length
data appear in the Drawings at the back of this manual.
Teledyne Analytical Instruments
Part I: 2-1
1 Introduction Model 5000B
Figure 2-3: Interface Panel of the Model 6600 Control Section
For safe connections, ensure that no uninsulated wire extends
outside of the terminal blocks. Stripped wire ends must insert completely
into terminal blocks. No uninsulated wiring should come in contact with
fingers, tools or clothing during normal operation.
Primary Input Power: The power supply in the Model 6600 will
accept a 115 Vac, 50/60 Hz power source. See Figure 2-4 for detailed
connections.
DANGER:Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
source. The standby function switches power on or
off to the displays and outputs only.
2-2: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
Figure 2-4: Primary Input Power Connections
115VAC,
Fuse Installation: The fuse holders accept 5 x 20 mm, 4.0 A, T type
(slow blow) fuses. Fuses are not installed at the factory. Be sure to install
the proper fuse as part of installation (See Fuse Replacement in chapter 4,
maintenance.)
Analog Outputs: There are eight DC output signal connectors on
the ANALOG OUTPUTS terminal block. There are two connectors per
output with the polarity noted. See Figure 2-5.
The outputs are:
0–1 V dc % of Range:Voltage rises linearly with increasing sample con-
centration, from 0 V at 0% to 1 V at 100%. (Full
scale = 100% programmed range.)
0–1 V dc Range ID:0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3.
4–20 mA dc % Range: Current increases linearly with increasing sample
concentration, from 4 mA at 0% to 20 mA at full
scale 100%. (Full scale = 100% of programmed
range.)
Teledyne Analytical Instruments
Part I: 2-3
1 Introduction Model 5000B
4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA =
Range 3.
Figure 2-5: Analog Output Connections
Examples:
The analog output signal has a voltage which depends on the sample
concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.
The signaloutput for concentration is linear over currently selected
analysis range. For example, if the analyzer is set on a range that was
defined as 0-10 %, then the output would be as shown in Table 2-1.
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 2-2 gives the
range ID output for each analysis range.
Table 2-2: Analog Range ID Output - Example
RangeVoltage (V)Current (mA)
Range 10.258
Range 20.5012
Range 30.7516
Alarm Relays:
There are three alarm-circuit connectors on the alarm relays block
(under RELAY OUTPUTS) for making connections to internal alarm relay
contacts. Each provides a set of Form C contacts for each type of alarm.
Each has both normally open and normally closed contact connections. The
contact connections are indicated by diagrams on the rear panel. They are
capable of switching up to 3 ampers at 250 V AC into a resistive load
(Figure 2-6).
Teledyne Analytical Instruments
Part I: 2-5
1 Introduction Model 5000B
N
Normally closed
Normally open
Moving contact
2-6: Types of Relay Contacts
ormally open
Moving contact
Figure
The connectors are:
Threshold Alarm 1:• Can be configured as high (actuates when
concentration is above threshold), or low
(actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2:• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm:Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be defeated. Actuates if self test fails.
To reset a System Alarm during installation, disconnect power to the instrument and then reconnect it
Further detail can be found in chapter 3, section 3-5.
Digital Remote Cal Inputs
Remote Zero and Span Inputs: The REMOTE SPAN and RE-
MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They
2-6: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See
Remote Calibration Protocol below.)
Zero:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the ZERO mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows ZERO mode to terminate when done. A
synchronous signal must open and close the external zero
valve appropriately. See Remote Probe Connector at end of
section 3.3. (With the -C option, the internal valves automatically operate synchronously).
Span:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the SPAN mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows SPAN mode to terminate when done. A
synchronous signal must open and close the external span
valve appropriately. See Remote Probe Connector at end of
section 3.3. (With the -C option, the internal valves automatically operate synchronously.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 5000B Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
Teledyne Analytical Instruments
Part I: 2-7
1 Introduction Model 5000B
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Bench terminal strip (section 3.6 Part III) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously.
Range ID Relays: Four dedicated RANGE ID CONTACT relays .
The first four ranges are assigned to relays in ascending order—Range 1 is
assigned to RANGE 1 ID, Range 2 is assigned to RANGE 2 ID, Range 3
is assigned to RANGE 3 ID, and Range 4 is assigned to RANGE 4 ID.
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future
versions of the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or
other digital device. The pinouts are listed in Table 2-3.
Table 2-3: RS-232 Signals
RS-232 Sig RS-232 Pin Purpose
DCD1Data Carrier Detect
RD2Received Data
TD3Transmitted Data
DTR4Data Terminal Ready
COM5Common
DSR6Data Set Ready
RTS7Request to Send
CTS8Clear to Send
RI9Ring Indicator
The data sent is status information, in digital form, updated every two
seconds. Status is reported in the following order:
•The concentration in percent
•The range is use (HI< MED< LO)
2-8: Part I
•The span of the range 0-100%, etc.
•Which alarm - if any - are disabled (AL-x DISABLED)
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
•Which alarms - if any - are tripped (AL-x ON)
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date. They
are described in Table 2-4.
Table 2-4: Commands via RS-232 Input
CommandDescription
as<enter>Immediately starts an autospan.
az<enter>Immediately starts an autozero.
st<enter>Toggling input. Stops/Starts any status message output
from the RS-232, Until st<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation. Table 2-
5 lists certain RS-232 values that are required by the 5000B.
Table 2-5: Required RS-232 Options
ParameterSetting
Baud2400
Byte8 bits
Paritynone
Stop Bits1
Message Interval2 seconds
Remote Bench and Solenoid Valves: The 5000B is a single-chassis
instrument. However, the REMOTE BENCH and SOLENOID RETURN
connectors are provided on the interface PCB. The Remote Bench is wired
at the factory as well as any optional solenoid valves included in the system.
2.3Testing the System
After The Control/Analysis Unit is both installed and interconnected, and
the system gas or liquid stream and electrical connections are complete, the
system is ready to test. Before plugging the unit into its power sources:
•Check the integrity and accuracy of the fluid connections. Make sure
there are no leaks.
•Check the integrity and accuracy of all electrical connections. Make
sure there are no exposed conductors
Teledyne Analytical Instruments
Part I: 2-9
1 Introduction Model 5000B
•Check that sample pressure typically between 0 and 30 psig,
according to the requirements of your process.
Power up the system, and test it by performing the following operation:
1. Repeat the Self-Diagnostic Test.
2. Zero the instrument.
3. Span the instrument.
Consult calibration recommendation in the addendum for your
particular application whether liquid or gas phase monitoring.
For steps 2 and 3, refer to part II for calibration.
2-10: Part I
Teledyne Analytical Instruments
Photometric AnalyzerOperation /Control Unit 3
Operation
3.1Introduction
Although the Model 5000B is usually programmed to your application
at the factory, it can be further configured at the operator level, or even,
cautiously, reprogrammed. Depending on the specifics of the application,
this might include all or a subset of the following procedures:
•Setting system parameters:
•Establish a security password, if desired, requiring Operator
to log in (secure in safe file for referrence).
•Establish and start an automatic calibration cycle, if desired.
•Routine Operation:
•Calibrate the instrument.
•Choose autoranging or select a fixed range of analysis.
•Set alarm setpoints, and modes of alarm operation (latching,
fail-safe, etc).
•Program/Reprogram the analyzer:
•Define new applications.
•Linearize your ranges.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.
3. 2Using the Data Entry and Function
Buttons
Data Entry Buttons: The < > buttons select options from the menu
currently being displayed on the VFD screen. The selected option blinks.
When the selected option includes a modifiable item, the DÑ arrow
buttons can be used to increment or decrement that modifiable item.
Teledyne Analytical Instruments
Part I 3-1
3 Operation/Control Unit Model 5000B
The Enter button is used to accept any new entries on the VFD screen.
The Escape button is used to abort any new entries on the VFD screen that
are not yet accepted by use of the Enter button.
Figure 3-1 shows the hierarchy of functions available to the operator via the
function buttons. The six functions of the buttons on the analyzer are:
•Analyze. This is the normal operating mode. The analyzer monitors
the concentration of the sample, displays the percent or parts-permillion of target fluid or contamination, and warns of any alarm
conditions.
•System. The system function consists of nine subfunctions.
Four of these are for ordinary setup and operation:
•Setup an Auto-Cal
•Assign Passwords
•Log out to secure system
•Initiate a Self-Test
Three of the subfunctions do auxiliary tasks:
•Checking model and software version
•Adjust electronic filter of the signal
•Display more subfunctions
•Display negative readings
Two of these are for programming/reprogramming the analyzer:
•Define fluid applications and ranges (Refer to programming
section, or contact factory.)
•Use the Curve Algorithm to linearize output. (Refer to
programming section, or contact factory.)
•Zero. Used to set up a zero calibration.
•Span. Used to set up a span calibration.
•Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching,
and/or fail-safe.
•Range. Used to set up three analysis ranges that can be switched
automatically with autoranging or used as individual fixed
ranges.
Any function can be selected at any time by pressing the appropriate
button (unless password restrictions apply). The order as presented in this
manual is appropriate for an initial setup.
Each of these functions is described in greater detail in the following procedures. The VFD screen text that accompanies each operation is reproduced, at
3-2 Part I
Teledyne Analytical Instruments
Loading...
+ 76 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.