Teledyne API Model M201E NH3 Analyzer Operator Manual M201E Ammonia Analyzer
SAFETY MESSAGES
Your safety and the safety of others is very important. We have provided many safety messages
throughout this manual; please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Safety messages
are associated with a safety alert symbol. These symbols are found in both the M200E and
M201E manuals and are also located inside the instrument. The definition of these symbols is
described below:
GENERAL WARNING/CAUTION: Refer to the instructions for details on the
specific danger.
CAUTION: Hot Surface Warning
WARNING: Electrical Shock Hazard
Technician Symbol: All operations marked with this symbol are to be performed
by qualified maintenance personnel only.
Electrical Ground: This symbol inside the instrument marks the central safety
grounding point for the instrument.
CAUTION The analyzer should only be used for the purpose
and in the manner described in this manual.
If you use the analyzer in a manner other than that for which
it was intended, unpredictable behavior could ensue with
This manual is to be used in conjunction with the M200E NOx Analyzer
manual, p/n 04410 supplied with this instrument. It is important that you
familiarize yourself with the workings of the NOx analyzer before proceeding
with NH3 measurements.
The M201E ammonia analyzer consists of a modified M200E nitrogen oxide
analyzer and an M501 NH3, ammonia converter. The analyzer measures the
following individual gas concentrations: TNx (NH
and NO in the sample gas.
and NOx), NOx (NO and NO2)
3
The
ammonia concentration is calculated by subtracting the NOX
reading from the TNx measurement.
The
nitrogen dioxide concentration is calculated by subtracting the NO
reading from the NOx measurement.
The instrument is available in three pneumatic configurations:
The M201E with External Calibrator Option (ECO) is shown in Figure
1-1.
The basic system is shown in Figure 1-2.
The M201E with Zero/Span valves is shown in Figure 1-3.
Note: For an M201E analyzer with Zero/Span option you must connect the
umbilical cord P/N 02255 between the analyzer and the external M501 NH3,
converter assembly.
If the analyzer is purchased without the Zero/Span option t hen the zero and span
calibration gases must be individually applied to the sample port. Remember, the
gases must be clean/dry and supplied at ambient pressure.
CAUTION
Do not pressurize the Sample/Span/Zero port above ambient
05206H DCN5910 7
M201E Ammonia Analyzer Teledyne API Model M201E NH3 Analyzer Operator Manual
Teledyne API Model M201E NH3 Analyzer Operator Manual M201E Ammonia Analyzer
Figure 1-2. M201E and M501 NH3, Pneumatic Flow
05206H DCN5910 9
M201E Ammonia Analyzer Teledyne API Model M201E NH3 Analyzer Operator Manual
Figure 1-3. M201E and M501 NH3, with Zero and Span Valve Options Pneumatic Flow
1.1 THEORY OF OPERATION
The Teledyne-API Model M201E analyzer measures ammonia by oxidizing it to
nitric oxide by the following reaction:
+ 5O2 4NO + 6H2O
4NH
3
The resulting nitric oxide is then measured by the chemiluminescent reaction with
ozone. Consult Section 10.1 in the M200E manual (provided with this
instrument) for more details on the nitric oxide measurement.
The analyzer uses two converters to oxidize the different sample gases. A high
temperature catalytic converter, the M501 NH3,, converts NH3 and NOx into NO
creating the TN
analyzer and consisting of heated molybdenum, converts all of the NOx in the
sample to NO producing the NOx channel. The nitric oxide channel is measured
while bypassing both the M501 NH
operated in reflux mode, is installed prior to the molybdenum converter and the
AutoZero valve. The drier removes ammonia and water from the switched
stream. The ammonia would have otherwise been converted by the molybdenum
channel. A second converter, housed inside the M201E
x
, and molybdenum converter. A Nafion® drier
3
10 05206H DCN5910
Teledyne API Model M201E NH3 Analyzer Operator Manual M201E Ammonia Analyzer
and registered as NOx. This location provides the drier with continuous flow,
thereby allowing it to stabilize faster.
Note: Due to the Nafion® drier, the analyzer has a maximum am monia range of
2.0 PPM.
The calculated gas concentrations, ammonia and nitrogen dioxide, are computed
from the differences between the measured channels, as outlined in the following
equations:
NO2 (Nitrogen Dioxide) Concentration = Calculated as follows:
[(NOx – NO) / (NO2_CE_FACTOR1)]
Note: The nitrogen dioxide efficiency factor (NO2_CE_FACTOR1) must be
calculated by conducting a GPT (gas phase titration) as described in Section
8.3.3 of the M200E manual and Section 7.0 of this manual. This value gives a
good indication of how well the molybdenum is converting nitrogen dioxide into
nitric oxide. In a properly functioning analyzer the value should be close to 1.0. It
is recommended that the molybdenum efficiency be checked every three months.
The ammonia converter efficiency factor (NH3_CE_FACTOR1) is discussed in
more detail in Section 2.3 of this manual. The conversion efficiency of the M501
NH3, should be checked prior to starting long term tests. Both efficiency factors
can be accessed through the analyzer VARS MENU.
The actual formula for computation of the gas concentrations is more
complicated than the above equations, as it includes sample, reaction cell
pressure changes and averaging the PMT signal. Then the zero offset and
slopes are applied to the separate channels (TNx, NOx, NO) to determine the
concentrations. Concentration compensation occurs while the variable
TP_FACTOR under the VARS MENU is switched ON. Otherwise the displayed
concentration is uncompensated. It is recommended that the variable
TP_FACTOR remain on at all times.
1.1.1 Minimizing PMT Drift
In order to account for PMT drift in the analyzer, the AutoZero valve switches
once a minute allowing the analyzer to read zero background. The AutoZero
valve directs the sample gas stream to completely bypass the reaction cell, while
simultaneously filling the reaction cell with Ozone for dark noise measurement.
This is then subtracted as a measurement offset from the raw PMT concentration
05206H DCN5910 11
M201E Ammonia Analyzer Teledyne API Model M201E NH3 Analyzer Operator Manual
signal. This process improves zero baseline stability by minimizing the effect of
PMT sensor drift.
1.1.2 Purging the Reaction Cell
As with many chemical reactions the conversion of ammonia in the presence of
other oxides of nitrogen is complicated. It is important to note that the valve
DWELL time for an AZERO measurement has a default setting of 8 seconds.
Shortening this value may not allow enough time to properly purge the reaction cell
of excess nitric oxide from the previous measurements.
In the molybdenum converter operating at 315oC the following significant reactions
are taking place:
Mo + NO2 MoO3 + NO ~100% Efficiency
The M501 NH3, ammonia converter operates at 825oC. At this high temperature,
several reactions occur:
NO NO Loss = ~ 3%
NO2 NO Efficiency = ~97%
NH3 NO Efficiency = ~97%
NH3 NO2 Efficiency = ~5%
As can be seen from the above reactions, the calculation of the ammonia
concentration and overall calibration of the instrument must be done carefully, if
accurate ammonia concentrations are to be measured.
1.2 SPECIAL CONSIDERATIONS FOR AMMONIA MEASUREMENT
Ammonia is a difficult gas to measure due to its chemical characteristics. The
gas tends to adsorb onto surfaces and diffuse into many materials. The following
precautions should be observed when designing ammonia sampling systems and
connecting them to the M201E analyzer:
Do NOT use copper tubing or fittings designed for household plumbing.
Use ONLY Chromatography grade (cleaned, passivated) stainless steel
tubing.
Use ONLY Glass tubing for sample inlet manifold.
ALWAYS keep the tubing from the M702SS calibrator to the analyzer as
short as possible. USE stainless steel tubing throughout, especially from the
ammonia calibration bottle to the calibrator.
If possible HEAT the sample line and DRY the sample gas.
Sample filters Will cause response delays on both the rise and fall of the
sample signal.
12 05206H DCN5910
Teledyne API Model M201E NH3 Analyzer Operator Manual M201E Ammonia Analyzer
These rules apply also to your calibrator’s internal plumbing. It is highly
recommended that you use a M702SS calibrator in conjunction with the M 201E.
The calibrator has stainless steel plumbing throughout. The M700 calibrator
contains Teflon tubing and internal MFC’s that can be affected by the ammonia
gas, therefore, the M700 calibrator is not recommend for use with the M201E
analyzer. It is important to keep the sampling system well maintained.
1.3 SAMPLE FILTRATION
The instrument can be provided with an optional stainless steel sample filter P/N
05571. For minimum response time operation, the instrument should be supplied
with clean sample gas at ambient pressure.
An additional filter resides in the M201E pneumatic pathway. The filter is a
¼” diameter stainless steel sintered filter used to protect the reaction cell flow
control orifice from plugging. This filter will plug rapidly if contaminated sample
gas is not filtered before entering the analyzer.
NOTE: In situations where the sample gas is known to be contaminated with
particulate material it is advisably to add extra filtration. However, it must be
noted that the response time is directly extended through filter additions.
05206H DCN5910 13
M201E Ammonia Analyzer Teledyne API Model M201E NH3 Analyzer Operator Manual
/min + 10% (500 cm3/min bypass to vacuum manifold,
3
/min to reaction cell)
NO calibration 1% of full scale
NH
calibration 2% of full scale
3
o
15-40
C
7” x 17” x 23.6” (18cm x 43cm x 61cm)
43 lbs (20 kg)
24 lbs (11 kg)
16 lbs (7 kg)
100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz, 125
watts
230V ~50 Hz, 125 watts
110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts
230 V ~50 Hz, 2.5 A peak
Installation Category Pollution Degree 2, Over-voltage Category II
4 user configurable outputs, each can be configured as
0.1, 1V, 5V or 10V. Three outputs convertible to 4-20mA isolated
current loop.
1 part in 4096 of selected full-scale voltage (12 bit)
1x RS-232, 1 x RS-485 or RS-232 or Ethernet(optional).
Communication speed: 300 – 115200 baud (user selectable)
8 status outputs from opto-isolators, 7 defined, 1 spare
14 05206H DCN5910
Loading...
+ 35 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.