Teledyne Foxy 200, Foxy Jr, Foxy R1, Foxy R2 User Manual

Clarity Controls
Teledyne Foxy 200
FC ENG
Code/Rev.: M165/70B
Date: 12/21/2016
Phone: +420 251 013 400 DataApex Ltd.
Fax: +420 251 013 401 Petrzilkova 2583/13
www.dataapex.com The Czech Republic
Clarity®, DataApex®and
®
are trademarks of DataApex Ltd. Microsoft®and Windows
TM
are
trademarks of Microsoft Corporation.
DataApex reserves the right to make changes to manuals without prior notice. Updated manuals can be downloaded from www.dataapex.com.
Author: DM
Teledyne Foxy 200 Table of Contents
Contents
1 Teledyne Foxy 200 Control Module 1 2 Requirements 2 3 Installation Procedure 3
3.1 Teledyne Foxy 200 fraction collector communication 3
3.2 Clarity Configuration 4
4 Using the control module 6
4.1 Method Setup - FC - Fraction Table 7
4.2 Method Setup - FC - Vial Numbers 12
4.3 Hardware Configuration 14
4.4 Device Monitor 15
4.5 Data Acquisition window 17
4.6 DataApex UNI Setup 18
5 Report Setup 20 6 Troubleshooting 21
- i -
Table of Contents Clarity Controls
To facilitate the orientation in the Teledyne Foxy 200 manual and Clarity chromatography station, different fonts are used throughout the manual. Meanings of these fonts are:
Instrument (blue text) marks the name of the window to which the text refers.
Open File(italics) describes the commands and names of fields in Clarity, parameters that can be entered into them or a window or dialog name (when you already are in the topic describing the window).
WORK1 (capitals) indicates the name of the file and/or directory. ACTIVE (capital italics) marks the state of the station or its part.
The bold text is sometimes also used for important parts of the text and the name of the Clarity station. Moreover, some sections are written in format other than normal text. These sections are formatted as follows:
Note: Notifies the reader of relevant information.
Caution: Warns the user of possibly dangerous or very important
information.
▌ Marks the problem statement or trouble question.
Description: Presents more detailed information on the problem, describes its causes,
etc.
Solution: Marks the response to the question, presents a procedure how to remove it.
- ii -
Teledyne Foxy 200 1 Teledyne Foxy 200 Control Module
1 Teledyne Foxy 200 Control Module
This manual describes the setting of the Teledyne Foxy 200 fraction collector. The control module enables direct control of the instrument over serial line.
Fig 1: Teledyne Foxy 200 fraction collector
Direct control means that the fraction collector can be completely controlled from the Clarity environment. The Instrument method controlling the analysis conditions will be saved in the measured chromatograms. The control is performed via the UNI Ruby control module and the Teledyne Foxy 200 script.
- 1 -
Teledyne Foxy 200 2 Requirements
2 Requirements
l Free serial COM port in the PC.
Note: Modern computers usually have only one (if any) serial (COM) port
installed. To use more devices requiring the RS232 port, the MultiCOM adapter (p/n MC01) is available.
l Serial DB9F-DB25M cable (p/n SK12).
Note: Cables are not part of Clarity Control Module. It is strongly
recommended that you order the required cables together with the Control Module.
- 2 -
Teledyne Foxy 200 3 Installation Procedure
3 Installation Procedure
3.1 Teledyne Foxy 200 fraction collector communication
The Teledyne Foxy 200 is controlled by serial (RS232) communication. It uses a serial DB9F-DB25M straight cable (p/n SK12) described in the picture below.
Fig 2: Serial DB9F-DB25M cable
On top of that, the fraction collector must be switched to the same Baud Rate as set in Clarity and the ID=5 to work with the control module. To do so, start the fraction collector and load the Program A, then select the Edit A option. Select page 3 from the Edit submenu and set the Serial function, then set the Baud Rate as desired and ID to 5 in the next screen.
- 3 -
3 Installation Procedure Clarity Controls
3.2 Clarity Configuration
Fig 3: How to Add UNI Ruby module
l
Start the Clarity station by clicking on the icon on the desktop.
l Invoke the System Configuration dialog accessible from the Clarity
window using the System - Configuration... command.
l Press the Add button (on Fig 3 on pg 4.) to invoke the Available Control
Modules dialog.
l You can specify the search filter to simplify the finding of the driver.
- 4 -
Teledyne Foxy 200 3 Installation Procedure
l Select the correct item and press the Add (on Fig 3 on pg 4.) button.
Each device with already created UNI profile should have its own item named accordingly in the Available Control Modules dialog.
l The DataApex UNI Setup dialog will appear.
Fig 4: DataApex UNI Setup
l Set the correct communication Port and click on the AutoDetect button to
establish communication with the device.
l You may fill in the custom Device Name.
Note: The DataApex UNI Setup dialog is described in detail in the chapter
"DataApex UNI Setup" on pg 18.
l The Teledyne Foxy 200 item will appear in the Setup Control Modules
list of the System Configuration dialog.
l Drag the appropriate item from the Setup Control Modules list on the left
side to the desired Instrument tab on the right side , or click on the
button .
l Set the Ext. Start Dig. Input and Ready Dig. Output numbers for your
acquisition card according to the wires used for synchronization.
- 5 -
Teledyne Foxy 200 4 Using the control module
4 Using the control module
After adding and setting up the new device one or more new tabs will appear in the Method Setup dialog depending on the type of the instrument. A new Teledyne Foxy 200 fraction collector section enabling the monitoring of the current fraction collector state will be also created in the Device Monitor window.
- 6 -
4 Using the control module Clarity Controls
4.1 Method Setup - FC - Fraction Table
The Method Setup - FC - Fraction Table tab is used for setting the automatic fraction collection program of the Teledyne Foxy 200 fraction collector. The actions of the fraction collector can be displayed in Data
Acquisition window or Chromatogram window, click the right mouse
button on the graph, choose Properties… to open the Graph Properties dialog and check the Show Events option.
Fig 5: Method Setup - FC - Fraction Table
Flow Rate
Sets the flow rate of the liquid phase going to the fraction collector. This parameter, together with the Fraction Volume parameter from the Fraction Table , helps to prevent collection vial overflowing. Flow rate units are taken from the Instrument Method Sending dialog.
Delay Volume Detector - Valve
Sets the volume of the tubing between the detector and waste/collect valve of the fraction collector. Volume units are devised from the Flow Rate Units set in the Instrument Method Sending dialog. The delayed volume is displayed in the Data Acquisition window as a vertical line trailing the signal front.
Delay Volume Valve - Vial
Sets the volume of the tubing between the waste/collect valve and the vial in the fraction collector tray. Volume units are devised from the Flow Rate Units set in the Instrument Method Sending dialog. If the fraction collector has no waste/collect valve, this field should be left empty and he total
- 7 -
Teledyne Foxy 200 4 Using the control module
tubing volume between the detector and fraction collector tray should be typed into the Delay Volume Detector - Valve field.
Fraction Table
Allows to set the fraction collector behavior. The table rows are run in time order, the time intervals should not overlap. If time interval on any row ends while the fraction collector is collecting, the collection will be terminated. Fraction collection based on negative peaks is not supported ­in case you would like to perform such fraction collection it is necessary to invert the detector signal prior to setting up the method.
Name
This name will be displayed in the Special Results table for all fractions collected in the given interval.
Start Time [min]
Beginning of the time interval where the Signal Condition will be evaluated. See the description of Signal Condition below.
Stop Time [min]
End of the time interval where the Signal Condition will be evaluated. When the Stop Time is reached, the fraction collection will be terminated regardless on the current Signal Condition state. See the description of Signal Condition below.
Fraction Volume
Collected amount can be collected to one or more vials/fractions. The Fraction Volume value should be lower or equal to the vial volume to prevent vial overflow. If the collected volume exceeds the set value, the vial will be automatically changed. Fraction Volume units are devised from the Flow Rate Units set in the Instrument Method
Sending dialog.
Detector Signal
Selects the detector used for triggering the collection of fractions according to the Signal Condition in the fraction table. Each row of the table can be related to a different detector signal. Only the signals that are enabled using the Enabled checkbox on the Method Setup -
Acquisition tab of the particular detector will work in the Fraction table.
Signal Condition
Signal Condition is evaluated within time interval set for each row in the Fraction Table . If these conditions are met more than once in a given time interval, the fraction collection will be performed more than once swapping the vial after each collection stop. If the start conditions are not met in the interval defined on fraction table row, the fraction collection will not occur in given time interval.
None - The fractions are collected in the defined time interval ruled by Start Time and Stop Time regardless of signal course. Level - The fraction collection:
- 8 -
4 Using the control module Clarity Controls
l starts when the detector signal increases above the Level Start l stops when the detector signal decreases below the Level Stop
Slope - The fraction collection:
l starts when the first derivative of the detector signal increases above
the Slope Start
l stops when the first derivative of the detector signal first decreases
and then increases above the negative of the Slope Stop. Following image displays the course of the first derivative of the signal and demonstrates Slope Stop activation. Read carefully the values of the slope in all End Fraction points and the course of the first derivative before it reaches these End Fraction points.
Fig 6: Fist derivate course
Note: It is strongly recommended not to use Slope Signal Condition for fraction
collection based on noisy detector signal.
Start L+S End L+S - The fraction collection:
l starts when both start signal conditions are fulfilled simultaneously,
detector signal increases above Level Start value and also the first derivative of the detector signal increases above Slope Start value. If only one of the start signal conditions is fulfilled the fraction collection will not start.
l stops when both stop signal conditions are fulfilled simultaneously,
detector signal decreases below Level Stop value and also the first derivative of the detector signal first decreases and then increases
- 9 -
Teledyne Foxy 200 4 Using the control module
above the negative of the Slope Stop value. If only one of the stop signal conditions is fulfilled the fraction collection will not stop.
Start L+S End L or S - The fraction collection:
l starts when both start signal conditions are fulfilled simultaneously,
detector signal increases above Level Start value and also the first derivative of the detector signal increases above Slope Start value. If only one of the start signal conditions is fulfilled the fraction collection will not start.
l stops when any of the stop signal conditions is fulfilled (whatever
occurs first), detector signal decreases below Level Stop value or the first derivative of the detector signal first decreases and then increases above the negative of the Slope Stop value. If none of the stop signal conditions is fulfilled the fraction collection will not stop before Stop
Time is reached.
Start L or S End L+S - The fraction collection:
l starts when any of the start signal conditions is fulfilled (whatever
occurs first), detector signal increases above Level Start value or the first derivative of the detector signal increases above Slope Start value. If none of the start signal conditions is fulfilled the fraction collection does not start.
l stops when both stop signal conditions are fulfilled simultaneously,
detector signal decreases below Level Stop value and also the first derivative of the detector signal first decreases and then increases above the negative of the Slope Stop value. If only one of the stop signal conditions is fulfilled the fraction collection does not stop.
Start L or S End L or S - The fraction collection:
l starts when any of the start signal conditions is fulfilled (whatever
occurs first), detector signal increases above Level Start value or the first derivative of the detector signal increases above Slope Start value. If none of the start signal conditions is fulfilled the fraction collection does not start.
l stops when any of the stop signal conditions is fulfilled (whatever
occurs first), detector signal decreases below Level Stop value or the first derivative of the detector signal first decreases and then increases above the negative of the Slope Stop value. If none of the stop signal conditions is fulfilled the fraction collection does not stop before Stop Time is reached.
Filter
Sets the data filter on the detector signal defined on the given row of the fraction table. Default value of 1 means no filtering of data will be
- 10 -
4 Using the control module Clarity Controls
performed, value of X means each X data points received will be averaged to 1 point used for fraction start/end conditions detection. The reaction to the fulfilled start/stop signal condition will be delayed the higher the Filter value is set, mainly on detectors with low data rate.
Note: Filter settings influence the value that should be set the Slope Start and
Slope Stop columns. To get the correct trigger value for the Slope Start or Slope Stop using the Show Slope/Level tool in the Chromatogram window, it is necessary to set the Global Bunching value in the
Integration Table to the same value as the desired filter.
- 11 -
Teledyne Foxy 200 4 Using the control module
4.2 Method Setup - FC - Vial Numbers
The Method Setup - FC - Vial Numbers tab is used for setting the behavior of the Teledyne Foxy 200 fraction collector.
Fig 7: Method Setup - FC - Vial Numbers
Start Vial Number
Vial Number
Sets the number of the vial where the first collected fraction will be stored. The current vial number is incremented each time the Next signal is activated. The vial number is stored in the Results Table.
Reset at
Specifies the conditions of the current vial number setting to Start Vial Number in three different situations:
l Each Run - the vial number is reset at the start of each acquisition. l Start Sequence - the vial number is reset only at the start of a
sequence.
l Open Instrument - the vial number is reset only once, when the
Instrument is opened.
Last Vial Number
Vial Number
Sets the number of the last available vial. The Teledyne Foxy 200 fraction collector can set the tray used in the DataApex UNI Setup dialog, so the value range of this field is limited.
- 12 -
4 Using the control module Clarity Controls
After Last Vial
Specifies the action executed when current vial number exceeds the Last Vial Number.
l Waste - the acquisition continues, but the fraction table is ignored and
the samples are automatically discarded.
l Shutdown - the acquisition is immediately stopped and the analysis
aborted.
Vial Change Options
Waste During Vial Change
This option prevents spilling when the vial is changed during collection. It temporary stops the fraction collection before changing the vial and restarts it again when the vial is swapped. The vial is changed in the middle of the wasting interval; the volume wasted during vial change can be specified in the Volume field. Volume units are the ones set in the
Instrument Method Sending dialog. This functionality is not applicable
(even if checked) when Collect All option is selected.
Change Vial when Collecting and Peak Top is Detected
If enabled, the vial is automatically swapped when the peak top is detected during collection. Be aware this option is not applied if the active row in the Fraction Table is set to None in the Signal Condition column.
Collect All
When checked, fractions are collected during the whole course of the analysis, regardless of the settings in the Fraction Table. The FC module then sends the Next Vial command only , which makes this mode of operation suitable for a fraction collector without a waste position. Collect All mode does not permit Waste During Vial Change. The maximum amount of eluent collected in a single vial is limited by the value in Vial Volume field. Once this volume is filled, it begins collecting into the next successive vial.
Note: With Collect All option the valve is not controlled.
Warning: Once error message " Vial number exceeded last vial number " is
displayed, it means that the FC was already at the last position when
command Next Vial was received. Collecting stops at position which is equal to the Last Vial Number - 1 because it's the last possible position to collect (due to Waste). Correct appropriatelyyour method.
- 13 -
Teledyne Foxy 200 4 Using the control module
4.3 Hardware Configuration
The Hardware Configuration dialog (invoked by using the FC Status button from the Method Setup - FC dialog) displays the configuration of the Teledyne Foxy 200 , namely the communication type and its parameters.
Fig 8: Hardware Configuration
- 14 -
4 Using the control module Clarity Controls
4.4 Device Monitor
The window with the fraction collector status can be invoked by the Monitor - Device Monitor command from the Instrument window or using the Device Monitor icon.
Device Monitor - Fraction Collector
Fig 9: Device Monitor - Fraction Collector
Current Position
Displays whether the fraction collector is currently wasting or the position to which it is currently collecting.
Next
Moves the collection vessel one vial forward. Using this button while the collector is performing its program set on the Method Setup - FC - Fraction
Table tab during an analysis run immediately overrides the automatic
collection program. To return to the automatic program, use the Resume Method button.
Collect / Waste
The description on the button changes according to the state in which the fraction collector is - while it is collecting, the Waste button forces the fraction collector to stop collecting, while it is in a waste cycle, the Collect button will force the collector to start collecting. Using any of the two actions while the collector is performing the program set on the Method
Setup - FC - Fraction Table tab during an analysis run, immediately
overrides the automatic collection program. To return to the automatic program, use the Resume Method button.
Position To Go
Allows to set the position to which the fraction collector should be switched. The switching itself is performed using the Go To Position button. This feature is only present for fraction collectors which support addressing particular collection vessels.
Resume Method
Resumes the actions of the automatic fraction table set on the Method
Setup - FC - Fraction Table tab. If the fraction collector is collecting at that
- 15 -
Teledyne Foxy 200 4 Using the control module
moment, it will start a waste cycle and then resume the function of the fraction table.
- 16 -
4 Using the control module Clarity Controls
4.5 Data Acquisition window
Fig 10: Data Acquisition window
The Data Acquisition window can display the fractions in the graph using the background color and also the start and stop fraction events and marker of the Fraction Delay. To see the fractions and markers right click on the graph, select Properties… to open the Graph Properties dialog and check the Show Events checkbox. When the Fraction Delay is set, there will be two colors present in the graph - lighter and darker red. The lighter red color shows the expected duration of the fraction (which Clarity only presumes based on data points it got so far), while the darker red color shows the placement of the real fractions.
Note: The Fraction Delay in the picture will appear if the Delay Volume field in
the Method Setup - FC - Fraction Table dialog is set to a nonzero value.
- 17 -
Teledyne Foxy 200 4 Using the control module
4.6 DataApex UNI Setup
The appearance of the DataApex UNI Setup dialog depends on the presence of the selected Ruby Script - if the script is not present, only the
Ruby Script field is visible.
Fig 11: DataApex UNI Setup
Ruby Script
Displays the selected Ruby Script. The correct TELEDYNEISCOFOXY200FC.RB script for the Teledyne Foxy 200 instrument can be found in the UTILS/UNI_ DRIVERS/ TELEDYNE subdirectory (accessible through the button) of the Clarity installation folder (C:\CLARITY\BIN by default).
Port
Defines the communication port used, possible values dependent on the type of communication of the device and/or available ports in the PC.
AutoDetect
It is used for verifying the device communication over the serial port selected above.
Instrument Name
Allows you to set the custom name of the instrument. This name (entered into the Value column) will be used throughout the Clarity station.
Device Type
Displays the subtype of the fraction collector.
Baud Rate
Sets the Baud Rate communication parameter for the Teledyne Foxy 200 fraction collector. Available values are 110, 300, 600 , 1200 , 2400 , 4800, 9600 , 19200 and 38400 baud. Please note that the value set in the
- 18 -
4 Using the control module Clarity Controls
DataApex UNI Setup dialog must be the same as the one set on the
fraction collector.
Rack Selection
Allows the user to select the correct rack type installed on the Teledyne Foxy 200 fraction collector. The Teledyne Foxy 200 fraction collector
supports two racks but both settings can be read from the fraction collector itself by using the Autodetect button.
Note: The setting of the racks to be used is performed in the Method A of the
fraction collector and cannot be changed from the side of Clarity - this is the limitation of the fraction collector communication protocol.
- 19 -
Teledyne Foxy 200 5 Report Setup
5 Report Setup
The fraction collector section on the method report can be enabled by checking the Instrument Control checkbox on the Method tab of the Report
Setup dialog.
Fig 12: Report Setup
All of the parameters set in the Method Setup - FC - Fraction Table and
Method Setup - FC - Vial Numbers tabs are reported, as well as the
custom FC Name and other parameters set in the DataApex UNI Setup dialog.
- 20 -
Teledyne Foxy 200 6 Troubleshooting
6 Troubleshooting
When the solution to a problem cannot be found easily, a recording of the communication between Clarity and the fraction collector will significantly help DataApex support. The data recording can be enabled by adding or amending the COMMDRV.INI file in the Clarity installation directory (C:\CLARITY\CFG by default). The file can be edited in any text editor (e.g. Notepad). The following section should be edited or added:
[COM1] echo=on textmode=on filename=CommDrvCOM1_%D.txt reset=off
Note: Instead of COM1, type the communication port used to communicate with
the Teledyne Foxy 200 fraction collector. This port number is displayed when the FC Status button in the Method Setup - FC dialog is invoked.
Note: %D (or %d) in the filename parameter means that the log will be created
separately for each day. The reset=off parameter disables deleting the content of the log each time the station is started during the same day.
The created *.TXT files will be of great help in the diagnosis of not documented errors and communication issues.
- 21 -
Loading...