Teledyne Foxy 200, Foxy Jr, Foxy R1, Foxy R2 User Manual

Clarity Controls
Teledyne Foxy 200
FC ENG
Code/Rev.: M165/70B
Date: 12/21/2016
Phone: +420 251 013 400 DataApex Ltd.
Fax: +420 251 013 401 Petrzilkova 2583/13
www.dataapex.com The Czech Republic
Clarity®, DataApex®and
®
are trademarks of DataApex Ltd. Microsoft®and Windows
TM
are
trademarks of Microsoft Corporation.
DataApex reserves the right to make changes to manuals without prior notice. Updated manuals can be downloaded from www.dataapex.com.
Author: DM
Teledyne Foxy 200 Table of Contents
Contents
1 Teledyne Foxy 200 Control Module 1 2 Requirements 2 3 Installation Procedure 3
3.1 Teledyne Foxy 200 fraction collector communication 3
3.2 Clarity Configuration 4
4 Using the control module 6
4.1 Method Setup - FC - Fraction Table 7
4.2 Method Setup - FC - Vial Numbers 12
4.3 Hardware Configuration 14
4.4 Device Monitor 15
4.5 Data Acquisition window 17
4.6 DataApex UNI Setup 18
5 Report Setup 20 6 Troubleshooting 21
- i -
Table of Contents Clarity Controls
To facilitate the orientation in the Teledyne Foxy 200 manual and Clarity chromatography station, different fonts are used throughout the manual. Meanings of these fonts are:
Instrument (blue text) marks the name of the window to which the text refers.
Open File(italics) describes the commands and names of fields in Clarity, parameters that can be entered into them or a window or dialog name (when you already are in the topic describing the window).
WORK1 (capitals) indicates the name of the file and/or directory. ACTIVE (capital italics) marks the state of the station or its part.
The bold text is sometimes also used for important parts of the text and the name of the Clarity station. Moreover, some sections are written in format other than normal text. These sections are formatted as follows:
Note: Notifies the reader of relevant information.
Caution: Warns the user of possibly dangerous or very important
information.
▌ Marks the problem statement or trouble question.
Description: Presents more detailed information on the problem, describes its causes,
etc.
Solution: Marks the response to the question, presents a procedure how to remove it.
- ii -
Teledyne Foxy 200 1 Teledyne Foxy 200 Control Module
1 Teledyne Foxy 200 Control Module
This manual describes the setting of the Teledyne Foxy 200 fraction collector. The control module enables direct control of the instrument over serial line.
Fig 1: Teledyne Foxy 200 fraction collector
Direct control means that the fraction collector can be completely controlled from the Clarity environment. The Instrument method controlling the analysis conditions will be saved in the measured chromatograms. The control is performed via the UNI Ruby control module and the Teledyne Foxy 200 script.
- 1 -
Teledyne Foxy 200 2 Requirements
2 Requirements
l Free serial COM port in the PC.
Note: Modern computers usually have only one (if any) serial (COM) port
installed. To use more devices requiring the RS232 port, the MultiCOM adapter (p/n MC01) is available.
l Serial DB9F-DB25M cable (p/n SK12).
Note: Cables are not part of Clarity Control Module. It is strongly
recommended that you order the required cables together with the Control Module.
- 2 -
Teledyne Foxy 200 3 Installation Procedure
3 Installation Procedure
3.1 Teledyne Foxy 200 fraction collector communication
The Teledyne Foxy 200 is controlled by serial (RS232) communication. It uses a serial DB9F-DB25M straight cable (p/n SK12) described in the picture below.
Fig 2: Serial DB9F-DB25M cable
On top of that, the fraction collector must be switched to the same Baud Rate as set in Clarity and the ID=5 to work with the control module. To do so, start the fraction collector and load the Program A, then select the Edit A option. Select page 3 from the Edit submenu and set the Serial function, then set the Baud Rate as desired and ID to 5 in the next screen.
- 3 -
3 Installation Procedure Clarity Controls
3.2 Clarity Configuration
Fig 3: How to Add UNI Ruby module
l
Start the Clarity station by clicking on the icon on the desktop.
l Invoke the System Configuration dialog accessible from the Clarity
window using the System - Configuration... command.
l Press the Add button (on Fig 3 on pg 4.) to invoke the Available Control
Modules dialog.
l You can specify the search filter to simplify the finding of the driver.
- 4 -
Loading...
+ 17 hidden pages