No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in
part, in any form or by any means, whether it be electronic, mechanical, optical, manual, or otherwise, without prior written consent of Teledyne Analytical Instruments,
16830 Chestnut Street, City of Industry, CA 91749-1580, USA.
Warranty
Teledyne warrants that the goods are free from defects of material and of construction
for a period of 2 years (or less, as the case may be, with consumables) from the date of
shipment from Teledyne. The liability of Teledyne, if any, shall be limited solely to the
replacement or repair of the goods, and shall not include shipping costs or other incidental damages as defined in Section 2-715 of the U.S. Uniform Commercial Code.
Teledyne cannot warrant any damage resulting from unauthorized repair, misuse, improper maintenance, negligence, or other events not caused by Teledyne or our goods.
CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR
ON THE ORDER OF A PHYSICIAN.
Teledyne Analytical Instruments ii
Portable Oxygen Analyzer
About This Manual
This Manual describes the installation, configuration, operation and
maintenance instructions for Model AX300 portable oxygen analyzer.
This Manual contains Cautions, Warnings, and Notes to ensure safe
and efficient performance of the analyzer.
For convenience with operating and maintaining your analyzer, this
manual includes Specifications, Troubleshooting and Maintenance Spare
Parts sections.
How To Use This Manual
This manual is designed to walk you through the initial set-up of
the AX300 Portable Oxygen Analyzer. After you have used it to initially
commission your analyzer, it becomes a quick reference guide to help
you with specific questions or operating challenges.
Before operating the analyzer for the first time, you are advised to
read and familiarize yourself with the contents of this manual.
Teledyne Analytical Instruments iii
AX300
Safety Messages
Your safety and the safety of others are very important. Please carefully read and understand the Caution, Warning and Notes found in several places in the manual.
CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE PUR-
POSES AND IN THE MANNER DESCRIBED IN THIS MANUAL.
IF YOU USE THE ANALYZER IN A MANNER OTHER THAN
THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE
BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED
WITH HAZARDOUS CONSEQUENCES.
Teledyne Analytical Instruments iv
Portable Oxygen Analyzer
Table of Contents
Safety Messages .......................................................................... iv
List of Figures .............................................................................. vii
List of Tables .............................................................................. viii
The Teledyne Analytical Instruments AX300 Oxygen Analyzer, hereinafter referred to as AX300 or analyzer, is a portable analyzer designed to analyze oxygen concentration in medical gas mixtures. Oxygen analysis is linear across the single range of 0-100% using Teledyne’s class R17MED oxygen sensor.
Figure 1-1: AX300 Front View
The AX300’s liquid crystal display (LCD) consists of up to 3.5
characters plus a decimal point indicator capable of displaying up to
105%. An integral battery life indicator is displayed continuously on the
AX300.
The analyzer is powered by three AA alkaline batteries and is designed to operate for 2000 hours on a single set of batteries in a Nonalarm State.
A sensor failure alarm is incorporated which warns the user if the
sensor signal is lost or low. When this occurs, the √SENSOR display
Teledyne Analytical Instruments 1
Introduction AX300
and a visual green light will flash, and the alarm buzzer will activate.
This alarm condition is also triggered when the oxygen concentration at
the point of analysis falls below 18% (Reference: ISO 7767 standard).
For safety reasons, the audible alarm cannot be silenced until the alarm
condition is resolved.
CAUTION: DURING NORMAL USE, ALARM CONDITIONS SHOULD
NOT BE OVERRIDDEN BY ACTIONS SUCH AS HOLDING
THE ALARM SILENCE KEY DEPRESSED OR REMOVING
THE BATTERIES. THE CAUSE FOR THE ALARM CONDITION SHOULD FIRST BE RESOLVED.
CAUTION: WHILE HOLDING DOWN THE ALARM SILENCE KEY MAY
MOMENTARILY SILENCE THE ALARM BUZZER, AUDIBLE
ALARM WILL RETURN UPON DISENGAGING THE ALARM
SILENCE KEY. HOLDING THE KEY FOR LONGER DURATIONS WILL INITIATE THE “STUCK KEY” ALARM, WHICH
CAN BE DEACTIVATED ONLY VIA REMOVING THE BATTERIES FROM THE ANALYZER. ANALYZER SHOULD BE
RECALIBRATED UP ON REPOWERING THE ANALYZER.
NOTE: THE ALARM SILENCE KEY DOES NOT PLAY ANY ROLE IN
THIS VERSION OF THE ANALYZER. IT IS DESIGNED IN
FOR ANY POSSIBLE FUTURE PRODUCT MODIFICATION.
1.1 Features
The following features are standard on the AX300:
Large easy to read 3½-digit LCD (see options)
Automatic LCD back lighting upon key press
Microprocessor controlled
Sensor fail/disconnect alarm indicator (audible and visual)
Non-adjustable concentration alarm below 18% O
2000 operating hours from 3 AA alkaline batteries
Battery status indicator
Stand for upright tabletop deployment
Optional hardware for pole clamping and V block support
Rugged high impact ABS plastic construction
2
Splash resistant case
Long life class R17MED sensor
Teledyne Analytical Instruments 2
Portable Oxygen Analyzer Introduction
0-1 VDC digital output (optional RS-232)
FDA 510(k) market clearance
1.2 Options
The following options are available for the AX 300-I analyzer:
A-Option—2½-digit LCD instead of 3½-digit
Note: Contact the factory for modifying your analyzer for 2½-digit LCD. For
RS-232 reconfiguration, see Section 2.1.5.
B-Option—RS 232 digital output instead of 0-1VDC
Universal Pole Mounting Clamp (P/N CP 2345)
V-Mount Pole Clamp (P/N CP 2344)
V-Mount Wall Adapter P/N B 647)
0-1 VDC Interface Cable (P/NB-75554)
RS 232 Interface Cable (P/N B-75555)
1.3 Applications
The AX300 analyzer is intended for spot-checking the concentration of oxygen in gas mixtures used in applications such as Anesthesia
gas delivery and Respiratory care, and is intended for adult, pediatric
and neonatal populations.
1.4 Operation Summary
The AX300 analyzer uses Teledyne Analytical Instruments’ Patented R17MED oxygen sensor. The sensor output is used to determine
the oxygen gas concentration that is displayed on the LCD screen. The
concentration alarm of AX300 analyzer is non-defeatable and will alarm
when the oxygen concentration falls below the threshold of 18%. This
alarm is tied to sensor fail/disconnect alarm that will also warn the user
of a possible sensor problem when the oxygen concentration is less than
18%. When a fault is detected, the SENSOR indicator is displayed on
the LCD, and the audible and visual alarms are activated.
The Class R17MED oxygen sensor is a disposable sensor. The sensor has a cathode (sensing electrode), an anode (counter electrode), and
a compatible electrolyte, and is packaged in a small plastic sensor body.
Oxygen entering the sensor reacts with the anode and a proportional cur-
Teledyne Analytical Instruments 3
Introduction AX300
rent is collected at the cathode, which is then converted into a digital
signal and displayed on the LCD screen on the analyzer.
Used with the R17MED sensor is a removable plastic diverter. This
diverter is used to facilitate the transport of gas mixtures through the
sensor. The diverter, packaged separately when shipped, is necessary
when the tee adapter is used to sample gas flowing through a tube.
The diverter is not necessary, and should not be used, when the
sensor is placed directly in a chamber or when the sensor is used in confined volume analysis such as incubators and inhalation tents.
CAUTION STATEMENTS:
THE R17MED SENSOR CONTAINS A CAUSTIC ELECTROLYTE AND LEAD. DO NOT TRY TO OPEN THE SENSOR
ASSEMBLY. CHECK THE SENSOR PERIODICALLY FOR
LEAKS. IF THE SENSOR IS LEAKING, REPLACE IT. DO
NOT TRY TO REPAIR IT. CONTACT YOUR DISTRIBUTOR
OR TELEDYNE FOR THE MATERIAL SAFETY DATA SHEET
RELATED TO HANDLING AND DISPOSAL OF THE SENSOR.
THE R17MED SENSOR CONTAINS A CAUSTIC ELECTROLYTE, AND IN THE EVENT OF A LEAK, THE ELECTROLYTE
MAY ATTACK CERTAIN MATERIALS IT IS EXPOSED TO,
SUCH AS ALUMINUM. USE PRECAUTION IN YOUR
CHOICE OF INTERFACE MATERIALS.
REMOVE AND SAVE THE DIVERTER WHEN THE SENSOR
IS USED IN CONFINED VOLUME APPLICATIONS.
THE AX300 OXYGEN ANALYZER, R17MED OXYGEN SENSOR AND ASSOCIATED HARDWARE ARE INTENDED FOR
USE AS SECONDARY DEVICES TO VERIFY THE CONCENTRATION OF OXYGEN IN GAS MIXTURES. DO NOT
USE THESE DEVICES AS PRIMARY LIFE-SUPPORT OR
PRIMARY MONITORING DEVICES.
THE INFORMATION OBTAINED FROM THE AX300
SHOULD NEVER BE USED TO ADJUST A LIFE-SUPPORT
SYSTEM. IT SHOULD ONLY BE USED AS AN INDICATION
THAT THE LIFE SUPPORT SYSTEM OR DEVICE IN THE
BREATHING SYSTEM THAT AX300 IS BEING USED WITH
MIGHT REQUIRE SERVICE AND/OR CALIBRATION.
Teledyne Analytical Instruments 4
Portable Oxygen Analyzer Introduction
Left intentionally blank
Teledyne Analytical Instruments 5
Operating Instructions AX300
Operating Instructions
Notes:
Upon receipt, INSPECT THE ENTIRE UNIT FOR ANY DAMAGE. Check the
unit and all included accessories for broken or loose parts. If damaged, DO NOT
USE. Notify the shipper, your distributor or Teledyne Analytical Instruments.
This equipment is internally powered using 3 AA batteries.
The LOCK/UNLOCK key helps prevent accidental change to the calibration set
point. At least two keys must be pressed in order to modify the critical calibra-
tion set point value.
If the keys are unlocked and no keys are pressed, the analyzer will revert to the
locked mode within nine seconds.
2.1 Setup
To use your AX300 analyzer, first install the batteries, then install the oxy-
gen sensor and then, calibrate the analyzer using the steps outlined in section
2.1.4.
2.1.1 Sensor Installation or Replacement
Note: Install the R17MED oxygen sensor before using the AX300 oxygen analyzer.
1. Remove the new sensor from its protective bag. Inspect the sensor for
damage or electrolyte leakage.
Warning: The sensor has a caustic electrolyte. Do not use a damaged or leaking
Warning: The sensor also contains lead. Leaking or exhausted sensors should be
Warning: The AX300 oxygen analyzer, R17MED oxygen sensor, and associated
sensor. Do not let the electrolyte come in contact with the skin. If it
does, flush affected area with water. Do not attempt to open or repair
the sensor.
handled and disposed of in accordance with local regulations.
hardware are non-sterile devices. Autoclaving or other chemical cleaning may damage the analyzer and the sensor.
Teledyne Analytical Instruments 6
Portable Oxygen Analyzer Operating Instructions
2. Plug one end of the coiled cable into the receptacle on the back of the
R17MED sensor and secure it in place with the capture nut located at the
base of the connector. See Figure 2-1.
Figure 2-1: Installing the R17MED Sensor
3. Plug the other end of the cable into the receptacle located on the side of the
unit and secure it using the capture nut. See Figure 2-2.
Figure 2-2: Sensor Cable Connection to Analyzer
Note: Cable shown is for illustration purposes only.
Note: Do not use the plastic flow diverter with the R17MED sensor when the sensor is
used for diffusion sampling (such as with oxygen tents, incubators, etc.). When
Teledyne Analytical Instruments 7
Operating Instructions AX300
the sensor is used with a Tee Adapter, such as in a breathing system, the diver-
ter must be used as shown in Figure 2-3.
Figure 2-3: Mounting the Sensor in the Tee Adapter
2.1.2 Mounting
The AX300 can be mounted in several ways depending on the optional
equipment ordered at the time of purchase. See Section 1.3.
2.1.2.1 V-Mount Adapter Installation
The V-Mount Adapter consists of a matching plastic plate with integral V-
grooves that attach to the rear of the analyzer.
To install V-Mount Adapter, remove the battery compartment door by prying up the hinged latch at the bottom of the cover, then slide the adapter plate
into grooves provided in rear case. Replace the battery compartment door and
secure door latch. See Figure 2-4.
Note: Door latch fits tightly on the battery cover. Use a coin to gently pry up the latch.
Figure 2-4: V-Mount Adapter Installation
Teledyne Analytical Instruments 8
Portable Oxygen Analyzer Operating Instructions
2.1.2.2 Universal Mounting Clamp Installation
The optional Universal Mounting Clamp comes with a 1/4-20 screw for securing the clamp to the rear of the analyzer. A threaded brass insert is installed on
the back of the analyzer for this purpose. See Figure 2-5.
Figure 2-5: Brass Insert for Universal Mounting Clamp
2.1.3 Battery Installation
Note: Three “AA” alkaline batteries must be installed in the unit before the analyzer
will operate. The unit must be recalibrated whenever new batteries are installed.
To install the batteries:
1. Turn the unit off (if it is on).
2. Hold the analyzer face down in the palm of your hand. Use a coin to
pry up the latch that secures the battery compartment door. Remove
the battery compartment door.
CAUTION: IMPROPER INSTALLATION OF THE BATTERIES MAY RESULT IN
DAMAGE TO THE UNIT AND BATTERIES.
Note: Use alkaline batteries only. Other battery types will produce erroneous battery
test readings.
3. Install 3 “AA” alkaline batteries into the holder as shown in Figure 2-
6. Each battery has its own slot. To insure proper polarity, place the
bottom (flat) or negative end of the battery in the end of the holder
marked “–”. Place the top (button) or positive end of the battery in the
end of the holder marked “+”. Do this for each battery.
Teledyne Analytical Instruments 9
Operating Instructions AX300
Figure 2-6: Installing Batteries
4. Re-install the battery compartment door. When the unit is first turned
on, the screen will momentarily display all LCD segments. During
this period, diagnostic tests are being conducted to ensure that the
circuits are functioning correctly. The unit will activate the audible
and visual alarms for about 1 second. The LCD will flash
continuously indicating the unit is in the unlocked position ready for
calibration.
Note: When batteries are first installed or power is lost for any reason, the analyzer
defaults to the calibration mode. All keys except the CAL and ON/OFF (I/O)
keys will be inoperable until a successful calibration is achieved.
5. Note the battery display located below the oxygen readout display. It
continuously shows a bar graph of the remaining life to the batteries.
Using new batteries will illuminate all 5 segments from the left to the
right end of the bar. The bar graph indicator is continuously on when
the analyzer is powered on.
2.1.4 Calibration
The AX300 should be calibrated before each use and every 8 hours to main-
tain accuracy.
When batteries are removed for any reason or when new batteries are installed, the oxygen analyzer defaults to the calibration mode with the LCD display flashing 00.0. Only the CAL and ON/OFF (I/O) keys are functional at this
point.
Note: For first time and for routine maintenance calibrations, make sure that the sen-
sor and sensor cable are installed correctly before attempting to calibrate the
analyzer.
Teledyne Analytical Instruments 10
Portable Oxygen Analyzer Operating Instructions
Figure 2-7: Calibration Sequence
To calibrate the analyzer (See Figure 2-7):
1. Turn the unit on by pressing the ON/OFF (I/O) key.
2. Check the battery status via the bar graph.
3. If the LCD is not flashing, press the LOCK/UNLOCK key to unlock
the keys. LCD will flash indicating that changes can be made to the
settings. If the batteries have just been installed the LCD will flash
00.0
4. As with most oxygen analyzers, the highest level of accuracy is
achieved when calibration is conducted using 100% oxygen. After installing the flow diverter as noted in Section 2.1.1, insert the sensor
into the plastic tee and connect to a supply of pure dry oxygen flowing at 1-2 liters per minute.
Note: An accessory calibration assembly (P/N C53790) is available from Teledyne for
use with the R17MED sensor
5. Wait about 20 seconds to ensure that the sample line is completely
purged with the calibration gas. Press the CAL key. The LCD will
count down from 9 to 0. The analyzer electronics will measure the
sensor output during this countdown to determine the gas concentra-
tion and selects the calibration range of either 100% or 20.9%. When
the calibration is complete, the LCD will display the gas value.
Note: At the end of the calibration countdown, the analyzer will display a solid number,
indicating acceptance of the calibration value. No further user action is required. If the LOCK/UNLOCK key is still pressed, the display will start blinking,
allowing the recalibration. If this LOCK/UNLOCK key-press is in error, either
Teledyne Analytical Instruments 11
Operating Instructions AX300
press the key once more or allow the software to revert to the locked mode automatically, which will occur wi thin nine seconds of not pressing any function al
keys. Once the analyzer reaches the locked mode, press the LOCK/UNLOCK
key and press the CAL key to repeat calibration.
Note: The AX300 can only be calibrated using certified medical grade 100% oxygen or
ambient level air (20.9% Oxygen). Improper calibration or use of other gas concentrations will either activate the
assume that the gas delivered for calibration was 100% Oxygen. This latter behavior of the analyzer, which is common across other peer group analyzers,
would introduce significant error in analysis. Hence, use only certified medical
grade 100% Oxygen gas or ambient level air (20.9% Oxygen) to calibrate your
analyzer.
SENSOR indicator or render the analyzer to
6. Remove the sensor from the oxygen supply and confirm the LCD
reads less than 22% in room air. It is not necessary for it to read exactly 20.9%.
7. It is important to perform the calibration carefully and thoroughly, us-
ing calibration gases that are free from contaminates. The accuracy of
the analyzer is only as good as the accuracy of calibration gases used
and the accuracy of the procedure used to calibrate the analyzer.
Note: A single point air calibration is not recommended unless the sensor can be ex-
posed to a known source of fresh outdoor air. Hospital room air may often have
oxygen at above normal ambient levels of 20.9%, which will introduce errors in
the calibration. Air calibration should only be used for analyzing oxygen levels
between 21% and 40% and should never be used where a high degree of accuracy is needed.
Note: Never calibrate the unit in humidified gas, as water vapor makes the oxygen
concentration appear lower than it really is. See Section 2.3.1: Humidity.
8. The analyzer is now ready for use.
2.1.5 Output 0-1 VDC or RS232
The AX300 Portable Oxygen Analyzer comes with optional 0-1 VDC and
RS232 signal outputs for use with compatible external recording, indicating or
computing equipment.
An optional 0-1 VDC Interface Cable (P/N B-75554) is available from Teledyne for accessing the voltage output.
To connect the analyzer to an analog recording/indicating device:
1. Insert one end of the interface cable into the output port on the side of
the analyzer. See Figure 2-8.
2. Insert the other end into the analog recorder/indicator device
equipped to handle a 0-1 VDC signal.
Teledyne Analytical Instruments 12
Portable Oxygen Analyzer Operating Instructions
When properly calibrated, the output signal generated by the analyzer is li-
near and proportional to the oxygen concentration.
If you requested Option-B (RS 232 digital output) at the time of purchase, a
digital RS 232 signal is output from the output port shown in Figure 2-8. Use the
optional RS 232 Interface Cable (P/N B-75555) available from Teledyne for
connection to a standard RS 232 port on a computer or other suitably equipped
digital device.
CAUTION: AX300 ANALYZER SHOULD ONLY BE CONNECTED TO AN
EN60601-1/IEC60606-1 APPROVED DEVICE.
Figure 2-8: 0-1 VDC or RS 232 Digital Output Port
RS232 signal output has been tested with Windows HyperTerminal. The setup
changes with different windows versions. If other software other than windows
HyperTerminal is used, consult with factory.
Note: Prior to every use, always test the batteries, check the sensor for any leaks or
damage, and check the analyzer calibration.
The AX300 analyzer can be used to measure a gas mixture for oxygen in
two basic modes:
In the inhalation side of breathing system upstream of any antibac-
terial filters, humidifiers and other such medicating devices
In confined volumes such as in incubators or tents
When analyzing for oxygen in breathing systems, the flow diverter must be
used. The diverter should be screwed onto the threaded front end of the
R17MED sensor. A tee adapter (plastic, P/N A268, or metal, P/N A283) should
be placed into the breathing loop, and the above sensor assembly plugged into
the tee adapter. See Figure 2-3.
CAUTION: CHECK THE BREATHING SYSTEM FOR LEAKS. BE CERTAIN THAT
THE SYSTEM DOWNSTREAM OF THE SENSOR DOES NOT PRODUCE ANY BACKPRESSURE OR RESTRICTION TO FLOW. ERRORS IN READINGS WILL RESULT IF THIS IS NOT FOLLOWED.
THE OXYGEN SENSOR IS A NON-STERILE DEVICE AND SHOULD
BE USED IN CONJUCTION WITH AN ANTI-BACTERIAL FILTER. ALWAYS INSTALL THE SENSOR ON THE INSPIRED LINE AHEAD OF
FILTERS, HUMIDIFIERS AND MEDICATING DEVICES. NEVER INSTALL THE SENSOR IN A LOCATION THAT WILL EXPOSE THE
SENSOR TO PATIENTS’ EXHALED BREATH OR SECRETIONS UNLESS YOU INTEND TO DISPOSE OF THE SENSOR AND FLOW
ADAPTER AFTER USE.
When analyzing for oxygen in confined volumes such as incubators, hoods,
etc., the flow diverter must be removed from the R17MED sensor so that it does
not interfere with the rapid exchange of gases to and from the sensing surface of
the sensor. Failure to remove the diverter in these applications may adversely
affect the response time of the sensor.
The R17MED sensor can be placed or hung inside incubators, tents, etc.
When it is necessary to thread the cable through a small hole in order to gain
access to the inside of a chamber, the cable should be disconnected at the sensor,
threaded through the hole, and reconnected inside the chamber.
Teledyne Analytical Instruments 15
Operating Instructions AX300
2.3 Factors Affecting Sampling Accuracy
2.3.1 Humidity
Humidity does not directly affect the accuracy of the sensor's measurement.
However, when a nebulizer or other device is used to increase moisture levels in
gas mixtures, the moisture actually dilutes the mixture. This dilution effect decreases the oxygen concentration. Your portable oxygen analyzer accurately
measures the decreased oxygen concentration.
As with all diffusion based oxygen sensors, excessive condensation on the
sensing surface of the R17MED will block the diffusion of oxygen to the sensor,
rendering it inoperative. We recommend installing the sensor on the dry side of
the breathing system at all times.
2.3.2 Temperature
The R17MED oxygen sensor compensates for ambient temperature changes
in the range of 0–40°C (32–104°F). Since the thermistor that compensates for
these changes is located in the rear of the sensor assembly, it is important that
gas mixtures, flowing over the front of the sensor, be at room temperature, which
is assumed to the temperature around the rear of the sensor. Reading errors may
occur if gases at higher temperature from devices such as heated humidifier are
used to measure oxygen with R17MED oxygen sensor.
A small thermal tracking error may be encountered in applications where
the entire sensor assembly is placed in the gas mixture to be analyzed (such as in
the incubators). Holding the sensor in your hand for more than a few minutes can
also affect the temperature tracking which appears as a slow drift on the LCD.
No adjustments should be made during this period since this error will be eliminated when both the thermistor and sensing electrode have had sufficient time to
come to thermal equilibrium. This can take up to 2 hours.
2.3.3 Pressure
Virtually all gas sensors and analyzers measure the partial pressure of the
gas that they sample. The only time that these analyzers can accurately read percentages is when the total pressure does not vary over time between calibrations
and use. For this reason, it is important to calibrate the AX300 oxygen analyzer
(with R17MED oxygen sensor) at regular intervals. It is recommended that the
unit be calibrated prior to each use or every 8 hours. It is recommended that the
analyzer be calibrated at the same pressure as the gas it will be used to analyze.
If the gas being analyzed flows through some length of tubing, use similar tubing/apparatus and same pressure/flow rates when calibrating the analyzer.
Teledyne Analytical Instruments 16
Portable Oxygen Analyzer Operating Instructions
When a ventilator is a part of the breathing system, the alternating “breathing” pressure cycles generated by the ventilator will be sensed as pressure pulses
by the sensor, and over time, such pulses manifest as pressure increases in the
fast-reacting sensors such as R17MED. This perceived pressure increase results
in correspondingly higher oxygen reading. While the concentration of oxygen is
not changing, it is the total pressure that is increasing producing a corresponding
increase in the partial pressure of oxygen.
2.3.4 Discrepancy in Readings
When a discrepancy in oxygen readings is detected, the oxygen analyzer’s
readings should be verified by checking the AX300 battery condition and calibration using 100% oxygen. If the analyzer can be calibrated, the unit can be assumed to be in good working order. If, after reinstalling the unit, the discrepancy
in oxygen readings persists, the problem is most likely elsewhere in the breathing
system. Resolve the cause(s) for the discrepancy before using the analyzer. The
troubleshooting section of this manual may provide additional assistance in locating the problem.
2.3.5 Anesthetic Gases
2.3.5.1 Gases That MAY INDUCE Reading Error
When using the R17MED sensor in the presence of anesthetic gases, the
oxygen reading may fall (see Table below). The magnitude of this error will depend upon the level of oxygen and the duration of exposure.
The anesthetic agents listed in the following table (Halothane, Enflurane,
Isoflurane, Sevoflurane, and Desflurane) were vaporized into a stream of 30%
oxygen / 70% nitrous oxide, and the resulting drops in oxygen level after an exposure of approximately two hours were noted.
Exposures in excess of two hours may produce slightly greater errors. The errors listed are typical for all oxygen sensors such as the R17MED. Exposing the
sensor to air or gases that do not contain anesthetic agents for a period of time
equal to or greater than the exposure interval will eliminate the reading error in
most cases.
Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas
Gas or Vapor Level
(Balance: Mixture of 30% O2 / 70% N2O, except where noted)
Gas or Vapor Test Level Oxygen Reading
Teledyne Analytical Instruments 17
Operating Instructions AX300
Error
Helium 50%, balance O2 0%
Nitrous Oxide 80%, balance O2 0%
Carbon Dioxide 10%, balance O2 0%
Halothane 4% < 1.5% O2 *
Enflurane 5% < 1.5% O2 *
Isoflurane 5% < 1.5% O2 *
Sevoflurane 5% < 1.5% O2 *
Desflurane 15% < 1.5% O2 *
* Errors are approximate and may vary based on exposure times and concentra-
tions.
These performances meet or exceed the requirements of ISO 7767: 1997 (E).
CAUTION: THE AX300 SHOULD NOT BE USED IN THE PRESENCE OF FLAM-
MABLE ANESTHETICS SUCH AS DIETHYL ETHER OR CYCLOPROPANE.
CAUTION: THE ANALYZER, THE OXYGEN SENSOR AND ASSOCIATED
HARDWARE ARE NON-STERILE DEVICES. DO NOT AUTOCLAVE
THE ANALYZER OR SENSOR, AS THIS WILL DAMAGE THE
EQUIPMENT.
2.3.5.2 Care After Use in Nitrous Oxide
CAUTION: THE R17MED SENSOR SHOULD NOT BE LEFT IN NITROUS OXIDE
MIXTURES ANY LONGER THAN NECESSARY.
After exposure to nitrous oxide mixtures, the sensor should be left in 100%
oxygen overnight (e.g., left in a breathing system that has been flushed with pure
oxygen). If this is not practical, when using the tee, remove the plastic flow diverter and leave the sensor in room air. If the oxygen reading continues to drop
after each use in nitrous oxide, the sensor should be removed from service. If the
sensor can no longer be calibrated or if there is any sign of electrolyte leakage,
the sensor should be disposed of in accordance with local regulations and the
Material Safety Data Sheet (MSDS) available through Teledyne.
Teledyne Analytical Instruments 18
Portable Oxygen Analyzer Operating Instructions
2.3.6 Cleaning
The AX300 and R17MED sensor are non-sterile devices.
The oxygen analyzer, oxygen sensor, and sensor interconnection cable may
be cleaned by wiping the surfaces with isopropyl alcohol or a mild cleaning solution.
When cleaning, do not allow liquids or moisture to enter the analyzer or
sensor internal cavities. Do not expose the electrical connections to the cleaning
solution. Do not immerse the analyzer, oxygen sensor or sensor interconnection
cable in water or any other liquid.
Do not expose the analyzer, oxygen sensor and interconnection cable to
steam, ethylene oxide, or radiation sterilization.
After cleaning, ensure that all surfaces are dry and the analyzer is properly
calibrated before using the analyzer.
2.4 Dos and Don’ts
– DO –
Read and understand the manual before using the analyzer
Calibrate every 8 hours or before every use.
Visually inspect the sensor for leakage before each use.
Calibrate using 100% oxygen and check in air.
Test batteries regularly and replace when battery indicator shows low
battery (no bars)
Keep the analyzer, sensor and connections dry, or on the dry side of
the breathing system.
Recalibrate after replacing the batteries.
Recalibrate after replacing the sensor.
Use only alkaline batteries.
Ensure that the R17MED sensor is properly connected to the analyz-
er.
Always use the plastic flow diverter when using the tee adapter.
Remove and save the plastic flow diverter when using the sensor in
non-flowing applications
Use only isopropyl alcohol or mild detergent, if cleaning is required,
Teledyne Analytical Instruments 19
Operating Instructions AX300
– DON’T –
Use this analyzer if you suspect any malfunction.
Use the analyzer in the presence of flammable gases.
Use anything but alkaline batteries.
Autoclave or freeze the sensor or analyzer.
Open or try to repair a leaking or broken sensor.
Immerse the unit or sensor in any liquid.
Pass hot or cold gas mixtures over the sensor.
Expose the unit to devices that produce high levels of radio, short
wave, microwave, x-ray, or high frequency interference.
Use cleaning agents or liquids in the cable receptacles or around the
battery compartment.
Place the unit in a water vapor-saturated environment.
Expose the LCD to excessive sunlight.
Expose the unit to a condensing water environment such as a mist
tent.
Teledyne Analytical Instruments 20
Portable Oxygen Analyzer Service Instructions
Service Instructions
3.1 General Service Information
Without access to special probes and test equipment, troubleshooting and repair
of the board or its components are not feasible. A factory replacement of the entire
PC board is recommended if necessary.
With the exception of replacing the sensor or batteries, there are no userserviceable components inside the unit. If a problem arises that cannot be corrected
by recalibration, changing the batteries or replacing the sensor as described in this
manual, the unit must be sent back to the factory for repair or replacement. See Section 3.10 for instructions on obtaining a Return Merchandise Authorization (RMA)
number before sending a unit back to Teledyne for repair.
3.2 Overall Maintenance
The AX300 analyzer requires very little maintenance, other than calibration,
checking the batteries and sensor. Occasional cleaning of the plastic surface can
be done with isopropyl alcohol. There are no user-serviceable components within
the analyzer.
3.3 Battery Maintenance
DO: Test batteries regularly. (Replace immediately when all battery
strength indicator bars are missing).
DO: Always use alkaline batteries.
DO: Recalibrate after replacing batteries.
The AX300 analyzer has a battery strength indicator that continuously displays the approximate amount of useful life remaining on the set of installed batteries. Excessive alarm activation will drain the batteries faster. The minimum
detectable change in battery voltage corresponds to an increment of about 50
hours, meaning that the battery voltage reading may not change for several hours
at a time.
If the analyzer is not used for a period of 30 days or more, the batteries
should be removed from the analyzer.
Teledyne Analytical Instruments 21
Service Instructions AX300
3.4 Sensor Maintenance
DO check the sensor for damage or leaks before use.
DON’T immerse the R17MED sensor in liquid.
DON’T autoclave the R17MED sensor.
DON’T open or try to repair the sensor.
Before every use, the sensor, cable and connections should be checked.
Check the sensor for leaks and condensation. Check the cable for splitting or
cracked insulation. Make sure the connections are tight and dry.
In the event that the sensor has been damaged, consult the Material Safety
Data Sheet in the Appendix for handling guidelines.
3.5 Calibration
Incorrect readings can often be traced to improper calibration. The AX300
should be calibrated before each use and every 8 hours to maintain accuracy. It
must be calibrated whenever new sensor or batteries are installed. Calibration
using methods other than described in section 2.1.4 will lead to improper operation.
Whenever new batteries are installed or removed for any reason, the oxygen
analyzer defaults to the calibration mode with the LCD display flashing 00.0.
Only the CAL and ON/OFF (I/O) keys are functional at this point.
3.6 Factors Affecting Sampling Accuracy
Refer to Section 2.3 of this manual.
3.7 Troubleshooting
The AX300 oxygen analyzer provides a variety of built-in safety features that
prevent its use when a fault is detected. When a unit displays the message √ SENSOR
and sounds the audible and visual alarm continuously, it is an indication of a faulty
connection between the sensor and the unit or an expired or faulty sensor. It also indicates when oxygen concentrations are below 18%.
For assistance with troubleshooting, see Table 3-1.
Teledyne Analytical Instruments 22
Portable Oxygen Analyzer Service Instructions
Table 3-1 Troubleshooting
Symptom Why What To Do
New sensor responds
slow or drifts.
Sensor will not read below 22 % after calibration
in 100% O2.
The sensor does not
react to changes in
oxygen concentration,
or the readings are unstable and drifting.
If the sensor is new
and was just removed
from its sealed bag it
may need to run for
several hours.
Calibration in 100%
was invalid or the
room air enriched with
excess oxygen.
Water may have condensed on the sensing surface.
Electrical interference
is disrupting the electronics
A) Wait 1–2 hours and recali-
brate.
A) Recalibrate using dry gas
B) Make sure that at least 6"
(30 cm.) of tubing is attached
to the exhaust side of the tee
adapter to prevent back filling. O2 flow rate should not
exceed 5 l/min.
C) Oxygen concentration at the
sensor is significantly higher
than 21%. Take the analyzer to a well-ventilated area
and check the reading
again.
D) Try calibrating with a known
good sensor, and if this fails,
contact customer support.
A) Remove the sensor from tee
adapter and unscrew the plastic flow diverter. Using absorbent tissue or cotton swab,
gently wipe off sensing surface inside threaded portion of
sensor assembly. Do not
damage the sensing surface,
and do not leave any tissue/swab residue on the
sensing membrane.
B) Relocate unit away from
sources of electrical noise
such as cauterizing equipment
and two-way radios.
The display is flashing
√ SENSOR
The unit has detected
a fault in the signal
from the sensor.
Sensor has expired.
The sensor has been
exposed to a gas containing less than 18%
A) Check sensor cable con-
nections to ensure they are
completely inserted into the
mating connector and the
capture nut is firmly in
place.
B) Expose the sensor in 100%
Teledyne Analytical Instruments 23
Service Instructions AX300
Symptom Why What To Do
oxygen. O2 and check calibration.
C) The sensor output has fal-
len to a level where it is no
longer usable. Replace
sensor.
The oxygen reading
fluctuates or appears to
be incorrect.
The unit has stopped
working and the LCD is
displaying alphanumeric characters.
The sample pressure
may be changing.
The AX300 unit is
equipped with an electronic "watch dog,"
which analyzes the
circuitry within the unit
for potential faults and
renders the unit inoperable until the condition is corrected. (See
Watch dog section)
Several conditions can
activate the "watch
dog." Dropping the unit,
poor battery connections, and radio frequency interference are
the most common
causes. See the watch
dog section for additional information
A) During calibration, make
sure there is no restriction
on exhaust side of sensor. If
the reading changes with
flow, the sensor is pressurized or there may be a leak
in the system.
B) If a high degree of accuracy
is desired, or the concentration of O2 being analyzed is
in excess of 40%, calibration
with 100% is recommended.
C) If humidified gas is used to
ventilate the patient, water
vapor actually dilutes the
gas. See Section 2.3. .
D) If a blender is used, check
its calibration.
A) Disconnect the batteries and
inspect the contacts for corrosion. Reconnect the batteries. If the unit functions
properly, calibrate the unit
and reset the alarm values.
B) Try a new set of batteries.
C) Increase the distance be-
tween the unit and any
source of radio frequency interference. The sensor cable is a prime source of
pickup as it can act like an
antenna. Relocate the sensor cable and if possible
change its coiled length to
“de-tune” its antenna effect.
Placing the cable in a different position may also help.
No display A) Batteries expired.
B) Bad battery connec-
A) Check/replace batteries.
B) Check battery connections.
Teledyne Analytical Instruments 24
Portable Oxygen Analyzer Service Instructions
Symptom Why What To Do
tion
C) Calibrate.
Note: In the event that none of these procedures produces desired results, remove the
batteries and return the unit to Teledyne for repair.
3.8 Watchdog
The AX300 is equipped with a watchdog circuit that continuously analyzes the
electronics for proper operation. If the watchdog detects a failure, one of the codes in
Table 3.3 will appear on the LCD.
The error codes can appear on the LCD when batteries are first installed,
during normal operation or if the unit is subjected to shock. In some cases, an
additional digit is used in the error code to supply additional information. For
example, the error code 6 and 7 are followed by another digit listed as (N) in Table 3.2. The error code 65 would indicate that a key is stuck and this key is the
Silence key.
In addition to supplying visual error codes, the audio device will beep a
number of times to indicate the general error in case the display is not functional.
Note: To reset the watchdog error code, remove one battery for 5 seconds and re-
place. If the error persists, contact your local representative or Teledyne fo r assistance.
Table 3.2 Error Codes
Error Code Audio Beeps Error
Indeterminate 2 The watchdog timer has timed out indicating
a serious software error
30 3 Analog output is different from the expected
value. May indicate a shorted or over-loaded
analog output or a failure of the analog to
digital converter or digital to analog converter circuit.
Teledyne Analytical Instruments 25
Service Instructions AX300
50 5 The ADC circuit failed during POST.
6 (5) 6
(There is no
indication of
which key is
stuck)
7(2) 7
(There is no
indication of
the type of
DAC failure)
A stuck key has been detected. The second
digit example (5) on the display shows
which key is stuck:
1 - Batt
2 - Cal
5 - Silence
8 - Key Lock
A failure has occurred during the automatic
calibration of the digital to analog converter
(DAC) circuit. The second digit example (2)
shows the type of digital to analog failure.
0 - Measurement
1 - High Test
2 - Low Test
3 - Offset Cal
4 - Gain Cal
3.9 Other Problems with the Analyzer
Most other problems arise from either mechanical damage to the
analyzer or electronic component failure.
Repair or troubleshooting the PCB or individual components on the
board is not feasible. Under most circumstances, a replacement of the
entire PC Board is recommended. If the repair diagnosis indicates problem with the PCB, the analyzer must be returned to the factory for PCB
replacement.
3.10 Return Authorization for Service
For any service beyond sensor and battery replacement, the analyzer must be returned to the factory. A return merchandise authorization
(RMA) number must be obtained from Teledyne Analytical Instruments
prior to returning the analyzer for service. You can request a RMA
number via email by contacting us at:
Teledyne Analytical Instruments 26
Portable Oxygen Analyzer Appendix
tetci_customerservice@teledyne.com
You can also contact us at the address below.
Customer Service Department
TELEDYNE Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91749-1580 USA
Phone (626) 934-1500, Fax (626) 961-2538
Or via the web at: www.teledyne-ai.com
Teledyne Analytical Instruments 27
Appendix AX300
Appendix
A.1 Specifications
Range: 0-100% oxygen
Accuracy: +
Response Time: 90% in less than 6 seconds at 25 °C
Battery Life: Approximately 2000 hours of continuous
System Power: 3 AA alkaline batteries.
Sensor Type: Class R17MED
Alarm: Non-adjustable SENSOR alarm
Expected Life: 36 months in air at 25°C, Atmospheric
Dimensions: 2.5" W × 1.25" D × 4.5" H (66 mm x 33-
Sensor Cable: Retracted: 2 ft / Extended: 10 ft.
2% of full scale (at constant temperature
and pressure)
use in non-alarm condition
activates upon sensor problem or O
centration falls below 18%
Pressure and 50% Relative Humidity.
(~7.5 months under similar conditions
when continuously exposed to 100% oxy-