TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
SAFETY MESSAGES
Your safety and the safety of others is very important. We have provided many important safety messages in
this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety message is
associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The
definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions for details on the
specific danger.
CAUTION: Hot Surface Warning
CAUTION: Electrical Shock Hazard
Technician Symbol: All operations marked with this symbol are to be performed
by qualified maintenance personnel only.
Electrical Ground: This symbol inside the instrument marks the central safety
grounding point for the instrument.
CAUTION
The analyzer should only be used for the purpose
and in the manner described in this manual.
If you use the analyzer in a manner other than that for which
it was intended, unpredictable behavior could ensue with
possible hazardous consequences.
ii
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
INTENTIONALLY BLANK
x
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
1 HOW TO USE THIS MANUAL
The Model 9110AH has been designed to produce accurate data, be serviceable, reliable and
easy to use. The M9110AH's microprocessor continually checks operating parameters such as
temperature, flow, and critical voltages. The instruments design uses top mounted, modular
components with captive screws to facilitate repair and ease of access. If you encounter any
difficulty refer to the Troubleshooting Section 9 - General Hints.
We recognize that the need for information from this manual changes as time passes. When the
instrument first arrives, it is necessary to get it up and running quickly and verify its correct
operation. As time passes, more detailed information is often required on special
configurations, calibration alternatives and other operational details. Finally there is the need
for periodic maintenance and to quickly troubleshoot problems to assure maximum reliability
and data integrity.
To address these needs, we have created three indexes to the information inside. They are:
Table of Contents:
Outlines the contents of the manual in the order the information is presented. This is a good
overview of the topics covered in the manual. There is also a list of Tables and a list of Figures.
Index to M9110AH Front Panel Menus:
The Menu Index briefly describes the front panel menus and refers you to other sections of the
manual that have a detailed explanation of each menu selection.
Troubleshooting Section 9:
The Troubleshooting Section allows you to diagnose and repair the instrument based on
variables in the TEST menu, the results of DIAGNOSTIC tests, and performance faults such as
excessive noise or drift. The troubleshooting section also explains the operation, adjustment,
diagnosis and testing of each instrument subsystem.
If you are unpacking the instrument for the first time, please refer to Getting Started in
Section 2.
1-1
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
INTENTIONALLY BLANK
1-2
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
2 GETTING STARTED
2.1 Unpacking
CAUTION
Your safety and the safety of others is very important. We
have provided many important safety messages in this manual.
Please read these messages carefully.
To avoid personal injury, always use two persons to
lift and carry the Model 9110AH.
1. Before operation it is necessary to remove the shipping hold-down screws. Remove the
instrument cover, then remove 2 screws as shown in Figure 2-1.
2. Also check for internal shipping damage, and generally inspect the interior of the instrument
to make sure all circuit boards and other components are in good shape.
3. Please check the voltage and frequency label on the serial number tag on the rear panel.
Compare that to your local power before plugging in the Instrument.
2.2 Electrical and Pneumatic Connections
1. Refer to Figure 2-2 to locate the rear panel electrical and pneumatic connections.
2. Attach the pump to the “Exhaust Out” port on the instrument rear panel. The exhaust from
the pump should also be vented to atmospheric pressure.
3. Attach the sample inlet line to the sample inlet port. For initial testing, sample gas can be
calibration gas or stack gas. The pressure of the sample gas at the inlet port should be at
ambient pressure and constant. See Figure 2-3.
4. If desired, attach the analog output connections to a strip chart recorder and/or datalogger.
Refer to Figure 9-4 - Analog Output Voltage Ranges - for switch settings. Factory default
setting is 0-5 VDC.
5. Connect the power cord to the correct voltage line, then turn to Section 2.3 Initial
Operation.
2-1
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
electrical subassemblies.
WARNING
Analyzer Exhaust – O3 Scrubber – Pump Pack
Danger – Analyzer exhaust contains ozone.
Ozone scrubber must always be present between
analyzer exhaust and pump.
Vent pump exhaust to well ventilated area at atmosphere
pressure FIRE or EXPLOSION HAZARD.
WARNING
Lethal voltages present inside case.
Do not operate with cover off during normal operation.
Before operation check for correct
input voltage and frequency.
Do not operate without proper chassis grounding.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting
2-2
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
Figure 2-1: Removal of Shipping Screws & Check for Correct Power
2-3
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
Figure 2-2: Rear Panel
2-4
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
Figure 2-3: Inlet and Exhaust Venting Recommendations
2-5
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
2.3 Initial Operation
1. Turn on the instrument power.
2. The display should immediately light, showing the computer’s memory configuration, then
the instrument type - M9110AH. If you are unfamiliar with the M9110AH, we recommend
that you read the overview Section 4 before proceeding. A diagram of the software menu
trees is in Figure 5-1 and Figure 5-2.
3. The M9110AH requires about 30 minutes for all internal components to come to
temperature. During this time the ozone generator power is OFF until the membrane dryer
has time to purge itself, therefore there will be no response from the instrument, even if
span gas is coming in the sample port. Many warning conditions are not displayed during
this time, even though temperatures and other conditions are out of specification. All
warning messages are enabled after 30 minutes of operation.
4. While waiting for instrument temperatures to stabilize, you can check for correct operation
by using some of the M9110AH's diagnostic and test features.
5. Examine the TEST functions by comparing the values listed in Table 2-1 to those in the
display. Remember that as the instrument warms up the values may not have reached their
final values yet. If you would like to know more about the meaning and utility of each
TEST function refer to Table 9-1. Also, now is a good time to verify that the instrument was
shipped with the options you ordered. Table 2-1 also contains the list of options. Section 6
covers setting up the options.
6. Electric Test and Optic Test both generate simulated signals in the M9110AH.
A. Electric Test checks the electronics of the PMT signal path. To operate Electric Test from
the front panel:
1) Scroll the TEST function to PMT.
2) Press SETUP-MORE-DIAG, then press ENTR to accept the default password. Scroll
to Electric Test and press ENTR to turn it on. Instrument responses should come to
the values indicated in Table 2-1. To turn off this test press EXIT. For more
information on the circuitry being tested refer to the Troubleshooting Section 9.1.3.2.
B. Optic Test is an "end to end" test of the analyzer HVPS-PMT-detector-electronics-
computer. It simulates a signal by turning on a LED in the Sensor Module. To operate
Optic Test from the front panel:
1) Scroll the TEST function to PMT.
2) Press SETUP-MORE-DIAG, then press ENTR to accept the default password. Scroll
to Optic Test and press ENTR to turn on optic test. Instrument response should come
up to the values indicated in Table 2-1. To turn off this test press EXIT. To return to
the SAMPLE mode press EXIT until SAMPLE is displayed in the upper left display.
For more information about OT operation see Section 9.1.3.3.
2-6
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
7. When the instrument is warmed up, re-check the TEST functions against Table 2-1. All of
the readings should compare closely with those in the table. If they do not see Section 9.1.1.
The next task is to calibrate the analyzer. There are several ways to do a calibration, they are
summarized in Table 7-1. For a preliminary checkout we recommend calibration with span
gas coming in through the sample port. The procedure is:
Step 1 - Set the range, then enter the expected NOx and NO span gas concentrations:
Step Number Action Comment
1. Press
CAL-CONC-NOX
2. Press ENTR ENTR stores the expected NOx span value. The internal
3. Press
CAL-CONC-NO
4. Press ENTR ENTR stores the expected NO span value. The internal
5. Press EXIT Returns instrument to SAMPLE mode.
6. Press
SETUP-RNGEMODE-SNGL
7. Press
SETUP-RNGESET
This key sequence causes the M9110AH to prompt for the
expected span concentration. Enter the span value by pressing
the key under each digit until the expected value is set.
formulas are adjusted to compute this number when span gas
concentration is input into the instrument.
In the same CAL-CONC sub menu press the NO button and
enter the expected NO span value. As before this value will be
used in the internal formulas to compute the NO concentration
value.
formulas are adjusted to compute this number when span gas
concentration is input into the instrument.
If necessary, you may want to change ranges. Normally the
instrument is shipped in single range mode set at 100 ppm. We
recommend doing the initial checkout on the 100 ppm range.
After SETUP-RNGE-SET, enter 100 and press ENTR. The
instrument will now be in the 100 ppm range.
2-7
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
Step 2 - Calibrate the instrument:
Zero/Span Calibration Procedure
Step Number Action Comment
1. Input Zero gas Allow Zero gas to enter the sample port on the rear of the
instrument.
2. Press CAL The M9110AH enters the calibrate mode from sample mode.
3. Wait 10 min Wait for reading to stabilize at the zero value. If you wait less
than 10 minutes the final zero value may drift.
4. Press ZERO The ZERO button will be displayed.
5. Press ENTR Pressing ENTR actually changes the equations and zeros the
instrument.
6. Press EXIT, input
Span gas
7. Wait 10 min Wait for reading to stabilize at the span value.
8. Press SPAN The SPAN button should be displayed. If there is no SPAN
9. Press ENTR Pressing ENTR actually changes the equations so that the
10. Press EXIT Pressing EXIT returns the instrument to SAMPLE mode.
M9110AH returns to the CAL menu. Now switch gas streams
to span gas.
button check the Troubleshooting Section 9.2.8 for
instructions on how to proceed. In certain circumstances at
low span gas concentrations both the ZERO and SPAN
buttons will appear.
concentration displayed is the same as the expected span
concentration you entered above, thus spanning the instrument.
2-8
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
Step 3 - Review Quality of calibration:
Calibration Quality Check Procedure
Step Number Action Comment
1. Scroll the TEST
function menu until
the NOx SLOPE is
displayed.
2. Scroll the TEST
function menu until
the NO SLOPE is
displayed.
4. Scroll the TEST
function menu until
the NOx OFFS is
displayed.
5. Scroll the TEST
function menu until
the NO OFFS is
displayed.
The SLOPE value for NOx should be 1.0 0.3. If the value is
not in this range, check Section 7.1 or 9. If the SLOPE value is
in the acceptable range the instrument will perform optimally.
The SLOPE value for NO should be 1.0 0.3. If the value is
not in this range, check Section 7.1 or 9. If the SLOPE is in the
acceptable range the instrument will perform optimally.
NOTE:
The NO and NOx slopes should be equal within 0.3.
The M9110AH will display the OFFSET parameter for the
NOx equation. This number should be near zero. A value of
0.0 50 indicates calibration in the optimal range. If the
OFFSET value is outside this range, check Section 7.1 or 9.1
for procedures to correct the OFFSET value to near zero.
The Instrument will now display the NO OFFSET value. It
should also have a value near zero (0.0 50).
Step 4 - The M9110AH is now ready to measure sample gas.
2-9
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
Figure 2-4: Front Panel
2-9
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
2-11
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
Figure 2-5: Assembly Layout
2-12
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
Power Voltage/Frequency
Rack Mount, w/ Slides
Rack Mount, w/ Ears Only
Rack Mount, External Pump w/o Slides
Stainless Zero/Span Valves
4-20 mA Current Loop Output, Isolated
Bypass flow 500 cc/min
Molybdenum Converter
Desiccant Canister - O3 generator
PROM # Serial #
Date Technician
2-14
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
INTENTIONALLY BLANK
2-15
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
3 SPECIFICATIONS, WARRANTY
3.1 Specifications
Operating Modes
RangesIn 1 ppm increments from 5 ppm to 5,000 ppm
Noise at zero0.02 ppm RMS
Noise at span<0.2% of reading RMS above 20 ppm
Detection Limit(Note 1)0.04 ppm RMS
Zero Drift (Note 2)<0.2% full scale/24 hours
Zero Drift (Note 2)<0.4% full scale/7 days
Span Drift (Note 2)<1% FS/24 hours
Lag Time
Switching Mode20 sec (Note 3)
NO
mode
x
Response Time
Switching Mode95% in < 40 sec (Note 3)
NO
mode
x
Sample Flow Rate290 10 cc/min (Including bypass)
Linearity1% of full scale
Precision0.5% of reading
Temperature Range5-40o C
Humidity0-95% RH non-condensing
Temp Coefficient< 0.1% per o C
Voltage Coefficient< 0.1% per V
Dimensions HxWxD7"x17"x23.6" (18 cm x 43 cm x 61 cm)
Weight, Analyzer43 lbs (20 kg)
Weight, Pump Pack16 lbs (7 kg)
Power, Analyzer100 V~ 50/60 Hz, 120 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 200 watts
Power, Analyzer
Power, Ext Pump110 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 295 watts
Power, Ext Pump
EnvironmentalInstallation Category (Over-voltage Category) IIPollution Degree 2
Analog Resolution 1 part in 2048 of selected voltage or current range
Recorder Output 0-100 mV, 0-1, 5, 10v, bipolar
Current Loop Option 4-20ma isolated
Status 12 Status Outputs from opto-isolator
Measurement Units ppm, mg/m3
4
230 V~ 50 Hz, 2.5A
4
230 V~ 50 Hz, 2.5A
NO/NOx switching mode, NO only mode, NOx only mode
Single range, independent ranges or autoranging
4 sec (Note 3)
95% in < 10 sec (Note 3)
1. Defined as twice the zero noise level.
2. At constant temperature and voltage.
3. Lag & response times longer for external converter option.
4. Electrical ratings for CE Mark compliance.
3-1
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F
3.2 Warranty
WARRANTY POLICY (02024c)
Prior to shipment, TELEDYNE equipment is thoroughly inspected and tested. Should
equipment failure occur, TELEDYNE assures its customers that prompt service and support
will be available.
COVERAGE
After the warranty period and throughout the equipment lifetime, TELEDYNE stands ready to
provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in
the industry. All maintenance and the first level of field troubleshooting is to be performed by
the customer.
NON-TELEDYNE MANUFACTURED EQUIPMENT
Equipment provided but not manufactured by TELEDYNE is warranted and will be repaired to
the extent and according to the current terms and conditions of the respective equipment
manufacturers warranty.
GENERAL
TELEDYNE warrants each Product manufactured by TELEDYNE to be free from defects in
material and workmanship under normal use and service for a period of one year from the date
of delivery. All replacement parts and repairs are warranted for 90 days after the purchase.
If a Product fails to conform to its specifications within the warranty period, TELEDYNE shall
correct such defect by, in TELEDYNE's discretion, repairing or replacing such defective
Product or refunding the purchase price of such Product.
The warranties set forth in this section shall be of no force or effect with respect to any
Product: (i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has
been used in any manner other than in accordance with the instruction provided by
TELEDYNE or (iii) not properly maintained.
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES
THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES
OF MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER
WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE
REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES
FOR BREACH OF ANY WARRANTY CONTAINED HEREIN. TELEDYNE SHALL
NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES
ARISING OUT OF OR RELATED TO THIS AGREEMENT OF TELEDYNE'S
PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR
OTHERWISE.
3-2
Loading...
+ 151 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.