Tektronix TLA5Fxxx, TLA6Fxx,,TLA7Fxx PowerFlex Field Upgrade Kits Field Install Instructions

Instructions
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlext Field Upgrade Kits for TLA5000, TLA600, and TLA700 Logic Analyzers
071-0591-05
This document applies to TLA System Software­version 3.2 and above.
www.tektronix.com
*P071059105*
071059105
Copyright © Tektronix, Inc. All rights reserved. Licensed software products are owned by Tektronix or its suppliers and are protected by United States copyright laws and international t reaty provisions.
Use, duplication, or disclosure by the Government is subject to restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227-7013, or subparagraphs (c)(1) and (2) of the Commercial Computer Software -- Restricted Rights clause at FAR 52.227-19, as applicable.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price c hange privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
PowerFlex is an unregistered trademark of Tektronix, Inc.
WARRANTY
Tektronix warrants that the media on which this software product is furnished and the encoding of the programs on the media will be free from defects in materials and workmanship for a period of three (3) months from the date of shipment. If a medium or encoding proves defective during the warranty period, Tektronix will provide a replacement in exchange for the defective medium. Except as to the media on which this software product is furnished, this software product is provided “as is” without warranty of any kind, either express or implied. Tektronix does not warrant that the functions contained in this software product will meet Customer’s requirements or that the operation of the programs will be uninterrupted or error-free.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period. If Tektronix is unable to provide a replacement that is free from defects in materials and workmanship within a reasonable time thereafter, Customer may terminate the license for this software product and return this software product and any associated materials for credit or refund.
THIS W ARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPLACE DEFECTIVE MEDIA OR REFUND CUSTOMER’S PAYMENT IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read the General Safety Summary in other system manuals for warnings and cautions related to operating the system.
ToAvoidFireor
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and marking on the product. Consult the product manual for further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do Not Operate Without Covers. Do not operate this product with covers or panels removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
1
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
WARNING
High Voltage
Protective Ground
(Earth) Terminal
CAUTION
Refer to Manual
2
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before servicing the instru-
ment.
Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then disconnect the power cord from the mains power.
Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
To avoid electric shock, do not touch exposed connections.
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
3
Contacting Tektronix
Phone 1-800-833-9200*
Address Tektronix, Inc.
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
Web site www.tektronix.com
Sales support 1-800-833-9200, select option 1*
Service support 1-800-833-9200, select option 2*
Technical support Email: techsupport@tektronix.com
1-800-833-9200, select option 3*
6:00 a.m. -- 5:00 p.m. Pacific time
* This phone number is toll free in North America. After office hours, please leave a
voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices.
4
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
PowerFlex Upgrade Instructions
These instructions describe how to upgrade your Tektronix logic analyzer or pattern generator module with the upgrade kit that you purchased from Tektro­nix. The instructions also provide steps to attach the labels to your Tektronix instrument to indicate that your upgrade kit has been installed.
Instruments
TLA600 Logic Analyzers All models supported
TLA700 Logic Analyzer Modules All models supported
TLA7PG2 Pattern Generators Modules All models supported
TLA5000 Logic Analyzers All models supported
Minimum Tool and Equipment List
No special tools are required to install the upgrades. However, your logic analyzer mainframe must have the PowerFlex Utility installed and should contain V3.2 of the TLA application software or higher.
Kit Contents
Your PowerFlex upgrade kit contains the following:
H Tektronix Upgrade Key Certificate
H Lexan configuration labels (not required for TLA600 series logic analyzers)
H Upgrade kit label
H Envelope for labels and certificate
H These instructions
NOTE. The upgrade kit will work on one instrument only! Each upgrade kit is made for a specific TLA700 series logic analyzer module, pattern generator module, TLA600 or TLA5000 series logic analyzer. Make sure that you have the correct upgrade kit for your instrument.
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
5
PowerFlex Upgrade Instructions
Install the Upgrade on Your Instrument
Complete the following steps to install the upgrade on your instrument. These instructions assume that your instrument is installed in the same mainframe where the PowerFlex Utility is installed.
1. If you have not already done so, power on your logic analyzer.
2. Exit the TLA and Pattern Generator (if installed) applications.
NOTE. You must exit the TLA and pattern generator applications before you start the PowerFlex Utility. You cannot run the PowerFlex Utility and the TLA or pattern generator application at the same time.
3. Start the PowerFlex Utility. Select Start → Program Files Tektronix
Logic Analyzer TLA PowerFlex Utility.
4. In the PowerFlex window (see Figure 1), select the instrument using the
Frame and Slot Number drop-down lists. Verify that the current configura­tion for your instrument is correct.
Figure 1: PowerFlex Utility
6
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
PowerFlex Upgrade Instructions
5. Click the Install Upgrade button in the PowerFlex window. The Install
Upgrade dialog box appears (see Figure 2).
Figure 2: Install Upgrade dialog box
6. Enter the upgrade key recorded on the Tektronix Upgrade Key Certificate;
hyphens are optional.
7. Click the Install button. The PowerFlex utility will install the upgrade and
display a message indicating that the upgrade was successful.
8. Click OK to dismiss the message. The updated configuration information
will appear in the PowerFlex window.
9. Exit the PowerFlex Utility.
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
7
PowerFlex Upgrade Instructions
Attaching Labels on the TLA600 or TLA5000 Logic Analyzers
After you install the upgrade kit on your TLA600 or TLA5000 logic analyzer, you must install the upgrade label on the rear of the instrument to indicate that the upgrade kit is installed.
1. Locate the big label on the rear of the instrument; it should look similar to
the label in Figure 3.
Kit label
Figure 3: TLA600 series logic analyzer label (located on the rear of the instrument). TLA5000 labels look similar.
2. Locate PowerFlex option label from the label sheet that you received with
this kit. Peel the label from the sheet and affix it to the location indicated by Kit labelin Figure 3.
Attaching Labels on the TLA700 Series Logic Analyzer Modules and Pattern Generator Modules
After you install the upgrade kit on your TLA700 series logic analyzer module or pattern generator module, you must install labels on the instrument indicating that the upgrade kit is installed. There are two types of labels that you must install on your instrument.
H Lexan label. This label is located on the front of the module and lists the
current configuration (see Figure 5 on page 10 or Figure 7 on page 12).
H PowerFlex kit label. This label is located on the side of the module; it lists
the current PowerFlex option and the upgrade key.
8
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
PowerFlex Upgrade Instructions
Service and Upgrade
Notes
TLA700 Logic Analyzer
Modules
To prevent personal injury or damage to the instrument, adhere to the following requirements before servicing the TLA700 series instruments:
H Read the General Safety Summary and the Service Safety Summary found at
the beginning of this document.
H Only qualified service personnel should perform the procedures in the
remaining sections of this document.
H Observe all warnings, notes, and cautions in the following sections.
Use the following instructions to install the labels on the TLA700 series logic analyzer modules:
CAUTION. Do not remove the logic analyzer module while the mainframe is powered on. Failure to do so, can damage the module or the mainframe. Always power down the mainframe and disconnect the power cord before removing any installed modules.
1. Power down the mainframe and disconnect the power cord.
2. Loosen the four retaining screws on the corners of the front panel that secure
the module to the mainframe.
3. Remove the module from the mainframe.
4. Locate the big label on the side of the logic analyzer module (see Figure 4).
Kit label
Figure 4: TLA700 series logic analyzer module label
5. Note the location of the long label (indicated by “Kit label” in Figure 4).
Depending on your instrument, this could be a label inside a rectangle or a string of text.
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
9
PowerFlex Upgrade Instructions
6. Locate PowerFlex option label from the label sheet that you received with
this kit. Peel the label from the sheet and affix it to the location indicated by Kit labelin Figure 4.
7. Locate the small lexan configuration label on the front panel of the module.
Figure 5 shows a TLA7Axx series logic analyzer module. The configuration label location is similar for other logic analyzer modules.
E2
10
Configuration label
Figure 5: Configuration label location on TLA logic analyzer modules
NOTE. Remove the old configuration label before installing the new label.
8. Carefully remove the old configuration label.
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
PowerFlex Upgrade Instructions
9. Locate the new configuration label from the sheet of labels that you received
with this kit. If necessary, refer to the big label to determine the correct label.
10. Affix the new configuration label in the hole from where you just removed
the older label.
11. Reinstall the module in the mainframe.
12. Use a screwdriver to tighten the four retaining screws to 2.5 in-lbs after
seating the module in place.
13. Reconnect the power cord to the mainframe.
TLA7PG2 Pattern
Generator Modules
Use the following instructions to install the labels on the TLA7PG2 Pattern Generator modules:
CAUTION. Do not remove the pattern generator module while the mainframe is powered on. Failure to do so, can damage the module or the mainframe. Always power down the mainframe and disconnect the power cord before removing any installed modules.
1. Power down the mainframe and disconnect the power cord.
2. Loosen the four retaining screws on the corners of the front panel that secure
the module to the mainframe.
3. Remove the module from the mainframe.
4. Locate the label on the side of the instrument; it looks similar to the label in
Figure 6.
5. Note the location of the option information on the center of the label.
6. Locate PowerFlex option label from the label sheet that you received with
this kit. Peel the label from the sheet and affix it to the location indicated by Kit labelin Figure 6.
7. Locate the new configuration label that you received with this kit.
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
11
PowerFlex Upgrade Instructions
Kit label
Figure 6: TLA7PG2 Pattern Generator module label
8. Affix the new configuration label over the old label near the bottom of the
front panel (see Figure 7).
9. Reinstall the module in the mainframe.
10. Use a screwdriver to tighten the four retaining screws to 2.5 in-lbs after
seating the module in place.
11. Reconnect the power cord to the mainframe.
Configuration label
12
Figure 7: Configuration label location on pattern generator modules
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
Troubleshooting
PowerFlex Upgrade Instructions
Refer to this section if you have problems installing the upgrade. Below is a list of possible problems or error messages that you may encounter and recom­mended solutions.
If these suggestions do not address your problem, contact Tektronix for help. Refer either to Contacting Tektronix on page NO TAG of this document or to Contacting Tektronix in the PowerFlex Utility online help.
Table 1: PowerFlex Utility problems and solutions
Problem or error message Possible solution
PowerFlex Utility will not start The TLA or pattern generator applications are running on the
mainframe. Exit these applications before trying to run the PowerFlex Utility.
Error: Invalid Upgrade Key. Upgrade failed. Possible typing error.
Error: Invalid Upgrade Key. Upgrade Failed. The Upgrade Key is made for the module with the following unique ID:
Error: The current module configuration is incompatible with the upgrade:
Error: The upgrade is already installed
TLA or pattern generator application will not start after installing the upgrade.
Try re-entering the upgrade key again. Make sure that you have the correct upgrade key for the module or instrument.
You entered an upgrade key for a different module. Make sure that you have the correct module selected in the PowerFlex Utility.
The upgrade key is correct, but you cannot upgrade the instrument from the current configuration. Verify that you ordered the correct upgrade kit for your instrument.
You attempted to install the upgrade twice. You can only install the upgrade once.
The PowerFlex Utility is still running. Exit the PowerFlex utility before starting the TLA or pattern generator application.
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits
13
PowerFlex Upgrade Instructions
14
TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Upgrade Kits