This document applies to TLA System Software Version
5.1SP1 and above
Warning
These servicing instructions are for use by qualified personnel
only. To avoid personal injury, do not perform any servicing
unless you are qualified to do so. Refer to all safety summaries
prior to performing service.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
MagniVu and iView are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
P.O . Bo x 50 0
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
World wi de, v i sit www.tektronix.com to find contacts in your area.
Table of Contents
General Safety Summary .........................................................................................iv
Service Safety Summary.......................................... ................................ ................vi
TLA5200 Series Product Specifications & Performance Verificationiii
General Safety Summary
General Safet
To Avoid Fire or Person al
Injury
ySummary
Review the fo
this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system.
Read the safety sections of the other component manuals for warnings and
cautions related to operating the system.
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to eart
terminals of the product, ensure that the product is properly grounded.
llowing safety precautions to avoid injury and prevent damage to
h ground. Before making connections to the input or output
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Power Disconnect. The power cord disconnects the product from the power source.
Donotblockthepowercord;itmustremain accessible to the user at all times.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Do Not Operate With Suspected Failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Use Proper Fuse. Useonlythefusetypeandratingspecified for this product.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
ivTLA5200 Series Product Specifications & Performance Verification
General Safety Summary
Terms in this Manual
Symbols and Terms on the
Product
These terms may
WAR N ING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER in
the marking.
WAR NI NG
read the marking.
CAUTIO
The following symbol(s) may appear on the product:
appear in this manual:
dicates an injury hazard immediately accessible as you read
indicates an injury hazard not immediately accessible as you
N indicates a hazard to property including the product.
TLA5200 Series Product Specifications & Performance Verificationv
Service Safety Summary
Service Safet
y Summary
Only qualifieSafety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
d personnel should perform service procedures. Read this Service
this product. Disconnect power, remove battery (if applicable), and
viTLA5200 Series Product Specifications & Performance Verification
Preface
Related Documentation
This document lists the characteristics and specifications of the TLA5200 and
TLA5200B series logic analyzers. It also includes the performance verification
procedures.
otherwise noted. Microprocessor-related products and individual logic analyzer
probes have their own documentation for characteristics and specifications.
To prevent personal injury or damage consider the following requirements before
attempting service:
The procedures in this manual should be performed only by qualified service
personnel.
Read the General Safety Summary and Service Safety Summary found at
the beginning of this manual.
Be sure to follow all warnings, cautions, and notes in this manual.
The following table lists related documentation available for your logic analyzer.
The documentation is available on the TLA Documentation CD included with
your logic analyzer, and on the Tektronix Web site (www.Tektronix.com).
All references to TLA5200 also apply to TLA5200B, unless
To obtain documentation not specified in the table, contact your local Tektronix
representative.
Table i: Related Documentation
ItemPurposeLocation
TLA Quick Start User Manual
Online Help
Installation Quick Reference CardsBasic installation information
Installation Manuals
YZs of Logic A nalyzers
X
TLA Product Speci ficationsSpecifications for other TLA products
Basic operational overview
In depth operation and UI help
Detailed first-time installation information
Introduction to logic analyzer basics
TLA5200 Series Product Specifications & Performance Verificationvii
Preface
Table i: Related Documentation (cont.)
ItemPurposeLocation
TPI.NET Documentation
Detailed information for controlling the
logic analyzer u sing .NET
Field upgrade kits
Optional Service ManualsSelf-service documentation for modules
TLA Application Software Release N otesSoftware description, compatibility,
Upgrade information for your logic
analyzer p
and mainframes
impact of
installation, upgrade, and operational
notes, and known issues.
roduct
changes, contact information,
Go to Start→All Programs→Tektronix
logic Analyzer→TLA Release Notes
viiiTLA5200 Series Product Specifications & Performance Verification
Specifications
The following tables list the specifications for the TLA5200B and TLA5200 series
logic analyzers. All references to TLA5200 also apply to TLA5200B, unless
otherwise no
ted. All specifications are guaranteed unless noted Typ i ca l . Typical
characteristics describe typical or average performance and provide useful
reference information.
Specifications that are marked with the
indirectly) in the Performance Verification chapter of this document.
The performance limits in this specification are valid with these conditions:
For op
documentation for any external oscilloscopes used with your Tektronix logic
analyzer to determine the warm-up period and signal-path compensation
requirements.
Table 1: Atmospheric characteristics
racteristic
Cha
Temperature
Relative humidity
Altitude
cription
Des
rating (no media in CD or DVD drive)
Ope
°C to +50 °C, 15 °C/hr m aximum gradient, noncondensing (derated 1 °C per 305 m (1000 ft)
+5
above 1524 m (5000 ft) altitude)
Nonoperating (no media in drive)
-20 °C to +60 °C, 15 °C/hr maximum gradient, noncondensing
Operating (no media in drive)
20% to 80% relative humidity, noncondensing. Maximum wet bulb temperature: +29 °C (derates
relative humidity to approximately 22% at +50 °C).
Nonoperating (no media in drive)
8% to 80% relative humidity, noncondensing. Maximum wet bulb temperature: +29 °C (derates relative
humidity to approximately 22% at +50 °C).
Operating
To 3000 m (9843 ft), (derated 1 °C per 305 m (1000 ft) above 1524 m (5000 ft) altitude.
Nonoperating
12,190 m (40,000 ft )
symbol are checked directly (or
The instr
ument must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these
specifications.
The instrument must have had a warm-up period of at least 30 minutes.
The ins
trument must have been calibrated/adjusted at an ambient temperature
between +20 °C and +30 °C.
timum performance using an external oscilloscope, please consult the
TLA5200 Series Product Specifications & Performance Verification1
Specifications
Table 2: TLA520
0 input parameters with probes
CharacteristicDescription
Threshold Accuracy
Threshold range and step size
±100 mV
Settable from +4.5 V to -2 V in 5 mV steps
Threshold channel selection16 threshold groups assigned to channels. P6410, P6417, P6418, and
P6419 probes have two threshold settings, one for the clock/qualifier
channel and one for the data channels. P6434 probes have four threshold
settings, one for each of the clock/qualifier channels and two for the data
channels (one per 16 data channels).
Channel-to-channel skew
Channel-to-channel skew (Typical)
Sample uncertainty
≤ ±150 ps maximum
≤ ±75 ps
Asynchronous
Sample period
Synchronous
125 ps
Input voltage range–2.5 to +5 V
Probe input resistance (Typical)20 kΩ
Probe in
(Typical)
put capacitance
P6410, P
P64181.4 pF data channels
6417, P6434
2pF
2pFCLK
/Qual channels
P6419< 0.7 pF
Minimum slew rate (Typical)0.2 V/ns
Maximum operating signal6.0 V
Probe
overdrive
, P6417, P6418,
P6410
P6419
P6434
p-p
±250 mV or ±25% of signal swing minimum required beyond threshold,
whichever is greater
±300 mV or ±25% of signal swing minimum required beyond threshold,
chever is greater
whi
±4 V maximum beyond threshold
imum nondestructive input signal to probe
Max
nimum input pulse width signal (single channel)
Mi
(Typical)
Delay time from probe tip to module input probe
V
±15
5 ns (P6434)
1.
1.25 ns (P6410, P6417, P6418, P6419)
33 ns ±100ps
7.
connector (Typical)
Table 3: TLA5200 timing latencies
CharacteristicDescription
System Trigger and
External Signal Input
Latencies
1
(Typical)
External System Trigger Input to LA Probe Tip
2
External Signal Input to LA Probe Tip via Signal 3, 4–656 ns + Clk
External Signal Input to LA Probe Tip via Signal 1, 2
4
–656 ns
–656 ns + Clk
3
3
2TLA5200 Series Product Specifications & Performance Verification