Contacting Tektronix
Tektronix, Inc.
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to find contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of
shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective
product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and
replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced
parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period
and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the
defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the
product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall
be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and
care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from a ttempts by personnel
other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or
connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or
d) to service a product that has been modified or integrated with other products when the effect of such modification or integration
increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES,
EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS
IS THE SOLE AND E XCLU S IVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX
AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES
IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
DAMAGES.
le Mode Transmitter Tests .................................................................................................24
mon Mode Operating DC Voltage Test (Receiver) .......................................................................38
Table of Content
s
TekExpress HEAC Quick Start User Manuali
Table of Content
s
iiTekExpress HEAC Quick Start User Manual
General Safety S
ummary
General Safet
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other
component manuals for warnings and cautions related to operating the system.
To Avoid Fire or Personal Injury
Use proper power cord. Use only the power cord specified for this product and certified for the country of use.
Use proper voltage setting. Before applying power, ensure that the line selector is in the proper position for the
source bei
Connect a
to a voltage source.
Ground the product. This product is indirectly grounded through the gr ounding conductor of the mainframe power cord.
To avoid electric shock, the grounding conductor must be c onnected to earth ground. Before making connections to the input
or outpu
Observe
product manual for further ratings information before making connections to the product.
ng used.
nd disconnect properly.
t terminals of the product, ensure that the product is properly grounded.
all terminal ratings.
y Summary
Do not connect or disconnect probes or test leads while they are connected
To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Power disconnect. The power cord disconnects the product from the power source. Do not block the power cord; it
must remain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by
fied service personnel.
quali
exposed circuitry.
Avoid
roper AC adapter.
Use p
Do not touch exposed connections and components when power is present.
Use only the AC adapter specified for this product.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
TekExpress HEAC Quick Start User Manualiii
General Safety S
Terms in This Manual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and Terms on the Product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ummary
ivTekExpress HEAC Quick Start User Manual
Compliance Info
rmation
Compliance In
This section lists the environmental standards with which the instrument complies.
formation
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life Handling
Observe the following guidelines when recycling an instrument or component:
Equipment
equipment may contain substances that could be harmful to the environment or human health if improperly handled at the
product’s end of life. In order to avoid release of such substances into the environment and to reduce the use of natural
resource
reused or recycled appropriately.
Recycling.
s, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are
This symb
to Directives 2002/96/EC and 2006/66/EC on waste electrical and electronic equipment (WEEE) and
batteries. For information about recycling options, check the Support/Service section of the Tektronix Web
site (ww
Production of this equipment required the extraction and use of natural resources. The
ol indicates that this product complies with the applicable European Union requirements according
w.tektronix.com).
Restriction of Hazardous Substances
This product has been classified as Monitoring and Control equipment, and is outside the scope of the 2002/95/EC RoHS
Directive.
TekExpress HEAC Quick Start User Manualv
Preface
Preface
TekExpress is the Tektronix Compliance Test Automation Framework, developed to support current and future test
automation needs of customers. Developed using National Instruments’ TestStand, TekExpress leverages on the capabilities
of the Microsoft .NE T framework. TekExpress HEAC provides a test and compliance solution that address the needs
of the of HDMI 1.4 and CTS 1.4 standards.
Key Features
The application provides the following features:
Reliable testing
Conformance to HDMI 1.4 (HEAC) standards and test specifications CTS 1.4 ensures reliable results
Accurate Transmitter (Tx) and Receiver (Rx) testing and violation testing ensures reliable tests
Sophisticated measurement techniques eliminate errors in jitter measurements
Closed loop measurements eliminate non-linearities of test setup enables precise testing
Significantly reduced test times for Sink and Audio tests
Optimal setups ensure accurate results
Fast testing
Remote control of signal sources automates Source and Sink tests and reduces test times
One-button selection of multiple tests ensures faster testing
Automatic mask fit, measurements, and Pass/Fail notification delivers quick results
Test margins and statistical information aid analysis to help find answers quickly
One-button CSV-format summary and MHT report saves time
Complete testing
Wide range of tests for Source and Sink devices enables complete validation
Complete compliance solution with a test fixture and signal sources
viTekExpress HEAC Quick Start User Manual
Documentation
This manual describes the basic installation and operation of the TekExpress application. For more detailed information, see
the online help. The following information is available for this product:
Table i: Related documentation
Preface
TekExpress HEAC Online Help, English
TekExpress
TekExpress HEAC Quick Start User Manual,
English (PDF)
HEAC User Manual, English (PDF)
Software Upgrades
Periodic software upgrades may become available. To check for upgrades:
Go to the Tektronix Web site (www.tektronix.com/software).
Enter the product name (TekExpress HEAC).
076-0191-xx
077-0388-xx
077-0389-x
x
TekExpress HEAC Quick Start User Manualvii
Preface
viiiTekExpress HEAC Quick Start User Manual
Install the Soft
ware
Install the So
The TekExpress application must be installe d on a computer or on a Tektronix oscilloscope.
ftware
Before Installation
Read the Readme.txt file provided with the product software.
Prerequisites
The product software installer automatically installs all the necessary software and components needed to run the
TekExpress HEAC solution. The following are the prerequisites:
5. Double-click the executables to extract the files. The audio patterns are extracted to
ll applications.
ad
HEACWebinstaller_<version>.EXE.
-click the executable to extract the files. After extraction, the HEAC installer launches and the software is installed
HEAC_AudioRX_Patterns_AWG5K.EXE and HEAC_AudioRX_Pat-
.
C:\HEAC Pattern s.
TekExpress HEAC Quick Start User Manual1
Install the Soft
ware
Required Test Equipment
Each test suite requires specific equipment setups. Click Show Schematic within the select pane to see how to connect the
device under test and test equipment to your oscilloscope.
The following Tektronix test fixtures are required, depending on the tests that you will be running:
TF-HEAC-TPA-KIT consisting of:
TF-HEAC-TPA-MAIN
TF-HEAC-TPA-AP
TF-HEAC-TPA-CP
TF-HEAC-TDR-AR (2 numbers)
TF-HEAC-TDR-CR (2 numbers)
TF-HDMI-TPA-CE (EDID board) consisting of:
EDID fixture PCBA
EEPROM with HDMI 1.4 software
Ribbon cable
TF-HDMID-TPA-KIT (required for HEAC testing of Type D devices) consisting of:
Plug assembly
Receptacle cable assembly (2 numbers)
The following arbitrary waveform generators are required for running common mode receiver tests:
Tektronix AWG5000 Series or AWG5000B Series
Tektronix AWG7000 Series or AWG7000B Series
Start the Software
To start the software, do one of the following:
From the Start menu, select Start > Programs > Tektronix > TekExpress > TekExpress HEAC.
Double-clickon the desktop.
Close the Software
lose the software, do one of the following:
To c
ect File > Exit.
Sel
Click.
2TekExpress HEAC Quick Start User Manual
Use the TekExpre
ss Software
Use the TekExp
Process Flow
Follow steps 1 through 11 to select and run a test to check for compliance. Details for each of these actions are given on
the following pages.
1. Set up the switches on the HEAC-TPA-MAIN fixture.
2. Set up the jumpers on the HEAC-TPA-MAIN fixture.
3. Select a device.
4. Select the test suite and choose the version.
5. Select the test.
6. Configure the test parameters.
7. View and select connected instruments.
8. Run the test.
9. View the progress of acquisition.
10. View the progress of analysis.
11. View and print the generated report.
ress Software
TekExpress HEAC Quick Start User Manual3
Use the TekExpre
ss Software
Set Up the Switches on the HEAC-TPA-MAIN Fixture
The following diagram shows the fixture configuration switches on the HEAC-TPA-MAIN fixture.
The following table shows the settings to configure on the HEAC-TPA-MAIN fixture.
Table 1: HEAC-TPA-MAIN switch setup
+5 V
Test IDDescriptionDUT
Source
Sink
Source
Sink
Sink
Sink
Source
Sink
Source
5-2to5-5
5-6 to 5-10
5-11 to 5-15
5-16
5-17
Differential Mode
Operating DC
Differential Mode
Transmitter
Common Mode
Transmitter
Single Mode
Transmitter
Differential Mode
Receiver
Common Mode
Receiver
fromS1S2S3S4S7
J6, J8RightRightRight
J8
J6RightRightRightRightRight
J8
J8
J8
J6RightRightRightRightRight
J8
J6RightRightRightRightRight
LeftLeftLeft
LeftLeftLeft
LeftLeftLeft
LeftLeftLeft
LeftLeftLeft
Left
Right
Right
Right
Right
Right
Right5-1
Left
Left
Left
Left
Left
4TekExpress HEAC Quick Start User Manual
Table 1: HEAC-TPA-MAIN switch setup (cont.)
+5 V
Test IDDescriptionDUT
5-18
5-19
5-20
Single Mode
Receiver
Common Mode
Operating DC
Single Mode
Operating DC
Source
Source
Source
fromS1S2S3S4S7
J6, J8RightRightRight
J6, J8RightRightRight
J6, J8RightRightRight
Set Up the Jumpers on the HEAC-TPA-MAIN Fixture
owing table shows the jumper settings to configure on the HEAC-TPA-MAIN fixture.
The foll
Table 2: HEAC-TPA-MAIN jumper setup
Use the TekExpre
Left
Left
Left
ss Software
Right
Right
Right
Test IDDescription
5-1to5-5
5-6 to 5-10
to 5-15
5-11
5-16
5-17 to 5-19
5-18 to 5-20
Differential Mode
mitter
Trans
n Mode
Commo
Transmitter
Single Mode
Transmitter
Differential Mode
eiver
Rec
Common Mode
eceiver
R
ingle Mode
S
Receiver
Termination
50 Ω
50 Ω
55 Ω
-
-
-
5 Ω
ource
ONOFFONOFF
ONOFFONOFF
OFFOFFOFFOFF
FF
O
FF
O
FF
O
FF
O
Signal S
output
impedanceJP1JP2JP3JP4
-
-
-
110 ΩOFFOFFOFFOFF
100 ΩONOFFONOFF
90 ΩONONONON
50 ΩONOFFONOFF
5
TekExpress HEAC Quick Start User Manual5
Use the TekExpre
ss Software
Select a Device, Test Suite, and Test
1. Click the Sele
2. Select a Devic
3. Select the Te
4. Choose the vesuite from the Version drop-down list.
5. Select the test to run. You can select
multiple tests from the list.
6. Click Select All to select all the tests in
the list.
7. Click Dese
8. Click Sho
schematic diagram of the test setup.
ct tab.
e type.
st Suite.
rsion of the selected test
lect All to disable the tests.
w Schematic to see the
Configure the Test Parameters
1. Click th
e test name to configure.
2. Click C
3. If multiple instruments are connected,
4. Verify schematic-related parameters
5. The Compliance Mode is selected by
onfigure.
choose the correct instrument from the
drop-down list in the parameter-value list.
such as DUT HEAC Signal connected to
Real Time Scope at and others. You can
also change other parameters such as
Minimum signal threshold.
default. You can generally use the
default settings in Compliance Mode.
6TekExpress HEAC Quick Start User Manual
6. Click the Acquire tab to view parameters
related to acquisition.
7. Click the Analyze tab to view the test
analysis parameters.
Use the TekExpre
ss Software
8. Click th
9. Click t
10. Click Apply.
11. Click Close.
NOTE
session.
e Limits tab to view the
measurement limits.
See the online help for various test limit
sons.
compari
he Comments tab to enter
test-specific comments.
. Use the File menu to save the
TekExpress HEAC Quick Start User Manual7
Use the TekExpre
ss Software
Connect and Configure the Equipment
The equipment connection is given in detail in the Application Examples section of this document. (See page 10, Application
Examples.)
You can cli ck Show Schematic to display a diagram of the test setup for each of the test suites.
Run the Test
1. EntertheIDoftheDevicetobetestedin the DUT ID field.
2.Click Run to run the test.
3. Click the Acquire tab to view the
waveform acquisition details for the tests
in the currently selected test suite.
4. Click the Analyze tab when the tests are
running to view the status and progress
of the tests. The application checks to
see if the device has passed the test.
The Status Messages window time
stamps all runtime messages and
displays them.
8TekExpress HEAC Quick Start User Manual
Interpret the Test Result
When the test completes, the report summary appears. Check the report to see if the device passed the test.
1. The report displays the DUT ID, the date
and time the test was run, and execution
time.
2. Test information like the Device Type,
Version, Execution Time, Compliance
Mode (Yes or No), and Overall Test
Result (Pass or Fail), MAC Address (in
case of Receiver tests) are shown.
3. Instrument information like the models,
serial numbers, and firmware/software
versions are shown.
4. A table displays the Test Name,
Measurement Details, Limit1 and Limit2,
CRC Status, CRC Details, Test Result,
Compliance Mode, Analysis Time, and
Comments.
Use the TekExpre
ss Software
Generate, Save, and Print a Report
ou have successfully run one or more test(s), a report summary is generated in MHT format.
After y
To save
To prev
the report summary in the desired location, select File > Save Report As.
iew the report before printing, select File > Print Preview Report. Select File > Print Report.
TekExpress HEAC Quick Start User Manual9
Application Exa
mples
Application Examples
Differential Mode Transmitter Tests
Required equipmentModel supported
Real-time os
Probe
Test fixtureTF-HEAC-TPA-P (A/C) or TF-HEAC-TPA-KIT,
50 Ω SMA T
cilloscope
erminators
Tektronix DPO7000 Series, DPO70000 Series, DPO70000B
Series, DSA 70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix P
DPO70000 Series, DPO70000B Series, DSA70000 Series,
DPO70000B, and MSO70000 Series Digital and Mixed
Signal Osc
Tektronix P6330, TDP1000, TDP 1500,, or TDP3500 for a
Tektronix DPO7000 Series Digital Oscilloscope
TF-HDMID
Any term
ies Digital and Mixed Signal Oscilloscopes
6247, P6248, P6330, or P7330 for a
illoscope
-TPA-KIT
inator that meets the following requirements:
50 Ω impe
Connect
DUT
SMA Cab
HDMI CableHDMI cable with HEAC capability
This tests the DUT for compliance with differential mode transmitter tests.
les
An HEAC
nix part number 174-1428-00 (1.5 meter)
Tektr o
Tektronix part number 174-1341-00 (1 meter)
dance ± 1% or better
s directly to the SMA female connector
sink or source device
10TekExpress HEAC Quick Start User Manual
Application Exa
1. Connect the DUT to the equipment as shown.
NOTE. This equipment setup diagram and connections are the same for all differential transmitter tests except the
Operating DC V
oltage test.
mples
Connect the CEC or CDC Controller to the CEC/CDC Control Port on the HEAC-TPA main board.
Connect the HEAC-TPA plug board to the HEAC connector on the Source/S ink DUT.
Connect the 50 Ω SMA terminators to the HEAC-TPA main board.
Connect a differential probe to the HEAC± differential signal probe point on the HEAC-TPA main board.
If you are testing a Sink DUT, set the HEAC-TPA main board to enable the Sink DUT test. Set switch S7 position to the
left (Off position) on the HEAC-TPA adapter.
If you are testing a Source DUT, set the H EAC-TPA main board to enable the Source DUT test. Activate +4 V bias using
the +5 V power from the Source DUT. Set switch S7 to the right (On position) on the HEAC-TPA main fixture.
Activate the HEC transmission on the HEAC Source/Sink DUT using CEC/CDC controller.
TekExpress HEAC Quick Start User Manual11
Application Exa
2. Select HEAC-Transmitter as the device
mples
type.
By default, th
as the Test Suite.
The default version CTS 1.4.
All the diffe
except the Template test, are selected
by default.
e Differential-Tx is selected
rential mode transmitter tests,
NOTE. The Te
test and not a part of the CTS 1.4.
3. Enter the D
ID field.
4. Click Show Schematic to display a
diagram of the test setup and confirm
that your
5. Click Co
Configuration panel.
6. Use the default values in the Compliance
Mode to run the tests for compliance with
the st
NOTE. To change acquisition parameter
values, click User Defined Mode.A
ge appears prompting you that if you
messa
make changes to a test, the test may no
longer be compliant. Click OK to proceed
the User Defined Mode.
with
7. Click Close.
mplate test is an informative
UT identification in the DUT
test setup is correct.
nfigure to open the
andards.
8. Click Run in the Select panel.
12TekExpress HEAC Quick Start User Manual
9. Click the Analyze tab to view test
analysis results.
10. The application displays the status of the
tests in the St
atus Messages window.
Click Save Status to save the status
messages.
11. When the tests complete, the application
shows a report in the Report panel.
Application Exa
mples
TekExpress HEAC Quick Start User Manual13
Application Exa
mples
Differential Mode Operating DC Voltage Test (Transmitter)
Required equipmentModel supported
Real-time osc
Probe
Test fixtureTF-HEAC-TPA-P (A/C) or TF-HEAC-TPA-KIT,
50 Ω SMA Te
illoscope
rminators
Tektronix DPO7000 Series, DPO70000 Series, DPO70000B
Series, DSA 70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix P6
DPO70000 Series, DPO70000B Series, DSA70000 Series,
DPO70000B, and MSO70000 Series Digital and Mixed
Signal Osci
Tektronix P6330, TDP1000, TDP1500, or TDP3500 for a
Tektronix DPO7000 Series Digital Oscilloscope
TF-HDMID-
Any termi
ies Digital and Mixed Signal Oscilloscopes
247, P6248, P6330, or P7330 for a
lloscope
TPA-KIT
nator that meets the following requirements:
50 Ω imped
Connect
DUT
SMA Cab
HDMI CableHDMI cable with HEAC capability
This tests the DUT for compliance with d ifferential mode operating DC voltage (Transmitter) tests.
les
An HEAC s
nix part number 174-1428-00 (1.5 meter)
Tektr o
Tektronix part number 174-1341-00 (1 meter)
ance ± 1% or better
s directly to the SMA female connector
ink or source device
14TekExpress HEAC Quick Start User Manual
1. Connect the equipment as shown.
Application Exa
mples
NOTE. Use two probes for DC voltage measurement. Use the differential mode configuration for other measurements.
Connect the CEC or CDC Controller to the CEC/CDC Control Port of the HEAC-TPA main board.
Connect the HEAC-TPA plug board to the HEAC connector of the Source/Sink DUT.
Connect the 50 Ω SMA terminators to the HEAC-TPA main board.
Connect the first single-ended probe from Ch1 of the digital oscilloscope to the HEAC+ probe point on the HEAC-TPA
main board, and the second single-ended probe from Ch2 to the HEAC- probe point.
If you are testing a Sink DUT, set the HEAC-TPA main board to enable the Sink DUT test. Set switch S7 to the left
(Off position) on the HEAC-TPA main fixture.
If you are testing a Source DUT, set the HEAC-TPA main board to enable the Source DUT test, and use +5 V power
from the DC Power Supply. Set switch S7 to the right (On position) and the switch S4 is set to the left (Off position)
on the HEAC-TPA main fixture.
NOTE. Do not use +5 V power from the Source DUT.
Connect and configure the DC Power Supply to s upply +5 V to the HEAC-TPA m ain board.
Power on the Source/Sink DUT.
Activate the HEC transmission capability for the HEAC Source/Sink DUT using the CEC/CDC controller.
TekExpress HEAC Quick Start User Manual15
Application Exa
2. Select HEAC-Transmitter as the device
mples
type.
By default, Di
the Test Suite.
The default version CTS 1.4.
fferential-Tx is selected as
3. Click Des
Operating DC Voltage test.
4. Enter the DUT identification in the DUT
ID field.
5. Click Show Schematic to display a
diagra
that your test setup is correct.
6. Click Configure to open the
Configuration panel.
7. Use the default values in the Compliance
Mode
the standards.
NOTE. To change acquisition parameter
es, click User Defined Mode.A
valu
message appears prompting you that If you
make changes to a test, the test may no
ger be compliant. Click OK to proceed
lon
with the User Defined Mode.
8. Click Close.
elect All and select only the
m of the test setup and confirm
to run the tests for compliance with
16TekExpress HEAC Quick Start User Manual
9. Click Run in the Select panel.
The application switches to the Acquire
tab.
10. Follow the on-screen prompts and the
schematics to change the setup. Click
OK to continue.
11. Click the Analyze tab to view test
analysis results.
12. The application displays the status of the
tests in the Status Messages window.
Click Save Status to save the status
messages.
Application Exa
mples
13. When the tests complete, the application
shows a report in the Report panel.
TekExpress HEAC Quick Start User Manual17
Application Exa
mples
Common Mode Transmitter Tests
Required equipmentModel supported
Real-time osc
Probe
illoscope
Tektronix DPO7000 Series, DPO70000 Series, DPO70000B
Series, DSA 70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix P6
Series, DPO70000B Series, DSA70000 Series, and
DSA70000B Series, and MSO7000 Series Digital and
Mixed Oscil
Tektronix P6243, P6245, or TAP1500 for Tektronix
DPO7000 Series Digital Oscilloscopes
ies Digital and Mixed Signal Oscilloscopes
243 or P6245 for Tektronix DPO70000
loscopes
NOTE. Yo u c
single-ended mode.
Test fixtureTF-HEAC-TPA-P (A/C) or TF-HEAC-TPA-KIT,
TF-HDMID-TPA-KIT
50 Ω SMA TerminatorsAny terminator that meets the following requirements:
DUT
SMA CablesTektronix part number 174-1428-00 (1.5 meter)
able
HDMI C
An HEAC sink device
onix part number 174-1341-00 (1 meter)
Tektr
HDMI c
an also use two differential probes in
50 Ω impedance ± 1% or better
Connects directly to the SMA female connector
able with HEAC capability
18TekExpress HEAC Quick Start User Manual
This tests the DUT for compliance with common mode (Transmitter) tests.
Application Exa
mples
1. Connect the eq
uipment as shown.
Connect the CEC or CDC Controller to the CEC/CDC Control Port of the HEAC-TPA main board.
Connect the HEAC-TPA plug board to the HEAC connector of the Sink DUT.
Connect the 50 Ω terminators to the HEAC-TPA main /board.
Connect the first single-ended probe from Ch1 of the digital oscilloscope to the HEAC+ probe point on the HEAC-TPA
main board, and the second single-ended probe from Ch2 to the HEAC– probe point.
Set the HEAC-TPA main board to enable the Sink DUT test.
Connect and configure DC Power Supply to drive +5 V to the HEAC-TPA main board.
Power on the Sink DUT.
Activate the ARC (Common mode) transmission on the HEAC Sink DUT using the CEC/CDC controller.
TekExpress HEAC Quick Start User Manual19
Application Exa
2. Select HEAC-Transmitter as the device
3. Select CommonMode-Tx.
4. Enter the DUT identification in the DUT
mples
type.
The default v
All the common mode transmitter tests
are selected by default.
ID field.
ersion CTS 1.4.
5. Click Show
diagram of the test setup and confirm
that your test setup is correct.
6. Click Configure to open the
Configur
7. Use the d
Mode to run the tests for compliance with
the standards.
NOTE. T
values, click User Defined Mode.A
message appears prompting you that If you
hanges to a test, the test may no
make c
longer be compliant. Click OK to proceed
with the User Defined Mode.
k Close.
8. Clic
Schematic to display a
ation panel.
efault v alues in the Compliance
o change acquisition parameter
9. Click Run in the Select panel.
The application switches to the Acquire
tab.
20TekExpress HEAC Quick Start User Manual
10. Click the Analyze tab to view test
analysis results.
11. The application displays the status of the
tests in the St
atus Messages window.
Click Save Status to save the status
messages.
12. Follow the on-screen prompts. For the
Rise-Fall Time test, configure the DUT to
transmit an MLT-3 signal and click OK.
Application Exa
mples
13. When prompted, configure the DUT to
disable the MLT-3 signal and click OK.
TekExpress HEAC Quick Start User Manual21
Application Exa
mples
14. The Transmit Ji
tter MAX test is run
at three frequencies. The application
prompts you to configure the DUT for
each of these f
requencies: 4.096 MHz,
5.688 MHz, and 6.144 MHz. If your
device supports the frequency, configure
the DUT, and c
lick OK.
22TekExpress HEAC Quick Start User Manual
15. The IEC Stream Verification test is run
under three conditions: With +5 V power
turned on and A
RC activated, With +5 V
power turned off, and With +5 V power
turned on and ARC not activated. Follow
the prompts a
nd click OK.
Application Exa
mples
16. When the
tests complete, the application
shows a report in the Report panel.
TekExpress HEAC Quick Start User Manual23
Application Exa
mples
Single Mode Transmitter Tests
Required equipmentModel supported
Real-time osc
Probe
illoscope
Tektronix DPO7000 Series, DPO70000 Series, DPO70000B
Series, DSA 70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix P6
Series, DPO70000B Series, DSA70000 Series, and
DSA70000B Series, and MSO7000 Series Digital and
Mixed Oscil
Tektronix P6243, P6245, or TAP1500 for Tektronix
DPO7000 Series Digital Oscilloscopes
ies Digital and Mixed Signal Oscilloscopes
243 or P6245 for Tektronix DPO70000
loscopes
NOTE. Yo u c
single-ended mode.
Test fixtureTF-HEAC-TPA-P (A/C) or TF-HEAC-TPA-KIT,
TF-HDMID-TPA-KIT
50 Ω SMA TerminatorsAny terminator that meets the following requirements:
DUT
SMA CablesTektronix part number 174-1428-00 (1.5 meter)
able
HDMI C
An HEAC sink device
onix part number 174-1341-00 (1 meter)
Tektr
HDMI c
an also use two differential probes in
50 Ω impedance ± 1% or better
Connects directly to the SMA female connector
able with HEAC capability
24TekExpress HEAC Quick Start User Manual
This tests the DUT for compliance with single mode (Transmitter) tests.
1. Connect the equipment as shown.
Application Exa
mples
Connect the CEC or CDC Controller to the CEC/CDC Control Port on the HEAC-TPA main board.
Connect the HEAC-TPA plug board to the HEAC connector on the Sink DUT.
Connect the 50 Ω SMA terminators to the HEAC-TPA main board.
Set the Impedance Conversion Circuit on the HEAC-TPA main board to 55 Ω.
Connect a single-ended probe from Ch1 of the digital oscilloscope to the HEAC+ probe point on the HEAC-TPA main
board.
Set the HEAC-TPA main board to enable the Sink DUT test.
Connect and set the DC Power Supply to supply +5 V to the HEAC-TPA main board. Set switch S7 to the left (Off
position) on the HEAC-TPA m ain fixture.
Power on the Sink DUT.
Activate the ARC (Single mode) transmission on the HEAC Sink D UT using the CEC/CDC c ontroller.
TekExpress HEAC Quick Start User Manual25
Application Exa
2. Select HEAC-Transmitter as the device
3. Select SingleMode-Tx.
4. Enter the DUT identification in the DUT
mples
type.
The default v
All the single mode transmitter tests are
selected by default.
ID field.
ersion CTS 1.4.
5. Click Show
diagram of the test setup and confirm
that your test setup is correct.
6. Click Configure to open the
Configuration panel.
7. Use the default values in the Compliance
Mode to r
the standards.
NOTE. To change acquisition parameter
,clickUser Defined Mode.A
values
message appears prompting you that If you
make changes to a test, the test may no
be compliant. Click OK to proceed
longer
with the User Defined Mode.
8. Click Close.
Schematic to display a
un the tests for compliance with
9. Click Run in the Select tab.
The application switches to the Acquire
tab.
26TekExpress HEAC Quick Start User Manual
10. Click the Analyze tab to view test
analysis results.
11. The application displays the status of the
tests in the S
tatus Messages window.
Click Save Status to save the status
messages.
12. The Transmit Jitter MAX test is run
at three frequencies. The application
prompts you to configure the DUT for
each of these frequencies: 4.096 MHz,
5.688 MHz, and 6.144 MHz. If your
device supports the frequency, configure
the DUT, and click OK.
Application Exa
mples
TekExpress HEAC Quick Start User Manual27
Application Exa
13. The IEC Stream Verification test is run
mples
under three conditions: with +5 V power
turned on and A
RC activated, with +5 V
power turned off, and with +5 V power
turned on and ARC not activated. Follow
the prompts a
nd click OK.
14. When the tests complete, the application
shows a report in the Report panel.
28TekExpress HEAC Quick Start User Manual
Differential Receiver Tests
Required equipmentModel supported
Real-time osc
Arbitrary wa
Probe
Test fixtu
DUT
SMA CablesTektronix part number 174-1428-00 (1.5 meter)
HDMI CableHDMI cable with H EAC capability
illoscope
veform generator
re
Application Exa
Tektronix DPO7000 Series, DPO70000 Series, DPO70000B
Series, DSA70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix AW
or AWG7000B Series Arbitrary Waveform Generator
Tektronix P6247, P6248, P6330, or P7330 for a
DPO70000 Series, DPO70000B Series, DSA70000 Series,
DPO70000B
Signal Oscilloscope
Tektronix P6330, TDP1000, TDP1500, or TDP3500 for a
Tektronix
TF-HEACTF-HDMID-TPA-KIT
An HEAC sink or source device
Tektronix part number 174-1341-00 (1 meter)
ies Digital and Mixed Signal Oscilloscopes
G5000 Series, AWG5000B, AWG7000 Series,
, and MSO70000 Series Digital and Mixed
DPO7000 Series Digital Oscilloscope
TPA-P (A/C) or TF-HEAC-TPA-KIT,
mples
ests the DUT for compliance with differential receiver tests.
This t
TekExpress HEAC Quick Start User Manual29
Application Exa
1. Connect the DUT to the equipment as shown.
mples
Connect a LAN cross-cable from the digital oscilloscope to the waveform generator, and enable the TekVISA connection
waveform generator and to the digital oscilloscope.
to the
ct an SMA cable to the Ch1 Marker1 output and to Ch2 on the digital oscilloscope.
Conne
ct the HEAC-TPA plug board to the HEAC connector of the Source/Sink DUT.
Conne
ect the Ch1 Analog(+) output of the arbitrary waveform generator to the HEAC+ SMA connector on the HEAC-TPA
Conn
main board and the Ch2 Analog(+) to HEAC–.
Set the Impedance Conversion Circuit on the HEAC-TPA main board to 50 Ω.
Connect a differential probe to the HEAC± differential signal probe point on the HEAC-TPA main board.
Connect the CEC or CDC Controller to the CEC/CDC Control Port of the HEA C-TPA main board.
If you are testing a Sink DUT, set the HEAC-TPA main board to enable the Sink DUT test. Set switch S7 to the left
(Off position) on the HEAC-TPA main board.
ou are testing a Source DUT, set the HEAC-TPA main board to enable the Source DUT test. Activate +4 V bias by
If y
using the +5 V power from the Source DUT. Set switch S7 to the right (On position) on the HEAC-TPA main fixture.
Activate the HEC transmission on the HEAC Source/Sink DUT using the CEC/CDC controller.
30TekExpress HEAC Quick Start User Manual
2. Select HEAC-Receiver as the device
type.
3. Select the Differential-Rx as the Test
Suite.
The default version CTS 1.4.
All the differential mode receiver tests
are selecte
4. Enter the DUT IP Address.
5. Enter the DUT identification in the DUT
ID field.
d by default.
Application Exa
mples
6. Click Sho
diagram of the test setup and confirm
that your test setup is correct.
7. Click Configure to open the
Configur
8. Use the
Mode to run the tests for compliance with
the standards.
Test fixtureT F-HEAC-TPA-P (A/C) or TF-HEAC-T PA-KIT,
DUT
SMA CablesTektronix part number 174-1428-00 (1.5 meter)
HDMI Cab
This tests the DUT for common mode receiver tests.
1. Connect the DUT to the equipment as shown.
veform generator
le
Application Exa
Series, DSA70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix AW
or AWG7000B Series Arbitrary Waveform Generator
TF-HDMID-TPA-KIT
An HEAC source device
Tektro n
HDMI cab
ies Digital and Mixed Signal Oscilloscopes
G5000 Series, AWG5000B Series, AWG7000,
ix part number 174-1341-00 (1 meter)
le with HEAC capability
mples
NOTE. T
Operating DC Voltage Test.
his equipment setup diagram and connections are the same for all common mode receiver tests except the
TekExpress HEAC Quick Start User Manual33
Application Exa
mples
Connect the HEAC-TPA main board to the HEAC connector on the Source DUT.
Connect the CEC or CDC Controller to the CEC/CDC Control Port of the HEA C-TPA main board.
Connect the Ch1 Analog(+) output from a waveform generator to the HEAC+ SMA connector of the HEAC-TPA main
board, and the C h2 Analog(+) to the HEAC– SMA connector.
Set the Impedance Conversion Circuit on the HEAC-TPA main board to 50 Ω.
Set the HEAC-TPA main board to enable the Source DUT test.
Activate +4 V bias using the +5 V power from the Source DUT.
Activate ARC (Common mode) transmission capability for the HEAC Source DUT using the CEC/CDC controller.
34TekExpress HEAC Quick Start User Manual
2. Select HEAC-Receiver as the device
type.
3. Select CommonMode-Rx as the Test
Suite.
All the common mode receiver tests are
selected by default.
The default
version CTS 1.4.
Application Exa
mples
4. Enter the DU
ID field.
5. Click Show Schematic to display a
diagram of the test setup and confirm
that your
6. Click Co
Configuration panel.
7. Use the default values in the C ompliance
Mode to run the tests for compliance with
the sta
NOTE. The User Defined Mode is not
available for HEAC Common Mode Receiver
.
tests
T identification in the DUT
test setup is correct.
nfigure to open the
ndards.
ck Close.
8. Cli
ick Run in the Select panel to run the
9. Cl
test.
The application switches to the Acquire
b.
ta
TekExpress HEAC Quick Start User Manual35
Application Exa
10. Click the Analyze tab.
11. The application displays the status of the
mples
tests in the Status Messages window.
Click Save Sta
tus.
36TekExpress HEAC Quick Start User Manual
Application Exa
mples
12. As the selected
tests are executed,
the application prompts you to verify
the audio response from the DUT.
Click Detecte
d or Not Detected as
appropriate for each test.
13. Follow the on-screen prompts and the
schematics to change the setup. Click
OK to continue.
14. When the tests complete, the application
shows a report in the Report panel.
TekExpress HEAC Quick Start User Manual37
Application Exa
mples
Common Mode Operating DC Voltage Test (Receiver)
Required equipmentModel supported
Real-time osc
Arbitrary wa
Probe
HEAC TPA-
Test fixt
DUT
SMA CablesTektronix part number 174-1428-00 (1.5 meter)
HDMI CableHDMI cable with HEAC capability
illoscope
veform generator
P adapter
ure
Tektronix DPO7000 Series, DPO70000 Series, DPO70000B
Series, DSA 70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix AW
and AWG7000B Series Arbitrary Waveform Generator
Tektronix P6243 or P6245 for Tektronix DPO70000
Series, DPO70000B Series, DSA70000 Series, and
DSA70000B
Mixed Oscilloscopes
Tektronix P6243, P6245, or TAP1500 for Tektronix
DPO7000 Se
Tektroni
TF-HEAC
TF-HDMID-TPA-KIT
An HEAC source device
Tektronix part number 174-1341-00 (1 meter)
G5000 Series, AWG5000B, AWG7000 Series,
Series, and MSO7000 Series Digital and
ries Digital Oscilloscopes
x HEAC-TPA-P
-TPA-P (A/C ) or TF-HEAC-TPA-KIT,
ies Digital and Mixed Signal Oscilloscopes
ests the DUT for compliance with common mode operating DC voltage (Receiver) tests.
This t
38TekExpress HEAC Quick Start User Manual
1. Connect the DUT to the equipment as shown.
Application Exa
mples
Connect the HEAC-TPA plug board to the HEAC connector on the Source DUT.
Connect the CEC or CDC Controller to the CEC/CDC Control Port on the HEAC-TPA main board.
Connect the Ch1 Analog(+) output from a waveform generator to the HEAC+ SMA connector of the HEAC-TPA main
board, and the C h2 Analog(+) to the HEAC– SMA connector.
Set the Impedance Conversion Circuit on the HEAC-TPA main board to 50 Ω.
Connect a single-ended probe from Ch1 of the Digital Oscilloscope to the HEAC+ probe point on the HEAC-TPA main
rd, and a second single-ended probe from Ch2 to the HEAC- probe point.
boa
the HEAC-TPA main board to enable the Source DUT test, and to use +5 V power from the DC Power Supply. Set
Set
switch S7 to the right (On position) on the HEAC-TPA main fixture.
NOTE. Do not use +5 V power from the Source DUT.
Connect and set the DC Power Supply to supply +5 V to the HEAC-TPA main board.
Power on the Source DUT.
Activate the ARC (Common mode) transmission on the HEAC Source DUT using the CEC/CDC controller.
TekExpress HEAC Quick Start User Manual39
Application Exa
2. Select HEAC-Receiver as the device
3. Select the CommonMode-Rx as the
4. Click Deselect All and select the
5. Enter the DUT identification in the DUT
mples
type.
Test Suite.
The default version CTS 1.4.
Operating DC VoltageTest.
ID field.
6. Click Show Schematic to display a
diagram of the test setup and confirm
that your test setup is correct.
7. In the Select panel, highlight the test and
click Configure.
40TekExpress HEAC Quick Start User Manual
8. Use the default values in the C ompliance
Mode to run the tests for compliance with
the standards
NOTE. To change acquisition parameter
values, click User Defined Mode.A
message appe
make changes to a test, the test may no
longer be compliant. Click OK to proceed
with the User
9. Click Close.
10. In the Select panel, click Run to start the
test.
The application switches to the Acquire
tab.
.
ars prompting you that If you
Defined Mode.
Application Exa
mples
TekExpress HEAC Quick Start User Manual41
Application Exa
11. Click the Analyze tab.
12. The application displays the status of the
13. When the tests complete, the application
mples
tests in the Status Messages window.
Click Save St
atus to save the status
messages.
shows a rep
ort in the Report panel.
42TekExpress HEAC Quick Start User Manual
Single Mode Receiver Tests
Required equipmentModel supported
Real-time osc
Arbitrary wa
Probe
HEAC TPA-
Test fixt
DUT
SMA CablesTektronix part number 174-1428-00 (1.5 meter)
HDMI CableHDMI cable with H EAC capability
illoscope
veform generator
P adapter
ure
Application Exa
Tektronix DPO7000 Series, DPO70000 Series, DPO70000B
Series, DSA70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix AW
Series, or AWG7000B Series Arbitrary Waveform Generator
Tektronix P6243 or P6245 for Tektronix DPO70000
Series, DPO70000B Series, DSA70000 Series, and
DSA70000B
Mixed Oscilloscopes
Tektronix P6243, P6245, or TAP1500 for Tektronix
DPO7000 Se
Tektroni
TF-HEAC
TF-HDMID-TPA-KIT
An HEAC sink or source device
Tektronix part number 174-1341-00 (1 meter)
ies Digital and Mixed Signal Oscilloscopes
G5000 Series, AWG5000B Series, AWG7000
Series, and MSO7000 Series Digital and
ries Digital Oscilloscopes
x HEAC-TPA-P
-TPA-P (A/C) or TF-HEAC-TPA-KIT,
mples
TekExpress HEAC Quick Start User Manual43
Application Exa
This tests the DUT for compliance with single mode (Receiver) tests.
mples
1. Connect the DU
T to the equipment as shown.
44TekExpress HEAC Quick Start User Manual
Application Exa
Connect the HEAC-TPA plug board to the HEAC connector on the Source DUT.
Connect the CEC or CDC Controller to the CEC/CDC Control Port on the HEAC-TPA main board.
Connect the Ch1 Analog(+) output from a waveform generator to the HEAC+ SMA connector of the HEAC-TPA main
board.
Set the Impedance Conversion Circuit on the HEAC-TPA main board to 55 Ω.
Connect a single-ended probe from Ch1 of the digital oscilloscope to the HEAC+ probe point on the HEAC-TPA main
board.
mples
Set the HEAC
NOTE. Do not use +5 V power from the Source DUT.
Load the waveform data on the arbitrary waveform generator and generate the waveforms.
Activate the ARC (Single mode) transmission on the HEAC Sink D UT using the CEC/CDC c ontroller.
2. Select HEAC-Receiver as the device
type.
3. Select the SingleMode-Rx as the Test
Suite.
The default version CTS 1.4.
All the single mode receiver tests are
select
4. Enter
ID field.
5. Click Show Schematic to display a
diagram of the test setup and confirm
your test setup is correct.
that
he Select panel, highlight the test and
6. In t
click Configure.
-TPA main to use +5 V power from the DC Power Supply.
ed by default.
the DUT identification in the DUT
TekExpress HEAC Quick Start User Manual45
Application Exa
7. Use the default values in the Compliance
NOTE. To change acquisition parameter
values, click User Defined Mode.A
message appe
make changes to a test, the test may no
longer be compliant. Click OK to proceed
with the User
8. Click Close.
mples
Mode to run the tests for compliance with
the standards
.
ars prompting you that If you
Defined Mode.
9. Click Run i
The application switches to the Acquire
tab.
10. Click the Analyze tab.
11. The application displays the status of the
tests in the Status Messages window.
Click Save Status to save the status
messages.
n the Select panel.
46TekExpress HEAC Quick Start User Manual
Application Exa
mples
12. Thesinglemode
receiver tests are run at
three terminal voltage levels (Vel): 0 V,
2.5 V, and 5 V.
13. The applica
tion prompts you to verify
the audio response from the DUT.
Click Detected or Not Detected as
appropria
te for each test.
14. Repeat Steps 13 and 14 for different
terminal voltage levels and audio
patterns of various frequencies.
15. When the tests complete, the application
shows a report in the Report panel.
TekExpress HEAC Quick Start User Manual47
Application Exa
mples
Single Mode Operating DC Voltage Test (Receiver)
Required equipmentModel supported
Real-time osc
Arbitrary wa
Probe
HEAC TPA-
Test fixt
DUT
SMA CablesTektronix part number 174-1428-00 (1.5 meter)
HDMI CableHDMI cable with HEAC capability
illoscope
veform generator
P adapter
ure
Tektronix DPO7000 Series, DPO70000 Series, DPO70000B
Series, DSA 70000 Series, DSA70000B Series, and
MSO70000 Ser
Tektronix AW
Series, or AWG7000B Series Arbitrary Waveform Generator
Tektronix P6243 or P6245 for Tektronix DPO70000
Series, DPO70000B Series, DSA70000 Series, and
DSA70000B
Mixed Oscilloscopes
Tektronix P6243, P6245, or TAP1500 for Tektronix
DPO7000 Se
x HEAC-TPA-P
Tektroni
TF-HEAC
TF-HDMID-TPA-KIT
An HEAC source device
Tektronix part number 174-1341-00 (1 meter)
ies Digital and Mixed Signal Oscilloscopes
G5000 Series, AWG5000B Series, AWG7000
Series, and MSO7000 Series Digital and
ries Digital Oscilloscopes
-TPA-P (A/C ) or TF-HEAC-TPA-KIT,
ests the DUT for compliance with single mode operating DC voltage (Receiver) tests.
This t
48TekExpress HEAC Quick Start User Manual
1. Connect the DUT to the equipment as shown.
Application Exa
mples
Connect the HEAC-TPA plug board to the HEAC connector on the Source DUT.
Connect the CEC or CDC Controller to the CEC/CDC Control Port on the HEAC-TPA main board.
Connect the Ch1 Analog(+) output from a waveform generator to the HEAC+ SMA connector of the HEAC-TPA main
board.
Set the Coupling & Impedance Conversion Circuit on the HEAC-TPA main board to 55 Ω.
Connect a single-ended probe from Ch1 of the digital oscilloscope to the HEAC+ probe point on the HEAC-TPA main
rd.
boa
the HEAC-TPA main board to enable the Source DUT test, and to use +5 V power from the DC Power Supply. Set
Set
switch S7 to the right (On position) on the HEAC-TPA main fixture.
NOTE. Do not use +5 V power from the Source DUT.
Connect and set the DC Power Supply to supply +5 V to the HEAC-TPA main board.
Power on the Source DUT.
Activate the ARC (Single mode) transmission on the HEAC Source DUT using the CEC/CDC controller.
Load the nominal audio waveform data on the arbitrary waveform generator and generate the waveform.
TekExpress HEAC Quick Start User Manual49
Application Exa
2. Select HEAC-Receiver as the device
3. Select the SingleMode-Rx as the Test
4. Click Deselect All and select the
mples
type.
Suite.
The default version CTS 1.4.
Operatin
g DC Voltage test.
5. Enter th
6. Click Show Schematic to display a
7. In the
e DUT identification in the DUT
ID field.
diagram of the test setup and confirm
ur test setup is correct.
that yo
Select panel, highlight the test and
click Configure.
50TekExpress HEAC Quick Start User Manual
8. Use the default values in the C ompliance
Mode to run the tests for compliance with
the standards
NOTE. To change acquisition parameter
values, click User Defined Mode.A
message appe
make changes to a test, the test may no
longer be compliant. Click OK to proceed
with the User
9. Click Close.
10. Click Run in the Select panel.
The application switches to the Acquire
tab.
.
ars prompting you that If you
Defined Mode.
Application Exa
mples
TekExpress HEAC Quick Start User Manual51
Application Exa
11. The operating DC voltage test runs at
mples
three terminal voltages: 0 V, 2.5 V, and
5 V. Adjust the
voltage and click OK to
continue.
52TekExpress HEAC Quick Start User Manual
12. Click the Analyze tab.
13. The application displays the status of the
tests in the Status Messages window.
Click Save St