BroadR-Reach 3.2 (100BASE-T1) Printable Application Help iii
Table of Contents
iv BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Welcome
The TekExpress BroadR-Reach (100BASE-T1) (BRR) is an automated
compliance test solution for performing electrical characterization of the BroadRReach transmitter as per OPEN Alliance BroadR-Reach (OABR) specification
version 3.2. The IEEE P802.3bw D3.3 (100BASE-T1) specification has a stated
objective to provide electrical interoperability for 100 Mb/s client interface in
reference to the OABR PHY v3.2 specification. BRR and 100BASE-T1 maintain
the same electrical test requirements as it relates to the Physical Medium
Attachment (PMA) transmitter electrical specifications.
BroadR-Reach (100BASE-T1) is an Ethernet-based point-to-point technology for
automotive applications. It provides five test mode signals for electrical
characterization of the Physical Medium Attachment (PMA) sublayer.
Figure 1: Setup panel
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help v
Welcome
Supported Tests
■
Transmitter Clock Frequency
■
Transmitter Timing Jitter Master and Slave
■
Transmitter Output Droop
■
Transmitter Power Spectral Density
■
Transmitter Peak Differential Output
■
Transmitter Distortion
■
Return Loss
Test modes
Table 1: Test modes
Required test modeBroadR-Reach measurement
1 Transmitter Output Droop
2 Transmitter Timing Jitter
See also:
■
Transmitter Timing Jitter-Master Jitter
■
Transmitter Timing Jitter-Slave Jitter
Transmitter Clock Frequency
4 Transmitter Distortion
Return Loss
5 Transmitter Power Spectral Density
5 Transmitter Peak Differential Output
Application Basics
Equipment Connection Setup
Install the Software
Application Directories and Usage
File Name Extensions
vi BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Operating Basics
AWG\AFG Automation
AWG\AFG automation is supported in the TekExpress BroadR-Reach
(100BASE-T1) application for the following measurements. Before selecting the
AWG\AFG automation feature, ensure that a GPIB connection is established
between the oscilloscope and the AWG\AFG. The AFG automation depends on
the measurement used for testing.
■
Transmitter Distortion with disturbing signal: AWG\AFG is used to transmit
a disturbing signal. CH1 and CH1_inverted are used.
■
Return Loss measurement and calibration: AWG\AFG is used to transmit a
wide band signal. CH1 and CH1_inverted are used for transmission of the
signal. A marker signal has to be connected to the auxiliary channel of the
oscilloscope (used as trigger source).
How BRR Automates AWG
\AFG
Automation of the AWG\AFG is taken care in the BRR application in the
following way:
1. Refresh the Instruments Control Settings to observe the connected AWG
\AFG through GPIB.
Figure 2: Instrument Control Settings
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 1
Operating Basics
2. Once the AWG\AFG is listed as shown above, go to the Global Settings tab
from the Configuration menu of the Tests. If the AWG\AFG is supported, it
will be listed as a drop-down menu option next to the Automate with AWG
\AFG label. By default, the application will consider the connected AWG to
be used for automation.
Figure 3: Test Selection tab
3. If you do not want to automate the use of an AWG\AFG, select the Do not
use option in the drop-down menu for that parameter, and then start the
testing.
NOTE. If Automate with AWG\AFG is set as Do not use, you will need to
manually copy the AWG\AFG waveforms from the oscilloscope to the AWG
\AFG. Waveforms are available at C:\Program Files\Tektronix\TekExpress
\TekExpress BroadR-Reach\AWG Waveforms.
2 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Operating Basics
Related Documentation
The following information is available as part of the BRR documentation set.
Table 2: Related documentation
ItemPurposeLocation
HelpApplication operation and User
Interface help
PDF of the helpPrintable version of the
compiled help
Application Help menu
PDF file that ships with BRR
software. Downloadable from
http://www.tek.com/
Application Directories and Usage
The application directory and associated files are organized in the C:\Program
Files\Tektronix\TekExpress\TekExpress BroadR-Reach folder
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 3
Operating Basics
The following table lists the default directory names and their usage.
TekExpress BroadR-Reach\Compliance Suites It contains compliance-specific files and filter
TekExpress BroadR-ReachIt contains the Manuals.
TekExpress BroadR-Reach\ExamplesIt contains various support files.
TekExpress BroadR-Reach\ICPIt contains instrument and BroadR-Reach
TekExpress BroadR-Reach\ImagesIt contains images required for the application.
TekExpress BroadR-Reach\LibIt contains utility files specific to the BroadR-
TekExpress BroadR-Reach\Report GeneratorIt contains style sheets for report generation.
TekExpress BroadR-Reach\ToolsIt contains instrument and BroadR-Reach
It contains the application and associated files.
AFG waveforms ) that get used in return loss
and distortion with disturbing signal .
application libraries.
files.
application-specific interface libraries.
Reach application.
application-specific files.
See Also
View Test-Related Files
4 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Operating Basics
View Test-Related Files
Files related to BRR tests are stored in the X:\BRR\Untitled Session shared
folder. In the BRR folder, each test setup has a test setup file and a test setup
folder, both with the test setup name. The test setup file is preceded by the BRR
icon and usually has no file extension displayed.
Inside the test setup folder is another folder named for the DUT ID used in the
test sessions (the default is DUT001).
Inside the DUT001 folder are the session folders and files. Each session has a
folder and file pair, both named for the test session using the naming convention
(date)_(time). Each session file is stored outside of its matching session folder.
See Also
Each session folder contains image files of any plots generated by the test session
and any waveform files if prerecorded waveform files were used during the
session.
The first time you run a new, unsaved session, the session files are stored in the
X:\BRR\Untitled Session folder. Once you name and save the session, the
Untitled Session folder name is changed to the one you specified.
NOTE. By default, test report files are saved in the session folder. You can change
the report file location for a specific test.
File Name Extensions
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 5
Operating Basics
File Name Extensions
The TekExpress BRR software uses the following file name extensions:
Table 4: File name extensions
File name extensionDescription
.TekXSession files are saved in this format but the
extensions may not be displayed
.pyThe test sequence file
.xmlThe encrypted XML file that contains the test-
specific configuration information
The log file extension is also xml
.wfmThe test waveform file
.mht
.pdf
.calCalibration file used with transmitter tests
Test result reports are saved in this format by
default. Test reports can also be saved in
MHTML format and .pdf.
See Also
Application Directories and Usage
6 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Getting Started
Compatibility
Supported Oscilloscopes
Supported AWG/AFGs
The TekExpress BRR application runs on the following Tektronix oscilloscopes
with bandwidths of 1 GHz or greater:
The TekExpress BRR application supports the following Tektronix arbitrary
waveform and arbitrary function generators:
■
AWG7102 (Option 2): 10 GS/s
■
AWG5014: 1.2 GS/s, 14-bit AWG
■
AWG7122C: 12 GS/s
■
AFG3252: 2 GS/s, 14-bit AFG
■
AFG3102: 1 GS/s, 14-bit AFG
NOTE.
■
You may use AWG automation for Return Loss and Transmitter Distortion
with Disturbing Signal measurements.
■
AFG3100 & AFG3200 (having limit up to 50 MHz) cannot be used for
Return Loss measurements.
■
Without an amplifier, the AWG7000 series generates a maximum voltage of
2 Volt peak-to-peak signal, which cannot meet the 5.4 Volt requirement for
the disturber signal for Transmitter Distortion test.
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 7
Getting Started
Probes
The TekExpress BRR application supports the following Tektronix differential
probes:
■
TDP1500, TDP3500, P6248, and P6247 differential probes
Test Fixtures
The TekExpress BRR application supports the TF-GBE-BTP and TF-BRR-CFD
test fixtures.
See Also
Minimum System Requirements
Minimum System Requirements
The following table shows the minimum system requirements for an oscilloscope
to run TekExpress.
Table 5: System requirements
OscilloscopeFor a list of compatible oscilloscopes, see
Compatibility.
ProcessorSame as the oscilloscope
Operating SystemSame as the oscilloscope
MemorySame as the oscilloscope
Hard DiskSame as the oscilloscope
DisplaySame as the oscilloscope
FirmwareTekScope 7.6.0 and later (Windows 7, 64-bit
only)
Software
■
TekExpress Framework version 4.0
■
Iron Python 2.7.3
■
PyVisa 1.0.0.25
■
Microsoft .NET 4.0 framework
■
Microsoft Internet Explorer 6.0 SP1 or later
■
Adobe Reader 8.0 or equivalent software
for viewing portable document format (PDF)
files
1
1
If TekExpress is running on an instrument having a video resolution lower than 800 x 600. it is recommended that you connect a secondary monitor,
which must be enabled before launching the application.
8 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Getting Started
ProbesFollowing differential probes are supported:
■
All measurements: P6247, P6248,
P6330,TDP1500, TDP3500. For more
details, refer to the BroadR-Reach
measurements section.
BRR FixturesTF-GBE-BTP, TF-BRR-CFD
Other Devices
Frequency DividerThe TF-BRR-CFD test fixture allows you to
CablesSix BNC cables/SMA cables with BNC to SMA
■
Microsoft compatible mouse or compatible
pointing device
■
Four USB ports (two USB ports minimum)
■
PCI-GPIB or equivalent interface for
instrument connectivity
2
phase lock frequency between DUT Transmitter
Clock, oscilloscope , and AWG/AFG sources.
Converts DUT Transmitter Clock from
66.666MHz to 10MHz.
connectors: one TCA-BNC or TCA-SMA.
See Also
Compatibility
2
If TekExpress is installed on a Tektronix oscilloscope, the virtual GPIB port will be used by TekExpress for communicating with oscilloscope
applications. If external GPIB communication devices such as USB-GPIB-HS or equivalent are used for instrument connectivity, ensure that the Talker
Listener utility is enabled in the DPO/DSA Oscilloscope's GPIB menu. For ease of use, connect to an external (secondary) monitor.
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 9
Getting Started
Install the Software
The software can be installed on any compatible instrument running Windows 7.
1. Close all applications (including the TekScope application).
2. Go to the www.tektronix.com Web site and search for BRR to locate the
installation and download the latest installation file.
3. Double-click the executable file to extract the installation files. After
extraction, the installer launches and displays the Install shield Wizard.
Follow the prompts to install the TekExpress BRR application. The software
automatically installs in C:\Program Files\Tektronix\TekExpress\TekExpress
BroadR-Reach.
The installer updates the TekScope Analyze menu to include TekExpress
BRR:
Figure 4: TekScope window
See Also
Minimum System Requirements
10 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Getting Started
Activate the License
Activate the license using the option installation wizard on the oscilloscope.
Follow these steps to activate the TekExpress BRR license:
1. From the oscilloscope menu bar, click Utilities > Option Installation.
The TekScope Option Installation wizard opens.
Figure 5: TekScope Option Installation wizard
2. Instructions for using the Options Installation window to activate licenses for
installed applications is provided in the oscilloscope online help. Press the F1
key on the oscilloscope keyboard to open the Option Installation help topic.
Follow the directions in the topic to activate the license.
See Also
View Version and License Information
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 11
Getting Started
View Version and License Information
Use the following instructions to view application version information and
license information for the application modules.
To view version information:
1. From the Options menu, select About TekExpress.
2. Click the View Version Details link to check the version numbers of the
installed test suites. Close the dialog box when finished.
To view license information:
1. From the oscilloscope Help menu, select About TekScope.
The Options section in the dialog box displays a list of installed options,
including TekExpress BRR.
2. To view the Option key, look in the Option Installation Key section. When
finished, click OK to close the dialog box.
See Also
Start the Application
Activate the License
Options Menu
The first time you open the application after installation, you are required to enter
a license key. The application also checks for a file called Resources.xml located
in the X: drive. If this file is not found, instrument discovery is performed before
launching TekExpress BRR. The Resources.xml file contains information
regarding instruments available on your network.
NOTE. When the application starts, it checks for the appropriate license key. If
the valid license key is not present, the application switches to the Evaluation
mode. If the application fails to detect the key, it continues to run in Evaluation
mode. You are allowed 10 free trials in Evaluation mode. Each time you open the
application without supplying a valid license key, one of the free trials is used.
To run the BRR application from the TekScope menu, select Analyze >
TekExpress BRR.
12 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Getting Started
A initializing screen is displayed, before the launch of the application.
While running the application, you can switch between the oscilloscope screen
and BRR application by clicking the desired window. To keep the application
window on top, select Keep On Top from the Options menu.
NOTE. If the application was not terminated properly during the last use, a dialog
box prompts you to recall the previously unsaved session.
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 13
Getting Started
Options Menu
Options menu
The Options menu is located in the upper right corner of the application.
It has the following selections:
Table 6: Options menu
MenuFunction
Default Test SetupIt opens an untitled test setup with defaults
selected.
Open Test SetupIt opens a saved test setup.
Save Test SetupIt saves the current test setup selections.
Save Test Setup AsIt creates a new test setup based on an existing
one.
Open RecentIt displays a menu of recently opened test
setups to select from.
Instrument Control SettingsIt shows the list of instruments connected to the
test setup and allows you to locate and refresh
connections to connected instruments.
Keep On TopIt keeps the TekExpress BRR application on top
of other open windows on the desktop.
Email settingsIt configures email options for test run and
results notifications.
HelpIt displays TekExpress Help.
About TekExpress
■
It displays application details such as
software name, version number, and
copyright.
■
It provides access to license information for
your BRR installation.
■
It provides a link to the Tektronix Web site.
14 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Getting Started
View Connected Instruments: the Instrument Control Settings
Use the Instrument Control Settings dialog box to view or search for connected
instruments required for the tests. The application uses TekVISA to discover the
connected instruments.
Figure 6: Instrument Control Settings
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 15
Getting Started
To refresh the list of connected instruments:
1. From the Options menu, select Instrument Control Setting.
2. In the Search Criteria section of the Instrument Control Settings dialog box,
select the connection types of the instruments to search for.
Instrument search is based on the VISA layer but different connected cables
determine the resource type, such as LAN, GPIB, USB. For example, if you
choose LAN, you can search for all the instruments supported by TekExpress
that are communicating over the LAN. If the search does not find any
instruments that match a selected resource type, a message appears telling
you that no such instruments were found. Click OK to close the message
window.
NOTE. Ensure that the GPIB option is enabled to detect the real time
oscilloscope in the Global Settings tab.
3. Click Refresh.
TekExpress searches for connected instruments.
4. After discovery, the dialog box lists the instrument-related details based on
the search criteria you selected. For example, if you selected LAN and GPIB
as the search criteria, the application checks for the availability of
instruments over LAN, then GPIB.
The details of the instruments are displayed in the Retrieved Instruments
table. The time and date of instrument refresh is displayed in the Last
Updated field.
16 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Getting Started
Application Controls
Table 7: Application controls descriptions
ItemDescription
Options menu It opens the Options menu for access to global
controls.
PanelsVisual frames with sets of related options. Some
panels are further divided into tabs and other
sections.
Start button
Use the Start button to continuously acquire and
accumulate measurements. If prior acquired
measurements have not been cleared, the new
measurements are added to the existing set.
Stop button
Use the Stop button to abort a test in progress.
Pause \ Continue button
Clear button
Use the Pause button to interrupt the current
acquisition. When a test is paused, the button
name changes to Continue.
Available only on the Results panel.
Use the Clear button to clear all existing
measurement results.
NOTE. Adding or deleting a measurement, or
changing a configuration parameter of an
existing measurement also clears
measurements. This is to prevent the
accumulation of measurement statistics or sets
of statistics that are not coherent.
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 17
Getting Started
ItemDescription
Application window move icon
Place the cursor over the three-dot pattern in
the upper left corner of the application window.
When the cursor changes to a hand, drag the
window to the desired location.
Mini view/ Normal view
Toggles the application between Mini view and
Normal view. Mini view displays the run
messages with the time stamp, progress bar,
Start/Stop button, and Pause/Continue button.
The application moves to Mini view when you
click the Start button.
Exit the Application
See Also
Options menu
Use the following method to exit the application:
NOTE. Using other methods to exit the application results in abnormal
termination of the application.
1.
Click on the application title bar.
2. Do one of the following:
■
A message box appears asking if you really want to exit TekExpress. To
exit, click Yes; otherwise, click No.
■
If you have an unsaved session or test setup open, you are prompted to
save it before exiting. To save it, click Yes. If you do not wish to save,
click No. To remain in the session, click Cancel..
18 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Application Basics
Application Basics
The TekExpress BRR software user interface is intuitive and easy to use. In
addition to the UI, a programmatic interface is available.
The user interface has four main panels, which allows you do the following:
■
■
■
■
■
■
Set up the tests.
Prepare to run the tests.
Run the tests and view the progress of analysis.
View the results of the tests.
Configure and view reports.
Save and recall test setups.
See Also
Overview and Key Specifications
BroadR-Reach 3.2 (100BASE-T1) Printable Application Help 19
Application Basics
20 BroadR-Reach 3.2 (100BASE-T1) Printable Application Help
Loading...
+ 148 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.