Related documentation..................................................................................................................................1
Minimum system requirements......................................................................................................................3
Instruments and accessories required...........................................................................................................4
Installing the software....................................................................................................................................5
File name extensions.....................................................................................................................................8
Operating basics
Launch the application...................................................................................................................................9
Global application controls...........................................................................................................................11
Set acquisition tab parameters...............................................................................................................20
Set configuration tab parameters...........................................................................................................21
Set preferences tab parameters.............................................................................................................23
Status panel overview..................................................................................................................................24
View a report..........................................................................................................................................30
Test setup files overview..............................................................................................................................37
Save a test setup.........................................................................................................................................37
Open (load) a saved test setup....................................................................................................................37
Create a test setup from default settings.....................................................................................................38
Create a test setup using an existing one....................................................................................................38
iv TekExpress® 100GBASE-SR4 Printable Application Help
Welcome
Welcome to the TekExpress 100GBASE-SR4 Automated Measurement Solution Software application. The TekExpress
100GBASE-SR4 provides turnkey testing and characterization of Tx Optical properties key to the SR4 Short Reach Ethernet
Specifications outlined in IEEE 802.3bm. Automation options help the customers to meet their conformance testing needs and
generate detailed reports. User-defined Mode lets customers make changes to the test limits and perform margin testing as part
of extended product characterization.
Key features of TekExpress 100GBASE-SR4 include:
■
80SSR4 is a comprehensive set of 100GBASE-SR4 (IEEE 802.3bm) optical transmitter characterization measurements,
including TDEC. It’s designed to offer a checklist of all relevant tests to verify conformance to the SR4 optical specification.
■
80STDEC offers streamlined high performance Transmitter and Dispersion Eye Closure (TDEC) on a variety of Tektronix
optical acquisition modules. Ideal for manufacturing settings and minimal conformance validation applications.
TekExpress
®
100GBASE-SR4 Printable Application Help v
Welcome
vi TekExpress® 100GBASE-SR4 Printable Application Help
Getting help and support
Related documentation
The following documentation is available as part of the TekExpress® 100GBASE-SR4 Solution application.
Table 1: Product documentation
ItemPurposeLocation
HelpApplication operation and User
Interface help
PDF of the helpPrintable version of the
compiled help
See also
PDF file that ships with 100GBASE-SR4 Solution software
distribution (TekExpress 100GBASE-SR4-Automated-Test-
Solution-Software-Printable-Help-EN-US.pdf).
Technical support
Conventions
Help uses the following conventions:
■
The term "Application," and "Software" refers to the TekExpress 100GBASE-SR4 Solution application.
■
The term “DUT” is an abbreviation for Device Under Test.
■
The term “select” is a generic term that applies to the two methods of choosing a screen item (button, control, list item):
using a mouse or using the touch screen.
Table 2: Icon descriptions
IconMeaning
This icon identifies important information.
This icon identifies conditions or practices that could result in loss of data.
This icon identifies additional information that will help you use the application more
efficiently.
TekExpress® 100GBASE-SR4 Printable Application Help 1
Getting help and support
Technical support
Tektronix values your feedback on our products. To help us serve you better, please send us your suggestions, ideas, or
comments on your application or oscilloscope. Contact Tektronix through mail, telephone, or the Web site. See Contacting
Tektronix at the front of this document for contact information.
When you contact Tektronix Technical Support, please include the following information (be as specific as possible):
General information
■
All instrument model numbers
■
Hardware options, if any
■
Modules used
■
Your name, company, mailing address, phone number, FAX number
■
Please indicate if you would like to be contacted by Tektronix about your suggestion or comments.
Application specific information
■
Software version number
■
Description of the problem such that technical support can duplicate the problem
■
If possible, save the setup files for all the instruments used and the application
■
If possible, save the TekExpress setup files, log.xml, *.TekX (session files and folders), and status messages text file
2 TekExpress® 100GBASE-SR4 Printable Application Help
Getting started
Minimum system requirements
The following table shows the minimum system requirements to install and run the TekExpress 100GBASE-SR4 Solution.
Table 3: System requirements
ComponentDescription
Oscilloscope
ProcessorSame as the oscilloscope
Operating SystemSame as the oscilloscope:
MemorySame as the oscilloscope
Hard DiskSame as the oscilloscope
DisplaySuper VGA resolution or higher video adapter (800 x 600 minimum video resolution for small
Firmware
■
Tektronix DSA8300 Digital Serial Analyzer
■
Frimware Version: 6.3.1.3 or greater
■
80SJNB Software Version: 3.2.4.0 or greater
■
Opt ADVTRIG
■
Opt JNB01/02
■
80CXX series Optical Sampling module
fonts or 1024 x 768 minimum video resolution for large fonts). The application is best viewed at
96 dpi display settings
■
TekScope 6.3.1.3 or greater (for Windows 7)
1
■
80SJNB Software Version: 3.2.4.0 or greater
Software
Other Devices
1
If TekExpress is running on an instrument that has a video resolution less than 800x600, connect and configure a second monitor to the instrument.
■
IronPython 2.7.3 installed
■
PyVisa 1.0.0.25 installed
■
Microsoft .NET 4.0 Framework
■
Microsoft Internet Explorer 7.0 SP1 or greater, or other Web browser for viewing reports
■
Adobe Reader software 7.0 or greater for viewing portable document format (PDF) files
■
Microsoft compatible mouse or compatible pointing device.
■
Two USB ports (four USB ports recommended).
TekExpress® 100GBASE-SR4 Printable Application Help 3
Getting started
Instruments and accessories required
100GBASE-SR4 application is launched on DSA8300 sampling scope. The following table lists the instruments and accessories
required for this application.
Table 4: Instruments and accessories required for 100GBASE-SR4 application
Instrument/AccessoryModel numberQuantity
Sampling OscilloscopeDSA83001
Clock Recovery UnitCR286A1
Optical Modules80C15-CRTP-MM for Compliance mode
80C10/80C14/80C15/80C15-CRTP-MM
for User Defined mode
Optical Power MeterAny make which supports 850 nm
wavelength
Phase Reference82A04B (Optional)
SMA Cables174-6023-00 : To connect sub rate clock
out of CRU to trigger
2
3
174-6023-00 : To connect clock out of
CRU to phase reference module
3
174-6023-00 : To connect electrical
output of the 80C15 module to CRU input.
Deskew the cables before use.
VBR (Variable Back Reflector)A Splitter and a Variable Back Reflector is
required for TDEC measurement, as
specified in Equipment setup. The splitter
should be the smallest split possible to
meet the back reflection requirement. For
most accurate result measure the back
reflection.
1
1
1
1
1
2
1
2
Required to reach jitter noise floors below 100fsec
3
Available with 80A08 accessory kit
4 TekExpress® 100GBASE-SR4 Printable Application Help
Getting started
Installing the software
Follow the steps to download and install the latest TekExpress 100GBASE-SR4 Solution. See Minimum system requirements for
compatibility.
1.Type the URL www.tek.com in the address bar of web browser and click Software Downloads
2.Enter TekExpress 100GBASE-SR4 Solution in the Enter your keywords field, and click Search
3.Select the latest version of software and follow the instructions to download. Copy the executable file into the oscilloscope.
4.Double-click the executable and follow the on-screen instructions. The software is installed at C:\Program Files
\Tektronix\TekExpress\TekExpress 100GBASE-SR4\
5.Map My TekExpress folder to drive X: and Set My TekExpress folder permissions
6.Select Application > 100GBASE-SR4 from the TekScope menu to launch the application
View software version
Use the following instructions to view version information for the application and for the application modules such as the
Programmatic Interface and the Programmatic Interface Client.
To view version information for 100GBASE-SR4, click
NOTE. This example shows a typical Version Details dialog box, and may not reflect the actual values as shown when you open
this item in the application.
My TekExpress folder settings
Map My TekExpress folder to drive X:
The first time you run TekExpress 100GBASE-SR4, it creates the following folders on the oscilloscope:
■
\My Documents\My TekExpress\100GBASE-SR4
button in the TekExpress application and select About TekExpress.
Shared My TekExpress folder is mapped to drive X: on the instrument running the 100GBASE-SR4 application. 100GBASESR4 uses this shared folder to save session waveform files and for other application file transfer operations.
Follow the below procedure to map the My TekExpress folder on the instrument to be drive X:
1.Open Microsoft Windows Explorer.
2.From the Windows Explorer menu, click Computer and select Map network drive.
TekExpress® 100GBASE-SR4 Printable Application Help 5
Getting started
3.Select the Drive letter as X: (if there is any previous connection on X:, disconnect it first through Tools > Disconnect
Network drive menu of Windows Explorer. If you do not see the Tools menu, press the Alt key).
4.In the Folder field, enter the remote My TekExpress folder path (for example, \\192.158.97.65\My
TekExpress).
To determine the IP address of the instrument where the My TekExpress folder exists, do the following:
1.On the instrument where the My TekExpress folder exists, click Start and select Run.
2.Enter cmd and press Enter.
3.At the command prompt, enter ipconfig and press Enter.
NOTE. The My TekExpress folder has the share name format
If the instrument is not connected to a domain, the share name format is
TekExpress
.
<domain><user ID>My TekExpress
<instrument name><user ID>My
.
NOTE. If the X: drive is mapped to any other shared folder, the application displays a warning message asking you to disconnect
the X: drive manually.
See also. Set My TekExpress folder permissions
Application directories
File name extensions
Set My TekExpress folder permissions
Make sure that the My TekExpress folder has read and write access. Also verify that the folder is not set to be encrypted:
1.Right-click the folder and select Properties.
2.Select the General tab and then click Advanced.
3.In the Advanced Attributes dialog box, make sure that the option Encrypt contents to secure data is NOT selected.
6 TekExpress® 100GBASE-SR4 Printable Application Help
Getting started
4.Click the Security tab and verify that the correct read and write permissions are set.
See also. Map My TekExpress folder to Drive X:
Application directories
File name extensions
Application directories
TekExpress 100GBASE-SR4 application
The TekExpress 100GBASE-SR4 application files are installed at the following location:
ICPContains instrument and TekExpress 100GBASE-SR4 application-specific interface
libraries
ImagesContains images of the TekExpress 100GBASE-SR4 application
LibContains utility files specific to the TekExpress 100GBASE-SR4 application
Report GeneratorContains style sheets for report generation
ToolsContains instrument and TekExpress 100GBASE-SR4 application-specific files
See also
View test-related files
File name extensions
File name extensions
The TekExpress 100GBASE-SR4 application uses the following file name extensions:
File name extensionDescription
.TekXApplication session files (the extensions may not be displayed)
.pyPython sequence file
.xmlTest-specific configuration information (encrypted) files
.csvTest waveform files
.mhtTest result reports (default)
.pdfTest result reports
.xsltStyle sheet used to generate reports
See also
View test-related files
Application directories
Application log files
Test reports can also be saved in HTML format
Application help document
8 TekExpress® 100GBASE-SR4 Printable Application Help
Operating basics
Launch the application
To launch the TekExpress 100GBASE-SR4 Solution, select Application > 100GBASE-SR4 from the TekScope menu.
When you launch the application for the first time, the file C:\Users\<username>\My Documents\My TekExpress
\100GBASE-SR4\Resources.xml is mapped to drive X:. This file contains information about available network-connected
instruments. The session files are stored in X:\100GBASE-SR4\. If this file is not found, then the application runs Instrument
Discovery Program to detect the connected instruments before launching 100GBASE-SR4 Solution.
NOTE. Map My TekExpress folder to drive X: and set My TekExpress folder permissions before running tests for the first time.
If the application goes behind the oscilloscope application, click Application > 100GBASE-SR4 to bring it to the front. To keep
the 100GBASE-SR4 application window on top, select Keep On Top from the 100GBASE-SR4 Options menu.
TekExpress
®
100GBASE-SR4 Printable Application Help 9
Operating basics
See also
Required My TekExpress folder settings
Application controls
Application panel overview
Application panels overview
TekExpress 100GBASE-SR4 Solution uses panels to group related configuration, test, and results settings. Click any button to
open the associated panel. A panel may have one or more tabs that list the selections available in that panel. Controls in a panel
can change depending on settings made in that panel or another panel.
10 TekExpress® 100GBASE-SR4 Printable Application Help
Operating basics
Table 6: Application panels overview
Panel NamePurpose
Setup panelThe Setup panel shows the test setup controls. Click the Setup button to open this panel.
Use this panel to:
■
Set DUT tab parameters
■
Select tests
■
Set acquisition tab parameters
■
Set configuration tab parameters
■
Set test notification parameters in the Preferences tab
Status panelView the progress and analysis status of the selected tests, and view test logs.
Results panelView a summary of test results and select result viewing preferences.
Reports panelBrowse for reports, save reports as specific file types, specify report naming conventions, select
report content to include (such as summary information, detailed information, user comments,
setup configuration, application configuration), and select report viewing options.
See also
Application controls
Global application controls
Application controls
This section describes the application controls.
Table 7: Application controls description
ItemDescription
Options menu
Test Panel buttons
Menu to display global application controls.
Controls that open panels for configuring test settings and options.
TekExpress® 100GBASE-SR4 Printable Application Help 11
Operating basics
ItemDescription
Start / Stop button
Use the Start button to start the test run of the measurements in the selected order. If prior
acquired measurements are not cleared, then new measurements are added to the existing set.
The button toggles to the Stop mode while tests are running. Use the Stop button to abort the
test.
Pause / Continue button
Use the Pause button to temporarily pause the acquisition. When a test is paused, this button
changes to “Continue.”
Clear button
Use the Clear button to clear all existing measurement results. Adding or deleting a
measurement, or changing a configuration parameter of an existing measurement, also clears
measurements. This is to prevent the accumulation of measurement statistics or sets of
statistics that are not coherent. This button is available only on Results panel.
Application window move icon
Place the cursor over the three-dot pattern in the upper left corner of the application window.
When the cursor changes to a hand, drag the window to the desired location.
Minimize icon
Click to minimize the application.
Close icon
Click to close the application.
See also. Application panel overview
12 TekExpress® 100GBASE-SR4 Printable Application Help
Operating basics
Options menu overview
To access Options menu, click in the upper-right corner of the application. It has the following:
Options menu
MenuFunction
Default Test SetupOpens an untitled test setup with defaults selected
Values displayed in Wavelength, Filter, and Bandwidth fields depend on the installed optical
module.
Open Test SetupOpens a saved test setup
Save Test SetupSaves the current test setup
Save Test Setup AsSaves the current test setup with a different file name or file type
Open RecentDisplays the recently opened test setups to open
Instrument Control SettingsDetects, lists, and refreshes the connected instruments found on specified connections (LAN,
GPIB, USB, and so on)
Keep On TopKeeps the TekExpress 100GBASE-SR4 application on top in the desktop
Email SettingsUse to configure email options for test run and results notifications
Open Current Suite
RunSession
HelpDisplays the TekExpress 100GBASE-SR4 help
About TekExpress
Allows the user to select the specific run setup, from the saved session
■
Displays application details such as software name, version number, and copyright
■
Provides a link to the end-user license agreement
■
Provides a link to the Tektronix Web site
TekExpress® 100GBASE-SR4 Printable Application Help 13
Operating basics
See also. Application controls
TekExpress instrument control settings
Use TekExpress Instrument Control Settings dialog box to search the instruments (resources) connected to the application. You
can use the Search Criteria to search the connected instruments depending on the connection type. The details of the connected
instrument is displayed in the Retrieved Instruments window.
You can access this dialog box from the Options menu.
The connected instruments displayed here can be selected for use under Global Settings in the test configuration section.
NOTE. Under Instrument Control Settings, select GPIB Option (Default setting), when using TekExpress 100GBASE-SR4
application.
See also. Options menu overview
View connected instruments
Use TekExpress Instrument Control Settings dialog box to search the instruments (resources) connected to the application. The
application uses TekVISA to discover the connected instruments.
NOTE. The correct instruments required for the test setup must be connected and recognized by the application before running
the test.
To refresh the list of connected instruments:
1.From the Options menu, select Instrument Control Settings.
2.In the Search Criteria section of the Instrument Control Settings dialog box, select the connection types of the instruments
to search.
Instrument search is based on the VISA layer, but different connections determine the resource type, such as LAN, GPIB,
and USB. For example, if you choose LAN, the search will include all the instruments supported by TekExpress that are
communicating over the LAN.
3.Click Refresh. TekExpress searches for connected instruments.
14 TekExpress® 100GBASE-SR4 Printable Application Help
Operating basics
4.After searching, the dialog box lists the instrument-related details based on the search criteria. For example, For the Search
Criteria as LAN and GPIB, the application displays all LAN and GPIB instruments connected to the application.
The details of the instruments are displayed in the Retrieved Instruments table. The time and date of instrument refresh is
displayed in the Last Updated field.
See also. Equipment connection setup
Configure email settings
Use the Email Settings utility to get notified by email when a measurement completes, or produces any error condition. Follow
the steps to configure email settings:
1.Select Options > Email Settings to open the Email Settings dialog box.
2.(Required) For Recipient email Address(es), enter one or more recipient email addresses. To include multiple addresses,
separate the addresses with commas.
3.(Required) For Sender’s Address, enter the email address used by the instrument. This address consists of the instrument
name, followed by an underscore, followed by the instrument serial number, then the @ symbol, and the email server ID.
For example: DSA8300_B130099@yourcompany.com.
TekExpress® 100GBASE-SR4 Printable Application Help 15
Operating basics
4.(Required) In the Server Configuration section, type the SMTP Server address of the Mail server configured at the client
location, and the SMTP Port number, in the corresponding fields.
If this server requires password authentication, enter a valid login name, password, and host name in the corresponding
fields.
NOTE. If any of the above required fields are left blank, the settings will not be saved and email notifications will not be sent.
5.In the Email Attachments section, select from the following options:
■
Reports: Select to receive the test report with the notification email.
■
Status Log: Select to receive the test status log with the notification email. If you select this option, then also select
whether you want to receive the full log or just the last 20 lines.
6.In the Email Configuration section:
■
Enter a maximum file size for the email message. Messages with attachments larger than this limit will not be sent. The
default is 5 MB.
■
Enter the number in the Number of Attempts to Send field, to limit the number of attempts that the system makes to
send a notification. The default is 1. You can also specify a timeout period.
7.Select the Email Test Results When complete or on error check box. Use this check box to quickly enable or disable
email notifications.
8.To test your email settings, click Test Email.
9.To apply your settings, click Apply.
10. Click Close when finished.
Email Settings
16 TekExpress® 100GBASE-SR4 Printable Application Help
Operating basics
Setup panel
Setup panel overview
The Setup panel contains sequentially ordered tabs that help you guide through the test setup and execution process.
Set DUT parameters
Use the DUT tab to select parameters for the device under test. This settings are global and apply to all tests for the current
session. DUT settings also affect the list of available tests in the Test Selection tab.
TekExpress® 100GBASE-SR4 Printable Application Help 17
Operating basics
Click Setup > DUT to access the DUT parameters:
Table 8: DUT tab settings
SettingDescription
DUT IDAdds an optional text label for the DUT to reports. The default value is DUT001. The maximum
number of characters is 32.
You cannot use the following characters in an ID name: (.,..,...,\,/:?”<>|*)
Opens a Comments dialog box in which to enter optional text to add to a report. Maximum size
Comments icon (to the
right of the DUT ID field)
Acquire TypeLive waveforms.
Device100GBASE-SR4 measurements are taken on Device only. This field is not editable.
Test PointTP2. This field is not editable.
Standard100GBASE-SR4, IEEE 802.3bm, Section 95. This field is not editable.
Device Profile
Optical Module Settings
Data Rate Controls - Data
Rate
WavelengthSelect the wavelength from the drop-down list.
is 256 characters. To enable or disable comments appearing on the test report. See Select
report options for details.
This application performs analysis on live waveforms only. This field is not editable.
Set the data rate to be tested.
18 TekExpress® 100GBASE-SR4 Printable Application Help
Operating basics
NOTE. The installed optical module sets the available wavelength, filter, and bandwidth selections.
You can set either the filter or bandwidth value, but not both.
See also. Select a test
Select tests
Use the Test Selection tab to select the TekExpress 100GBASE-SR4 tests.
Table 9: Test Selection tab settings
SettingDescription
TestsClick on a test to select or unselect. Highlight a test to show
details in the Test Description pane.
Test DescriptionShows a brief description of the highlighted test in the Test
field.
See also. Set acquisition tab parameters
TekExpress® 100GBASE-SR4 Printable Application Help 19
Operating basics
Set acquisition tab parameters
Use the Acquisitions tab to view test acquisition parameters. The contents displayed on this tab depends on the DUT type and
the tests selected.
NOTE. 100GBASE-SR4 acquires all waveforms needed by each test group before performing analysis.
Table 10: Acquisitions tab settings
SettingDescription
View Optical Modules buttonClick to view the detected optical modules that are installed in
the instrument.
Calibration buttonClick to view the status of Oscilloscope calibration, External
attenuation calibration, and Instrument noise calibration.
Update these parameters by clicking the associated Refresh or
Measure button. See Pre-measurement calibration for details.
Show Acquire ParametersSelect to view the acquisition parameters.
20 TekExpress® 100GBASE-SR4 Printable Application Help
Loading...
+ 82 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.