Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077-0001
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
TekConnect, TekVISA, FastFrame, and VocalLink are registered trademarks of Tektronix, Inc.
WARRANTY
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and
workmanship for a period of one (1) year from the date of shipment. If this product proves defective during its
warranty period, Tektronix, at its option, will either repair the defective product without charge for parts and labor,
or provide a replacement in exchange for the defective product.
This warranty applies only to products returned to the designated Tektronix depot or the Tektronix authorized
representative from which the product was originally purchased. For products returned to other locations,
Customer will be assessed an applicable service charge. The preceding limitation shall not apply within the
European Economic Area, where products may be returned for warranty service to the nearest designated service
depot regardless of the place of purchase.
In order to obtain service under this warranty, Customer must provide the applicable office of Tektronix or its
authorized representative with notice of the defect before the expira tion of the warranty period and make suitable
arrangements for the performance of service. Customer shall be responsible for packaging and shipping the
defective product to the service center designated by Tektronix or its representative, with shipping charges
prepaid. Tektronix or its representative shall pay for the return of the product to Customer. Customer shall be
responsible for paying any associated taxes or duties.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty:
a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or
service the product;
b) to repair damage resulting from improper use or connection to incompatible equipment;
c) to repair any damage or malfunction caused by the use of non-Tektronix supplies or consumables;
d) to repair a product that has been modified or integrated with other products when the effect of such
modification or integration increases the time or difficulty of servicing the product; or
e) to repair damage or malfunction resulting from failure to perform user maintenance and cleaning at the
frequency and as prescribed in the user manual (if applicable).
THE ABOVE WARRANTIES ARE GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’
RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE
REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS
VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE
POSSIBILITY OF SUCH DAMAGES.
Table of Contents
Getting Started
General Safety Summaryxiii...................................
Table A--12: Certifications and compliancesA--39...................
Table B--1: Supported measurements and their definitionB--1........
Table B--2: Supported measurements and their definitionB--9........
Table C--1: File menu commandsC--1............................
Table C--2: Edit menu commandsC --3............................
Table C--3: Vertical menu commandsC--4.........................
T able C--4: Horiz/Acq menu commandsC--5.......................
Table C--5: Trig menu commandsC--7............................
Table C--6: Display menu commandsC-- 9.........................
Table C--7: Cursor menu commandsC--11.........................
Table C--8: Measure menu commandsC--11........................
Table C--9: Masks menu commandsC --13..........................
T able C--10: Math menu commandsC--14..........................
Table C--11: Utilities menu commandsC--15........................
Table C--12: Help menu commandsC-- 16..........................
xii
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read
the General Safety Summary in other system manuals for warnings and cautions
related to operating the system.
ToAvoidFireor
Personal Injury
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and Disconnect Properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
Wear Eye Protection. Wear eye protection if exposure to high-intensity rays or
laser radiation exists.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Provide Proper Ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
xiii
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
CAUTION
Refer to Manual
Mains Disconnected
OFF (Power)
WARNING
High Voltage
Mains Connected
ON (Power)
Protective Ground
(Earth) Terminal
Standby
xiv
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Preface
About This Manual
This user manual covers the following information:
HDescribes the capabilities of the instrument, how to install it and how to
reinstall its software
HExplains how to operate the instrument: how to control acquisition of,
processing of, and input/output of information
HLists specifications and accessories of the instrument
This manual is composed of the following chapters:
HGetting Started shows you how to configure and install your instrument and
provides an incoming inspection procedure.
HOperating Basics uses maps to describe the various interfaces for controlling
the instrument, including the front panel and the software user interface.
These maps provide overviews of the product and its functions from several
viewpoints.
HReference comprises an encyclopedia of topics (see Overview on page 3--1)
that describe the instrument interface and features, and gives background and
basic information on how to use them. (The online help onboard the
instrument application describes the interface, features, and their usage;
detailed descriptions of all programming commands are found in the
Programmer Online Guide.)
HAppendices provides additional information including the specifications,
measurements, and cleaning information.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
xv
Preface
Related Manuals and Online Documents
This manual is part of a document set of standard-accessory manuals and online
documentation; this manual mainly focuses on installation, background, and user
information needed to use the product features. See the following list for other
documents supporting instrument operation and service. (Manual part numbers
are listed in Accessories & Options on page 1--37.)
Document nameDescription
Online HelpAn online help system, integrated with the User Interface appl ication that ships with this
product. The help is preinstalled in the instrument.
ReferenceA quick reference to major features of the instrument and how they operate.
Programmer Online GuideAn alphabetical listing of the programming commands and other information related to
controlling the instrument over the GPIB
help program and as a PDF manual.
Service ManualDescribes how to service the instrument to the module level. This optional manual must
be ordered separately.
Option SM Serial Mask Testing
Option ST Serial Pattern Trigger User Manual
1Located on the Product Software CD. See CD instructions for installation instructions.
Describes how to use serial mask testing and serial pattern triggers.
1
. This document is available as both an online
For more information on how the product documentation relates to the instrument operating interfaces and features, see Documentation Map on page 2--2.
xvi
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Contacting Tektronix
Preface
Phone1-800-833-9200*
AddressTektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Web sitewww.tektronix.com
Sales support1-800-833-9200, select option 1*
Service support1-800-833-9200, select option 2*
Technical supportEmail: techsupport@tektronix.com
1-800-833-9200, select option 3*
6:00 a.m. -- 5:00 p.m. Pacific time
*This phone number is toll free in North America. After office hours, please leave a
voice mail message.
Outside North America, contact a Tektronix sales office or distributor; see the
Tektronix web site for a list of offices.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
xvii
Preface
xviii
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Product Description
This chapter describes the CSA7000 Series Communications Signal Analyzers,
the TDS7000 Series Digital Phosphor Oscilloscopes, and the TDS6000 Series
Digital Storage Oscilloscopes and their options. Following this description are
three sections:
HInstallation shows you how to configure and install the instrument, as well
as how to reinstall the system software included with the product.
HIncoming Inspection provides a procedure for verifying basic operation and
functionality.
HAccessories & Options lists the standard and optional accessories for this
product.
Models
This manual supports the following instruments:
HCSA7404 Communications Signal Analyzer
HCSA7154 Communications Signal Analyzer
HTDS7404 Digital Phosphor Oscilloscope
HTDS7254 Digital Phosphor Oscilloscope
HTDS7154 Digital Phosphor Oscilloscope
HTDS7054 Digital Phosphor Oscilloscope
HTDS7104 Digital Phosphor Oscilloscope
HTDS6604 Digital Storage Oscilloscope
HTDS6404 Digital Storage Oscilloscope
Differences between the instruments will be called out when necessary.
Otherwise, the material applies to all instruments. The word “instrument” refers
to all products.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 1
Product Description
Key Features
CSA7000 Series, TDS7000 Series, and TDS6000 Series instruments are high
performance solutions for verifying, debugging, and characterizing sophisticated
electronic designs. The series features exceptional signal acquisition performance, operational simplicity, and open connectivity to the design environment.
Classic analog-style controls, a large touch-sensitive display, and graphical
menus provide intuitive control. Open access to the Windows operating system
enables unprecedented customization and extensibility. Key features include:
HUp to 6 GHz bandwidth and 20 GS/s real time sampling rate, depending on
the model
HRecord lengths up to 32,000,000 samples, depending on model and option
HFast acquisition at up to 400,000 acquisitions per second for analog
instrument emulation and isolation of data-dependent failures during
conformance/performance testing and for examination of very low-level
signals in pseudo-random bit streams
HCSA7000 Series: Communication signal analysis, serial mask testing, serial
pattern triggering, and triggering on communications signals. Use of these
features is described in the Option SM Serial Mask Testing and Option ST
and clock recovery provides single-connection convenience, protects
integrity of system calibration, and increases versatility
HUp to 1% DC vertical gain accuracy, depending on the model
HFour input channels (each with 8-bit resolution), CH 3 signal output, and
auxiliary trigger input and output
HSample, envelope, peak-detect, high-resolution, waveform database, and
average acquisition modes
HFull programmability, with an extensive GPIB-command set and a message-
based interface
1- 2
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Product Description
HTrigger modes include edge, logic, pulse, serial (CSA7000 Series, optional
on TDS7000 Series and TDS6000 Series, and not available on TDS7104 and
TDS7054), communication (CSA7000 Series and optional on TDS7000
Series and TDS6000 Series), and sequence at up to 4 GHz bandwidth,
depending on the model
HPowerful built-in measurement capability, including histograms, automatic
measurements, eye pattern measurements (CSA7000 Series and optional on
TDS7000 Series and TDS6000 Series), and measurement statistics
HA large 10.4 inch (264.2 mm) color display that supports color grading of
waveform data to show sample density
HMultiView Zoom to view and compare up to four zoom areas at a time. Lock
and manually or automatically scroll up to four zoom areas.
HAn intuitive, graphical user interface (UI), with online help that is built in
and available on screen
HInternal, removable disk storage
Product Software
HWide array of probing solutions
The instrument includes the following software:
HSystem Software, which includes a specially configured version of
Windows 2000, comes preinstalled on the instrument. Windows 2000 is the
operating system on which the user-interface application of this product runs,
and provides an open desktop for you to install other compatible applications. Do not attempt to substitute any version of Windows that is not
specifically provided by Tektronix for use with your instrument.
HProduct Software. Comes preinstalled on the instrument. This software,
running on Windows 2000, is the instrument application. This software starts
automatically when the instrument is powered on, and provides the user
interface (UI) and all other instrument control functions. You can also
minimize the instrument application.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 3
Product Description
HSupport Software. Not preinstalled on the instrument. The compact discs,
included with the instrument, contain additional software and files that may
be useful to you:
HReadme file. This file contains release notes and updates that could not
be included in other product documentation.
HGPIB Programmer Online Help software. This software, in an online
help format, contains the information that you need to program the
instrument through its GPIB interface. A printable PDF file of this
information is also available on the compact disc.
HPerformance Verification Procedures. The compact disc contains
instructions to perform a performance verification.
See the instructions for the compact discs for information about installing the
support software.
Occasionally new versions of software for your instrument may become
available at our web site. See Contacting Tektronix on page xvii in Preface.
Software Upgrade
Tektronix may offer software upgrade kits for the instrument. Contact your
Tektronix service representative for more information (see Contacting Tektronix
on page xvii).
1- 4
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Installation
This chapter covers installation of the instrument, addressing the following
topics:
HUnpacking on page 1--6
HChecking the Environment Requirements on page 1--7
HConnecting Peripherals on page 1--7
HPowering On the Instrument on page 1--9
HShutting Down the Instrument on page 1--10
HCreating an Emergency Startup Disk on page 1--11
HBacking Up User Files on page 1--11
HInstalling Software on page 1--12
HEnabling Your LAN and Connecting to a Network on page 1--15
HSetting up a Dual Display on page 1--17
CAUTION. Be sure to create your emergency startup disk as described on
page 1--11. You may need that disk if you ever need to reinstall Windows 2000
from the instrument hard drive.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 5
Installation
Unpacking
Verify that you have received all of the parts of your instrument. The graphical
packing list shows the standard accessories that you should find in the shipping
carton (probes depend on the option you ordered.) You should also verify that
you have:
HThe correct power cord for your geographical area.
HThe compact discs that include copies of the software installed on the
instrument and additional support software that may be useful to you: the
Operating System Restore, Product Software, and Optional Applications
Software. Store the product software in a safe location where you can easily
retrieve it.
NOTE. The certificate of authenticity (Windows 2000 licence agreement) is
attached to the rear of your instrument. This certificate proves your ownership of
the Windows operating system in your instrument. Without this certificate, you
might have to purchase a new Windows license if the hard disk in your instrument ever needs rebuilding or replacement.
HAll the standard and optional accessories that you ordered.
Fill out and send in the customer registration card.
1- 6
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Checking the Environment Requirements
Read this section before attempting any installation procedures. This section
describes site considerations, power requirements, and ground connections for
your instrument.
Installation
Site Considerations
Operating Requirements
Connecting Peripherals
The instrument is designed to operate on a bench or on a cart in the normal
position (on the bottom feet). For proper cooling, at least three inches (7.62 cm)
of clearance is required on both sides of the instrument, and the bottom requires
the clearance provided by the instrument feet.
If you operate the instrument while it is resting on the rear feet, make sure that
you properly route any cables coming out of the rear of the instrument to avoid
damaging them.
CAUTION. To prevent damage to the instrument, keep the bottom and sides of the
instrument clear of obstructions for proper cooling.
The Specifications in Appendix A list the operating requirements for the
instrument. Power source, temperature, humidity, and altitude are listed.
The peripheral connections are the same as those you would make on a personal
computer. The connection points are shown in Figure 1--1. See Table 1--1 on
page 1--9 for additional connection information.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 7
Installation
CAUTION. To avoid product damage, either power off the instrument or place the
instrument in Standby power mode before installing any accessories except a
USB mouse or keyboard to the instrument connectors. (You can connect and
disconnect USB devices with the power on.) See Shutting Down the Instrument
on page 1--10.
DescriptionIcon/LabelLocations
Monitor (PC only, for
dual display
operation) ......
Printer.............
RS-232..........
Network................
Keyboard ................
Mouse.................
USB..................
Audio line out......... ....
Audio line in .............
Removableharddrive...........
Compactdiskdrive.......
GPIB........
Instrument monitor
(large-screen instrument
display)...........
CardSlot......
Figure 1- 1: Locations of peripheral connectors on rear panel
1- 8
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Installation
Table 1- 1: Additional accessory connection information
ItemDescription
MonitorIf you use a nonstandard monitor, you may need to change the
Windows 2000 display settings to achieve the proper resolution
for your monitor. To set up a dual display, see page 1--17.
PrinterConnect the printer to the EPP (enhanced parallel port)
connector directly. If your printer has a DB-25 connector, use
the adapter cable that came with your printer to connect to the
EPP connector. For information on printer usage, see PrintingWaveforms on page 3--277.
RackmountRefer to the Rackmount Installation Instructions for information
on installing the rackmount kit.
OtherRefer to the Readme file on the Product Software CD for
possible additional accessory installation information not
covered in this manual.
Powering On the Instrument
Follow these steps to power on the instrument for the first time.
Either one of the following fuse sizes can be used, each size requires a different
fuse cap. Both fuses must be the same type. See Table 1--2 and Figure 1--2.
Table 1- 2: Line fuses
Line voltageDescriptionPart number
100 V to 250 V operationUL198G and CSA C22.2, No.
CAUTION. Connect the keyboard, mouse, and other accessories before applying
power to the product.
59, fast acting: 8 A, 250 V
IEC127, sheet 1, fast acting
“F”, high breaking capacity:
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 9
Installation
Rear panelFront panel
3
Turn on
the power.
Check the
fuses.
Figure 1- 2: Powering on the instrument
Shutting Down the Instrument
When you push the front-panel On/Standby switch, the instrument starts a
shutdown process (including a Windows shutdown) to preserve settings and then
removes power from most circuitry in the instrument. Avoid using the rear-panel
power switch or disconnecting the line cord to power off the instrument.
NOTE. If you do not use the On/Standby switch to shut down the instrument
before powering off the instrument, the instrument will be in the factory Default
Setup when powered on the next time.
To completely remove power to the instrument, perform the shutdown just
described, set the power switch on the rear panel to off, and then remove the
power cord from the instrument.
21
Connect the
power cord.
4
If needed, push the On/Standby
switch to power on the instrument.
1- 10
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Creating an Emergency Startup Disk
Now that you have completed the basic installation process, you should create an
emergency startup disk that you can use to restart your instrument in case of a
major hardware or software failure. Store this disk in a safe place.
CAUTION. Create this disk and store it in a safe place. It may allow you to
recover your Windows 2000 installation without rebuilding the entire instrument
hard disk.
The emergency startup disk contains basic files to restart your instrument. It also
contains files to check and format the hard disk.
Follow these steps to create the emergency startup disk:
1. Minimize the instrument application by selecting Minimize in the File
menu.
2. Click the Windows Start button, point to Programs, Accessories, System
Tools, and click Backup.
Installation
Backing Up User Files
3. On the Tools menu, click Create an Emergency Repair Disk.
4. Insert a floppy disk into the disk drive and follow the on-screen instructions
to create the startup disk.
You should always back up your user files on a regular basis. Use the Back Up
tool to back up files stored on the hard disk. The Back Up tool is located in the
System Tools folder in the Accessories folder.
1. Minimize the instrument application by selecting Minimize in the File
menu.
2. Click the Windows Start button.
3. Select Programs, Accessories, System Tools, Backup in the Start menu.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 11
Installation
Installing Software
4. Use the backup tool that displays to select your backup media and to select
the files and folders that you want to back up. Use the Windows online help
for information on using the Backup tool. You can back up to the floppy
drive or to a third-party storage device over the printer port (rear panel).
The instrument system and application software is preinstalled at the factory. If
you have to reinstall the software for any reason, refer to the instructions that
accompany the CDs that are shipped with the instrument. If you need to restore
the operating system, you also need the Windows licence information from the
Certificate of Authenticity that is shipped with the instrument.
Software Release Notes
Accessory Software
Read the software release notes README.TXT ASCII file on the product-software CD before performing installation procedures. This file contains additional
installation and operation information that supercedes other product documentation.
To view the README.TXT file, open the Notepad Windows accessory. Then
open the file on the Product Software CD.
The Product Software CD also contains accessory software and files that you can
choose to install in the instrument or in another computer. Refer to the instructions that accompany the CD for installation information.
GPIB Programmer Online Help Software. You can install the GPIB Programmer
online help on the instrument, but it may be more convenient to install it on the
PC that is functioning as the GPIB system controller. From the system controller,
you can copy and paste commands from the help directly into your test programs. The programmer information contains the following content:
HGPIB configuration information for the instrument
HLists of the command groups and the commands they contain
1- 12
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Installation
HDetailed command descriptions including syntax and examples
HStatus and error messages
HProgramming examples
The CD also contains a printable version of the programmer information in the
form of a PDF file.
Semi-Automated Performance Verification Procedure. This software (TDS7104,
TDS7054, and TDS6000 Series Only) provides a semiautomated method to
verify the oscilloscope performance. The installer installs the software and a PDF
file that provides instructions to perform the procedure. The PDF file also lists
the specific test equipment required to perform the procedure.
You should not install this software on the oscilloscope, but rather on the PC you
plan to use as a GPIB controller. The GPIB controller must be equipped with a
National Instruments GPIB Controller card and software.
Manual Performance Verification Procedure. This is a printable PDF file that
describes how to verify the instrument performance using generic test equipment.
User manual. This is a PDF file of this user manual.
Serial Mask User Manual. This is a PDF file that describes how to use the serial
mask features of the instrument.
Optional Accessory Software. The Optional Applications Software CD contains
programs that you can install and run five times per application. You can then
purchase an upgrade from Tektronix if you decide that you want to continue to
use the application. Refer to the instructions that accompany the CD for
installation information.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 13
Installation
Desktop Applications
You can install desktop application software on the instrument. The instrument
has been tested with the following software products installed:
HMicrosoft Office 2000 (including Word, Excel, Powerpoint, and Access)
HMathCad
HMATLAB
Other software products may be compatible but have not been tested by
Tektronix. If the instrument malfunctions after you install software, you should
uninstall the software, and then reinstall the instrument application to restore
proper operation.
Exiting the Instrument Application. Before installing other desktop applications,
you should exit the instrument application. Follow these steps to exit the
instrument application:
NOTE. If you are not using a USB keyboard and mouse, you must power on the
instrument after attaching your keyboard and mouse.
1. Connect a keyboard and mouse to the instrument.
Options
2. While holding down the CTRL and ALT keys, press the DELETE key.
3. Select Task Manager.
4. In the Applications tab, select TekScope.exe, and then select End Process to
stop the instrument application.
The instrument application will restart after you restart the entire system,
following the installation of the desktop application software.
Some options contain software that must be installed and/or enabled. To do the
installation, follow the specific instructions that come with the option.
Tektronix provides a key that you must enter (one time) to enable all the options
that you have purchased for your instrument. To enter the key, select OptionInstallation in the Utilities menu, and then follow the on-screen instructions.
1- 14
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Enabling Your LAN and Connecting to a Network
You can connect the instrument to a network to enable printing, file sharing,
internet access, and other communications functions. Before you make the
connection, do the following steps to enable network access to the instrument:
1
Power
down
Installation
Rear panelFront panel
3
Connect a keyboard
and mouse
2
Power
on
Figure 1- 3: Enabling your LAN and connecting to a network
4. As the instrument begins to boot, press the F2 key on the keyboard
repeatedly until the message “Entering SETUP” (“Loading SETUP” on
some instruments) appears.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 15
Installation
5. In the BIOS Setup Utility, use the right-arrow key on the keyboard to
highlight the Advanced menu at the top of the screen.
6. Use the arrow down key to highlight PCI Configuration (Peripheral
Configuration on some instruments) in the Advanced screen, and then press
Enter.
7. Use the arrow down key to highlight Embedded Ethernet Controller (LAN
Device on some instruments) in the Peripheral Configuration screen, and
then press Enter.
8. Use the arrow up or down key to highlight Enabled, and then press Enter.
9. Press the F10 key to save and exit. Confirm the Save of Configuration
changes when you are prompted on screen.
10. Use the Windows network setup utility to define the instrument as a network
client and configure it for your network. You can find the network setup
utility in the Windows Start menu if you select Settings > Control Panel
and then double click Network. You should consult your network administrator for specific instructions to make these settings.
NOTE. If you want to disable network access for the instrument, perform the
above procedure except substitute Disabled for the command listed in step 8. The
instrument will boot faster with network access disabled.
1- 16
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Setting up a Dual Display
Use the following steps to set up the instrument for dual display operation. You
can operate the instrument while having full use of Windows and other applications on the external monitor.
1
Use the On/Standby switch to power down.
2
Connect a
keyboard and
mouse.
3
Connect an
external monitor.
Installation
4
Power on.
Figure 1- 4: Setting up a dual display
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
5
Power on.
1- 17
Installation
6. Watch for a message on the external monitor telling you that Windows has
successfully initialized the display adapter.
7. The instrument should detect that the new monitor was connected. Follow
the instructions on the instrument display to install new drivers for the
monitor.
8. Type a Control-M to minimize the instrument application.
9. In the Windows desktop, right-click the mouse, and then select Properties to
display the Display Properties dialog box.
10. Select the Settings tab, and click the grayed-out monitor in the display box.
11. Click yes when you are prompted to enable the new monitor.
12. Set the resolution that you want to use on the external monitor.
13. Click on the external monitor in the display box, and drag it to the correct
orientation.
CAUTION. Do not change the resolution or color settings for the internal LCD
monitor. The internal resolution must be 640 x 480 and the color setting must be
High Color (24 bit).
14. Click OK to apply the settings. The new monitor will display additional
desktop area.
To make the best use of the new display area, do these additional steps to move
the Windows controls to the external monitor:
1. Click (and hold) on the Windows task bar in the area shown in Figure 1--5,
and then drag it upwards and toward the external monitor. The task bar will
first go to the side of the internal monitor, then to the side of the external
monitor, and finally to the bottom of the external monitor.
1- 18
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Figure 1- 5: Drag area for Windows task bar
2. Release the mouse when the task bar is where you want it to be.
Installation
Click here to drag task bar.
Select all
3
Internal monitor
Drag
Drop
External monitor
Figure 1- 6: Moving Windows desktop icons to the external monitor
4. If you use the instrument help system, you can drag the help windows to the
external monitor so that you can read them while you operate the instrument.
5. When you open any Windows application, drag the windows from the
application to the external monitor.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 19
Installation
1- 20
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Incoming Inspection
This chapter contains instructions for performing the Incoming Inspection
Procedure. This procedure verifies that the instrument is operating correctly after
shipment, but does not check product specifications. This procedure contains the
following parts:
HSelf Tests on page 1--22 provides instructions for performing the internal
self tests.
HFunctional Tests on page 1--23 measures the time- and amplitude-reference
signals at the PROBE COMPENSATION connector.
HPerform the Extended Diagnostics on page 1--35 provides instructions for
performing internal self calibration and the extended diagnostics.
If the instrument fails any test within this section, it may need service. To contact
Tektronix for service, see Contacting Tektronix on page xvii.
Make sure you have put the instrument into service as detailed in Installation
starting on page 1--5. Then assemble the following test equipment and proceed
with the procedures that follow.
Assemble Equipment
Self tests do not require any test equipment. The functional tests require the
following test equipment:
HOne BNC cable, such as Tektronix part number 012-0076-xx
The functional tests for the CSA7000 Series, TDS6000 Series, TDS7404,
TDS7254, and TDS7154 instrument require the following additional test
equipment:
HA P7240 probe (P7260 probe with TDS6604)
HA probe calibration and deskew fixture, Tektronix part number 067-0405-xx
(067-0848-xx for TDS6604)
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 21
Incoming Inspection
Self Tests
HOne TCA-BNC TekConnect adapter, or one TCA-SMA TekConnect adapter
and one SMA male-to-BNC female adapter, such as Tektronix part number
015-1018-xx
This procedure uses internal routines to verify that the instrument functions and
was adjusted properly. No test equipment or hookups are required.
Equipment
required
PrerequisitesPower on the instrument and allow a 20 minute warm-up before doing
1. Verify that internal diagnostics pass: Do the following substeps to verify
passing of internal diagnostics.
a. Display the System diagnostics menu:
HIf the instrument is in tool-bar mode, click the MENU button to put
the instrument into menu bar mode.
HFrom the Utilities menu, select Instrument Diagnostics . . . . This
displays the diagnostics control window.
b. Run the System Diagnostics:
HFirst disconnect any input signals and probes from all four channels.
HClick the Run button in the diagnostics control window.
c.Wait: The internal diagnostics do an exhaustive verification of proper
instrument function. This verification will take five to fifteen minutes.
When the verification is finished, the resulting status will appear in the
diagnostics control window.
None
this procedure.
1- 22
d. Verify that no failures are found and reported on-screen. All tests should
pass.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Functional Tests
Incoming Inspection
e.Run the signal-path compensation routine:
HFrom the Utilities menu, select Instrument Calibration . . . . This
displays the instrument calibration control window.
HIf required because the instrument is in service mode, select the
Signal Path button under Calibration Area.
HTouch the Calibrate button to start the routine.
f.Wait: Signal-path compensation may take five to ten minutes to run.
g. Confirm signal-path compensation returns passed status: Verify that the
word Pass appears in the instrument calibration control window.
2. Return to regular service: C lick the Close button to exit the instrument
calibration control window.
The purpose of these procedures is to confirm that the instrument functions
properly. A list of required test equipment is shown on page 1--21.
NOTE. These procedures verify that the instrument features operate. They do not
verify that they operate within limits.
Therefore, when the instructions in the functional tests that follow call for you to
verify that a signal appears on-screen “that is about five divisions in amplitude”
or “has a period of about six horizontal divisions,” and so forth, do not
interpret the quantities given as limits.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 23
Incoming Inspection
NOTE. Do not make changes to the front-panel settings that are not called out in
the procedures. Each verification procedure will require you to set the instrument
to certain default settings before verifying functions. If you make changes to
these settings, other than those called out in the procedure, you may obtain
invalid results. In this case, redo the procedure from step 1.
When you are instructed to push a front-panel button or screen button, the button
may already be selected (its label will be highlighted). If this is the case, it is not
necessary to push the button.
1. Initialize the instrument: Push the front-panel DEFAULT SETUP button.
2. Hook up the signal source: Connect the equipment as shown in Figure 1--7
to the channel input you want to test (beginning with CH 1).
3. Turn off all channels: If any of the front-panel channel buttons are lighted,
push those buttons to turn off the displayed channels. See Figure 1--8.
4. Select the channel to test: Push the channel button for the channel you are
currently testing. The button lights and the channel display comes on.
1- 24
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Incoming Inspection
TDS7104 & TDS7054
BNC cable from the
PROBE COMPENSATION
output to the GAIN CAL
SIG input on the fixture.
BNC cable from PROBE
COMPENSATION output
to channel input
TDS6604
BNC cable from PROBE
COMPENSATION output to
the A input of the Probe
Calibration and Deskew
Fixture
CSA7404, CSA7154, TDS7404,
TDS7254, TDS7154, & TDS6404
A
Remove
the jumper
Connect the probe tip to
the short pin and the
probe ground to the long
pin as shown.
NOTE. If a P7240 probe is
not available, connect the
probe compensation output
to the channel input using a
BNC cable and adapters.
GAIN CAL connections
Figure 1- 7: Universal test hookup for functional tests - CH 1 shown
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 25
Incoming Inspection
TDS
&
Channel buttons
Figure 1- 8: Channel button location
5. Set up the instrument:
HPush the front-panel AUTOSET button. This sets the horizontal and
vertical scale and vertical offset for a usable display and sets the trigger
source to the channel that you are testing.
HTouch the Vert button and then touch Offset. Confirm that the Ch1
Offset is 1.8 V (0.0 V if not using a probe).
6. Verify that the channel is operational: Confirm that the following statements
are true.
HVerify that the vertical scale readout and the waveform amplitude for the
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Incoming Inspection
HThe front-panel vertical POSITION knob (for the channel you are
testing) moves the signal up and down the screen when rotated.
HTurning the vertical SCALE knob counterclockwise (for the channel you
are testing) decreases the amplitude of the waveform on-screen, turning
the knob clockwise increases the amplitude, and returning the knob to
the original scale setting returns the amplitude to that shown in
Table 1--3 for that scale setting.
7. Verify that the channel acquires in all acquisition modes: From the
TDS6404:
Touch the Vert button and then touch Offset. Adjust the Ch1 Offset to 0.8 V
using the multipurpose knob.
5. CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, &
TDS6404:
Set the Vertical SCALE to 100 mV per division.
6. Set the time base: Set the horizontal SCALE to 200 s/div. The time-base
readout is displayed at the bottom of the graticule.
1- 28
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Incoming Inspection
7. Verify that the time base operates: Confirm the following statements.
HOne period of the square-wave probe-compensation signal is about five
horizontal divisions on-screen for the 200 s/div horizontal scale setting.
HRotating the horizontal SCALE knob clockwise expands the waveform
on-screen (more horizontal divisions per waveform period), counterclockwise rotation contracts it, and returning the horizontal scale to
200 s/div returns the period to about five divisions.
HThe horizontal POSITION knob positions the signal left and right
on-screen when rotated.
8. Verify horizontal delay:
a. Center a rising edge on screen:
HSet the horizontal POSITION knob so that the rising edge where the
waveform is triggered is lined up with the center horizontal
graticule.
HChange the horizontal SCALE to 20 s/div. The rising edge of the
waveform should remain near the center graticule and the falling
edge should be off screen.
b. Turn on and set horizontal delay:
HFrom the Horiz/Acq menu, select Horizontal/Acquisition
Setup....
HClick the Horizontal tab in the control window that displays.
HClick the Delay Mode button to turn delay on.
HDouble click the Horiz Delay control in the control window to
display the pop-up keypad. Click the keypad buttons to set the
horizontal delay to 500 s, and then click the ENTER key.
c.Verify the waveform: Verify that a falling edge of the waveform is within
a few divisions of center screen.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 29
Incoming Inspection
d. Adjust the horizontal delay: Rotate the upper multipurpose knob to
change the horizontal delay setting. Verify that the falling edge shifts
horizontally. Rotate the front-panel horizontal POSITION knob. Verify
that this knob has the same effect (it also adjusts delay, but only when
delay mode is on).
e.Verify the delay toggle function:
HRotate the front-panel horizontal POSITION knob to center the
falling edge horizontally on the screen.
HChange the horizontal SCALE to 40 ns/div. The falling edge of the
waveform should remain near the center graticule. If not, readjust the
delay setting to center the falling edge.
HPush the front-panel DELAY button several times to toggle delay off
and on and back off again. Verify that the display switches quickly
between two different points in time (the rising and falling edges of
this signal).
Check Trigger Operation
9. Remove the test hookup: Disconnect the equipment from the channel input
and the probe compensation output.
1. Initialize the instrument: Push the front-panel DEFAULT SETUP button.
2. Hook up the signal source: Connect the equipment to the CH 1 input as
shown in Figure 1--10.
3. Set up the instrument: Push the front-panel AUTOSET button.
1- 30
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Incoming Inspection
TDS7104 & TDS7054
BNC cable from PROBE
COMPENSATION
output to CH 1 input
CSA7404, CSA7154, TDS7404, TDS7254,
TDS7154, TDS6604, & TDS6404
BNC cable from PROBE
COMPENSATION output
to CH 1 input
Figure 1- 10: Setup for trigger test
4. CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, &
TDS6404:
Touch the Vert button and then touch Offset. Adjust the Ch1 Offset to 0.8 V
using the multipurpose knob.
5. CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, &
TDS6404:
Set the Vertical SCALE to 100 mV per division.
6. Verify that the main trigger system operates: Confirm that the following
statements are true.
HThe trigger level readout for the A (main) trigger system changes with
the trigger-LEVEL knob.
HThe trigger-LEVEL knob can trigger and untrigger the square-wave
signal as you rotate it. (Leave the signal untriggered).
HPushing the front-panel trigger LEVEL knob sets the trigger level to the
50% amplitude point of the signal and triggers the signal that you just
left untriggered. (Leave the signal triggered.)
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 31
Incoming Inspection
7. Verify that the delayed trigger system operates:
a. Set up the delayed trigger:
HFrom the Trig menu, select A→B Sequence . . . . This displays the
A→B Sequence tab of the trigger setup control window.
HClick the Trig After Time button under A Then B.
HClick the B Trig Level control in the control window.
b. Confirm that the following statements are true:
HThe trigger-level readout for the B trigger system changes as you
turn the lower multipurpose knob.
HAs you rotate the lower multipurpose knob, the square-wave
probe-compensation signal can become triggered and untriggered.
(Leave the signal triggered.)
c.Verify the delayed trigger counter:
Check File System
HDouble click the Trig Delay control to pop up a numeric keypad for
that control.
HClick on the keypad to enter a trigger delay time of 1 second,and
then click Enter.
HVerify that the trigger READY indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
8. Remove the test hookup: Disconnect the equipment from the channel input
and the Probe Compensation output.
Equipment
required
PrerequisitesNone
One BNC cable
One 1.44 Mbyte, 3.5 inch DOS-compatible formatted disk.
TDS6404:
Touch the Vert button and then touch Offset. Adjust the Ch1 Offset to 0.8 V
using the multipurpose knob.
6. CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, &
TDS6404:
Set the Vertical SCALE to 100 mV per division.
7. Set the time base: Set the horizontal SCALE to 1ms/div. The time-base
readout is displayed at the bottom of the graticule.
8. Save the settings:
a. Pull down the File menu to select Instrument Setup . . . . This displays
the instrument setups control window.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 33
Incoming Inspection
b. Click the Save button under Save settings to file in the control window.
This displays a familiar Windows dialog box for choosing a destination
directory naming the file.
c.In the Save Instrument Setup As dialog box, select the 3
1
/2Floppy (A:)
icon in the Save in: drop-down list to set the save destination to the
floppy disk.
d. Note the default file name, and then click the Save buttontosavethe
setup to the default file name.
9. Change the settings again: Set the horizontal SCALE to 200 s/div.
10. Verify the file system works:
a. Click the Recall Setups tab in the control window.
b. Click the Recall button under Recall settings from file in the control
window. This displays a familiar Windows dialog box for locating the
settings file that you want to recall.
1
c.In the Recall Instrument Setup dialog box, select the 3
/2Floppy (A:)
icon in the Look in: drop-down list.
d. Locate and double click in the dialog box on the setup file that you
previously stored.
e.Verify that the instrument retrieved the saved setup from the disk. Do
this by noticing the horizontal SCALE is again 1 ms and the waveform
shows ten cycles just as it did when you saved the setup.
1- 34
11. Remove the test hookup:
a. Disconnect the equipment from the channel input and the probe
compensation output.
b. Remove the floppy disk from the floppy disk drive.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Perform the Extended Diagnostics
Extended diagnostics and self calibration perform a more detailed functionality
check than the incoming inspection and Power-on diagnostics.
NOTE. Allow a 20-minute warm-up before running the self calibration.
Disconnect any attached probes from the instrument. Then select the Utilities
menu. Run the self calibration followed by the extended diagnostics by first
selecting the Instrument Calibration (see page 3-- 171 for additional information
on signal path compensation) and then the Instrument Diagnostics tabs. Results
of the tests display on their property pages.
Incoming Inspection
Checking the Underlying
System (Optional)
To check the hardware and Windows software underlying the instrument UI (user
interface), run the CheckIt Utilities from the Windows Start menu:
1. Exit the instrument application before running the external diagnostics.
a. Display the Task Manager by pressing CTRL, ALT,andDelete,and
then touch Task Manager.
b. Select TekScope and then touch End Task.
2. Click Start, then click Programs in the Start Menu. Finally, click CheckIt
Utilities.
3. From the Go To menu, select and run the tests you want to perform.
4. Check test results. All tests except the Modem and CD-ROM tests should
pass. The CD-ROM test requires data from the CheckIt Utilities CD.
5. Dismiss the CheckIt Utilities: Select Exit in the File menu.
6. Restart your instrument UI software: Click Start, then click Programs in the
Start Menu. Finally, click Tektronix Oscilloscope and then TekScope.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 35
Incoming Inspection
1- 36
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Accessories & Options
This section lists the standard and optional accessories available for the
instrument, as well as the product options.
Options
The following options can be ordered for the instrument:
Table 1- 4: Options
CSA7154
CSA7404
OptionDescription
1KK4000 Instrument Cartnnnnnnnnn
1RRack Mount Kit (includes: hardware and instructions for converting to
rackmount configuration)
Acquisition memory
2MLong record length: 2 Ms per channel, 8 Ms maximumnnnnnnn
3MLong record length: 4 Ms per channel, 16 Ms maximumnnnnnnn
4MLong record length: 8 M per channel, 32 Ms maximumnnnnn
Probes
30No probesnn
33Add one P6158, 3 GHz, low capacitance, 20x divider probenn
34Add one P6247, 1 GHz differential probenn
35Add one P6243, 1 GHz active probenn
36Add one P6139A, 500 MHz passive probenn
37Add one P6245, 1 GHz Active Probennnn
39Add one P6248, 1.7 GHz Differential Probennnn
51Add one P7240, 4 GHz, active probennn
52Add one P7330, 3 GHz, differential probennnnn
nnnnnnnnn
TDS7404
TDS7154
TDS7254
TDS7104
TDS7054
TDS6604
TDS6404
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 37
Accessories & Options
Table 1- 4: Options (Cont.)
CSA7154
CSA7404
TDS7254
TDS7404
TDS7154
TDS7104
TDS6604
TDS7054
TDS6404
Option
Power cords
A1Universal European 230 V, 50 Hz power cordnnnnnnnnn
A2United Kingdom 230 V, 50 Hz power cordnnnnnnnnn
A3Australian 230 V, 50 Hz power cordnnnnnnnnn
A5Switzerland 230 V, 50 Hz power cordnnnnnnnnn
ACChina 230 V, 50 Hz power cordnnnnnnnnn
A99No power cordnnnnnnnnn
Software applications
J1TDSJIT2 Jitter Analysis applicationnnnnnnnnn
J2TDSDDM2 Disk Drive Measurements applicationnnnnnnn
JT3TDSJIT3 Advanced Jitter Analysis application with random and
CP2TDSCPM2 Compliance Testing for ITU-T G.703 and ANSI T1.102
USBUSB2 Compliance Testing Software for USB1.0/USB2.0 standards
SMSerial Communications Mask Testing (Standard on CSA7000 Series)nnnnnnn
STSerial Pattern Triggering (Standard on CSA7000 Series)nnnnnnn
The accessories in Table 1--6 are orderable for use with the instrument at the time
this manual was originally published. Consult a current Tektronix catalog for
additions, changes, and details.
NOTE. The P6339A probe is not supported by this instrument.
1- 44
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Operational Maps
This chapter acquaints you with how the instrument functions and operates. It
consists of several maps that describe the system, its operation, and its documentation:
HDocumentation Map, on page 2--2, lists the documentation that supports the
HSystem Overview Maps on page 2--4, describe the high-level operating blocks
HUser-Interface Map, on page 2--7, describes the elements of the User
HFront-Panel Map, on page 2--8, describes the elements of the instrument
HDisplay Map, on page 2--9, describes elements and operation of single-grati-
HFront Panel I/O Map, on page 2--10, describes inputs, outputs, and
instrument.
and operating cycle of the instrument.
Interface (UI) application, which provides complete control of the instrument.
front panel and cross references information relevant to each element.
cule and multiple-graticule displays.
peripherals.
HRear Panel I/O Map, on page 2--11, describes input/output ports and
peripherals.
Tutorial (How to?) procedures are available online, as part of the online help.
For information on configuring and installing your instrument, refer to Chapter 1, Getting Started.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
2- 1
Documentation Map
This instrument ships with documents individually tailored to address different
aspects or parts of the product features and interface. The table below cross
references each document to the instrument features and interfaces it supports.
To read about…Refer to these documents:Description
Read the Reference for a quick overview of
instrument features and their usage.
Read the User Manual for general information about your instrument—procedures on
how to put it into service, specifications of
its performance, maps of its user interface
controls, overviews and background on its
features.
For more usage information, see OnlineHelp System below.
Access online help from the instrument for
context-sensitive information on virtually all
controls and elements on screen.
Online help includes procedures for using
instrument functions. See Accessing OnlineHelp on page 3--283.
Quickly find the syntax of a command, and
copy the command if desired. Read about
communication, error handling, and other
information on GPIB usage. This guide is
on the product software CD.
Analysis and Connectivity Tools
2- 2
Oscilloscope Analysis and Connectivity
Made Easy
TekVISA Programming Manual
VXIplug&play Driver Help
TekVISA Excel Toolbar Help
These documents consists of various
connectivity and analysis tools that you can
install and configure for your instrument.
For more information, see the Analysis andConnectivity Support topic in the instrument
online help.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Documentation Map
You may also want to obtain the optional service manual for this product if you
self-service or performance test this instrument. S ee Accessories & Options on
page 1--37.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
2- 3
System Overview Maps
The instrument is a highly capable waveform acquisition, test, and measurement
system. The following model provides background information on its operation,
which, in turn, may provide you insight on how the instrument can be used.
Functional Model Map
CH 1-- 4
Recovered clock output
Recovered data output
Electrical output
Optical input
Digital Signal Acquisition
Input channels
Acquisition
Amplitude
scaling
page 3--20
Trigger
system
page 3--71
Optical-to-electrical
converter
CSA7000 Series
page 3--26
system
Timebase
system
pages
3--124,
3--185
Signal Processing
& T ransformation
DSP
pages
3--147,
3--185,
Output &
Storage
CH 1-- 4
Ref 1--4
Math 1 --4
Display & UI
Page 3--123
2- 4
The model comprises four high-level subsystems or processes (embodying a
variety of hardware and software functions) and the data that connects them:
HDigital Signal Acquisition System. Acquires a waveform record from each
signal that you input to each channel using the following subsystems:
HInput Channels. Conditions the input signal, primarily through the use
of analog hardware, before the signal is converted to digital form.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
System Overview Maps
HTrigger System. Recognizes a specific event of interest on the input trigger
signal and informs the Timebase of the occurrence of the event. Also
provides recovered clock and data signals (optional on TDS 7000 Seriesand
TDS6000 Series instruments).
HTimebase System. Tells the Acquisition system to start an acquisition
cycle (that is, to convert from analog to digital). In more general terms,
synchronizes the capturing of digital samples in the Acquisition system
to the trigger events generated from the Trigger system.
HAcquisition System. Performs the actual A/D conversion and storing of
HDSP Transformation System. Performs a variety of transformations or
operations, beginning with the most fundamental data element(s) in the
system, the Channel Waveform(s). Waveform math operations, automatic
measurements, spectral waveforms, and histogram generation are examples.
HInput/Output Systems. Provides output (and sometimes input) of instru-
ment-data elements in a form suitable to the user and also provides user
input control.
The process overview that follows describes each step in the top-level cycle of
instrument operation.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
2- 5
System Overview Maps
Yes
Process Overview Map
Process Overview
Yes
Stop condition?
No
Implement
pretrigger points
No
Idling...
setup
Acquire
Trigger
accepted
Yes
Reset
Abort
Power on
Power down
Arm
Process Block Description
1.The instrument starts in the idle state; it enters this state
upon power up, upon receiving most control setting changes,
or upon finishing acquisition tasks.
2.Control settings are implemented as they are requested.
When you toggle the RUN/STOP control to RUN, the
instrument starts the hardware.
3.The instrument acquires samples until the pretrigger portion
of the waveform record (channel) being acquired is filled.
4.The instrument then begins waiting for a trigger. Acquiring
continues to take place, keeping the pretrigger points current,
until triggering criteria are met or a trigger is forced (Auto
trigger mode only) and the instrument accepts a trigger.
2- 6
No
Acquire
posttrigger
points
5.The instrument acquires samples until the posttrigger portion
of the waveform record (channel) being acquired is filled.
6.If averaging, enveloping, or waveform database is on, the
record becomes part of the multi-acquisition record that these
Waveform
record
complete?
Yes
modes produce. The process loops back to step 3 above to
acquire additional records until the number of acquisitions
required for the acquisition mode currently set are processed,
and then processing continues to step 7 below.
7.At this point the acquisition record is in DSP memory and is
Waveform
available
available to the instrument for measurement of its
parameters, display, and so on.
The instrument then checks for a user-specified stop
condition and either returns to its idle state or continues at
step 3, according to what it finds.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
User Interface Map -- Complete Control and Display
Menu Bar: Access to data I/O,
printing, online help system, and
instrument functions here
Status Bar: Display of
acquisition status, mode, and
number of acquisitions; trigger
status; warnings; date; and time
Buttons/Menu. Touch to
toggle between tool-bar
and menu-bar modes
Multipurpose Knob
Readouts. Adjust
parameters controlled by
Display: Live, reference, & math
waveforms display here, along
with cursors
Waveform Handle: Touch and drag
to change vertical position of
waveform. Touch the handle and
change the position and scale using
the multipurpose knobs
Controls Status: Quick reference to
vertical, horizontal, and trigger
selections, scale, and parameters
multipurpose knobs
Readouts: Display
cursor readouts and
measurement
readouts in this area,
selectable from the
menu bar or tool bar
If a control menu is
displayed, these
readouts move to the
graticule area
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
2- 7
Front-Panel Map -- Quick Access to Most Often Used Features
Use these buttons to start and stop acquisition or start a single
acquisition sequence. The ARM, READY, and TRIG’D lights
Turn knob to adjust waveform intensity. Page 3--53.
Push button to turn Fast Acquisition on or off (7000 Series only).
Use these knobs and buttons to set the trigger parameters. Push
ADVANCED to display additional trigger functions. Pages 3--71
Push to return settings to default values. Page 3 --18.
Horizontally scale, position, delay, and set record length
show the acquisition status. Page 3 --82.
Page 3--47.
and 3--88.
Push to turn cursors on or off. Page 3 --160.
Push to make a hard copy. Page 3--277.
(resolution) of selected waveform. Page 3--23.
Push to automatically set up the vertical, horizontal, and trigger
Turn the multipurpose knobs to adjust parameters selected from
Push ZOOM to add a magnified graticule to the display. Push
Turn channel displays on and off. Vertically scale or position the
waveform. Change the input termination (TDS7104 & TDS7054).
controls based on selected channels. Page 3--17.
the screen interface. Push a Fine button to toggle between
normal and fine adjustment with its multipurpose knob.
Page 3--165.
Push to turn the touch screen on and off. Page 3 --129.
HORIZ or VERT to select the axis to magnify. Page 3--131.
Page 3--8.
2- 8
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Display Map -- Single Graticule
Drag icon to change the trigger
level
Drag cursors to measure
waveforms on screen
Drag the position icons to
reposition a waveform
Click icon to assign multipurpose
knobs to waveform vertical
position and scale
Drag across the waveform area to
zoom the boxed waveform segment.
Also enable/disable histograms and
measurement gating
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
2- 9
Front Panel I/O Map
CSA7000 Series
Floppy disk drive
Probe compensation
output
Recovered clock output
Recovered data output
O/E converter electrical output
Probe compensation
output
Auxiliary trigger input
CH 3 SIGNAL OUTPUT; scale and
offset controlled by CH3 controls
Optical input
Channel
inputs
TDS7404, TDS7254, TDS7154, TDS6604, & TDS6404
Auxiliary trigger output
Channel
inputs
TDS7104 & TDS7054
Ground terminal
Ground terminal
Auxiliary trigger input
Auxiliary trigger output
CH 3 SIGNAL OUTPUT; scale and
offset controlled by CH3 controls
2- 10
Ground terminal
Probe
compensation
output
Channel
inputs
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Rear Panel I/O Map
Removable hard disk drive to provide
individual environment for each user or to
secure data. Press to release
CDROM-RW drive accessible from
Windows. Press to open
USB connector for mouse or
keyboard and mouse
PS-2 connectors for mouse and keyboard
Upper VGA port to connect a
monitor for side-by-side display
Lower VGA port to connect a
monitor for instrument display
Parallel port (Centronics) to
connect printer or other device
GPIB port to connect to controller
RJ-45 connector to connect to network
COM1 serial port
Connectors for speaker and microphone
CSA7000 Series: Auxiliary (external
trigger) input. Auxiliary (trigger) output
signal. Analog (CH 3) signal output; scale
and offset controlled by CH3 controls.
External reference input and internal
reference output
PCMCIAslotsfortwotype-1cards,
TDS7000 Series & TDS6000 Series:
twotype-2cards,oronetype-3card
External reference input and internal
reference output
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
2- 11
Rear Panel I/O Map
2- 12
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Overview
This chapter describes in depth how the many features of the instrument operate.
Please note the following points on using this chapter:
HEach section in this chapter provides background information needed to
operate your instrument effectively as well as the higher-level procedures for
accessing and using the features. These procedures emphasize using the front
panel when possible.
HLower-level, detailed usage procedures are in the online help system.
The table that follows on page 3--2 lists operating tasks and the sections in this
chapter that document those tasks.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 1
Overview
Tasks or topics
Signal InputAcquiring waveformsAcquiring WaveformsOverview of section contents3--7
Subtasks or subtopicsSection titleContentsPage no.
Signal Connection and
Conditioning
To Set Up Signal InputSetting up signal input3--13
To Autose t the
Instrument
To Reset the InstrumentResetting the instrument3--18
To Get More HelpAccessing the online help3--19
Input Conditioning
Background
Setting Acquisition
Controls
To Set Acquisition ModesSet up of the acquisition system3--34
To Start and Stop
Acquisition
To Set Roll ModeSetting up roll mode3--38
Acquisition Control
Background
Overview of signal connection and
conditioning techniques and setups
Automatic setup of the acquisition,
triggering systems and input channels
Background information on input signal
conditioning
Background and setup of acquisition
controls and roll mode
Starting and stopping acquisitions3--37
Background information on acquisition
controls
3--8
3--17
3--19
3--26
3--39
Using Fast Acquisition
Mode
To Turn Fast Acquisitions
On and Off
Using FastFrameOverview and setup of FastFrame3--56
To Set FastFrame ModeSetting up FastFrame mode3--58
Time Stamping FramesSetting up FastFrame time stamps3--60
O/E Converter
CSA7000 Series only
Front Panel Connectors
CSA7000 Series only
Wavelength, Filter, and
Bandwidth Selection
CSA7000 Series only
Optical bandwidth
CSA7000 Series only
Overview and setup of Fast Acquisition
mode
Turning fast acquisitions on and off3--50
Overview and reference to documentation
on using optical signals and the O/E
Converter
Overview of O/E Converter front-panel
connectors
Overview of optical controls3--67
Overview of optical bandwidth3--68
3--47
3--63
3--64
3- 2
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Overview
Tasks or topics
Signal InputTriggering waveformsTriggeringBackground on basic trigger operation3--71
Triggering ConceptsBackground on triggering concepts3--72
Triggering From the Front
Panel
Access ProceduresSetting up triggers with front-panel
To Check Trigger StatusChecking trigger status3--82
Additional Trigger
Parameters
Advanced TriggeringOverview and setup for triggering on
To Trigger on a GlitchSetting up glitch triggering3--94
To Trigger on a Runt PulseSetting up runt pulse triggering3--96
Trigger Based on Pulse
Width
To Trigger Based on
Transition Time
Trigger Based on Pulse
Timeout
Using the front-panel edge trigger controls 3--78
controls
Using additional trigger controls3--83
specific pulse and logic conditions
Setting up pulse width triggering3--99
Setting up transition time triggering3--101
Setting up pulse timeout triggering3--104
Page no.ContentsSection titleSubtasks or subtopics
3--78
3--88
Trigger on a PatternSetting up pattern triggering3--105
To Trigger on a StateSetting up state triggering3--109
To Trigger on Setup/Hold
Time Violations
Sequential TriggeringOverview and setup f or triggering after
To Trigger on a SequenceSetting up triggering on a sequence3--118
Comm Triggering
(optional on TDS7000
Series)
Serial Pattern Triggering
(optional on TDS7000
Series)
Setting up setup/hold time violation
triggering
time delay and triggering on events
Overview and reference to documentation
on using communications triggering
Overview and reference to documentarian
on using serial pattern triggering
3--110
3--113
3--122
3--122
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 3
Overview
Tasks or topics
Display FeaturesUsing the DisplayDisplaying WaveformsUsing display features and customizing
the display
Data Processing
(Calculation)
Using the Waveform
Display
To Display Waveforms in
theMainGraticule
Setting Zoom ControlsOverview and use of Zoom3--131
To Zoom WaveformsSetup to zoom a waveform3--132
Customizing the DisplayCustomizing display elements3--138
Set Display StylesOverview of display adjustments you can
Customize Graticule and
Waveforms
Taking MeasurementsMeasuring WaveformsSetup for measuring parameters3--147
Taking Automatic
Measurements
To Take Autom a t i c
Measurements
To Localize a
Measurement
Overview of display features and setup of
the display
Display adjustments you can make3--129
make
Overview of graticule and waveform
adjustments you can make
Overview and setup for taking automatic
measurements
Set for automatic measurements3--153
Setup to measure a segment of a
waveform
Page no.ContentsSection titleSubtasks or subtopics
3--123
3--124
3--141
3--144
3--148
3--159
Taking Cursor Measurements
To Set the Cursor SourcesSelecting the cursor sources3--165
Taking HistogramsOverview and setup for taking Histograms 3--168
To Start and Reset
Histogram Counting
Optimizing Measurement
Accuracy
To Compensate the
Instrument
Overview and setup for taking cursor
measurements
Setup for histogram counting3--169
Increase accuracy by compensating the
instrument and adjusting deskew
Setup to compensate the instrument3--172
3--160
3--171
3- 4
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Cos
Cosh, Tanh, Average,
Invert, Sqrt, Log 10, Log e,
Phase, and Imag)
Spectral Math operationsDefining Spectral Math
Sinh
To Connect the Probe
Calibration Fixture
To Calibrate ProbesImproving measurement accuracy by
To Compensate Passive
Probes
To Deskew ChannelsAdjusting delay between channels3--181
Serial Mask TestingPointer to mask testing information3--184
Creating and Using Math
Waveforms
Defining Math WaveformsOverview of performing math operation on
,
ToDefineaMath
Waveform
,
Operations on Math
Waveforms
To Use Math WaveformsExample setups for measuring and
Waveforms
Page no.ContentsSection titleSubtasks or subtopics
Connecting the probe calibration and
deskew fixture to the instrument
calibrating your probes
Compensating passive probes3--180
Functions for processing waveforms,
extracting segments of waveforms, etc.
waveforms
Defining math waveforms3--194
Overview of measuring and adjusting
math waveforms
adjusting math waveforms
Functions for frequency domain analysis
of your waveforms.
3--173
3--177
3--185
3--186
3--197
3--198
3--202
Optimizing
Measurement
Accuracy
Instrument compensation
and compensation or
calibration of probes
To Select a Predefined
Spectral Math Waveform
To Define A Spectral Math
Waveform
Spectral Math ExampleAn example of performing spectral
Optimizing Measurement
Accuracy
Selecting a predefined spectral math
waveform
Setup to define a spectral math waveform 3 --231
analysis of a signal
Procedures for compensating the instrument, compensating passive probes, and
calibrating active, voltage probes
3--230
3--239
3--171
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 5
Overview
,
Tasks or topics
Data Input and
Output
All uploading and downloading of waveforms,
calculation results, and
other data
Page no.ContentsSection titleSubtasks or subtopics
Data Input/OutputPrimary reference for uploading and
downloading of any data to or from the
instrument
Saving and Recalling a
Setup
To Save Your SetupSavi ng an instrument setup3--248
To Recall Your SetupRecalling an instrument setup3--251
Saving and Recalling
Waveforms
To Save Your WaveformSaving a waveform3--254
To Recall Your WaveformRecalling a saved waveform3--257
To Clear ReferencesSetup to clear references3--260
Exporting and Copying
Waveforms
To Export Your WaveformExporting a waveform to a file3--263
To Use an Exported
Waveform
To Copy Your WaveformCopying a waveform to the clipboard3 --273
Background and setup for saving and
recalling setups
Background and setup for saving and
recalling waveforms
Background and setup for exporting
waveforms and using exported waveforms
Using an exported waveform in Excel3--271
3--245
3--245
3--253
3--262
Printing WaveformsBackground and setup for printing
waveforms
To Date/Time Stamp
Hardcopies
Remote Communication
HelpUsing Online HelpAccessing Online HelpInformation on accessing the information
GPIBRemote CommunicationInformation on programming the
How to Use Online HelpUsing the online help3--284
Background and setup to display the date
and time
instrument over the GPIB
in the online help
3--277
3--281
3--282
3--283
3- 6
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Acquiring Waveforms
Before you can do anything (display, print, measure, analyze, or otherwise
process) to a waveform, you must acquire the signal. This instrument comes
equipped with the features that you need for capturing your waveforms before
further processing them according to your requirements. The following topics
cover capturing signals and digitizing them into waveform records:
HSignal Connection and Conditioning: How to connect waveforms to the
instrument channels; how to scale and position the channels and timebase for
acquiring waveforms.
HSetting Acquisition Controls: How to choose the appropriate acquisition mode
for acquiring your waveforms; how to start and stop acquisition.
HAcquisition Control Background: Background information on the data
sampling and acquisition process.
HUsing Fast Acquisition Mode (CSA7000 Series & TDS7000 Series): Using
Fast Acquisition mode to capture and display transient events, such as
glitches or runt pulses, often missed during longer dead times that accompany normal DSO operation.
HUsing FastFrame (CSA7000 Series & TDS7000 Series): Using FastFrame to
capture many records in a larger record, and then view and measure each
record individually.
HO/E Converter (CSA7000 Series): Using the Optical-to-Electrical Converter
to test optical signals.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 7
Acquiring Waveforms
Storage
Channel
inputs
Auxiliary
trigger input
InputDisplay
Trigger
Acquisition
system
Waveform
transform
system
Horizontal
time base
NOTE. This section describes how the vertical and horizontal controls define the
acquisition of live waveforms. These controls also define how all waveforms are
displayed, both live and derived waveforms (math waveforms, reference
waveforms, and so on). The sections that follow cover display-related usage:
HDisplaying Waveforms on page 3--123.
HCreating and Using Math Waveforms on page 3--185.
Signal Connection and Conditioning
This section presents overviews of the instrument features related to setting up
the input signal for digitizing and acquisition. It addresses the following topics:
HHow to turn on channels and adjust vertical scale, position, and offset
HHow to set horizontal scale, position, and access record-length and trigger-
position controls
HHow to get a basic trigger on your waveform
3- 8
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
NOTE. Terminology: This manual uses the terms vertical acquisition window and
horizontal acquisition window throughout this section and elsewhere. These
terms refer to the vertical and horizontal range of the segment of the input signal
that the acquisition system acquires. The terms do not refer to any windows or
display windows on screen.
Figure 3--1 shows the model for each input channel.
Figure 3- 1: Input and Acquisition Systems and Controls
Use input conditioning to ensure the instrument acquires the data that you want
to display, measure, or otherwise process. To ensure the best possible data for
displaying and further processing, do the following:
HSet the vertical scale to control the size of the vertical acquisition window
for each channel to capture part or all of the vertical amplitude of the input
signal. When vertical scaling is set to capture only a fraction of the input
signal range (for increased detail), the vertical offset control may be used to
determine which portion of the input signal is captured by the vertical
acquisition window.
50 Ω
K3
Vertica l
scale
+
+
Vertica l
position
To the
Σ
Bandwidth
limit
remainder
of the
acquisition
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 9
Acquiring Waveforms
HSet horizontal scale to control the size of the horizontal acquisition window
to capture as much as you want of the input signal(s). Set the horizontal
position to delay the window relative to a trigger and to control where in the
input signal (data stream) that the horizontal acquisition window acquires.
For more background on acquisition window concepts, see Input ConditioningBackground on page 3--19.
The instrument can automatically obtain and display a stable waveform of usable
size. Pushing the Autoset button automatically sets up the instrument controls
based on the characteristics of the input signal. Autoset is much faster and easier
than a manual control-by-control setup.
The instrument can also be reset to its factory default settings.
Usage of some input conditioning controls or features may be limited when other
control settings are in effect. Voltage offset is incompatible with reference
waveforms because offset is an acquisition control.
Connecting and
Conditioning Your Signals
Read the following topics related to waveform acquisition; they provide details
that can make it easier to set up and acquire your waveforms.
Probes and Signal Connection. Select the probe or cable that brings the signal into
the instrument. Choose the probe or cable that best fits your acquisition task,
whether it is connecting an active probe to test a digital circuit, or connecting to
a test fixture through SMA cables to characterize a device. The connection to the
instrument depends on your application.
Tektronix provides a variety of probes and cables for this product. For a list of
probes available for use, see Accessories and Options on page 1--37. You can
also check your Tektronix catalog for connection accessories that may support
your application. More information about your probes can be found in the user
manual for your probes.
Four acquisition channels are available. Each channel can be displayed as a
waveform or can contribute waveform data to other waveforms (math and
reference waveforms for example).
3- 10
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Acquiring Waveforms
Coupling. All instruments and probes specify a maximum signal level. (See
Specifications in your user manuals for exact limits.) Exceeding the limit, even
momentarily, may damage the input channel. Use external attenuators, if
necessary, to prevent exceeding the limits.
Coupling determines whether an input signal is directly connected to the input
channel, connected through a DC blocking capacitor (TDS7104 & TDS7054), or
not connected at all. These three choices are refereed to as DC coupling, AC
coupling, and GND coupling.
The input resistance of each input channel is either 50 Ω or you can select 50 Ω
or 1 MΩ, depending on the model. To properly terminate signals in other
impedance environments, use an adapter.
All probes expect a specific coupling and input termination. Both coupling and
input termination resistance are displayed on screen.
Scaling and Positioning. These key controls determine the portion of the input
signal presented to the acquisition system:
HSet vertical scaling, positioning, and DC offsets to display the features of
interest on your waveform and avoid clipping. (See Note that follows.)
Vertical Acquisition Window Considerations on page 3--20 describes the
vertical acquisition window.
Clipped
Acquired waveformDisplayed waveform
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 11
Acquiring Waveforms
HSet horizontal scale, position, and resolution (record length) so that the
acquired waveform record includes the waveform attributes of interest with
good sampling density on the waveform. The settings that you make define
the horizontal acquisition window (see Horizontal Acquisition WindowConsiderations on page 3--23).
NOTE. Waveform data outside the vertical acquisition window is clipped; that is,
the data is limited to the minimum or maximum boundaries of the vertical
acquisition window. This limiting causes inaccuracies in amplitude-related
measurements. Note that the acquisition window also includes 1 division above
and below the displayed graticule area. See Vertical Acquisition WindowConsiderations on page 3--20.
Trigger and Display. Set basic trigger controls to gate waveform acquisition, and
use the display to interactively scale, position, and offset waveforms. See the
sections Triggering on page 3--71 and Displaying Waveforms on page 3--123.
Flexible Control Access. This manual focuses on basic setup through the front
panel, and then through use of the User Interface (UI) Application displayed on
screen. The online help system also documents the UI.
3- 12
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Acquiring Waveforms
To Set Up Signal Input
Use the procedure that follows when setting up the instrument to scale and
position input signals for acquisition. For more information, display online help
while performing the procedure.
OverviewTo set up si gnal inputRelated control elements and resources
Prerequisites
Connect input
signal
1.The acquisition system should be set to run
continuously.
See page 3--34 for acquisition setup and
page 3--71 for trigger setup.
2.Connect to the signal to be acquired using proper
probing and connecting techniques.
Note. For more details on control over input setup, push
the Vert button to display the Vertical control window,
and then touch the HELP button.
Select the
input signal
channel
3.Push a channel button (CH 1--CH 4) to select the
signal channel.
A channel button lights when its channel is on.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 13
Acquiring Waveforms
OverviewRelated control elements and resourcesTo set up signal input (Cont.)
Select input
termination
Select input
coupling
TDS7104 and TDS7054 only:
4.Push an input termination button to toggle between
1MΩ and 50 Ω input termination.
Hint. Some probes force the instrument to set the
termination that the probe requires.
5.Touch Vert to display the Vertical control window. To
change the input coupling, select the channel tab and
then select from:
HDC to couple both the AC and DC components of
an input signal
HTDS7104 and TDS7054: AC to couple only the AC
components of an input signal
HGND to disconnect the input signal from the
acquisition
Touch Close to close the window.
Note. CSA7000 Series: If the O/E Electrical Out--to-CH1 adapter is installed, the Termination, Coupling, and
Bandwidth selections are replaced with Wavelength and
Dark Level selections.
3- 14
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
OverviewRelated control elements and resourcesTo set up signal input (Cont.)
Acquiring Waveforms
Set vertical
acquisition
window
6.Use the vertical knobs to scale and position the
waveform on screen.
Positioned verticallyScaled vertically
Dragging the waveform
handle also positions the
waveform.
7.Touch Vert to display the Vertical control window.
To change the offset, touch the Offset control and
turn the multipurpose knob to adjust the offset.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 15
Acquiring Waveforms
OverviewRelated control elements and resourcesTo set up signal input (Cont.)
Set horizontal
acquisition
window
For help
8.Use horizontal knobs to scale and position the waveform
on screen and to set record length.
Dragging the reference icon
also positions the waveform.
Positioned horizontallyScaled horizontally
The Resolution knob sets the record length. (See
discussion on page 3--24.)
If required to stabilize the display, push LEVEL to set the
trigger level to 50%.
9.For more information on the controls described in this
procedure, push the Vert or Horiz button. Touch the
HELP button in the toolbar.
Continue with
acquisition
setup
10. To finish acquisition setup, you must set the
acquisition mode and start the acquisition.
See To Set Acquisition Modes on page 3--34.
3- 16
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Acquiring Waveforms
To Autoset the Instrument
Autoset automatically sets up the instrument controls (acquisition, display,
horizontal, trigger, and vertical) based on the characteristics of the input signal.
Autoset is much faster and easier than a manual control-by-control setup. When
the input signal is connected, do an autoset to automatically set up the instrument:
OverviewTo autoset the instrumentControl elements and resources
Prerequisites
Execute
1.Signals must be connected to channels. A triggering
source must be provided.
See page 3--71 in this manual for trigger setup
information.
2.Push the Autoset button to execute an Autoset.
If you use Autoset when one or more channels are
displayed, the instrument selects the lowest numbered
channel for horizontal scaling and triggering. All
channels in use are individually vertically scaled.
If you use Autoset when no channels are displayed, the
instrument turns on channel one (CH 1) and scales it.
Execute
Autoset Undo
3.The Autoset Undo control window opens automatically
after an Autoset operation. Click Undo if you want to
undo the last Autoset.
Only the parameters that were set by the last Autoset
are undone. Parameters that you changed that are not
controlled by Autoset retain their settings.
If you don’t want this window to appear, set the
Utilities/User Preferences/Autoset Undo button to Off.
You can still do an Autoset Undo using the Horiz/Acq
menu.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 17
Acquiring Waveforms
OverviewControl elements and resourcesTo autoset the instrument (Cont.)
Prompt
Done
4.Select User Preferences in the Utilities menu to display
the Prompt Before Action window. Touch Autoset to
toggle between ON and OFF:
HOFF to set up for performing an autoset when t he
AUTOSET button is pushed
HON to set up for displaying a prompt before
performing an autoset when the AUTOSET button
is pushed
Touch Close to save your prompt selection.
To Reset the Instrument
NOTE. Autoset may change the vertical position to position the waveform
appropriately. It sets vertical offset to 0 V unless an offsetable probe is attached.
If you are not using an offsetable probe and your signal contains offset (such as
the probe compensation signal), you may need to adjust the Vertical Offset and
SCALE to display the signal.
You may want to revert to the factory default setup; if so, reset the
instrument:
OverviewTo reset the instrumentControl elements and resources
Prerequisites
Execute
1.The instrument is powered up and running.
See Powering On the Instrument on page 1--9.
2.Push the DEFAULT SETUP button.
3- 18
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Acquiring Waveforms
To Get More Help
You can get help on the vertical and acquisition controls by accessing online
help:
OverviewTo get more helpControl elements and resources
Prerequisites
Access
vertical set up
1.Instrument powered up and running.
2.Touch the Help button in toolbar mode or select Help on
Window from the Help menu in menu bar mode.
help
3.You can also select topics related to the vertical controls
from the online help Contents/Index/Find window: select
Contents and Index in the Help menu, as shown at right.
See Powering On the Instrument on page 1--9.
Input Conditioning
Background
This section contains background information that can help you more effectively
set up the acquisition window of each channel.
Input. This instrument samples in real-time or random equivalent-time; both
sampling systems provide pretrigger information by using the trigger to stop an
already running acquisition. Both sampling systems also sample the input after it
is scaled, providing improved input protection and dynamic range.
CAUTION. To prevent damage to the acquisition system, do not overdrive the
inputs and observe static-safe procedures
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 19
Acquiring Waveforms
Autoset Considerations. Autoset acquires samples from the input signal and
attempts to take the following actions based on the input data:
HEvaluate the amplitude range of the input signals and set the size and vertical
offset of the vertical acquisition window to acquire the signal with good
resolution, but without clipping.
HSet the trigger to the approximate midlevel of the signal being autoset and
switches to edge trigger mode.
HEvaluate the signal transitions and set the horizontal scale to produce a
waveform display of 2 or 3 cycles of the input signal.
Sometimes Autoset cannot produce a correct display due to the nature of the
input signal; if so, you may have to adjust the scale, trigger, and acquisition
controls manually. Some conditions that can cause Autoset to fail are:
HNo signal present
HSignals with extreme or variable duty cycles
HSignals with multiple or unstable signal periods
HSignals with too low amplitude
HNo recognizable trigger signal
HSignals with a frequency <20 Hz
HSignals with a frequency above the bandwidth of the instrument
HSignals with high offset and low peak-to-peak variations
Vertical Acquisition Window Considerations. You can set the vertical size,
position, and offset of each channel independently of other channels. Vertical
scale and offset specify the vertical acquisition window for each channel. Parts of
the signal amplitude that fall within the vertical window are acquired; parts
outside (if any) are not.
The offset control subtracts a constant DC level from the input signal before the
vertical scale factor is applied, and the vertical position control adds a constant
number of divisions of signal after the scale factor is applied to the resulting
difference.
3- 20
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Acquiring Waveforms
The vertical scale and position controls have the following effects on the vertical
acquisition window and the displayed waveform:
HThe vertical volts per division that you set determines the vertical size of the
acquisition window, allowing you to scale it to contain all of a waveform
amplitude or only part. Figure 3--2 on page 3--22 shows two vertical
acquisition windows that contain the entire waveform, but only one window
contains the entire waveform in the graticule on screen.
NOTE. Amplitude-related automatic measurements (for example, peak-to-peak
and RMS) will be accurate for vertical windows like those shown in Figure 3--2a
and b because neither waveform is clipped (that is, both waveforms are
acquired). But if signal amplitude were to extend outside the vertical acquisition
window, the data acquired is clipped. Clipped data causes inaccurate results if
used in amplitude-related automatic measurements. Clipping also causes
inaccurate amplitude values in waveforms that are stored or exported for use in
other programs.
If the scale of a math waveform is changed so that the math waveform is clipped,
it will not affect amplitude measurements on that math waveform.
HThe vertical position adjusts the acquisition, and therefore the display, of the
acquired waveform relative to the vertical acquisition window. Figure 3--2b
shows how vertical position moves the acquired waveform vertically in the
acquisition window to place the acquired waveform in the graticule display.
HAs you vary the vertical position, the vertical acquisition window moves up
and down (±5 divisions) on the waveform. With input signals that are
smaller than the window, it appears the waveform moves in the window.
Actually, the position moves the vertical acquisition window up and down
on the input signal.
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 21
Acquiring Waveforms
a. SCALE setting determines
the vertical acquisition
window size; here
100 mV/div x 10 divisions
(8 graticule divisions and
ᐔ1 division of position)
b. Vertical offset and position
can change the location of
the acquired waveform
within the acquisition
window, repositioning it so
its waveform appears in the
graticule
Vertica l
window
Channel
reference
indicator
Vertica l
window
Channel
reference
Indicator
1
1
Figure 3- 2: Setting vertical range and position of input channels
+0.5 Volt
+0.4 Volt
Graticule
--0.4 Volt
--0.5 Volt
+1.0 Volt
+0.7 Volt
Graticule
--0.1 Volt
--1.0 Volt
The vertical offset control affects the vertical acquisition window and the
displayed waveform as follows:
HThe vertical range (window) is always centered around the offset value. It is
the voltage level at the middle of the vertical acquisition window. With no
(zero) offset, as shown in Figure 3--2 a and b, that voltage level is zero
(ground).
HAs you vary vertical offset, the middle voltage level moves relative to zero.
This moves the vertical acquisition window up and down on the waveform.
With input signals that are smaller than the window, it appears the waveform
moves in the window. Actually, a larger signal shows what really happens:
the offset moves the middle of the vertical acquisition window up and down
on the input signal. Figure 3--3 shows how offset moves the acquisition
window to control the portion of the waveform amplitude the window
captures.
3- 22
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.