Tektronix TDS7000B Datasheet

Digital Phosphor Oscilloscopes
TDS7000B Series
Superior Measurement Fidelity, Unrivaled Analysis, Uncompromised Usability
The TDS7000B Series’ unique combina­tion of superior measurement fidelity, unrivaled analysis and uncompromised usability makes it the ultimate test machine to simplify and speed the design of high-speed, complex systems. This family offers the industry’s best solution to the challenging signal integrity issues faced by designers verifying, character­izing and debugging sophisticated electronic designs. They deliver up to 7.25 GHz (typical) true analog bandwidth, down to 43 ps rise time (20% to 80%), and 20 GS/s maximum real-time sample rate to capture critical events with fine detail.
Exclusive DPX enables waveform capture rates up to 400,000 wfms/s to quickly find rare glitches in seconds or minutes, instead of hours or days. Innovative software solutions deliver domain expertise for advanced analysis and compliance testing, while the OpenChoice enables users to integrate their expertise through the ability to easily write custom programs or utilize popular commercial software. The intuitive graphical user interface delivers sophisticated capability to advanced users without intimidating occasional users.
®
acquisition technology
®
architecture
Features & Benefits
Up to 7.25 GHz True Analog Bandwidth and Down to 43 ps Rise Time (20% to 80%)
>400,000 wfms/s Waveform Capture Rate, Powered by Exclusive DPX®Acquisition Technology
20 GS/s Maximum Real-time Sample Rate
Exceptional Delta-time Accuracy for High Confidence in Critical Timing Measurements
MyScope®Custom Control Windows Enhance Productivity
Right Mouse Click Menus for Exceptional Efficiency
Powerful Triggering Features for Fast Detection of Relevant Faults
Communications Mask Testing Up to 4.25 Gb/s Rates
Clock Recovery from Serial Data Streams up to 3.125 Gb/s
Up to 64 MB Record Length with MultiView Zoom™for Quick Navigation of Long Records
TekConnect®Interface for High Fidelity Connection
Classic Direct Controls, Touch-sensitive Display or Mouse Navigation
OpenChoice®with Microsoft Windows 2000 Delivers Built-in Networking and Analysis
XGA 1024x768 Display
Applications
Signal Integrity, Jitter and Timing Analysis
Verification, Debug and Characterization of Sophisticated Designs
Design, Development and Compliance Testing of Serial Data Streams Up to 4.25 Gb/s Rates
Debug of Telecom, Datacom and Storage Area Network Equipment Designs, and High-speed Backplanes
Spectral Analysis Investigation of
Transient Phenomena
Digital Phosphor Oscilloscopes
TDS7000B Series
MyScope®Custom Control Windows
MyScope control windows allow you to build your own control windows with only the controls, features and capabilities that you care about and are important in your job. For the first time you can create your own personalized “toolbox” of oscilloscope features. No longer do you need to search through menus for features or re-learn how to drive the oscilloscope after a break from the lab. MyScope control windows are easily created in a matter of minutes using a simple, visual, drag-and-drop process. Once created, these customized windows are easily accessed through a dedicated MyScope button and menu selection on the oscilloscope button/menu bar, just like any other control window. You can make an unlimited number of custom control windows, enabling each person who uses the oscilloscope in a shared
environment to have their own unique control window. Since the contr ol windows are stored as files on the hard drive, they can easily be transferred to other CSA7000B, TDS5000B or TDS6000B/C Series oscilloscopes, or they can even be e-mailed to a co-worker around the world when the need arises. MyScope control windows will benefit all oscilloscope users, from eliminating the ramp-up time that many face when returning to the lab after not using an oscilloscope for a while, to the power user who can now operate far more efficiently. Everything you need is found in one control window rather than having to constantly navigate through menu after menu to repeat similar tasks.
Right Clicks
Right mouse click menus make simple things as they should be – simple. Right click menus are context sensitive, meaning the choices presented in the menu depend on where you right clicked
the mouse. This makes right click menus extremely intuitive. Want to change the cursor type? Right click on a cursor or the cursor readouts. Want to change the reference levels of an automatic measure­ment? Right click on the measurement. Want to change trigger parameters? Right click on the trigger readouts. Want to change a waveform’s color? Right click on the waveform handle. Virtually all objects on the oscilloscope display have right click menus associated with them that include all the appropriate actions or features relative to those objects. There are also right click menus for regions of the display in addition to just objects. For example, right clicking in the main graticule brings up a menu with choices such as Clear Data, Default Setup, Autoset, Screen Captures, Save All Waveforms and Add Screen Text, providing single-click access to many of your most commonly performed tasks.
Digital Phosphor Oscilloscopes • www.tektronix.com/oscilloscopes
2
Digital Phosphor Oscilloscopes
TDS7000B Series
Characteristics
Vertical System
TDS7054 TDS7104 TDS7154B TDS7254B TDS7404B TDS7704B
Input Channels 4 4 4 4 4 4 Hardware Analog 500 MHz 1 GHz 1.5 GHz
*1
Bandwidth (–3 dB) (typical), 7GHz
Rise Time 10% 800 ps 400 ps 200 ps 130 ps 100 ps 62 ps to 90% (typical)
Rise Time 20% 135 ps 83 ps 72 ps 43 ps to 80% (typical)
DC Gain Accuracy 1% ±(2% + ±2.5% +
(2% x offset) (2% x offset)
Hardware Bandwidth 250 MHz or 20MHz Requires TCA-1 MEG Limits
Input Coupling AC, DC, GND DC, GND Input Impedance 1MΩ ±0.5% or 50 Ω ±1% 50 Ω ±2.5% Input Sensitivity,1 MΩ 1 mV/div to 10 V/div — Input Sensitivity,50 Ω 1 mV/div to 1 V/div 2 mV/div to 1 V/div Vertical Resolution 8-Bit (>11-Bit with averaging) 8-Bit (>11-Bit with averaging) Max Input Voltage, ±150 V CAT I Derate at 20 dB/decade
1MΩ to 9 V Max Input Voltage, 5V
, with peaks less than ±30 Volts <1 V
RMS
above 200 kHz
RMS
for <100 mV/div, <5 V
RMS
50 Ω Also determined by TekConnect®accessory Offset Range 1 mV/div to 100 mV/div: ±1 V 2 mV to 50 mV/div: ±0.5 V
101 mV/div to 1 V/div: ±10 V 50.5 mV to 99.5 mV: ±0.25 V
1.01 V/div to 10V/div: ±100 V 100 mV to 500 mV: ±5 V
Channel-to-Channel 100:1 at 100 MHz and 80:1 at 1.5 GHz and Isolation Any Two 30:1 at the rated bandwidth 15:1 at the rated bandwidth Channels at Equal Vertical Scale Settings ≥≥≥≥
*1
At 10 mV/div.
*1
2.5 GHz
for 100 mV/div settings
RMS
505 mV to 1 V/div: ±2.5 V
4 GHz
*1
7.25 GHz
*1
guaranteed
*1
Note:Typical system bandwidth of TDS7404B with P7240:4GHz. Note:Typical system bandwidth of TDS7404B with P7330:3.5GHz. Note:Typical system bandwidth of TDS7704B with P7260:6GHz. Note:Typical system bandwidth of TDS7704B with P7350:5GHz. Note:Typical system bandwidth of TDS7704B with P7380:7GHz.
Digital Phosphor Oscilloscopes • www.tektronix.com/oscilloscopes
3
Digital Phosphor Oscilloscopes
TDS7000B Series
Time Base System
TDS7054/TDS7104 TDS7154B/TDS7254B/
TDS7404B/TDS7704B
Time Base Range 200 ps/div to 40 s/div 50 ps to 10 s/div Time Base Delay Time Range 16 ns to 250 s 5 ns to 250 s Channel-to-Channel Deskew Range ±25ns ±75 ns Delta Time Measurement Accuracy ±((0.06/sample rate) + (15 ppm x reading)) RMS ±((0.06/sample rate) + (2.5 ppm x reading)) RMS Trigger Jitter (RMS) 8 ps
Long Term Sample Rate and Delay ±15 ppm over 1 ms interval 2.5 ppm over any 100 ms interval Time Accuracy
Acquisition System
TDS7054 TDS7104 TDS7154B/ TDS7254B/
Real-time Sample Rates
1 channel (max) 5 GS/s 10 GS/s 20 GS/s 2 channels (max) 5GS/s 5 GS/s 10 GS/s 3-4 channels (max) 2.5 GS/s 2.5 GS/s 5 GS/s Equivalent Time Sample Rate (max) 250 GS/s 250 GS/s 1TS/s Maximum Record Length per Channel 2 Mb (1-CH.), 1 Mb (2-CH.), 500 Kb (4-CH.) 4 Mb (1-CH.), 2 Mb (2-CH.),
with Standard Memory 1 Mb (4-CH.)
With Memory Opt. 2M 8 Mb (1-CH.), 4 Mb (2-CH.),
With Memory Opt. 3M 16 Mb (1-CH.), 8 Mb (2-CH.),
With Memory Opt. 4M 32 Mb (1-CH.), 16 Mb (2-CH.),
With Memory Opt. 5M 64 Mb (1-CH.), 32 Mb (2-CH.),
(typical) 2 ps
RMS
2 Mb (4-CH.)
4 Mb (4-CH.)
(typical) (7254B/7154B)
RMS
1.5 ps
1.2 ps
(typical) (7404B)
RMS
(typical) (7704B)
RMS
TDS7404B/ TDS7704B
8 Mb (4-CH.)
16 Mb (4-CH.)
Maximum Duration at Highest Real-time Resolution (1-CH)
TDS7054 TDS7104 TDS7154B/TDS7254B/
TDS7404B/TDS7704B
Time Resolution (Single-shot) 200 ps (5 GS/s) 100 ps (10 GS/s) 50 ps (20 GS/s) Max Duration with Standard Memory 400 µs 200 µs 200 µs Max Duration with Opt. 2M 1.6 ms 800 µs 400 µs Max Duration with Opt. 3M 3.2 ms 1.6 ms 800 µs Max Duration with Opt. 4M 1.6 ms Max Duration with Opt. 5M 3.2 ms
4
Digital Phosphor Oscilloscopes • www.tektronix.com/oscilloscopes