Please check for change information at the rear
of this manual.
First Printing: July 1993.
Instrument Serial Numbers
Each instrument manufactured by Tektronix has a serial number on a panel insert or tag, or
stamped on the chassis. The first letter in the serial number designates the country of
manufacture. The last five digits of the serial number are assigned sequentially and are unique
to each instrument. Those manufactured in the United States have six unique digits. The
country of manufacture is identified as follows:
B010000Tektronix, Inc., Beaverton, Oregon, USA
E200000Tektronix United Kingdom, Ltd., London
J300000Sony/Tektronix, Japan
H700000Tektronix Holland, NV, Heerenveen, The Netherlands
Instruments manufactured for Tektronix by external vendors outside the United States are
assigned a two digit alpha code to identify the country of manufacture (e.g., JP for Japan, HK
for Hong Kong, IL for Israel, etc.).
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
Printed in U.S.A.
Copyright E Tektronix, Inc., 1993. All rights reserved. Tektronix products are covered by U.S.
and foreign patents, issued and pending. The following are registered trademarks:
TEKTRONIX, TEK, TEKPROBE, and SCOPEĆMOBILE.
WARRANTY
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from
the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair
the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty
period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and
shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall
pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service
center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for
products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance
and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by
personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from
improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with
other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER
WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR
REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR
BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Welcome
This is the Performance Verification for the TDS 520A, 524A, 540A, and 544A
Oscilloscope. It contains procedures suitable for determining if the instruĆ
ment functions, was adjusted properly, and meets the performance characĆ
teristics as warranted.
Also contained in this document are technical specifications for these oscilĆ
loscopes.
Related Manuals
The following documents are related to the use or service of the digitizing
oscilloscope.
HThe TDS 520A, 524A, 540A, & 544A User Manual (Tektronix part number
070-8710-01).
HThe TDS Family Programmer Manual (Tektronix part number
070-8709-01) describes using a computer to control the digitizing
oscilloscope through the GPIB interface.
Please take a moment to review these safety precautions. They are provided
for your protection and to prevent damage to the digitizing oscilloscope.
This safety information applies to all operators and service personnel.
Symbols and Terms
These two terms appear in manuals:
H
statements identify conditions or practices that could result in
damage to the equipment or other property.
Hstatements identify conditions or practices that could result in
personal injury or loss of life.
These two terms appear on equipment:
HCAUTION indicates a personal injury hazard not immediately accessible
as one reads the marking, or a hazard to property including the equipĆ
ment itself.
HDANGER indicates a personal injury hazard immediately accessible as
Observe all of these precautions to ensure your personal safety and to
prevent damage to either the digitizing oscilloscope or equipment conĆ
nected to it.
Power Source
The digitizing oscilloscope is intended to operate from a power source that
will not apply more than 250 V
tween either supply conductor and ground. A protective ground connection,
through the grounding conductor in the power cord, is essential for safe
system operation.
between the supply conductors or beĆ
RMS
Grounding the Digitizing Oscilloscope
The digitizing oscilloscope is grounded through the power cord. To avoid
electric shock, plug the power cord into a properly wired receptacle where
earth ground has been verified by a qualified service person. Do this before
making connections to the input or output terminals of the digitizing oscilloĆ
scope.
Without the protective ground connection, all parts of the digitizing oscilloĆ
scope are potential shock hazards. This includes knobs and controls that
may appear to be insulators.
Use the Proper Power Cord
Use only the power cord and connector specified for your product. Use only
a power cord that is in good condition.
Use the Proper Fuse
To avoid fire hazard, use only the fuse specified in the parts list for your
product, matched by type, voltage rating, and current rating.
Do Not Remove Covers or Panels
To avoid personal injury, do not operate the digitizing oscilloscope without
the panels or covers.
Electric Overload
Never apply to a connector on the digitizing oscilloscope a voltage that is
outside the range specified for that connector.
Do Not Operate in Explosive Atmospheres
The digitizing oscilloscope provides no explosion protection from static
discharges or arcing components. Do not operate the digitizing oscilloscope
in an atmosphere of explosive gases.
vi
Safety
Performance Verification
Procedures
Brief Procedures
The Self Tests use internal routines to confirm basic functionality and proper
adjustment. No test equipment is required to do these test procedures.
The Functional Tests utilize the probeĆcompensation output at the front panel
as a testĆsignal source for further verifying that the oscilloscope functions
properly. A standardĆaccessory probe, included with this oscilloscope, is the
only equipment required.
General Instructions
Besides the Brief Procedures, the set of procedures that can be used to
verify oscilloscope performance includes the Performance Tests, found later
in this section. You may not need to perform all of these procedures, deĆ
pending on what you want to accomplish:
HTo rapidly confirm that this oscilloscope functions and was adjusted
properly, just do the procedures under Self Tests, which begin on
page 1Ć4.
Advantages: These procedures are quick to do, require no external
equipment or signal sources, and perform extensive functional and
accuracy testing to provide high confidence that the oscilloscope will
perform properly. They can be used as a quick check before making a
series of important measurements.
HTo further check functionality, first do the Self Tests just mentioned; then
do the procedures under Functional Tests that begin on page 1Ć6.
Advantages: These procedures require minimal additional time to
perform, require no additional equipment other than a standardĆaccesĆ
sory probe, and more completely test the internal hardware of this
oscilloscope. They can be used to quickly determine if the oscilloscope
is suitable for putting into service, such as when it is first received.
HIf more extensive confirmation of performance is desired, do the PerĆ
formance Tests, beginning on page 1Ć15, after doing the Functional and
Self Tests just referenced.
Advantages: These procedures add direct checking of warranted
specifications. They require more time to perform and suitable test
equipment is required. (See Equipment Required on page 1Ć16.)
If you are not familiar with operating this oscilloscope, read the TDS 520A,
524A, 540A, 544A, & 644A Reference or the TDS 520A, 524A, 540A, & 544A
User manual. These contain instructions that acquaint you with the use of
the frontĆpanel controls and the menu system.
Throughout these procedures the following conventions apply:
HEach test procedure uses the following general format:
Title of Test
Equipment Required
Prerequisites
Procedure
HEach procedure consists of as many steps, substeps, and subparts as
required to do the test. Steps, substeps, and subparts are sequenced
as follows:
1. First Step
a. First Substep
HFirst Subpart
HSecond Subpart
b. Second Substep
2. Second Step
HIn steps and substeps, the leadĆin statement in italics instructs you what
to do, while the instructions that follow tell you how to do it: in the examĆ
ple step below, Initialize the oscilloscope" by doing Press save/recall
SETUP. Now, press the mainĆmenu button...".
Initialize the oscilloscope: Press save/recall SETUP. Now, press the
mainĆmenu button Recall Factory Setup; then the sideĆmenu
button OK Confirm Factory Init.
HWhere instructed to use a frontĆpanel button or knob, or select from a
main or side menu, or verify a readout or status message, the name of
the button or knob appears in boldface type: press SHIFT; then ACĆQUIRE MENU", press the mainĆmenu button Coupling", or verify that
the status message is Pass.
The symbol at the left is accompanied by information you must read
to do the procedure properly.
1Ć2
Performance Verification Procedures
Brief Status
Information
Graticule and Waveforms
Waveform Reference
Symbols: Ground Levels
and Waveform Sources
Vertical Scale, Horizontal
Scale, and Trigger Level
Readouts
Brief Procedures
HRefer to Figure 1Ć1: Main menu" refers to the menu that labels the
seven menu buttons under the display; side menu" refers to the menu
that labels the five buttons to the right of the display. PopĆup menu"
refers to a menu that pops up when a mainĆmenu button is pressed.
Position of Waveform
Record Relative to
the Screen and Display
General Purpose
Knob Readout
PopĆup Menu
Side menu area.
Readouts for
measurements
move here when
CLEAR MENU is
pressed.
Main menu display area. Readouts in
lower graticule area move here when
This procedure uses internal routines to verify that this oscilloscope funcĆ
tions and was adjusted properly. No test equipment or hookups are reĆ
quired.
Verify Internal Adjustment, Self Compensation, and
Diagnostics
Equipment Required: None.
Prerequisites: Power on the Digitizing Oscilloscope and allow a 20 minute
warmĆup before doing this procedure.
Procedure:
1. Verify that internal diagnostics pass: Do the following substeps to verify
passing of internal diagnostics.
a. Display the System diagnostics menu:
HPress SHIFT; then press UTILITY.
HRepeatedly press the mainĆmenu button System until Diag/Err
is highlighted in the popĆup menu.
b. Run the System Diagnostics: Press the mainĆmenu button Execute;
then press the sideĆmenu button OK Confirm Run Test.
c. Wait: The internal diagnostics do an exhaustive verification of proper
oscilloscope function. This verification will take up to two minutes. At
some time during the wait, a clock" icon (shown at left) will appear
onĆscreen. When the verification is finished, the resulting status will
appear on the screen.
d. Confirm no failures are found: Verify that no failures are found and
reported onĆscreen.
e. Confirm the three adjustment sections have passed status:
HPress SHIFT; then press UTILITY.
HPress the main menu button System until Cal is highlighted in
the popĆup menu.
HVerify that the word Pass appears in the main menu under the
following menu labels: Voltage Reference, Frequency ReĆsponse, and Pulse Trigger. (See Figure 1Ć2.)
1Ć4
Performance Verification Procedures
First, the CAL menu is displayed.
Second, the adjustment
sections are verified.
Brief Procedures
compensation is run and is
Third, a signal path
verified.
Figure 1Ć2:ăVerifying Adjustments and SignalĆPath Compensation
When doing steps f and g, do not turn off the oscilloscope until
signalĆpath compensation completes. If you interrupt (or lose) power
to the instrument while signalĆpath compensation is running, a
message is logged in the oscilloscope error log. If such a case
occurs, rerun signalĆpath compensation.
f.Run the signalĆpath compensation: Press the mainĆmenu button
Signal Path; then press the sideĆmenu button OK Compensate
Signal Paths.
g. Wait: SignalĆpath compensation runs in about one to two minutes.
While it progresses, a clock" icon (shown at left) is displayed onĆ
screen. When compensation completes, the status message will be
updated to Pass or Fail in the main menu (see step h).
h. Confirm signalĆpath compensation returns passed status: Verify the
word Pass appears under Signal Path in the main menu. (See
Figure 1Ć2.)
2. Return to regular service: Press CLEAR MENU to exit the system meĆ
nus.
The purpose of these procedures is to confirm that this oscilloscope funcĆ
tions properly. The only equipment required is one of the standardĆaccessoĆ
ry probes and, to check the file system, a 3.5 inch. 720 K or 1.44 Mbyte
floppy disk.
These procedures verify functions; that is, they verify that oscilloĆ
scope features operate. They do not verify that they operate within
limits.
Therefore, when the instructions in the functional tests that follow
call for you to verify that a signal appears onĆscreen that is about
five divisions in amplitude" or has a period of about six horizontal
divisions", etc., do NOT interpret the quantities given as limits.
Operation within limits is checked in Performance Tests, which begin
on page 1Ć15.
DO NOT make changes to the frontĆpanel settings that are not
called out in the procedures. Each verification procedure will require
you to set the oscilloscope to certain default settings before verifying
functions. If you make changes to these settings, other than those
called out in the procedure, you may obtain invalid results. In this
case, just redo the procedure from step 1.
When you are instructed to press a menu button, the button may
already be selected (its label will be highlighted). If this is the case, it
is not necessary to press the button.
Verify All Input Channels
Equipment Required: One P6139A probe.
Prerequisites: None.
Procedure:
1. Install the test hookup and preset the oscilloscope controls:
Figure 1Ć3:ăUniversal Test Hookup for Functional Tests
a. Hook up the signal source: Install the probe on CH 1. Connect the
probe tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND.
1Ć6
Performance Verification Procedures
Brief Procedures
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
2. Verify that all input channels operate: Do the following substeps Ċ test
CH 1 first, skipping substepa since CH 1 is already set up for verification
from step 1.
a. Select an unverified channel:
HPress WAVEFORM OFF to remove from display the channel just
verified.
HPress the frontĆpanel button that corresponds to the channel
you are to verify.
HMove the probe to the channel you selected.
b. Set up the selected channel:
HPress AUTOSET to obtain a viewable, triggered display in the
selected channel.
HSet the horizontal SCALE to 200 ms. Press CLEAR MENU to
remove any menu that may be on the screen.
c. Verify that the channel is operational: Confirm that the following
statements are true.
HThe vertical scale readout for the channel under test shows a
setting of 200 mV, and a squareĆwave probeĆcompensation
signal about 2.5 divisions in amplitude is onĆscreen. (See FigĆ
ure 1Ć1 on page 1Ć3 to locate the readout.)
HThe vertical POSITION knob moves the signal up and down the
screen when rotated.
HTurning the vertical SCALE knob counterclockwise decreases
the amplitude of the waveform onĆscreen, turning the knob
clockwise increases the amplitude, and returning the knob to
200 mV returns the amplitude to about 2.5 divisions.
d. Verify that the channel acquires in all acquisition modes: Press
SHIFT; then press ACQUIRE MENU. Use the side menu to select, in
turn, each of the five hardware acquire modes and confirm that the
following statements are true. Refer to the icons at the left of each
statement as you confirm those statements.
HSample mode displays an actively acquiring waveform onĆ
screen. (Note that there is noise present on the peaks of the
square wave.)
HPeak Detect mode displays an actively acquiring waveform
onĆscreen with the noise present in Sample mode peak deĆ
tected."
HHi Res mode displays an actively acquiring waveform onĆscreen
with the noise that was present in Sample mode reduced.
HEnvelope mode displays an actively acquiring waveform onĆ
screen with the noise displayed.
HAverage mode displays an actively acquiring waveform onĆ
screen with the noise reduced like in Hi Res mode.
TDS 520A and 524A only: Substep e will have you repeat the preĆ
vious substeps to check all input channels. Be sure to check only
CH 1 and CH 2 when testing the TDS 520A or 524A. (Step 3 will test
the AUX 1 and AUX 2 inputs.) When testing the TDS 540A or 544A,
test all four channels, CH 1 through CH 4.)
e. Test all channels: Repeat substeps a through d until all four input
channels are verified.
3. TDS 520A and 524A Only: Verify auxiliary inputs operate: Perform the
following substeps when checking the AUX 1 and AUX 2 inputs only.
a. Select an auxiliary channel:
HPress WAVEFORM OFF to remove from display the channel just
verified.
HPress the frontĆpanel button that corresponds to the channel
you are to verify.
HMove probe to the channel you selected.
b. Set up the selected channel: Press AUTOSET to obtain a viewable
display in the selected channel.
The display obtained might not trigger stably because autoset
cannot provide more than about ½ division of probe compensation
signal amplitude in an auxiliary channel equipped with a 10X probe.
This amount is less than the minimum trigger sensitivity for auxiliary
channel trigger sources; therefore, triggering is not required.
c. Verify that the channel is operational: Confirm that the following
statements are true.
HThe vertical scale readout for the channel under test shows a
setting of 1 V, and a squareĆwave probeĆcompensation signal
about ½ division in amplitude is onĆscreen. (See Figure 1Ć1 on
page 1Ć3 to locate the readout.)
HThe vertical POSITION knob moves the signal up and down the
screen when rotated.
HTurning the vertical SCALE knob counterclockwise to 10 V
decreases the amplitude of the waveform onĆscreen. (The
amplitude will drop to near zero when doing this substep.)
1Ć8
HReturning the knob to 1 V returns the amplitude to about ½
division.
d. Verify that the channel acquires in all acquisition modes: Disconnect
the probe ground lead from the probeĆcompensation terminal. Do
step 2, substep d to verify the five acquire modes.
Performance Verification Procedures
Brief Procedures
e. Test all channels: Repeat substeps a through d to verify AUX 2.
4. Remove the test hookup: Disconnect the probe from the channel input
and the probeĆcompensation terminals.
Verify the Time Base
Equipment Required: One P6139A probe.
Prerequisites: None.
Procedure:
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the
probe tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND. (See
Figure 1Ć3 on page 1Ć6.)
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup; then press
the sideĆmenu button OK Confirm Factory Init.
c. Modify default settings:
HPress AUTOSET to obtain a viewable, triggered display.
HSet the horizontal SCALE to 200 ms.
HPress CLEAR MENU to remove the menus from the screen.
2. Verify that the time base operates: Confirm the following statements.
a. One period of the squareĆwave probeĆcompensation signal is about
five horizontal divisions onĆscreen for the 200 ms horizontal scale
setting (set in step 1c).
b. Rotating the horizontal SCALE knob clockwise expands the waveĆ
form onĆscreen (more horizontal divisions per waveform period), and
that counterclockwise rotation contracts it, and that returning the
horizontal scale to 200 ms returns the period to about five divisions.
c. The horizontal POSITION knob positions the signal left and right
onĆscreen when rotated.
3. Remove the test hookup: Disconnect the probe from the channel input
and the probeĆcompensation terminals.
Verify the File System (Optional on TDS 520A and 540A)
Equipment Required: One P6139A probe and one 720 K or 1.44 Mbyte, 3.5
inch DOS compatible disk. You can use a disk of your own or you can use
the Programming Examples Software 3.5 inch disk (Tektronix part number
063-1134-00) contained in the TDS Family Programmer Manual (Tektronix
part number 070-8709-01).
Prerequisites: None.
Procedure:
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the
probe tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND. (See
Figure 1Ć3 on page 1Ć6.)
b. Insert the test disk: Insert the disk in the disk drive to the left of the
monitor.
HPosition the disk so the metal shutter faces the drive.
HPosition the disk so the stamped arrow is on the top right side.
In other words, place the angled corner in the front bottom
location.
HPush the disk into the drive until it goes all the way in and clicks
into place.
c. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
d. Modify default settings:
HPress AUTOSET to obtain a viewable, triggered display.
HSet the horizontal SCALE for the M (main) time base to 200 ms
(one click clockwise). Notice the waveform on the display now
shows two cycles instead of five.
HPress CLEAR MENU to remove the menus from the screen.
e. Save the settings:
HPress SETUP.
HPress the mainĆmenu button Save Current Setup; then press
the sideĆmenu button To File.
HTurn the general purpose knob to select the file to save. Choose
TEK?????.SET (or fdo:). With this choice, you'll save a file
starting with TEK, then containing 5Ćnumbers, and a .SET
extension. For example, the first time you run this on a blank,
formatted disk or on the Example Programs Disk, the TDS will
assign the name TEK00000.SET to your file. If you ran the
procedure again, the TDS would increment the name and call
the file TEK00001.SET.
1Ć12
Performance Verification Procedures
HPress the sideĆmenu button Save To Selected File.
2. Verify the file system works:
HPress AUTOSET to restore the 500 ms time base and the five
cycle waveform.
HPress the mainĆmenu button Recall Saved Setup; then press
the sideĆmenu button From File.
HTurn the general purpose knob to select the file to recall. For
example, if you followed the instructions above and used a
blank disk, you had the TDS assign the name TEK00000.SET to
your file.
HPress the sideĆmenu button Recall From Selected File.
HVerify that Digitizing Oscilloscope retrieved the saved setup from
the disk. Do this by noticing the horizontal SCALE for the M
(main) time base is again 200 ms and the waveform shows only
two cycles just as it was when you saved the setup.
3. Remove the test hookup:
Brief Procedures
HDisconnect the standardĆaccessory probe from the channel
input and the probeĆcompensation terminals.
HRemove the disk from the disk drive. Do this by pushing in the
This subsection contains a collection of procedures for checking that TDS
520A, 524A, 540A, and 544A Digitizing Oscilloscopes perform as warranted.
Since the procedures cover models with both two fullĆfeatured channels
(TDS 520A and 524A) and four fullĆfeatured models (TDS 540A and 544A),
instructions that apply only to one of the model types are clearly identified.
Otherwise, all test instructions apply to both the two and four channel
models.
The procedures are arranged in four logical groupings: Signal AcquisitionSystem Checks, Time Base System Checks, Triggering System Checks, and
Output Ports Checks. They check all the characteristics that are designated
as checked in Section 2, Specifications. (The characteristics that are
checked appear in boldface type under Warranted Characteristics in
Section 2.)
These procedures extend the confidence level provided by the basic
procedures described on page 1Ć1. The basic procedures should be
done first, then these procedures performed if desired.
Prerequisites
The tests in this subsection comprise an extensive, valid confirmation of
performance and functionality when the following requirements are met:
HThe cabinet must be installed on the Digitizing Oscilloscope.
HYou must have performed and passed the procedures under Self Tests,
found on page 1Ć4, and those under Functional Tests, found on
page 1Ć6.
HA signalĆpath compensation must have been done within the recomĆ
mended calibration interval and at a temperature within ±5_ĂC of the
present operating temperature. (If at the time you did the prerequisite
Self Tests, the temperature was within the limits just stated, consider this
prerequisite met.)
HThe Digitizing Oscilloscope must have been last adjusted at an ambient
temperature between +20_ĂC and +30_ĂC, must have been operating
for a warmĆup period of at least 20 minutes, and must be operating at an
ambient temperature between +4_ĂC and +50_ĂC. (The warmĆup reĆ
quirement is usually met in the course of meeting the first prerequisite
listed above.)
Related Information Ċ Read General Instructions and Conventions that
These procedures use external, traceable signal sources to directly check
warranted characteristics. The required equipment list is shown in Table 1Ć1.
Instrument Serial Number:Certificate Number:
Temperature:RH %:
Date of Calibration:Technician:
Performance TestMaximumOutgoingIncomingMinimum
Probe Compensator Output Signal
Frequency (CH1 Freq.)950 Hz____________________1050 Hz
Voltage (difference)495 mV____________________505 mV
1Ć22
Performance Verification Procedures
Performance Tests
Signal Acquisition
System Checks
These procedures check those characteristics that relate to the signalĆacquiĆ
sition system and are listed as checked under Warranted Characteristics in
Section 2, Specifications.
Check Accuracy of Offset (Zero Setting)
Equipment Required: None.
Prerequisites: The oscilloscope must meet the prerequisites listed on
page 1Ć15.
1. Preset the instrument controls:
a. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
HPress CLEAR MENU to remove the menus from the screen.
b. Modify the default settings:
HSet the horizontal SCALE to 1 ms.
HPress SHIFT; then ACQUIRE MENU.
HPress the mainĆmenu button Mode; then press the sideĆmenu
button Hi Res.
HPress DISPLAY.
HPress the mainĆmenu button Graticule; then press the sideĆmeĆ
nu button Frame.
HPress CURSOR.
HPress the mainĆmenu button Function; then press the sideĆmeĆ
nu button H Bars.
HPress CLEAR MENU.
2. Confirm input channels are within limits for offset accuracy at zero offset:
Do the following substeps Ċ test CH 1 first, skipping substep a since
CH 1 is already set up to be checked from step 1.
a. Select an unchecked channel: Press WAVEFORM OFF to remove
the channel just confirmed from the display. Then, press the frontĆ
panel button that corresponds to the channel you are to confirm.
Follow these rules to match this procedure to the model of the
oscilloscope under test:
Models TDS 540A, 544A OnlyĊWhen using Table 1Ć2 to test
CH 1ĊCH 4; ignore the columns for AUX 1 & AUX 2 settings and
limits.
Model TDS 520A, 524A OnlyĊUse Table 1Ć2 to test input chanĆ
nels; use the columns for CH 1ĊCH 4 when testing CH 1 and CH 2;
use the columns for AUX 1 and AUX 2 when testing those channels.
TableĂ1Ć2:ăDC Offset Accuracy (Zero Setting)
Vertical Scale
Setting
CH 1 CH 4
1ĂmV
AUX 1 &
AUX 2
100ĂmV0±1.6ĂmV±10ĂmV
Vertical
Position
and Offset
Setting
1
Offset Accuracy Limits
CH1-CH4AUX1&AUX2
100ĂmV1V0±25ĂmV±100ĂmV
1ĂV10ĂV0±250ĂmV±1ĂV
1
Vertical position is set to 0 divisions and vertical offset to 0 V when the oscilloscope is
initialized in step 1.
b. Set the vertical scale: Set the vertical SCALE to one of the settings
listed in Table 1Ć2 that is not yet checked. (Start with the first setting
listed.)
c. Display the test signal: The baseline DC test level was initialized for
all channels in step 1 and is displayed as you select each channel
and its vertical scale. Be sure not to use the vertical POSITION knob
while checking any channel for accuracy of offset, since varying the
position invalidates the check.
d. Measure the test signal: Rotate the general purpose knob to superĆ
impose the active cursor over the baseline DC test level. (Ignore the
other cursor.)
1Ć24
e. Read the measurement results at the absolute (@:) cursor readout,
not the delta (D:) readout on screen (see Figure 1Ć4).
f.Check against limits: Do the following subparts in the order listed.
HCHECK that the measurement results are within the limits listed
for the current vertical scale setting.
HRepeat substeps b through f until all vertical scale settings
settings listed in Table 1Ć2 are checked for the channel under
test.
g. Test all channels: Repeat substeps a through f for all input channels.
3. Disconnect the hookup: No hookup was required.
Performance Verification Procedures
Ignore the
inactive cursor.
Align the active
cursor to the DC
baseline (no input).
Then read the
offset relative to
ground reference.
Figure 1Ć4:ăMeasurement of DC Offset Accuracy at Zero Setting
Performance Tests
Check DC Gain and Voltage Measurement Accuracy
WARNING
Performance of this procedure requires input voltages up to
130 VDC. Be sure to set the DC calibration generator to 0 volts
before connecting, disconnecting, and/or moving the test hookup
during the performance of this procedure.
Equipment Required: Two dualĆbanana connectors (Item 5), one BNC T
connector (Item 6), one DC calibration generator (Item 8), and two precision
coaxial cables (Item 4).
Prerequisites: The oscilloscope must meet the prerequisites listed on
page 1Ć15.
Figure 1Ć6:ăMeasurement of the DC Accuracy for Delta Measurements
e. Test all channels: Repeat substeps a through d for all four channels.
3. Reestablish the initial test hookup setup:
a. Hook up the testĆsignal source:
HSet the output of a DC calibration generator to 0 volts.
HMove the BNC T connector back to CH 1.
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
c. Modify the default settings:
HPress SHIFT; then ACQUIRE MENU.
HPress the mainĆmenu button Mode; then press the sideĆmenu
button Average 16.
HPress DISPLAY.
HPress the mainĆmenu button Graticule; then press the sideĆ
menu button Frame.
1Ć28
Performance Verification Procedures
Performance Tests
4. Confirm input channels are within limits for DC accuracy at maximum
offset and position: Do the following substeps Ċ test CH 1 first, skipping
substep a since CH 1 is already selected from step 3.
a. Select an unchecked channel:
HPress WAVEFORM OFF to remove the channel just confirmed
from the display.
HPress the frontĆpanel button that corresponds to the channel
you are to confirm.
HSet the generator output to 0 V.
HMove the test hookup to the channel you select.
b. Turn on the measurement Mean for the channel:
HPress MEASURE, then press the mainĆmenu button Select
Measrmnt for CHx.
HPress the side menu button more until the menu label Mean
appears in the side menu (its icon is shown at the left). Press the
sideĆmenu button Mean.
HPress CLEAR MENU.
Follow these rules to match this procedure to the model of the
oscilloscope under test:
Models TDS 540A, 544A OnlyĊUse Table 1Ć3 to test CH 1ĊCH 4;
ignore Table 1Ć4 AUX 1 & AUX 2 settings and limits.
Model TDS 520A, 524A OnlyĊUse Table 1Ć3 to test CH 1 and CH 2
only; use Table 1Ć4 to test AUX 1 and AUX 2 only.
c. Set its vertical scale: Set the vertical SCALE to one of the settings
listed in Table 1Ć3 (and Table 1Ć4 for the TDS 520A and 524A) that is
not yet checked. (Start with the first setting listed.)
HPress VERTICAL MENU. Press the mainĆmenu button Position.
HUse the keypad to set vertical position to -5 divisions (press
-5, then ENTER, on the keypad). The baseline level will move
off screen.
HPress the mainĆmenu button Offset.
HUse the keypad to set vertical offset to the positiveĆpolarity
setting listed in the table for the current vertical scale setting.
The baseline level will remain off screen.
HSet the generator to the level and polarity indicated in the table
for the vertical scale, position, and offset settings you have
made. The DC test level should appear on screen. (If it doesn't
return, the DC accuracy check is failed for the current vertical
scale setting of the current channel.)
e. Measure the test signal: Press CLEAR MENU. Read the measureĆ
ment results at the Mean measurement readout. See Figure 1Ć7.
1Ć30
Performance Verification Procedures
First set vertical and position offsets
to maximum (no input). Note gnd ref
indicator bounded onĆscreen for the
offset baseline below screen.
Second, input a DC
level equal to the offset
plus 3 divisions.
Third, turn on the
Measurement called
mean and read the
results here.
Figure 1Ć7:ăMeasurement of DC Accuracy at Maximum Offset and Position
Performance Tests
f.Check against limits:
HCHECK that the readout for the measurement Mean readout on
screen is within the limits listed for the current vertical scale and
position/offset/generator settings.
HRepeat step d, reversing the polarity of the position, offset, and
generator settings as is listed in the table.
HCHECK that the Mean measurement readout on screen is within
the limits listed for the current vertical scale setting and position/
offset/generator settings.
HRepeat substeps c through f until all vertical scale settings
settings listed in Table 1Ć3 (and Table 1Ć4 for the TDS 520A and
524A) are checked for the channel under test.
g. Test all channels: Repeat substeps a through f for all four channels.
5. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Then disconnect the cable from the generator output at the input
Equipment Required: One highĆfrequency leveled sine wave generator and
its leveling head (Item 11), plus two10X attenuators (Item 1).
Prerequisites: See page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
b. Modify the default settings:
HSet the horizontal SCALE to 50 ns.
HNow press SHIFT, then ACQUIRE MENU.
HPress the mainĆmenu button Mode; then press the sideĆmenu
button Average 16.
HPress TRIGGER MENU.
HPress the mainĆmenu button Coupling; then press the sideĆ
menu button Noise Rej.
HPress Measure. Now press the mainĆmenu button High-Low
Setup; then press the sideĆmenu button Min-Max.
c. Hook up the testĆsignal source: Connect, through its leveling head,
the sine wave output of a highĆfrequency leveled sine wave generaĆ
tor to CH 1. Set the output of the generator to a reference frequency
of 6ĂMHz. See Figure 1Ć8.
High
Frequency
Sine Wave
Generator
Output
Leveling Head
1Ć32
Figure 1Ć8:ăInitial Test Hookup
Performance Verification Procedures
Performance Tests
2. Confirm the input channels are within limits for analog bandwidth: Do the
following substeps Ċ test CH 1 first, skipping substepsa and b since
CH 1 is already set up for testing from step 1.
a. Select an unchecked channel:
HPress WAVEFORM OFF to remove the channel just confirmed
from display.
HPress the frontĆpanel button that corresponds to the channel
you are to confirm.
HMove the leveling head to the channel you select.
b. Match the trigger source to the channel selected:
HPress TRIGGER MENU.
HPress the mainĆmenu button Source.
HPress the sideĆmenu button that corresponds to the channel
selected.
c. Set its input impedance:
HPress VERTICAL MENU; then press the mainĆmenu button
Coupling.
HPress the sideĆmenu W button to toggle it to the 50 W setting.
d. Set the vertical scale: Set the vertical SCALE to one of the settings
listed in Table 1Ć5 not yet checked. (Start with the 100 mV setting.)
e. Display the test signal: Do the following subparts to first display the
reference signal and then the test signal.
HPress MEASURE; then press the mainĆmenu button Select
Measrmnt for CHx.
HNow press the side menu button more until the menu label
PkĆPk appears in the side menu (its icon is shown at the left).Press the sideĆmenu button PkĆPk.
HPress CLEAR MENU.
HSet the generator output so the CHx PkĆPk readout equals the
reference amplitude in Table 1Ć5 that corresponds to the vertical
scale set in substep d.
HPress the frontĆpanel button SET LEVEL TO 50% as necessary
to trigger a stable display.
f.Measure the test signal:
HIncrease the frequency of the generator output to the test freĆ
quency in Table 1Ć5 that corresponds to the vertical scale set in
substep d.
First, increase the reference
frequency to the test
frequency; then decrease the
horizontal scale.
Second, read the results from
the readout of measurement
PkĆPk.
HSet the horizontal SCALE to 1 ns. Press SET LEVEL TO 50% as
necessary.
HRead the results at the CHx PkĆPk readout, which will automatiĆ
cally measure the amplitude of the test signal. See Figure 1Ć9.
1Ć34
Figure 1Ć9:ăMeasurement of Analog Bandwidth
Performance Verification Procedures
Performance Tests
g. Check against limits:
HCHECK that the PkĆPk readout on screen is within the limits
listed in Table 1Ć5 for the current vertical scale setting.
HWhen finished checking, set the horizontal SCALE back to the
50 ns setting.
Checking each channel's bandwidth at all vertical scale settings is
time consuming and unnecessary. You may skip checking the
remaining vertical scale settings in Table 1Ć5 (that is, skip the followĆ
ing substep, h) if this digitizing oscilloscope has performed as
follows:
HPassed the 100 mV vertical scale setting just checked in this
procedure.
HPassed the Verify Internal Adjustment, Self Compensation, and
Diagnostics procedure found under Self Tests, on page 1Ć4.
NOTE
Passing the signal path compensation confirms the signal path for
all vertical scale settings for all channels. Passing the internal
diagnostics ensures that the factoryĆset adjustment constants that
control the bandwidth for each vertical scale setting have not
changed.
h. Check remaining vertical scale settings against limits (optional):
HIf desired, finish checking the remaining vertical scale settings
for the channel under test by repeating substeps d through g for
each of the remaining scale settings settings listed in Table 1Ć5
for the channel under test.
HWhen doing substep e, skip the subparts that turn on the CHx
PkĆPk measurement until you check a new channel.
HInstall/remove 10X attenuators between the generator leveling
head and the channel input as is needed to obtain the six diviĆ
sion reference signals listed in the table.
i.Test all channels: Repeat substeps a through g for all four channels.
3. Disconnect the hookup: Disconnect the test hook up from the input
connector of the channel last tested.
Equipment Required: One mediumĆfrequency leveled sineĆwave generator
(Item 10), one precision, 50 W, coaxial cable (Item 4), one 50 W terminator
(Item 3), and a dualĆinputĆcoupler (Item 7).
Prerequisites: See page 1Ć15.
Procedure:
DO NOT use the vertical position knob to reposition any channel
while doing this check. To do so invalidates the test.
1. Install the test hookup and preset the instrument controls:
a. Initialize the front panel;
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
b. Modify the initialized frontĆpanel control settings:
HDo not adjust the vertical position of any channel during this
procedure.
HSet the horizontal SCALE to 500 ps.
HPress SHIFT; then ACQUIRE MENU.
HPress the mainĆmenu button Mode, and then press the sideĆmeĆ
nu button Average 16.
c. Hook up the testĆsignal source:
HConnect, through a 50 W precision coaxial cable, followed by a
50 W termination, the sine wave output of a mediumĆfrequency
sine wave generator to a dualĆinput coupler. See Figure 1Ć10.
Medium
Frequency
Sine Wave
Generator
Output
1Ć36
50 W Cable
50 W Terminator
Figure 1Ć10:ăInitial Test Hookup
Performance Verification Procedures
Dual Input Coupler
Performance Tests
HConnect the coupler to both CH 1 and CH 2.
2. Confirm CH 1 through CH 4 (CH 2 for 520A and 524A) are within limits for
channel delay:
a. Set up the generator: Set the generator frequency to 250 MHz and
the amplitude for about five divisions in CH 1.
Hint: as you are adjusting the generator amplitude, push SET LEVĆEL TO 50% frequently to speed up the updating of the waveform
amplitude on screen.
TDS 520A and 524A only: Press CH 2; then skip to substep e and
continue this check. If testing a TDS 540A or 544A model, continue
with the next substep, b.
b. Save a CH 2 waveform: Press CH 2; then press save/recall WAVEĆ
FORM. Now, press the mainĆmenu button Save Wfm Ch2; then
press the sideĆmenu button To Ref 2.
c. Save CH 3 waveform: Move the coupler from CH 2 to CH 3, so that
CH 1 and CH 3 are driven. Press CH 3; then press the sideĆmenu
button To Ref 3.
d.Display all test signals:
HPress WAVEFORM OFF twice to remove CH 2 and CH 3 from
the display.
HMove the coupler from CH 3 to CH 4, so that CH 1 and CH 4 are
driven. Press CH 4.
HNow, press the frontĆpanel button MORE. Press the mainĆmenu
buttons Ref 2 and Ref 3.
e. Measure the test signal:
HLocate the point on the rising edge of the leftĆmost waveform
where it crosses the center horizontal graticule line. This is the
time reference point for this waveform. Note the corresponding
time reference point for rightĆmost waveform. See Figure 1Ć11.
HPress CURSOR; then press the sideĆmenu button V Bars.
HPress CLEAR MENU.
HRotate the General Purpose knob to align one cursor to the time
reference point of the leftĆmost waveform edge and the other
cursor to the time reference point of the rightĆmost waveform
edge. (Press SELECT to switch between the two cursors.) See
Figure 1Ć11.
HRead the measurement results at the D: cursor readout, not the
@: readout on screen.
Second, identify the time reference
points of those waveforms.
Third, turn on the cursor and align
the V bar cursors to the time
reference points.
Fourth, read the results
here.
Figure 1Ć11:ăMeasurement of Channel Delay
f.Check against limits: CHECK that the cursor readout on screen is
v250 ps.
3. Disconnect the hookup: Disconnect the cable from the generator output
at the input connectors of the channels.
1Ć38
Performance Verification Procedures
Performance Tests
Time Base System
Checks
These procedures check those characteristics that relate to the Main and
Delayed time base system and are listed as checked under WarrantedCharacteristics in Section 2, Specifications.
Check Accuracy for LongĆTerm Sample Rate, Delay Time,
and Delta Time Measurements
Equipment Required: One timeĆmark generator (Item 12) and one precision
coaxial cable (Item 4).
Prerequisites: See page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Hook up the testĆsignal source: Connect, through a 50 W precision
coaxial cable, the timeĆmark output of a timeĆmark generator to
CH 1. Set the output of the generator for 10 ms markers. See FigĆ
ure 1Ć12.
TimeĆMark
Generator
Output
50 W Coaxial Cables
Figure 1Ć12:ăInitial Test Hookup
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
c. Modify the initialized frontĆpanel control settings:
HSet the vertical SCALE to 500 mV.
HPress VERTICAL MENU; then press the mainĆmenu button
Coupling. Press the sideĆmenu W button to 50 W.
HPress SET LEVEL TO 50%.
HUse the vertical POSITION knob to center the test signal on
These procedures check those characteristics that relate to the Main and
Delayed trigger systems and are listed as checked under Warranted CharacĆteristics in Section 2, Specifications.
Check Accuracy (Time) for PulseĆGlitch or PulseĆWidth
Triggering
Equipment Required: One mediumĆfrequency leveled sine wave generator
(Item 10), one 10X attenuator (Item 1), and one precision, 50 W, coaxial
cable (Item 4).
Prerequisites: See page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Initialize the instrument:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
b. Modify the default setup:
HPress vertical MENU.
HPress the mainĆmenu button Coupling; then press the sideĆmeĆ
nu W button to select 50W coupling.
HSet the horizontal SCALE to 10 ns.
c. Hook up the testĆsignal source: Connect, through a 50 W precision
coaxial cable, followed by a 10X attenuator, the output of a mediumĆ
frequency leveled sine wave generator (Item 10) to CH 1. See FigĆ
ure 1Ć14.
Medium
Frequency
Sine Wave
Generator
Output
50 W Coaxial Cable
10X Attenuator
Figure 1Ć14:ăInitial Test Hookup
1Ć42
Performance Verification Procedures
Performance Tests
2. Confirm the trigger system is within timeĆaccuracy limits for pulseĆglitch
or pulseĆwidth triggering (Horizontal Scale v1 ms):
a. Display the test signal: Set the output of the sine wave generator for
a 100ĂMHz, fiveĆdivision sine wave on screen. Press SET LEVEL TO50%.
b. Set the trigger mode: Press TRIGGER MENU. Now press the mainĆ
menu button Mode & Holdoff; then the sideĆmenu button Normal.
c. Set upper and lower limits that ensures triggering:
HPress the mainĆmenu button Type; then repeatedly press the
same button until Pulse is highlighted in the menu that pops up.
HPress the mainĆmenu button Class; then repeatedly press the
same button until Width is highlighted in the menu that pops up.
HPress the mainĆmenu button Trig When; then press the sideĆmeĆ
nu button Within Limits.
HPress the sideĆmenu button Upper Limit. Use the keyboard to
set the upper limit to 10 ns: press 10; then SHIFT; then n; thenENTER.
HPress the sideĆmenu button Lower Limit. Use the keypad to set
the lower limit to 2 ns.
d. Check against limits:
HPress SET LEVEL TO 50%.
HWhile doing the following subparts, monitor the display (it will
stop acquiring) and the frontĆpanel light TRIG (it will extinguish)
to determine when triggering is lost.
HUse the general purpose knob to increase the Lower Limit
readout until triggering is lost.
HCHECK that the Lower Limit readout is within 2.5 ns to 7.5 ns,
inclusive.
HUse the keypad to return the Lower Limit to 2 ns and reestabĆ
lish triggering.
HPress the sideĆmenu button Upper Limit; then use the general
purpose knob to slowly decrease the the Upper Limit
until triggering is lost.
HCHECK that the Upper Limit readout is within 2.5 ns to 7.5 ns,
Figure 1Ć15:ăMeasurement of Time Accuracy for Pulse and Glitch Triggering
stops.
3. Confirm the trigger system is within timeĆaccuracy limits for pulseĆglitch
or pulseĆwidth triggering (horizontal scale >1 ms):
a. Set upper and lower limits that ensure triggering at 250 kHz:
HPress the sideĆmenu button Upper Limit. Use the keyboard to
set the upper limit to 4 ms.
HPress the sideĆmenu button Lower Limit. Use the keypad to set
the lower limit to 500 ns.
b. Display the test signal:
HSet the horizontal SCALE to 5 ms.
HSet the output of the sineĆwave generator for a 250 kHz, fiveĆdiviĆ
sion sine wave on screen. Set the vertical SCALE to 20 mV (the
waveform will overdrive the display).
HPress SET LEVEL TO 50%.
c. Check against limits: Do the following subparts in the order listed.
HUse the general purpose knob to increase Lower Limit readout
until triggering is lost.
HCHECK that the Lower Limit readout is within 1 msto3 ms,
inclusive.
1Ć44
HUse the keypad to return the Lower Limit to 500 ns and reesĆ
tablish triggering.
Performance Verification Procedures
Performance Tests
HPress the sideĆmenu button Upper Limit; then use the general
purpose knob to slowly decrease the the Upper Limit readout
until triggering is lost.
HCHECK that the Upper Limit readout is within 1 msto3 ms,
inclusive.
4. Disconnect the hookup: Disconnect the cable from the generator output
at the input connector of CH 1.
Check Accuracy, TriggerĆlevel or Threshold, DC Coupled
Equipment Required: One DC calibration generator (Item 8), one BNC T
connector (Item 6), and two precision, 50 W, coaxial cables (Item 4).
Prerequisites: The oscilloscope must meet the prerequisites listed on
page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Hook up the testĆsignal source:
Dual Banana to
BNC Adapter
HSet the output of a DC calibration generator to 0 volts.
HConnect the output of a DC calibration generator through a
HConnect the Sense output of the generator, through a second
DC Calibrator
Output Sense
dualĆbanana connector followed by a 50 W precision coaxial
cable to one side of a BNC T connector. See Figure 1Ć16.
dualĆbanana connector followed by a 50 W precision coaxial
cable, to other side of the BNC T connector. Now connect the
BNC T connector to CH 1. See Figure 1Ć16.
HCHECK that the Level readout in the side menu is within 9.940 V
to 10.060 V, inclusive.
4. Disconnect the hookup:
a. First set the output of the DC calibration generator to 0 volts.
b. Then disconnect the cable from the generator output at the input
connector of CH 1.
Sensitivity, Edge Trigger, DC Coupled
Equipment Required: One mediumĆfrequency leveled sine wave generator
(Item 10), one highĆfrequency leveled sine wave generator (Item 11), one
precision 50ĂW coaxial cable (Item 4), and one 10X attenuator (Item 1).
When checking the TDS 540A and 544A, a BNC T connector (Item 6), a 5X
attenuator (Item 2), and a second precision 50ĂW coaxial cable (Item 4) are
also required.
Prerequisites: See page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
HPress the sideĆmenu button OK Confirm Factory Init.
b. Modify the initialized frontĆpanel control settings:
HSet the horizontal SCALE for the M (main) time base to 20 ns.
HPress HORIZONTAL MENU; then press the mainĆmenu button
Time Base.
HPress the sideĆmenu button Delayed Only; then the sideĆmenu
button Delayed Triggerable.
HSet the horizontal SCALE for the D (delayed) time base to 20 ns;
then press the sideĆmenu button Main Only.
HPress TRIGGER MENU; then press the mainĆmenu button
Mode & Holdoff. Now press the sideĆmenu button Normal.
HPress VERTICAL MENU; then press the mainĆmenu button
Coupling. Now press the sideĆmenu W button and select the
50 W setting.
1Ć48
HPress SHIFT; then press ACQUIRE MENU. Now press the
mainĆmenu button Mode; then the sideĆmenu Average 16
button.
Performance Verification Procedures
Performance Tests
c. Hook up the testĆsignal source:
Medium
Frequency
Sine Wave
Generator
Output
50 W Coaxial Cables
Figure 1Ć18:ăInitial Test HookupĊTDS 520A or 524A Only
HTDS 520A, 524A only: Connect, through a 50 W precision coaxĆ
ial cable, the signal output of a mediumĆfrequency sine wave
generator to CH 1. See Figure 1Ć18.
To AUX TRIG INPUT
on Rear Panel
Medium
Frequency
Sine Wave
Generator
Output
50 W Coaxial Cables
Figure 1Ć19:ăInitial Test HookupĊTDS 540A or 544A Only
HTDS 540A or 544A only: Connect the signal output of a mediumĆ
frequency sine wave generator to a BNC T connector. Connect
one output of the T connector to CH 1 through a 50 W precision
coaxial cable; connect the other output of the T connector to the
AUX TRIG INPUT at the rear panel. See Figure 1Ć19.
2. Confirm Main and Delayed trigger systems are within sensitivity limits
HPress the mainĆmenu button HighĆLow Setup; then press the
sideĆmenu button MinĆMax.
HPress the mainĆmenu button Select Measrmnt for Ch1.
HPress the sideĆmenu button -more- until Amplitude appears
in the side menu (its icon is shown at the left). Press the sideĆ
menu button Amplitude.
HPress SET LEVEL TO 50%.
HPress CLEAR MENU.
HSet the test signal amplitude for about three divisions on screen.
Now fine adjust the generator output until the CH 1 Amplitude
readout indicates the amplitude is 350 mV. (Readout may fluctuĆ
ate around 350 mV.)
HDisconnect the 50 W precision coaxial cable at CH 1 and reconĆ
nect it to CH 1 through a 10X attenuator.
b. Check for Main trigger system for stable triggering at limits:
HRead the following definition: A stable trigger is one that is
consistent; that is, one that results in a uniform, regular display
triggered on the selected slope (positive or negative). This
display should not have its trigger point switching between
opposite slopes, nor should it roll" across the screen. At horiĆ
zontal scale settings of 2Ăms/division and faster, TRIG'D will
remain constantly lit. It will flash for slower settings.
HPress TRIGGER MENU; then press the mainĆmenu button
Slope.
HPress SET LEVEL TO 50%. CHECK that a stable trigger is
obtained for the test waveform on both the positive and negative
slopes. (Use the side menu to switch between trigger slopes;
use the TRIGGER LEVEL knob to stabilize the trigger if reĆ
quired.) See Figure 1Ć20.
1Ć50
Performance Verification Procedures
First, set a signal with an
amplitude at the minimum
trigger sensitivity.
Second, check for a stable
trigger at both the positive
and negative slope settings.
Figure 1Ć20:ăMeasurement of Trigger Sensitivity
Performance Tests
HLeave the Main trigger system triggered on the positive slope of
the waveform before continuing to the next step. (The Main
trigger system must be triggered to check the delayed trigger
system in the next step.)
c. Check delayed trigger system for stable triggering at limits: Do the
following subparts in the order listed.
HPress HORIZONTAL MENU; then press the mainĆmenu button
Time Base. Now press the sideĆmenu button Delayed Only.
HPress SHIFT; then press DELAYED TRIG. Press the mainĆmenu
button Level.
HPress the sideĆmenu button SET TO 50%. CHECK that a stable
trigger is obtained for the test waveform for both the positive and
negative slopes of the waveform. (Use the General Purpose
knob to stabilize the trigger if required.) Press the mainĆmenu
button Slope; then use the side menu to switch between trigger
slopes.
HLeave the delayed trigger system triggered on the positive slope
of the waveform before continuing to the next step. Also, return
to the main time base: Press HORIZONTAL MENU; then press
the mainĆmenu button Time Base. Now press the sideĆmenu
button Main Only.
TDS 520A or 524A only: Skip to step 4 since the TDS 520A and
524A are not equipped with an AUX Trigger input. If testing the TDS
540A or 544A, continue with step 3.
HRemove the 10X attenuator and reconnect the cable to CH 1.
HSet the test signal amplitude for about 2.5 divisions on screen.
HNow fine adjust the generator output until the CH 1 Amplitude
readout indicates the amplitude is 250 mV. (Readout may fluctuĆ
ate around 250 mV.)
b. Check the AUX trigger source for stable triggering at limits: Do the
following in the order listed.
HUse the definition for stable trigger from step 2.
HPress TRIGGER MENU; then press the mainĆmenu button
Source.
HPress the sideĆmenu button -more- until the sideĆmenu label
Auxiliary appears; then press Auxiliary.
HPress SET LEVEL TO 50%. CHECK that a stable trigger is
obtained for the test waveform on both the positive and negative
slopes. Press the mainĆmenu button Slope; then use the side
menu to switch between trigger slopes. Use the TRIGGERLEVEL knob to stabilize the trigger if required.
HLeave the Main trigger system triggered on the positive slope of
the waveform before proceeding to the next check.
HPress the mainĆmenu button Source; then press the sideĆmenu
button -more- until CH 1 appears. Press CH 1.
4. Confirm that the Main and Delayed trigger systems are within sensitivity
limits (500 MHz):
a. Hook up the testĆsignal source: Disconnect the hookup installed in
step 1. Connect, through its leveling head, the signal output of a
highĆfrequency leveled sineĆwave generator to CH 1.
b. Set the Main and Delayed Horizontal Scales:
HSet the horizontal SCALE to 500 ps for the M (Main) time base.
HPress HORIZONTAL MENU. Now press the mainĆmenu button
Time base; then press the sideĆmenu button Delayed TriggerĆ
able.
HPress the sideĆmenu button Delayed Only.
HSet the horizontal SCALE to 500 ps for the D (Delayed) time
base. Press the sideĆmenu button Main Only.
c. Display the test signal:
1Ć52
H
Set the generator frequency to 500 MHz.
Performance Verification Procedures
Performance Tests
HSet the test signal amplitude for about five divisions on screen.
Now fine adjust the generator output until the CH 1 Amplitude
readout indicates the amplitude is 500 mV. (Readout may fluctuĆ
ate around 500 mV.)
HDisconnect the leveling head at CH 1 and reconnect it to CH 1
through a 5X attenuator.
d. Repeat step 2, substeps b and c only, since only the 500 MHz
frequency is to be checked here.
5. Confirm that the Main and Delayed trigger systems couple trigger signalsfrom all channels: Doing the procedure Check Analog Bandwidth, which
begins on page 1Ć32, checks coupling. If you have not done that proceĆ
dure, do so after finishing this procedure. See the following note.
NOTE
Steps 1 through 4 confirmed trigger sensitivity for the Main and
Delayed triggering systems using the CH 1 input. Doing the proceĆ
dure Check Analog Bandwidth ensures that trigger signals are
coupled from all four channels.
When checking delayed triggering sensitivity at 500 MHz, the
waveform record may have some missing interpolated record
points. The waveform is still stably triggered. (See definition of a
stable trigger earlier in this procedure.)
6. Disconnect the hookup: Disconnect the cable from the generator output
at the input connector of the channel last tested.
The procedure that follows checks those characteristics of the output sigĆ
nals that are listed as checked under Warranted Characteristics in Section 2,Specifications. The oscilloscope outputs these signals at its front and rear
panels.
Check Outputs Ċ CH 3 and Main and Delayed Trigger
ą(TDS 540A and 544A only)
Equipment Required: Two 50 W precision cables (Item 4), and one calibraĆ
tion generator (Item 9).
Prerequisites: See page 1Ć15. Also, this Digitizing Oscilloscope must have
passed Check Accuracy Ċ LongĆTerm Sample Rate, Delay time, Time MeaĆ
surement on page 1Ć39 and Check Accuracy for DC Gain and Voltage MeaĆ
surements on page 1Ć25.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Hook up testĆsignal source 1:
HConnect the standard amplitude output of a calibration generaĆ
tor through a 50 W precision coaxial cable to CH 3. See Figure
1Ć21.
HSet the output of the calibration generator to 0.500 V.
b. Hook up testĆsignal source 2: Connect the Main Trigger Out at the
rear panel to CH 2 through a 50 W precision cable. See Figure 1Ć21.
To Main
Trigger Out
Calibration
Generator
Output
50 W Coaxial Cables
Figure 1Ć21:ăInitial Test Hookup
c. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the mainĆmenu button Recall Factory Setup.
1Ć54
HPress the sideĆmenu button OK Confirm Factory Init.
Performance Verification Procedures
Performance Tests
d. Modify the initialized frontĆpanel control settings:
HSet the horizontal SCALE to 200 ms.
HPress SHIFT; then ACQUIRE MENU.
HPress the mainĆmenu button Mode. Then press the sideĆmenu
button Hi Res.
2. Confirm Main and Delayed Trigger outputs are within limits for logic
levels:
a. Display the test signal:
HPress WAVEFORM OFF to turn off CH 1; then press CH 2 to
display that channel. Set the vertical SCALE to1V.
HPress TRIGGER MENU.
HPress the mainĆmenu button Source; then press the sideĆmenu
button CH 3. Press SET LEVEL TO 50%.
HUse the vertical POSITION knob to center the display on
screen.
First, turn on the
measurements
high and low.
Second, read the
measurement
results here.
b. Measure logic levels:
HPress MEASURE; then press the mainĆmenu button Select
Measrmnt for Ch2.
HRepeatedly press the sideĆmenu button -more- until High and
Low appear in the side menu (their icons are shown at the left).Press both sideĆmenu buttons High and Low. See Figure 1Ć22.
Figure 1Ć22:ăMeasurement of Main Trigger Out Limits
HCHECK that the Ch2 High readout is w2.5 volts and that the
Ch2 Low readout is v700 mV.
HPress VERTICAL MENU; then press the mainĆmenu button
Coupling. Now press the sideĆmenu button W to toggle it to the
50 W setting.
HCHECK that the Ch2 High readout is w1.0 volt and that the Ch2
Low readout v250 mV.
d. Check Delayed Trigger output against limits:
HMove the precision 50 W cable from the Main Trigger Output
BNC to the Delayed Trigger Output BNC.
HCHECK that the Ch2 High readout is w1.0 volt and that the Ch2
Low readout v250 mV.
HPress the sideĆmenu button W select the 1 MW setting.
HPress CLEAR MENU.
HCHECK that the Ch2 High readout is w2.5 volts and that the
Ch2 Low readout is v700 mV.
3. Confirm CH 3 output is within limits for gain:
a. Measure gain:
HMove the precision 50 W cable from the rearĆpanel DELAYED
TRIGGER OUTPUT BNC to the rearĆpanel SIGNAL OUTPUT
BNC.
HPush SHIFT. Then push DELAYED TRIG.
HPress the mainĆmenu button Source. Then press the sideĆmenu
button Ch3.
HPush HORIZONTAL MENU.
HPush the mainĆmenu button Time Base. Then press the sideĆ
menu button Delayed Triggerable.
HSet vertical SCALE to 100 mV.
HPress MEASURE; then press the mainĆmenu button Select
Measrmnt for Ch2.
HRepeatedly press the sideĆmenu button -more- until PkĆPk
appears in the side menu (its icon is shown at the left). Press the
sideĆmenu button PkĆPk.
HPress CLEAR MENU.
1Ć56
Performance Verification Procedures
Performance Tests
b. Check against limits:
HCHECK that the readout Ch2 PkĆPk is between 90ĂmV and
110 mV, inclusive.
HPress VERTICAL MENU; then press the sideĆmenu button W to
toggle to the 50 W setting.
HSet vertical SCALE to 10 mV; then press CLEAR MENU.
HCHECK that the readout Ch2 PkĆPk is between 45 mV and
55 mV, inclusive.
4. Disconnect the hookup: Disconnect the cable from the generator output
at the input connector of the channel last tested.
Check Probe Compensator Outputs
One female BNC to clip adapter (Item 15), two dualĆbanana connectors
(Item 5), one BNC T connector (Item 6), two 50 W precision cables (Item 4),
and one DC calibration generator (Item 8).
Prerequisites: See page 1Ć15. Also, this Digitizing Oscilloscope must have
passed Check Accuracy Ċ LongĆTerm Sample Rate, Delay time, Time MeaĆ
surement on page 1Ć39 and Check Accuracy for DC Gain and Voltage MeaĆ
surements on page 1Ć25.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Hook up testĆsignal:
HConnect CH 1 to PROBE COMPENSATION SIGNAL and to
PROBE COMPENSATION GND through a precision coaxial
HPress the sideĆmenu button OK Confirm Factory Init.
c. Modify the initialized frontĆpanel control settings:
HSet the vertical SCALE to 100 mV as required.
HSet the horizontal SCALE to 200 ms.
HPress Set Level to 50% and use the VERTICAL POSITION
knob to center the display on screen.
HPress SHIFT; then ACQUIRE MENU.
HPress the mainĆmenu button Mode; then press the sideĆmenu
Average button.
HSelect 128 averages. On the keypad, type 128; then press
ENTER.
2. Confirm that the Probe Compensator signal is within limits for frequency:
a. Measure the frequency of the probe compensation signal:
HPress MEASURE; then press the mainĆmenu button Select
Measrmnt for Ch1.
HRepeatedly press the sideĆmenu button -more- until FrequenĆ
cy appears in the side menu (its icon is shown at the left). Pressthe sideĆmenu button Frequency.
HPress CLEAR MENU to remove the menus from the display. See
Figure 1Ć24.
1Ć58
Figure 1Ć24:ăMeasurement of Probe Compensator Frequency
Performance Verification Procedures
Performance Tests
b. Check against limits: CHECK that the CH 1 Freq readout is within
950 Hz to 1.050 kHz, inclusive.
3. Confirm that the Probe Compensator signal is within limits for amplitude:
a. Save the probe compensation signal in reference memory:
HPress SAVE/RECALL WAVEFORM; then press the mainĆmenu
button Save Wfm Ch 1.
HPress the sideĆmenu button to Ref 1 to save the probe compenĆ
sation signal in reference 1.
HDisconnect the cable from CH 1 and the clips from the probe
compensation terminals.
HPress MORE; then press the mainĆmenu button Ref 1 to disĆ
played the stored signal.
HPress CH 1.
b. Hook up the DC standard source:
HSet the output of a DC calibration generator to 0 volts.
Dual Banana to
BNC Adapters
HConnect the output of a DC calibration generator through a
dualĆbanana connector followed by a 50 W precision coaxial
cable to one side of a BNC T connector. See Figure 1Ć25.
HConnect the Sense output of the generator through a second
dualĆbanana connector followed by a 50 W precision coaxial
cable to the other side of the BNC T connector. Now connect the
BNC T connector to CH 1. See Figure 1Ć25.
DC Calibrator
Output Sense
HI
LO
50 W Coaxial Cables
BNC T
Connector
Figure 1Ć25:ăSubsequent Test Hookup
c. Measure amplitude of the probe compensation signal:
HPress SHIFT; then ACQUIRE MENU. Then use the keypad to
a. Setup equipment for Jitter Test (See Figure 1Ć27).
HConnect one of the rear panel composite outputs marked
COMPST on the TSG121 through a 75 W cable and a 75 W
terminator to the oscilloscope CH1 input.
HPress the PAL signal source 100% FIELD control (the fourth
TSG121 frontĆpanel button from the left).
PAL Signal
Source
TSG121
COMPST
75 W Cable
75 W
Terminator
Figure 1Ć27:ăJitter Test Hookup
b. CHECK that the oscilloscope lights up its front panel TRIG'D LED
and it displays the waveform on screen (See Figure 1Ć28).
1Ć62
Figure 1Ć28:ăJitter Test Displayed Waveform
Performance Verification Procedures
Performance Tests
HCONFIRM that the TRIG'D LED is lit and the waveform is disĆ
played on screen.
c. Press SHIFT; then ACQUIRE MENU.
d. Press the mainĆmenu Mode.
e. Select the sideĆmenu Average. It should be already set to 16.
f.Press the mainĆmenu Create Limit Test Template.
g. Press the sideĆmenu V Limit.
h. Use the keypad to set V Limit to 100 mdiv (press 100, SHIFT, m,
then ENTER)
i.Press the sideĆmenu OK Store Template.
j.Press MORE.
k. Press the mainĆmenu Ref1.
l.Press CH1.
m. Press SHIFT; then ACQUIRE MENU.
n. Press the mainĆmenu Limit Test Setup.
o. Toggle the sideĆmenu Limit Test to ON.
p. Toggle the sideĆmenu Ring Bell if Condition Met to ON.
q. Press the mainĆmenu Mode.
r.Press the sideĆmenu Envelope.
s. Use the keypad to set envelope to use 100 acquisitions (press 100,
then ENTER).
t.Press the mainĆmenu Stop After button.
u. Press the sideĆmenu Single Acquisition Sequence.
v.CONFIRM that the oscilloscope successfully makes 100 acquisiĆ
tions. If not successful, the oscilloscope bell will ring. When the word
Run in the top left corner of the display changes to STOP, the test is
complete (See Figure 1Ć29).
x.Toggle the sideĆmenu Ring Bell if Condition Met to OFF.
y. Toggle the sideĆmenu Limit Test to OFF.
4. Check Triggered Signal Range.
Setup oscilloscope for Triggered Signal Test.
a. Press MORE.
b. Press WAVEFORM OFF.
c. Press HORIZONTAL MENU.
d. Use the keypad to set the horizontal scale timeĆperĆdivision (Horiz
Scale (/div))to50ms (press 50, SHIFT, m, then ENTER).
e. Press SHIFT; then ACQUIRE MENU.
f.Press the mainĆmenu Stop After.
g. Press the sideĆmenu RUN/STOP button only.
h. Press the mainĆmenu Mode.
i.Press the sideĆmenu Sample.
j.Press RUN/STOP.
1Ć64
k. Press VERTICAL MENU.
l.Use the keypad to set fine scale to 300 mV (press 300, SHIFT, m,
then ENTER).
Performance Verification Procedures
Positive pulse
Negative pulses
Performance Tests
m. CONFIRM that the TRIG'D LED stays on and that the waveform on
screen is stable. ie; does not move horizontally or vertically. Also,
CONFIRM that the waveform on the screen has one positive pulse
and a number of negative pulses (See Figure 1Ć30).
Figure 1Ć30:ăTriggered Signal Range Test - 300 mV
n. Use the keypad to set the fine scale to 75 mV (press 75, SHIFT, m,
then ENTER).
o. CONFIRM that the TRIG'D LED stays lit and that the waveform on
screen is stable. ie; does not move horizontally or vertically. Also,
CONFIRM that the waveform on the screen has one positive pulse
and a number of negative pulses (See Figure 1Ć31).
p. Disconnect all test equipment (TSG121) from the digitizing oscilloĆ
scope.
5. Check 60 Hz Rejection.
a. Setup oscilloscope for 60 Hz Rejection Test.
HUse the keypad to set the Ch1 Fine Scale to 282 mV (press 282,
SHIFT m, then ENTER).
HPress WAVEFORM OFF.
HPress CH2.
HPress VERTICAL MENU.
HUse the keypad set the fine scale to 2 V (press 2, then ENTER).
HPress HORIZONTAL MENU.
HUse the keypad to set the horizontal scale timeĆperĆdivision
(Horiz Scale (/div)) to 5 ms (press 5, SHIFT, m, then ENTER).
b. Setup 60 Hz signal generator (SG 502).
HConnect the output of the SG 502 to the CH2 input through a
50 W cable (See Figure 1Ć32).
1Ć66
Performance Verification Procedures
Performance Tests
Signal
Generator
SG 502
50 W Cable
Figure 1Ć32:ă60 Hz Rejection Test Hookup
HAdjust the SG 502 for three vertical divisions of 60 Hz signal
(See Figure 1Ć33). The signal will not be triggered. That is, it will
run free.
Figure 1Ć33:ă60 Hz Rejection Test Setup Signal
c. Check 60 Hz rejection.
HUse the keypad to set the the horizontal scale timeĆperĆdivision
(Horiz Scale (/div)) to 50 ms (press 50, SHIFT, m, then ENTER).
HReconnect the output of the signal generator (SG 502). Connect
the PAL signal source's composite signal connector (labelled
COMPST on the TSG 121) to a 75 W cable and a 75 W terminaĆ
tor. Connect both signals to the CH1 input through a BNC T
(See Figure 1Ć34).
HUse the keypad to set fine scale to 500 mV (press 500, SHIFT,
m, then ENTER).
HConnect another the PAL signal source's composite signal
connector (labelled COMPST on the TSG 121) through a 75 W
cable and a 75 W terminator to the CH2 input (See Figure 1Ć34).
Signal
Generator
SG 502
PAL Signal
Source
TSG121
COMPST
COMPST
75 W Cable
75 W Cable
50 W Cable
75 W
Terminator
BNC T
Connector
75 W
Terminator
Figure 1Ć34:ăSubsequent 60 Hz Rejection Test Hookup
HCONFIRM that the TRIG'D LED stays on and that the waveform
on screen is stable. Stable means the waveform does not move
horizontally or vertically. Also, confirm that the waveform on the
screen has one positive pulse and a number of negative pulses
(See Figure 1Ć35).
1Ć68
Performance Verification Procedures
Performance Tests
Figure 1Ć35:ă60 Hz Rejection Test Result
HDisconnect all test equipment from the digitizing oscilloscope.
6. Check Line Count Accuracy.
a. Setup oscilloscope for Line Count Accuracy Test.
HPress WAVEFORM OFF.
HPress CH1.
HPress HORIZONTAL MENU.
HPress the mainĆmenu Record Length.
HPress, if needed, the sideĆmenu -more- 1 of 2.
HPress the sideĆmenu 5000 points in 100divs.
HPress the mainĆmenu Horiz Scale (/div).
HUse the keypad to set the horizontal scale to 200 ns (press 200,
SHIFT, n, then ENTER).
b. Check Line Count Accuracy.
HConnect a composite output signal from the PAL signal source
(on the TSG 121 this refers to the signal at the rear labelled
COMPST) to the CH1 input through a 75 W cable and a 75 W
terminator (See Figure 1Ć36).
HTurn the pulse generator PULSE DURATION variable control to
adjust the negative pulse so the oscilloscope CH1 - Width
measurement displays 400ns +/-10 ns.
HTurn the HORIZONTAL SCALE knob to set the oscilloscope
time base to 5 ms/div.
HTurn the pulse generator PERIOD variable control to adjust the
period until the oscilloscope CH1 Period measurement reads
21.000ms -25/+50 ns (See Figure 1Ć41). Read note shown
below.
NOTE
The pulse duration and period adjustments are critical in making
this measurement. If the pulse duration and/or the duty cycle are
not stable, the FLEXFMT function may not function. You must take
care when making these adjustments.
HDisconnect all test equipment (TSG 121) from the oscilloscope.
HPress save/recall SETUP, press the mainĆmenu button Recall
Factory Setup, and the sideĆmenu OK Confirm Factory Init.
1Ć76
Performance Verification Procedures
Specifications
Specifications
This subsection begins with a general description of the traits of the
TDS 520A, 524A, 540A and 544A Digitizing Oscilloscopes. Three subsecĆ
tions follow, one for each of three classes of traits: nominal traits, warĆranted characteristics, and typical characteristics.
General Product
Description
The Tektronix TDS 520A, 524A, 540A and 544A Digitizing Oscilloscopes are
portable, fourĆchannel instruments suitable for use in a variety of test and
measurement applications and systems. Key features include:
H500 MHz maximum analog bandwidth.
H1 Gigasample/second maximum digitizing rate (TDS 540A and 544A);
500 Megasamples/second maximum digitizing rate (TDS 520A and
524A).
HFourĆchannel acquisition Ċ the TDS 540A and 544A offer four fullĆfeaĆ
tured channels; the TDS 520A and 524A offer two fullĆfeatured channels
and two channels with limited vertical scale selections: 100ĂmV, 1ĂV, and
10ĂV.
HExtensive triggering capabilities: such as edge, logic, and glitch. Video
trigger (Option 05) is also available. The video trigger modes are NTSC,
PAL, SECAM, HDTV, and FlexFormat
HWaveform Math Ċ Invert a single waveform and add, subtract, and
multiply two waveforms. On the TDS 524A, 544A, and other TDS 500A
equipped with option 2F, integrate or differentiate a single waveform or
perform an FFT (fast fourier transform) on a waveform to display its
magnitude or phase versus its frequency.
HEightĆbit digitizers.
TM
(user definable format).
HUp to 15,000Ćpoint record length per channel (50,000Ćpoint with option
1M).
HFull GPIB software programmability. Hardcopy output using GPIB,
RSĆ232, or Centronics ports (RSĆ232 and Centronics is standard on the
TDS 524A and 544A and optional, as option 13, on the TDS 520A and
540A).
HComplete measurement and documentation capability.
HIntuitive graphic icon operation blended with the familiarity of traditional
horizontal and vertical knobs.
HOnĆline help at the touch of a button.
HA full complement of advanced functions, like continuouslyĆupdated
measurements, results and local pass/fail decision making.
HSpecialized display modes, such as variable persistence (with color
coding on the TDS 524A and 544A), dot or vector mode, sin(x)/x or
linear display filters, and, on the TDS 524A and 544A, user selectable
color palettes. The Fit to Screen" feature compresses the entire
waveform record to fit on the screen.
User Interface
HA complement of advanced acquisition modes such as peakĆdetect,
highĆresolution, sample, envelope, and average. The FastFrame
feature acquires waveforms in rapid succession with a minimum of dead
time between acquisitions. FastFrame
50,000 frames per second.
HA unique graphical user interface (GUI), an onĆboard help mode, and a
logical frontĆpanel layout which combine to deliver a new standard in
usability.
HVGA output for driving remote monitors.
HA 1.44 Mbyte, DOS 3.3 or later, floppy disk drive for saving waveforms,
color images, and oscilloscope setups (standard on the TDS 524A and
544A and optional, as option 1F, on the TDS 520A and 540A). Also, all
TDS 500A have builtĆin NVRAM storage for saving waveforms and
setups.
HOn the 524A and 544A, a color display for distinguishing among waveĆ
forms, their measurements, and associated text.
Use a combination of frontĆpanel buttons, knobs, and onĆscreen menus to
control the many functions of these oscilloscopes. The frontĆpanel controls
are grouped according to function: vertical, horizontal, trigger, and special.
Set a function you adjust often, such as vertical positioning or the time base
setting, directly by its own frontĆpanel knob. Set functions which you change
less often, such as vertical coupling and horizontal mode, indirectly using
selected menus.
TM
allows acquisition rates of up to
TM
2Ć2
Menus
Pressing one (sometimes two) frontĆpanel button(s), such as vertical menu,
displays a main menu of related functions, such as coupling, bandwidth,
etc., at the bottom of the screen. Pressing a mainĆmenu button, such as
coupling, displays a side menu of settings for that function, such as AC, DC,
or GND (ground) coupling, at the right side of the screen. Pressing a sideĆ
menu button selects a setting such as DC.
Indicators
OnĆscreen readouts help you keep track of the settings for various functions,
such as vertical and horizontal scale and trigger level. Some readouts use
the cursors or the automatic parameter extraction feature (called measure)
to display the results of measurements made or the status of the instrument.
Specifications
Specifications
General Purpose Knob
Assign the general purpose knob to adjust a selected parameter function.
More quickly change parameters by toggling the SHIFT button. Use the
same method as for selecting a function, except the final sideĆmenu selecĆ
tion assigns the general purpose knob to adjust some function, such as the
position of measurement cursors on screen, or the setting for a channels
fine gain.
GUI
The user interface also makes use of a GUI, or Graphical User Interface, to
make setting functions and interpreting the display more intuitive. Some
menus and status are displayed using iconic representations of function
settings such as those shown here for full, 100 MHz, and 20 MHz
bandwidth. Such icons allow you to more readily determine status or the
available settings.
Signal Acquisition
System
TDS 540A and 544A: The signal acquisition system provides four vertical
channels with calibrated vertical scale factors from 1 mV to 10 V per division.
All four channels can be acquired simultaneously.
Each of the four TDS 540A and 544A channels can be displayed, vertically
positioned, and offset, can have their bandwidth limited (100 MHz or
20 MHz) and their vertical coupling specified. Fine gain can also be adĆ
justed.
TDS 520A and 524A: The signal acquisition system provides four vertical
channels. Two are fullĆfeatured vertical channels (CH1 and CH2) with caliĆ
brated vertical scale factors from 1 mV to 10 V per division. The other two
are auxiliary channels (AUX1 and AUX2) with three calibrated deflection
factors of 100 mV, 1 V, and 10 V per division. Any two of the four channels
can be acquired simultaneously.
Each of the four TDS 520A and 524A channels can be displayed, vertically
positioned, and offset. CH1 and CH2 can also have their bandwidth limited
(100 MHz or 20 MHz) and their vertical coupling specified. Fine gain can
also be adjusted for CH1 and CH2.
On all TDS 520A, 524A, 540A and 544A: Besides the four channels, up to
three math waveforms and four reference waveforms are available for disĆ
play. (A math waveform results when dual waveform operations, such as
add, are specified on any two channels. A reference waveform results when
you save a live waveform in a reference memory.)
There are three horizontal display modes: main only, main intensified, and
delayed only. You can select among various horizontal record length setĆ
tings.
A feature called Fit to Screen" allows the user to view entire waveform
records within the 10 division screen area. In other words, waveforms are
compressed to fit on the screen. The exception is that, with Option 1M,
50,000 point records are displayed over a 15 division time span. (see
TableĂ2Ć1)
TableĂ2Ć1:ăRecord Length vs. Divisions per Record,
Samples Per Division, and Sec/Div Sequence
Divisions per Records
R
50000
15000
h
1
2
Samples/Division (Sec/Div Sequence)
Fit to Screen OFF
50 (1-2-5)
Fit to Screen ON
50 (1-2-5)
1000 divisions10 divisions
300 divisions15 divisions
5000100 divisions10 divisions
250050 divisions10 divisions
100020 divisions10 divisions
50010 divisions10 divisions
1 The 50,000 samples in 1,000 division record is only available with Option 1M.
2 The maximum record length of 15,000 samples (50,000 samples with Option 1M) is selectĆ
able with all acquisition modes except Hi Res. In Hi Res, the maximum record length is
5,000 samples (15,000 samples with Option 1M).
FastFrameTMallows multiple triggered acquisitions to occur with a minimum
of dead time" between acquisitions. A maximum of 910 acquisition seĆ
quences (frames) are possible with 50 points per record (frame length)
when a 50,000 point waveform record length is available (with option 1M).
You can delay by time with respect to the main trigger both the delayed
display and the intensified zone on the main intensified display. You can set
them to display immediately after the delay (delayed runs after main mode).
The delayed display can also be set to display at the first valid trigger after
the delay (delayed triggerable mode).
You can also delay by a selected number of events the delayed display (or
the intensified zone). In this case, the events source is the delayed trigger
source. For any events signal, the delayedĆtrigger system conditions the
signal by determining the source, coupling, etc., of that signal.
2Ć4
Specifications
Specifications
Trigger System
The triggering system supports a varied set of features for triggering the
signalĆacquisition system. Trigger signals recognized include:
HEdge (mainĆ and delayedĆtrigger systems): This familiar type of triggerĆ
ing is fully configurable for source, slope, coupling, mode (auto or
normal), and holdoff.
HLogic (mainĆtrigger system): This type of triggering can be based on
pattern (asynchronous) or state (synchronous). In either case, logic
triggering is configurable for sources, for boolean operators to apply to
those sources, for logic pattern or state on which to trigger, for mode
(auto or normal), and for holdoff. TimeĆqualified logic triggering can also
be specified.
HPulse (mainĆtrigger system): Pulse triggering is configurable for triggerĆ
ing on runt or glitch pulses, or on pulse widths or periods inside or
outside limits that you specify. It is also configurable for source, polarity,
mode, and holdoff.
HVideo (with option 05: Video Trigger): Video triggering is compatible with
standard NTSC, PAL, SECAM, and HDTV formats. An additional feature
called FlexFormat
format on which to trigger.
You can choose where the trigger point is located within the acquired waveĆ
form record by selecting the amount of pretrigger data displayed. Select
presets of 10%, 50%, and 90% of pretrigger data in the horizontal menu, or
assign the general purpose knob to set pretrigger data to any value within
the limits of trigger position resolution.
TM
(flexible format) allows the user to define the video
You can specify a mode and manner to acquire and process signals.
HSelect equivalentĆtime sampling on repetitive signals or interpolation of
points sampled on nonĆrepetitive signals. Both can increase apparent
sample rate on the waveform when maximum realĆtime rates are
reached. The apparent sample rate can be increased, even further, by
using the FastFrame
cond are possible using FastFrame
HUse peakĆdetect, highĆresolution, sample, envelope, and average
modes can be used to acquire signals.
HSet the acquisition to stop after a single acquisition (or sequence of
acquisitions if acquiring in average or envelope modes).
HSelect channel sources for compliance with limit tests. You can direct
the TDS to signal you or generate hard copy output based on the reĆ
sults. Also, you can create templates for use in limit tests.
Help and autoset can assist you in setting up the Digitizing Oscilloscope to
make your measurements.
TM
feature. Acquisition rates of 50,000 Frames/SeĆ
TM
.
Measurement
Assistance
Help
Help displays operational information about any frontĆpanel control. When
help mode is in effect, manipulating any frontĆpanel control causes the
oscilloscope to display information about that control. When help is first
invoked, an introduction to help is displayed on screen.
Autoset
Autoset automatically sets up the oscilloscope for a viewable display based
on the input signal.
Once you have set up to make your measurements, the cursor and measure
features can help you quickly make those measurements.
Cursor
Three types of cursors are provided for making parametric measurements
on the displayed waveforms. Horizontal bar cursors (H Bar) measure vertical
parameters (typically volts). Vertical bar cursors (V Bar) measure horizontal
parameters (typically time or frequency) and now extend to the top and
bottom of the screen. Paired cursors measure both amplitude and time
simultaneously. These are delta measurements; that is, measurements
based on the difference between two cursors.
2Ć6
Specifications
Specifications
Both H Bar and V Bar cursors can also be used to make absolute measureĆ
ments; that is measurements relative to a defined level or event. For the
H Bars, either cursor can be selected to read out its voltage with respect to
any channels ground reference level. For the V Bars, it's time with respect to
the trigger point (event) of the acquisition, and the cursors can control the
portion of the waveform on which automatic measurements are made.
For time measurements, units can be either seconds or Hertz (for 1/time).
When the video trigger option installed (Option 05), the video line number
can be selected using the vertical cursors. IRE amplitude (NTSC) can be
measured using the horizontal cursors with or without the video trigger
option installed.
Measure
Measure can automatically extract parameters from the signal input to the
Digitizing Oscilloscope. Any four out of the more than 20 parameters availĆ
able can be displayed to the screen. The waveform parameters are meaĆ
sured continuously with the results updated onĆscreen as the Digitizing
Oscilloscope continues to acquire waveforms.
Storage and I/O
Digital Signal Processing (DSP)
An important component of the multiprocessor architecture of this Digitizing
Oscilloscope is Tektronix's proprietary digital signal processor, the DSP. This
dedicated processor supports advanced analysis of your waveforms when
doing such computeĆintensive tasks as interpolation, waveform math, and
signal averaging. It also teams with a custom display system to deliver
specialized display modes (See Display, later in this description.)
Acquired waveforms may be saved in any of four nonvolatile REF (reference)
memories or, if available, on a 3.5 inch, DOS 3.3ĆorĆlater compatible disk.
The disk is standard on the TDS 524A and 544A. It is available as option 1F
on the 520A and 540A. Any or all of the saved waveforms may be displayed
for comparison with the waveforms being currently acquired.
The source and destination of waveforms to be saved may be chosen.
Assignment can be made to save any of the four channels to any REF
memory or to move a stored reference from one REF memory to another.
Reference waveforms may also be written into a REF memory location via
the GPIB interface.
The Digitizing Oscilloscope is fully controllable and capable of sending and
receiving waveforms over the GPIB interface (IEEE Std 488.1-1987/IEEE
Std 488.2-1987 standard). This feature makes the instrument ideal for
making automated measurements in a production or research and developĆ
ment environment that calls for repetitive data taking. SelfĆcompensation
and selfĆdiagnostic features built into the Digitizing Oscilloscope to aid in
fault detection and servicing are also accessible using commands sent from
a GPIB controller.
Another standard feature is hardcopy. This feature allows you to output
waveforms and other onĆscreen information to a variety of graphic printers
and plotters from the Digitizing Oscilloscope front panel, providing hard
copies without requiring you to put the Digitizing Oscilloscope into a systemĆ
controller environment. You can make hardcopies in a variety of popular
output formats, such as PCX, TIFF, BMP, RLE, EPS, Interleaf, and EPS mono
or color. You can also save hardcopies in a disk file in any of the formats
above. The hardcopies obtained are based on what is displayed onĆscreen
at the time hardcopy is invoked. The hardcopies can be stamped with date
and time and spooled to a queue for printing at a later time. You can output
screen information via GPIB, RSĆ232C, or Centronics interfaces.
Display
The TDS 520A, 524A, 540A and 544A Digitizing Oscilloscopes offer flexible
display options. You can customize the following attributes of your display:
HColor: Waveforms, readouts, graticule, etc. on the TDS 524A and 544A;
HIntensity: waveforms, readouts, and graticule;
HStyle of waveform display(s): vectors or dots, intensified or nonĆintensiĆ
fied samples, infinite persistence, and variable persistence with color
coding;
HInterpolation method: Sin(x)/x or Linear;
HDisplay format: xy or yt with various graticule selections including NTSC
and PAL to be used with video trigger.
Zoom
This Digitizing Oscilloscope also provides an easy way to focus in on those
waveform features you wish to examine up close. By invoking zoom, you
can magnify the waveform parameter using the vertical and horizontal
controls to expand (or contract) and position it for viewing.
2Ć8
Specifications
Nominal Traits
This subsection contains tables that list the electrical and mechanical
nominal traits that describe the TDS 520A, 524A, 540A and 544A Digitizing
Oscilloscopes.
Nominal traits are described using simple statements of fact such as Four,
all identical" for the trait Input Channels, Number of", rather than in terms of
limits that are performance requirements.
TableĂ2Ć2:ăNominal Traits Ċ Signal Acquisition System
NameDescription
Bandwidth Selections20 MHz, 100 MHz, and FULL (500 MHz)
Digitizers, Number ofTDS 540A and 544A: Four, all identical
TDS 520A and 524A: Two, both identical
Digitized Bits, Number of8 bits
1
Input Channels, Number ofTDS 540A and 544A: Four, all identical, called CH1 - CH4
TDS 520A and 524A: Two fullĆfeatured (CH1 and CH2), plus two
limited, auxiliary inputs (AUX1 and AUX2)
Input Coupling
Input Impedance Selections
Ranges, Offset, TDS 540A, 544A, and
CH1 and CH2 on TDS 520A and 524A
Ranges, Offset, AUX1 and AUX2 on
TDS 520A and 524A
2
DC, AC, or GND
1MW or 50 W
Volts/Div Setting
1 mV/div - 99.5 mV/div
100 mV/div - 995 mV/div
1 V/div - 10 V/div
Volts/Div Setting
100 mV/div
1 V/div
10 V/div
Offset Range
±1 V
±10 V
±100 V
Offset Range
±0.5 V
±5.0 V
±50 V
Range, Position±5 divisions
Range, SensitivityTDS 540A and 544A: 1 mV/div to 10 V/div
3
TDS 520A and 524A: CH1 and CH2: 1 mV/div to 10 V/div
TDS 520A and 524A: AUX1 and AUX2: 100 mV/div, 1 V/div, 10 V/div
3
1 Displayed vertically with 25 digitization levels (DLs) per division and 10.24 divisions dynamic range with zoom off. A DL is the
smallest voltage level change that can be resolved by the 8Ćbit AĆD Converter, with the input scaled to the volts/division setting of
the channel used. Expressed as a voltage, a DL is equal to 1/25 of a division times the volts/division setting.
2 The input characteristics (Input Coupling, Input Impedance Selections, etc.) apply to both fullĆfeatured and auxiliary inputs except
where otherwise specified.
3 The sensitivity ranges from 1 mV/div to 10 V/div in a 1-2-5 sequence of coarse settings. Between a pair of adjacent coarse setĆ
tings, the sensitivity can be finely adjusted. The resolution of such a fine adjustment is 1% of the more sensitive of the pair. For
example, between 50 mV/div and 100ĂmV/div, the volts/division can be set with 0.5 mV resolution.
Range, Equivalent Time or Interpolated
Waveform Rate
2,3
Description
Number of
Channels On
1
2
3or4
Number of
Channels On
1
2
SampleĆRate
Range
5 Samples/s - 1 GSamples/s
5 Samples/s - 500 MSamples/s
5 Samples/s - 250 MSamples/s
SampleĆRate
Range
5 Samples/s - 500 MSamples/s
5 Samples/s - 250 MSamples/s
500 MSamples/s to 100 GSamples/s
Range, Seconds/Division500 ps/div to 10 s/div
Record Length
4
500 samples, 1000 samples, 2500 samples, 5000 samples,
15000 samples. A record length of 50000 samples is available with
Option 1M. Up to four 50 K waveform records may be saved in
NVRAM with Option 1M installed.
FastFrame
TM
Maximum Frame Rate: 50,000 Frames/Second
Frame Length Range: 50 points/Frame to 5,000 Points/Frame
Maximum Number of Frames:
ą910 Frames at 50 Points/Frame
ą(with Option 1M, 50,000 Record Length)
ą227 Frames at 50 Points/Frame
ą(standard configuration of 15,000 Record Length)
1 The range of realĆtime rates, expressed in samples/second, at which a digitizer samples signals at its inputs and stores the samĆ
ples in memory to produce a record of timeĆsequential samples
2 The range of waveform rates for equivalent time or interpolated waveform records.
3 The Waveform Rate (WR) is the equivalent sample rate of a waveform record. For a waveform record acquired by realĆtime samĆ
pling of a single acquisition, the waveform rate is the same as the realĆtime sample rate; for a waveform created by interpolation of
realĆtime samples from a single acquisition or by equivalentĆtime sampling of multiple acquisitions, the waveform rate is faster than
the real time sample rate. For all three cases, the waveform rate is 1/(Waveform Interval) for the waveform record, where the waveĆ
form interval (WI) is the time between the samples in the waveform record.
4 The maximum record length of 15,000 samples (50,000 samples with Option 1M) is selectable with all acquisition modes except
Hi Res. In Hi Res, the maximum record length is 5,000 samples (15,000 samples with Option 1M).
2Ć10
Specifications
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.