Tektronix TDS520A, 524A, 540A,,544A Performance Verification Performance Verification

Performance Verification
TDS 520A, 524A, 540A, & 544A Digitizing Oscilloscopes
070-8712-01
Please check for change information at the rear of this manual.
First Printing: July 1993.
Instrument Serial Numbers
B010000 Tektronix, Inc., Beaverton, Oregon, USA E200000 Tektronix United Kingdom, Ltd., London J300000 Sony/Tektronix, Japan H700000 Tektronix Holland, NV, Heerenveen, The Netherlands
Instruments manufactured for Tektronix by external vendors outside the United States are assigned a two digit alpha code to identify the country of manufacture (e.g., JP for Japan, HK for Hong Kong, IL for Israel, etc.).
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
Printed in U.S.A.
Copyright E Tektronix, Inc., 1993. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. The following are registered trademarks: TEKTRONIX, TEK, TEKPROBE, and SCOPEĆMOBILE.
WARRANTY
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Welcome
This is the Performance Verification for the TDS 520A, 524A, 540A, and 544A Oscilloscope. It contains procedures suitable for determining if the instruĆ ment functions, was adjusted properly, and meets the performance characĆ teristics as warranted.
Also contained in this document are technical specifications for these oscilĆ loscopes.
Related Manuals
The following documents are related to the use or service of the digitizing oscilloscope.
H The TDS 520A, 524A, 540A, & 544A User Manual (Tektronix part number
070-8710-01).
H The TDS Family Programmer Manual (Tektronix part number
070-8709-01) describes using a computer to control the digitizing oscilloscope through the GPIB interface.
H The TDS 520A, 524A, 540A, 544A, & 644A Reference (Tektronix part
number 070-8711-01) gives you a quick overview of how to operate your digitizing oscilloscope.
H The TDS 520A, 524A, 540A, & 544A Service Manual (Tektronix part
number 070-8713-01) provides information for maintaining and servicĆ ing your digitizing oscilloscope to the module level.
TDS 520A, 524A, 540A, & 544A Performance Verification
i
Welcome
ii
Welcome
Safety Summary v.............................................
Performance Verification Procedures
Brief Procedures 1Ć1............................................
General Instructions 1Ć1.....................................
Conventions 1Ć2.............................................
Self Tests 1Ć4...............................................
Verify Internal Adjustment, Self Compensation, and
Diagnostics 1Ć4.......................................
Functional Tests 1Ć6.........................................
Verify All Input Channels 1Ć6...............................
Verify the Time Base 1Ć9...................................
Verify the Main and Delayed Trigger Systems 1Ć10.............
Verify the File System (Optional on TDS 520A and 540A) 1Ć12...
Performance Tests 1Ć15...........................................
Prerequisites 1Ć15.........................................
Equipment Required 1Ć16...................................
Test Record 1Ć19..........................................
Signal Acquisition System Checks 1Ć23........................
Check Accuracy of Offset (Zero Setting) 1Ć23..................
Check DC Gain and Voltage Measurement Accuracy 1Ć25......
Check Analog Bandwidth 1Ć32..............................
Check Delay Between Channels 1Ć36........................
Time Base System Checks 1Ć39...............................
Check Accuracy for LongĆTerm Sample Rate, Delay Time,
and Delta Time Measurements 1Ć39......................
Trigger System Checks 1Ć42...................................
Check Accuracy (Time) for PulseĆGlitch or PulseĆWidth
Triggering 1Ć42........................................
Check Accuracy, TriggerĆlevel or Threshold, DC Coupled 1Ć45...
Sensitivity, Edge Trigger, DC Coupled 1Ć48....................
Output Signal Checks 1Ć54....................................
Check Outputs Ċ CH 3 and Main and Delayed Trigger
(TDS 540A and 544A only) 1Ć54..........................
Check Probe Compensator Outputs 1Ć57.....................
Option 05 Video Trigger Checks 1Ć61..........................
Check Video Trigger 1Ć61...................................
Table of Contents
TDS 520A, 524A, 540A, & 544A Performance Verification
iii
Table of Contents
Specifications
Specifications 2Ć1...............................................
General Product Description 2Ć1..............................
User Interface 2Ć2...........................................
Menus 2Ć2...............................................
Indicators 2Ć2............................................
General Purpose Knob 2Ć3.................................
GUI 2Ć3..................................................
Signal Acquisition System 2Ć3................................
Horizontal System 2Ć4.......................................
Trigger System 2Ć5..........................................
Acquisition Control 2Ć6......................................
OnĆBoard User Assistance 2Ć6...............................
Help 2Ć6.................................................
Autoset 2Ć6..............................................
Measurement Assistance 2Ć6.................................
Cursor 2Ć6...............................................
Measure 2Ć7.............................................
Digital Signal Processing (DSP) 2Ć7.........................
Storage and I/O 2Ć7..........................................
Display 2Ć8..................................................
Zoom 2Ć8................................................
Nominal Traits 2Ć9...............................................
Warranted Characteristics 2Ć15...................................
Typical Characteristics 2Ć21.......................................
iv
Contents
Safety Summary
Please take a moment to review these safety precautions. They are provided for your protection and to prevent damage to the digitizing oscilloscope. This safety information applies to all operators and service personnel.
Symbols and Terms
These two terms appear in manuals:
H
statements identify conditions or practices that could result in
damage to the equipment or other property.
H statements identify conditions or practices that could result in
personal injury or loss of life.
These two terms appear on equipment:
H CAUTION indicates a personal injury hazard not immediately accessible
as one reads the marking, or a hazard to property including the equipĆ ment itself.
H DANGER indicates a personal injury hazard immediately accessible as
one reads the marking.
This symbol appears in manuals:
These symbols appear on equipment:
DANGER
High Voltage
TDS 520A, 524A, 540A, & 544A Performance Verification
StaticĆSensitive Devices
Protective
ground (earth)
terminal
ATTENTION
Refer to
manual
v
Safety Summary
Specific Precautions
Observe all of these precautions to ensure your personal safety and to prevent damage to either the digitizing oscilloscope or equipment conĆ nected to it.
Power Source
The digitizing oscilloscope is intended to operate from a power source that will not apply more than 250 V tween either supply conductor and ground. A protective ground connection, through the grounding conductor in the power cord, is essential for safe system operation.
between the supply conductors or beĆ
RMS
Grounding the Digitizing Oscilloscope
The digitizing oscilloscope is grounded through the power cord. To avoid electric shock, plug the power cord into a properly wired receptacle where earth ground has been verified by a qualified service person. Do this before making connections to the input or output terminals of the digitizing oscilloĆ scope.
Without the protective ground connection, all parts of the digitizing oscilloĆ scope are potential shock hazards. This includes knobs and controls that may appear to be insulators.
Use the Proper Power Cord
Use only the power cord and connector specified for your product. Use only a power cord that is in good condition.
Use the Proper Fuse
To avoid fire hazard, use only the fuse specified in the parts list for your product, matched by type, voltage rating, and current rating.
Do Not Remove Covers or Panels
To avoid personal injury, do not operate the digitizing oscilloscope without the panels or covers.
Electric Overload
Never apply to a connector on the digitizing oscilloscope a voltage that is outside the range specified for that connector.
Do Not Operate in Explosive Atmospheres
The digitizing oscilloscope provides no explosion protection from static discharges or arcing components. Do not operate the digitizing oscilloscope in an atmosphere of explosive gases.
vi
Safety
Performance Verification
Procedures
Brief Procedures
The Self Tests use internal routines to confirm basic functionality and proper adjustment. No test equipment is required to do these test procedures.
The Functional Tests utilize the probeĆcompensation output at the front panel as a testĆsignal source for further verifying that the oscilloscope functions properly. A standardĆaccessory probe, included with this oscilloscope, is the only equipment required.
General Instructions
Besides the Brief Procedures, the set of procedures that can be used to verify oscilloscope performance includes the Performance Tests, found later in this section. You may not need to perform all of these procedures, deĆ pending on what you want to accomplish:
H To rapidly confirm that this oscilloscope functions and was adjusted
properly, just do the procedures under Self Tests, which begin on page 1Ć4.
Advantages: These procedures are quick to do, require no external equipment or signal sources, and perform extensive functional and accuracy testing to provide high confidence that the oscilloscope will perform properly. They can be used as a quick check before making a series of important measurements.
H To further check functionality, first do the Self Tests just mentioned; then
do the procedures under Functional Tests that begin on page 1Ć6.
Advantages: These procedures require minimal additional time to perform, require no additional equipment other than a standardĆaccesĆ sory probe, and more completely test the internal hardware of this oscilloscope. They can be used to quickly determine if the oscilloscope is suitable for putting into service, such as when it is first received.
H If more extensive confirmation of performance is desired, do the PerĆ
formance Tests, beginning on page 1Ć15, after doing the Functional and Self Tests just referenced.
Advantages: These procedures add direct checking of warranted specifications. They require more time to perform and suitable test equipment is required. (See Equipment Required on page 1Ć16.)
If you are not familiar with operating this oscilloscope, read the TDS 520A, 524A, 540A, 544A, & 644A Reference or the TDS 520A, 524A, 540A, & 544A User manual. These contain instructions that acquaint you with the use of the frontĆpanel controls and the menu system.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć1
Brief Procedures
Conventions
Throughout these procedures the following conventions apply:
H Each test procedure uses the following general format:
Title of Test
Equipment Required
Prerequisites
Procedure
H Each procedure consists of as many steps, substeps, and subparts as
required to do the test. Steps, substeps, and subparts are sequenced as follows:
1. First Step
a. First Substep
H First Subpart
H Second Subpart
b. Second Substep
2. Second Step
H In steps and substeps, the leadĆin statement in italics instructs you what
to do, while the instructions that follow tell you how to do it: in the examĆ ple step below, Initialize the oscilloscope" by doing Press save/recall SETUP. Now, press the mainĆmenu button...".
Initialize the oscilloscope: Press save/recall SETUP. Now, press the mainĆmenu button Recall Factory Setup; then the sideĆmenu button OK Confirm Factory Init.
H Where instructed to use a frontĆpanel button or knob, or select from a
main or side menu, or verify a readout or status message, the name of the button or knob appears in boldface type: press SHIFT; then ACĆ QUIRE MENU", press the mainĆmenu button Coupling", or verify that the status message is Pass.
The symbol at the left is accompanied by information you must read to do the procedure properly.
1Ć2
Performance Verification Procedures
Brief Status Information
Graticule and Waveforms
Waveform Reference
Symbols: Ground Levels
and Waveform Sources
Vertical Scale, Horizontal
Scale, and Trigger Level
Readouts
Brief Procedures
H Refer to Figure 1Ć1: Main menu" refers to the menu that labels the
seven menu buttons under the display; side menu" refers to the menu that labels the five buttons to the right of the display. PopĆup menu" refers to a menu that pops up when a mainĆmenu button is pressed.
Position of Waveform
Record Relative to
the Screen and Display
General Purpose
Knob Readout
PopĆup Menu
Side menu area. Readouts for measurements move here when CLEAR MENU is pressed.
Main menu display area. Readouts in lower graticule area move here when
CLEAR MENU is pressed.
Figure 1Ć1:ăMap of Display Functions
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć3
Brief Procedures
Self Tests
This procedure uses internal routines to verify that this oscilloscope funcĆ tions and was adjusted properly. No test equipment or hookups are reĆ quired.
Verify Internal Adjustment, Self Compensation, and Diagnostics
Equipment Required: None.
Prerequisites: Power on the Digitizing Oscilloscope and allow a 20 minute
warmĆup before doing this procedure.
Procedure:
1. Verify that internal diagnostics pass: Do the following substeps to verify passing of internal diagnostics.
a. Display the System diagnostics menu:
H Press SHIFT; then press UTILITY.
H Repeatedly press the mainĆmenu button System until Diag/Err
is highlighted in the popĆup menu.
b. Run the System Diagnostics: Press the mainĆmenu button Execute;
then press the sideĆmenu button OK Confirm Run Test.
c. Wait: The internal diagnostics do an exhaustive verification of proper
oscilloscope function. This verification will take up to two minutes. At some time during the wait, a clock" icon (shown at left) will appear onĆscreen. When the verification is finished, the resulting status will appear on the screen.
d. Confirm no failures are found: Verify that no failures are found and
reported onĆscreen.
e. Confirm the three adjustment sections have passed status:
H Press SHIFT; then press UTILITY.
H Press the main menu button System until Cal is highlighted in
the popĆup menu.
H Verify that the word Pass appears in the main menu under the
following menu labels: Voltage Reference, Frequency ReĆ sponse, and Pulse Trigger. (See Figure 1Ć2.)
1Ć4
Performance Verification Procedures
First, the CAL menu is displayed.
Second, the adjustment
sections are verified.
Brief Procedures
compensation is run and is
Third, a signal path
verified.
Figure 1Ć2:ăVerifying Adjustments and SignalĆPath Compensation
When doing steps f and g, do not turn off the oscilloscope until signalĆpath compensation completes. If you interrupt (or lose) power to the instrument while signalĆpath compensation is running, a message is logged in the oscilloscope error log. If such a case occurs, rerun signalĆpath compensation.
f. Run the signalĆpath compensation: Press the mainĆmenu button
Signal Path; then press the sideĆmenu button OK Compensate Signal Paths.
g. Wait: SignalĆpath compensation runs in about one to two minutes.
While it progresses, a clock" icon (shown at left) is displayed onĆ screen. When compensation completes, the status message will be updated to Pass or Fail in the main menu (see step h).
h. Confirm signalĆpath compensation returns passed status: Verify the
word Pass appears under Signal Path in the main menu. (See Figure 1Ć2.)
2. Return to regular service: Press CLEAR MENU to exit the system meĆ nus.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć5
Brief Procedures
Functional Tests
The purpose of these procedures is to confirm that this oscilloscope funcĆ tions properly. The only equipment required is one of the standardĆaccessoĆ ry probes and, to check the file system, a 3.5 inch. 720 K or 1.44 Mbyte floppy disk.
These procedures verify functions; that is, they verify that oscilloĆ scope features operate. They do not verify that they operate within limits.
Therefore, when the instructions in the functional tests that follow call for you to verify that a signal appears onĆscreen that is about five divisions in amplitude" or has a period of about six horizontal divisions", etc., do NOT interpret the quantities given as limits. Operation within limits is checked in Performance Tests, which begin on page 1Ć15.
DO NOT make changes to the frontĆpanel settings that are not called out in the procedures. Each verification procedure will require you to set the oscilloscope to certain default settings before verifying functions. If you make changes to these settings, other than those called out in the procedure, you may obtain invalid results. In this case, just redo the procedure from step 1.
When you are instructed to press a menu button, the button may already be selected (its label will be highlighted). If this is the case, it is not necessary to press the button.
Verify All Input Channels
Equipment Required: One P6139A probe.
Prerequisites: None.
Procedure:
1. Install the test hookup and preset the oscilloscope controls:
Figure 1Ć3:ăUniversal Test Hookup for Functional Tests
a. Hook up the signal source: Install the probe on CH 1. Connect the
probe tip to PROBE COMPENSATION SIGNAL on the front panel; connect the probe ground to PROBE COMPENSATION GND.
1Ć6
Performance Verification Procedures
Brief Procedures
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
2. Verify that all input channels operate: Do the following substeps Ċ test CH 1 first, skipping substep a since CH 1 is already set up for verification
from step 1.
a. Select an unverified channel:
H Press WAVEFORM OFF to remove from display the channel just
verified.
H Press the frontĆpanel button that corresponds to the channel
you are to verify.
H Move the probe to the channel you selected.
b. Set up the selected channel:
H Press AUTOSET to obtain a viewable, triggered display in the
selected channel.
H Set the horizontal SCALE to 200 ms. Press CLEAR MENU to
remove any menu that may be on the screen.
c. Verify that the channel is operational: Confirm that the following
statements are true.
H The vertical scale readout for the channel under test shows a
setting of 200 mV, and a squareĆwave probeĆcompensation signal about 2.5 divisions in amplitude is onĆscreen. (See FigĆ ure 1Ć1 on page 1Ć3 to locate the readout.)
H The vertical POSITION knob moves the signal up and down the
screen when rotated.
H Turning the vertical SCALE knob counterclockwise decreases
the amplitude of the waveform onĆscreen, turning the knob clockwise increases the amplitude, and returning the knob to 200 mV returns the amplitude to about 2.5 divisions.
d. Verify that the channel acquires in all acquisition modes: Press
SHIFT; then press ACQUIRE MENU. Use the side menu to select, in turn, each of the five hardware acquire modes and confirm that the following statements are true. Refer to the icons at the left of each statement as you confirm those statements.
H Sample mode displays an actively acquiring waveform onĆ
screen. (Note that there is noise present on the peaks of the square wave.)
H Peak Detect mode displays an actively acquiring waveform
onĆscreen with the noise present in Sample mode peak deĆ tected."
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć7
Brief Procedures
H Hi Res mode displays an actively acquiring waveform onĆscreen
with the noise that was present in Sample mode reduced.
H Envelope mode displays an actively acquiring waveform onĆ
screen with the noise displayed.
H Average mode displays an actively acquiring waveform onĆ
screen with the noise reduced like in Hi Res mode.
TDS 520A and 524A only: Substep e will have you repeat the preĆ vious substeps to check all input channels. Be sure to check only CH 1 and CH 2 when testing the TDS 520A or 524A. (Step 3 will test the AUX 1 and AUX 2 inputs.) When testing the TDS 540A or 544A, test all four channels, CH 1 through CH 4.)
e. Test all channels: Repeat substeps a through d until all four input
channels are verified.
3. TDS 520A and 524A Only: Verify auxiliary inputs operate: Perform the following substeps when checking the AUX 1 and AUX 2 inputs only.
a. Select an auxiliary channel:
H Press WAVEFORM OFF to remove from display the channel just
verified.
H Press the frontĆpanel button that corresponds to the channel
you are to verify.
H Move probe to the channel you selected.
b. Set up the selected channel: Press AUTOSET to obtain a viewable
display in the selected channel.
The display obtained might not trigger stably because autoset cannot provide more than about ½ division of probe compensation signal amplitude in an auxiliary channel equipped with a 10X probe. This amount is less than the minimum trigger sensitivity for auxiliary channel trigger sources; therefore, triggering is not required.
c. Verify that the channel is operational: Confirm that the following
statements are true.
H The vertical scale readout for the channel under test shows a
setting of 1 V, and a squareĆwave probeĆcompensation signal about ½ division in amplitude is onĆscreen. (See Figure 1Ć1 on page 1Ć3 to locate the readout.)
H The vertical POSITION knob moves the signal up and down the
screen when rotated.
H Turning the vertical SCALE knob counterclockwise to 10 V
decreases the amplitude of the waveform onĆscreen. (The amplitude will drop to near zero when doing this substep.)
1Ć8
H Returning the knob to 1 V returns the amplitude to about ½
division.
d. Verify that the channel acquires in all acquisition modes: Disconnect
the probe ground lead from the probeĆcompensation terminal. Do step 2, substep d to verify the five acquire modes.
Performance Verification Procedures
Brief Procedures
e. Test all channels: Repeat substeps a through d to verify AUX 2.
4. Remove the test hookup: Disconnect the probe from the channel input and the probeĆcompensation terminals.
Verify the Time Base
Equipment Required: One P6139A probe.
Prerequisites: None.
Procedure:
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the
probe tip to PROBE COMPENSATION SIGNAL on the front panel; connect the probe ground to PROBE COMPENSATION GND. (See Figure 1Ć3 on page 1Ć6.)
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup; then press
the sideĆmenu button OK Confirm Factory Init.
c. Modify default settings:
H Press AUTOSET to obtain a viewable, triggered display.
H Set the horizontal SCALE to 200 ms.
H Press CLEAR MENU to remove the menus from the screen.
2. Verify that the time base operates: Confirm the following statements.
a. One period of the squareĆwave probeĆcompensation signal is about
five horizontal divisions onĆscreen for the 200 ms horizontal scale setting (set in step 1c).
b. Rotating the horizontal SCALE knob clockwise expands the waveĆ
form onĆscreen (more horizontal divisions per waveform period), and that counterclockwise rotation contracts it, and that returning the horizontal scale to 200 ms returns the period to about five divisions.
c. The horizontal POSITION knob positions the signal left and right
onĆscreen when rotated.
3. Remove the test hookup: Disconnect the probe from the channel input and the probeĆcompensation terminals.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć9
Brief Procedures
Verify the Main and Delayed Trigger Systems
Equipment Required: One P6139A probe.
Prerequisites: None.
Procedure:
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the
probe tip to PROBE COMPENSATION SIGNAL on the front panel; connect the probe ground to PROBE COMPENSATION GND. (See Figure 1Ć3 on page 1Ć6.)
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
c. Modify default settings:
H Press AUTOSET to obtain a viewable, triggered display.
H Set the horizontal SCALE for the M (main) time base to 200 ms.
H Press TRIGGER MENU.
H Press the mainĆmenu button Mode & Holdoff.
H Press the sideĆmenu button Normal.
H Press CLEAR MENU to remove the menus from the screen.
2. Verify that the main trigger system operates: Confirm that the following statements are true.
H The trigger level readout for the main trigger system changes
with the trigger LEVEL knob.
H The triggerĆlevel knob can trigger and untrigger the squareĆwave
signal as you rotate it. (Leave the signal untriggered.)
H Pressing SET LEVEL TO 50% triggers the signal that you just
left untriggered. (Leave the signal triggered.)
3. Verify that the delayed trigger system operates:
a. Select the delayed time base:
1Ć10
H Press HORIZONTAL MENU.
H Press the mainĆmenu button Time Base.
H Press the sideĆmenu button Delayed Triggerable; then press
the sideĆmenu button Delayed Only.
H
Set the horizontal SCALE for the D (delayed) time base to 200 ms.
Performance Verification Procedures
Brief Procedures
b. Select the delayed trigger level menu:
H Press SHIFT; then press DELAYED TRIG.
H Press the mainĆmenu button Level; then press the sideĆmenu
button Level.
c. Confirm that the following statements are true:
H The triggerĆlevel readout for the delayed trigger system changes
as you turn the general purpose knob.
H The general purpose knob can trigger and untrigger the squareĆ
wave probeĆcompensation signal as you rotate it. (Leave the signal untriggered.)
H Pressing the sideĆmenu button Set to 50% triggers the probeĆ
compensation signal that you just left untriggered. (Leave the signal triggered.)
d. Verify the delayed trigger counter:
H Press the mainĆmenu button Delay by Time.
H Use the keypad to enter a delay time of 1 second (press 1 then
press ENTER).
H Verify that the trigger READY indicator on the front panel flashes
about once every second as the waveform is updated onĆ screen.
4. Remove the test hookup: Disconnect the standardĆaccessory probe from the channel input and the probeĆcompensation terminals.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć11
Brief Procedures
Verify the File System (Optional on TDS 520A and 540A)
Equipment Required: One P6139A probe and one 720 K or 1.44 Mbyte, 3.5
inch DOS compatible disk. You can use a disk of your own or you can use the Programming Examples Software 3.5 inch disk (Tektronix part number 063-1134-00) contained in the TDS Family Programmer Manual (Tektronix part number 070-8709-01).
Prerequisites: None.
Procedure:
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the
probe tip to PROBE COMPENSATION SIGNAL on the front panel; connect the probe ground to PROBE COMPENSATION GND. (See Figure 1Ć3 on page 1Ć6.)
b. Insert the test disk: Insert the disk in the disk drive to the left of the
monitor.
H Position the disk so the metal shutter faces the drive.
H Position the disk so the stamped arrow is on the top right side.
In other words, place the angled corner in the front bottom location.
H Push the disk into the drive until it goes all the way in and clicks
into place.
c. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
d. Modify default settings:
H Press AUTOSET to obtain a viewable, triggered display.
H Set the horizontal SCALE for the M (main) time base to 200 ms
(one click clockwise). Notice the waveform on the display now shows two cycles instead of five.
H Press CLEAR MENU to remove the menus from the screen.
e. Save the settings:
H Press SETUP.
H Press the mainĆmenu button Save Current Setup; then press
the sideĆmenu button To File.
H Turn the general purpose knob to select the file to save. Choose
TEK?????.SET (or fdo:). With this choice, you'll save a file starting with TEK, then containing 5Ćnumbers, and a .SET extension. For example, the first time you run this on a blank, formatted disk or on the Example Programs Disk, the TDS will assign the name TEK00000.SET to your file. If you ran the procedure again, the TDS would increment the name and call the file TEK00001.SET.
1Ć12
Performance Verification Procedures
H Press the sideĆmenu button Save To Selected File.
2. Verify the file system works:
H Press AUTOSET to restore the 500 ms time base and the five
cycle waveform.
H Press the mainĆmenu button Recall Saved Setup; then press
the sideĆmenu button From File.
H Turn the general purpose knob to select the file to recall. For
example, if you followed the instructions above and used a blank disk, you had the TDS assign the name TEK00000.SET to your file.
H Press the sideĆmenu button Recall From Selected File.
H Verify that Digitizing Oscilloscope retrieved the saved setup from
the disk. Do this by noticing the horizontal SCALE for the M (main) time base is again 200 ms and the waveform shows only two cycles just as it was when you saved the setup.
3. Remove the test hookup:
Brief Procedures
H Disconnect the standardĆaccessory probe from the channel
input and the probeĆcompensation terminals.
H Remove the disk from the disk drive. Do this by pushing in the
tab at the bottom of the disk drive.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć13
Brief Procedures
1Ć14
Performance Verification Procedures
Performance Tests
This subsection contains a collection of procedures for checking that TDS 520A, 524A, 540A, and 544A Digitizing Oscilloscopes perform as warranted. Since the procedures cover models with both two fullĆfeatured channels (TDS 520A and 524A) and four fullĆfeatured models (TDS 540A and 544A), instructions that apply only to one of the model types are clearly identified. Otherwise, all test instructions apply to both the two and four channel models.
The procedures are arranged in four logical groupings: Signal Acquisition System Checks, Time Base System Checks, Triggering System Checks, and Output Ports Checks. They check all the characteristics that are designated as checked in Section 2, Specifications. (The characteristics that are checked appear in boldface type under Warranted Characteristics in Section 2.)
These procedures extend the confidence level provided by the basic procedures described on page 1Ć1. The basic procedures should be done first, then these procedures performed if desired.
Prerequisites
The tests in this subsection comprise an extensive, valid confirmation of performance and functionality when the following requirements are met:
H The cabinet must be installed on the Digitizing Oscilloscope.
H You must have performed and passed the procedures under Self Tests,
found on page 1Ć4, and those under Functional Tests, found on page 1Ć6.
H A signalĆpath compensation must have been done within the recomĆ
mended calibration interval and at a temperature within ±5_ĂC of the present operating temperature. (If at the time you did the prerequisite Self Tests, the temperature was within the limits just stated, consider this prerequisite met.)
H The Digitizing Oscilloscope must have been last adjusted at an ambient
temperature between +20_ĂC and +30_ĂC, must have been operating for a warmĆup period of at least 20 minutes, and must be operating at an ambient temperature between +4_ĂC and +50_ĂC. (The warmĆup reĆ quirement is usually met in the course of meeting the first prerequisite listed above.)
Related Information Ċ Read General Instructions and Conventions that
start on page 1Ć1.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć15
Performance Tests
Equipment Required
These procedures use external, traceable signal sources to directly check warranted characteristics. The required equipment list is shown in Table 1Ć1.
TableĂ1Ć1:ăTest Equipment
Item Number and
Minimum Requirements Example Purpose
Description
1 Attenuator,10X
(three required)
Ratio: 10X; impedance 50 W; connectors: female BNC inĆ put, male BNC output
2 Attenuator, 5X Ratio: 5X; impedance 50 W;
connectors: female BNC inĆ put, male BNC output
3 Terminator, 50 W Impedance 50 W; connectors:
female BNC input, male BNC output
4 Cable, Precision
50 W Coaxial
50 W, 36 inch, male to male BNC connectors
(two required)
5 Connector, DuĆ
Female BNC to dual banana Tektronix part number alĆBanana (two required)
6 Connector, BNC
Male BNC to dual female BNC Tektronix part number T"
7 Coupler, DualĆ
Input
Female BNC to dual male
BNC
Tektronix part number 011-0059-02
Tektronix part number 011-0060-02
Tektronix part number 011-0049-01
Tektronix part number 012-0482-00
103-0090-00
103-0030-00
Tektronix part number 067-0525-02
Signal Attenuation
Signal Attenuation
Signal Termination for Channel Delay Test
Signal InterconnecĆ tion
Various Accuracy Tests
Checking Trigger Sensitivity
Checking Delay Between Channels
8 Generator, DC
Calibration
9 Generator, CaliĆ
bration
Variable amplitude to ±110 V;
accuracy to 0.1%
500 mV square wave calibraĆ
tor amplitude; accuracy to
Data Precision 8200, with 1 kV option installed
PG 506A
0.25%
10 Generator, LevĆ
eled Sine Wave, MediumĆFreĆ
200 kHz to 250 MHz; Variable
amplitude from 5ĂmV to
5.5 V
into 50 W
pĆp
Tektronix SG 503 Leveled Sine Wave Generator
quency
11 Generator, LevĆ
eled Sine Wave, HighĆFrequency
250 MHz to 500 MHz; Variable
amplitude from 500 mV to
V
into 50 W; 6 MHz
4
pĆp
Tektronix SG 504 Leveled Sine Wave Generator ing Head
reference
12 Generator, Time
Mark
Variable marker frequency
from 10 ms to 10 ns; accuraĆ
Tektronix TG 501 Time Mark Generator
cy within 2 ppm
13 Generator, CaliĆ
bration
500 mV square wave calibraĆ
tor amplitude; accuracy to
PG 506A
0.25%
1 Requires a TM 500 or TM 5000 Series Power Module Mainframe.
Checking DC Offset, Gain, and MeasureĆ ment Accuracy
1
To check accuracy of the CH 3 Signal Out (TDS 540A and 544A only)
Checking Trigger
1
Sensitivity at low freĆ quencies
Checking Analog
1
with its LevelĆ
Bandwidth and TrigĆ ger Sensitivity at high frequencies
Checking SampleĆ
1
Rate and DelayĆtime Accuracy
1
Use to check accuĆ racy of the CH 3 Output
1Ć16
Performance Verification Procedures
TableĂ1Ć1:ăTest Equipment (Cont.)
Performance Tests
Item Number and Description
14 Probe, 10X, inĆ
cluded with this instrument
15 Adapter, BNC
female to Clip Leads
16 Power Supply,
Dual Output
17 3.5 inch, 720 K
or 1.44 Mbyte, DOSĆcompatible floppy disk
18 Generator,
Video Signal
19 Oscillator, LevĆ
eled Sinewave Generator
PurposeExampleMinimum Requirements
A P6139A probe Tektronix number P6139A Signal InterconnecĆ
tion
BNC female to Clip Leads Tektronix part number
013-0076-00
Signal Coupling for Probe Compensator Output Check
0-35 V and 60 V 2 A; current
limit without foldback
Tektronix PS 280 Power Supply Power Supply
Troubleshooting
Programming Examples SoftĆ ware Disk (Tektronix part numĆ ber 063-1134-00) that comes
Checking File SysĆ tem Basic FunctionĆ
ality with the TDS Family ProgramĆ mer Manual (Tektronix part number 070-8709-01)
Provides NTSC compatible outputs.
Tektronix TSG 121 Used to Test Video
Option 05 Equipped
Instruments Only
60 Hz. Sine Wave Tektronix part number SG 502 Used to Test Video
Option 05 Equipped
Instruments Only
20 Pulse Generator Tektronix part number PG 502 Used to Test Video
Option 05 Equipped
Instruments Only
21 Cable, 75 W
Coaxial
75 W, 36 inch, male to male BNC connectors
Tektronix part number 012-1338-00
Used to Test Video
Option 05 Equipped
Instruments Only
22 Termination,
75 W
Impedance 75 W; connectors: female BNC input, male BNC output
Tektronix part number 011-0102-01
Used to Test Video
Option 05 Equipped
Instruments Only
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć17
Performance Tests
1Ć18
Performance Verification Procedures
Performance Tests
Test Record
Photocopy the next four pages and use them to record the performance test results for your instrument.
TDS 500A Test Record
Instrument Serial Number: Certificate Number: Temperature: RH %: Date of Calibration: Technician:
Performance Test Minimum Incoming Outgoing Maximum
Offset Accuracy
CH1 Offset +1 mV
+100 mV +1 V
CH2 Offset +1 mV
+100 mV +1 V
CH3 Offset +1 mV
+100 mV +1 V
CH4 Offset +1 mV
+100 mV +1 V
DC Voltage Measurement Accuracy (Averaged)
CH1 5 mV Vert scale setting,
-5 Div position setting
CH1 5 mV Vert scale setting,
+5 Div position setting
CH1 200 mV Vert scale setting,
-5 Div position setting
- 1.6 mV
- 25 mVąą
- 250 mV
- 1.6 mV
-25mVąą
- 250 mV
- 1.6 mV
-25mVąą
- 250 mV
- 1.6 mV
-25mVąą
- 250 mV
+ 1.0355 V __________ __________ + 1.0445 V
- 1.0445 V __________ __________ - 1.0355 V
+ 11.525 V __________ __________ + 11.675 V
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
__________ __________ __________
+ 1.6 mV +25mV + 250 mV
+ 1.6 mV +25mV + 250 mV
+ 1.6 mV + 25mV + 250 mV
+ 1.6 mV +25mV + 250 mV
CH1 200 mV Vert scale setting,
+5 Div position setting
CH1 1 V Vert scale setting,
-5 Div position setting
CH1 1 V Vert scale setting,
+5 Div position setting
CH2 5 mV Vert scale setting,
-5 Div position setting
CH2 5 mV Vert scale setting,
+5 Div position setting
TDS 520A, 524A, 540A, & 544A Performance Verification
- 11.675 V __________ __________ - 11.525 V
+ 107.450 V __________ __________ + 108.550 V
- 108.550 V __________ __________ - 107.450 V
+ 1.0355 V __________ __________ + 1.0445 V
- 1.0445 V __________ __________ - 1.0355 V
1Ć19
Performance Tests
TDS 500A Test Record (Cont.)
Instrument Serial Number: Certificate Number: Temperature: RH %: Date of Calibration: Technician:
Performance Test MaximumOutgoingIncomingMinimum
CH2 200 mV Vert scale setting,
-5 Div position setting
CH2 200 mV Vert scale setting,
+5 Div position setting
CH2 1 V Vert scale setting,
-5 Div position setting
CH2 1 V Vert scale setting,
+5 Div position setting
CH3 5 mV Vert scale setting,
-5 Div position setting
CH3 5 mV Vert scale setting,
+5 Div position setting
CH3 200 mV Vert scale setting,
-5 Div position setting
CH3 200 mV Vert scale setting,
+5 Div position setting
CH3 1 V Vert scale setting,
-5 Div position setting
CH3 1 V Vert scale setting,
+5 Div position setting
+ 11.525 V __________ __________ + 11.675 V
- 11.675 V __________ __________ - 11.525 V
+ 107.450 V __________ __________ + 108.550 V
- 108.550 V __________ __________ - 107.450 V
+ 1.0355 V __________ __________ + 1.0445 V
- 1.0445 V __________ __________ - 1.0355 V
+ 11.525 V __________ __________ + 11.675 V
- 11.675 V __________ __________ - 11.525 V
+ 107.450 V __________ __________ + 108.550 V
- 108.550 V __________ __________ - 107.450 V
CH4 5 mV Vert scale setting,
-5 Div position setting
CH4 5 mV Vert scale setting,
+5 Div position setting
CH4 200 mV Vert scale setting,
-5 Div position setting
CH4 200 mV Vert scale setting,
+5 Div position setting
CH4 1 V Vert scale setting,
-5 Div position setting
CH4 1 V Vert scale setting,
+5 Div position setting
1Ć20
+ 1.0355 V __________ __________ + 1.0445 V
- 1.0445 V __________ __________ - 1.0355 V
+ 11.525 V __________ __________ + 11.675 V
- 11.675 V __________ __________ - 11.525 V
+ 107.450 V __________ __________ + 108.550 V
- 108.550 V __________ __________ - 107.450 V
Performance Verification Procedures
Performance Tests
TDS 500A Test Record (Cont.)
Instrument Serial Number: Certificate Number: Temperature: RH %: Date of Calibration: Technician:
Performance Test MaximumOutgoingIncomingMinimum
Analog Bandwidth
CH1 100 mV 424 mV __________ __________ N/A
CH2 100 mV 424 mV __________ __________ N/A
CH3 100 mV 424 mV __________ __________ N/A
CH4 100 mV 424 mV __________ __________ N/A
Delay Between Channels
Delay Between Channels N/A __________ __________ 250 ps
Time Base System
Long Term Sample Rate/ Delay Time @ 500 ns/10 ms
-2.5 Div __________ __________ +2.5 Div
Delta Time@5ns(100 MHz) 19.760 ns __________ __________ 20.240 ns
Trigger System Accuracy
PulseĆGlitch or PulseĆWIdth, Hor. scale 1 ms
Lower Limit Upper Limit
2.5 ns
2.5 ns
__________ __________
__________ __________
7.5 ns
7.5 ns PulseĆGlitch or PulseĆWIdth, Hor. scale > 1 ms
Lower Limit Upper Limit
Main Trigger, DC Coupled) Delayed Trigger, DC Coupled)
1 ms 1 ms
9.940 V
9.940 V
__________ __________
__________ __________
__________ __________
__________ ___________
3 ms 3 ms
10.060 V
10.060 V
Output Signal Checks
MAIN TRIGGER OUTPUT, 1 M W High 2.5 V __________ __________ Low 0.7 V
MAIN TRIGGER OUTPUT, 50 W High 1.0 V __________ __________ Low 0.25 V
DELAYED TRIGGER OUTPUT, 50 W High 1.0 V __________ __________ Low 0.25 V
DELAYED TRIGGER OUTPUT, 1 M W High 2.5 V __________ __________ Low 0.7 V
CH 3 SIGNAL OUTPUT, 1 M W PkĆPk 90 mV __________ __________ PkĆPk 110 mV
CH 3 SIGNAL OUTPUT, 50 W PkĆPk 45 mV __________ __________ PkĆPk 55 mV
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć21
Performance Tests
TDS 500A Test Record (Cont.)
Instrument Serial Number: Certificate Number: Temperature: RH %: Date of Calibration: Technician:
Performance Test MaximumOutgoingIncomingMinimum
Probe Compensator Output Signal
Frequency (CH1 Freq.) 950 Hz __________ __________ 1050 Hz
Voltage (difference) 495 mV __________ __________ 505 mV
1Ć22
Performance Verification Procedures
Performance Tests
Signal Acquisition System Checks
These procedures check those characteristics that relate to the signalĆacquiĆ sition system and are listed as checked under Warranted Characteristics in Section 2, Specifications.
Check Accuracy of Offset (Zero Setting)
Equipment Required: None.
Prerequisites: The oscilloscope must meet the prerequisites listed on
page 1Ć15.
1. Preset the instrument controls:
a. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
H Press CLEAR MENU to remove the menus from the screen.
b. Modify the default settings:
H Set the horizontal SCALE to 1 ms.
H Press SHIFT; then ACQUIRE MENU.
H Press the mainĆmenu button Mode; then press the sideĆmenu
button Hi Res.
H Press DISPLAY.
H Press the mainĆmenu button Graticule; then press the sideĆmeĆ
nu button Frame.
H Press CURSOR.
H Press the mainĆmenu button Function; then press the sideĆmeĆ
nu button H Bars.
H Press CLEAR MENU.
2. Confirm input channels are within limits for offset accuracy at zero offset: Do the following substeps Ċ test CH 1 first, skipping substep a since
CH 1 is already set up to be checked from step 1.
a. Select an unchecked channel: Press WAVEFORM OFF to remove
the channel just confirmed from the display. Then, press the frontĆ panel button that corresponds to the channel you are to confirm.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć23
Performance Tests
Follow these rules to match this procedure to the model of the oscilloscope under test:
Models TDS 540A, 544A OnlyĊWhen using Table 1Ć2 to test CH 1ĊCH 4; ignore the columns for AUX 1 & AUX 2 settings and limits.
Model TDS 520A, 524A OnlyĊUse Table 1Ć2 to test input chanĆ nels; use the columns for CH 1ĊCH 4 when testing CH 1 and CH 2; use the columns for AUX 1 and AUX 2 when testing those channels.
TableĂ1Ć2:ăDC Offset Accuracy (Zero Setting)
Vertical Scale
Setting
CH 1 ­CH 4
1ĂmV
AUX 1 & AUX 2
100ĂmV 0 ±1.6ĂmV ±10ĂmV
Vertical Position
and Offset Setting
1
Offset Accuracy Limits
CH1-CH4 AUX1&AUX2
100ĂmV 1V 0 ±25ĂmV ±100ĂmV
1ĂV 10ĂV 0 ±250ĂmV ±1ĂV
1
Vertical position is set to 0 divisions and vertical offset to 0 V when the oscilloscope is
initialized in step 1.
b. Set the vertical scale: Set the vertical SCALE to one of the settings
listed in Table 1Ć2 that is not yet checked. (Start with the first setting listed.)
c. Display the test signal: The baseline DC test level was initialized for
all channels in step 1 and is displayed as you select each channel and its vertical scale. Be sure not to use the vertical POSITION knob while checking any channel for accuracy of offset, since varying the position invalidates the check.
d. Measure the test signal: Rotate the general purpose knob to superĆ
impose the active cursor over the baseline DC test level. (Ignore the other cursor.)
1Ć24
e. Read the measurement results at the absolute (@:) cursor readout,
not the delta (D:) readout on screen (see Figure 1Ć4).
f. Check against limits: Do the following subparts in the order listed.
H CHECK that the measurement results are within the limits listed
for the current vertical scale setting.
H Repeat substeps b through f until all vertical scale settings
settings listed in Table 1Ć2 are checked for the channel under test.
g. Test all channels: Repeat substeps a through f for all input channels.
3. Disconnect the hookup: No hookup was required.
Performance Verification Procedures
Ignore the
inactive cursor.
Align the active
cursor to the DC
baseline (no input).
Then read the
offset relative to
ground reference.
Figure 1Ć4:ăMeasurement of DC Offset Accuracy at Zero Setting
Performance Tests
Check DC Gain and Voltage Measurement Accuracy
WARNING
Performance of this procedure requires input voltages up to 130 VDC. Be sure to set the DC calibration generator to 0 volts before connecting, disconnecting, and/or moving the test hookup during the performance of this procedure.
Equipment Required: Two dualĆbanana connectors (Item 5), one BNC T connector (Item 6), one DC calibration generator (Item 8), and two precision coaxial cables (Item 4).
Prerequisites: The oscilloscope must meet the prerequisites listed on page 1Ć15.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć25
Performance Tests
Dual Banana to
BNC Adapters
Procedure:
1. Install the test hookup and preset the instrument controls (see FigĆ ure 1Ć5):
DC Calibrator
Output Sense
HI
LO
50 W Coaxial Cables
BNC T
Connector
Figure 1Ć5:ăInitial Test Hookup
a. Hook up the testĆsignal source:
H Set the output of a DC calibration generator to 0 volts.
H Connect the output of a DC calibration generator through a
dualĆbanana connector followed by a 50 W precision coaxial cable to one side of a BNC T connector.
H Connect the Sense output of the generator through a second
dualĆbanana connector followed by a 50 W precision coaxial cable to the other side of the BNC T connector. Now connect the BNC T connector to CH 1.
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
1Ć26
c. Modify the default settings:
H Press SHIFT; then ACQUIRE MENU.
H Press the mainĆmenu button Mode; then press the sideĆmenu
button Average 16.
H Press CURSOR.
H Press the mainĆmenu button Function; then press the sideĆmeĆ
nu button H Bars.
H Press DISPLAY.
Performance Verification Procedures
Performance Tests
H Press the mainĆmenu button Graticule; then press the sideĆmeĆ
nu button Frame.
2. Confirm input channels are within limits for DC delta voltage accuracy: Do the following substeps Ċ test CH 1 first, skipping substep a since
CH 1 is already selected from step 1.
a. Select an unchecked channel:
H Set the generator output to 0 V.
H Press WAVEFORM OFF to remove the channel just confirmed
from the display.
H Press the frontĆpanel button that corresponds to the next chanĆ
nel you are to confirm.
H Move the test hook up to the channel you select.
b. Display the test signal:
H Press VERTICAL MENU. Press the mainĆmenu button Position.
H Use the keypad to set vertical position to -2.5 divisions (press
-2.5, then ENTER, on the keypad).
c. Measure the test signal:
H Press CLEAR MENU. Use the general purpose knob to preciseĆ
ly align the active cursor to the DC baseline level on screen.
H Set the generator output to 500 mV.
H Press SELECT. Use the general purpose knob to precisely align
the alternate cursor to the 500 mV DC test level on screen.
H Press CLEAR MENU. Read the measurement results from the
delta (D:) readout, not the absolute (@:) readout. See Figure 1Ć6 on page 1Ć28.
d. Check against limits: CHECK that the D: readout on screen is within
485 mV to 515 mV (see Figure 1Ć6).
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć27
Performance Tests
First align a cursor to the
DC baseline (no input).
Second align the second
cursor to the DC test level
Third read the results
that you input.
of the
measurement here.
DDC
Figure 1Ć6:ăMeasurement of the DC Accuracy for Delta Measurements
e. Test all channels: Repeat substeps a through d for all four channels.
3. Reestablish the initial test hookup setup:
a. Hook up the testĆsignal source:
H Set the output of a DC calibration generator to 0 volts.
H Move the BNC T connector back to CH 1.
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
c. Modify the default settings:
H Press SHIFT; then ACQUIRE MENU.
H Press the mainĆmenu button Mode; then press the sideĆmenu
button Average 16.
H Press DISPLAY.
H Press the mainĆmenu button Graticule; then press the sideĆ
menu button Frame.
1Ć28
Performance Verification Procedures
Performance Tests
4. Confirm input channels are within limits for DC accuracy at maximum offset and position: Do the following substeps Ċ test CH 1 first, skipping substep a since CH 1 is already selected from step 3.
a. Select an unchecked channel:
H Press WAVEFORM OFF to remove the channel just confirmed
from the display.
H Press the frontĆpanel button that corresponds to the channel
you are to confirm.
H Set the generator output to 0 V.
H Move the test hookup to the channel you select.
b. Turn on the measurement Mean for the channel:
H Press MEASURE, then press the mainĆmenu button Select
Measrmnt for CHx.
H Press the side menu button more until the menu label Mean
appears in the side menu (its icon is shown at the left). Press the sideĆmenu button Mean.
H Press CLEAR MENU.
Follow these rules to match this procedure to the model of the oscilloscope under test:
Models TDS 540A, 544A OnlyĊUse Table 1Ć3 to test CH 1ĊCH 4; ignore Table 1Ć4 AUX 1 & AUX 2 settings and limits.
Model TDS 520A, 524A OnlyĊUse Table 1Ć3 to test CH 1 and CH 2 only; use Table 1Ć4 to test AUX 1 and AUX 2 only.
c. Set its vertical scale: Set the vertical SCALE to one of the settings
listed in Table 1Ć3 (and Table 1Ć4 for the TDS 520A and 524A) that is not yet checked. (Start with the first setting listed.)
TableĂ1Ć3:ăDC Accuracy: CH 1-CH 4
Scale Setting
Position Setting
Offset
Setting
Generator Setting
Accuracy Limits
(Divs)
5ĂmV -5 +1 V +1.040 V +1.0355 V to +1.0445 V
+5 -1 V -1.040 V -1.0355 V to -1.0445 V
200ĂmV -5 +10 V +11.6 V +11.525 V to +11.675 V
+5 -10 V -11.6 V -11.525 V to -11.675 V
1ĂV -5 +100 V +108 V +107.450 V to +108.550 V
+5 -100 V -108 V -107.450 V to -108.550 V
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć29
Performance Tests
TableĂ1Ć4:ăDC Accuracy: AUX 1-AUX 2
Scale Setting
Position Setting
Offset Setting
Generator Setting
Accuracy Limits
(Divs)
100ĂmV -5 +0.5 V +1.3 V +1.271 V to +1.329 V
+5 -0.5 V -1.3 V -1.271 V to -1.329 V
1V -5 +5 V +13 V +12.710 V to +13.290 V
+5 -5 V -13 V -12.710 V to -13.290 V
10ĂV -5 +50 V +130 V +127.10 V to +132.90 V
+5 -50 V -130 V -127.10 V to -132.90 V
d. Display the test signal:
H Press VERTICAL MENU. Press the mainĆmenu button Position.
H Use the keypad to set vertical position to -5 divisions (press
-5, then ENTER, on the keypad). The baseline level will move off screen.
H Press the mainĆmenu button Offset.
H Use the keypad to set vertical offset to the positiveĆpolarity
setting listed in the table for the current vertical scale setting. The baseline level will remain off screen.
H Set the generator to the level and polarity indicated in the table
for the vertical scale, position, and offset settings you have made. The DC test level should appear on screen. (If it doesn't return, the DC accuracy check is failed for the current vertical scale setting of the current channel.)
e. Measure the test signal: Press CLEAR MENU. Read the measureĆ
ment results at the Mean measurement readout. See Figure 1Ć7.
1Ć30
Performance Verification Procedures
First set vertical and position offsets to maximum (no input). Note gnd ref indicator bounded onĆscreen for the
offset baseline below screen.
Second, input a DC
level equal to the offset
plus 3 divisions.
Third, turn on the
Measurement called
mean and read the
results here.
Figure 1Ć7:ăMeasurement of DC Accuracy at Maximum Offset and Position
Performance Tests
f. Check against limits:
H CHECK that the readout for the measurement Mean readout on
screen is within the limits listed for the current vertical scale and position/offset/generator settings.
H Repeat step d, reversing the polarity of the position, offset, and
generator settings as is listed in the table.
H CHECK that the Mean measurement readout on screen is within
the limits listed for the current vertical scale setting and position/ offset/generator settings.
H Repeat substeps c through f until all vertical scale settings
settings listed in Table 1Ć3 (and Table 1Ć4 for the TDS 520A and 524A) are checked for the channel under test.
g. Test all channels: Repeat substeps a through f for all four channels.
5. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Then disconnect the cable from the generator output at the input
connector of the channel last tested.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć31
Performance Tests
Check Analog Bandwidth
Equipment Required: One highĆfrequency leveled sine wave generator and
its leveling head (Item 11), plus two10X attenuators (Item 1).
Prerequisites: See page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
b. Modify the default settings:
H Set the horizontal SCALE to 50 ns.
H Now press SHIFT, then ACQUIRE MENU.
H Press the mainĆmenu button Mode; then press the sideĆmenu
button Average 16.
H Press TRIGGER MENU.
H Press the mainĆmenu button Coupling; then press the sideĆ
menu button Noise Rej.
H Press Measure. Now press the mainĆmenu button High-Low
Setup; then press the sideĆmenu button Min-Max.
c. Hook up the testĆsignal source: Connect, through its leveling head,
the sine wave output of a highĆfrequency leveled sine wave generaĆ tor to CH 1. Set the output of the generator to a reference frequency of 6ĂMHz. See Figure 1Ć8.
High Frequency Sine Wave
Generator
Output
Leveling Head
1Ć32
Figure 1Ć8:ăInitial Test Hookup
Performance Verification Procedures
Performance Tests
2. Confirm the input channels are within limits for analog bandwidth: Do the following substeps Ċ test CH 1 first, skipping substeps a and b since
CH 1 is already set up for testing from step 1.
a. Select an unchecked channel:
H Press WAVEFORM OFF to remove the channel just confirmed
from display.
H Press the frontĆpanel button that corresponds to the channel
you are to confirm.
H Move the leveling head to the channel you select.
b. Match the trigger source to the channel selected:
H Press TRIGGER MENU.
H Press the mainĆmenu button Source.
H Press the sideĆmenu button that corresponds to the channel
selected.
c. Set its input impedance:
H Press VERTICAL MENU; then press the mainĆmenu button
Coupling.
H Press the sideĆmenu W button to toggle it to the 50 W setting.
d. Set the vertical scale: Set the vertical SCALE to one of the settings
listed in Table 1Ć5 not yet checked. (Start with the 100 mV setting.)
TableĂ1Ć5:ăAnalog Bandwidth
Vertical Scale
Attenuators (10X)
Reference Amplitude (at 6 MHz)
Horizontal Scale
Test Frequency
Limits
100ĂmV none 600ĂmV (6 divisions) 1ns 500 MHz w424ĂmV
1V none 5 V (5 divisions) 1ns 500 MHz w3.535 V
500ĂmV none 3 V (6 divisions) 1ns 500 MHz w2.121 V
200ĂmV none 1.2ĂV (6 divisions) 1ns 500 MHz w848 mV
50ĂmV 1 300 mV (6 divisions) 1ns 500 MHz w212 mV
20ĂmV 1 120ĂmV (6 divisions) 1ns 500 MHz w84 mV
10ĂmV 1 60ĂmV (6 divisions) 1ns 500 MHz w42ĂmV
5ĂmV 2 30ĂmV (6 divisions) 1ns 500 MHz w21ĂmV
2ĂmV 2 12ĂmV (6 divisions) 2ns 350 MHz w8.48ĂmV
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć33
Performance Tests
e. Display the test signal: Do the following subparts to first display the
reference signal and then the test signal.
H Press MEASURE; then press the mainĆmenu button Select
Measrmnt for CHx.
H Now press the side menu button more until the menu label
PkĆPk appears in the side menu (its icon is shown at the left). Press the sideĆmenu button PkĆPk.
H Press CLEAR MENU.
H Set the generator output so the CHx PkĆPk readout equals the
reference amplitude in Table 1Ć5 that corresponds to the vertical scale set in substep d.
H Press the frontĆpanel button SET LEVEL TO 50% as necessary
to trigger a stable display.
f. Measure the test signal:
H Increase the frequency of the generator output to the test freĆ
quency in Table 1Ć5 that corresponds to the vertical scale set in substep d.
First, increase the reference
frequency to the test
frequency; then decrease the
horizontal scale.
Second, read the results from
the readout of measurement
PkĆPk.
H Set the horizontal SCALE to 1 ns. Press SET LEVEL TO 50% as
necessary.
H Read the results at the CHx PkĆPk readout, which will automatiĆ
cally measure the amplitude of the test signal. See Figure 1Ć9.
1Ć34
Figure 1Ć9:ăMeasurement of Analog Bandwidth
Performance Verification Procedures
Performance Tests
g. Check against limits:
H CHECK that the PkĆPk readout on screen is within the limits
listed in Table 1Ć5 for the current vertical scale setting.
H When finished checking, set the horizontal SCALE back to the
50 ns setting.
Checking each channel's bandwidth at all vertical scale settings is time consuming and unnecessary. You may skip checking the remaining vertical scale settings in Table 1Ć5 (that is, skip the followĆ ing substep, h) if this digitizing oscilloscope has performed as follows:
H Passed the 100 mV vertical scale setting just checked in this
procedure.
H Passed the Verify Internal Adjustment, Self Compensation, and
Diagnostics procedure found under Self Tests, on page 1Ć4.
NOTE
Passing the signal path compensation confirms the signal path for all vertical scale settings for all channels. Passing the internal diagnostics ensures that the factoryĆset adjustment constants that control the bandwidth for each vertical scale setting have not changed.
h. Check remaining vertical scale settings against limits (optional):
H If desired, finish checking the remaining vertical scale settings
for the channel under test by repeating substeps d through g for each of the remaining scale settings settings listed in Table 1Ć5 for the channel under test.
H When doing substep e, skip the subparts that turn on the CHx
PkĆPk measurement until you check a new channel.
H Install/remove 10X attenuators between the generator leveling
head and the channel input as is needed to obtain the six diviĆ sion reference signals listed in the table.
i. Test all channels: Repeat substeps a through g for all four channels.
3. Disconnect the hookup: Disconnect the test hook up from the input connector of the channel last tested.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć35
Performance Tests
Check Delay Between Channels
Equipment Required: One mediumĆfrequency leveled sineĆwave generator
(Item 10), one precision, 50 W, coaxial cable (Item 4), one 50 W terminator (Item 3), and a dualĆinputĆcoupler (Item 7).
Prerequisites: See page 1Ć15.
Procedure:
DO NOT use the vertical position knob to reposition any channel while doing this check. To do so invalidates the test.
1. Install the test hookup and preset the instrument controls:
a. Initialize the front panel;
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
b. Modify the initialized frontĆpanel control settings:
H Do not adjust the vertical position of any channel during this
procedure.
H Set the horizontal SCALE to 500 ps.
H Press SHIFT; then ACQUIRE MENU.
H Press the mainĆmenu button Mode, and then press the sideĆmeĆ
nu button Average 16.
c. Hook up the testĆsignal source:
H Connect, through a 50 W precision coaxial cable, followed by a
50 W termination, the sine wave output of a mediumĆfrequency sine wave generator to a dualĆinput coupler. See Figure 1Ć10.
Medium Frequency Sine Wave
Generator
Output
1Ć36
50 W Cable
50 W Terminator
Figure 1Ć10:ăInitial Test Hookup
Performance Verification Procedures
Dual Input Coupler
Performance Tests
H Connect the coupler to both CH 1 and CH 2.
2. Confirm CH 1 through CH 4 (CH 2 for 520A and 524A) are within limits for
channel delay:
a. Set up the generator: Set the generator frequency to 250 MHz and
the amplitude for about five divisions in CH 1.
Hint: as you are adjusting the generator amplitude, push SET LEVĆ EL TO 50% frequently to speed up the updating of the waveform amplitude on screen.
TDS 520A and 524A only: Press CH 2; then skip to substep e and continue this check. If testing a TDS 540A or 544A model, continue with the next substep, b.
b. Save a CH 2 waveform: Press CH 2; then press save/recall WAVEĆ
FORM. Now, press the mainĆmenu button Save Wfm Ch2; then press the sideĆmenu button To Ref 2.
c. Save CH 3 waveform: Move the coupler from CH 2 to CH 3, so that
CH 1 and CH 3 are driven. Press CH 3; then press the sideĆmenu button To Ref 3.
d. Display all test signals:
H Press WAVEFORM OFF twice to remove CH 2 and CH 3 from
the display.
H Move the coupler from CH 3 to CH 4, so that CH 1 and CH 4 are
driven. Press CH 4.
H Now, press the frontĆpanel button MORE. Press the mainĆmenu
buttons Ref 2 and Ref 3.
e. Measure the test signal:
H Locate the point on the rising edge of the leftĆmost waveform
where it crosses the center horizontal graticule line. This is the
time reference point for this waveform. Note the corresponding time reference point for rightĆmost waveform. See Figure 1Ć11.
H Press CURSOR; then press the sideĆmenu button V Bars.
H Press CLEAR MENU.
H Rotate the General Purpose knob to align one cursor to the time
reference point of the leftĆmost waveform edge and the other cursor to the time reference point of the rightĆmost waveform edge. (Press SELECT to switch between the two cursors.) See Figure 1Ć11.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć37
Performance Tests
First, display the live/reference
waveforms for channels. Note their
overlapping ground reference
indicators.
H Read the measurement results at the D: cursor readout, not the
@: readout on screen.
Second, identify the time reference
points of those waveforms.
Third, turn on the cursor and align
the V bar cursors to the time
reference points.
Fourth, read the results
here.
Figure 1Ć11:ăMeasurement of Channel Delay
f. Check against limits: CHECK that the cursor readout on screen is
v250 ps.
3. Disconnect the hookup: Disconnect the cable from the generator output at the input connectors of the channels.
1Ć38
Performance Verification Procedures
Performance Tests
Time Base System Checks
These procedures check those characteristics that relate to the Main and Delayed time base system and are listed as checked under Warranted Characteristics in Section 2, Specifications.
Check Accuracy for LongĆTerm Sample Rate, Delay Time, and Delta Time Measurements
Equipment Required: One timeĆmark generator (Item 12) and one precision
coaxial cable (Item 4).
Prerequisites: See page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Hook up the testĆsignal source: Connect, through a 50 W precision
coaxial cable, the timeĆmark output of a timeĆmark generator to CH 1. Set the output of the generator for 10 ms markers. See FigĆ ure 1Ć12.
TimeĆMark Generator
Output
50 W Coaxial Cables
Figure 1Ć12:ăInitial Test Hookup
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
c. Modify the initialized frontĆpanel control settings:
H Set the vertical SCALE to 500 mV.
H Press VERTICAL MENU; then press the mainĆmenu button
Coupling. Press the sideĆmenu W button to 50 W.
H Press SET LEVEL TO 50%.
H Use the vertical POSITION knob to center the test signal on
screen.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć39
Performance Tests
H Set the horizontal SCALE of the Main time base to 1 ms.
H Press TRIGGER MENU; then press the mainĆmenu button
Mode & Holdoff. Now press the sideĆmenu button Normal.
H Press SET LEVEL TO 50%.
H Press horizontal MENU. Press the mainĆmenu button Record
Length; then press the sideĆmenu button 1000 points in 20 divs.
H Press the mainĆmenu button Trigger Position. Press the sideĆ
menu button Pretrigger; then set pretrigger to 20%: press 20, then ENTER, on the keypad.
2. Confirm Main and Delayed time bases are within limits for accuracies:
a. Display the test signal:
H Adjust the horizontal POSITION so the trigger T is aligned to the
center vertical graticule line.
H Press the mainĆmenu button Time Base.
H Press the sideĆmenu buttons Delayed Only and Delayed Runs
After Main.
First, the trigger T is aligned
to the center graticule line.
Second, the horizontal
modes are set.
Third, the horizontal scale for
D time base is set and a
10 ms delay is entered.
Fourth, the waveforms rising edge is
checked to be within ±2.5 horizontal
divisions of the center horizontal
graticule line.
b. Measure the test signal:
H Set the horizontal SCALE of the D (delayed) time base to
100 ns.
H Use the keypad to set delayed time to 10 ms. (Press 10, then
SHIFT, then m followed by ENTER.)
1Ć40
Figure 1Ć13:ăMeasurement of Accuracy Ċ LongĆTerm and DelayĆTime
Performance Verification Procedures
Performance Tests
c. Check longĆterm sample rate and delay time accuracies against
limits: CHECK that the rising edge of the marker crosses the center
horizontal graticule line at a point within ±2.5 divisions of center graticule. See Figure 1Ć13.
d. Check deltaĆtime accuracy against limits:
H Press the sideĆmenu button Main Only. Set horizontal SCALE to
2 ns.
H Set the output of the generator for 20 ns markers.
H Press SET LEVEL TO 50%.
H Press SHIFT; then ACQUIRE MENU. Next, press the mainĆmenu
button Mode. Finally, press the sideĆmenu button Average.
H Enter 8, for eight averages, on the keypad.
H Press MEASURE.
H Press the mainĆmenu button HighĆLow Setup; then press the
sideĆmenu button MinĆMax.
H Press the mainĆmenu button Select Measrmnt for Ch1.
H Press the sideĆmenu button -more-, until PERIOD appears in
the side menu. Press PERIOD.
H Press CLEAR MENU.
H CHECK that the readout for CH 1 Per is within 19.760 ns to
20.240 ns.
3. Disconnect the hookup: Disconnect the cable from the generator output at the input connector of CH 1.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć41
Performance Tests
Trigger System Checks
These procedures check those characteristics that relate to the Main and Delayed trigger systems and are listed as checked under Warranted CharacĆ teristics in Section 2, Specifications.
Check Accuracy (Time) for PulseĆGlitch or PulseĆWidth Triggering
Equipment Required: One mediumĆfrequency leveled sine wave generator
(Item 10), one 10X attenuator (Item 1), and one precision, 50 W, coaxial cable (Item 4).
Prerequisites: See page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Initialize the instrument:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
b. Modify the default setup:
H Press vertical MENU.
H Press the mainĆmenu button Coupling; then press the sideĆmeĆ
nu W button to select 50 W coupling.
H Set the horizontal SCALE to 10 ns.
c. Hook up the testĆsignal source: Connect, through a 50 W precision
coaxial cable, followed by a 10X attenuator, the output of a mediumĆ frequency leveled sine wave generator (Item 10) to CH 1. See FigĆ ure 1Ć14.
Medium Frequency Sine Wave
Generator
Output
50 W Coaxial Cable
10X Attenuator
Figure 1Ć14:ăInitial Test Hookup
1Ć42
Performance Verification Procedures
Performance Tests
2. Confirm the trigger system is within timeĆaccuracy limits for pulseĆglitch or pulseĆwidth triggering (Horizontal Scale v1 ms):
a. Display the test signal: Set the output of the sine wave generator for
a 100ĂMHz, fiveĆdivision sine wave on screen. Press SET LEVEL TO 50%.
b. Set the trigger mode: Press TRIGGER MENU. Now press the mainĆ
menu button Mode & Holdoff; then the sideĆmenu button Normal.
c. Set upper and lower limits that ensures triggering:
H Press the mainĆmenu button Type; then repeatedly press the
same button until Pulse is highlighted in the menu that pops up.
H Press the mainĆmenu button Class; then repeatedly press the
same button until Width is highlighted in the menu that pops up.
H Press the mainĆmenu button Trig When; then press the sideĆmeĆ
nu button Within Limits.
H Press the sideĆmenu button Upper Limit. Use the keyboard to
set the upper limit to 10 ns: press 10; then SHIFT; then n; then ENTER.
H Press the sideĆmenu button Lower Limit. Use the keypad to set
the lower limit to 2 ns.
d. Check against limits:
H Press SET LEVEL TO 50%.
H While doing the following subparts, monitor the display (it will
stop acquiring) and the frontĆpanel light TRIG (it will extinguish) to determine when triggering is lost.
H Use the general purpose knob to increase the Lower Limit
readout until triggering is lost.
H CHECK that the Lower Limit readout is within 2.5 ns to 7.5 ns,
inclusive.
H Use the keypad to return the Lower Limit to 2 ns and reestabĆ
lish triggering.
H Press the sideĆmenu button Upper Limit; then use the general
purpose knob to slowly decrease the the Upper Limit until triggering is lost.
H CHECK that the Upper Limit readout is within 2.5 ns to 7.5 ns,
inclusive.
readout
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć43
Performance Tests
First, the upper and lower limits
are set so the test waveform
triggers within limits.
Second, a limit (here, the lower
limit) is increased until triggering
Figure 1Ć15:ăMeasurement of Time Accuracy for Pulse and Glitch Triggering
stops.
3. Confirm the trigger system is within timeĆaccuracy limits for pulseĆglitch or pulseĆwidth triggering (horizontal scale >1 ms):
a. Set upper and lower limits that ensure triggering at 250 kHz:
H Press the sideĆmenu button Upper Limit. Use the keyboard to
set the upper limit to 4 ms.
H Press the sideĆmenu button Lower Limit. Use the keypad to set
the lower limit to 500 ns.
b. Display the test signal:
H Set the horizontal SCALE to 5 ms.
H Set the output of the sineĆwave generator for a 250 kHz, fiveĆdiviĆ
sion sine wave on screen. Set the vertical SCALE to 20 mV (the waveform will overdrive the display).
H Press SET LEVEL TO 50%.
c. Check against limits: Do the following subparts in the order listed.
H Use the general purpose knob to increase Lower Limit readout
until triggering is lost.
H CHECK that the Lower Limit readout is within 1 msto3 ms,
inclusive.
1Ć44
H Use the keypad to return the Lower Limit to 500 ns and reesĆ
tablish triggering.
Performance Verification Procedures
Performance Tests
H Press the sideĆmenu button Upper Limit; then use the general
purpose knob to slowly decrease the the Upper Limit readout until triggering is lost.
H CHECK that the Upper Limit readout is within 1 msto3 ms,
inclusive.
4. Disconnect the hookup: Disconnect the cable from the generator output at the input connector of CH 1.
Check Accuracy, TriggerĆlevel or Threshold, DC Coupled
Equipment Required: One DC calibration generator (Item 8), one BNC T
connector (Item 6), and two precision, 50 W, coaxial cables (Item 4).
Prerequisites: The oscilloscope must meet the prerequisites listed on page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Hook up the testĆsignal source:
Dual Banana to
BNC Adapter
H Set the output of a DC calibration generator to 0 volts.
H Connect the output of a DC calibration generator through a
H Connect the Sense output of the generator, through a second
DC Calibrator
Output Sense
dualĆbanana connector followed by a 50 W precision coaxial cable to one side of a BNC T connector. See Figure 1Ć16.
dualĆbanana connector followed by a 50 W precision coaxial cable, to other side of the BNC T connector. Now connect the BNC T connector to CH 1. See Figure 1Ć16.
HI
LO
50 W Coaxial Cables
BNC T Connector
TDS 520A, 524A, 540A, & 544A Performance Verification
Figure 1Ć16:ăInitial Test Hookup
1Ć45
Performance Tests
b. Initialize the oscilloscope:
H Press save/recall Setup.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
c. Select Delayed Triggerable:
H Press HORIZONTAL MENU.
H Press the mainĆmenu button Time Base.
H Press the sideĆmenu button Delayed Triggerable.
2. Confirm Main trigger system is within limits for TriggerĆlevel/Threshold accuracy:
a. Display the test signal:
H Press VERTICAL MENU. Press the mainĆmenu button Position.
H Use the keypad to set vertical position to -3 divisions (press
-3, then ENTER, on the keypad.) The baseline level will move down three divisions.
H Press the mainĆmenu button Offset.
H Use the keypad to set vertical offset to +10 volts. Press 10, then
ENTER. The baseline level will move offĆscreen.
H Set the standard output of a DC calibration generator to
+10 volts. The DC test level will appear on screen.
b. Measure the test signal:
H Press SET LEVEL TO 50%.
H Press TRIGGER MENU.
H Read the measurement results from the readout below the label
Level in the menu; not the trigger readout in the graticule area.
c. Check against limits:
H CHECK that the Level readout in the main menu is within
9.940 V to 10.060 V, inclusive.
H Press TRIGGER MENU. Press the mainĆmenu button Slope;
then press the sideĆmenu button for negative slope. (See icon at left.) Repeat substep b.
H CHECK that the Level readout in the main menu is within
9.940 V to 10.060 V, inclusive. See Figure 1Ć17.
1Ć46
Performance Verification Procedures
First, set vertical offset to maximum
and vertical position to -3 divisions.
Second, set input equal to the offset to return the DC
level to the screen.
Third, push SET LEVEL to 50%
and check the results in the
main menu under Level."
Figure 1Ć17:ăMeasurement of TriggerĆlevel Accuracy
Performance Tests
3. Confirm Delayed trigger system is within limits for TriggerĆlevel/Threshold accuracy:
a. Select the Delayed time base:
H Press HORIZONTAL MENU.
H Press the mainĆmenu button Time Base.
H Press the sideĆmenu buttons Delayed Only and Delayed TrigĆ
gerable.
H Set D (delayed) horizontal SCALE to 500 ms.
b. Select the Delayed trigger system:
H Press SHIFT; then press the frontĆpanel button DELAYED TRIG.
H Press the mainĆmenu button Level.
c. Measure the test signal: Press the sideĆmenu button SET TO 50%.
The TRIG'D indicator should be lit. Read the measurement results in the side menu below the label Level.
d. Check against limits: Do the following subparts in the order listed.
H CHECK that the Level readout in the side menu is within 9.940 V
to 10.060 V, inclusive.
H Press the mainĆmenu button Slope; then press the sideĆmenu
button for negative slope. (See icon at left.) Press the mainĆmeĆ nu button Level. Repeat substep c.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć47
Performance Tests
H CHECK that the Level readout in the side menu is within 9.940 V
to 10.060 V, inclusive.
4. Disconnect the hookup:
a. First set the output of the DC calibration generator to 0 volts.
b. Then disconnect the cable from the generator output at the input
connector of CH 1.
Sensitivity, Edge Trigger, DC Coupled
Equipment Required: One mediumĆfrequency leveled sine wave generator
(Item 10), one highĆfrequency leveled sine wave generator (Item 11), one precision 50ĂW coaxial cable (Item 4), and one 10X attenuator (Item 1). When checking the TDS 540A and 544A, a BNC T connector (Item 6), a 5X attenuator (Item 2), and a second precision 50ĂW coaxial cable (Item 4) are also required.
Prerequisites: See page 1Ć15.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
H Press the sideĆmenu button OK Confirm Factory Init.
b. Modify the initialized frontĆpanel control settings:
H Set the horizontal SCALE for the M (main) time base to 20 ns.
H Press HORIZONTAL MENU; then press the mainĆmenu button
Time Base.
H Press the sideĆmenu button Delayed Only; then the sideĆmenu
button Delayed Triggerable.
H Set the horizontal SCALE for the D (delayed) time base to 20 ns;
then press the sideĆmenu button Main Only.
H Press TRIGGER MENU; then press the mainĆmenu button
Mode & Holdoff. Now press the sideĆmenu button Normal.
H Press VERTICAL MENU; then press the mainĆmenu button
Coupling. Now press the sideĆmenu W button and select the 50 W setting.
1Ć48
H Press SHIFT; then press ACQUIRE MENU. Now press the
mainĆmenu button Mode; then the sideĆmenu Average 16 button.
Performance Verification Procedures
Performance Tests
c. Hook up the testĆsignal source:
Medium Frequency Sine Wave Generator
Output
50 W Coaxial Cables
Figure 1Ć18:ăInitial Test HookupĊTDS 520A or 524A Only
H TDS 520A, 524A only: Connect, through a 50 W precision coaxĆ
ial cable, the signal output of a mediumĆfrequency sine wave generator to CH 1. See Figure 1Ć18.
To AUX TRIG INPUT
on Rear Panel
Medium
Frequency
Sine Wave Generator
Output
50 W Coaxial Cables
Figure 1Ć19:ăInitial Test HookupĊTDS 540A or 544A Only
H TDS 540A or 544A only: Connect the signal output of a mediumĆ
frequency sine wave generator to a BNC T connector. Connect one output of the T connector to CH 1 through a 50 W precision coaxial cable; connect the other output of the T connector to the AUX TRIG INPUT at the rear panel. See Figure 1Ć19.
2. Confirm Main and Delayed trigger systems are within sensitivity limits
(50 MHz):
a. Display the test signal:
H Set the generator frequency to 50 MHz.
H Press MEASURE.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć49
Performance Tests
H Press the mainĆmenu button HighĆLow Setup; then press the
sideĆmenu button MinĆMax.
H Press the mainĆmenu button Select Measrmnt for Ch1.
H Press the sideĆmenu button -more- until Amplitude appears
in the side menu (its icon is shown at the left). Press the sideĆ menu button Amplitude.
H Press SET LEVEL TO 50%.
H Press CLEAR MENU.
H Set the test signal amplitude for about three divisions on screen.
Now fine adjust the generator output until the CH 1 Amplitude readout indicates the amplitude is 350 mV. (Readout may fluctuĆ ate around 350 mV.)
H Disconnect the 50 W precision coaxial cable at CH 1 and reconĆ
nect it to CH 1 through a 10X attenuator.
b. Check for Main trigger system for stable triggering at limits:
H Read the following definition: A stable trigger is one that is
consistent; that is, one that results in a uniform, regular display triggered on the selected slope (positive or negative). This display should not have its trigger point switching between opposite slopes, nor should it roll" across the screen. At horiĆ zontal scale settings of 2Ăms/division and faster, TRIG'D will remain constantly lit. It will flash for slower settings.
H Press TRIGGER MENU; then press the mainĆmenu button
Slope.
H Press SET LEVEL TO 50%. CHECK that a stable trigger is
obtained for the test waveform on both the positive and negative slopes. (Use the side menu to switch between trigger slopes; use the TRIGGER LEVEL knob to stabilize the trigger if reĆ quired.) See Figure 1Ć20.
1Ć50
Performance Verification Procedures
First, set a signal with an
amplitude at the minimum
trigger sensitivity.
Second, check for a stable
trigger at both the positive
and negative slope settings.
Figure 1Ć20:ăMeasurement of Trigger Sensitivity
Performance Tests
H Leave the Main trigger system triggered on the positive slope of
the waveform before continuing to the next step. (The Main trigger system must be triggered to check the delayed trigger system in the next step.)
c. Check delayed trigger system for stable triggering at limits: Do the
following subparts in the order listed.
H Press HORIZONTAL MENU; then press the mainĆmenu button
Time Base. Now press the sideĆmenu button Delayed Only.
H Press SHIFT; then press DELAYED TRIG. Press the mainĆmenu
button Level.
H Press the sideĆmenu button SET TO 50%. CHECK that a stable
trigger is obtained for the test waveform for both the positive and negative slopes of the waveform. (Use the General Purpose knob to stabilize the trigger if required.) Press the mainĆmenu button Slope; then use the side menu to switch between trigger slopes.
H Leave the delayed trigger system triggered on the positive slope
of the waveform before continuing to the next step. Also, return to the main time base: Press HORIZONTAL MENU; then press the mainĆmenu button Time Base. Now press the sideĆmenu button Main Only.
TDS 520A or 524A only: Skip to step 4 since the TDS 520A and 524A are not equipped with an AUX Trigger input. If testing the TDS 540A or 544A, continue with step 3.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć51
Performance Tests
3. Confirm the AUX Trigger input:
a. Display the test signal:
H Remove the 10X attenuator and reconnect the cable to CH 1.
H Set the test signal amplitude for about 2.5 divisions on screen.
H Now fine adjust the generator output until the CH 1 Amplitude
readout indicates the amplitude is 250 mV. (Readout may fluctuĆ ate around 250 mV.)
b. Check the AUX trigger source for stable triggering at limits: Do the
following in the order listed.
H Use the definition for stable trigger from step 2.
H Press TRIGGER MENU; then press the mainĆmenu button
Source.
H Press the sideĆmenu button -more- until the sideĆmenu label
Auxiliary appears; then press Auxiliary.
H Press SET LEVEL TO 50%. CHECK that a stable trigger is
obtained for the test waveform on both the positive and negative slopes. Press the mainĆmenu button Slope; then use the side menu to switch between trigger slopes. Use the TRIGGER LEVEL knob to stabilize the trigger if required.
H Leave the Main trigger system triggered on the positive slope of
the waveform before proceeding to the next check.
H Press the mainĆmenu button Source; then press the sideĆmenu
button -more- until CH 1 appears. Press CH 1.
4. Confirm that the Main and Delayed trigger systems are within sensitivity
limits (500 MHz):
a. Hook up the testĆsignal source: Disconnect the hookup installed in
step 1. Connect, through its leveling head, the signal output of a highĆfrequency leveled sineĆwave generator to CH 1.
b. Set the Main and Delayed Horizontal Scales:
H Set the horizontal SCALE to 500 ps for the M (Main) time base.
H Press HORIZONTAL MENU. Now press the mainĆmenu button
Time base; then press the sideĆmenu button Delayed TriggerĆ able.
H Press the sideĆmenu button Delayed Only.
H Set the horizontal SCALE to 500 ps for the D (Delayed) time
base. Press the sideĆmenu button Main Only.
c. Display the test signal:
1Ć52
H
Set the generator frequency to 500 MHz.
Performance Verification Procedures
Performance Tests
H Set the test signal amplitude for about five divisions on screen.
Now fine adjust the generator output until the CH 1 Amplitude readout indicates the amplitude is 500 mV. (Readout may fluctuĆ ate around 500 mV.)
H Disconnect the leveling head at CH 1 and reconnect it to CH 1
through a 5X attenuator.
d. Repeat step 2, substeps b and c only, since only the 500 MHz
frequency is to be checked here.
5. Confirm that the Main and Delayed trigger systems couple trigger signals from all channels: Doing the procedure Check Analog Bandwidth, which begins on page 1Ć32, checks coupling. If you have not done that proceĆ dure, do so after finishing this procedure. See the following note.
NOTE
Steps 1 through 4 confirmed trigger sensitivity for the Main and Delayed triggering systems using the CH 1 input. Doing the proceĆ dure Check Analog Bandwidth ensures that trigger signals are coupled from all four channels.
When checking delayed triggering sensitivity at 500 MHz, the waveform record may have some missing interpolated record points. The waveform is still stably triggered. (See definition of a stable trigger earlier in this procedure.)
6. Disconnect the hookup: Disconnect the cable from the generator output at the input connector of the channel last tested.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć53
Performance Tests
Output Signal Checks
The procedure that follows checks those characteristics of the output sigĆ nals that are listed as checked under Warranted Characteristics in Section 2, Specifications. The oscilloscope outputs these signals at its front and rear panels.
Check Outputs Ċ CH 3 and Main and Delayed Trigger ą(TDS 540A and 544A only)
Equipment Required: Two 50 W precision cables (Item 4), and one calibraĆ
tion generator (Item 9).
Prerequisites: See page 1Ć15. Also, this Digitizing Oscilloscope must have passed Check Accuracy Ċ LongĆTerm Sample Rate, Delay time, Time MeaĆ
surement on page 1Ć39 and Check Accuracy for DC Gain and Voltage MeaĆ surements on page 1Ć25.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Hook up testĆsignal source 1:
H Connect the standard amplitude output of a calibration generaĆ
tor through a 50 W precision coaxial cable to CH 3. See Figure 1Ć21.
H Set the output of the calibration generator to 0.500 V.
b. Hook up testĆsignal source 2: Connect the Main Trigger Out at the
rear panel to CH 2 through a 50 W precision cable. See Figure 1Ć21.
To Main
Trigger Out
Calibration
Generator
Output
50 W Coaxial Cables
Figure 1Ć21:ăInitial Test Hookup
c. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
1Ć54
H Press the sideĆmenu button OK Confirm Factory Init.
Performance Verification Procedures
Performance Tests
d. Modify the initialized frontĆpanel control settings:
H Set the horizontal SCALE to 200 ms.
H Press SHIFT; then ACQUIRE MENU.
H Press the mainĆmenu button Mode. Then press the sideĆmenu
button Hi Res.
2. Confirm Main and Delayed Trigger outputs are within limits for logic
levels:
a. Display the test signal:
H Press WAVEFORM OFF to turn off CH 1; then press CH 2 to
display that channel. Set the vertical SCALE to1V.
H Press TRIGGER MENU.
H Press the mainĆmenu button Source; then press the sideĆmenu
button CH 3. Press SET LEVEL TO 50%.
H Use the vertical POSITION knob to center the display on
screen.
First, turn on the
measurements
high and low.
Second, read the
measurement
results here.
b. Measure logic levels:
H Press MEASURE; then press the mainĆmenu button Select
Measrmnt for Ch2.
H Repeatedly press the sideĆmenu button -more- until High and
Low appear in the side menu (their icons are shown at the left). Press both sideĆmenu buttons High and Low. See Figure 1Ć22.
Figure 1Ć22:ăMeasurement of Main Trigger Out Limits
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć55
Performance Tests
c. Check Main Trigger output against limits:
H CHECK that the Ch2 High readout is w2.5 volts and that the
Ch2 Low readout is v700 mV.
H Press VERTICAL MENU; then press the mainĆmenu button
Coupling. Now press the sideĆmenu button W to toggle it to the
50 W setting.
H CHECK that the Ch2 High readout is w1.0 volt and that the Ch2
Low readout v250 mV.
d. Check Delayed Trigger output against limits:
H Move the precision 50 W cable from the Main Trigger Output
BNC to the Delayed Trigger Output BNC.
H CHECK that the Ch2 High readout is w1.0 volt and that the Ch2
Low readout v250 mV.
H Press the sideĆmenu button W select the 1 MW setting.
H Press CLEAR MENU.
H CHECK that the Ch2 High readout is w2.5 volts and that the
Ch2 Low readout is v700 mV.
3. Confirm CH 3 output is within limits for gain:
a. Measure gain:
H Move the precision 50 W cable from the rearĆpanel DELAYED
TRIGGER OUTPUT BNC to the rearĆpanel SIGNAL OUTPUT
BNC.
H Push SHIFT. Then push DELAYED TRIG.
H Press the mainĆmenu button Source. Then press the sideĆmenu
button Ch3.
H Push HORIZONTAL MENU.
H Push the mainĆmenu button Time Base. Then press the sideĆ
menu button Delayed Triggerable.
H Set vertical SCALE to 100 mV.
H Press MEASURE; then press the mainĆmenu button Select
Measrmnt for Ch2.
H Repeatedly press the sideĆmenu button -more- until PkĆPk
appears in the side menu (its icon is shown at the left). Press the sideĆmenu button PkĆPk.
H Press CLEAR MENU.
1Ć56
Performance Verification Procedures
Performance Tests
b. Check against limits:
H CHECK that the readout Ch2 PkĆPk is between 90ĂmV and
110 mV, inclusive.
H Press VERTICAL MENU; then press the sideĆmenu button W to
toggle to the 50 W setting.
H Set vertical SCALE to 10 mV; then press CLEAR MENU.
H CHECK that the readout Ch2 PkĆPk is between 45 mV and
55 mV, inclusive.
4. Disconnect the hookup: Disconnect the cable from the generator output at the input connector of the channel last tested.
Check Probe Compensator Outputs
One female BNC to clip adapter (Item 15), two dualĆbanana connectors (Item 5), one BNC T connector (Item 6), two 50 W precision cables (Item 4), and one DC calibration generator (Item 8).
Prerequisites: See page 1Ć15. Also, this Digitizing Oscilloscope must have passed Check Accuracy Ċ LongĆTerm Sample Rate, Delay time, Time MeaĆ
surement on page 1Ć39 and Check Accuracy for DC Gain and Voltage MeaĆ surements on page 1Ć25.
Procedure:
1. Install the test hookup and preset the instrument controls:
a. Hook up testĆsignal:
H Connect CH 1 to PROBE COMPENSATION SIGNAL and to
PROBE COMPENSATION GND through a precision coaxial
cable and a BNC to clip adapter. See Figure 1Ć23.
50 W Coaxial Cable
Female BNC to Clip Adapter
Figure 1Ć23:ăInitial Test Hookup
Black Lead to GND
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the mainĆmenu button Recall Factory Setup.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć57
Performance Tests
H Press the sideĆmenu button OK Confirm Factory Init.
c. Modify the initialized frontĆpanel control settings:
H Set the vertical SCALE to 100 mV as required.
H Set the horizontal SCALE to 200 ms.
H Press Set Level to 50% and use the VERTICAL POSITION
knob to center the display on screen.
H Press SHIFT; then ACQUIRE MENU.
H Press the mainĆmenu button Mode; then press the sideĆmenu
Average button.
H Select 128 averages. On the keypad, type 128; then press
ENTER.
2. Confirm that the Probe Compensator signal is within limits for frequency:
a. Measure the frequency of the probe compensation signal:
H Press MEASURE; then press the mainĆmenu button Select
Measrmnt for Ch1.
H Repeatedly press the sideĆmenu button -more- until FrequenĆ
cy appears in the side menu (its icon is shown at the left). Press the sideĆmenu button Frequency.
H Press CLEAR MENU to remove the menus from the display. See
Figure 1Ć24.
1Ć58
Figure 1Ć24:ăMeasurement of Probe Compensator Frequency
Performance Verification Procedures
Performance Tests
b. Check against limits: CHECK that the CH 1 Freq readout is within
950 Hz to 1.050 kHz, inclusive.
3. Confirm that the Probe Compensator signal is within limits for amplitude:
a. Save the probe compensation signal in reference memory:
H Press SAVE/RECALL WAVEFORM; then press the mainĆmenu
button Save Wfm Ch 1.
H Press the sideĆmenu button to Ref 1 to save the probe compenĆ
sation signal in reference 1.
H Disconnect the cable from CH 1 and the clips from the probe
compensation terminals.
H Press MORE; then press the mainĆmenu button Ref 1 to disĆ
played the stored signal.
H Press CH 1.
b. Hook up the DC standard source:
H Set the output of a DC calibration generator to 0 volts.
Dual Banana to
BNC Adapters
H Connect the output of a DC calibration generator through a
dualĆbanana connector followed by a 50 W precision coaxial cable to one side of a BNC T connector. See Figure 1Ć25.
H Connect the Sense output of the generator through a second
dualĆbanana connector followed by a 50 W precision coaxial cable to the other side of the BNC T connector. Now connect the BNC T connector to CH 1. See Figure 1Ć25.
DC Calibrator
Output Sense
HI
LO
50 W Coaxial Cables
BNC T
Connector
Figure 1Ć25:ăSubsequent Test Hookup
c. Measure amplitude of the probe compensation signal:
H Press SHIFT; then ACQUIRE MENU. Then use the keypad to
set AVERAGE to 16 in the side menu.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć59
Performance Tests
H Adjust the output of DC calibration generator until it precisely
overlaps the top (upper) level of the stored probe compensation signal. (This value will be near 500 mV.)
H Record the setting of the DC generator.
H Adjust the output of DC calibration generator until it precisely
overlaps the base (lower) level of the stored probe compensaĆ tion signal. (This value will be near zero volts.)
H Record the setting of the DC generator.
d. Press CLEAR MENU to remove the menus from the display. See
Figure 1Ć26.
1Ć60
Figure 1Ć26:ăMeasurement of Probe Compensator Amplitude
e. Check against limits:
H Subtract the value just obtained (base level) from that obtained
previously (top level).
H CHECK that the difference obtained is within 495ĂmV to 505ĂmV,
inclusive.
4. Disconnect the hookup: Disconnect the cable from CH 1.
Performance Verification Procedures
Performance Tests
Option 05 Video Trigger Checks
Equipment Required: PAL signal source (Item 18), 60 Hz. sine wave generĆ
ator (Item 19), pulse generator (Item 20), 75 W Cable (Item 21), 75 W termiĆ nator (Item 22), 50 W cable (Item 4), 50 W terminator (Item 3).
Check Video Trigger
Prerequisites: See page 1Ć15. These prerequisites include running the
signal path compensation routine.
Procedure:
1. Setup digitizing oscilloscope to factory defaults by completing the followĆ ing steps:
a. Press save/recall SETUP.
b. Press the mainĆmenu Recall Factory Setup.
c. Press the sideĆmenu OK Confirm Factory Init.
d. Wait for the Clock Icon to leave the screen.
e. CONFIRM the digitizing oscilloscope is setup as shown below.
Channel: CH1
Volt/div: 100 mV
Horizontal scale: 500 ms/div
2. Setup digitizing oscilloscope for TV triggers by completing the following steps:
a. Press TRIGGER MENU.
b. Press the mainĆmenu Type popĆup until you select Video.
c. Press the mainĆmenu Standard popĆup until you select 625/PAL.
d. Press the mainĆmenu Line.
e. Use the keypad to set the line number to 7 (press 7, then ENTER).
f. Press VERTICAL MENU.
g. Press the mainĆmenu Bandwidth.
h. Select 100 MHz from the side menu.
i. Press the mainĆmenu Fine Scale.
j. Use the keypad to set the fine scale to 282mV (press 282, SHIFT,
m, then ENTER).
k. Press HORIZONTAL MENU.
l. Press the mainĆmenu Horiz Scale.
m. Use the keypad to set the horizontal scale to 200 ns (press 200,
SHIFT, n, then ENTER).
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć61
Performance Tests
3. Check Jitter vs. Signal Amplitude
a. Setup equipment for Jitter Test (See Figure 1Ć27).
H Connect one of the rear panel composite outputs marked
COMPST on the TSG121 through a 75 W cable and a 75 W terminator to the oscilloscope CH1 input.
H Press the PAL signal source 100% FIELD control (the fourth
TSG121 frontĆpanel button from the left).
PAL Signal
Source
TSG121
COMPST
75 W Cable
75 W
Terminator
Figure 1Ć27:ăJitter Test Hookup
b. CHECK that the oscilloscope lights up its front panel TRIG'D LED
and it displays the waveform on screen (See Figure 1Ć28).
1Ć62
Figure 1Ć28:ăJitter Test Displayed Waveform
Performance Verification Procedures
Performance Tests
H CONFIRM that the TRIG'D LED is lit and the waveform is disĆ
played on screen.
c. Press SHIFT; then ACQUIRE MENU.
d. Press the mainĆmenu Mode.
e. Select the sideĆmenu Average. It should be already set to 16.
f. Press the mainĆmenu Create Limit Test Template.
g. Press the sideĆmenu V Limit.
h. Use the keypad to set V Limit to 100 mdiv (press 100, SHIFT, m,
then ENTER)
i. Press the sideĆmenu OK Store Template.
j. Press MORE.
k. Press the mainĆmenu Ref1.
l. Press CH1.
m. Press SHIFT; then ACQUIRE MENU.
n. Press the mainĆmenu Limit Test Setup.
o. Toggle the sideĆmenu Limit Test to ON.
p. Toggle the sideĆmenu Ring Bell if Condition Met to ON.
q. Press the mainĆmenu Mode.
r. Press the sideĆmenu Envelope.
s. Use the keypad to set envelope to use 100 acquisitions (press 100,
then ENTER).
t. Press the mainĆmenu Stop After button.
u. Press the sideĆmenu Single Acquisition Sequence.
v. CONFIRM that the oscilloscope successfully makes 100 acquisiĆ
tions. If not successful, the oscilloscope bell will ring. When the word Run in the top left corner of the display changes to STOP, the test is complete (See Figure 1Ć29).
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć63
Performance Tests
STOP shows the
test is complete
Figure 1Ć29:ăJitter Test When Completed
w. Press the mainĆmenu Limit Test Setup.
x. Toggle the sideĆmenu Ring Bell if Condition Met to OFF.
y. Toggle the sideĆmenu Limit Test to OFF.
4. Check Triggered Signal Range.
Setup oscilloscope for Triggered Signal Test.
a. Press MORE.
b. Press WAVEFORM OFF.
c. Press HORIZONTAL MENU.
d. Use the keypad to set the horizontal scale timeĆperĆdivision (Horiz
Scale (/div))to50ms (press 50, SHIFT, m, then ENTER).
e. Press SHIFT; then ACQUIRE MENU.
f. Press the mainĆmenu Stop After.
g. Press the sideĆmenu RUN/STOP button only.
h. Press the mainĆmenu Mode.
i. Press the sideĆmenu Sample.
j. Press RUN/STOP.
1Ć64
k. Press VERTICAL MENU.
l. Use the keypad to set fine scale to 300 mV (press 300, SHIFT, m,
then ENTER).
Performance Verification Procedures
Positive pulse
Negative pulses
Performance Tests
m. CONFIRM that the TRIG'D LED stays on and that the waveform on
screen is stable. ie; does not move horizontally or vertically. Also, CONFIRM that the waveform on the screen has one positive pulse and a number of negative pulses (See Figure 1Ć30).
Figure 1Ć30:ăTriggered Signal Range Test - 300 mV
n. Use the keypad to set the fine scale to 75 mV (press 75, SHIFT, m,
then ENTER).
o. CONFIRM that the TRIG'D LED stays lit and that the waveform on
screen is stable. ie; does not move horizontally or vertically. Also, CONFIRM that the waveform on the screen has one positive pulse and a number of negative pulses (See Figure 1Ć31).
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć65
Performance Tests
Negative pulses
Positive pulse
Figure 1Ć31:ăTriggered Signal Range Test - 75 mV
p. Disconnect all test equipment (TSG121) from the digitizing oscilloĆ
scope.
5. Check 60 Hz Rejection.
a. Setup oscilloscope for 60 Hz Rejection Test.
H Use the keypad to set the Ch1 Fine Scale to 282 mV (press 282,
SHIFT m, then ENTER).
H Press WAVEFORM OFF.
H Press CH2.
H Press VERTICAL MENU.
H Use the keypad set the fine scale to 2 V (press 2, then ENTER).
H Press HORIZONTAL MENU.
H Use the keypad to set the horizontal scale timeĆperĆdivision
(Horiz Scale (/div)) to 5 ms (press 5, SHIFT, m, then ENTER).
b. Setup 60 Hz signal generator (SG 502).
H Connect the output of the SG 502 to the CH2 input through a
50 W cable (See Figure 1Ć32).
1Ć66
Performance Verification Procedures
Performance Tests
Signal
Generator
SG 502
50 W Cable
Figure 1Ć32:ă60 Hz Rejection Test Hookup
H Adjust the SG 502 for three vertical divisions of 60 Hz signal
(See Figure 1Ć33). The signal will not be triggered. That is, it will run free.
Figure 1Ć33:ă60 Hz Rejection Test Setup Signal
c. Check 60 Hz rejection.
H Use the keypad to set the the horizontal scale timeĆperĆdivision
(Horiz Scale (/div)) to 50 ms (press 50, SHIFT, m, then ENTER).
H Reconnect the output of the signal generator (SG 502). Connect
the PAL signal source's composite signal connector (labelled COMPST on the TSG 121) to a 75 W cable and a 75 W terminaĆ tor. Connect both signals to the CH1 input through a BNC T (See Figure 1Ć34).
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć67
Performance Tests
H Press VERTICAL MENU.
H If needed, press the mainĆmenu Fine Scale.
H Use the keypad to set fine scale to 500 mV (press 500, SHIFT,
m, then ENTER).
H Connect another the PAL signal source's composite signal
connector (labelled COMPST on the TSG 121) through a 75 W cable and a 75 W terminator to the CH2 input (See Figure 1Ć34).
Signal
Generator
SG 502
PAL Signal
Source
TSG121
COMPST
COMPST
75 W Cable
75 W Cable
50 W Cable
75 W
Terminator
BNC T
Connector
75 W
Terminator
Figure 1Ć34:ăSubsequent 60 Hz Rejection Test Hookup
H CONFIRM that the TRIG'D LED stays on and that the waveform
on screen is stable. Stable means the waveform does not move horizontally or vertically. Also, confirm that the waveform on the screen has one positive pulse and a number of negative pulses (See Figure 1Ć35).
1Ć68
Performance Verification Procedures
Performance Tests
Figure 1Ć35:ă60 Hz Rejection Test Result
H Disconnect all test equipment from the digitizing oscilloscope.
6. Check Line Count Accuracy.
a. Setup oscilloscope for Line Count Accuracy Test.
H Press WAVEFORM OFF.
H Press CH1.
H Press HORIZONTAL MENU.
H Press the mainĆmenu Record Length.
H Press, if needed, the sideĆmenu -more- 1 of 2.
H Press the sideĆmenu 5000 points in 100divs.
H Press the mainĆmenu Horiz Scale (/div).
H Use the keypad to set the horizontal scale to 200 ns (press 200,
SHIFT, n, then ENTER).
b. Check Line Count Accuracy.
H Connect a composite output signal from the PAL signal source
(on the TSG 121 this refers to the signal at the rear labelled COMPST) to the CH1 input through a 75 W cable and a 75 W terminator (See Figure 1Ć36).
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć69
Performance Tests
PAL Signal
Source TSG121
COMPST
75 W Cable
75 W
Terminator
Figure 1Ć36:ăLine Count Accuracy Test Hookup
H Press the mainĆmenu Trigger Position.
H Press the sideĆmenu to Set to 50%.
H Press the mainĆmenu Horiz Pos.
H Press the sideĆmenu to Set to 50%.
H Use the HORIZONTAL POSITION knob to move the falling
edge of the sync pulse to two divisions to the left of center screen (See Figure 1Ć37).
1Ć70
Figure 1Ć37:ăLine Count Accuracy Test Setup Waveform
H Press CURSOR.
H Press the mainĆmenu Function.
Performance Verification Procedures
Performance Tests
H Press the sideĆmenu V Bars.
H Using the General Purpose knob, place the left cursor directly
over the trigger T" icon.
H Press SELECT.
H Turn the General Purpose knob to adjust the right cursor for a
cursor delta reading of 6.780us.
H Use the HORIZONTAL POSITION knob to position the right
cursor to center screen.
H Verify that the cursor is positioned on a positive slope of the
burst signal (See Figure 1Ć38).
Figure 1Ć38:ăLine Count Accuracy Correct Result Waveform
H Disconnect all test equipment (TSG 121) from the digitizing
oscilloscope.
H Turn off cursors by pressing CURSOR, then the mainĆmenu
Function button, and, finally, Off from the side menu.
7. Check the Sync Duty Cycle.
a. Setup digitizing oscilloscope for Sync Duty Cycle Test.
H Press TRIGGER MENU.
H Press the Standard popĆup to select FlexFmt. (Trigger Type
should already be set to Video)
H Press the mainĆmenu Setup.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć71
Performance Tests
H Press the sideĆmenu Field Rate.
H Use the keypad to set the field rate to 60.05 Hz (press 60.05,
then ENTER).
H Press the sideĆmenu Lines.
H Use the keypad to set the field rate to 793 lines (press 793, then
ENTER).
H Press the sideĆmenu Fields.
H Use the keypad to set the number of fields to 1 (press 1, then
ENTER).
H Press the sideĆmenu Sync Width.
H Use the keypad to set the width to 400 ns (press 400, SHIFT, n,
then ENTER).
H Press the sideĆmenu -more- 1 of 2. Then press V1 Start
Time.
H Use the keypad to set V1 start time to 10.10 ms (press 10.10,
SHIFT, m, then ENTER).
H Press the sideĆmenu V1 Stop Time.
H Use the keypad to set V1 stop time to 10.50 ms (press 10.50,
SHIFT, m, then ENTER).
H Press the mainĆmenu Type popĆup to select Edge.
H Press HORIZONTAL MENU.
H Press the mainĆmenu Record Length.
H Select the sideĆmenu 1000 points in 20div. (if needed, first
press the sideĆmenu -more- 2 of 2
).
H Turn the HORIZONTAL POSITION knob to position the trigger
`T' two divisions to the left of the center screen.
H Press MEASURE.
H If needed, press the mainĆmenu Select Measrmnt.
H Press the sideĆmenu Negative Width.
H Press the sideĆmenu Period.
b. Setup the pulse generator (PG502) for Sync Duty Cycle Test
H Set PULSE DURATION to 50 ns.
1Ć72
H Set PERIOD to 10 ms.
H Set OUTPUT (VOLTS) to -1 for LOW LEVEL and +1 for HIGH
LEVEL.
H Depress the COMPLEMENT button.
H Be sure BACK TERM is depressed (in).
Performance Verification Procedures
Performance Tests
c. Check Sync Duty Cycle
H Connect the pulse generator (PG502) through a 50 W cable and
a50W terminator to the oscilloscope CH1 input (See FigĆ ure 1Ć39).
Pulse
Generator
PG502
50 W Cable
50 W
Terminator
Figure 1Ć39:ăPG502 Setup for Sync Duty Cycle Test
H Turn the pulse generator OUTPUT (VOLTS) control until the
signal on the oscilloscope shows a one division negative going pulse (See Figure 1Ć40).
NOTE
You may need to adjust the trigger level control to obtain a stable trigger.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć73
Performance Tests
Figure 1Ć40:ăSync Duty Cycle Test: OneĆDiv Neg Pulse Waveform
H Turn the pulse generator PULSE DURATION variable control to
adjust the negative pulse so the oscilloscope CH1 - Width measurement displays 400ns +/-10 ns.
H Turn the HORIZONTAL SCALE knob to set the oscilloscope
time base to 5 ms/div.
H Turn the pulse generator PERIOD variable control to adjust the
period until the oscilloscope CH1 Period measurement reads
21.000ms -25/+50 ns (See Figure 1Ć41). Read note shown below.
NOTE
The pulse duration and period adjustments are critical in making this measurement. If the pulse duration and/or the duty cycle are not stable, the FLEXFMT function may not function. You must take care when making these adjustments.
1Ć74
Performance Verification Procedures
Performance Tests
Figure 1Ć41:ăSync Duty Cycle Test: Critically Adjusted Pulse
H Press TRIGGER MENU.
H Press the mainĆmenu Type popĆup until you select Video.
If the TRIG'D LED is not on, check that the CH1 - Width and CH1 Period measurements are adjusted correctly (see note
above). CONFIRM that the setup is correct and the oscilloscope will trigger.
H CONFIRM that the TRIG'D LED is on and the waveform is
stable.
H Disconnect the signal source from CH1, wait a few seconds,
then reconnect the signal.
H CONFIRM that the TRIG'D LED is on and the waveform is
stable.
H Press Sync Polarity.
H Press Pos Sync.
H Push the pulse generator COMPLEMENT button out.
H CONFIRM that the TRIG'D LED is on and the waveform is
stable.
H Disconnect the signal source from CH1, wait a few seconds,
then reconnect the signal.
H CONFIRM that the TRIG'D LED is on and the waveform is
stable.
TDS 520A, 524A, 540A, & 544A Performance Verification
1Ć75
Performance Tests
H Disconnect all test equipment (TSG 121) from the oscilloscope.
H Press save/recall SETUP, press the mainĆmenu button Recall
Factory Setup, and the sideĆmenu OK Confirm Factory Init.
1Ć76
Performance Verification Procedures
Specifications
Specifications
This subsection begins with a general description of the traits of the TDS 520A, 524A, 540A and 544A Digitizing Oscilloscopes. Three subsecĆ tions follow, one for each of three classes of traits: nominal traits, warĆ ranted characteristics, and typical characteristics.
General Product Description
The Tektronix TDS 520A, 524A, 540A and 544A Digitizing Oscilloscopes are portable, fourĆchannel instruments suitable for use in a variety of test and measurement applications and systems. Key features include:
H 500 MHz maximum analog bandwidth.
H 1 Gigasample/second maximum digitizing rate (TDS 540A and 544A);
500 Megasamples/second maximum digitizing rate (TDS 520A and 524A).
H FourĆchannel acquisition Ċ the TDS 540A and 544A offer four fullĆfeaĆ
tured channels; the TDS 520A and 524A offer two fullĆfeatured channels and two channels with limited vertical scale selections: 100ĂmV, 1ĂV, and 10ĂV.
H Extensive triggering capabilities: such as edge, logic, and glitch. Video
trigger (Option 05) is also available. The video trigger modes are NTSC, PAL, SECAM, HDTV, and FlexFormat
H Waveform Math Ċ Invert a single waveform and add, subtract, and
multiply two waveforms. On the TDS 524A, 544A, and other TDS 500A equipped with option 2F, integrate or differentiate a single waveform or perform an FFT (fast fourier transform) on a waveform to display its magnitude or phase versus its frequency.
H EightĆbit digitizers.
TM
(user definable format).
H Up to 15,000Ćpoint record length per channel (50,000Ćpoint with option
1M).
H Full GPIB software programmability. Hardcopy output using GPIB,
RSĆ232, or Centronics ports (RSĆ232 and Centronics is standard on the TDS 524A and 544A and optional, as option 13, on the TDS 520A and 540A).
H Complete measurement and documentation capability.
H Intuitive graphic icon operation blended with the familiarity of traditional
horizontal and vertical knobs.
H OnĆline help at the touch of a button.
H A full complement of advanced functions, like continuouslyĆupdated
measurements, results and local pass/fail decision making.
TDS 520A, 524A, 540A, & 544A Performance Verification
2Ć1
Specifications
H Specialized display modes, such as variable persistence (with color
coding on the TDS 524A and 544A), dot or vector mode, sin(x)/x or linear display filters, and, on the TDS 524A and 544A, user selectable color palettes. The Fit to Screen" feature compresses the entire waveform record to fit on the screen.
User Interface
H A complement of advanced acquisition modes such as peakĆdetect,
highĆresolution, sample, envelope, and average. The FastFrame feature acquires waveforms in rapid succession with a minimum of dead time between acquisitions. FastFrame 50,000 frames per second.
H A unique graphical user interface (GUI), an onĆboard help mode, and a
logical frontĆpanel layout which combine to deliver a new standard in usability.
H VGA output for driving remote monitors.
H A 1.44 Mbyte, DOS 3.3 or later, floppy disk drive for saving waveforms,
color images, and oscilloscope setups (standard on the TDS 524A and 544A and optional, as option 1F, on the TDS 520A and 540A). Also, all TDS 500A have builtĆin NVRAM storage for saving waveforms and setups.
H On the 524A and 544A, a color display for distinguishing among waveĆ
forms, their measurements, and associated text.
Use a combination of frontĆpanel buttons, knobs, and onĆscreen menus to control the many functions of these oscilloscopes. The frontĆpanel controls are grouped according to function: vertical, horizontal, trigger, and special. Set a function you adjust often, such as vertical positioning or the time base setting, directly by its own frontĆpanel knob. Set functions which you change less often, such as vertical coupling and horizontal mode, indirectly using selected menus.
TM
allows acquisition rates of up to
TM
2Ć2
Menus
Pressing one (sometimes two) frontĆpanel button(s), such as vertical menu, displays a main menu of related functions, such as coupling, bandwidth, etc., at the bottom of the screen. Pressing a mainĆmenu button, such as coupling, displays a side menu of settings for that function, such as AC, DC, or GND (ground) coupling, at the right side of the screen. Pressing a sideĆ menu button selects a setting such as DC.
Indicators
OnĆscreen readouts help you keep track of the settings for various functions, such as vertical and horizontal scale and trigger level. Some readouts use the cursors or the automatic parameter extraction feature (called measure) to display the results of measurements made or the status of the instrument.
Specifications
Specifications
General Purpose Knob
Assign the general purpose knob to adjust a selected parameter function. More quickly change parameters by toggling the SHIFT button. Use the same method as for selecting a function, except the final sideĆmenu selecĆ tion assigns the general purpose knob to adjust some function, such as the position of measurement cursors on screen, or the setting for a channels fine gain.
GUI
The user interface also makes use of a GUI, or Graphical User Interface, to make setting functions and interpreting the display more intuitive. Some menus and status are displayed using iconic representations of function settings such as those shown here for full, 100 MHz, and 20 MHz bandwidth. Such icons allow you to more readily determine status or the available settings.
Signal Acquisition System
TDS 540A and 544A: The signal acquisition system provides four vertical channels with calibrated vertical scale factors from 1 mV to 10 V per division. All four channels can be acquired simultaneously.
Each of the four TDS 540A and 544A channels can be displayed, vertically positioned, and offset, can have their bandwidth limited (100 MHz or 20 MHz) and their vertical coupling specified. Fine gain can also be adĆ justed.
TDS 520A and 524A: The signal acquisition system provides four vertical channels. Two are fullĆfeatured vertical channels (CH1 and CH2) with caliĆ brated vertical scale factors from 1 mV to 10 V per division. The other two are auxiliary channels (AUX1 and AUX2) with three calibrated deflection factors of 100 mV, 1 V, and 10 V per division. Any two of the four channels can be acquired simultaneously.
Each of the four TDS 520A and 524A channels can be displayed, vertically positioned, and offset. CH1 and CH2 can also have their bandwidth limited (100 MHz or 20 MHz) and their vertical coupling specified. Fine gain can also be adjusted for CH1 and CH2.
On all TDS 520A, 524A, 540A and 544A: Besides the four channels, up to three math waveforms and four reference waveforms are available for disĆ play. (A math waveform results when dual waveform operations, such as add, are specified on any two channels. A reference waveform results when you save a live waveform in a reference memory.)
TDS 520A, 524A, 540A, & 544A Performance Verification
2Ć3
Specifications
ecordLengt
Record Length
Horizontal System
There are three horizontal display modes: main only, main intensified, and delayed only. You can select among various horizontal record length setĆ tings.
A feature called Fit to Screen" allows the user to view entire waveform records within the 10 division screen area. In other words, waveforms are compressed to fit on the screen. The exception is that, with Option 1M, 50,000 point records are displayed over a 15 division time span. (see TableĂ2Ć1)
TableĂ2Ć1:ăRecord Length vs. Divisions per Record,
Samples Per Division, and Sec/Div Sequence
Divisions per Records
R
50000
15000
h
1
2
Samples/Division (Sec/Div Sequence)
Fit to Screen OFF 50 (1-2-5)
Fit to Screen ON 50 (1-2-5)
1000 divisions 10 divisions
300 divisions 15 divisions
5000 100 divisions 10 divisions
2500 50 divisions 10 divisions
1000 20 divisions 10 divisions
500 10 divisions 10 divisions
1 The 50,000 samples in 1,000 division record is only available with Option 1M.
2 The maximum record length of 15,000 samples (50,000 samples with Option 1M) is selectĆ
able with all acquisition modes except Hi Res. In Hi Res, the maximum record length is 5,000 samples (15,000 samples with Option 1M).
FastFrameTMallows multiple triggered acquisitions to occur with a minimum of dead time" between acquisitions. A maximum of 910 acquisition seĆ quences (frames) are possible with 50 points per record (frame length) when a 50,000 point waveform record length is available (with option 1M).
You can delay by time with respect to the main trigger both the delayed display and the intensified zone on the main intensified display. You can set them to display immediately after the delay (delayed runs after main mode). The delayed display can also be set to display at the first valid trigger after the delay (delayed triggerable mode).
You can also delay by a selected number of events the delayed display (or the intensified zone). In this case, the events source is the delayed trigger source. For any events signal, the delayedĆtrigger system conditions the signal by determining the source, coupling, etc., of that signal.
2Ć4
Specifications
Specifications
Trigger System
The triggering system supports a varied set of features for triggering the signalĆacquisition system. Trigger signals recognized include:
H Edge (mainĆ and delayedĆtrigger systems): This familiar type of triggerĆ
ing is fully configurable for source, slope, coupling, mode (auto or normal), and holdoff.
H Logic (mainĆtrigger system): This type of triggering can be based on
pattern (asynchronous) or state (synchronous). In either case, logic triggering is configurable for sources, for boolean operators to apply to those sources, for logic pattern or state on which to trigger, for mode (auto or normal), and for holdoff. TimeĆqualified logic triggering can also be specified.
H Pulse (mainĆtrigger system): Pulse triggering is configurable for triggerĆ
ing on runt or glitch pulses, or on pulse widths or periods inside or outside limits that you specify. It is also configurable for source, polarity, mode, and holdoff.
H Video (with option 05: Video Trigger): Video triggering is compatible with
standard NTSC, PAL, SECAM, and HDTV formats. An additional feature called FlexFormat format on which to trigger.
You can choose where the trigger point is located within the acquired waveĆ form record by selecting the amount of pretrigger data displayed. Select presets of 10%, 50%, and 90% of pretrigger data in the horizontal menu, or assign the general purpose knob to set pretrigger data to any value within the limits of trigger position resolution.
TM
(flexible format) allows the user to define the video
TDS 520A, 524A, 540A, & 544A Performance Verification
2Ć5
Specifications
Acquisition Control
OnĆBoard User Assistance
You can specify a mode and manner to acquire and process signals.
H Select equivalentĆtime sampling on repetitive signals or interpolation of
points sampled on nonĆrepetitive signals. Both can increase apparent sample rate on the waveform when maximum realĆtime rates are reached. The apparent sample rate can be increased, even further, by using the FastFrame cond are possible using FastFrame
H Use peakĆdetect, highĆresolution, sample, envelope, and average
modes can be used to acquire signals.
H Set the acquisition to stop after a single acquisition (or sequence of
acquisitions if acquiring in average or envelope modes).
H Select channel sources for compliance with limit tests. You can direct
the TDS to signal you or generate hard copy output based on the reĆ sults. Also, you can create templates for use in limit tests.
Help and autoset can assist you in setting up the Digitizing Oscilloscope to make your measurements.
TM
feature. Acquisition rates of 50,000 Frames/SeĆ
TM
.
Measurement Assistance
Help
Help displays operational information about any frontĆpanel control. When help mode is in effect, manipulating any frontĆpanel control causes the oscilloscope to display information about that control. When help is first invoked, an introduction to help is displayed on screen.
Autoset
Autoset automatically sets up the oscilloscope for a viewable display based on the input signal.
Once you have set up to make your measurements, the cursor and measure features can help you quickly make those measurements.
Cursor
Three types of cursors are provided for making parametric measurements on the displayed waveforms. Horizontal bar cursors (H Bar) measure vertical parameters (typically volts). Vertical bar cursors (V Bar) measure horizontal parameters (typically time or frequency) and now extend to the top and bottom of the screen. Paired cursors measure both amplitude and time simultaneously. These are delta measurements; that is, measurements based on the difference between two cursors.
2Ć6
Specifications
Specifications
Both H Bar and V Bar cursors can also be used to make absolute measureĆ ments; that is measurements relative to a defined level or event. For the H Bars, either cursor can be selected to read out its voltage with respect to any channels ground reference level. For the V Bars, it's time with respect to the trigger point (event) of the acquisition, and the cursors can control the portion of the waveform on which automatic measurements are made.
For time measurements, units can be either seconds or Hertz (for 1/time).
When the video trigger option installed (Option 05), the video line number can be selected using the vertical cursors. IRE amplitude (NTSC) can be measured using the horizontal cursors with or without the video trigger option installed.
Measure
Measure can automatically extract parameters from the signal input to the Digitizing Oscilloscope. Any four out of the more than 20 parameters availĆ able can be displayed to the screen. The waveform parameters are meaĆ sured continuously with the results updated onĆscreen as the Digitizing Oscilloscope continues to acquire waveforms.
Storage and I/O
Digital Signal Processing (DSP)
An important component of the multiprocessor architecture of this Digitizing Oscilloscope is Tektronix's proprietary digital signal processor, the DSP. This dedicated processor supports advanced analysis of your waveforms when doing such computeĆintensive tasks as interpolation, waveform math, and signal averaging. It also teams with a custom display system to deliver specialized display modes (See Display, later in this description.)
Acquired waveforms may be saved in any of four nonvolatile REF (reference) memories or, if available, on a 3.5 inch, DOS 3.3ĆorĆlater compatible disk. The disk is standard on the TDS 524A and 544A. It is available as option 1F on the 520A and 540A. Any or all of the saved waveforms may be displayed for comparison with the waveforms being currently acquired.
The source and destination of waveforms to be saved may be chosen. Assignment can be made to save any of the four channels to any REF memory or to move a stored reference from one REF memory to another. Reference waveforms may also be written into a REF memory location via the GPIB interface.
The Digitizing Oscilloscope is fully controllable and capable of sending and receiving waveforms over the GPIB interface (IEEE Std 488.1-1987/IEEE Std 488.2-1987 standard). This feature makes the instrument ideal for making automated measurements in a production or research and developĆ ment environment that calls for repetitive data taking. SelfĆcompensation
TDS 520A, 524A, 540A, & 544A Performance Verification
2Ć7
Specifications
and selfĆdiagnostic features built into the Digitizing Oscilloscope to aid in fault detection and servicing are also accessible using commands sent from a GPIB controller.
Another standard feature is hardcopy. This feature allows you to output waveforms and other onĆscreen information to a variety of graphic printers and plotters from the Digitizing Oscilloscope front panel, providing hard copies without requiring you to put the Digitizing Oscilloscope into a systemĆ controller environment. You can make hardcopies in a variety of popular output formats, such as PCX, TIFF, BMP, RLE, EPS, Interleaf, and EPS mono or color. You can also save hardcopies in a disk file in any of the formats above. The hardcopies obtained are based on what is displayed onĆscreen at the time hardcopy is invoked. The hardcopies can be stamped with date and time and spooled to a queue for printing at a later time. You can output screen information via GPIB, RSĆ232C, or Centronics interfaces.
Display
The TDS 520A, 524A, 540A and 544A Digitizing Oscilloscopes offer flexible display options. You can customize the following attributes of your display:
H Color: Waveforms, readouts, graticule, etc. on the TDS 524A and 544A;
H Intensity: waveforms, readouts, and graticule;
H Style of waveform display(s): vectors or dots, intensified or nonĆintensiĆ
fied samples, infinite persistence, and variable persistence with color coding;
H Interpolation method: Sin(x)/x or Linear;
H Display format: xy or yt with various graticule selections including NTSC
and PAL to be used with video trigger.
Zoom
This Digitizing Oscilloscope also provides an easy way to focus in on those waveform features you wish to examine up close. By invoking zoom, you can magnify the waveform parameter using the vertical and horizontal controls to expand (or contract) and position it for viewing.
2Ć8
Specifications
Nominal Traits
This subsection contains tables that list the electrical and mechanical nominal traits that describe the TDS 520A, 524A, 540A and 544A Digitizing Oscilloscopes.
Nominal traits are described using simple statements of fact such as Four, all identical" for the trait Input Channels, Number of", rather than in terms of limits that are performance requirements.
TableĂ2Ć2:ăNominal Traits Ċ Signal Acquisition System
Name Description
Bandwidth Selections 20 MHz, 100 MHz, and FULL (500 MHz)
Digitizers, Number of TDS 540A and 544A: Four, all identical
TDS 520A and 524A: Two, both identical
Digitized Bits, Number of 8 bits
1
Input Channels, Number of TDS 540A and 544A: Four, all identical, called CH1 - CH4
TDS 520A and 524A: Two fullĆfeatured (CH1 and CH2), plus two limited, auxiliary inputs (AUX1 and AUX2)
Input Coupling
Input Impedance Selections
Ranges, Offset, TDS 540A, 544A, and CH1 and CH2 on TDS 520A and 524A
Ranges, Offset, AUX1 and AUX2 on TDS 520A and 524A
2
DC, AC, or GND
1MW or 50 W
Volts/Div Setting
1 mV/div - 99.5 mV/div
100 mV/div - 995 mV/div
1 V/div - 10 V/div
Volts/Div Setting
100 mV/div
1 V/div
10 V/div
Offset Range
±1 V
±10 V
±100 V
Offset Range
±0.5 V
±5.0 V
±50 V
Range, Position ±5 divisions
Range, Sensitivity TDS 540A and 544A: 1 mV/div to 10 V/div
3
TDS 520A and 524A: CH1 and CH2: 1 mV/div to 10 V/div TDS 520A and 524A: AUX1 and AUX2: 100 mV/div, 1 V/div, 10 V/div
3
1 Displayed vertically with 25 digitization levels (DLs) per division and 10.24 divisions dynamic range with zoom off. A DL is the
smallest voltage level change that can be resolved by the 8Ćbit AĆD Converter, with the input scaled to the volts/division setting of the channel used. Expressed as a voltage, a DL is equal to 1/25 of a division times the volts/division setting.
2 The input characteristics (Input Coupling, Input Impedance Selections, etc.) apply to both fullĆfeatured and auxiliary inputs except
where otherwise specified.
3 The sensitivity ranges from 1 mV/div to 10 V/div in a 1-2-5 sequence of coarse settings. Between a pair of adjacent coarse setĆ
tings, the sensitivity can be finely adjusted. The resolution of such a fine adjustment is 1% of the more sensitive of the pair. For example, between 50 mV/div and 100ĂmV/div, the volts/division can be set with 0.5 mV resolution.
TDS 520A, 524A, 540A, & 544A Performance Verification
2Ć9
Nominal Traits
TableĂ2Ć3:ăNominal Traits Ċ Time Base System
Name
Range, SampleĆRate
1,3
TDS 540A, 544A
Range, SampleĆRate
1,3
TDS 520A, 524A
Range, Equivalent Time or Interpolated Waveform Rate
2,3
Description
Number of Channels On
1
2
3or4
Number of Channels On
1
2
SampleĆRate Range
5 Samples/s - 1 GSamples/s
5 Samples/s - 500 MSamples/s
5 Samples/s - 250 MSamples/s
SampleĆRate Range
5 Samples/s - 500 MSamples/s
5 Samples/s - 250 MSamples/s
500 MSamples/s to 100 GSamples/s
Range, Seconds/Division 500 ps/div to 10 s/div
Record Length
4
500 samples, 1000 samples, 2500 samples, 5000 samples, 15000 samples. A record length of 50000 samples is available with Option 1M. Up to four 50 K waveform records may be saved in NVRAM with Option 1M installed.
FastFrame
TM
Maximum Frame Rate: 50,000 Frames/Second
Frame Length Range: 50 points/Frame to 5,000 Points/Frame
Maximum Number of Frames: ą910 Frames at 50 Points/Frame ą(with Option 1M, 50,000 Record Length)
ą227 Frames at 50 Points/Frame ą(standard configuration of 15,000 Record Length)
1 The range of realĆtime rates, expressed in samples/second, at which a digitizer samples signals at its inputs and stores the samĆ
ples in memory to produce a record of timeĆsequential samples
2 The range of waveform rates for equivalent time or interpolated waveform records.
3 The Waveform Rate (WR) is the equivalent sample rate of a waveform record. For a waveform record acquired by realĆtime samĆ
pling of a single acquisition, the waveform rate is the same as the realĆtime sample rate; for a waveform created by interpolation of realĆtime samples from a single acquisition or by equivalentĆtime sampling of multiple acquisitions, the waveform rate is faster than the real time sample rate. For all three cases, the waveform rate is 1/(Waveform Interval) for the waveform record, where the waveĆ form interval (WI) is the time between the samples in the waveform record.
4 The maximum record length of 15,000 samples (50,000 samples with Option 1M) is selectable with all acquisition modes except
Hi Res. In Hi Res, the maximum record length is 5,000 samples (15,000 samples with Option 1M).
2Ć10
Specifications
Loading...