Attach a probe to CH 1
and hook the probe to your
signal.
Press
CH 1.
Press
AUTOSET.
Adjust VERTICAL and HORIZONTAL
POSITION and SCALE.
To Preview a Waveform:
1
Press ZOOM.
2
Press Mode in the main menu. Then press Preview in the side menu to turn on Dual Window Zoom.
Upper graticule zooms the boxed area
on the selected waveform.
Lower graticule displays the selected waveform
unzoomed with the zoomed area in box.
Device under test
3
Use the Selected Graticule menu to select the upper or lower waveform. Use the vertical and horizontal knobs
to adjust the waveform in the graticule you select.
To Capture Infrequent Events (TDS 500C & 700C Models):
Press InstaVu to toggle between InstaVu and Normal waveform
capture rates.
When in InstaVu mode:
H Waveforms displayed are updated thousands of times
faster than normal.
H Very brief changes in waveforms are captured.
H Certain features, such as Limit Testing, Math Waveforms, Zoom, and
record lengths longer than 500 points, are not available.
Copyright E Tektronix, Inc., 1997. Printed in U.S.A.
Page 3
To Save a Hardcopy to the File System:
To Set Up Using a Menu:
12
Press SHIFT, and then press HARDCOPY.
34
Press Port in the main menu, press File in the side
menu, and then press CLEAR MENU.
Disk drive
Press Format in the main menu, and select a hardcopy
format from the side menu.
Press HARDCOPY anytime to save a copy of the current
screen to a unique file in the oscilloscope file system.
1
Press any of the front panel menu buttons.
2
Select an item from the main (bottom) menu.
3
Select an item from the side menu, if displayed.
Use SHIFT for
the alternate
(blue) menus.
The readout indicates the
value that you can adjust
with the general purpose
knob or keypad.
To Perform Other File System Operations:
H Press SA VE/RECALL WAVEFORM, and use the menu buttons to save a waveform to a file or recall it from a file.
H Press SA VE/RECALL SETUP, and use the menu buttons to save a setup to a file or recall it from a file.
H Press File Utilities in the Save/Recall Waveform, Save/Recall Setup, or Hardcopy menus to access utilities that create
directories, copy files, and do other operations in the oscilloscope file system.
4
Adjust menu item values using the general purpose
knob or by entering numbers on the keypad.
Adjustable menu item value
General purpose knob
Keypad
End your entry by
pressing ENTER ( ).
Page 4
To Select a Trigger:
To Take Measurements Automatically:
1
Press TRIGGER MENU.
2
Select a trigger type or parameter from the
main menu.
This symbol
indicates a
pop-up menu
Set TRIGGER MAIN LEVEL.
3
Trigger Selections
TYPE
<Edge>
Select any one of
Ch 1 thru Ch 4,
Line, or DC Aux
Positive
Negative
Define levels High,
Low, or Don’t Care
for Ch 1 thru Ch 4
(On some models, Ax 1 & Ax 2 replace Ch 3 & Ch 4)
TYPE
<Logic>
CLASS
<Pattern>
AND
OR
NAND
NOR
CLASS
<State>
Define levels High,
Low, or Don’t Care
for Ch 1 thru Ch 3
Select edge for the
clock (always Ch 4)
AND
OR
NAND
NOR
CLASS
<Setup/Hold>
Select one of
Ch 1 thru Ch 4 as
the data source
Do not select the
same channel used
as the clock source
Select one of
Ch 1 thru Ch 4 as
the clock source
Select the clock
Clock Source
edge
Do not select the
same channel used
as the data source
12
Press MEASURE.
34
Select up to four measurements.
Press Select Measrmnt or Snapshot in the main menu.
Press CLEAR MENU to move the measurement
readouts away from the graticule.
“T” shows the
trigger position
Press to display the pop-up menus
Press again to make a selection
A pop-up selection changes the other main
menu items
Arrow shows
the trigger level
Title of the
side menu
Removes the
menus from
the screen
Level
LevelSlopeSource
Set level or select
preset level based
on TTL or ECL logic
DC
Coupling
AC
HF
Reject
LF
Reject
Noise Rej (DC
Low Sensitivity)
Set a threshold
level for each of
Ch 1 thru Ch 4
Set ThresholdsDefine LogicDefine InputsTrigger When
Goes TRUE
Goes FALSE
TRUE for
less than
TRUE for
more than
1
Qualification by time
Set a threshold
level for each of
the pattern
channels, Ch 1
thru Ch 3, and
the clock, Ch 4.