Y ou can use the dedicated, front-panel knobs and buttons to do the most common operations.
Turn INTENSITY to adjust waveform intensity.
Push
FastAcq to turn Fast Acquisition
on and off.
Push CURSORS to turn cursors on and off.
Push PRINT to make a hard copy .
Push
DEFAULT SETUP to return settings to the
factory default values.
Push
AUTOSET to automatically set the vertical,
horizontal, and trigger controls to produce a
usable display.
Use the multipurpose knobs to control
parameters selected from the screen interface.
FINE button to make small changes with
Push a
a multipurpose knob.
Push OFF to turn the optional touch
screen on and off.
Push
ZOOM to add a magnified graticule to the
display . Push the
HORIZ or VERT button to
select the axis that you want to magnify.
Use these knobs and buttons to set horizontal
SCALE and POSITION for the waveforms. Push
DELAY to turn on horizontal delay, and then use
POSITION to set the delay time. Adjust
RESOLUTION to change the number of acquired
points in the waveform.
Use these knobs and buttons to set the basic
trigger parameters. Push
ADVANCED to display
a menu of additional trigger functions.
Use these buttons to start and stop
acquisition or start a single acquisition
sequence. The ARM, READY, and TRIG’D
lights show the acquisition status.
Push
CH<X> to turn channel displays on and off.
Adjust the channel
POSITION and SCALE or
change the input termination with dedicated
knobs and buttons.
Copyright E Tektronix, Inc. Beaverton, OR 97077
To Use the Screen Interface
Y ou can control all oscilloscope functions, including powering of f the instrument, through the screen interface.
Choose the Toolbar
Click here to change to menu-bar operation.
Click a button in the toolbar to display a
control window at the bottom of the display.
Click a screen control to change a setting.
Click a numerical control to assign that control
to a multipurpose knob. Turn the multipurpose
knob to adjust the parameter value.
Or Choose the Menu Bar
Use some menu items to display a control
window at the bottom or side of the display.
Use some menu items to directly
change settings.
Click here to change to toolbar operation.
Click here to close a control window.
More Operating Tips:
H Use the optional touch screen to control the oscilloscope when bench space is
unavailable, such as on a cart or in an equipment rack.
H Plug in a mouse and keyboard if you have the bench space to use them. Y ou can
plug in a USB mouse or keyboard anytime, even while the oscilloscope is
running.
H Use the menu bar to access PC-related functions such as, Page Setup, Export,
and Copy.
To Display a Waveform
Attach a probe to CH 1 and connect the probe
to your signal.
Push CH 1 if channel 1 is not
already displayed.
Push AUTOSET.
1
2
3
Adjust VERTICAL and HORIZONTAL
POSITION
Adjust RESOLUTION to change the record
length and sample rate. Y ou can acquire
more samples in the waveform to see more
detail, or acquire fewer samples with a
Adjust INTENSITY to change the brightness,
vector fill, and display persistence of acquired
and SCALE if necessary to
optimize the display.
faster update rate.
points.
Recommended Probe
Accessories
4
5
6
P5050 Passive Probes for
general-purpose applications
P6243 and P6245 Active Probes or P6158
Low-C Probe for high-speed applications
P6246, P6247, P6248 Differential Probes
for differential signals and low-noise
applications
TCP202 Current Probe for
general-purpose applications
To See More Waveform Detail
Use Zoom
Push the ZOOM button to split the
screen and add a zoom graticule.
Push the HORIZ button or the VERT button to
select the axis you want to magnify in the zoom
graticule. Use the multipurpose knobs to adjust
scale and position of the magnified waveform.
Y ou can also set up a zoom graticule
from the screen interface. Click and drag
across the segment of the waveform that
you want to see in greater detail.
Then select Zoom from the drop-down list to
magnify the highlighted waveform segment.
Use Horizontal Delay
Use the Zoom function to magnify an acquisition vertically, horizontally, or in both
waveform dimensions.
affect only the display , not the actual waveform that is acquired.
SCALE or POSITION changes you make to the Zoom graticule
1
2
Main graticuleZoom graticule
3
4
Use horizontal
the trigger location by a significant interval of time.
DELAY to acquire waveform detail in a region that is separated from
Push the front-panel DELAY button.
Adjust the delay time with the horizontal
POSITION control, or enter the delay time
in the control window.
Adjust the horizontal SCALE to acquire
the detail you need around the delay
expansion point.
Trigger point
Delay time
1
Acquired waveform
Expansion point
2
3
More Operating Tips:
H Y ou can use Zoom and Horizontal Delay together to magnify a delayed
acquisition.
H Toggle Horizontal Delay on and off to quickly compare signal details at two
different areas of interest, one near the trigger location and the other centered at
the delay time.
To Use Fast Acquisition
Turn Fast Acquisition on to acquire up to 100,000 waveforms per second.
Push the front-panel FastAcq button.
Or select
Adjust INTENSITY to optimize the intensity or
color grading for the signal you want to analyze.
Increasing intensity can make less-frequently
acquired points brighter in the display.
Fast Acquisitions in the
Horiz/Acq menu.
How DPOs Work
Unlike DSOs that have long dead times between
acquisitions, Digital Phosphor Oscilloscopes
(DPOs) continuously acquire waveforms at rates
comparable to analog oscilloscopes. DPOs
continuously overlay the acquired information
into a three-dimensional database that is
updated on the display 30 times per second. For
each pixel in the display, the intensity (or color)
of the pixel is proportional to the number of
actual samples that the pixel represents.
1
2
Analog Real Time sweeps
Digital Storage acquisitions
Digital Phosphor acquisitions
More Operating Tips:
H Choose one of the color grading palettes in the Display Colors control window to
see different sample densities represented in different colors.
H Turn
AutoBright on in the Display Appearance control window. When you use
AutoBright, the displayed waveforms remain visible even at low trigger repetition
rates.
To Choose an Acquisition Mode
Select Acquisition Mode... in the
Horiz/Acq menu.
Or click the Horiz button, and then select
an acquisition mode in the
horizontal/acquisition control window.
For Average or Envelope acquisition modes,
click the
number of waveforms with the multipurpose
knob. Y ou can also double-click the control and
# of Wfms control and then set the
use the pop-up keypad.
How the Acquisition
Modes Work
Sample mode retains one sampled point from
each acquisition interval.
1
2
acquisition interval +
Interval 1234Interval 1234
record duration
number of points in record
Sample
Displayed record points (at
maximum horizontal magnification)
Peak Detect mode uses the highest and
lowest of all the samples contained in two
consecutive acquisition intervals.
Hi Res mode calculates the average of all
the samples for each acquisition interval.
Envelope mode finds highest and lowest record
points over many acquisitions. Envelope uses
Peak Detect for each individual acquisition.
Average mode calculates the average value for
each record point over many acquisitions.
Average uses Sample mode for each individual
acquisition.
Peak Detect
Hi Res
Three acquisitions from one source
Acquisition 123
Envelope
Average
To Select a Trigger
Select the
source, coupling, slope and mode with these
front-panel controls. Push
EDGE trigger type and then set the
ADVANCED to select
one of the other trigger types.
You can also select a trigger type in the
Trigger menu.
Or click the
trigger type in the trigger control window
Trig button and then select a
that is displayed.
Trigger Selections
Trigger typeLevelsTrigger conditions
Edge
Video
Glitch
Width
Runt
WindowTwo levels to define
Timeout
Single level
Single levelNoneTrigger on NTSC, PAL, SECAM, Analog HDTV, and
Single level
Single level
Two levels to define the
logic transition region
signal transition region
Single level
Timers
None
One to specify glitch
width
Two to specify minimum
and maximum pulse
widths
One to specify an
optional minimum
runt-pulse duration
One to specify an
optional minimum
violation width
One to specify time-out
time
Trigger on rising or falling edge, as defined by slope
control. Coupling choices are DC, AC, AC LF Reject, AC
HF Reject, and Noise Reject.
Custom video formats.
Trigger on glitches narrower than the specified width or
ignore glitches narrower than the specified width.
Trigger on pulses that have widths between the range of
the two timers or outside the range of the two timers.
Trigger on a pulse that enters the transition region from
one side but does not leave the region from the other side.
Pulses can be qualified with the logical states of other
channels (4-channel models only).
Trigger when a signal violates either upper or lower
boundaries set outside the normal range of the signal.
Events can be qualified with the logical states of other
channels (4-channel models only).
Trigger when a signal does not make a transition for a
specified length of time.
Transition
Setup/Hold
Pattern
State
Two levels to define the
logic transition region
Independent levels for
Data and Clock
Independent levels for
each channel
Independent levels for
each channel
One to specify transition
time
One to specify setup
time and one to specify
hold time
One to specify pattern
duration
None
Trigger when a logic signal spends more time or less time
in the transition region than a specified amount of time.
Trigger on violations of setup or hold time between a Data
signal and a Clock signal. The specified setup and hold
times can be positive or negative values.
Trigger when a Boolean combination of up to four
channels becomes true. Trigger immediately or only after
the combination is true for a specified time duration.
Trigger on transition of one channel when a Boolean
combination of up to three other channels is true.
To Use A (Main) and B (Delayed) Triggers
Y ou can use the A Event (Main) trigger alone or combine it with the B Event (Delayed) trigger to capture more complex signals.
Set the A trigger type and source in the
A Event (Main) tab of the trigger control
window.
Choose a function in the A→B Sequence tab of
the trigger control window.
Set the trigger delay time or the number
of B events, as appropriate.
Set the B trigger characteristics in the B Event
(Delayed) tab of the trigger control window.
Trigger on B Event
1
2
3
4
Posttrigger
record
A trigger point
Pretrigger
record
Acquired waveform record
The A trigger arms the oscilloscope.
Posttrigger acquisition starts on nth B event.
B Trigger After Delay Time
The A trigger arms the oscilloscope.
Posttrigger acquisition starts on the first B
edge after the trigger delay time.
A trigger
source
B trigger
source
A trigger
source
B trigger
source
Waiting for nth
event (Where n=5)
A trigger point
Trigger
delay time
B trigger point
B trigger point
More Operating Tips:
H B-trigger delay time and horizontal delay time are independent functions. When
you establish a trigger condition using either the A trigger alone or the A and B
triggers together, you can also use horizontal delay to delay the acquisition by an
additional amount.
H When using the B trigger, the A trigger can be any of the following types: Edge,
Glitch, or Width. The B trigger type is always Edge type.
To Take Measurements with Cursors
Push the front-panel CURSORS button.
Select the waveform you want to measure and
a cursor type in the cursor control window.
Or you can activate cursors on the selected
waveform directly in the Cursor menu.
Place cursors with the multipurpose
knobs or enter the cursor locations
numerically.
1
2
3
If you choose split cursors to take
measurements between waveforms,
select the source for each cursor.
Read cursor measurement
results in the display.
4
T1: 356.4 ms
T2: 352.5 ms
DT: 3.92 ms
1/DT: 255 Hz
V1: 5.120 V
V2: 4.886 V
DV: 234 mV
DV/DT: 59.7 V/s
5
Other Cursor Measurement Tips:
H You can set the cursors to move together in tandem if you choose the Tracking
cursor mode. The cursors move independently if you choose the Independent
cursor mode.
H If you use the zoom graticule, you can place a cursor directly on a specific
waveform point to take precision measurements.
H Y ou can also move cursors by dragging them to a new position.
To Take Automated Measurements
Click the Meas button, and then select up to
eight measurements using the measurement
control window.
Use the tabs to access measurements in
four different categories.
Or choose a measurement for the selected
waveform directly in the
Measure menu.
Automated Measurement Selections
AmplitudeTimeMoreHistogram (Option 2A)
High
Low
Max
Mean
Pk–Pk
Amplitude
Positive
Overshoot
Negative
Overshoot
RMS
Period
Frequency
Positive
Width
Rise Time
Fall Time
Negative
Width
Delay
Positive
Duty Cycle
Negative
Duty Cycle
Burst
Width
Phase
Area
Wfm
Count
Hits in
Box
Peak
Hits
Max
Min
Pk–Pk
Std
Deviation
m ± 1s
m ± 2s
Min
Cycle
Mean
Cycle
RMS
Cycle Area
Median
Mean
m± 3s
To Customize an Automated Measurement
Use Gating to confine the measurement
to a certain portion of the waveform.
Turn on measurement statistics to
characterize the stability of the measurement.
Adjust the measurement reference levels to
different relative or different fixed values.
Select snapshot to see a one-time
view of all valid measurements.
To Set Up a Histogram (Option 2A)
Drag across the segment of the waveform that
you want the histogram to cover. To set up a
horizontal histogram, for example, make the
box wider than it is tall.
Select Histogram from the drop-down list.
View the histogram at the top or edge of
the graticule.
1
2
3
If you need to make any adjustments
to the histogram, use the histogram
setup control window.
Take automated measurements on histogram
data. See previous page for information.
4
5
To Use Standard Math Waveforms
Select Math Setup in the Math menu.
Choose the math equation that you want to
define. (In this example, it is Math 1.)
Select a waveform, an operator, and a
waveform to build your waveform expression. In
this example, Math 1 is defined as Channel 1
minus Channel 2.
1
2
3
Use the Apply button to create the math
waveform after defining your math expression.
Use the Source, Scale, and Window Type FFT
functions to analyze the frequency components
of your signal.
4
5
To Use Advanced Math Waveforms (Option 2A)
Math Setup in the Math menu.
Select
1
Choose a predefined math espression.
Or click Editor to define a more advanced math
waveform. Then build the waveform expression
using sources, operators, constants,
measurements, and functions.
To Use Spectral Analysis
Choose Spectral Analysis to define FFT
magnitude and phase waveforms. When an
FFT waveform is selected, you can use the
multipurpose knobs to adjust the FFT waveform
just as you would on a spectrum analyzer.
2
Adjust FFT center
frequency
Y ou can view time-domain and
frequency-domain waveforms
simultaneously. You can also use
gating to select only a portion of the
time-domain waveform for spectral
analysis.
Adjust FFT span
To Store Information
To Save and Recall Waveforms
To save or recall waveforms, select
Reference Waveforms and then Save Wfm... or
Recall Wfm... in the File menu.
1
Or click the
Use the reference setup control window to copy
a live waveform into one of four nonvolatile
reference waveform storage locations. Y ou can
also display these waveforms as reference
Select Save Wfm to File to store the live
waveform as a file on a disk drive. Y ou can
recall a waveform stored on disk into one of the
internal reference waveform locations for
Refs button.
waveforms.
display.
To Save and Recall Instrument
Setups
To save an instrument setup, select
Instrument Setup in the File menu.
Or click the Setups button.
2
3
1
Use the settings control window to save the
current setup into one of ten internal storage
locations. Use the pop-up keyboard to label the
setups for easy identification.
Or select
current setup on a disk drive. Y ou can recall any
setup stored on disk and then save it in an
internal setup storage location for faster access.
To restore the oscilloscope to a known initial
state, push the front-panel
Or select
Save Settings to File to store the
DEFAULT SETUP
button.
Recall Default Setup in the File menu.
2
3
To Print a Hard Copy
To print a hard copy to the internal
printer, an attached printer, or a
network printer, push the
front-panel
changes to the page orientation in
the Page Setup dialog box. Choose
screenshots or Banner to print long,
PRINT button or select
Print in the File menu.
If necessary , you can make
Screen-copy to print basic
continuous-feed waveforms.
The Page Setup dialog box also includes
selectors for the print palette and a feature
called Ink Saver. Ink Saver optimizes the
display colors and shades for printing hard
copies on white paper.
To Export Your Results
Y ou can use the Windows clipboard to
copy information. Simply select the item
to copy , copy it, and then paste it into
another Windows application.
Y ou can export waveform data into a
comma-separated ASCII file for use in a
spreadsheet or data analysis program. Select
Export Setup in the File menu to set the output
content and format for images, waveforms, or
measurements.
To Run Application Software
Y ou can install and run optional application software on your oscilloscope. These software packages provide advanced capability
supporting many applications. Two examples are shown below; additional packages may be available. Contact your Tektronix
representative for more information.
Use TDSJIT2 Jitter Analysis Software to
characterize timing performance. Analyze jitter
on contiguous clock cycles using single-shot
acquisitions.
Use TDSDDM2 Disk Drive Measurement
Software to measure disk drive signals
according to IDEMA standards.
Follow the instructions provided with the
application software to install it. To run
the software, select the application in the
File / Run Application menu.
To Connect to a Network
Like any other Windows computer, you can
connect the oscilloscope to a network to enable
printing, file sharing, internet access, and other
communications functions.
To make a network connection, consult with
your network administrator , and then use the
standard Windows utilities to configure the
oscilloscope for compatibility with your network.
Network
To Use a Dual Monitor
Connect a keyboard, mouse, and monitor to the
oscilloscope and configure Windows for
dual-monitor mode. Y ou can operate the
oscilloscope while having full use of Windows
and other installed applications on the external
monitor.
Connect the monitor to the upper SVGA port on
the oscilloscope side panel. Use the Settings
tab in the Windows Display Properties dialog
box to set up a dual-monitor configuration.
To Access the Help System
Click the HELP button or select Help on
in the Help menu to receive
Window
context-sensitive help on the current setup.
Contents and Index in the Help menu to
Select
access any topic in the help system. Select the
topic, and then click the Display button in the
dialog box.
Click an outlined control shown in the help
window to receive more specific information
about the control.
Click a tab in a help window to navigate
between the Overview and specific topics.
Click the Minimize button in a help window to
move help out of the way so you can
operate the oscilloscope. Click the Restore
button to see the last help topic again.
Help
To Use the Oscilloscope I/O
On the Front Panel
Probe compensator output to
compensate and deskew probes.
Auxiliary input
Channel inputs
On the Left Side Panel
USB connector for mouse, keyboard, or other
peripherals
PS-2 connectors for mouse and keyboard
Upper SVGA port to connect a monitor for
dual-monitor operation
Floppy disk drive
CDROM drive on the side of
the oscilloscope; press the
button on the cover to open the
drive
Ground terminal
Lower VGA port to replicate the oscilloscope display
on an external monitor
Parallel port (Centronics) to
connect printer or other device
GPIB port to connect to controller
RJ-45 connector to connect to network
COM1 serial port
Miniature phone jacks for microphone
input and line input and output
Game controller port
On the Rear Panel
External reference input
CH 3 signal output
(4-channel models only)
Trigger signal output
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.