Tektronix TCA-1MEG Primary User

ZZZ
xx
TCA-1MEG High Impedance Buffer Amplier
Instruction Manual
Rev. A
*P071101003*
071-1010-03
xx
ZZZ
TCA-1MEG High Impedance Buffer A
mplier
Instruction Manual
Warning
The servicing instructions are for use by qualied personnel only. To avoid personal injury, do not perform any servicing unless you are qualied to do so. Refer to all safety summaries prior to performing service.
www.tektronix.com
071-1010-03
Copyright © Tektronix. All rights re owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international t reaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and p
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
served. Licensed software products are
rice change privileges reserved.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical su pport:
In North America, call 1-800-83 3-920 0. Worldwide, visit www.tektronix.com to nd contacts in your area.
Warra nty
Tektronix warrants that this produ workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this w the d efective product without charge for parts and labor, or will provide a replacement in exchange for the defective Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or integrated with other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT I TEKTRONIX AN D ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHAN RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AN BREACH OF THIS WARRANTY. TEKTRONIX AN D ITS VENDORS WILL NOT BE LIABL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADV
[W2 – 15AUG04]
N LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED.
TABILITY OR FITNESS FOR A PARTICULAR P URPOSE. TEKTRONIX'
D EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR
E FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
ANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
ct will be free from defects in materials and
arranty period, Tektronix, at its option, either will repair
product. Parts, modules and replacement products used by
products become the property of Tektronix.
Table of Contents
Important safety information ................................... ......... v
General safety summary..... ....................................... v
Service safety summary....................................... .... vii
Terms in this manual ............................................. viii
Symbols and terms on the product .............................. viii
Compliance information ....... .......................................... x
Safety compliance............... .................................... x
Environmental considerations................................... xiii
Preface................................................... ................ xv
Getting Started..................... ....................................... 1
Product Description ........................................ ......... 1
Accessories ......................... ................................. 3
Input Voltage Derating ..................................... ......... 4
Probe Calibration for High Accuracy............ .................. 4
Functionality Not Supported ................................. ...... 5
Installation ........... ................................................ 5
Connecting the TCA-1MEG to the Host Instrument ............. 7
Connecting a Probe to the TCA-1MEG ............... ............ 8
Attaching and Grounding a TCA-1MEG System................. 9
Functional Overview.............................................. 10
Cleaning............................................................ 11
Functional Checks.................................................. .... 13
Prerequisites ..................... .................................. 13
Required Equipment ........................... ................... 14
Run the SPC (Signal Path Compensation) on the Host
Instrument ................................. ................... 16
Calibrate the TCA-1MEG to the Host Instrument.............. 16
Calibrate the P6139B Passive Voltage Probe to the TCA-1MEG
and Host Instrument.............. ............................ 18
LF Probe Compensation .. ........................................ 20
Check Vertical Input Coupling .......... ......................... 22
Check Bandwidth Limit ....................... ................... 26
TCA-1MEG High Impedance Buffer A m plier Instruction Manual i
Table of Contents
Specications ............................. .............................. 31
Performance Verication............................................... 37
TCA-1MEG System Bandwidth ........................... ...... 38
Replaceable Parts. . . ... . ... . ... . ... . . . ... . ... . ... . ... . . . ... . ... . . . ... . .. 43
ii TCA-1MEG High Impedance Buffer Amplier Instruction Manual
List of Figures
Figure 1: TCA-1MEG High Impedance Buffer Amplier............ 1
Figure 2: Connecting and disconnecting the TCA-1MEG .. . ... . ... . . 7
Figure 3: Calibration data message .................................. ... 8
Figure 4: TekConnect and TEKPROBE Interface features . . ... . ... 10
Figure 5: Simplied block diagram. .................................. 11
Figure 6: Probe Compensation Output of the TDS6604 . . ... .. ... . . 15
Figure 7: Setup with the P6139B attached to the TCA-1MEG . .. .. 20
Figure 8: Probe compensation waveforms ................. .......... 21
Figure 9: Test equipment hookup for the Vertical Input Coupling and
Bandwidth Limit checks............................. ............. 22
Figure 10: Typical waveform at GND ................................ 23
Figure 11: Typical waveform at AC Coupling ....................... 24
Figure 12: Typical waveform at DC Coupling....................... 25
Figure 13: Typical waveform at Full (500 MHz) bandwidth . ... . .. 27
Figure 14: Typical waveform at 100 MHz bandwidth .............. 28
Figure 15: Typical waveform at 20 MHz bandwidth................ 29
Figure 16: Typical Input Impedance and Phase versus Freque ncy . 33
Figure 17: Voltage Derating versus Frequency ......... ............. 34
Figure 18: TCA-1MEG dimensions .................................. 35
Figure 19: TCA-1MEG System Bandwidth equipment hookup . ... 39
Figure 20: TCA-1MEG replaceable parts . . .. ... . ... . ... . . . ... . ... . .. 44
TCA-1MEG High Impedance Buffer A m plier Instruction Manual iii
Table of Contents
List of Tables
Table 1: Compatible devices that require high impedance input .. . . . 2
Table 2: Required equipment
Table 3: Warranted electrical specications........... ............... 31
Table 4: Typical electrical characteristics ......................... ... 32
Table 5: Physical charact
Table 6: Environmental characteristics ............................... 35
Table 7: Equipment required for performance verication ......... 37
Tab le 8: TCA - 1MEG test
Table 9: TCA-1MEG replaceable parts. . .. .. . .. .. . . ... .. ... . ... .. ... . 44
.................................... ...... 14
eristics ....................... ............... 35
record...................................... 42
iv TCA-1MEG High Impedance Buffer Amplier Instruction Manual
Important safety information
This manual contains informat by the user for safe operation and to keep the prod uct in a safe condition.
To safely perform service on t is provided at the end of this section. (See page vii, Service safety summary.)
General safety summary
Use the product only as specied. Review the following safety precautions to avoid i products connected to it. Carefully read all instructions. Retain these instructions for future reference.
Comply with local and national safety codes.
For correct and saf follow generally accepted safety procedures in addition to the safety precautions specied in this manual.
The product is designed to be used by trained personnel only.
Only qualied remove the cover for repair, maintenance, or adjustment.
Before use, a is operating correctly.
lways check the product with a known source to be sure it
njury and prevent damage to this product or any
e operation of the pro duct, it is essential that you
personnel who are aware of the hazards involved should
ion and warnings that must be f o ll owed
his product, additional information
This produc
Use personal protective equipment to prevent shock and arc blast injury where haz
While using this product, you may need to access other parts of a larger system. warnings and cautions related to operating the system.
When in system is the responsibility of the assembler of t he system.
TCA-1MEG High Impedance Buffer A m plier Instruction Manual v
t is not intended for detection of hazardous voltages.
ardous live conductors are exposed.
Read the safety sections of the other component manuals for
corporating this equipment into a system, the safety of that
Important safety information
To avoid re or personal injury
Ground the prod uct . This product is indirectly grounded through the
grounding conductor of th e mainframe power cord. To avoid electric shock, the grou nding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, make sure that the product is properly grounded.
Do not disable the power cord grounding connection.
Connect and disconnect properly. Do not connect or disconnect prob es
or test leads while they are connected to a voltage source.
Use o nly insulated voltage probes, test leads, and adapters supplied with the product, or indicated by Tektronix to be suitable for the product.
Observe all terminal ratings. To av o id re or shock hazard, observe all
ratings and markings on the product. Consult the produ ct manual for further ratings information before making connections to the product. Do not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product, probe, or accessory. Use caution when using 1:1 test leads because the probe tip voltage is directly transm itted to the p rod uct.
Do not apply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Do not oat the common terminal above the rated voltage for that terminal.
The measuring terminals on this pro duct are not rated for connection to mains Category III or IV circuits, or to mains Category II circuits greater than 100 V.
When connected to circuits other than mains, the maximum transient overvoltage rating is 400 V peak.
Do not operate without covers. Do not operate this product with covers
or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and
components when power is present.
vi TCA-1MEG High Impedance Buffer Amplier Instruction Manual
Important safety information
Do not operate with susp ecte d failures. If you suspect that there is damage
to this product, have it inspected by qual ied service personnel.
Disable the produ ct if it is dam a ged . Do not use the p ro duct if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not use probes or test leads if they are dam aged , if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before yo u use it. Look for cracks or missing pieces.
not operate in wet/damp conditions.
Do
occurifaunitismovedfromacold to a warm environment.
Do not operate in an explosive atmosphere.
Keep produ ct surfaces clean and dry.
you clean the product.
Service safety summary
The Service safety sum mary section contains additional information required to safely perform service on the product. Only qualied personnel should perform service procedures. Read this Service safety summary and the General safety summary before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of
this product unless another person capable of rendering rst aid and resuscitation is present.
Use care when servicing with power on. Dangerous voltages or currents
may exist in this product. Disconnect power, remove battery (if
Be aware that condensation may
Remove the input signals before
TCA-1MEG High Impedance Buffer A m plier Instruction Manual vii
Important safety information
applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
Verify safety after repair. Alw
performing a repair.
ays recheck ground continuity after
Terms in this manual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could
result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could
result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not imm ediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
When this symbol is marked o n the product, be sure to consult
the manual to nd out the nature of the potential hazards and
any actions which have to be taken to avoid them. (This symbol
may also be used to refer the user to ratings in the manual.)
viii TCA-1MEG High Impedance Buffer Amplier Instruction Manual
Important safety information
The following symbol(s) may appear on the product:
TCA-1MEG High Impedance Buffer A m plier Instruction Manual ix
Compliance information
Compliance information
This section lists the safet y and environmental standards wi th which the instrument complies.
Safety compliance
This section lists the safety standards with which the product complies and other safety compliance information.
EU declaration of conformity – low voltage
Compliance was demonstrated t o the following specication as listed in the Ofcial Journal of the European Union:
Low Voltage Directive 2006/95/EC.
EN 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
EN 61010-2-030. Safety Requirements for Electrical Equipm ent for Measurement, Control, and Laboratory Use – Part 2-030: Particular requirements for testing and measuring circuits.
nationally recognized testing laboratory listing
U.S.
UL 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
UL 61010-2-030. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-030: Particular requirements for testing and measuring circuits.
Canadian certication
CAN/CSA-C22.2 No. 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory U se – Part 1: General Requirements.
CAN/CSA-C22.2 No. 61010-2-030. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory
x TCA-1MEG High Impedance Buffer Amplier Instruction Manual
Compliance information
Use – Part 2-030: Particular requirements for testing a n d measuring circuits.
Additional compliances
IEC 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
IEC 61010-2-030. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 2-030: Particular requirements for testing and measuring circuits.
Equipment type
Test and measuring equipment.
Safety class
Class 1 – grounded product.
Pollution degree descriptions
A measure of the contaminants that could o ccur in the environment around and within a product. Typically the internal environment insid e a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution degree 1. No po llution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. T his location is a typical ofce/home environment. Temporary condensation occurs only when the product is out of service.
Pollution degree 3. Conductive pollution, or dry, nonconductive pollution that beco mes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected fro m direct sunshine, rain, or direct wind.
Pollution degree 4. Po llution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
TCA-1MEG High Impedance Buffer A m plier Instruction Manual xi
Compliance information
Pollution degree rating
Pollution degree 2 (as dened in IEC 61010-1). Rated for indoor, dry location use only.
IP rating
IP20 (as dened in IEC 60529).
Measurement and overvoltage category descriptions
Measurement terminals on this product may be rated fo r measuring mains voltages from one or more of the following categ or ies (see specic ratings marked on the product and in the manual).
Category II. Circuits directly connected to th e building wiring at utilization points (socket outlets and similar points).
Category III. In the b uilding wiring and distribution system.
Category IV. At the source of the electrical supply to the building.
NOTE. Only mains power supply circuits have an overvoltage category
rating. Only measurement circuits have a measurement category rating. Other circuits within the product do not have either rating.
xii TCA-1MEG High Impedance Buffer A mplier Instruction Manual
Environmental considerations
This section provides information about the environmental impact of the product.
Product end-of-life handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the
extraction and use of nat substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. To avoid release of such substances into th resources, we encourage you to recycle t his product in an appropriate system that will ensure that m ost of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002/96/EC and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Support/Service section of the Tektronix Web site (ww w.tektronix.com).
ural resources. The equipment may contain
e environment and to reduce the use of natural
Compliance information
Restriction o
This product is classied as an industrial monitoring and control instrument a restrictions of the recast RoHS Directive 2011/65/EU until July 22, 2017.
TCA-1MEG High Impedance Buffer A m plier Instruction Manual xiii
f hazardous substances
ccessory, and is not required to comply with the substance
Compliance information
xiv TCA-1MEG High Impedance Buffer Amplier Instruction Manual