Tektronix SPECMON3, SPECMON6 Primary User

x
SPECMON3 Real Time Spectrum Analyzer and SPECMON6 Real Time Spectrum Analyzer
ZZZ
Quick Start User Manual
*P071306400*
071-3064-00
xx
SPECMON3 Real Time Spectrum Analyzer and SPECMON6 Real Time Spectrum Analyzer
ZZZ
Quick Start User Manual
www.tektronix.com
071-3064-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by na
tional copyright laws and international treaty provisions.
Tektronix pro previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
ducts are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tektronix.com to nd contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from a ttempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modied or integrated with other products when the effect of such modication or integration increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND E XCLU S IVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
[W2 – 15AUG04]
Table of Contents
General safety summary ..... . . . . . ..... . . . . . ....... . . . . ..... . . . . . ....... . . . . ..... . . . . . ...... . . . . . ...... . . . . ...... . . . . . ..... . . . . . ...... iii
Compliance Information............................................................................................................... v
EMC Compliance................................................................................................................. v
Safety Compliance............................................................................................................... vi
Environmental Considerations................................................................................................. viii
Preface................................................................................................................................. ix
Key Features .....................................................................................................................ix
Documentation ...................................................................................................................x
Software Upgrades. ....... . . . . ...... . . . . . ....... . . . . ....... . . . . ...... . . . . . ..... . . . . . ...... . . . . . ....... . . . . ....... . . . . ...... . . . . . . x
Conventions Used in This Manual..... . . . . . ..... . . . . . ....... . . . . ..... . . . . . ..... . . . . . ....... . . . . ..... . . . . . ....... . . . . ..... . . . . . .. x
Installation.............................................................................................................................. 1
Before Installation................................................................................................................ 1
Standard Accessories. . . ....... . . . . ..... . . . . . ...... . . . . . ...... . . . . ...... . . . . . ...... . . . . ...... . . . . . ..... . . . . . ....... . . . . ..... . . . . . 1
Optional Accessories ....... . . . . ....... . . . . . ...... . . . . . ....... . . . . ....... . . . . ....... . . . . ....... . . . . ........ . . . . ....... . . . . ....... 3
Options ........................................................................................................................... 3
Operating Considerations........................................................................................................ 4
Connecting to the Instrument.... . . . . . ..... . . . . . ....... . . . . ..... . . . . . ...... . . . . . ..... . . . . . ....... . . . . ..... . . . . . ...... . . . . . ..... . . 5
Powering On the Instrument..................................................................................................... 6
Powering Off the I
Removing the Power............................................................................................................. 6
Operating System Restore....................................................................................................... 7
Adding an External Monitor...................................................................................................... 8
Inspecting the Instrument....................................................................................................... 12
User Maintenance. . . . ...... . . . . ..... . . . . . ...... . . . . ..... . . . . ...... . . . . ....... . . . . ...... . . . . ..... . . . . . ...... . . . . ..... . . . . . ..... . . 13
Operation.............................................................................................................................. 14
Getting Acquainted with Your Instrument . . ...... . . . . ...... . . . . . ..... . . . . . ...... . . . . . ..... . . . . . ...... . . . . . ..... . . . . . ...... . . . . . 14
Basic Concepts.................................................................................................................. 23
Navigating Displays............................................................................................................. 25
Selecting Displays...............................................................................................................26
Connecting a Signal. . ..... . . . . . ....... . . . . ...... . . . . . ..... . . . . . ...... . . . . . ..... . . . . . . ....... . . . . ..... . . . . . ....... . . . . ...... . . . . . 31
Starting and Stopping Acquisitions ..... . . . . . ...... . . . . . ..... . . . . . ...... . . . . . ....... . . . . ....... . . . . ...... . . . . . ....... . . . . ....... 32
Markers.......................................................................................................................... 34
Touchscreen Actions............................................................................................................ 37
Printing .......................................................................................................................... 39
Saving Data ..................................................................................................................... 41
Recalling Data...................................................................................................................42
Plotting Measurements ona Map .............................................................................................. 43
Advanced Techniques. . . . . ...... . . . . ..... . . . . . ....... . . . . ...... . . . . ..... . . . . . ..... . . . . . ..... . . . . . ..... . . . . . ...... . . . . ....... . . . . 58
Index
nstrument ..................................................................................................... 6
Table of Content
s
SPECMON3 & SPECMON6 Quick Start User Manual i
Table of Content
s
ii SPECMON3 & SPECMON6 Quick Start User Manual
General safety s
ummary
General safet
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
To avoid re or personal injury
Use proper power cord. Use only the power cord specied for this product and certied for the country of use.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric
shock, the terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the
product manual for further ratings information before making connections to the product.
The inputs are not rated for connection to mains or Category II, III, or IV circuits.
Power disconnect. The power cord disconnects the product from the power source. Do not block the power cord; it
must rem
grounding conductor must be connected to earth ground. Before making connections to the input or output
ain accessible to the user a t all times.
y summary
Do not op
Do not o
qualied service personnel.
erate without covers.
perate with suspected failures.
Do not operate this product with covers or panels removed.
If you suspect that there is damage to this product, have it inspected by
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Replace batteries properly. Replace batteries only with the specied type and rating.
Use proper fuse. Use only the fuse type and rating specied for this product.
Do not operate in wet/damp conditions.
Do not operate in an explosive a tmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation.
proper ventilation.
Refer to the manual's installation instructions for details on installing the product s o it has
SPECMON3 & SPECMON6 Quick Start User Manual iii
General safety s
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ummary
iv SPECMON3 & SPECMON6 Quick Start User Manual
Compliance Info
rmation
Compliance In
This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument complies.
EMC Compliance
EC Declaration of Conformity – EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following specications as listed in the Ofcial Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment for measurement, control, and
laboratory use.
CISPR 11:2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:20
IEC 61000-4-3:2002. RF electromagnetic eld immunity
IEC 61000-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
1234
formation
01. Electrostatic discharge immunity
5
6
7
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, uctuations, and icker
European Contact.
Tektronix UK, Ltd. Western Peninsula Western Road Bracknell, RG12 1RF United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standar
3
To ensure compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
At the IEC 61000-4-11 Voltage-Interruption Transient Immunity test at 0% for 250 cycles, the EUT reboots and can take greater than 10 seconds to recover.
5
Performance degradation information for the IEC 61000-4-3 test: Residual spurious signals can typically increase to -55 dBm with exposure to the disturbance levels of this test.
6
Performance degradation inf either 55 dB below reference level or -75 dBm, with exposure to the disturbance levels of this test.
7
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels (IEC 61000-4-11).
ormation for the IEC 61000-4-6 test: Residual spurious signals can typically increase to the worse of
d may occur when this e quipmen t is connected to a test object.
SPECMON3 & SPECMON6 Quick Start User Manual v
Compliance Info
Australia / New Zealand Declaration of Conformity – E MC
Complies with the EMC provision of the Radiocommunications Act per the following standard, in accordance with ACMA:
CISPR 11:2003. Radiated and Conducted Emissions, Group 1, Class A, in accordance with EN 61326-1:2006 and EN 61326-2-1:2006.
Australia / New Zealand contact.
rmation
Baker & McKe
Level 27, AM
50 Bridge St
Sydney NSW
nzie
P Centre
reet
2000, Australia
Safety Compliance
EC Declaration of Conformity – Low Voltage
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC.
EN 61010-1: 2001. Safety requirements for electrical equipment for measurement control and laboratory use.
U.S. Nationally Recognized Testing Laboratory Listing
UL 61010-1:2004, 2ndEdition. Standard for electrical measuring and test equipment.
Canadian Certication
CAN/CSA-C22.2 No. 61010-1:2004. Safety requirements for electrical equipment for measurement, control, and
ratory use. Part 1.
labo
Additional Compliances
IEC 61010-1: 2001. Safety requirements for electrical equipment for measurement, control, and laboratory use.
Equipment Type
Test and measuring equipment.
Safety Class
Class 1 – grounded product.
vi SPECMON3 & SPECMON6 Quick Start User Manual
Compliance Info
rmation
Pollution Degree Description
A measure of the contaminants that could occur in the environment around and within a product. Typically the internal environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical ofce/home environment. Temporary condensation occurs only when the product is out of service.
Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
Pollution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
Pollution Degree
Pollution Degree 2 (as dened in IEC 61010-1). Note: Rated for indoor use only.
SPECMON3 & SPECMON6 Quick Start User Manual vii
Compliance Info
rmation
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life Handling
Observe the following guidelines when recycling an instrument or component:
Equipment Recycling. Production of this equipment required the extraction and use of natural resources. The
equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. In order to avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2002/96/EC and 2006/66/EC on waste ele ctrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com).
Perchlor
state of California, CR lithium coin cells are classied as perchlorate materials and require special handling. See www.dtsc.ca.gov/hazardouswaste/perchlorate for additional information.
Restric
This product has been classied as Monitoring and Control equipment, and is outside the scope of the 2002/95/EC RoHS Directive.
ate Materials.
tion of Hazardous Substances
This product contains one or more type CR lithium coin cell batteries. According to the
viii SPECMON3 & SPECMON6 Quick Start User Manual
Preface
This manual describes the installation and basic operation of the SPECMON3 Real-Time Spectrum Analyzers and SPECMON6 Real-Time Spectrum Analyzers. For more detailed information, see the online help for your instrument.
SPECMON3
SPECMON6
Key Features
Leading real time technologies help to troubleshoot the toughest transient interferences in the eld:
Unique Swept DPX enables “Real Time Scanning” of the whole 3 GHz/6.2 GHz frequency range for transient interference discovery
Preface
(Opt. 200)
Up to 110 MH
Best in cla Intercept (POI) (Opt 200)
Patented DPX Density Trigger (Opt. 200), Frequency Mask Trigger (Opt. 52) and advanced triggering allows capturing signals with as short as 3.7 μs in the frequency domain and 9.1 ns in the time domain
Multiple-in-one design reduces total cost of ownership with less initial purchase cost and annual maintenance cost
Built-in mapping supports both manual and automatic “drive test” (standard feature). Commercial off-the-shelf third-party GPS rece
Standar
Large r advanced triggering (Opt. 200), the system improves the data post-analysis efciency by 10x. It eliminates many needs of a standalone data recorder by offering up to 12 years storage of DPX Spectrogram/Waterfall Traces (Opt. 53) or up to 7secon of interest according to the customer specied trigger conditions
Leading 3-in-1 multiple-domain correlation and analysis capability simplies the instrumentation needs for time domain, frequency domain and modulation domain analysis
Supports signal demodulation for over 20 general-purpose analog and digital signal types, including AM/FM demod and fle
Comp Power, ACPR, OBW, and Spurious Search
z real time BW for close-in signal discovery and capture
ss real-time capability to discover and capture signals with as short as 3.7us duration with 100% Probability of
iver supported via USB or Bluetooth connection
d Automated Pulse Analysis suite makes eld pulse analysis (for example, airport radar) easier than ever
eal-time memory simplies the long time signal monitoring tasks (standard and Opt. 53). When working with
ds of IQ data at 110 MHz BW (Opt. 110). It also saves post-analysis time by only triggering and storing the signal
xible OFDM signal analysis
rehensive built-in standard eld measurements such as Field Strength, Signal Strength, EMI test, Channel
Standard preamplier provides simplied external connections and solid-state hard drive increases instrument ruggedness
Industry-standard platform lowers system training cost; open data format improves asset utilization by providing compatibility
h industry standard products
wit
dows 7 Ultimate (64-bit) OS with support for Microsoft standard language localization
Win
tured IQ data can be saved into Matlab format for use with third-party software analysis tools
Cap
SPECMON3 & SPECMON6 Quick Start User Manual ix
Preface
Documentation
The following documentation is available with your Tektronix SPECMON3 or SPECMON6 Real Time Spectrum Analyzers. For the most current documentation, refer to the Tektronix Web site.
To read about Use these documents
Installation and operation (overviews)
In-depth operation and user interface help
Application examples Application Examples Manual
Programming commands Programmer Manual
Specications and performance verication
User service
Quick Start User Manual
The quick start user manual contains general information about how to put your instrument into service and guides to user interface controls.
Online help
The online help contains detailed information about how to operate the instrument.
This manual provides practical examples of using the analyzer to accomplish tasks.
This manual contains descriptions of programming commands and their use. This manual islocatedinaPDFle located on the Documents CD.
Specications and Performance Verication Manual
This manual contains the instrument specications and a procedure to check instrument performance against warranted characteristics. This manual is in a PDF le located on the Documents CD.
Service Manual.
A manual that provides a list of replaceable parts, care and maintenance information, and information for servicing the instrument to the module level. This manual is in a PDF le located on the Product Documentation CD.
Software Upgrades
are option upgrades are available. Software upgrades for options become operational only after you enter a valid
Softw option key for the specic analyzer model and serial number.
To check for upgrades:
1. Use your Web browser to go to www.tektronix.com/software.
2. Enter the product name (for example SPECMON3 or SPECMON6) to nd available software upgrades.
Conventions Used in This Manual
The following icons are used throughout this manual:
Sequence Step
Front panel power
Connect power
Network
PS2 SVGA USB
x SPECMON3 & SPECMON6 Quick Start User Manual
Installation
Before Installation
Unpack the instrument, and check that you received all items listed as Standard Accessories. Optional accessories and instrument options are also listed in this section. Check the Tektronix Web site (www.tektronix.com) for the most current information.
Standard Accessories
Your instrument comes with the following accessories: quick start user manual (languages are listed below), programmer manual (on CD), specications and performance verication manual (on CD), declassication and security instructions (on CD), servi
Product Documentation
ce manual (on CD), power cord, BNC-N adapter, U SB keyboard, USB mouse, and instrument front cover.
SPECMON3 Real-Time Spectrum Analyzers and SPECMON6 Real-Time Spectrum Analyzers Quick Start User Manual
is available in the following languages:
Installation
English, Tektronix part number 071-3064-XX.
Simplied Chinese, Tektronix part number 071-3066-XX.
Japanese, Tektronix part number 071-3068-XX.
Russian, Tektronix part number 071-3070-XX.
SPECMON3 Real-Time Spectrum Analyzers and SPECMON6 Real-Time Spectrum Analyzers Application Examples Manual is available in the following languages:
English, Tektronix part number 071-3065-XX.
Simplied Chinese, Tektronix part number 071-3067-XX.
Japanese, Tektronix part number 071-3069-XX.
Russian, Tektronix part number 071-3071-XX.
SPECMON3 and SPECMON6 Product Documentation CD, Tektronix part number 063-4468-XX.
SPECMON3 & SPECMON6 Quick Start User Manual 1
Installation
The Product Documentation CD contains PDF versions of the following manuals:
SPECMON3 Real-Time Spectrum Analyzers and SPECMON6 Real-Time Signal Analyzers Quick Start User Manual, in English, Ja
panese, Russian and Simplied Chinese.
SPECMON3 Real Manual, in English, Japanese, Russian and Simplied Chinese.
SPECMON3 Real-Time Spectrum Analyzers and SPECMON6 Real-Time Spectrum Analyzers Printable Online Help,
Tektronix part number 077-0740-XX.
SPECMON3 Real-Time Spectrum Analyzers and SPECMON6 Real-Time Spectrum Analyzers Programmer Manual, Tektronix p
SPECMON3 an
Tektronix part number 077-0754-XX.
SPECMON3 and SPECMON6 Real-Time Spectrum Analyzers Specications and Performance Verication Manual, Tektronix part number 077-0742-XX.
SPECMON3 and SPECMON6 Real-Time Spectrum Analyzers Declassication and Security Instructions, Tektronix
-Time Spectrum Analyzers and SPECMON6 Real-Time Signal Analyzers Application Examples
art number 077-0741-XX.
d SPECMON6 Real-Time Spectrum Analyzers Service Manual,
part number 077-0754-XX.
Power Cords
The analyzer is shipped with one of the following power cord options. Power cords for use in North America are UL listed and CSA certied. Cords for use in areas other than North America are approved by at least one authority acceptable in the country to which the product is shipped.
International Power Plugs
Opt. A0
Opt. A
Opt. A
Opt. A
Opt.
Opt.
Opt.
Opt.
Opt
Opt
Opt
- North America power
1 - Universal EURO power
2 - United Kingdom power
3 - Australia power
A4 - 240 V, North America power
A5 - S witzerland power
A6 - Japan power
A10 - China power
. A11 - India power
. A12 - Brazil power
. A99 - No power cord
2 SPECMON3 & SPECMON6 Quick Start User Manual
Optional Accessories
SPECMON3 and SPECMON6
RTPA2A - Real-Time Spectrum Analyzer Probe Adapter; Supports TekConnect probes P7225, P7240, P7260, P7330,
P7313, P7350
, P7350SMA, P7380, P7380SMA.
Installation
Options
You can add the following options to your analyzer:
RSAVu - Ofin
WLAN (IEEE 802.11a/b/g/n).
065-0852-XX – Additional Removable Hard Drive for use with Opt. 56 (Windows 7 and instrument software pre-installed).
Transit Case – Tektronix part number 016-1963-00.
Option 10 - Adds Audio Analysis.
Option 11 - Adds Phase Noise and Jitter Measurement.
Option 12 - Adds Settling Time Measurement (Frequency and Phase).
Option 21 - Adds general-purpose digital modulation analysis.
Option 22 - Adds exible OFDM analysis.
Option 40 – 40 MHz Acquisition BW (SPECMON3 only).
Option 52 - Adds frequency mask triggering.
Option 53 - Adds Memory, 4 GB Acquisition Memory Total.
Option 55 - Adds Digital I and Q output
e analysis software based on the RSA3000 platform for analysis supporting 3G wireless standards,
Option 110 – 110 MHz Acquisition BW.
Option 200 - Adds Advanced DPX with D ensity, Time Qualied and Runt Triggers and Zero-Span.
SPECMON3 & SPECMON6 Quick Start User Manual 3
Installation
Operating Considerations
Power supply requirements
Source voltage and frequency Power consumption
100 – 240 V
115 V
RMS
RMS
,400
, 50/60 Hz
Hz
450 W maximum (all models)
Environmen
Characteristic Description
Operating temperature +10 °C to +40 °C
Non-opera
Humidity
Altitude:
Operati
Non-operating
Cooling clearance
Bottom
Sides a
tal requirements
ting temperature
ng
nd back
-20 °C to +60 °C
90% up to 30 °C (non-condensing) 80% relative humidity maximum when accessing the DVD drive Above 30 °
Up to 300
12190 m (40,000 ft)
20 mm (0.79 in)
50 mm (1.97 in)
C it is limited by maximum wet-bulb temperature 29 °C
0 m (9843 ft)
4 SPECMON3 & SPECMON6 Quick Start User Manual
Installation
CAUTION. To ensure proper cooling, do not operate the instrument on its side. Operate the instrument only when it is resting
on its feet
WARNING. To avoid personal injury, take extra care when lifting or moving the spectrum analyzer. The instrument is
heavy and requires extra care when moving it.
, located on the bottom of the case. Maintain at least 2 inches (5 cm.) clearance around ventilation openings.
Connecting to the Instrument
Connecting to a Network
Connect your keyboard, mouse, printer, and other accessories to your instrument before applying power. (Accessories with USB connectors can be co
You can connect your in network for printing, le sharing, Internet access, and other functions. Consult your network administrat Windows utilities to congure the instrument for your network.
Quick Tip
If you connect your
regular approved operating system updates, and using up-to-date antivirus software.
nnected before or after applying power.)
strument to a
or and use the standard
instrument to a network, you should protect your instrument by using an internet rewall, installing
SPECMON3 & SPECMON6 Quick Start User Manual 5
Installation
Powering On the Instrument
Powering Off the Instrument
Removing the Power
6 SPECMON3 & SPECMON6 Quick Start User Manual
Operating System Restore
The instrument contains an operating system restore le on a separate partition of the hard drive.
The preferred method to restore the instrument operating system is to use the hard disk restore le.
CAUTION. Using the restore process reformats the hard drive and reinstalls the operating system. All saved data is lost. If
possible, save important les to external media before performing a system restore.
1. Restart the instrument. During the boot-up process you will see the following message at the top of the screen: Starting Acronis Loader... press F5 for Acronis Startup Recovery Manager
NOTE. To successfully complete the system restore, you must use the Windows version of the Acronis software. Using a
generic MAC keyboard starts the DOS version of the Acronis software. Do not use a MAC keyboard.
2. Repeatedly press the F5 key until the Acronis True Image Tool opens. There is a 5-second time period from when the message appears until the instrument proceeds with the normal instrument startup. If the instrument does not open the Acronis application, power off the instrument, then power on the instrument and try again.
3. Click Restore.
Installation
4. In the C onrmation dialog box, click Yes to restore the instrument operating system, or No to exit the restore process. The restore process takes approximately 30 minutes; the actual time depends on the instrument conguration.
SPECMON3 & SPECMON6 Quick Start User Manual 7
Installation
Adding an External Monitor
Use the following procedure to add an external monitor for dual monitor conguration. Both the analyzer and the second monitor must have the color set to True Color.
1. Power off the analyzer and the external monitor.
2. Connect the external monitor to the analyzer.
3. Power on the analyzer and the external monitor.
8 SPECMON3 & SPECMON6 Quick Start User Manual
4. Open the W indows Control Panel.
5. Double-click Intel(R) Graphics and
Media to open the Intel Graphics and Media Control
Panel.
6. Set the Display setting to Built-in
Display.
Installation
SPECMON3 & SPECMON6 Quick Start User Manual 9
Installation
7. Click Multiple Displays.
8. Set the Operating Mode to Clone Displays or Extended Displays as
preferred.
9. If you select Extended Desktop,besure to set the Positioning to align the two desktop displays.
10 SPECMON3 & SPECMON6 Quick Start User Manual
Installation
Adjusting the Windows Display Settings
To change the instrument display settings, use the Intel Graphics Driver instead of the default Windows display properties. The Intel Graphics Driver provides additional capabilities that are not accessib le from the Windows display properties dialog box.
1. Open the W indows Control Panel.
2. Double-click Intel(R) Graphics and
Media to open the Intel Graphics and Media Control Panel.
3. Adjust the settings as necessary.
NOTE. Not all of the selections in the
y page will be available if there is no
propert external monitor connected to the instrument.
4. Click O
NOTE. A lthough it is not recommended, you can attach a second monitor after the analyzer is powered on; however, after
connecting, you will still need to enable the Extended Desktop from the Intel Extreme Graphics 2M control panel.
K to apply the settings and to
close the properties page.
SPECMON3 & SPECMON6 Quick Start User Manual 11
Installation
Inspecting the Instrument
Run the diagnostics application (Tools > Diagnostics). If failures occur, perform the following steps to get more information about them. You can also use the following steps as a detailed incoming inspection to verify the functionality of your instrument. If you want to check the accuracy specications of your instrument, see the SPECMON3 and SPECMON6 Real-Time Spectrum Analyzers Specications and Performance Verication Technical Reference manual PDF (Tektronix part number 077-0742-XX) located on the Product Documentation CD.
1. Select Tools > Diagnostics.
2. On the Diag Modules, All Tests.
3. Click Single Test.
4. Click RUN.
The instrument will run through the tests one or X icon will appear to the right of each check box as that test is complet the test was passed. The X means the test was failed.
5. For tests that require manual interv instructions to complete the tests.
NOTE. Select the Diagnostics Failure Information tab to see basic diagnostic failure information. Use the Windows Event
r (Control Panel > System and Security > Administrative Tools > Event Viewer) to view failure history and nondiagnostic
Viewe failures reported by the application.
nostics tab, click All
at a time. A check mark
ed. The check icon means
ention, follow the on-screen
12 SPECMON3 & SPECMON6 Quick Start User Manual
User Maintenance
Cleaning Your Instrument
Clean the exterior surfaces of the chassis with a dry lint-free cloth or a soft-bris tle brush. If any dirt remains, use a cloth or swab dipped in a 75% isopropyl alcohol solution. Use a swab to clean narrow spaces around controls and connectors. Do not use abrasi
CAUTION. Avoid getting moisture inside the instrument during e xterior cleaning; use just enough moisture to dampen the
cloth or swab. Do not wash the front-panel On/Standby switch. Cover the switch while washing the instrument. Use only deionized o or distilled water. Do not use chemical cleaning agents; they might damage the chassis. Avoid chemicals that contain benzene, toluene, xylene, acetone, or similar solvents.
ve compounds on any part of the instrument because they might damage the instrument.
r distilled water when cleaning. Use a 75% isopropyl alcohol solution as a cleanser and rinse with deionized
Installation
CAUTION. T
abrasive cleaners or commercial glass cleaners to clean the display surface. Avoid spraying liquids directly on the display surface. Avoid scrubbing the display with excessive force.
Clean the with distilled water or a 75% isopropyl alcohol solution and gently rub the display surface. Avoid using excess force; this might damage the display surface.
Upgradi
Software upgrades are available from Tektronix. They can either be downloaded from the Tektronix Web site or they can be ordered from your local Tektronix representative. To add additional software options or features, you will need an option key from Tektronix. When you receive the software from Tektronix, install it on your instrument. When prompted, enter the option key, and then follow the online instructions to complete the upgrade.
o prevent damage to the at panel display, do not use improper cleaning agents or methods. Avoid using
display surface by gently rubbing the display with a clean-room wipe. If the display is very dirty, moisten the wipe
ng the Instrument Software
Returning Your Instrument
u return your instrument to Tektronix:
If yo
up any user software on the hard disk. You may need to reinstall the software after your instrument is returned.
Back
repacking the instrument for shipment, use the original packaging. If the packaging is unavailable or unt for use,
When
contact your local Tektronix representative to obtain new packaging.
Seal the shipping carton with an industrial stapler or strapping tape.
SPECMON3 & SPECMON6 Quick Start User Manual 13
Operation
Operation
Getting Acquainted with Your Instrument
Controls and display elements are shown in the following illustrations and tables.
Front Panel Controls
Most front panel controls are shortcuts for opening control panels.
Ref number Item Description Menu equivalent
1 Media Removable solid-state drive.
2 Displays
3
4 Trigger
5
6 Analysis
7
8
9 Amplitude
Settings Opens the Settings control panel for the selected display. Setup > Settings
Acquire
Freq
Span (Spectrum) Adjusts the span or frequency range shown in the Spectrum
Opens the Displays dialog box to select measurement displays.
Opens the Trigger control panel. Setup > Trigger
Opens the Acquire control panel. Setup > Acquire
Opens the Analysis control panel. Setup > Analysis
Adjusts the measurement frequency. Setup > Analysis >
1
display.
Adjusts the reference level. Setup >
Setup > Displays
Frequency tab
Setup > Settings > Freq & Span
Amplitude
2
12
2
14 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
Ref number Item Description Menu equivalen
10
1
Applicable only when the selected display is Spectrum, DPX Spectrum, or Spectrogram.
2
Press and hold
BW (Spectrum) Adjust the res
for one second to open the associated control panel.
olution bandwidth (RBW).
1
Setup > Settin BW tab
gs >
12
Ref number Item Description Menu equivalent
12
Run (Stop)
13 Markers, Peak
Push to start and stop acquisitions. Run > Run
Moves the active marker to the maximum peak of the trace in the selected display. If markers are turned off, the MR (marker reference) marker will appear at the maximum peak.
14
15
17
Markers, Select Selects the next marker.
Markers, Dene Opens the Markers control panel.
Control knob Increments/decrements numbers with ne resolution.
18 Arrow keys Move the Markers. The Up arrow moves the selected
marker 10 trace points to the right. The down arrow moves the selected marker 10 trace points to the left. The left and right arrows move the selected marker to the next peak. (Turning the knob moves the marker one trace point left or right.)
19
Increment /
Increments or decrements numbers with coarse resolution.
decrement keys
20 Markers, Delete Deletes the highest-numbered marker.
21 Markers, Add Adds a marker to the selected trace.
22 Replay Replays the current acquisition record.
23
Single Sets the Run mode to Single Sequence and initiates a single
sequence acquisition cycle.
t
SPECMON3 & SPECMON6 Quick Start User Manual 15
Operation
Ref number Item Description Menu equivalent
24 Keypad Enters values in controls.
25 Enter
Completes the data entry in controls. Same as pressing the Enter key on the keyboard.
Ref number Item Description Menu equivalent
26 Recall
27
28
Save Saves working le using the same user selections as for the
Touc
h Screen Off
Opens the Recall dialog box.
ous Save action.
previ
s the touch screen on and off.
Turn
File > Recall
File > Save
16 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
Ref number Item Description Menu equivalen
29 Help Displays the online help. Help >
User Manual
31 Application
32 DPX
33 User
34 Preset
Opens the Appl the effect of pressing Application using the settings at Setup >More presets > Preset Options > Preset type/action.
For instruments with Option 200, opens swept DPX over the full fre without Option 200, opens the DPX display over the maximum real time acquisition bandwidth of the instrument. You can chan settings at Setup > More presets > Preset Options > Preset type/action.
Opens the User preset dialog box. You can change the effect of presets > Preset Options > Preset type/action.
Resets the instrument to its factory default settings. You can change the effect of pressing Preset using the settings at Setup > M
ication preset dialog box. You can change
quency range of the instrument. For instruments
ge the effect of pressing DPX using the
pressing User using the settings at Setup > More
ore presets > Preset Options > Preset type/action.
Setup > Preset
Setup > More Presets > Application
Setup > More Presets >
Setup > More Presets > User
DPX
t
SPECMON3 & SPECMON6 Quick Start User Manual 17
Operation
Front Panel Connectors
Ref number Item Description
1
2 Trig In External T
3
4
5
Trig Out Trigger output connector. 50 , BNC, High > 2.0 V, Low < 0.4 V, (output current 1 mA)
trigger level is user adjustable
USB port USB 2.0 connector
USB port USB 2.0 connector
RF Input
RF input connector 50
rigger input connector, –2.5 V to +2.5 V range,
18 SPECMON3 & SPECMON6 Quick Start User Manual
Rear Panel
Ref number Connector, Description
Operation
1
2
4, 5
6
7
8 External Trigger 2 Input
9
10
11
12
13
14
15 LAN, Ethernet network connector
AC Input, main power connector
GPIB
Real Time IQ Out (Option 55)
+28 VDC output, switched
Microphone in; Headphone, audio output; and Line In connectors
COM 2, serial port for connecting peripherals
VGA external monitor output (resolution not limited to VGA)
PS2 Keyboard input
USB 2.0 ports for mouse and other peripherals (printers, external hard disks)
Ref Out, reference frequency output
Ref In, reference frequency input
SPECMON3 & SPECMON6 Quick Start User Manual 19
Operation
Interface and Display Elements
Commonly-used buttons and controls are located in the tool bar. Most of the buttons open control panels for changing the instrument setups. The contents of the control panels vary depending on the selected display. You can also access the control panels from the front panel buttons or from the keyboard.
Ref number Item Description
1 Displays
2Markers
3
4 Trigger
5
6 Analysis
7
20 SPECMON3 & SPECMON6 Quick Start User Manual
Settings Opens the Settings control panel for the selected display. Each display has
Acquire
Measurement Frequency
Opens the Select Displays dialog box so that you can select measurement displays.
Opens or closes the Marker toolbar at the bottom of the window.
its own control panel.
Opens the Trigger control panel so that you can dene the trigger settings.
Opens the Acquire control panel so that you can dene the acquisition settings.
Opens the Analysis control panel so that you can dene the analysis settings such as frequency, analysis time, and units.
Displays the Measurement Frequency. To change the value, click the text and use the front panel knob to dial in a frequency. You can also enter a frequency with the front panel keypad or use the front panel up and down buttons.
Operation
Ref number Item Description
8
9 Amplitude
10 Replay Runs a new measurement cycle on the existing acquisition data record using
11 Run
12
13
14
Reference Lev
Check Mark indicator
el
Displays the r number from the keypad or use the front panel up and down buttons.
Opens the Amplitude control panel so that you can dene the Reference Level, congure internal attenuation, and enable/disable the (optional) Preamplier.
any new settings.
Starts and stops data acquisitions. When the instrument is acquiring data, the button labe
You can specify the run conditions in the Run menu. For example, if you select Single Sequence in the Run menu, when you click the Run button, the instru Continuous, the instrument will run continuously until you stop the acquisitions.
The check mark indicator ( indicates the display for which the acquisition hardware is optimized. When Best for multiple windows is selected in the Amplitude control p anel's RF & IF Optimization control, none of the measurement displays shows a checkmark, as there is not a single optimized measurement.
Displays the Open dialog box so you can recall a setup le or an acquisition data le.
Opens the Save As dialog box so you can save a setup, screen capture, acquisition data le, or export results or settings.
eference level. To change the value, click the text and enter a
l has green lettering. When stopped, the label has black lettering.
ment will run a single measurement cycle and stop. If you select
) in the upper, left-hand corner of the display
15
Displays the Main Presets dialog window or recalls Preset (Main), depending on current settings.
SPECMON3 & SPECMON6 Quick Start User Manual 21
Operation
Display-Specic Controls
Most displays have commonly-used controls arranged around the graph. They provide quick access to common settings; the full control set for the display is in the Settings control panel. The following illustration and table show some of the common icons that appear in most displays.
Table 1:
Ref number Control icon Description
1
2
3 6.337μs
Common icons
Drop-down List. Click to select a value from the drop-down list.
Knob. Click anywhere on the icon or on the text readout to select the control. You
can then change the value with the front panel control knob or arrow keys, or enter a
from the keypad.
value
icized numbers. These indicate automatically selected control values. Several
Ital
controls in the instrument (for example, RBW and Analysis Length) allow you to select an Auto mode. In Auto mode, the instrument software automatically sets the
e. You can change the control to Manual mode by changing the value. To return
valu to Auto mode, clear the value, and then press the Enter key.
22 SPECMON3 & SPECMON6 Quick Start User Manual
Basic Concepts
Real Time Analyzers
Tektronix real-time analyzers have an optional frequency mask trigger. The frequency mask trigger allows you to set up a spectral mas 100% of the input signal before acquiring, and then capture precisely the intermittent RF event of interest.
DPX™ technology allows you to view the RF signal as a live display. You can view very short transient signals, multilevel signals, and time-varying events. Option 200 includes the ability to measure and trigger on signal density values in the DPX display
k to capture a signal when a spectral anomaly occurs. This allows the real-time spectrum analyzer to check
.
Operation
The analyze troubleshooting and signal characterization applications. All of the measurements, other than the DPX display, are based on the same underlying set of time domain sample data.
Measureme
The analyzer window can show one or any combination of general purpose and application-specic displays.
This example shows a DPX Spectrum Split display. A DPXogram display appears on the top and a DPX Spectrum display appears on the bottom.
rs allow you to view the captured signals in a wide variety of time-correlated displays. This is useful for device
nt Displays
SPECMON3 & SPECMON6 Quick Start User Manual 23
Operation
This example sh displays, a Spectrum display and a DPX spectrum display.
When you have m open, the display with the check mark icon has control over the acquisition hardware. Togiveadisp click its title bar.
This example shows four displays in the window.
ows the window with two
ore than one display
lay priority over any others,
quisition data source can be a live acquisition or a saved acquisition data le. The sharing of acquisition data between
The ac displays allows you to view the same data in multiple ways. For example, you can display a power versus frequency (spectrum) display and a spectrogram diagram at the same time. Because the data source is shared among the different
ays, the displays are correlated. That is, a point or set of points within one display is associated with a point or set of
displ points in every other display in the application window.
24 SPECMON3 & SPECMON6 Quick Start User Manual
Navigating Displays
The analyzers use multiple displays to keep operations simple, and yet provide advanced functionality when needed. Some of the display categories include:
General Signal Viewing displays. These displays show graphs for basic signal parameters such as amplitude, phase, and frequency plotted against frequency or time. Examples of these displays include: Spectrum, DPX, Spectrogram, Time Overview, and Phase vs Time.
Analog Modulation. These displays show graphs and measurements for Analog Modulation, Frequency Modulation, and Phase Modulation.
General Purpose Digital Modulation displays. Available with Option 21, General Purpose Modulation Analysis. These displays show the results of demodulating s ignals and analyzing them in multiple domains. Examples of these displays include: Constellation, EVM vs Time, Eye, Symbol Table, and Trellis diagrams.
RF Measurements displays. These displays show the results of automated measurements on RF signals. Examples of these displays include: Channel Power/Field Strength and ACPR, MCPR, Phase Noise, and CCDF.
OFDM Analysis. These displays provide OFDM analysis for WLAN 802.11a/j/g and WiMax 802.16-2004 signals.
Pulsed RF displays. Available with Option 20, Advanced Signal Analysis. These displays show the results of advanced
analysis for pulsed RF signals. Examples of these displays include: Pulse Statistics, Pulse Table, and Pulse Trace.
Audio Analysis. These displays measure basic time- and frequency-domain parameters of analog audio signals modulated on a carrier (AM, FM and PM modulation) or unmodulated (non-carrier) audio signals (Direct).
Operation
SPECMON3 & SPECMON6 Quick Start User Manual 25
Operation
Selecting Displays
The analyzer provides an easy way to select displays.
1. Click the Displays icon to open the Select Displays dialog box.
There are two ways to select displays: choose any combination of displays or use an Application Preset.
Choose any Combination of Displays. Select displays one by one as needed.
1. Click the desired application folder
in the Select Displays dialog box.
The available d isplays for each folder are listed in the Available Displays pane.
2. Click one of the display icons.
3. Click Add (or double-click the
icon) to add the selected display to the S elected displays list. Repeat the procedure to add other displays.
4. Click OK to close the dialog box. The selected displays will open.
26 SPECMON3 & SPECMON6 Quick Start User Manual
Use an Application Preset. Select displays from the presets for your application.
Operation
1. Click Applica
Displays dialog box.
2. Select one
NOTE. S ome presets appear only if the
associated instrument option is installed.
3. Click OK to close the dialog box. The
preset di
tion Presets in the Select
of the presets.
splays will open. (See Table 2.)
Table 2: Application presets
Presets Displays
Modulation Analysis (Option 21 only) Constellation, DPX Spectrum, Signal Quality, Symbol Table
Pulse Analysis (Option 20 only) DPX Spectrum, Pulse Table, Pulse Tr ace, Time Overview
Spectrum Analysis Spectrum
Spur Search Multi Zone 9k-1GHz Spurious
Time-Frequency Analysis
SPECMON3 & SPECMON6 Quick Start User Manual 27
Frequency vs Time, Spectrogram, Spectrum, Time Overview
Operation
Quick Tips
Push the Displays button on the front panel to open the Select Displays dialog box.
To add a display, double-click the icon.
To remove displays, do one of the following:
Select a display icon in the Selected Displays list, and then click Remove.
Double-click a display icon in the Selected Displays list.
Click the Close button in the upper right hand corner of the selected display.
You can mix displays from multiple application folders.
Common Information Messages
The following table lists some of the common information messages that might appear during normal operation.
Table 3: Common information messages
Message Explanation and recommended action
Acq BW too small for current setup
The displ contains. This can be due to any of the following reasons:
ay needs a wider acquisition bandwidth than the current data record
A display other than the one you intended has been selected. The selected display has requested a smaller acquisition bandwidth to achieve a better
y or dynamic range for its particular measurement.
accurac
Select the display that contains the message. Click Run if the instrument is not already acquiring data.
Acquisitions are not running and the measurement now requests a wider
dth than the last acquisition.
bandwi
Click Run to perform a new acquisition with a wider bandwidth.
The data is from a recalled le.
There is no way to increase the acquisition bandwidth for saved data. You
djust the measurement settings so that less bandwidth is required.
must a
Need swept acq or larger Acq BW When any display requires a swept acquisition, the other displays are unable to
ess the swept data. This is because the swept data is customized for the
proc selected display.
Select the display that you are interested in and the acquisition settings will
matically change to meet the requirements of the selected display.
auto
RBW increased to 100 kHz
current Spectrum Length or Analysis Length is not long enough to allow
The a narrower RBW lter.
If the Length control is set to Manual, try increasing it or setting it to Auto so
t the Analysis Length will increase to the required value.
tha
28 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
Changing Measurement Settings
The displays have settings that control how signals are acquired and processed. The settings that you can change depend on the selected display.
To access the settings (in this case, for the Spectrum display):
1. Click Settings to open the Settings
control panel.
2. Adjust the controls as needed for your
signal or test. For example, use the Freq & Span tab to set frequency and span. Most of the control panels have tabs that provide easy access to other related controls.
3. Yo u can cha
display screen in the Prefs tab. Not all of the following controls apply to every display. controls not described here.
To turn on or off the graticule, select
the Show graticule check box.
To turn on or off the marker readout,
select t graph check box.
To turn on or off the trace legend,
select the Show trace legend check box.
nge the appearance of the
Some displays have additional
he Show Marker readout in
Quick Tip
have multiple displays on the screen, click anywhere inside one of the other displays to change to the control
If you
panel for that display.
SPECMON3 & SPECMON6 Quick Start User Manual 29
Operation
Presetting the Instrument
You can preset the instrument to restore the setups to their default values.
1. To start with the default instrument settings, push the Preset button on the front panel or select Preset from the Setup menu.
Quick Tip
Preset does not change any of the settings specied from the Cong In/Out control panel or from the Tools menu, such as GPIB
address, Save/Export preferences, or the Auto Alignment setting.
30 SPECMON3 & SPECMON6 Quick Start User Manual
Connecting a Signal
The analyzer has a single RF signal input located on the front panel. Make sure that you observe the input signal requirements before connecting a signal to the input.
CAUTION. To prevent static damage to the instrument or to any instrument accessories, discharge to ground any
electrostatic charge that may be present on the center and outer conductors of cables, before attaching the cables to the instrument inputs. Do not create an ESD (electrical static damage) antenna by leaving cables attached to the instrument with the other ends of the cables open.
Required Signal Levels
The following table shows the input signal levels over which the analyzer can be used for measurements. The accuracy is guaranteed at a signal amplitude and frequency (normalization point). Measurements can be made on signals within the entire ran signal to be measured (see the SPECMON3 and SPECMON6 data sheet for more information).
ge of allowable inputs, but the accuracy of the measurement is affected by the frequency and amplitude of the
Operation
RF signal
Modulation measurement requirement Spectral display requirement
-50 dBm t
Input si
Characteristic Description
Input i
Input frequency range
Maxi
Maxi
CAU
voltage rating is ±5 V DC. The maximum pulse specications are 50 W peak, <10 µs pulse width, 0.001 duty cycle and 30 dB attenuation.
input range (CW or peak envelope power)
o +30 dBm
DANL to +
30 dBm
gnal requirements
mpedance
SPECMON3
ON6
SPECM
SPECMON3/SPECMON6
mum measurable continuous input power
mum voltage rating
TION. To prevent damage to the instrument, do not connect CW signals greater than 1 Watt (+30 dBm). The maximum
50
9 kHz to 3 GHz (RF Band)
9 kHz to 6.2 GHz (RF Band)
o 32 MHz (LF Band)
1Hzt
1 W (+30 dBm)
±5 V DC, AC coupled
SPECMON3 & SPECMON6 Quick Start User Manual 31
Operation
To connect an RF signal:
Connect the si connector on the front panel.
gnal to the RF INPUT
Quick Tip
If the source signal is greater than 1 Watt (+30 dBm), connect the source signal to an attenuator. Connect the attenuator output to the RF Input connector.
Starting and Stopping Acquisitions
1. ClicktheRunbuttontostartan
acquisition.
To select the Run mode, click the drop-list icon to the right of the Run button.
Select Single or Contin uo us to set the Run mode as necessary.
e analyzer is in Continuous
If th Run mode, the Replay button is dimmed while acquisitions are
ve. If there are no acquisitions
acti active, the Replay button is blue.
32 SPECMON3 & SPECMON6 Quick Start User Manual
2. Click the Stop button again to stop
an acquisition. If an analysis is in process, it wi and then stop.
ll nish its sequence
Quick Tip
Push the front panel Run button to start and stop acquisitions.
Run Control
The Run menu provides access to commands that control the acquisition and the measurement sequence. Refer to the online help for information on all of the Run menu selections. Continuous is the default selection. When you select Continuo single acquisition every time you click Run or push the Run button on the front panel.
us, the instrument will continuously acquire data until you stop the acquisitions. Select Single Sequence to run a
Operation
SPECMON3 & SPECMON6 Quick Start User Manual 33
Operation
Markers
Use markers to measure time, frequency, power and other results. Markers measure absolute values and can also measure the difference between markers.
Using Markers to Measure Frequency and Power
The following procedure shows how to use markers in the Spectrum display to measure frequency and power. The example assumes that
1. Click Markers to open the
2. Select Add Marker from the
all markers are turned off and that a signal is on the screen.
Marker toolbar at the bottom of the Analysi
drop-down list on the Marker Toolbar.
The rst marker is designated the marker reference (MR).
s window.
3. Position the marker on the screen u the Marker toolbar. You can also drag the marker to the desired positi or the mouse.
4. Read the frequency and power level from the marker readout on the dis
sing the arrow buttons in
on using the touch screen
play.
34 SPECMON3 & SPECMON6 Quick Start User Manual
Markers Display
Rectangular markers show absolute values in the readout. The readout of the selected marker appears in the upper left
corner of the display and in the Marker toolbar.
Diamond markers show delta values (relative to the reference marker). The delta readout of the selected marker appears
in the upper right corner of the display and in the Marker toolbar.
When you drag markers along a trace with either the touch screen or the mouse, the marker will seek the high points. The marker stays within 20 trace points of the mouse position. If there are no peaks within the current range of trace points, the marker nds the highest point.
When you attach the marker to the bitmap trace in the DPX spectrum display, the marker readout indicates the hit count for pixels (maximum 65,536) along with frequency and power (see the following gure). You can position the marker at any point in the graph.
In an instrument with Option 200, markers attached to the DPX bitmap display the signal density in percent at the selected pixel, rather then the hit count.
Operation
Quick Tips
If markers are turned off, click Peak in the Marker toolbar or on the front panel to add the marker reference (MR) and to
ion it on the maximum peak level on the trace in the selected display.
posit
Tab le in the Marker toolbar to open a tabular display showing the marker information.
Click
SPECMON3 & SPECMON6 Quick Start User Manual 35
Operation
Marker Toolbar
The Marker toolbar has several controls to work with markers. (See Table 4.) Some of these controls are similar to those on the front panel.
Table 4: Marker t oolbar
Icon Description
Opens the Dene Markers control panel. Use the control panel to add and remove markers and to set marker parameters.
Sets the Measurement Frequency of the instrument to the frequency of the selected marker. This button is only present when the selected display has frequency on the horizontal axis.
Moves the marker to the maximum peak level on the trace. The maximum peak can be either the highest or the lowest point in displays with both positive and negative values on their vertical axis (such as Frequency Error or Phase).
Moves the selected marker to the next peak to the left of the current position.
Moves the selected marker to the next peak to the right of the current position.
Moves the selected marker to the next lower peak value. The peak value refers to the absolute value of the peak amplitude. When repeatedly moving the marker, it can move in any direction depending on the location of the next peak. If the marker is attached to the DPX bitmap trace, this button moves it to the next-lower-amplitude density peak at the same frequency.
Moves the selected marker to the next higher peak value. The peak value refers to the absolute value of the peak amplitude. When repeatedly moving the marker, it can move in any direction depending on the location of the next peak. If the marker is attached to the DPX bitmap trace, this button moves it to the next-higher-amplitude density peak at the same frequency.
Opens or closes the marker table in the display.
Adding Markers
an have up to ve markers. The rst marker (MR) is a reference marker. The other markers (M1, M2, M3, and M4) can
You c show absolute values or can show both absolute values and the difference values between them and the reference marker.
To add another marker, select Add Marker from the drop-down list in the toolbar.
36 SPECMON3 & SPECMON6 Quick Start User Manual
Dening Markers
Use the Dene Markers control panel to dene the marker settings.
1. Click Dene to open the Dene Markers control panel.
Use the control panel to:
Add or delete markers.
Turn markers off.
Assign markers to specictraces.
Set the marker readouts to absolute or delta.
Dene the threshold and excursion values.
Operation
Quick Tip
You can add or delete markers from the drop-down list next to Dene in the M arker toolbar.
Touchscreen Actions
You can use the touchscreen to change marker settings and how waveforms are displayed by using the Touchscreen Actions menu. To use the Touchscreen Actions menu, touch the display and hold for one second, then remove your nger. You can also use a mouse to display the Touchscreen Action menu by clicking the right mouse button.
Some selections in the Touchscreen Actions menu are selectable only when a marker is dened.
Touchscreen Actions menu
SPECMON3 & SPECMON6 Quick Start User Manual 37
Operation
Table 5: Touchscreen actions menu
Icon Menu item Description
Select Selects markers and adjusts their position.
Span Zoom
CF Pan Adjusts the Center Frequency according to horizontal movement.
Zoom
Pan
Reset Scale Returns the horizontal and vertical scale and position settings to their default
Marker to peak
Next Peak
Add marker
Delete marker Removes the last added marker.
All markers off
Trigger On This
Zooms the graph area about the selected point. Touch the graph display at a point of interest and drag to increase or decrease the span about the point of interest. Span Zoom adjusts the span control and can affect the acquisition bandwidth.
Adjusts horizontal and vertical scale of the graph. The rst direction with enough movement becomes the primary scale of adjustment. Adjustment in the secondary direction does not occur until a threshold of 30 pixels of movement is crossed.
Dragging to the left o r down zooms out and displays a smaller waveform (increases the scale value). Dragging to the right or up zooms in and displays a larger waveform (decreases the scale value).
Adjusts horizontal and vertical position of the waveform. The rst direction with enough movement becomes the primary direction of movement. Movement in the secondary direction does not occur until a threshold of 30 pixels of movement is crossed.
values.
Moves the selected marker to the highest peak. If no marker is turned on, this control automatically adds a marker.
Moves the selected marker to the next peak. Choices are Next left, Next right, Next lower (absolute), and Next higher (absolute).
Denes a new marker located at the horizontal center of the graph.
Removes all markers.
Positions the DPX Density measurement box at the click point, measures the current signal density, sets up the DPX Density Trigger function based on the measurement result, and starts acquisitions.
38 SPECMON3 & SPECMON6 Quick Start User Manual
Printing
You can print a copy of the screen and its contents (a screen shot) to a printer or a le.
1. Push the Print button on the front panel
Operation
or select Print from the File menu.
SPECMON3 & SPECMON6 Quick Start User Manual 39
Operation
2. Fill out the det and then click OK.
The Print dialog box is the standard Microsoft Win printer controls are dependent on your printer.
Ink Saver M
You can print screen images with a white background.
1. Select Options from the Tools menu.
ails in the Print dialog box,
dows Print dialog box. The
ode
2. Select the Colors tab in the control panel.
3. Set the Color Scheme to B lizzard.
The background color changes from black to white in the graphs.
40 SPECMON3 & SPECMON6 Quick Start User Manual
Saving Data
You can save different types of data for future use.
1. Select Save As. from the File menu to
open the Save As dialog box.
2. Navigate to the folder where you want
to save the location.
3. Enter a le name.
4. Select the type of le to save. (See
Tabl e 6. )
Operation
setups, or use the default
5. Click Save.
Table 6
File ty
Setup
Results Export les Saves the trace and numeric data for the selected display. The trace and numeric data
Pic
Selected Trace Saves a trace for later recall back into the display from which it was saved.
Data (acquisition data and ac
: File types for save
pes
les
ture (PNG/JPG/BMP)
quisition data export)
ption
Descri
all of the setup information for all settings displays, except those settings that
Saves are specied under the Tools menu (Alignments and Options) or in the CongIn/Out control panel.
aved in CSV format as text les.
are s
es a graphical representation of the screen in the s pecied format. This option is
Sav useful for including the g raphic in reports or other applications. Marker readouts and other information are included.
Saves data for reanalysis by the SPECMON3 and SPECMON6 analyzers or as CSV
omma-separated value) or MAT (Matlab) format to use with external software.
(c
SPECMON3 & SPECMON6 Quick Start User Manual 41
Operation
Quick Tips
Select File > Save or press the Save button to save the information to a le using the same name, path, and data type as the last time that you saved. If the le already exists, the le name sufx number will increment if you previously selected that option in the Tools > Options > Auto Filename control panel tab.
Select File > Save As to open the standard Windows S ave dialog box where you can edit the le name, choose what data type to s ave, and select a folder for storing the le.
Recalling Data
You can recall and reanalyze acquisition data and setup les previously saved by the analyzer.
1. Push the Re or select Recall from the File menu.
call button on the front panel
42 SPECMON3 & SPECMON6 Quick Start User Manual
2. In the Open dialog box, browse to the
location of the le.
3. If you do not see the le, select the type
of le from the
drop-down list.
Operation
4. Double click
the le name to load the le.
Quick Tip
If you recall a trace that has an x-axis range that does not fall within the x-axis range of the current graph, the trace will not be visible. A small arrow will point off-screen in the direction of the recalled trace. Use the Frequency and Span (or Span and Offset) c (which can also be used in the Spectrum displa y). These controls are on the Settings > Scale tab.
ontrols to bring the trace on the screen in the Spectrum display. In other displays, use Horizontal Scale and Offset
Plotting Measurements on a Map
The analyzer includes a program named RSA Map that allows you to record m easurement results and their location on a map.
Map Files
RSA Map uses MIF format map les or Windows bitmap les (.bmp) to indicate location. The .bmp format map les can
her geo-referenced (using world map le format to specify geographic location) or non geo-referenced (which does
be eit not have a geographic location information).
Launching RSA Map
To launch RSA Map:
1. In the SP ECMO N application, select
Tools > RSA Map from the menu bar.
SPECMON3 & SPECMON6 Quick Start User Manual 43
Operation
RSA Map will launch in a separate window.
Loading a Map
Before you can use the RSA Map tool to record measurements you must load a map. If you loaded a map in a previous session, that map will remain loaded until it is replaced by a new map.
1. Select F
2. Navigat
ile > Load Map .
e to the location of them map le to be loaded, and select the map le to load (you may need to specify the le
the map.
type of
44 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
3. Click Open. The RSA Map application
will load the selected map le.
Setting Up a GPS Receiver
To include GPS-derived coordinates with measurements when they are placed on a map, you must enable GPS in RSA Map.
NOTE. An external GPS receiver with USB interface is recommended. The GPS receiver must output NMEA 0183 serial
stall any necessary software for the GPS receiver before running RSA Map.
data. In
1. Connec
2. Click on the GPS icon.
3. Click the Enable check box.
4. Click the COM port drop-down list arrow
5. Set the Baud Rate as appropriate for the
6. Click Apply to enable GPS.
t a GPS receiver with USB interface to one of the USB ports on the instrument.
isplays the Setup GPS window.
This d
and scroll through the list. Select the port that displays the name of the attached GPS receiver.
connected GPS receiver.
SPECMON3 & SPECMON6 Quick Start User Manual 45
Operation
When the GPS receiver is communicating with the analyzer, the readouts in the Status area will upda
te to display current location
information.
46 SPECMON3 & SPECMON6 Quick Start User Manual
7. Click OK to accept the GPS settings you have entered and close the Setup GPS window.
When the analyzer is locked onto the GPS signal, the GPS icon displays a green lock icon.
Mapping a Measurement
Use the following procedure to map measurement results. Before you use the following procedure, do the following:
Load a map. (See page 44, Loading a Map.).
Select Tools > SPECMON to switch to the SPECMON application and set up the measurement you wish to map in the analyzer application. If you want to map Signal Strength, set up the Chan Power/STR and ACPR display on the analyzer as required for the signal you will be measuring.
1. In the RSA Map application, select one of the measurement buttons. The selected measurement is surrounded with a blue highlight and the name of the measurement appears below the Thumbnail area.
Operation
k the Single Measurement button.
2. Clic
SPECMON3 & SPECMON6 Quick Start User Manual 47
Operation
3. Click or tap on the map.
If a GPS receiver is active, and you are using a geo-referenced map, a single measurement i GPS coordinates.
If a GPS receiver is not active, or if you are using an image map, a single measurement tapped the map
4. You can select a different measurement button, and then tap the map to add a new measurement to the map. Multiple measurement icons at the same location stack on top of each other.
s placed at the current
is placed where you
5. To move an measurement result (for example stacked), tap the Move Result button.
6. Tap and to a new position. If the icon is m oved to a new stack of measurements, then t time-of-measurement order.
, if two or more results are
drag the measurement result
he icon is placed in the stack in
48 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
NOTE. The number of measurements that can be placed on a map is limited to 200. When the limit is reached, RSA Map
shows a message As to save the map and measurements to a le, or select File > Clear All Measurements to clear the map and resume adding measurements to the map.
stating that the limit has been reached, and discards any following measurements. Select File > Save
NOTE. To add m
move to the new location. If you want to add a measurement at a point other than the current GPS location, disable GPS before placing the measurement.
NOTE. To move measurement icons on a map, select the M ove Result button ( ). Click on the measurement and
drag it to the new location.
easurements at a new location on a geo-referenced map with an active GPS receiver, you must physically
Repeat RSA Map Measurements
Repeat measurements lets you automatically map a single measurement at a set time interval or change in GPS position. To perform r
NOTE. The RSA Map program maps only completed measurement results. If the specied time interval is less than that
required to take a measurement, RSA Map ignores the specied time interval and maps results as soon as the measurement acquisi 10 seconds, then the instrument maps results every 20 seconds.
1. Set up the measurement parameters
epeat measurements, GPS must be enabled and locked. The map used must be a geo-referenced map.
tion is done. For example, if the instrument requires 20 seconds to take a measurement, but the time interval is set to
(frequency, bandwidth, and so on) in the
ON application.
SPECM
2. Selec
3. Enable GPS as previously described.
4. Tap the Repeat Measurements button.
5. Tap the Settings button display the
t Tools > RSA Map to return to the
RSA Map application.
ttings window.
Se
SPECMON3 & SPECMON6 Quick Start User Manual 49
Operation
6. Set the type of measurement interval (time or distance between measurement positions) as
needed.
7. Ta p Done.
8. Tap the map at any point to start repeat
measurements. The instrument adds a measurements to the map at the current GPS position.
The instrument will repeat measurements based on time or distance as specied on the
(Settings) > Repeat Measurement tab.
9. Tap the map to stop measurements:
50 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
10. Tap the Single Measurement or the
Select icons to exit the RSA Map logging measurements
NOTE. The number of measurements that can be placed on a map is limited to 200. For repeat measurements, when the
instrument reaches the measurement limit, a message is briey displayed, the measurements and map are automatically saved to the c to the map.
mode.
urrent Save Results directory, the map is cleared, and then RSA Map continues adding measurements
RSA Map Measurement Icons
The RSA Map application uses icons on the map to represent a measurement taken at that location. You can perform various actions on measurement icons, including viewing the measurement results of an icon, moving the icon, deleting an icon, and attaching a measurement direction arrow to an icon. See the following sections, and the RSA Map online help for more information.
Viewing and Deleting Measurement Icons
To view or delete the measurement results associated with an RSA Map measurement icon:
1. Ta p t he
2. Tap a measurement icon to open the
3. Tap the measurement of interest in the
RSA Map Select button.
Map Measurements dialog box.
.
list
SPECMON3 & SPECMON6 Quick Start User Manual 51
Operation
4. Ta p Open Measurement to open the measurement screen for the selected measurement. to return to the RSA Map screen.
5. Ta p t h e Delete Measurement button to delete the selected measurement. If there are multiple measurements to delete, repeatedly select and tap Delete Measurement.
Select Tools > RSA Map
6. TapOKtoclosethedialogbox.
NOTE. To delete all the measurement icons from a map, select File > Clear All Measurements.
Draw a M easurements Azimuth (Direction) Arrow
The RSA Map azimuth direction arrow function lets you draw an arrow on a mapped measurement to indicate the direction your antenna was pointing when you took a measurement.
1. Tap the Change Azimuth button.
52 SPECMON3 & SPECMON6 Quick Start User Manual
2. Tap and drag from the center of a
measurement icon in the direction the antenna wa measurement was made. RSA Map draws an arrow from the measurement icon.
The RSA Map measurement thumbnail display shows the direction of the arrow as you draw it. This display shows the True Azimuth and Magnetic Azimuth. As you change the direction of the arrow, the angles shown in the thumbnail display are updated to help you set the direction accurately. The information shown depends on the map type (geo-referenced or image). The example shown is for a geo-referenced map.
s pointing when the
Operation
3. To change the arrow direction, draw a
new line. RSA Map replaces the existing arrow with the new direction arrow.
You can also change the direction of the arrow by tapping the measurement, selecting Change Azimuth from the Map Measurements window and entering a Magnetic Azimuth value for the result.
4. To change the style of the azimuth line,
select View > Azimuth line style – Line or Azimuth line style – A rro w .
Delete an Azimuth Measurement Direction Arrow
1. Tap the RSA Map Select button.
SPECMON3 & SPECMON6 Quick Start User Manual 53
Operation
2. Single-tap the measurement icon with the direction arrow to delete. RSA Map opens the Map M
easurements dialog
box.
54 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
3. Tap the measure
direction arrow to delete.
4. Ta p t he Delete Azimuth button.
5. Ta p OK. RSA Map closes the dialog box
and deletes the direction arrow.
ment name that has the
Saving Measurement Results
RSA Map all results. RSA Map saves results in a compressed .zip format. The saved results contain several types of les:
Measurement data les (exported results)
Map le used for the measurements
Google Earth KMZ le
Recallable results les (trace and setup les)
ows you to save measurement results to a le for documentation. Saved results can also be recalled to review
MapInfo-compatible MIF/MID les
For complete details on the saved archive contents, see the RSA Map Help.
To save measurement results:
1. Select File > Save As.
SPECMON3 & SPECMON6 Quick Start User Manual 55
Operation
2. Navigate to the to save the results and type a name for the le.
3. Select Save to save the results.
location where you want
Recalling Measurement Results
To recall saved measurement results:
1. Select Fil
e > Recall.
2. Navigate to the location where you want
to save t the le.
3. Select Save to save the results.
he results and type a name for
56 SPECMON3 & SPECMON6 Quick Start User Manual
RSA Map loads the contents of the recalled measurement archive.
Operation
SPECMON3 & SPECMON6 Quick Start User Manual 57
Operation
Advanced Techniques
This section contains advanced operating techniques. For examples of practical applications, see the SPECMON3 Real-Time Spectrum Analyzers and SPECMON6 Real-Time Spectrum Analyzers Application Examples Manual.
Setting Up Triggering
The analyzers offer real-time frequency domain triggering (option 52) and frequency edge triggering in addition to power-level triggers, li acquisition of the signal. This allows the analyzer to store both pretrigger and posttrigger information.
The analyzer has two triggering modes:
Free Run mode
is a fast and easy way to see the signals. Free Run is usually adequate for the Spectrum display unless you need to specify a particular time at which to collect the data record.
Triggered mode. In Triggered mode, the instrument initiates an acquisition when a trigger event is recognized. The
conditions that dene a trigger event depend on the selected trigger source. There are several source selections available for choosing the signal to monitor for a trigger event.
ne trigger, runt trigger, and external triggers. It uses a trigger event as a reference point in time for the seamless
.
In Free run mode, the instrument initiates acquisitions without considering any trigger conditions. It
Once you selec
RF Input
Trig In
Trig2In
Gated
Line
Once you select the trigger source, you can set additional parameters to specify the trigger event. If you specify the RF Input as the trigger source, you can chose from two trigger types:
t the Triggered mode, you must select the source. The trigger sources are:
DPX Density Triggering. This trigger type enables the instrument to capture any signal you can see in the DPX display,
including difcult to see signals hiding beneath other signals that have a higher amplitude or a greater repetition rate. Using the Trigger On This™ feature, you can draw a box around the area of interest in the graph, set a Density threshold, and the instrument will trigger when the measured Density exceeds your threshold.
Time Qualied Triggering. Time Qualied triggers allow you to trigger the instrument based on when signal transitions
occur in time. You can apply time qualication to the basic trigger types. For example, if you select the Power type and congure it to trigger on a rising edge, Time Qualication allows you to screen rising-edge events based on how long the signal stays high after it exceeds the trigger threshold. You may wish to trigger only on pulses that last longer than a specied time or shorter than a specied time. You can set Time Qualied to accept only pulse widths between two specied values, or outside of these two values.
You can also time qualify Frequency Edge, Frequency Mask, DPX D ensity, and Runt trigger events. For each trigger type, the instrument nds all events that meet the basic trigger criteria, then further tests them against your time qualication parameters.
Frequency Edge Triggering. Frequency Edge triggering triggers the instrument on time-domain signal characteristics.
The incoming data is compared to a user-selected level in Hertz. Frequency edge trigger is similar to Power triggering, except you also need to set a power threshold to avoid triggering on the apparent frequency uctuations that occur at low amplitude. You can select the time-domain bandwidth and trigger on the rising or falling edge.
58 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
Frequency Mask Triggering. Frequency Mask Triggering allows you to trigger the instrument when a signal in the
frequency doma generate the trigger event. It allows you to trigger on weak signals in the presence of strong signals. This triggering is also useful for capturing intermittent signals.
in violates the mask. You can draw a mask to dene the conditions within the real-time bandwidth that will
Power Triggering. Power triggering triggers the instrument on time-domain signal characteristics. The incoming data is
compared to a user-selected level in dBm. You can select the time-domain bandwidth and trigger on the rising o r falling edge.
ing.
Runt Trigger
to cross a second threshold before recrossing the rst.
In addition to dening the trigger event, you can dene the following parameters by selecting the appropriate tab in the Trigger control panel:
Position and Delay parameters that dene where a trigger occurs within the acquired data record and whether to
delay the t
Runt triggering denes a trigger event based on a pulse amplitude that crosses one threshold but fails
rigger for a set amount of time.
Whether a s
swept acquisition mode.
ingle event can trigger an entire sweep, or each s egment of the sweep requires a new trigger event in
SPECMON3 & SPECMON6 Quick Start User Manual 59
Operation
Accessing Trigger Parameters
1. Click Trig to o
panel.
2. As shown in the following steps, click one of the tab panel to dene the trigger parameters.
3. Use the Event tab to set parameters that dene trigger events. The parameters vary depend and Type selections.
4. Use the Tim parameters that qualify trigger events based on their time duration.
pen the Trigger control
s in the Trigger control
ing on the trigger Source
e Qualied tab to set
5. Use the A for trigger position and holdoff settings.
6. Use the the Save on trigger function. The Save on trigger function allows you to save an acq capture when a trigger event occurs.
dvanced tab to set parameters
Actions tab to set parameters for
uisition data le and/or a screen
60 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
Quick Tip
Other ways to access the Trigger control panel:
Push the Trigger button on the front panel.
Select Trigger. from the Setup menu.
Setting Up a DPX Density Trigger
The DPX Density trigger captures transients partially obscured by other higher-amplitude or more frequent signals. If you can see it in interest in the display and to measure its current density. You set a density threshold, and the instrument triggers when the density measurement exceeds your threshold.
The quickest way to set up a DPX Density trigger is to use a mouse and right-click on a signal of interest in the DPX graph, or p This action performs several tasks:
Places the density measurement box at the click point
Sets the measurement box to its default size
Measures the average density in the box
the DPX bitmap, you can trigger on it. The DPX Density area measurement is used to select the region of
ress on the screen and hold with your nger for one second. A menu will pop up. Select Trigger On This.
Sets the Trigger mode to Triggered
Selects the DPX Density "Higher" trigger type
Adjusts the trigger threshold to a density value slightly below the current measured value
Starts the acquisition process
To optimize DPX Density trigger settings for your particular signals, open the Trigger control panel. Use the controls for changing the polarity of the trigger's threshold comparison (Higher versus Lower), the Threshold value, and the size and
on of the measurement box. Use the Frequency and Amplitude settings to locate the center of the box, and set the +/-
locati values to adjust the width and height of the measurement box.
For events that are partially masked by stronger signals, you will often need to adjust the density Threshold based on density measurements of the event of interest and of the obscuring signal. For example, if the density of the background
representing only the stronger signal is 7% and the density measurement including both signals is 9%, you would
color set the Threshold to 8%.
SPECMON3 & SPECMON6 Quick Start User Manual 61
Operation
If your expected signal is a continuous tone and you want to capture events when it hops to a different frequency or drops to a lower amplitude, move the Density measurement box to surround the peak of the signal, select Lower in the Density control, and set Threshold to 100%. The instrument will trigger when the signal is absent from the measurement box.
Hint. The density measurement is the average of the densities for each column of pixels in the box. If your measurement
box is wider than the peak of a CW signal, the density reading will be less than 100%. To get a 100% reading, you can broaden signal should enter the left side of the box and exit the right side. It should not cross the bottom edge of the box.
the peak by increasing RBW or narrow the box until there are no colored pixels either above or below the box. The
Creating a Frequency Mask
Use the Mask Editor to create a Frequency Mask. The mask is a set of points dened by frequency and amplitude. Access the Mask Editor from the Trigger control panel. (This procedure uses the Spectrum display.)
You can create a Frequency Mask by using either the manual or Auto Draw method. The following procedure explains the manual method. For a procedure on how to use the Auto Draw method, see the topic Mask E ditor (Frequency Mask Trigger) in the online help.
1. Click Trig to open the Trigger control panel.
62 SPECMON3 & SPECMON6 Quick Start User Manual
2. Select the Event tab in the control panel.
3. Set the Source to RF Input.
4. SettheTypetoFrequency Mask.
5. Click Mask Editor to open the Mask
Editor.
The Mask Edit
or opens with the traces of
the Spectrum display in the graph.
6. Click New in the tool bar. This removes
the existing mask and then creates a default 6-point mask.
Operation
SPECMON3 & SPECMON6 Quick Start User Manual 63
Operation
7. Click any point in the mask to select it.
The frequency and amplitude of the selected poin graph.
8. Edit the point by doing one of the following:
Drag and drop the point with the
mouse.
t are highlighted in the
Enter the fr
the table.
Adjust amplitude with the up/down
arrow keys and frequency with the left/righ
equency and amplitude in
tarrowkeys.
Quick Tips
Click Auto draw to generate a mask automatically adjusted to the Reference.
Right click on the display screen to add a new point.
Click Add Points and then click on the graph to add a new point in the graph.
Click Next to select the next point in the graph.
Click the X (delete button) to delete the selected mask point.
Click Save to save the mask le and continue working with the mask editor.
Click Open to load and edit a previously saved mask le.
Dening Trigger Condition s
After you have set up the mask, you can select whether to trigger on the presence or absence of mask violations.
1. Select a condition.
hoices are described in the
The c following table. (See Table 7.) The icons at the bottom of the control panel identify
t is considered true or false.
wha
ck Triggered.
2. Cli
The instrument will trigger when the trigger conditions are met.
64 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
Table 7: Trigger violations
Violation Description
T
F The instrument triggers whenever the signal has no points in the mask.
T>F
F>T
T>F>T
F>T>F
The instrument triggers whenever the signal violates the mask (above the line).
Onetransitiondefines the trigger event. The signal must go out of the mask (no violation) after being inside of the mask (violation).
Onetransitiondefines the trigger event. T he signal must go into the mask after being outside of t
Two transi outside of the mask, and then cross back into the mask.
Two transitions dene the trigger event. The signal must start outside of the mask ( no violation), cross into the mask, and then cross back outside of the m ask.
he mask.
tions dene the trigger event. The signal must start inside of the mask, cross
Controlling Acquisitions
When two or more displays are open, one display controls the acquisition system. The acquisition requirements for the selected display set the acquisition hardware parameters. While the acquisition parameters are optimized for the selected display, they might not be ideal for other displays. In this case, the other displays might be unable to maintain optimum performance and consequently might be unable to produce results. (This also happens when you analyze recalled data: measurements attempt to deliver results, but may not be able to comply with all their settings.)
You can specify which display controls the hardware acquisition parameters by clicking anywhere in that display. When there are multiple displays, the highlighted display controls the hardware acquisition parameters.
In this example, there are two displays, a Spectrum display and a DPX Spectrum display. The highlighted title bar on the Spectrum display indicates that it is the selected display.
The check mark indicator in the upper left corner of the display indicates that acquisition hardware is optimized for the Spectrum display.
Some displays do not control the acquisition hardware. These displays just show the captured data. Some displays share the same parameters. If any of one of these displays in such a family controls the acquisition hardware, then the acquisition
rameters are optimized for all of the displays in that family.
pa
SPECMON3 & SPECMON6 Quick Start User Manual 65
Operation
Measurements with Conictin g Acquisition Requirements
The analyzers let you open any combination of measurements. You can select the displays that you are interested in and the instrument will automatically set up the appropriate analysis and acquisition parameters.
If measurements place conicting requirements on the acquisition parameters, the current selected display takes priority. The following example shows incompatible measurement settings when one display needs real-time data (such as the Amplitude vs Time measurement) while another display requires swept data (such as a Spectrum graph with a very wide span).
1. Push the Preset button on the front panel to set the instrument to the default settings.
2. Tune the instrument to the frequency of your signal and then set the span.
3. Click the Displays icon to open the Select Displays dialog box.
4. Select t
5. Select the Amplitude vs Time icon.
6. Click Add to add the icon to the Selected
NOTE. A
already in the Selected Displays list.
7. Click OK to accept the changes and
he General Signal Viewing
folder.
Displays list.
dd the Spectrum display if it is not
the dialog box.
close
66 SPECMON3 & SPECMON6 Quick Start User Manual
8. Increase the span in the Spectrum
display to 200 MHz.
Because Spect display, the acquisition parameters adjust automatically to meet the needs of the displa exceeds the real-time bandwidth of the instrument. As a result, the instrument will change t
rum is the selected
y. The 200 MHz span
o swept mode.
Operation
9. Look at the S
“Swept” indicates when acquisitions are swept.
When the ac all measurements usually run. If the acquisitions are swept, only the selected display ca
10. Click the Amplitude vs Time display to
make it t
The acquisition switches back to real-time because the selected display require suitable for the Amplitude vs Time display but doesn't contain the full range of freq display.
In this case, the Spectrum is unable to run.
tatus bar. The readout
quisitions are in “Real Time,”
nrun.
he active display.
s real-time. The data is now
uencies needed by the Spectrum
SPECMON3 & SPECMON6 Quick Start User Manual 67
Operation
11. Decrease the sp
an to 25 MHz in the Spectrum display. Now the acquisitions are real time again.
12. Check the status bar readout to verify that the acqui
sition is real-time.
Both of the displays can run now.
68 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
Searching Signals
The analyzers let you search the Spectrum display's trace results for changes in power level. You can identify matches crossing a power level or you can create a mask that species power level and frequency range limits. You can select which trace to search and specify an action to take when a search match occurs. Pass/Fail results from the Spurious and Settling Time measurements are also searchable, but we will use the Spectrum trace example here.
To search a trace:
1. Open the Spectrum display or select the
Spectrum display if it is already open.
2. Select Too ls > Search.
3. Select the Enable Search check box.
4. Select the trace you wish to search from
the Search in drop-down list box.
5. Verify that the trace you selected in the
Search in list is the selected trace and
that the Show box is checked.
SPECMON3 & SPECMON6 Quick Start User Manual 69
Operation
6. Specify what to search for by making a selection from the Search for drop-down list.
To search for signal levels above or below a specied level, select either is
greater than
or is less than.
To search for signal levels that vary with frequency, select either is outside mask or is inside m
ask. To specify the limits in the search mask, click the Edit limits button.
The orange s
haded area in the graph display shows the levels which, if occupied by the signal, will constitute a match.
Display with greater than search selected
y with outside mask search selected
Displa
70 SPECMON3 & SPECMON6 Quick Start User Manual
7. Click the Actions tab to specify what
actions the analyzer should take when a match occurs.
The analyzer can take up to ve actions when a match occurs, depending on the run mode. In R can sound a beep, stop acquisitions, save acquisition data, save trace data, and save a pic Replay mode, it can sound a beep and stop analyzing.
8. To have the analyzer take an action
when a matc action.
You can select any combination of actions to select all actions, they will all occur when a match is found (in Run mode).
9. Click the close box to remove the Search
control p
un mode, the analyzer
ture of the display. In
h occurs, select the desired
perform on a match. If you
anel.
Operation
10. Press the Run button to initiate a search.
Red line when there is a match on incoming data. Click the Replay button on the toolbar to initi record.
s appear in the graph display
ate a search on the current data
SPECMON3 & SPECMON6 Quick Start User Manual 71
Operation
Using Correction Tables
The analyzers let you correct the signal displays and measurement results to account for gains and losses in external equipment. You can specify a single gain value or you can use an external loss table to adjust the gain/loss value based on frequency.
To specify external gain/loss corrections:
1. Press the Amplitude icon button to display the Amplitude control panel.
2. Select the External Gain/Loss Correction tab.
3. To apply a constant correction factor,
select the External Gain value check
ter a value into the text entry box.
box. En
NOTE. You can enter either positive or
negative gain values. Positive values
ent a gain. Negative values represent
repres aloss.
4. To u s e
5. To ed
a table that corrects for gain/loss values depending on frequency, click a check box under External Loss Tables.
an enable all three external loss
You c tables at the same time. Thus, you could use different external loss tables for an
nna, a preamplier, and for cabling.
ante
it the contents of an external loss table, click the Edit button for the table youwanttoedit.
72 SPECMON3 & SPECMON6 Quick Start User Manual
6. Edit the contents of the table to match
the loss characteristics of the equipment you are correc you are nished.
NOTE. You can enter either positive or
negative val to the External Gain value setting, positive values represent a loss and negative values represent a g
For detailed information on using the Table Editor, select Help > User Manual and look in the Inde
7. You can change the title for each external
loss table. Select the title box and type in a new title.
Note that the title is only a label. It is not tied to the le name of any table you have loaded or saved.
ting for. Click OK when
ues into the table. But in contrast
ain.
xforExternal Loss Tables.
Operation
8. Specify whether table corrections are
applied to all acquired data or only to traces in the Spectrum, Spectrogram, Spurious and Amplitude vs Time displays.
Corrections applied to data are performed in real time using the digital lter capabilities of the instrument. While this has the advantage of correcting the data for use in all measurements, digital ltering is limited in its ability to apply corrections that change rapidly in amplitude over small frequency ranges, as may occur when switching between two antennas. When this happens, the lters may result in an amplitude correction that appears to ring in its response, resulting in poorly corrected data. For that reason, the instrument offers trace-only correction for spectrograms, spectrum, spurious, and amplitude vs. time displays. These trace corrections are similar to those applied in traditional spectrum analyzers, and should be used when discontinuities occur in the amplitude/frequency correction table.
SPECMON3 & SPECMON6 Quick Start User Manual 73
Operation
When choosing w sample data or traces, keep the following in mind:
If sample data is acquired while data correcti record is permanently corrected. If the acquisition is saved and recalled, the recalled acq Corrections are not saved separately from the raw data; they are used during the acquisi values.
Trace corrections can be applied at any time, as they are part of the measureme Replay will apply trace corrections to a trace if this control is enabled (and an externa enabled).
Be aware that if you recall acquisition data that was saved with data corrections and you us corrections selected and an external loss table enabled, the trace will be
ted using corrected data and also
calcula corrected with trace corrections. This usually creates incorrect traces.
hether to correct
on is enabled, that data
uisition includes corrections.
tion process to adjust the data
nt's computations. Using
l loss table is specied and
e Replay with trace
9. Click the close box to remove the ude control panel when you are
Amplit nished making changes.
74 SPECMON3 & SPECMON6 Quick Start User Manual
Operation
Using Audio Demodulation
The analyzers can demodulate audio signals, which can help you to identify unknown radio signals. You access the Audio Demod tab from the Settings control panel of the DPX Spectrum plot display.
To demodulate an audio signal:
1. Press the Displays button to display the Select Displays dialog box.
2. Select the General Signal Viewing folder.
3. Double-click the DPX Spectrum icon to add it to the Selected Displays list.
4. Click OK to close the dialog box. The DPX display will open.
5. If necessary, select the Spectrum plot from the drop-down list.
6. Press the Settings front-panel button to display the DPX Settings control panel.
7. Click the Audio Demod tab to display the Audio Demod control panel.
SPECMON3 & SPECMON6 Quick Start User Manual 75
Operation
8. Click the On button to enable audio demodulation.
9. To specify the demodulation type, click either the AM o
NOTE. Note that the AM and FM buttons
select only the demodulation type; they do not specify a
10. S et the center frequency as needed.
The receiver frequency (frequency to be demodulated) is specied with either markers or front panel Frequency button or the Freq control in the toolbar). Use the Tune with setting t frequency control is used to specify the receiver frequency.
11. Select either a marker or frequency control list.
r FM button.
frequency band.
the frequency control (the
o specify whether markers or the
from the Tune wit h drop-down
12. If the marker you select is not enabled (it will be visible in the display if it is
d), click the Markers icon Markers
enable button in the toolbar to display the Markers toolbar.
13. Click the Dene button in the Marker toolbar to display the Dene Markers
trol panel.
con
ck Add to add markers as needed.
14. Cli
76 SPECMON3 & SPECMON6 Quick Start User Manual
15. Click the Settings button to return to the
Audio Demod control panel.
16. Set the required receiver bandwidth by
entering a value into the Receiver BW value entry box.
Operation
SPECMON3 & SPECMON6 Quick Start User Manual 77
Operation
17. S e t the receiver frequency.
If you speciedamarkerwiththeTune with setting, required frequency. Fine tune the marker position with the front-panel knob or your mouse wh Frequency control in the Tune with setting, press the front-panel FREQ button, then use the Freq setting in the toolbar to set the required frequency.
The receive Receiver Freq between the Tune with and Receiver BW settings.
18. Adjust the Audio Gain setting so you can hear the de the sound level is also affected by the setting of the Windows volume control.
drag the marker to the
eel. If you selected
turn the front-panel knob or
r frequency is displayed as
modulated signal. Note that
78 SPECMON3 & SPECMON6 Quick Start User Manual
Index
Index
A
Accessories
optional, 3 power cords, 2 Product Docu standard, 1
Acquisition
conicts, 6 controlling, 65 data source, 24 starting,
stopping, 32 Analysis window, 23, 25 Applicati audio demodulation, 75
on presets, 27
mentation, 1
6
32
C
CF Pan
Touchscr check mark indicator, 21, 65 Cleaning procedures, 13 Conict
measurement, 66 Connectors
front p
rear panel, 19
RF signal input, 31, 32 Contro Controls
display-specic, 22
front
interface, 20 Cooling clearances, 4
ction tables, 72
corre
een actions menu, 37
s
anel, 18
lling acquisitions, 65
panel, 14
D
Data
recalling, 42
ing, 41
sav Default settings, 30 Diagnostics, 12
ital demodulation displays, 25
Dig Display
conicts, 66
splay properties
Di
changing, 11
Display settings
changing, 29
Displays
general signal viewing, 25 GP digital modulation, 25 printing, 39 Pulsed RF, 25 RF measurement, 25 selecting,
Documentation, x
Tektronix part numbers, 1
DPX Densit
triggering, 58
26
y
E
Environmental specications, 4
,4
altitude cooling clearances, 4 humidity, 4
tempera Extended desktop, 8 external gain/loss correction, 72 Externa
ture, 4
l monitor, 8
F
Features, ix File types, 41
un trigger mode, 58
Free R Frequency edge
triggering, 58
ency mask
Frequ
Mask Editor, 62
triggering, 58
t panel
Fron
connectors, 18
controls, 14
G
eral signal viewing displays, 25
Gen Graphics driver, 11
I
iMap
unching, 43
la Incoming inspection, 12
Information messages, 28 Ink saver mode, 40 Input requir
Installati
Italicized numbers, 22
ements frequency range, 31 input impedance, 31 power, 31 RF signal range, 31 voltage, 31
on, 1
network connections, 5
K
Key featur
es, i x
M
Maintenance
cleaning procedures, 13
ging, 13
repacka
Marker toolbar, 36 Markers, 34
e, 35
absolut adding, 36 dening, 37
35
delta, DPX bitmap trace, 35 shapes, 35
ditor
Mask E
adding points, 62 editing points, 64
rement
Measu
conicts, 66
Mechanical specications, 4
ages
Mess
on-screen, 28
MR, 35
N
work connections, 5
Net
O
Operating system restore, 7 Optional accessories, 3
tions
Op
Power cords, 2
SPECMON3 & SPECMON6 Quick Start User Manual 79
Index
P
Pan
Touchscreen actions menu, 37 Part numbers, Power cord options, 2 Power level triggering, 59 Power supply Preset, 30
selecting displays, 27
system setti Printing, 39
ink saver mode, 40 Product fea Pulsed RF displays, 25
manuals, 1
,4
ngs, 30
tures, ix
R
Real-Time spectrum analyzer, 23 Rear panel Recalling data, 42 Repackaging, 13 Reset sca
Returning the instrument, 13 RF measur RF signal connections, 31, 32 RF signal input range, 31 Run cont Run menu, 33 Runt triggering, 59
connectors, 19
le
Touchscreen actions menu, 37
ement displays, 25
rol, 33
S
Summary, iii
Safety Save options, 41 Saving data, 41
ns, printing, 39
Scree Search, 69
Select
Touchscreen ac Selected window, 65 Selecting displays, 26 Settings
changing, 29
default, 30
preset, 30 Setups
recalling, 42
saving, 41 Software upgrades, x
installing, 13 Span Zoom
Touchscreen actions menu, 37 Specications
altitude,
cooling clearances, 4
environmental, 4
humidity,
input requirements, 31
mechanical, 4
power sup
product dimensions, 4
temperature, 4
m analyzer
Spectru
real-time concepts, 23 Standard accessories, 1
power co
Product Documentation, 1 Starting acquisitions, 32
ng acquisitions, 32
Stoppi
tions menu, 37
4
4
ply, 4
rd, 2
T
Time Qualied
triggering, 58
screen actions menu, 37
touch
Trace legend, 29 Trigger
control panel, 60 mask, 62 violations, 6
Trigger Event parameters, 59
accessing, 60 Triggered tr Triggering
DPX Density, 58
free run, 58
frequency edge, 58
frequency mask, 58
parameters
power level, 59
Runt, 59
time quali
triggered mode, 58
5
igger mode, 58
,59
ed, 58
U
Upgrades
software
,x,13
V
Violations
trigger mask, 65
W
Window
analysis, 23
selected, 65
Z
Zoom
Touchscreen actions menu, 37
80 SPECMON3 & SPECMON6 Quick Start User Manual
Loading...