Tektronix Series 2200 Multichannel Programmable DC Power Supplies Specifications,Performance Verification Reference manual

www.keithley.com
Series 2200
Multichannel Programmable DC Power Supplies
Specifications and Performance Verification
Technical Reference
2220S-905-01 Rev. B
/ Dec 2013
ECNEDIFNOCFOERUSAEMRETAERGA
Series 2200
Specifications and Performance Verification
Technical Reference
© 2013, Keithley Instruments, Inc.
Cleveland, Ohio, U.S.A.
All rights reserved.
Any unauthorized reproduction, photocopy, or use the information herein, in whole or in part,
without the prior written approval of Keithley Instruments, Inc. is strictly prohibited.
All Keithley Instruments product names are trademarks or registered trademarks of Keithley
Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective
holders.
Document number: 2220S-905-01 Rev. B / Dec 2013
Safety precautions
The following safety precautions should be observed before using this product and an y associated instrumentation. Although some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all install a tion, operation, and maintenance information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are measurement, control, and data I/O connections, with low transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high transient overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient overvoltages often associated with local AC mains connections. Certain Keithley measuring inst ruments may be connected to mains. These instruments will be marked as category II or higher.
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains. Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground)
connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation.
If a
The user documentation in all cases where the symbol is marked on the instrument.
The contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables. To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If yo u are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
symbol on an instrument means caution, risk of danger. The user must refer to the operating instructions located in the
symbol on an instrument means caution, risk of electric shock. Use standard safety precautions to avoid personal
symbol indicates a connection terminal to the equipment frame.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do n ot app ly cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
Safety precaution revision of January 2013.
ii 2220S-900-01 Rev. C/December 2013
Table of Contents
Preface .............................................................................................................. iii
Welcome ........................... ................................ .................................. .......... iii
Products.............................. .................................. ................................ ........ iii
Extended Wa
Contact Information .......................................................................................... iv
Specications .............................. ................................ .................................. ....... 1
Performance Verication ................. ................................ .................................. ...... 13
Test Record ................ ................................ .................................. .................. 15
Performance Verication Procedures ....................... ................................ ................ 23
rranty ......................... .................................. ................................ iv
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference i
Table of Contents
List of Tables
Table 1: Series 2220 Dual Channel specications.............................................................. 1
Table 2: Ser
Table 3: Series 2230 Triple Channel specications for Channel 3........................... ................. 5
Table 4: Combined channel characteristics............................... ................................ ....... 7
Table 5: Series 2230 Triple Channel unit characteristics....................................................... 7
Table 6: Mains power characteristics ................. ................................ ........................... 8
Table 7: Common characteristics................................................................................. 8
Table 8: T
Table 9: Interfaces and output ports ......... ................................ ................................ ..... 9
Table 10: Mechanical characteristics............................................................................ 10
Table 11: Environment performance .... ................................ .................................. ...... 10
Table 12: Keypad special operations................ ................................ ............................ 10
Table 13: Safety characteristics.................................................................................. 11
Table
family .......................... ................................ .................................. .............. 12
Table 15: DC voltage accuracy with remote sense .................... ................................ ........ 16
Table 16: DC voltage accuracy without remote sense......................................................... 17
Table 17: DC voltage readback accuracy ....................................................................... 18
Table 18: DC voltage line regulation............................................................................ 19
le 19: DC voltage load regulation ......... .................................. ................................ 19
Tab
Table 20: DC current accuracy................................................................................... 20
Table 21: DC current readback accuracy ....................................................................... 21
Table 22: DC current line regulation ............................................................................ 22
Table 23: DC current load regulation ..... .................................. ................................ .... 22
Table 24: Voltage noise at 7 MHz ....... ................................ ................................ ........ 22
Table 25: Voltage noise at 20 MHz.............................. .................................. .............. 22
Table 26: Current noise at 20 MHz.......... ................................ ................................ .... 22
Table 27: Test equipment ........... ................................ ................................ .............. 24
ies 2230 Triple Channel specications for Channels 1 and 2 .............. ..................... 3
erminal strip ................................ ................................ ............................. 9
14: Electromagnetic Compatibility (EMC) — Measurement, laboratory, and control product
ii Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
Preface
Welcome
Preface
Products
Thank you fo Multichannel Programmable DC Power Supplies are exible DC sources designed to power a wide range of applications. The model 2230-30-1 and its variants offer three power channels and the model 2220-30-1 and its variants provide two channels. The output channels on both models are independent and isolated, allowing you to power circuits with different references or polarities. Each channel c feature remote sense capability which can be used to reduce the effect of lead resistance, delivering 0.03% basic voltage accuracy even when using long leads. Basic current ac curacy is 0.1% for all channels and linear regulation delivers low noise – less than 3 mVp-p. Flexible display modes make it easy to use the two 30 V outputs in combination, and the USB interface makes it easy to build PC-bas model include a GPIB interface in combination with the USB interface.
compact power supplies cover a wide range of applications without
These covering a lot of bench space. Versions of these power supplies are available for use at 100 VAC nominal line voltage which is common in Japan. These versions are indicated by the "J" sufx.
This manual contains information about the following products:
r using a Keithley Instruments product. The Series 2200
an be enabled or disabled as your application requires. All outputs
ed systems without converters or special cables. The G vers ions of each
Model Description
2220-30-1 Programmable Dual Channel DC Power Supply
2220G-30-1 Programmable Dual Channel DC Power Supply with GPIB Interface
2220J-30-1 Programmable Dual Channel DC Power Supply for Japan
2220GJ-30-1 Programmable Dual Channel DC Power Supply with GPIB Interface for Japan
2230-30-1 Programmable
2230G-30-1 Programmable Triple Channel DC Power Supply with GPIB Interface
2230J-30-1 Triple Channel Programmable DC Power Supply for Japan
2230GJ-30-1 Programmable Triple Channel DC Power Supply with GPIB Interface for Japan
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference iii
Triple Channel DC Power Supply
Preface
Extended Warranty
Contact Information
Additional years of warranty coverage are available on many products. These valuable contracts protect you from unbudgeted service expenses and provide additional years of protection at a fraction of the price of a repair. Extended warranties are available on new and existing products. Contact your local Keithley Instrument
s representative for details.
If you have following sources:
1. Keithley
2. Keithley web forum (http://forum.keithley.com)
3. Call Keithley Instruments corporate headquarters (toll-free inside the U.S. and
Canada only) at 1-888-KEITHLEY (1-888-534-8453), or from outside the U.S. at Instruments website (http://www.keithley.com).
any questions after reviewing this information, please use the
Instruments website (http://www.keithley.com)
+1-440-248-0400. For worldwide contact numbers, visit the Keithley
iv Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
Specications
This section contains specications for the Series 2200 Multichannel Programmable DC Power Supplies. All specications are guaranteed unless noted as "typical. not guaranteed. Specications that are marked with the in Performance Verication.
Table 1: Series 2220 Dual Channel specications
Parameter Description
Channel 1 and 2, nominal
Constant voltage operation, nominal
Voltage range
This is the range over which the output voltage is adjustable
Remote sense voltage range
Voltage resolution, nominal
age setting
Volt accuracy with remote sense
Voltage setting accuracy without remote sense
Current limit, nominal
Voltage temperature coefcient, typical
Constant current operation, nominal
Current range
This is the range over which the output current is adjustable
Channels 1 and 2 are symmetrical and isolated. They are primarily targeted at powering analog circuits operating around ground.
The unit may be set to a constant voltage over a range of currents
0to30V
The total voltage from the – terminal to the + terminal may not exceed the rated full scale (FS) voltage + 0.2 V, allowing a minimum of 0.1 V/line for remote sense.
The maximum voltage difference between a terminal and its associated sense line may not exceed 1 V.
The available overhead above full scale voltage to support remote sense is limited to 0.1 V/line. However, this widens as the output voltage decreases to the ultimate limit of 1 V/line. Thus, the maximum difference specication controls operation except for the 2 V of output below full scale output.
1mV
±(0.03% of settling + 10 mV) at 25 °C ±5 °C with remote sense.
±(0.05% of settling + 10 mV) at 25 °C ±5 °C without remote sense.
Dened by the c onstant current setting. The changeover is automatic.
±(0.03% + 10 m V) per °C outside the 25 °C ±5 °C standard range
The unit may be set to a constant current over a range of voltages.
0Ato1.5A
" Typical specications are provided for your convenience but are
symbol are checked
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference 1
Specications
Table1:Series2220DualChannelspecifications (cont.)
Parameter Description
Current resolution, nominal
Current set
ting
accuracy
Voltage li
mit,
nominal
Current temperature
nt, typical
coefcie
Meter, nominal Internal readback capability. Voltage and current can be measured internally.
Current readback
ion, nominal
resolut
Current
readback accuracy
Current readback temperature coefcient, typical
Voltage
readback accuracy
Voltage readback resolution, nominal
Voltage readback temperature
cient, typical
coef
Line regulation
– voltage
Line regulation
rent
–cur
Load regulation
– voltage
Load regulation
rrent
–cu
Voltage ripple,
20 MHz bandwidth
Voltage ripple,
Hz bandwidth
7M
Current ripple,
20 MHz bandwidth
Common Mode Current Noise
1mA
±(0.1% + 5 mA) at 25 °C ±5 °C
0to30V
±(0.1% + 5 mA) per °C outside the 25 °C ±5 °C standard range
1mA
of reading + 5 mA) at 25 °C ±5 °C
±(0.1%
±(0.1% + 5 mA) per °C outside of the 25 °C ±5 °C standard range
±(0.03% of reading + 10 mV) at 25 °C ±5 °C standard range
1mV
±(0.03% of reading + 10 mV) per °C outside the 25 °C ±5 °C standard range
ge over the full rated line voltage range: ±(0.01% + 3 mV)
Chan
Change over full rated line voltage range: ±(0.1% + 3 mA)
nge from 0 to 100% of full scale current: ±(0.01% + 3 mV)
Cha
Change from 10 to 90% of full scale voltage: ±(0.01% + 3 mA)
<3mV <1.
<1mV
<5mA
<1.5μA
and 20 mV
RMS
5mV
and 12 mV
RMS
and 3 mV
RMS
RMS
, 700 Hz bandwidth.
RMS
p-p
p-p
,typical
p-p
2 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
Table 1: Series 2220 Dual Channel specications (cont.)
Parameter Description
Voltage transient response settling time, load change, typical
Voltage transient response settling time, setting change, rising, typical
Voltage transient response settling time, setting change, falling, typical
Unit SpecicMains Requirements
<150 ms to within 75 mV following a change from 0.1 A to 1 A.
<150 ms from beginning of excursion to within 75 mV of terminal value following a change from 1 V to 11 V intoa10resistor.
<150 ms from beginning of excursion to within 75 mV of terminal value following a change from 11 V to 1Vintoa10resistor.
2220-30-1 and 2220G-30–1
Line Selection Switch
110V
220V
Voltage Frequency Fuse Rating Max VA +
110/115/120 50/60 Hz
220/230/240 50/60 Hz
Specications
6.3A TH 250V 350
3.15A TH 250V 350
2220J-30-1 and 2220GJ-30–1
Voltage Frequency Fuse Rating Max VA +
50/60 Hz
50/60 Hz
6.3A TH 250V 350
3.15A TH 250V 350
Maximum Input VA, typical
Line Selection Switch
110V 100
220V 200
350 VA
Table2:Series2230TripleChannelspecifications for Channels 1 and 2
Parameter Description
Channel 1 and 2, nominal Channels 1 and 2 are symmetrical and isolated. They are primarily targeted at powering analog
circuits operating around ground.
Constant Voltage Operation, nominal
Voltage Range
The unit may be set to be a constant voltage over a range of currents.
The output voltage may be adjusted from 0V to 30V.
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference 3
Specications
Table2:Series2230TripleChannelspecifications for Channels 1 and 2 (cont.)
Parameter Description
Remote Sense Voltage Range The total voltage from - terminal to + terminal may not exceed the rated full scale voltage +0.2V,
allowing a minimum of 0.1V/line for remote sense.
The Maximum voltage difference between a terminal and its associated sense line m ay not exceed 1V.
The available overhead above full scale voltage to support remote sense is limited to 0.1 V/line. However, this widens as the output voltage decreases to the ultimate l imit of 1 V/line. Thus, the maximum difference specication controls operation except for the 2 V of output below full scale output.
Voltage Resolution, nominal 1 mV Voltage Setting Accuracy with
Remote Sense VoltageSetting Accuracy without
Remote Sense
Current Limit, nominal Dened by Constant Current setting. Changeover is automatic.
Voltage Temperature Coefcient, typical
Constant Current Operation, nominal
Current Range The output voltage may be adjusted from 0 A to 1.5 A.
Current Resolution, nominal Current Setting Accuracy
Voltage Limit, nominal 0 to 30 V.
Current Temperature Coefcient, typical
Meter, nominal Internal readback capability. Voltage and current are measured internally.
Current Readback Resolution, nominal
urrent Readback Accuracy
C
Current Readback Temperature Coefcient, typical
Voltage Readback Accuracy
Voltage Readback Resolution, nominal
Voltage Readback Temperature Coefcient, typical
Line regulation – Voltage
Line regulation – Current
Load regulation – Voltage
+/-(0.03% of setting + 10 mV) at 25 °C ±5 °C with Remote Sense
+/-(0.05% of setting + 10 mV) at 25 °C ±5 °C without Remote Sense
+/-(0.03% of setting + 10 mV) per °C outside of the 25 °C ±5 °C standard range
The unit may be set to be a constant current over a range of voltages.
1mA
+/-(0.1% + 5 mA) at 25 °C ±5 °C.
+/-(0.1% + 5 mA) per °C outside of the 25 °C ±5 °C standard range
1mA
±(0.01% of reading + 5 mA) at 25 °C ±5 °C
±(0.01% + 5 mA) per °C outside the 25 °C ±5 °C standard range
±(0.03% of reading + 1 0 mV) at 25 °C ±5 °C
1mV
±(0.03% of reading + 10 mV) per °C outside the 25 °C ±5 °C standard range
Change over the full rated line voltage range: ±(0.01% + 3 mV)
Change over full rated line voltage range: ±(0.1% + 3 mA)
Change from 0 to 100% of full scale. Voltage: ±(0.01% + 3 mV)
4 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
Table 2: Series 2 230 Triple Channel specications for Channels 1 and 2 (cont.)
Specications
Load regulat
ion – Current
Change from 10 to 90% of full scale. Current: ±(0.01% + 3 mA)
While not readily testable, similar performance is expected over the 0% to 100% of full scale range of the product.
Voltage Ripple, 20 MHz bandwidth
h
<3mV <1.5mV
Voltage Ripple, 7 MHz bandwidth < 1 mV
Current Ripple, 20 MHz
<5mA
and 20 mV
RMS
RMS
and3mV
RMS
RMS
and 12 mV
p-p
p-p
p-p
,typical
bandwidth
Common Mode Current Noise
Voltage Transient Response
<1.5μA
<150 ms to within 75 mV following a change from 0.1 A to 1 A.
. 700 Hz bandwidth.
RMS
Settling Time, Load Change, typical
Voltage Transient Response Settling Time, Setting Change,
<150 ms from beginning of excursion to within 75 mV of terminal value following a change from1Vto11Vintoa10resistor.
Rising, typical
Voltage Transient Response Settling time, Setting Change,
<150 ms from beginning of excursion to within 75 mV of terminal value following a change from11Vto1Vintoa10resistor.
Falling, typical
Table 3: Series 2230 Triple Channel specications for Channel 3
Parameter Description
Channel 3, nominal
Constant Voltage Operation,
nal
nomi
Voltage Range
ote Sense Voltage Range
Rem
ltage Resolution, nominal
Vo
Voltage Setting Accuracy with
Remote Sense VoltageSetting Accuracy without
Remote Sense
Current Limit, nominal Dened by Constant Current setting. Chageover is automatic.
Voltage Temperature Coefcient, typical
A third channel is provided. It is primarily targeted at logic power.
The unit may be set to be a constant voltage over a range of currents.
utput voltage may be adjusted from 0 V to 6 V.
The o
total voltage from - terminal to + terminal may not exceed the rated full scale voltage +0.2V,
The allowing a minimum of 0.1V/line for remote sense.
The maximum voltage difference between a terminal and its associated sense line may not
eed 1V.
exc
The available overhead above full scale voltage to support remote sense is limited to 0.1 V/line. However, this widens as the output voltage decreases to the ultimate limit of 1 V/line.
us, the maximum difference specication controls operations except for the 2 V of output
Th below full scale output.
V
1m
/-(0.03% of setting + 10 mV) at 25 °C ±5 °C with Remote Sense
+
+/-(0.05% of setting + 10 mV) at 25 °C ±5 °C without Remote Sense
+/-(0.03% of setting + 10 mV) per °C outside of the 25 °C ± 5 °C standard range
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference 5
Specications
Table 3: Series 2230 Triple Channel specications for Channel 3 (cont.)
Constant Current Operation,
The unit may be set to be a constant current over a range of voltages.
nominal
Current Range The output voltage m ay be adjusted from 0 A to 5 A.
Current Resolution, nominal Current Setting Accuracy
1mA
+/-(0.1% + 5 mA) at 25 °C ±5 °C.
Voltage Limit, nominal 0 to 6 V.
Current Temperature Coefcient,
+/-(0.1% + 5 mA) per °C outside of the 25 °C ±5 °C standard range
typical
Meter, nominal Internal readback capability. Voltage and current can be measured internally. DVM connections
exist to allow measuring external voltages.
Current Readback Resolution,
1mA
nominal
Current Readback Accuracy
Current Readback Temperature
cient, typical
Coef
Voltage Readback Accuracy
Voltage Readback Resolution,
±(0.1% of reading + 5 mA) at 25 °C ±5 °C
±(0.1% +5mA) per °C outside of the 25 °C ± 5 °C standard range
3% of reading + 1 0 mV) at 25 °C ±5 °C
±(0.0
0.1 mV
nominal
Voltage Readback Temperature
±(0.03% of reading + 10 mV) per °C outside of the 25 °C ±5 °C standard range
Coefcient, typical
Line Regulation – Voltage
Line Regulation – Current
Load Regulation – Voltage
Load Regulation – Current
ge Ripple, 20 MHz
Volta
Bandwidthh
Voltage Ripple, 7 MHz
Change over the full rated line voltage range: ±(0.01% + 3 mV)
Change over full rated line voltage range: ±(0.1% + 3 mA)
Change from 0 to 100% of full scale. Voltage: ±(0.01% + 3 mV)
Change from 10 to 90% of full scale. C urrent: ±(0.01% + 3 mA)
<3mV <1.5mV
<1mV
and 20 mV
RMS
RMS
and3mV
RMS
and 12 mV
p-p
.
p-p
p-p
,typical
bandwidth
nt Ripple, 20 MHz
Curre
Less than 6 mA
RMS
Bandwidth
μA
Common Mode Current Noise
age Transient Response
Volt
<1.5
<150 ms to within 75 mV following a change from 0.1 A to 1 A.
. 700 Hz bandwidth.
RMS
Settling Time, Load Change, typical
Voltage Transient Response Settling Time, Setting Change,
ing, typical
Ris
Voltage Transient Response
ttling Time, Setting Change,
Se Falling, typical
<150 ms from beginning of excursion to within 75 mV of terminal value following a change
m0.4Vto4Vintoa4Ω resistor.
fro
50 ms from beginning of excursion to within 75 mV of terminal value following a change
<1 from0.4Vto4Vintoa4Ω resistor.
The specication does not include Command Decode Time.
6 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
Specications
Table 4: Combin
Characteristic Description
Combined Channel Characteristics, nominal
Series Combinations, nominal Deliver up to 60 V when CH1 and CH2 are wired in series. Voltage Readback is the combined
Available Series Congurations, nominal
Series Co Accuracy, typical
Series Combination Current Accuracy, typical
Paralle
Available Parallel Congurations, nominal
Parallel Combination Voltage Accur
Para Accuracy, typical
mbination Voltage
l Combination, nominal
acy (typical)
llel Combination Current
ed channel characteristics
The unit can be congured to provide functionality on Channel 1 and Channel 2 together.
voltage. Se only. If the currents in the two channels are signicantly different there is no indication of CC/CV status for Channel 2.
Channels 1 and 2 are connected in series. Either channel 1 or channel 2 m ay be the most positive s
±(0.10% o
±(0.1% + 5 mA) at 25 °C ±5 °C for Channel 1's current.
The current is only measured for Channel 1. The error is the same as for the Current Measure
Deliver combined current.
CH1 and CH2 are connected in parallel (+ to + and – to –)
±(0.05% of setting + 10 mV) at 25 °C ±5 °C
±(0.
ment Accuracy.
s up to 3 A when CH1 and CH2 are wired in parallel. The meter reads back the
2% +10 mA) at 25 °C ±5 °C
ttings changes are to the combined voltage. Current Readback is for Channel 1
ource.
f the setting + 50 mV) at 25 °C ±5 °C
Table 5: Series 2230 Triple Channel unit characteristics
Characteristic Description
Unit Characteristics, nominal Characteristics and specications of features unique to this product, but unrelated to a
particular channel.
Unit SpecicMains Requirements
Maximum Input VA, typical 450 VA
2230-30-1 and 2230G-30–1
Line Selection Switch
110V
220V
2230J-30-1 and 2230GJ-30–1
Line Selection Switch
110V 100
220V 200
Voltage Frequency Fuse Rating Max VA
110/115/120 50/60 Hz
220/230/240 50/60 Hz
Voltage Frequency Fuse Rating Max VA
50/60 Hz
50/60 Hz
6.3A TH 250V 450
3.15A TH 250V 450
6.3A TH 250V 450
3.15A TH 250V 450
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference 7
Specications
Table 6: Mains p
ower characteristics
Characteristic Description
Source Voltage
The 2220-30-1, 2220G-30-1, 2230-30-1, and 2230G-30-1 each have two ranges, selectable with a bottom panel switch. Each range is capable of ±10% excursions.
Marked Allowed nominal voltages
110
220
The 2220J-30-1, 2220G J-30-1, 2230J-30-1, and 2230GJ-30-1 each have two ranges, selectable with a bottom panel switch. Each range is capable of ±10% excursions.
Marked Allowed nominal voltage
110
220
Source Frequency 50/60 Hz
Fuse Ra
ting, nominal
Model
2230, 2230G and 2220, 2220G
Table 7: Common characteristics
110/115/120 VAC
220/230/240 VAC
100 VAC
200 VAC
RMS
RMS
110 VAC setting 220 VAC setting
250V
6.3A T 159-0399-00
T 250V
3.15A 159-0493-00
RMS
RMS
Item Description
Setup Memory, nominal
30 setup memory locations
Output Timer, nominal The product includes a timer that can be used to disable the output after a predetermined
period of time.
Timer Set Time, nominal
The Timer may be set to any value between 1 and 999,999 seconds.
Timer Accuracy ±100 ppm
Software Voltage Limit, nominal The product offers the opportunity to change the maximum settable voltage on a temporary
basis. The MAX VOLT menu entry, when set below hardware maximum limit, disallows adjusting the product to voltages in excess of the MAX VOLT limit.
Cleaning Requirements, nominal Inspect the power supply as often as operating conditions require. To clear the exterior
surface, perform the following steps:
1. Remove loose dust on the outside of the power supply with a lint free cloth. Use care to avoid scratching the display.
2. Use a soft cloth dampened with water to clear the power supply. Use an aqueous solution of 75% isopropyl alcohol for more efcient cleaning.
Caution: To avoid damage to the surface of the power supply, do not use any abrasive or chemical cleaning agents.
Caution: Avoid getting moisture inside the unit during external cleaning. Use only enough cleaning solution to dampen the cloth or swab.
8 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
Specications
Table 7: Common characteristics (cont.)
Item Description
Isolation Voltage, Output to Chassis
Any output can be oated up to 240 V (DC + peak AC with AC limited to a maximum 3 V maximum of 60 Hz) relative to the Earth Ground terminal. Note that remote sense terminals should remain within 1 V of their respective outputs to avoid possible damage.
Isolation Voltage, Output to Output Any output can be oated up to 240 V (DC + pk AC with AC limited to a maximum of 3 V
and a maximum of 60 Hz) relative to any other output terminal. Remote sense terminals should remain within 1 V of their respective outputs to avoid possible damage.
Table 8: Terminal strip
Item Description
The rear panel has a customer accessible terminal strip. This strip is comprised of a board-mounted section and a technician-removeable section. The technician-removeable section offers the customer a safe set of screw terminals to connect
us instrument functions detailed below.
to vario
pk-pk
and
pk-pk
Table 9: Interfaces and output ports
Parameter (instrument model) Description
USB device, nominal
(All models)
GPIB interface
(G models only)
Command processing time, typical
(All models)
USB 2.0 Full Speed device. 12 Mb/sec maximum.
IEEE-488.2 compliant
About 400 ms per command.
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference 9
Specications
Table 10: Mecha
Parameter
Weight, instrument, nominal
Dimensions, nominal
Cooling method, nominal
Required cooling clearance
Construction materials, nominal
nical characteristics
9kg (19.8 lbs)
Model
2230
Fan
Rear of instrument must be separated by at least 5 cm (2 in) from any airow restrictions to assure adequate cooling.
Bottom of instrument must have 1 cm (0.4 in) from any airow restrictions. The protective boot maintains the proper bottom clearance.
For rackm instructions.
Chassis – Aluminum Front Panel Plastics – PC/ABS blend
ount conguration, follow the clearance requirements provided with the rackmount installation
Table 11: Environment performance
Parameter All models
Temperature
Humidity
Altitude
Operating: +0 °C to +40 °C
Nonoperating: -20 °C to 70 °C
Operating:
5% to 95% relative humidity (% RH) at up to 40 °C, noncondensing
Nonoperating:
5% to 95% RH (relative humidity) at up to +40 °C,
5% to 60% RH above +40 °C up to +70 C, noncondensing
Operating: 100% capability up to 2,000 meters.
Nonoperating: Up to 4,000 meters
Dimensions (Width * Height * Depth
214 mm * 88 mm * 355 mm (8.45” * 3.47” * 13.9”)
Table 12: Keypad special operations
aracteristic
Ch
Display the heat sink temperature Enter + Esc + 1
Reset the unit to the factory default
ettings
s
scription
De
Enter+Esc+2
10 Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
Loading...
+ 46 hidden pages