TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
•In North America, call 1-800-833-9200.
•Worldwide, visit to www.tek.com find contacts in your area.
Table of Contents
Table of Contents
List of Figures................................................................................................................................................................................7
List of Tables................................................................................................................................................................................. 8
Getting help and support.............................................................................................................................................................10
Virtual keypad and multi purpose knob................................................................................................................................ 11
File name extensions........................................................................................................................................................... 13
Launching the application.................................................................................................................................................... 14
Updating the database..................................................................................................................................................18
Result tab.............................................................................................................................................................................18
Querying Scope RN values for Custom or Automatic Mode.........................................................................................19
Querying Scope RN values for Current Scope Setting Mode.......................................................................................19
Saving and recalling setup.......................................................................................................................................................... 20
Saving a setup..................................................................................................................................................................... 20
Recalling a setup................................................................................................................................................................. 20
Performing characterization with Automatic or Custom Mode............................................................................................. 21
Performing characterization with Current Scope Setting mode........................................................................................... 25
About Parameters................................................................................................................................................................ 29
Result tab parameters..........................................................................................................................................................30
About the GPIB program......................................................................................................................................................31
Figure 5: Starting the Scope Noise Application from Analyze menu...........................................................................................14
Figure 10: View of the Result tab................................................................................................................................................ 19
Figure 13: Probe calibration information pop-up window............................................................................................................ 22
Figure 14: Probe-tip information pop-up window........................................................................................................................ 22
Figure 15: Pop-up for 50 Ohm termination for channel...............................................................................................................22
Figure 16: Database update during on-going noise characterization..........................................................................................23
Figure 17: Characterization completion pop-up while probe is connected..................................................................................23
Figure 19: Showing Math expression in "Additional Info" box.....................................................................................................24
Figure 20: Noise Compensation for RN measurement in the configuration panel DPOJET.......................................................24
Figure 21: Noise Compensation for TN@BER measurement in the configuration panel DPOJET............................................ 24
Figure 25: Disconnect DUT and connect probe with nominal tip................................................................................................ 26
Figure 26: Pop-up for 50 Ohm termination for channel...............................................................................................................26
Figure 28: Result tab showing probe model and SL no.............................................................................................................. 27
Figure 29: Showing Math expression in "Additional Info" box.....................................................................................................27
Figure 30: Noise Compensation for RN measurement in the configuration panel DPOJET.......................................................28
Figure 31: Noise Compensation for TN@BER measurement in the configuration panel DPOJET............................................ 28
Table 4: File name extensions.....................................................................................................................................................13
Table 5: Options menu settings...................................................................................................................................................15
Table 6: Example of scope noise values in Custom mode..........................................................................................................17
Table 7: Characterization tab options, parameters with default values.......................................................................................29
Table 8: Result tab options, parameters with default values.......................................................................................................30
The Oscilloscope's vertical noise arises mainly from the analog front-end and digitizing process. The oscilloscope noise will degrade the
actual measurement accuracy, especially when we are measuring low-amplitude signals. Hence, it is important to remove the oscilloscope
noise from the measurement results.
Scope Noise Characterization is a standalone application used to characterize noise for Tektronix Performance Digital Oscilloscopes
(DPO7000C, DPO70000C/D/DX, DSA70000C/D, MSO70000/C/DX, and DPO70000SX Series).
Figure 1: Scope Noise Characterization home screen
Key features:
•Supports Tektronix Performance Digital Oscilloscopes (DPO7000C, DPO70000C/D/DX, DSA70000C/D, MSO70000/C/DX, and
DPO70000SX Series).
•Supports characterizing the noise for the given range of values.
•Supports Noise characterization on both ATI and TekConnect channels.
•Supports Math expressions.
•Characterize the noise of the oscilloscope only or with the probe (and tip).
•Supports PI interface for all user controls.
•Interface to read the noise value for a specific oscilloscope configuration.
•Save/Recall of application setup file.
•Indicates the validity of the noise values.
Scope Noise Characterization Help 9
Getting help and support
Getting help and support
Product documents
Use the product documents for more information on the application functions, understand the theory of operation, how to remotely program
or operate the application, and do other tasks.
How to use the applicationScope Noise Characterization Help
PDF version of this document can be downloaded from www.tek.com/downloads
Compiled HTML (CHM) version is integrated with the application. Press F1 key from the
keyboard to launch the help.
Conventions
This document uses the following conventions:
•When steps require a sequence of selections using the application interface, the “>” delimiter marks each transition between a menu
and an option.
•For example, Analyze > Scope Noise Characterization
•The term “application” refers to Scope Noise Characterization application.
•The abbreviation used for “Scope Noise Characterization” will be “SNC”.
•The term “DUT” is an abbreviation for Device Under Test.
•The term “select” is a generic term that applies to the methods of choosing an option: with a mouse or with the touch screen.
Technical support
Tektronix values your feedback on our products. To help us serve you better, please send us your suggestions, ideas, or comments on
your application or oscilloscope. Contact Tektronix through mail, telephone, or the Web site. See Contacting Tektronix at the front of this
document for contact information.
When you contact Tektronix Technical Support, please include the following information (be as specific as possible):
General information
•All instrument model numbers
•Hardware options, if any
•Modules used
•Your name, company, mailing address, phone number, FAX number
•Please indicate if you would like to be contacted by Tektronix about your suggestion or comments.
Application specific information
•Software version number
•Description of the problem such that technical support can duplicate the problem
•If possible, save the setup files for all the instruments used and the application
Scope Noise Characterization Help 10
Getting started
Getting started
Software requirements
Scope Noise Characterization application requires DPOJET version 10.4.0.1 or later.
Supported probes
The application supports the following probes:
•P76xx
•P77xx
View software version
Click Options > About to view application details such as the software version number, application name, and copyright details. Refer to
Figure 2 on page 11
Figure 2: Scope Noise Characterization Version Number and Copyright information
The version displayed above is indicative only, the version number displayed will vary depending upon the exact version of
Note:
the application installed.
Virtual keypad and multi purpose knob
Select the
reference IDs.
icon on any of the entry text boxes and use the virtual keypad to enter alphanumeric values, such as voltage levels and
Scope Noise Characterization Help 11
Figure 3: Virtual numeric keypad
Getting started
Figure 4: Virtual keyboard
The MP Knob Reference and are the identifiers that show which Multi-Purpose Knob (MPK) may be used as an alternate means
to control a parameter; turn the knob on the oscilloscope front panel to adjust the corresponding parameter, refer to Figure 3 on page 12
and Figure 4 on page 12. Also, the value can be entered directly on the MPK display component.
Icon descriptions
This section lists the icons displayed in the Scope Noise Characterization application and their description.
Table 2: Icon descriptions
ICONDescription
This icon identifies warning / error.
This icon displays the description for various characterization
modes.