Tektronix PRISM MPI2-10, PRISM MPX2-10 Instructions Manual

xx
PRISM MPI2-10 and MPX2-10 MediaAnalysisPlatform
ZZZ
Field Upgrade
Instructions
*P075110600*
075-1106-00
xx
PRISM MPI2-10 and MPX2-10 Media Analysis Platform
ZZZ
Field Upgrade
Instructions
www.tek.com
075-1106-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverto USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200. Worldwide, visit www.tek.com to nd contacts in your area.
Table of Contents
Important safety information ............... ................................ ................................ ....... ii
General safety summary ...................................................................................... ii
Service safety summary........................ ................................ .............................. iv
Terms in thi
Symbols and terms on the product ........................................................................... v
Kit description ..... ................................ ................................ ................................. 1
Supported upgrades. ................................ ................................ ........................... 1
Kit parts list..................................................................................................... 1
MPI2-10 hardware upgrades . .................................. ................................ ................... 3
Recomme
Upgrade the PRISM MPI2-10 Eye mezzanine board....... ................................ ............... 4
MPX2-10 hardware upgrades .................................................................................... 11
Recommended tool list ........................... .................................. .......................... 11
Upgrade the PRISM MPX Eye mezzanine board......................................................... 12
Upgrade the instrument rmware................................ ................................ ................ 21
ade the software license..................................................................................... 24
Upgr
Verify the hardware upgrade ..................................................................................... 30
s manual . ................................ ................................ ........................ iv
nded tool list ............................. ................................ ........................... 3
PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions i
Important safety information
Important saf
ety information
This manual c for safe operation and to keep the product in a safe condition.
To safely perform service on this product, additional information is provided at the end of this section. (See page iv, Servicesafetysummary.)
General safety summary
Use the product only as specied. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
This product shall be used in accordance with loca l and national codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety precautions specied in this manual.
The product is designed to be used by trained personnel only.
Only qualied personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
ontains information and warnings that must be followed by the user
To avoid re or personal
injury
Before use, always check the product with a known source to be sure it is operating correctly.
This product is not intended for detection of hazardous voltages.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the oth cautions related to operating the system.
When incorporating this equipment into a system, the safety of that system is the responsibility of the assembler of the system.
Use proper power cord. Use only the power cord specied for this product and certied for the country of use.
Do not use the provided power cord for other products.
Power disconnect. The power cord disconnects the product from the power source. See instructions for the location. Do not p osition the equipment so that it is difcult to operate the power cord; it must remain accessible to the u ser at all times to allow for quick disconnection if needed.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product.
er component manuals for warnings and
ii PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions
Important safety information
Do not apply a po
tential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. Hazardous voltage exposure is possible.
Avoid expos
ed circuitry. Do not touch exposed connections and components
when power is present.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operati
on.
Before use, inspect test leads and accessories for mechanical damage and replace when da
maged. Do not use test leads if they are damaged, if there is exposed
metal, or if a wear indicator shows.
ne the exterior of the product before you use it. Look for cracks or missing
Exami pieces.
nly specied replacement parts.
Use o
Do not operate in wet/damp conditions. Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before you clean
the product.
rovide proper ventilation. Refer to the installation instructions in the manual for
P
details on installing the product so it has proper ventilation.
Slots and openings are provided for ventilation and should never be covered or otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or prolonged keyboard or pointer use may result in serious injury.
Be sure your work area meets applicable ergonomic standards. Consult with an ergonomics professional to avoid stress injuries.
Use only the Tektronix rackmount hardware specied for this product.
PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions iii
Important safety information
Service safet
ysummary
The Service safety summary section contains additional information required to safely perform service on the product. Only qualied personnel should perform service proc summary before performing any service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of this
product un present.
Disconnect power. To avoid electric shock, switch off the product power and disconnect the power cord from the mains power before removing any covers or panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components.
Verify safety after repair. Always recheck ground continuity and mains dielectric stre
edures. Read this Service safety summary and the General safety
less another person capable of rendering rst aid and resuscitation is
ngth after performing a repair.
Termsinthismanual
se terms may appear in this manual:
The
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
iv PRISM MPI2-10 and MPX2-10 Field Upgrade K it Instructions
Important safety information
Symbols and te
rms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
When this symbol is marked on the product, be sure to consult the manual to nd out the nature of the potential hazards and any actions which have to be taken to avoid them. (This symbol may also be used to refer the user to ratings in the manual.)
The follo
wing symbol(s) may appear on the product:
PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions v
Important safety information
vi PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions
Kit description
Supported upgrades
This document provides instructions for installing the Eye mezzanine board, upgrading the rmware, and updating software licenses. The k it applies to the following Tektronix instruments:
PRISM MPI2-10 Media Analysis Platform
PRISM MPX2-10 Media Analysis Platform
The following hardware upgrades are supported by this document:
MPI2-10-UP PHY. Add SDI Physical Layer Measurement Package to the MPI2-10 product; includes automated measurement of 12G/3G/HD/SD-SDI Eye pattern parameters, jitter p arameters and jitter waveform display.
(Option MP2-FMT-4K required for 12G support)
MPX2-10-UP PHY. Add SDI Physical Layer Measurement Package t o the MPX2-10 product; includes automated measurement of 12G/3G/HD/SD-SDI Eye pattern parameters and jitter parameters and jitter waveform display.
Kit parts list
(Option MP2-FMT-4K required for 12G support)
The following table lists the parts that are supplied with the upgrade kit.
Quantity Part Number Description
1 878-1285-XX
2 211-1117-XX
1 129-1889-XX
1 075-1095-XX
1
The spacer post provided in the kit will not be needed for the MPI2-10 12G Eye board upgrade. The MPI2-10 Mixed Media board comes with a second spacer post installed.
1
EMBLY, CIRCUIT BOARD,12G EYE MEZZ, 3895134XX
ASS
SCREW, MACHINE; 4-40 X 0.187, PAN HE AD, STEEL, ZINC
NISH, T-10, TORX DR
FI
SPACER POST .188 HEX X .325L BNC, MEZZ BOARD
OUNTING
M
PRISM FIELD UPGRADE INSTRUCTIONS
PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions 1
Kit description
2 PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions
MPI2-10 hardware upgrades
These instructions are for the qualied, trained personnel or personnel familiar with servicing the product. Contact your nearest Tektronix, Inc., Service Center or Tektronix Factory Service for installation assistance.
WAR NI NG . Dangerous voltages may be present. To prevent electrical shock,
disconnect Failure to do so could cause serious injury or death.
the power cord from the unit before starting the upgrade procedure.
Recommended tool list
CAUTION. M
any components within the instrument are susceptible to static discharge damage. To prevent static discharge damage, service the product only in a static-free environment. Observe standard handling precautions for static-sensitive devices while installing this kit. Always wear a grounded wrist strap, grounded foot strap, and static resistant apparel while installing this kit.
Table 1
Name Descri
9/16 i
Scre
T-10
T-15 TORX tip Used for removing instrument screws.
Torque driver
: Recommended tool list
nch socket nut driver
wdriver handle
TORX tip
ption
o removed BNC nuts.
Used t
Accepts Torx-driver bits
for removing instrument screws.
Used TORX-driver bit for T-10 size screw heads.
TORX-driver bit for T-15 size screw heads.
Used to apply precisely a specic torque to
astener.
af
PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions 3