This document applies to firmware version 1.00 and above.
Warning
The servicing instructions are for use by qualified personnel
only. To avoid personal injury, do not perform any servicing
unless you are qualified to do so. Refer to all safety summaries
prior to performing service.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O . Bo x 50 0
Beaverto
USA
For product information, sales, service, and technical support:
n, OR 97077
In North America, call 1-800-833-9200.
World wide , vis i t www.tektronix.com to find contacts in your area.
Table of Contents
General Safety Summary .........................................................................................iv
eplaceable parts – cable and connectors, rear chassis ............... ..............................40
Table of Contents
MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manualiii
General Safety Summary
General Safet
To Avoid Fire or Personal
Injury
ySummary
Review the fo
this product or any products connected to it.
To avoid potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
Use proper power cord. Use only the power cord specified for this product and
certified for the country of use.
Connect and disconnect properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Connect and disconnect properly. De-energize the circuit under test before
connecting or disconnecting the current probe.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To a void fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
llowing safety precautions to avoid injury and prevent damage to
Connect the probe reference lead to earth ground only.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating o f that termi
Power disconnect. The power cord disconnects the product from the power source.
Donotblockthepowercord;itmustremain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualified service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual’s installation instructions for
details on installing the product so it has proper ventilation.
nal.
ivMSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual
General Safety Summary
TermsinThisManual
Symbols and Terms on the
Product
These terms may
WAR N ING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
These terms may appear on the product:
DANGER in
the marking.
WAR N IN G
read the marking.
CAUTIO
The following symbol(s) may appear on the product:
appear in this manual:
dicates an injury hazard immediately accessible as you read
indicates an injury hazard not immediately accessible as you
N indicates a hazard to property including the product.
MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manualv
General Safety Summary
viMSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual
Preface
This service manual provides information that you need to troubleshoot,
disassemble, and replace parts on the following Tektronix oscilloscopes:
ModelBandwidthAnalog Channels
Manual Conventions
MSO4104B1 GHz
MSO4104B-L1 GHz
MSO4102B1 GHz
MSO4102B-L1 GHz
MSO4054B
MSO4034B
DPO4104B1 GHz
DPO4104B-L1 GHz
DPO4102B1 GHz
DPO4102B-L1 GHz
DPO4054B
DPO4034B
500 MHz4
350 MHz4
500 MHz4
350 MHz4
4
4
2
2
4
4
2
2
This manual uses certain conventions that you should become familiar with
before performing service.
Modules
Throughout this manual, any replaceable component, assembly, or part is referred
to by the term module.
Replaceable Parts
This manual refers to any field-replaceable assembly or mechanical part
specifically by its name or generically as a r eplaceable part. In general, a
replaceable part is any circuit board or assembly, such as the hard disk drive, or a
mechanical part, such as the I/O port connectors, that is listed in the replaceable
parts list.
Safety
Symbols and terms related to safety appear in the General Safety Summary.
Information for service procedures appears in both the General Safety Summary
and the Service Safety Summary.
MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manualvii
Preface
viiiMSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual
WheretoFindOperatingInformation
For information on installing and operating your DPO4000B or your MSO4000B
Series Digital Phosphor Oscilloscope, refer to the Tektronix MSO4000B Series
and DPO4000B
provided with your oscilloscope. This manual is also available, in 11 languages,
at www.tektronix.com/manuals.
Series Digital Phosphor Oscilloscopes User Manual, which was
MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual1
Where to Find Operating Information
2MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual
Theory of Operation
This chapter describes the electrical operation of the oscilloscope to the module
level. The block diagrams show four-channel and two-channel oscilloscope
module inter
Power Supply
connections. (See Figure 1 on page 5.) (See Figure 2 on page 6.)
The Power S
internal circuits.
I/O Board
The I/O board contains USB ports, an Ethernet port (LAN), a VGA Video port, an
AUX OUT BNC connector, and an EXT REF IN BNC connector.
Main Board and Analog Board
The Main and Analog boards contain the following functions:
Acquisition System
Trigger System
The Ac
digitized signal in memory. The signal enters a channel input, and then passes
through an attenuator and preamplifier. The analog signal from each preamplifier
goes through a digitizer, a time base controller, and then into acquisition memory.
The analog signal from each preamplifier is also distributed to a trigger circuit.
The Trigger system processes the analog signals from the preamplifiers and sends
trigger information to the time-base controller. Advanced trigger functions are
enabled only when the appropriate application modules and supporting software
are installed.
quisition system begins with the analog signal path and ends with a
upply board converts AC line voltage to +12 V to power for all
splay System
Di
Processor System
MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual3
The Display system combines live waveform data from acquisition memory with
menus and text, and stores this information in display memory. It then uses this
data to refresh the XGA display module (LCD).
The Processor system contains a 460EX Power PC microprocessor that controls
the entire instrument. The processor system also contains FLASH ROM, system
RAM, and interfaces to USB ports and the Ethernet port.
Theory of Operation
Power Converter
Front-Panel Board
The Power Conve
main power supply and generates voltages for the analog and digital circuitry on
the Main board. The front panel receives the +5 V standby power, and uses input
from the power switch to turn the oscilloscope on and off.
The Front Panel board contains a microprocessor that reads the front-panel buttons
and controls, and then sends this information to the processor system on the Main
board. The Front Panel board also generates the probe compensation output signal
and provid
es an interface to the application modules.
rter receives +12 V power and +5 V standby power from the
4MSO4000B and DPO4000B Series Digital Phosphor Oscilloscopes Service Manual