Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a
period of 2 years from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defecti v e.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in
Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation
prepaid, to the indicated service facility . Repairs will be made and the product returned, transportation prepaid.
Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMIT A TION OF W ARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written
consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE.
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR
ANY DIRECT , INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF
THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS
BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION,
LOSSES SUSTAINED AS THE RESULT OF INJURY T O ANY PERSON, OR DAMAGE TO PROPERTY.
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered
sequentially. When a new Revision is created, all Addenda associated with the previous Revision
of the manual are incorporated into the new Revision of the manual. Each new Revision includes
a revised copy of this print history page.
Revision A (Document Number 2303-902-01)............................................................January 1998
Revision B (Document Number 2303-902-01) .................................................................June 1998
Revision C (Document Number 2303-902-01) ........................................................November 1999
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
SafetyPrecautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous
voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar
with the safety precautions required to avoid possible injury. Read the operating information carefully
before using the product.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for
ensuring that the equipment is operated within its specifications and operating limits, and for ensuring
that operators are adequately trained.
Operatorsuse the product for its intended function. They must be trained in electrical safety procedures
and proper use of the instrument.They must be protected from electric shock and contact with hazardous
livecircuits.
Maintenance personnel perform routine procedures on the product to keep it operating, for example,
setting the line voltage or replacing consumable materials. Maintenance procedures are described in the
manual. The procedures explicitly state if the operator may perform them. Otherwise, they should be
performed only by service personnel.
Service personnel are trained to work on live circuits, and perform safe installations and repairs of products. Only properly trained service personnel may perform installation and service procedures.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety
practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Users of this product must be protected from electric shock at all times. The responsible body must ensure that users are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product users in these circumstances must be
trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or
above 1000 volts, no conductive part of the circuit may be exposed.
As described in the International Electrotechnical Commission (IEC) Standard IEC 664, digital multimeter measuring circuits (e.g., Keithley Models 175A, 199, 2000, 2001, 2002, and 2010) are Installation
Category II. All other instruments’ signal terminals are Installation Category I and must not be connected to mains.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with
impedance limited sources. NEVER connect switching cards directly to AC mains. When connecting
sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any
capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching
cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power
line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions
or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications
and operating information, and as shown on the instrument or test fixture panels, or switching card.
Whenfuses are usedin a product, replacewith same type and rating for continued protectionagainstfire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth
ground connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should refer to the operating instructions located in
the manual.
The symbol on an instrument shows that it can source or measure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact
with these voltages.
The WARNING heading in a manual explains dangers that might result in personal injury or death. Always
read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may
invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the
power transformer, test leads, and input jacks, must be purchased from Keithley Instruments. Standard fuses,
with applicable national safety approvals, may be used if the rating and type are the same. Other components
that are not safety related may be purchased from other suppliers as long as they are equivalent to the original
component. (Note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component,
call a Keithley Instruments office for information.
To clean an instrument, use a damp cloth or mild, water based cleaner. Clean the exterior of the instrument
only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a
computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
Use the procedures in this section to verify that Model 2303, Model 2303B, or Model
2303-PJ accuracy is within the limits stated in the accuracy specifications. You can perform these
verification procedures:
•When you first receive the unit to make sure that it was not damaged during shipment.
•To verify that the unit meets factory specifications.
•To determine if calibration is required.
•Following calibration to make sure it was performed properly.
WARNINGThe information in this section is intended only for qualified service per-
sonnel. Do not attempt these procedures unless you are qualified to do so.
NOTEIf the power supply is still under warranty, and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the correct
course of action.
Verificationtestrequirements
Be sure that you perform the verification tests:
•Under the proper environmental conditions.
•After the specified warm-up period.
•Using the correct line voltage.
•Using the proper test equipment.
•Using the specified output signals and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
•An ambient temperature of 18-28°C (65-82°F).
•A relative humidity of less than 70% unless otherwise noted.
Warm-upperiod
Allow the Model 2303 to warm up for at least one hour before conducting the verification
procedures.
If the unit has been subjected to temperature extremes (those outside the ranges stated above),
allow additional time for the instrument’s internal temperature to stabilize. Typically, allow one
extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
Also,allowthe test equipment to warmup for theminimumtimespecifiedbythe manufacturer.
Line power
The Model 2303 requires a line voltage of 100 to 240V and a line frequency of 50 to 60Hz.
Verification tests must be performed within this range.
Recommended testequipment
Table 1-1 summarizes recommended verification equipment. You can use alternate equipment as long as that equipment has specifications at least four times better than corresponding
Model 2303, 2303B, or 2303-PJ specifications. Keep in mind, however, that test equipment
accuracy will add to the uncertainty of each measurement.
Table 1-1
Recommended verification equipment
DescriptionManufacturer/ModelSpecifications
Digital MultimeterKeithley 2001DC Voltage1 20V: ±22ppm
1. Full-range, 90-day, 23˚C various measurement points.
2. Characterize resistor using 4-wire ohms function and 20Ω range of DMM before use.
3. Used only for Model 2303-PJ. Characterize resistor using 4-wire ohms function and 200Ω range of DMM
before use. See Resistor considerations for temperature coefficient.
4. Used only for Models 2303 and 2303B. Characterize resistor using 4-wire ohms function and 20kΩ range of
DMM before use.
Resistorconnections
When performing the verification tests, make connections to the test resistors as shown in
Figure 1-1. Be sure to connect the Model 2303 SENSE leads and DMM test leads as close to the
resistor body as possible.
1-4Performance Verification
Figure 1-1
Test resistor connections
Resistorconsiderations
The test resistors should be characterized using the lowest possible range and the 4-wire ohms
function of the DMM recommended in Table 1-1 to measure the resistance values. Use the measured resistance values to calculate the actual currents during the test procedures.
NOTEThe temperature coefficient and temperature change of the 1Ω and 30Ω resistors
when passing current at full load must be low enough so that the change in resistance
does not cause incorrect readings as follows:
V
OUT
voltage, and ∆R is the change in resistance caused by heating.)
Verificationlimits
The verification limits stated in this section have been calculated using only the Model 2303
accuracy specifications, and they do not include test equipment uncertainty . If a particular mea-
surement falls outside the allowable range, recalculate new limits based both on Model 2303
specifications and corresponding test equipment specifications.
Examplelimitscalculation
As an example of how verification limits are calculated, assume you are testing the power
supply using a 10V output value. Using the Model 2303 voltage output accuracy specification
of ±(0.05% of output + 10mV offset), the calculated output limits are:
If the Model 2303 is not within specifications and not under warranty, see the calibration pro-
cedures in Section 2 for information on calibrating the unit.
Testconsiderations
When performing the verification procedures:
•Make sure that the test equipment is properly warmed up and connected to the correct
Model 2303 terminals on the rear panel. Also be sure the test equipment is set up for the
proper function and range.
•Do not connect test equipment to the Model 2303 through a scanner, multiplexer, or
other switching equipment.
•Be sure that the power supply output is turned on before making measurements.
•Allow the power supply output signal to settle before making a measurement.
Instrumentprogramming
Refer to Appendix D for basic information on programming the Model 2303, 2303B, or
2303-PJ from the front panel or over the bus.
Outputvoltageaccuracy
Follow the steps below to verify that Model 2303, 2303B, or 2303-PJ output voltage accuracy
is within specified limits. This test involves setting the output voltage to specific values and measuring the voltages with a precision digital multimeter.
1.With the power off, connect the digital multimeter to the Model 2303 OUTPUT
SOURCE terminals, as shown in Figure 1-2. Be sure to observe proper polarity
(SOURCE + to INPUT HI; SOURCE - to INPUT LO).
1-6Performance Verification
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG
POWER
INFO LOCALEXIT ENTER
Figure 1-2
Connections for voltage verification tests
STORE RECALL
FILTER MATH
CHAN SCAN
CONFIGMENU
Model 2001 DMM
2001 MULTIMETER
FREQ TEMP
RANGE
RANGE
AUTO
Input HI
SENSE
Ω 4 WIRE
350V
PEAK
INPUTS
F R
FRONT/REAR
CAL
Input LO
INPUT
HI
1100V
!
PEAK
LO
500V
PEAK
2A 250V
AMPS
Source +Source -
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
ISOLATION FROM EARTH:
22 VOLTS MAX.
_ _ _ _
+ + +
SOURCE SENSED VM INSOURCE
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
U.S.A.
CAUTION: FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
OUTPUT
IEEE-488
0-20V, 0-5A
+
Model 2303
LINE FUSE
SLOWBLOW
2.5A, 250V
LINE RATING
100-240VAC
50, 60 HZ
185VA MAX
REMOTE
DISPLAY
OPTION
2.Select the multimeter DC volts measuring function, and enable auto-ranging.
3.Make sure the Model 2303 output is turned on.
4.Verify output voltage accuracy for each of the voltages listed in Table 1-2. For each test
point:
• Set the Model 2303 output voltage to the indicated value. When setting the voltage,
set the compliance current to 3A.
• Allow the reading to settle.
• Verify that the multimeter reading is within the limits given in the table.
Table 1-2
Output voltage accuracy limits
Model 2303 output
voltage setting
Output voltage limits
(1 Year, 18°C-28°C)
5.00V4.9875 to 5.0125V
10.00V9.985 to 10.015V
15.00V14.9825 to 15.0175V
Voltagereadbackaccuracy
Follow the steps below to verify that Model 2303, 2303B, or 2303-PJ voltage readback accuracy is within specified limits. The test involves setting the source voltage to specific values, as
measured by a digital multimeter, and then verifying that voltage readback readings are within
required limits.
1.With the power off, connect the digital multimeter to the Model 2303 OUTPUT
SOURCE terminals, as shown in Figure 1-3. Be sure to observe proper polarity
(SOURCE + to INPUT HI; SOURCE - to INPUT LO).
2.Select the multimeter DC volts function, and enable auto-ranging.
3.Make sure actual voltage readings are being displayed (use DISPLAY), and turn on the
Model 2303 output.
4.Verify voltage readback accuracy for each of the voltages listed in Table 1-3. For each
test point:
• Set the Model 2303 output voltage to the indicated value as measured by the digital
multimeter. Note that it may not be possible to set the voltage source precisely to the
specified value. Use the closest possible setting, and modify reading limits accordingly. When setting the voltage, set the compliance current to 3A.
• Allow the reading to settle.
• Verify that the actualModel 2303 voltagereading is withinthe limits givenin the table.
Performance Verification 1-7
Table 1-3
Voltage readback accuracy limits
Model 2303 output
voltage setting*
5.00V4.995 to 5.005V
10.00V9.992 to 10.008V
14.00V13.990 to 14.010V
* As measured by digital multimeter. See procedure.
Voltage readback limits
(1 Year, 18°C-28°C)
Compliancecurrentaccuracy
Follow the steps below to verify that Model 2303, 2303B, or 2303-PJ compliance current
accuracy is within specified limits. The test involves setting the compliance current to specific
values and determining the actual current by measuring the voltages across a characterized 1Ω
resistor with a precision digital multimeter.
1.With the power off, connect the digital multimeter and characterized 1Ω resistor to the
Model 2303 OUTPUT SOURCE terminals, as shown in Figure 1-3. Be sure to observe
1-8Performance Verification
proper polarity (SOURCE + to INPUT HI; SOURCE - to INPUT LO). Also be sure to
use 4-wire connections to the resistor terminals.
PREV
DISPLAY
NEXT
POWER
2001 MULTIMETER
DCV ACV DCI ACI Ω2 Ω4
REL TRIG
STORE RECALL
INFO LOCALEXIT ENTER
CHAN SCAN
FREQ TEMP
FILTER MATH
CONFIGMENU
RANGE
RANGE
AUTO
Input HI
SENSE
Ω 4 WIRE
350V
PEAK
INPUTS
F R
FRONT/REAR
CAL
INPUT
HI
!
LO
Input LO
1100V
PEAK
500V
PEAK
2A 250V
AMPS
1 Ω Resistor
Model 2001 DMM
22 VOLTS MAX.
_ _ _ _
Sense -
+
LINE FUSE
SLOWBLOW
2.5A, 250V
LINE RATING
100-240VAC
50, 60 HZ
185VA MAX
Note: Use 4-wire connections
to resistor terminals.
Sense +
Source +Source -
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
ISOLATION FROM EARTH:
+ + +
SOURCE SENSEDVM INSOURCE
OUTPUT
0-20V, 0-5A
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
U.S.A.
CAUTION: FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Model 2303
Figure 1-3
Connections for output current and 5A range current verification tests
2.Select the multimeter DC voltage measuring function, and enable auto-ranging.
3.Turn on the Model 2303 output.
4.Verify compliance current accuracy for the currents listed in Table 1-4. For each test
point:
• Set the Model 2303 output voltage to 9V, and set the compliance current to the value
being tested.
• Note and record the digital multimeter voltage reading.
• Calculate the current from the voltage reading and actual 1Ω resistor value: I = V/R
• Verify that the current is within the limits given in the table.
REMOTE
DISPLAY
OPTION
Table 1-4
Compliance current accuracy limits
Performance Verification 1-9
Model 2303 compliance
current setting
1.000A0.993 to 1.007A
2.000A1.992 to 2.008A
3.000A2.990 to 3.010A
4.000A3.989 to 4.011A
5.000A4.987 to 5.013A
Compliance current limits
(1 Year, 18°C-28°C)
Current readbackaccuracy
Follow the steps below to verify that Model 2303, 2303B, or 2303-PJ current readback accuracy is within specified limits. The test involves setting the output current to specific values as
measured with a resistor and precision digital multimeter.
5Arangereadbackaccuracy
1.With the power off, connect the digital multimeter and characterized 1Ω resistor to the
Model 2303 OUTPUT SOURCE terminals, as shown in Figure 1-3. Be sure to observe
proper polarity (SOURCE + to INPUT HI; SOURCE - to INPUT LO). Also be sure to
use 4-wire connections to the resistor terminals.
2.Select the multimeter DC volts measuring function, and enable auto-ranging.
3.Select the 5A readback range. Also make sure actual current readings are displayed.
4.Turn on the Model 2303 output.
5.Verify 5A range current readback accuracy for the currents listed in Table 1-5. For each
test point:
• By changing the output voltage, adjust the current to the correct value, as determined
from the multimeter voltage reading and characterized resistance value. When setting
the voltage, be sure to set the compliance current to 5A.
• Notethat it may not be possibleto set the outputcurrent to the exactvalue.In that case,
set the current to the closest possible value, and modify readinglimits accordingly.
• Allow the reading to settle.
• Verify that the actual Model2303 current reading is withinthe limits givenin the table.
1-10Performance Verification
Table 1-5
5A range current readback accuracy limits
Nominal output
Voltage
Model 2303
Output current*
Current readback limits
(1 Year, 18°C-28°C)
1V1.000A0.9976 to 1.0024A
2V2.000A1.9956 to 2.0044A
3V3.000A2.9936 to 3.0064A
4V4.000A3.9916 to 4.0084A
4.9V4.900A4.8898 to 4.9102A
* As determined from digital multimeter and 1Ω resistor. See procedure.
Models 2303and2303B5mArange readbackaccuracy
The following test applies only to the Models 2303 and 2303B:
1.With the power off, connect the digital multimeter and characterized 3kΩ resistor to the
Model 2303 or 2303B OUTPUT SOURCE terminals, as shown in Figure 1-4. Be sure
to observe proper polarity and connections (3kΩ resistor between SOURCE + and DMM
INPUT HI; SOURCE - to DMM INPUT LO).
Input HI
Input LO
1100V
PEAK
500V
PEAK
2A 250V
AMPS
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG
POWER
STORE RECALL
INFO LOCALEXIT ENTER
CHAN SCAN
FILTER MATH
CONFIGMENU
2001 MULTIMETER
FREQ TEMP
RANGE
RANGE
SENSE
INPUT
Ω 4 WIRE
HI
350V
!
PEAK
LO
INPUTS
F R
FRONT/REAR
AUTO
CAL
3kΩ Resistor
Model 2001 DMM
Sense +
Source +Source -
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
ISOLATION FROM EARTH:
22 VOLTS MAX.
_ _ _ _
+ + +
SOURCE SENSEDVM INSOURCE
OUTPUT
0-20V, 0-5A
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
Note: Use 4-wire connections
to resistor terminals.
U.S.A.
CAUTION: FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Model 2303
Figure 1-4
Connections for Model 2303 and 2303B 5mA current verification tests
+
Sense -
SLOWBLOW
LINE RATING
LINE FUSE
2.5A, 250V
100-240VAC
50, 60 HZ
185VA MAX
REMOTE
DISPLAY
OPTION
Performance Verification 1-11
2.Select the multimeter DC volts measuring function, and enable auto-range.
3.Select the 5mA readback range and actual current reading data.
4.Turn on the Model 2303 output.
5.Verify 5mA range current readback accuracy for the currents listed in Table 1-6. For each
test point:
• Set the Model 2303 output current to the correct value, as determined from the digital
multimeter voltage reading and 3kΩ resistance value. Note that it may not be possible
to set the output current to the exact value. In that case, set the current to the closest
possible value, and modify reading limits accordingly.
• Allow the reading to settle.
• Verify that the actual Model2303 current reading is withinthe limits givenin the table.
Table 1-6
Model 2303 and 2303B 5mA range current readback accuracy limits
Nominal output
Voltage
3V1.0000mA0.9970 to 1.0030mA
6V2.0000mA1.9950 to 2.0050mA
9V3.0000mA2.9930 to 3.0070mA
12V4.0000mA3.9910 to 4.0090mA
14.7V4.9000mA4.8892 to 4.9108mA
* As determined from digital multimeter voltage reading and 3kΩ resistance
value. See procedure.
Model 2303
output current*
Current readback limits
(1 Year, 18°C-28°C)
Model2303-PJ500mArangereadback accuracy
The following tests apply only to the Model 2303-PJ:
1.With the power off, connect the digital multimeter and characterized 30Ω resistor to the
Model 2303-PJ OUTPUT SOURCE terminals, as shown in Figure 1-5. Be sure to
observe proper polarity and connections (30Ω resistor between SOURCE + and DMM
INPUT HI; SOURCE - to DMM INPUT LO).
1-12Performance Verification
DCV ACV DCI ACI Ω2 Ω4
PREV
DISPLAY
NEXT
REL TRIG
POWER
INFO LOCALEXIT ENTER
Note: Use 4-wire connections
to resistor terminals.
STORE RECALL
FILTER MATH
CHAN SCAN
CONFIGMENU
Model 2001 DMM
2001 MULTIMETER
FREQ TEMP
RANGE
AUTO
RANGE
Input HI
SENSE
Ω 4 WIRE
350V
PEAK
INPUTS
FR
FRONT/REAR
CAL
INPUT
HI
!
LO
2A 250V
AMPS
Input LO
1100V
PEAK
500V
PEAK
Sense +
Source +Source -
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
ISOLATIONFROM EARTH:
22 VOL TS MAX.
_ _ _ _
+ + +
SOURCE SENSEDVM INSOURCE
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
U.S.A.
CAUTION: FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
OUTPUT
IEEE-488
+
0-20V, 0-5A
Model 2303
Sense -
SLOWBLOW
LINE RATING
3kΩ Resistor
LINE FUSE
2.5A, 250V
100-240VAC
50, 60 HZ
185VA MAX
REMOTE
DISPLAY
OPTION
Figure 1-5
Connections for Model 2303-PJ 500mA current verification tests
2.Select the multimeter DC volts measuring function, and enable auto-range.
3.Select the 500mA readback range and actual current reading data.
4.Turn on the Model 2303-PJ output.
5.Verify 500mA range current readback accuracy for the currents listed in Table 1-7. For
each test point:
• Set the Model 2303-PJ output current to the correct value, as determined from the dig-
ital multimeter voltage reading and 30Ω resistance value. Note that it may not be pos-
sible to set the output current to the exact value. In that case, set the current to the
closest possible value, and modify reading limits accordingly.
• Allow the reading to settle.
• Verify that the actual Model 2303-PJ current reading is within the limits given in the
table.
Table 1-7
Model 2303-PJ 500mA range current readback accuracy limits
Performance Verification 1-13
Nominal output
voltage
Model 2303-PJ
output current*
Current readback limits (1 Year,
18˚C-28˚C)
3V100.00mA99.76 to 100.24mA
6V200.00mA199.56 to 200.44mA
9V300.00mA299.36 to 300.64mA
12V400.00mA399.16 to 400.84mA
14.7V490.00mA488.98 to 491.02mA
Digital voltmeter input accuracy
Follow the steps below to verify that Model 2303, 2303B, and 2303-PJ digital voltmeter input
accuracy is within specified limits. The test involves setting the voltage applied to the DVM
input to accurate values and then verifying that the Model 2303 digital voltmeter input readings
are within required limits.
1.With the power off, connect the Model 2303 DVM IN terminals to OUTPUT SOURCE
terminals and the digital multimeter, as shown in Figure 1-6. Be sure to observe proper
polarity (DVM IN + to SOURCE + and DMM INPUT HI; DVM IN - to SOURCE - and
DMM INPUT LO).
Input HI
SENSE
Ω 4 WIRE
350V
PREV
DCV ACV DCI ACI Ω2 Ω4
DISPLAY
NEXT
REL TRIG
POWER
STORE RECALL
INFO LOCALEXIT ENTER
CHAN SCAN
2001 MULTIMETER
FREQ TEMP
FILTER MATH
CONFIGMENU
RANGE
AUTO
RANGE
PEAK
F R
INPUTS
FRONT/REAR
CAL
INPUT
HI
!
LO
Input LO
1100V
PEAK
500V
PEAK
2A 250V
AMPS
DVM IN DVM IN +
Model 2001 DMM
Figure 1-6
Connections for DVM accuracy verification
Source +
Source -
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
ISOLATION FROM EARTH:
22 VOLTS MAX.
_ _ _ _
+ + +
SOURCE SENSEDVM INSOURCE
OUTPUT
0-20V, 0-5A
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
MADE IN
U.S.A.
CAUTION: FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
LINE FUSE
SLOWBLOW
2.5A, 250V
LINE RATING
100-240VAC
50, 60 HZ
+
185VA MAX
Model 2303
REMOTE
DISPLAY
OPTION
1-14Performance Verification
2.Select the DMM DC volts function, and enable auto-ranging.
3.Enable the Model 2303 DVM input.
4.Turn on the Model 2303 source output.
5.Verify digital voltmeter input accuracy for each of the voltages listed in Table 1-8. For
each test point:
• Set the voltage to the indicated value as measured by the digital multimeter.
• Set compliance to 3A.
• Allow the reading to settle.
• Verify that the Model 2303 voltage reading is within the limits given in the table.
Table 1-8
Digital voltmeter input accuracy limits
Model 2303 output
voltage setting*
Digital voltmeter input reading limits
(1 Year, 18°C-28°C)
14.00V13.990 to 14.010V
3.00V2.996V to 3.004V
* As measured by digital multimeter. See procedure.
2
Calibration
2-2Calibration
Introduction
Use the procedures in this section to calibrate the Model 2303, 2303B, or 2303-PJ. These procedures require accurate test equipment to measure precise DC voltages and currents. Calibration can be performed either from the front panel (Models 2303 and 2303-PJ only), or by
sending SCPI calibration commands over the IEEE-488 b us with the aid of a computer (Models
2303, 2303B, and 2303-PJ).
WARNING The information in this section is intended only for qualified service person-
nel. Do not attempt these procedures unless you are qualified to do so.
Environmental conditions
T emperature and relative humidity
Conduct the calibration procedures at an ambient temperature of 18-28°C (65-82°F) with a
relative humidity of less than 70% unless otherwise noted.
W arm-up period
Allow the Model 2303 to warm up for at least one hour before performing calibration.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument’s internal temperature to stabilize. Typically,
allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature
range.
Also, allow the test equipment to warm up for the minimum time specified by the
manufacturer.
Line power
The Model 2303 requires a line voltage of 100 to 240V at line frequency of 50 to 60Hz. The
instrument must be calibrated while operating from a line voltage within this range.
Calibration considerations
When performing the calibration procedures:
•Make sure that the test equipment is properly warmed up and connected to the appropriate Model 2303 terminals.
•Always allow the source signal to settle before calibrating each point.
•Do not connect test equipment to the Model 2303 through a scanner or other switching
equipment.
20 Ω
Ω
Ω
Ω
Calibration 2-3
•Calibration must be performed in the sequence outlined in this manual, or an error will
occur.
•If an error occurs during calibration, the Model 2303 will generate an appropriate error
message. See Appendix B for more information.
WARNING The maximum common-mode voltage (voltage between LO and chassis
ground) is 22VDC. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.
Calibration cycle
Perform calibration at least once a year to ensure the unit meets or exceeds its specifications.
Recommended calibration equipment
Table 2-1 lists the recommended equipment for the calibration procedures. You can use
alternate equipment as long as that equipment has specifications at least four times better than
corresponding Model 2303 specifications. See also Resistor considerations for important temperature coefficient information.
1. Full-range, 90-day, 23˚C ±5˚C accuracy specifications of ranges required for various measurement points.
2. Characterize resistor using 4-wire ohms function and 20 Ω range of DMM before use.
3. Required only for Model 2303-PJ. Characterize resistor using 4-wire ohms function and 200 Ω range of DMM
before use. See Resistor considerations for temperature coefficient.
4. Required only for Models 2303 and 2303B. Characterize resistor using 4-wire ohms function and 20k Ω range
of DMM before use. See Resistor considerations for temperature coefficient.
, 0.1%, 100W
1
20V: ±22ppm
1
200
20k
2
3
: ±59ppm
: ±43ppm
: ±36ppm
4
Resistor connections
When performing calibration, make connections to the test resistors as shown in Figure 2-1.
Be sure to connect the Model 2303 SENSE leads and DMM test leads as close to the resistor
body as possible.
2-4Calibration
2303 Source +
2303 Sense +
2303 Sense -
2303 Source -
Figure 2-1
Test resistor connections
Resistor considerations
The test resistors should be characterized using the 4-wire ohms function of the DMM recommended in Table 2-1 to measure the resistance values. Use the measured resistance values to
calculate the actual currents during the calibration procedures.
NOTE The temperature coefficient and temperature change of the 4
when passing current at full load must be low enough so that the c hange in r esistance
does not cause incorrect readings as follows:
/
∆
R <25% of Model 2303 amps specification. (V
V
OUT
voltage, and
DMM Input HI
Test Resistor
DMM Input LO
OUT
∆
R is the change in resistance caused by heating.)
Ω
and 30
Ω
resistors
is the Model 2303 output
Front panel calibration
Model 2303 and Model 2303-PJ front panel calibration are covered separately below.
Model 2303 front panel calibration
Use the following procedure to calibrate the Model 2303 from the front panel. T able 2-2 summarizes calibration steps.
NOTE Calibration must be performed in the following sequence, or an error will occur. To
abort calibration and re vert to previous calibration constants at any time during the
procedure, press the MENU key.
Calibration 2-5
Step 1: Prepare the Model 2303 for calibration
1.Turn on the Model 2303 and the digital multimeter, and allow them to warm up for at
least one hour before performing calibration.
2.Press the MENU key, then choose CALIBRATE UNIT, and press ENTER. The instrument will display the date last calibrated:
CALIBRATE UNIT
LAST ON 02/01/98
3.Press the up arrow key. The instrument will display the number of times it was
calibrated:
CALIBRATE UNIT
TIMES= 01
4.Press the up arrow key. The unit will then prompt you to run calibration:
CALIBRATE UNIT
RUN
5.Press ENTER. The unit will then prompt for the calibration code:
CALIBRATE UNIT
Cal Code KI002303
6.Using the edit keys, set the display to the current calibration code, then press ENTER.
(Default: KI002303.) The unit will then prompt you as to whether or not to change the
code:
CALIBRATE UNIT
Change Code NO
7.Be sure NO is selected (use the up and down arrow k eys), press ENTER, then follo w the
steps below to calibrate the unit. (See Changing the calibration code at the end of this
section if you wish to change the code.)
Step 2: Perform calibration steps
NOTE The unit will display the most recently calibrated values. Factory defaults are shown
in this manual.
1.Connect both the OUTPUT SOURCE and DVM IN terminals to the digital multimeter,
as shown in Figure 2-2. (Connect SOURCE + and D VM IN + to DMM INPUT HI; connect SOURCE - and DVM IN - to DMM INPUT LO.)
Loading...
+ 74 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.