The Model 2290-PM-200 protection module (PM) is a stand-alone module that protects certain lower-voltage
source-measure units (SMUs) that are part of a testing configuration from damage by voltage sources that are
greater than 220 V.
Figure 1: Model 2290-PM-200
The protection module is for use with the following lower-voltage SMUs:
• Model 2611B Single-Channel System SourceMeter
• Model 2612B Dual-Channel System SourceMeter Instrument
• Model 2635B Single-Channel System SourceMeter Instrument
• Model 2636B Dual-Channel System SourceMeter Instrument
• Model 4200-SCS Semiconductor Characterization System SMUs:
• Model 4200-SMU Medium Power Source-Measure Unit
• Model 4210-SMU High Power Source-Measure Unit
071317001 / May 2015 *P071317001* 1
®
Instrument
Model 2290-PM-200 Protection Module User's Guide
The protection module is intended for use in applications where a device breakdown or other potential failure
could connect the high-voltage output of the Model 2290-5 or Mod el 229 0-10 hi g h-voltage power supply
instrument to a lower-voltage SMU.
The Model 2290-PM-200 does not protect the Model 2601B Single-Channel System SourceMeter
Instrument or the Model 2602B Dual-Channel System SourceMeter Instrument. Do not use the
protection module with these source-measure instruments.
To prevent damage to the Model 2290-PM-200 in the case of device failure, the current from all
power sources, including SourceMeter instruments and power supplies, should be limited to 100 mA.
The Model 2290-PM-200 protection module must be connected to protectiv e earth terminals
(safety grounds) using the supplied green-yellow ground cables (Model CA-568). Connection
of ground cables to protective ear th ter minals is necessary for safety.
Figure 2: Model 2290-PM-200 grounding
The standard triaxial IN connectors are used to connect the SMU. Maximum clamped voltage on the IN
connectors is 240 V DC.
• The center conductor of the LO connector inner shield is connected to LO. The outer shield (shell) is
chassis ground.
• The center conductor of the HI connector is HI and the inner shield is guard. The outer shield (shell) is
chassis ground.
2 071317001 / May 2015
Model 2290-PM-200 Protection Module User's Guide
There are two high-voltage coaxial connectors used for connection to the device under test (DUT).
• The smaller OUT connector is intended for use in applications that include the Model 2290-5 high-voltage
power supply only. This connector is rated to 5000 V.
• The larger OUT connector is intended for use in applications that include the Model 2290-10 high-voltage
power supply only. This connector is rated to 10000 V.
Electrical char ac ter i stics
• Leakage: < 10 pA at 200 V typical
• DC current carrying capability: 100 mA maximum in unclamped condition
• Pulsed current carrying capability: 10 A maximum at 1% duty cycle in unclamped condition
• Maximum protection active current: You must limit current to <120 mA in clamped condition (~230 V)
• Protection voltage:
• Minimum: 220 V DC
• Typical: 230 V DC
• Maximum: 240 V DC
• Series resistance: 470 Ω ± 10 percent
Mechanical dimensions
Figure 3: Model 2290-PM-200 dimensions
071317001 / May 2015 3
Model 2290-PM-200 Protection Module User's Guide
Connections using a M odel 2611B or 2612B SMU
Required accessories:
• One Model 2600-TRIAX adapter per channel
• Two standard triaxial cables (either Model 7078-TRX or 4200-TRX) per source-measure unit (SMU)
channel
• One high-voltage (HV) coaxial cable
• Use Model 2290-5-SHV or 2290-5-MHV if you are using the 2290-5 high-voltage power supply
Connection summar y
•Use Model 2290-10-SHVUC or 2290-10-SHV if you are using the 2290-10 high-voltage power supply
Refer to the next figure to connect the Model 2611B or 2612B source-measure unit (SMU) channels to the
Model 2290-PM-200.
Only 2-wire applications are supported.
Connecting the Model 2600-TRIAX to the Model 2290-PM-200 connects the LO terminal of the
Model 2611B or 2612B to protective earth (safety ground).
Do not convert triaxial cables to BNC cables. Using BNC cables may compromise overvoltage
protection and can result in damage to the SMU. You must use triaxial cables to connect to the
lower-voltage SMU.
4 071317001 / May 2015
Model 2290-PM-200 Protection Module User's Guide
Figure 4: Model 2290-PM-200 connections for Model 261xB using Model 2290-5 power supply
071317001 / May 2015 5
Model 2290-PM-200 Protection Module User's Guide
Figure 5: Model 2290-PM-200 connections for Model 261xB using Model 2290-10 power supply
6 071317001 / May 2015
Model 2290-PM-200 Protection Module User's Guide
Connections using a M odel 2635B or 2636B SMU
Required accessories:
• Two standard triaxial cables (either Model 7078-TRX or 4200-TRX) for each source-measure unit (SMU)
channel
• One high-voltage (HV) coaxial cable
• Use Model 2290-5-SHV or 2290-5-MHV if you are using the Model 2290-5 hi gh-voltage power supply
• Use Model 2290-10-SHVUC or 2290-10-SHV if you are using the Model 2290-10 hig h-voltage power
Connection summar y
Refer to the next figure to connect source-measure unit (SMU) channels of the Model 2635B or 2636B to the
Model 2290-PM-200.
Only 2-wire applications are supported.
supply
Do not convert triaxial cables to BNC cables. Using BNC cables may compromise overvoltage
protection and can result in damage to the SMU. You must use triaxial cables to connect to the
lower-voltage SMU.
071317001 / May 2015 7
Model 2290-PM-200 Protection Module User's Guide
Figure 6: Model 2290-PM-200 connections for Model 263xB using Model 2290-5 power supply
8 071317001 / May 2015
Model 2290-PM-200 Protection Module User's Guide
Figure 7: Model 2290-PM-200 connections for Model 263xB using Model 2290-10 power supply
071317001 / May 2015 9
Model 2290-PM-200 Protection Module User's Guide
Connections using a M odel 4200-SCS SMU
Required accessories to connect one Model 4200-SMU or 4210-SMU (without the Model 4200-PA Remote
PreAmp):
• One Model 4200-TRX cable
• One Model 4200-MTRX cables
• One high-voltage (HV) coaxial cable
• Use Model 2290-5-SHV or 2290-5-MHV if you are using the Model 2290-5 hi gh-voltage power supply
• Use Model 2290-10-SHVUC or 2290-10-SHV if you are using the Model 2290-10 hig h-voltage power
Connection summar y
Refer to the next figure to connect a Model 4200-SMU or 4210-SMU and the GNDU (ground unit) to the Model
2290-PM-200. Make sure to use the IN connectors of the Model 2290-PM-200 for connection to the Model
4200-SCS and use the OUT connector for connection to the device under test (DUT).
Only 2-wire applications are supported.
supply
When using multiple source-measure units (SMUs) that are installed in the Model 4200-SCS, each
SMU needs its own protection module. To achieve adequate voltage protection, connect the GNDU
Sense terminal to the LO, SL (LO and sense LO) connector of each protection module using triaxial
tees (such as the Keithley Model 237-TRX-T) and additional triaxial cables (Model 4200-TRX).
Connecting the Model 4200-SMU or 4210-SMU to the Model 2290-PM-200 connects the LO terminal
of the SMU to protective earth (safety ground).
10 071317001 / May 2015
Model 2290-PM-200 Protection Module User's Guide
Figure 8: Model 2290-PM-200 connections for Model 4200-SCS SMU using Model 2290-5 power supply
071317001 / May 2015 11
Model 2290-PM-200 Protection Module User's Guide
Figure 9: Model 2290-PM-200 connections for Model 4200-SCS SMU 7 using Model 2290-10 power
supply
12 071317001 / May 2015
Model 2290-PM-200 Protection Module User's Guide
Model 2290-PM-200 verification
A Keithley SourceMeter instrument capable of sourcing 100 V at 10 mA and 1 A at >10 V and a high voltage
power supply are required for verification purposes. Refer to the next figure when performing Verify module
continuity (on page 13), Verify the HI to LO leakage of the Model 2290-PM-200, a nd Verify module clamp
voltage (on page 14) steps in this section.
Figure 10: Model 2290-PM-200 Connections for verification procedures using a Model 2290-5
high-voltage power supply
Item Description
1 Model 2290-5 or 2290-10 high-v oltage power
supply
2 Model 2290-5 or 2290-10 SHV high-voltage cable
3 Model 2290-PM-200 high-voltage SMU protection
module
4 Triaxial cables
5 Source-measure unit (SMU)
Verify module continuity
1. Make sure that the high-voltage power supply is off.
2. Connect a high-voltage cable to the protection module OUT terminal (see the previous figure).
3. Short the HI and LO conductors on the high-voltage output cable.
4. Set the source-measure unit (SMU) to source 10 V and program the current limit to 1 A.
5. Turn on SMU output and verify that SMU is in current limit.
071317001 / May 2015 13
Model 2290-PM-200 Protection Module User's Guide
Verify the HI to LO leakage of the Model 2290-PM-200
1. Turn off the high-voltage power supply.
2. Remove the high-voltage cable from the protection module.
3. As shown in the previous figure, connect the protection module inputs to HI and LO terminals of a
source-measure unit (SMU) using two triaxial cables.
4. Program the SMU to source 100 volts or greater.
5. Turn on SMU output and allow to settle for 10 seconds minimum.
6. Observe current measurement on the SMU.
7. Current indication should be below 10 pA module specification.
Verify module clamp voltage
1. Make sure that the high-voltage power supply is off.
2. Turn the source-measure unit (SMU) output off and disconnect the triaxial cables (see the previous figure).
3. Connect a high-voltage cable from the protection module OUT terminal to the high-voltage power supply.
4. Engage the interlock on the back of the high-voltage power supply by connecting pin 1 to pin 2.
The high-voltage power supply is capable of sourcing hazardous live high voltages th at can
cause personal injury or death due to electric shock. This unit should be used only by
qualified personnel who recognize the dangers of high voltages.
Make certain that the source is turned off and that high voltage is completely discharged
before removing the high-voltage cable. High-voltage cables can store a charge if they are
disconnected from the supply while the high voltage is on. The charge on the cable can
cause injury or damage even after the cable is disconnected from the unit.
The high-voltage power supply is provided with an interlock circuit that must be engaged to
allow high-voltage output. The interlock helps facilitate s afe operation of the equipment in a
test system. Bypassing the interlock could expose the operator to hazardous voltages that
could result in personal injury o r death.
5. Turn on the high-voltage power supply.
6. Set the current limit of the high-voltage power supply to 1 mA.
7. Set the high-voltage power supply for 300 V DC.
8. Turn the high-voltage output on and verify that the high-voltage power supply is in current limit.
9. Voltage should never exceed 250 V DC.
14 071317001 / May 2015
Safety precaut ions
The following safety precautions should be observed before using this product and any associated instrumentation. Although
some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous
conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions
required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using
the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the
equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line
voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures
explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained
service personnel may perform installation and servi ce proc edures.
Keithley Instruments products are designed for use with electrical signals that are measurement, control, and data I/O
connections, with low transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high
transient overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient
overvoltages often associated with local AC mains connections. Certain Keithley measuring instruments may be connected to
mains. These instruments will be marked as category II or higher.
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than
30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any
unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators
are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If
the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited
sources. NEVER connect switching cards directly to AC mains. When connecting sourc es to sw itch ing card s, inst all protective
devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the
connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input
power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under
test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting
cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth)
ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the
voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or
accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be
impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifi cat ion s and operating
information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground)
connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
If a
The
user documentation in all cases where the symbol is mark ed on the instru ment .
The
contact with these voltages.
The
The
If this
properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read
the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains h az ards that coul d dama ge the instr ume nt. Such dam age may
invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national
safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be
purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be
purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the
applicability of a replacement component, call a Keithley Instruments office for information.
screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation.
symbol on an instrument means caution, risk of danger. The user must refer to the operating instructions located in the
symbol on an instrument means caution, risk of electric shock. Use standard safety precautions to avoid personal
symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
symbol indicates a connection terminal to the equipment frame.
symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with
no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled
according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the
factory for proper cleanin g/s er v icing.
Safety precaution revision as of January 2013.